Omega Products FPD1201B-IP Installation Manual

FPD1001B thru FPD1003B,
FPD1102B and FPD1103B,
FPD1201B thru FPD1203B
SERIES LOW FLOW
Positive Displacement Flowmeters
To the owner:
Red - Supply
Black - Ground
White - Output
X
V
cc
1 2 3
Please take a few minutes to read through the manual before installing and operating your meter. Always retain this manual for future reference. If you have any problems with the meter, refer to the Maintenance and Troubleshooting sections of the manual.
This manual contains connection and op­erating instructions for the OMEGA FPD Series meters. This includes the following models:
FPD1001B FPD1202B-IP FPD1201B FPD1003B FPD1201B-IP FPD1203B FPD1002B FPD1103B FPD1202B FPD1203B-IP FPD1102B
Part breakdowns for each model are locat­ed at the back of this manual. For models with displays and /or 4-20 mA output, an additional instruction manual is provided.
The flowmeter has incorporated the oval ro­tor principal into its design. This has proven to be a reliable and highly accurate method of measuring flow. Exceptional repeatability and high accuracy over a wide range of fluid viscosities and flowrates are features of the oval rotor design. With low pressure drop and high pressure rating, oval rotor flowmeters are suitable for both gravity and pump (in-line) applications.
NOTE: To prevent damage to the meter,
slowly fill the system with fluid (this will prevent damage caused by air purge). Failure to do this could damage the meter.
To reduce pressure build up, turn off the pump at the end of each day.
INSTALLATION
1. Use thread sealant on all pipe threads.
2. Ensure the meter is installed so that rotor shafts are always in a horizontal plane. Flow is bi-directional.
3. OMEGA recommends use of flexible connections.
4. Extreme care must be taken when installing the meter. Pipe strain or overtightening the meter connections can cause meter damage.
PULSER DETAILS
Your meter is equipped with both a Hall Effect sensor and a Reed Switch sensor. The equipment you are sending the pulse signal to will determine which sensor you use. The wires for the unused sensor will not be used.
OPERATION
Please read this information
!
carefully before use!
Before use, confirm the fluid to be used is compatible with the meter. Refer to Industry fluid compatibility charts or consult your local representative for advice.
To prevent damage from dirt or foreign matter it is recommended that a Y or bas­ket type 200 mesh strainer be installed as close as possible to the inlet side of the meter. Contact your local representative for advice.
Hall Effect Sensor Specifications:
4.5V to 24V (4.6 ~ 9 mA) operation needs only an unregulated supply.
Open collector 25 mA output NPN
compatible with digital logic.
Figure 1
Hall Effect Sensor Wiring Details
1
Reed Relay Specifications:
Alignment Marks
Body
Rotors
O-Ring
Inner Plate & PCB Assembly
O-Ring
Meter Cap
Screws
Inner Plate & PCB Assembly
Wire Clip
Seal
Wiring: Yellow wire / Green wire.
Two wire SPST N/O.
Switching voltage: 30 VDC; Maximum
current: 0.5 AMPS.
MAINTENANCE
CAUTION
Ensure the fluid is isolated from the meter to be repaired.
CAUTION
Ensure the fluid line is depressurized before commencing repairs.
CAUTION
Ensure electrical wiring is isolated and disconnected before commenc­ing repairs.
CAUTION
To prevent damage to the meter dur­ing re-commissioning, slowly fill the piping system with fluid before start­ing the pumping system.
CAUTION
Refer to the Sensor Wiring section for details.
Pulse Meter Disassembly
1. Loosen and remove four Phillips head or cap head screws (see Figure 2).
2. Remove the meter cap and O-ring.
3. Remove the inner plate and PCB as­sembly. Refer to the PCB Removal section for additional details.
4. Remove the O-ring.
5. Remove the rotors. Note the position of the rotor with the magnet or grub screws.
6. Clean and inspect all components, replace as necessary.
Pulse Meter Reassembly
1. Replace the rotors. Refer to Figure 3 or Figure 4 for correct orientation. Rotate the rotors by hand to ensure correct engagement.
Figure 2
2
Figure 3
Rotors with
Magnets
Alignment Marks
Magnets
NOT Visible
Magnets
Visible
GM001 Meters
Figure 4
GM002 & GM003 Meters
Models: FPD1001B & FPD1201B Models: FPD1002B, FPD1003B, FPD1102B,
2. Fit the O-ring into the O-ring groove in the meter body.
3. Combine the inner plate assembly (Items 4-7). Form the O-ring into the in­ner plate and replace the meter cap.
NOTE: Ensure all the alignment marks are
lined up with the mark on the body (see Figure 2).
4. Replace and tighten the four bolts to the required torque. Refer to Bolt Torque in the Specifications section for details.
5. Check meter function using low air pressure.
6. Restore the fluid and reconnect the wiring as detailed in the Installation and Pulser Details sections.
PCB Removal
CAUTION
The PCB contains components that are sensitive to static shock. Ensure all items and maintenance staff are grounded before working on the PCB.
1. Using a thin blade or similar tool, remove the wire clip and seal (see Figure 2).
2. While lifting the PCB with the blade inserted near the alignment marks, carefully push the cable into the inner plate. Continue to lift the back of the PCB until all components on the PCB clear the inner plate.
FPD1103B, FPD1202B & FPD1203B
3. Grasp the cable near the PCB and pull the cable through the opening in the inner plate until the cable ferrules are near the inner plate opening.
4. Push the cable ferrules into the PCB opening, one at a time, and remove the cable.
PCB Replacement
CAUTION
The PCB contains components that are sensitive to static shock. Ensure all items and maintenance staff are grounded before working on the PCB.
1. Feed the cable ferrules, one at a time, through the cable opening in the inner plate.
2. Gently pull the cable through the inner plate opening until the PCB almost touches the inner plate.
3. While pulling the cable, tilt the PCB into the recess in the inner plate. Ensure components underneath the PCB do not come into contact with the inner plate.
4. Pull the cable until the cable ties con­tact the inner plate. Gently push the PCB down into the housing.
5. Press the seal into the inner plate cable entry, followed by the wire clip. Ensure they both are seated correctly.
3
Loading...
+ 11 hidden pages