Please take a few minutes to read through
the manual before installing and operating
your meter. Always retain this manual for
future reference. If you have any problems
with the meter, refer to the Maintenance and
Troubleshooting sections of the manual.
This manual contains connection and operating instructions for the OMEGA FPD
Series meters. This includes the following
models:
Part breakdowns for each model are located at the back of this manual. For models
with displays and /or 4-20 mA output, an
additional instruction manual is provided.
The flowmeter has incorporated the oval rotor principal into its design. This has proven
to be a reliable and highly accurate method
of measuring flow. Exceptional repeatability
and high accuracy over a wide range of
fluid viscosities and flowrates are features
of the oval rotor design. With low pressure
drop and high pressure rating, oval rotor
flowmeters are suitable for both gravity and
pump (in-line) applications.
NOTE: To prevent damage to the meter,
slowly fill the system with fluid (this will
prevent damage caused by air purge).
Failure to do this could damage the
meter.
To reduce pressure build up, turn off the
pump at the end of each day.
INSTALLATION
1. Use thread sealant on all pipe threads.
2. Ensure the meter is installed so that
rotor shafts are always in a horizontal
plane. Flow is bi-directional.
3. OMEGA recommends use of flexible
connections.
4. Extreme care must be taken when
installing the meter. Pipe strain or
overtightening the meter connections
can cause meter damage.
PULSER DETAILS
Your meter is equipped with both a Hall
Effect sensor and a Reed Switch sensor.
The equipment you are sending the pulse
signal to will determine which sensor you
use. The wires for the unused sensor will
not be used.
OPERATION
Please read this information
!
▲
carefully before use!
Before use, confirm the fluid to be used is
compatible with the meter. Refer to Industry
fluid compatibility charts or consult your
local representative for advice.
To prevent damage from dirt or foreign
matter it is recommended that a Y or basket type 200 mesh strainer be installed as
close as possible to the inlet side of the
meter. Contact your local representative
for advice.
Hall Effect Sensor Specifications:
•
4.5V to 24V (4.6 ~ 9 mA) operation
needs only an unregulated supply.
•
Open collector 25 mA output NPN
compatible with digital logic.
Figure 1
Hall Effect Sensor Wiring Details
1
Page 4
Reed Relay Specifications:
Alignment Marks
Body
Rotors
O-Ring
Inner Plate &
PCB Assembly
O-Ring
Meter
Cap
Screws
Inner Plate &
PCB Assembly
Wire
Clip
Seal
•
Wiring: Yellow wire / Green wire.
•
Two wire SPST N/O.
•
Switching voltage: 30 VDC; Maximum
current: 0.5 AMPS.
MAINTENANCE
CAUTION
Ensure the fluid is isolated from the
meter to be repaired.
CAUTION
Ensure the fluid line is depressurized
before commencing repairs.
CAUTION
Ensure electrical wiring is isolated
and disconnected before commencing repairs.
CAUTION
To prevent damage to the meter during re-commissioning, slowly fill the
piping system with fluid before starting the pumping system.
CAUTION
Refer to the Sensor Wiring section for
details.
Pulse Meter Disassembly
1. Loosen and remove four Phillips head
or cap head screws (see Figure 2).
2. Remove the meter cap and O-ring.
3. Remove the inner plate and PCB assembly. Refer to the PCB Removal
section for additional details.
4. Remove the O-ring.
5. Remove the rotors. Note the position
of the rotor with the magnet or grub
screws.
6. Clean and inspect all components,
replace as necessary.
Pulse Meter Reassembly
1. Replace the rotors. Refer to Figure 3 or
Figure 4 for correct orientation. Rotate
the rotors by hand to ensure correct
engagement.
2. Fit the O-ring into the O-ring groove in
the meter body.
3. Combine the inner plate assembly
(Items 4-7). Form the O-ring into the inner plate and replace the meter cap.
NOTE: Ensure all the alignment marks are
lined up with the mark on the body
(see Figure 2).
4. Replace and tighten the four bolts
to the required torque. Refer to Bolt
Torque in the Specifications section
for details.
5. Check meter function using low air
pressure.
6. Restore the fluid and reconnect the
wiring as detailed in the Installation and
Pulser Details sections.
PCB Removal
CAUTION
The PCB contains components that
are sensitive to static shock. Ensure
all items and maintenance staff are
grounded before working on the PCB.
1. Using a thin blade or similar tool, remove
the wire clip and seal (see Figure 2).
2. While lifting the PCB with the blade
inserted near the alignment marks,
carefully push the cable into the inner
plate. Continue to lift the back of the
PCB until all components on the PCB
clear the inner plate.
