GENERAL INFORMATION
Use the OMEGA Standard Display in indoor
or outdoor applications where occasional
exposure to moisture is common. This
manual covers two models:
• Standard Display – Local (mounted to
the meter)
• Standard Display – Remote (where
remote monitoring is required)
Product differences in this manual are
identified by either, Local or Remote as
necessary.
Both models are battery powered; no external power source is required.
SAFETY INSTRUCTIONS
• When measuring flammable liquids,
observe precautions against fire or
explosion.
• When working in hazardous environments, always exercise appropriate
safety precautions.
• Be sure O-rings and seals are kept in
good repair.
INSTALLATION
CAUTION
Installation should be performed only by
qualified personnel, and in accordance
with local governing regulations.
Remote Installation
Choose a mounting location suitable for
the Standard Display. The ideal mounting
location is where the:
• FPD Series meter is as close as possible.
• mounting surface has minimal vibration.
• ambient temperature is 0°F to 140°F
(-18°C to 60°C) when using remote
display.
• cable lengths are minimal.
Avoid mounting locations where the Standard Display is:
• subject to constant exposure to water
or other liquids (occasional low-pressure splashing will not harm unit).
• subject to > 5g shock loading.
• facing the sun directly for long periods
of time.
Mount the OMEGA Standard Display Remote using standard U-bolts. Mounting
options include:
• Wall
• Pipe
This product comes with 20 feet of shielded
cable. OMEGA offers an optional cable kit
with 100 feet of cable (see spare parts list).
OPERATION
Computer Display
All operations are reflected in the LCD
readout. The large center digits indicate
amounts, where smaller words or “icons”
located above and below indicate specific
information regarding totals, flow, calibration
and units of measure.
Computer is on continuously and always
ready to perform. The computer is powered
by a field replaceable battery. When display
becomes dim, faded or the low battery
message appears (see below), the battery
needs to be replaced. Reference the
Maintenance Section for details.
1
Batch and Cumulative Totals
The computer maintains two totals. The
Cumulative Total provides continuous measurement and cannot be manually reset. The
Batch Total can be reset to measure flow
during a single use. The Cumulative Total
is labeled TOTAL 1, Batch Total is labeled
TOTAL 2 BATCH.
When the Cumulative Total reaches a
display reading of 999,999 the computer
will highlight an X10 icon. This indicates to
the operator that a zero must be added to
the 6 digits shown. When the next rollover
occurs, the computer will highlight an X100
icon. This indicates to the operator that two
zeros must be added to the 6 digits shown.
Press the DISPLAY button briefly to switch
between the TOTAL 1, TOTAL 2 BATCH
and FLOWRATE. Press DISPLAY briefly
to display the TOTAL 2 BATCH. Hold the
DISPLAY button for 3 seconds to reset the
Batch Total to zero.
When fluid is flowing through the meter, a
small propeller icon is highlighted.
NOTE: Totalization counts total units
without differentiating between gallons,
liters or field calibrated units.
Flowrate Feature
To use this feature, press and release
DISPLAY until FLOWRATE icon appears.
The factory set time base will be highlighted
to the right of FLOWRATE (M = minutes,
H = hours, D = days). When FLOWRATE
is invoked, the display will be indicating
rate of flow.
Factory and Field Calibration
All calibration information is visible to the
user as icons on the top line of the display,
above the numeric digits.
All units are configured with a “factory”
calibration. Both gallons and liters are
available (“GL” or “LT” will be displayed).
While holding the CALIBRATE button, briefly
press DISPLAY to toggle between gallons
and liters. This factory calibration (indicated
with FAC) is permanently programmed into
the computer and is not user adjustable.
NOTE: Your computer may have other units
of measure programmed into it. If so,
holding the CALIBRATE button and momentarily pressing the DISPLAY button
will toggle through all factory set units.
Other possible units are: IGL (imperial
gallon), QT (quart), CF (cubic feet), CM
(cubic meter), BL (42 gal. barrel), CC
(cubic centimeter) or OZ (ounce).
Switching between different units will not
corrupt the Total’s contents. For example,
in GL mode, the computer totalizes 10.00
gallons, if the user switches to LT mode,
the display will read 37.85 liters (the same
volume, different unit).
The “field” calibration may be set by the
user, and can be changed or modified at
any time using the calibration procedure
described in the Calibration Section. Totals
or flowrate derived from the field calibration
are invoked when the FAC icon is no longer
visible on the top line of the display.
2
CONFIGURATION
Configuration determines what information is displayed on the screen. The configuration string is a 6-digit number that
programs the computer electronics to a
specific configuration. Using the information below, the end user can change the
information displayed on the screen by
changing the configuration string.
Local Mount Display
The factory calibration and display configuration features are preprogrammed and
further user programming is not required.
When replacing electronic displays, they
must be configured and calibration information entered before use. If desired, the
calibration or features can be changed in
the field using the procedures described
below.
Remote Mount Display
The default calibration and display configuration features may not be appropriate for
the user installation.
ATTENTION
ALL remote mount and replacement
displays must be configured AND calibrated before use!
Factory Default Configuration
The configuration strings below are the
default configurations for the local mounted
display and can be used for the final configuration for replacement displays or remote
mounted displays purchased separately.
NOTE: For replacement and remote elec-
tronics, this configuration does NOT
determine the K-factor for your specific
meter. That information must still be
entered using the calibration procedure. For replacement and remote
configurations, please see Calibration
Section for more information.
FACTORY CONFIGURATION STRINGS
MODEL CONFIGURATION STRINGS
FPD1001B
FPD1002B
FPD1003B
FPD1102B
FPD1103B
FPD1201B
FPD1202B
FPD1203B
FPD1004
FPD1204
FPD1105
FPD1005
FPD1205
FPD1006
FPD1007
B 2 2 0 1 3
B 1 2 0 1 3
3