FPD1103B, FPD1202B & FPD1203B
3. Grasp the cable near the PCB and pull
the cable through the opening in the
inner plate until the cable ferrules are
near the inner plate opening.
4. Push the cable ferrules into the PCB
opening, one at a time, and remove
the cable.
PCB Replacement
CAUTION
The PCB contains components that
are sensitive to static shock. Ensure
all items and maintenance staff are
grounded before working on the PCB.
1. Feed the cable ferrules, one at a time,
through the cable opening in the inner
plate.
2. Gently pull the cable through the inner
plate opening until the PCB almost
touches the inner plate.
3. While pulling the cable, tilt the PCB into
the recess in the inner plate. Ensure
components underneath the PCB do
not come into contact with the inner
plate.
4. Pull the cable until the cable ties contact the inner plate. Gently push the
PCB down into the housing.
5. Press the seal into the inner plate cable
entry, followed by the wire clip. Ensure
they both are seated correctly.
3
Page 6
TROUBLESHOOTING
Symptom Probable Cause Corrective Action
FLUID WILL NOT 1. Foreign matter blocking Dismantle meter, clean rotors. Strainer must
FLOW THROUGH rotors be fitted in-line.
THE METER
2. Line strainer blocked Clean strainer.
3. Damaged rotors Replace rotors. Strainer must be fitted in-line.
4. Meter connections over- Readjust connections.
tightened
REDUCED FLOW 1. Line strainer partially Clean strainer.
THROUGH THE blocked
METER
2. Fluid is too viscous Maximum viscosity 1000 centipoise.
METER READING 1. Fluid flowrate is too low See specifications for minimum and maximum
INACCURATE or too high flowrates.
2. Air in fluid Bleed air from system.
3. Excess wear caused by Check meter body and rotors.
incorrect installation
METER NOT 1. Faulty Hall Effect sensor Replace meter cap.
GIVING A PULSE or Reed Switch
SIGNAL
2. Faulty magnet Replace rotors.
3. Rotors installed in wrong Refer to correct rotor positioning and
position assembly instructions.
5 1 BS029VS O-Ring (Viton)
5 1 BS029TES O-Ring (Perfluoroelastomer)
1 u N359S O-Ring (NBR)
1 u N359S Wire Clip
1 u MS271AS PCB Assembly
1 u MS428S Meter Cap (FPD1001B)
1 u MS248-1S Meter Cap (FPD1201B & FPD1201B-IP)
6
Page 9
SPECIFICATIONS
Stainless Steel Models Aluminum Models
Port Size 1/8 in. 1/8 in.
Accuracy of Reading ± 1% ± 1%
Maximum Viscosity 1000 Centipoise 1000 Centipoise
Flow Ranges (LPH or GPH)
Above 5 centipoise 0.5 to 50 / 0.132 to 13.2 0.5 to 50 / 0.132 to 13.2
Below 5 centipoise 2 to 50 / 0.528 to 13.2 2 to 50 / 0.528 to 13.2
K-Factor (Pulses per Litre/Gallon)1552 / 5875.32 1552 / 5874.32
Max. Operating Pressure SS = 1000 kPa / 150 PSI / 10 Bar 500 kPa / 75 PSI / 5 Bar
Min. Operating Temperature -10°C / +14°F -10°C / +14°F
Max. Operating Temperature 120°C / 248°F 60°C / 176°F
Pulse Type Hall Effect Sensor/Reed Switch Hall Effect Sensor/Reed Switch
Cable Length 1 meter / 3.28 feet 1 meter / 3.28 feet
Weight 602 g / 21.23 oz. 308 g / 10.86 oz.
Dimensions (W x D x H)60mm x 50mm x 50mm 60mm x 50mm x 50mm
Face to Face 67mm / 2.63 in. 60mm / 2.36 in.
7
Page 10
DISPLAY PARTS LISTING
Models: FPD1002B
FPD1202B
FPD1102B
FPD1202B-IP
1
2
5
4
3
u = Recommended Spare Parts (See Figure 2).
Bold Text = Indicates Stainless Steel model parts.
Item Rec. Part or Set
No. Qty. Parts (Order from this column only)Part Description
1 1 MS1S-2S Meter Body & Shafts NPT Stainless Steel
1 1 MS1AL-2S Meter Body & Shafts NPT Aluminum
1 1 MS1R-2S Meter Body & Shafts NPT PPS
1 1 MS1R-2C Meter Body & Hast Shafts NPT PPS
5 1 BS029VS O-Ring (Viton)
5 1 BS029TES O-Ring (Perfluoroelastomer)
1 u N359S Wire Clip
1 u MS271AS PCB Assembly
1 u N359S O-Ring (NBR)
1 u MS428S Meter Cap
1 u MS428-1S Meter Cap
8
Page 11
SPECIFICATIONS
Stainless Steel & PPS Models Aluminum Models
Port Size 1/4 in. 1/4 in.
Accuracy of Reading ± 1% ± 1%
Maximum Viscosity 1000 Centipoise 1000 Centipoise
Flow Ranges (LPH or GPH)
Above 5 centipoise 2 to 100 / 0.53 to 26.4 2 to 100 / 0.53 to 26.4
Below 5 centipoise 5 to 100 / 1.32 to 26.4 5 to 100 / 1.32 to 26.4
K-Factor (Pulses per Litre/Gallon)1000 / 3785 1000 / 3785
Max. Operating Pressure SS = 1000 kPa / 150 PSI / 10 Bar 500 kPa/ 75 PSI / 5 Bar
I.P. = 5515 kPa /800 PSI / 55 Bar
PPS = 500 kPa / 75 PSI / 5 Bar
Min. Operating Temperature -10°C / +14°F -10°C / +14°F
Max. Operating Temperature SS = 120°C / 248°F 80°C / 176°F
PPS = 80°C / 176°F
I.P. = 120°C / 248°F
Pulse Type Hall Effect Sensor/Reed Switch Hall Effect Sensor/Reed Switch
Cable Length 1 meter / 3.28 feet 1 meter / 3.28 feet
Weight SS = 240 g / 8.5 oz. 310 g / 11 oz.
PPS = 600 g / 21.2 oz.
Dimensions (W x D x H)60mm x 50mm x 50mm 60mm x 50mm x 50mm
Face to Face SS = 67mm / 2.63 in. 60mm / 2.36 in.
PPS = 64mm / 2.52 in.
9
Page 12
DISPLAY PARTS LISTING
1
4
5
2
Models: FPD1003B
FPD1203B
FPD1103B
FPD1203B-IP
u = Recommended Spare Parts (See Figure 2).
Bold Text = Indicates Stainless Steel model parts.
Item Rec. Part or Set
No. Qty. Parts (Order from this column only)Part Description
1 1 MS2S-2S Meter Body & Shafts NPT Stainless Steel
1 1 MS2AL-2S Meter Body & Shafts NPT Aluminum
1 1 MS2R-2S Meter Body & Shafts NPT PPS
1 1 MS2R-2C Meter Body & Hast Shafts NPT PPS
2 1 MS427S Inner Plate
4 1 MS814HS Rotor Assembly Stainless Steel High Viscosity
5 1 BS029VS O-Ring (Viton)
5 1 BS029TES O-Ring (Perfluoroelastomer)
1 u N359S O-Ring (NBR)
1 u N359S Wire Clip
1 u MS271AS PCB Assembly
1 u MS428S Meter Cap
1 u MS428-1S Meter Cap
3
10
Page 13
SPECIFICATIONS
Stainless Steel & PPS Models Aluminum Models
Port Size 1/4 in. 1/4 in.
Accuracy of Reading ± 1% ± 1%
Max. Viscosity 1000 Centipoise 1000 Centipoise
High Viscosity Option Yes Yes
Flow Ranges (LPH or GPH)
Above 5 centipoise 15 to 500 / 4 to 32 15 to 500 / 4 t o 32
Below 5 centipoise 25 to 500 / 6 to 132 25 to 500 / 6 to 132
K-Factor (Pulses per Litre/Gallon)400 / 1514 400 / 1514
Max. Operating Pressure SS = 1000 kPa/150 PSI/10 Bar 500 kPa/75 PSI/5 Bar
PPS = 500 kPa/75 PSI/5 Bar
I.P. = 5515 kPa/800 PSI/55 Bar
Min. Operating Temperature -10°C / +14°F -10°C / +14°F
Max. Operating Temperature SS = 120°C / 248°F 80°C / 176°F
PPS = 80°C / 176°F
I.P. = 120°C / 248°F
Pulse Type Hall Effect Sensor/Reed Switch Hall Effect Sensor/Reed Switch
Cable Length 1 meter / 3.28 feet 1 meter / 3.28 feet
Weight 240 g / 8.5 oz. 320 g / 12 oz.
600 g / 21.2 oz. (PPS)
Dimensions (W x D x H)60mm x 50mm x 50mm 60mm x 50mm x 50mm
Face to Face SS = 67mm / 2.63 in. 60mm / 2.36 in.
PPS = 64mm / 2.52 in.
11
Page 14
Page 15
Page 16
M-4 686 / 0808
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.