Omega Products FL-1902 Installation Manual

omega.com
®
®
FL-1900 Series Rotameters
Operator’s Manual:
INSTRUCTION
M0360/0201
®
6
4
6
0.5
SLMP AIR 70° 7 14.7 PSIA
SECTION 1 INTRODUCTION
1.1 GENERAL DESCRIPTION
OMEGA®Series FL-1900 Rotameters provide an
economical and efficient means of flow rate indication when accuracy is not critical. Typical applications include purging, seal oil systems, bearing lubrication, and cooling water flow indication.
Engineering design makes them especially suitable for use with corrosive liquids and gases, and installation in corrosive atmospheres.
The Series FL-1900 Rotameters are furnished with a control valve, but are available without the valve. Indicator body and supported frame are constructed from stainless steel. Sealing parts are manufactured from vinylidene fluoride resin (Kynar). O-ring seals are Viton-A. The meter is totally enclosed within a cover of high impact polycarbonate (Lexan).
1.2 SPECIFICATION
The following specifications apply to all models.
CAUTION
Do not use this meter in excess of the specified values
PRESSURE RATINGS: 200 psig maximum TEMPERATURE RATING: 250
°
F maximum REPEATABILITY: ± 0.5 % of full scale ACCURACY: Meter specified to have an accuracy
of ± 10 % of maximum scale from 100 % to 10% of scale reading. Conforms to ISA R.P. 16.1,
Specification 10-S-10 DIMENSION: See Figure 1.1 METERING TUBE: Borosilicate glass END FITTINGS: 316SS WINDOW SHIELD: High Impact Plastic
FLOAT STOPS: Teflon TUBE PACKING: Viton O - RINGS: Viton SCALE: 65mm direct reading for air or water
SECTION 2 INSTALLATION
2.1 RECEIPT OF EQUIPMENT
When the equipment is received, the outside packing cases should be checked for any damage incurred
during shipment. If the packing case is damaged, the local carrier should be notified at once regarding his liability.
Remove the envelope containing the shipping list. Carefully remove the equipment from the packing case. Make sure spare or replacement parts are not discarded with the packing material. Inspect for damaged or missing parts.
2.2 RETURN SHIPMENT
Do not return any merchandise without an Authorized Return (AR) number. Call OMEGACustomer Service
Department at (203) 359-1660.
CAUTION
It is recommended that this publication be read in its entirety before performing any operation. Failure to understand and follow these instructions could result in serious personal injury and/or damage to the equipment
WARNING
Glass metering tubes are designed for operation up to the maximum operating pressures and temperatures as specified herein. Due to the inherent brittle characteristic of glass and conditions beyond our control, tube breakage could result below specified operating conditions. Possible glass tube breakage represents a potential hazard to operating personnel; therefore, operator protection should be supplied where operating pressures may exceed 50 psig.. A safety shield constructed of 1/2 inch acrylic plastic may be used or the glass tube meter maybe replaced with an all metal (armored) meter.
Figure 1.1 Dimension Diagram
2.3 FILTER INSTALLATION
It is recommended that a filter with a rating of two (2) microns be installed on the inlet side of the meter. This filter will reduce the amount of foreign matter that could impair operation of meter.
2.4 INSTALLATION OF METER
a. Check the control valve is fully closed. b. Check the valve body is fully seated in meter
frame. Finger tighten if necessary.
c. Slowly tip meter back and forth and check that
float has freedom to move. d. Remove dust covers from the inlet and outlet connections. e. Install inlet and outlet line connections. Inlet
connection is at the zero end of the meter
.
2.5 CONTROL VALVE REPOSITIONING PROCEDURE
The user may desire to change the control valve from the inlet side of the meter to the outlet side. If meter is installed in the line, make sure that flow is stopped to the meter. Failure to do this may result in a safety hazard. a. Disconnect inlet and outlet lines to meter. b. Remove metering tube and seal spindle. Refer to
paragraph 4.3 c. Remove check ball from inside seal spindle.
Retain check ball for future use. d. Install seal spindle in meter body. Refer to
paragraph 4.4 e. Turn meter 180
°
from original position so control
valve is now on outlet and seal spindle is on inlet. f. Install metering tube. Refer to paragraph 4.4 g. Connect inlet and outlet lines to meter.
SECTION 3 OPERATION
3.1 GENERAL
After the meter has been installed it is ready for operation. During operation make sure that meter temperature and pressure specifications are not exceeded, as stated in paragraph 1.2
When the valve stem knob is turned counterclockwise to the stop, the valve is fully open. Flow indication is measured using the center of the float as the reference point. A check valve is incorporated within the seal spindle. The check valve prevents reverse flow through the meter when upstream flow is stopped. When the control valve is installed on the outlet of the meter, the check valve is not included.
SECTION 4 MAINTENANCE
4.1 GENERAL
No periodic maintenance, oiling, or cleaning is required for the meter or meter components when it is used in normal applications. It may be necessary to remove the metering tube and components for cleaning when various liquids have stained or left a deposit.
It is recommended that the meter and meter components be ultrasonically cleaned if possible. If ultrasonic cleaning is not available, use a suitable solvent. Tube and float may be cleaned with a pipe cleaner or similar accessory.
4.2 SERVICE INFORMATION
Should this equipment require repair or adjustment, contact OMEGA ENGINEERING Customer Service Department at (203) 359-1660.
It is important that servicing be performed only by trained and qualified service personnel. If this equipment is not properly serviced, serious personal injury and/or damage to the equipment could result.
4.3 DISASSEMBLY PROCEDURE (SEE FIGURE 5.1)
a. Remove shield (8) by squeezing sides at top and
bottom, and pulling forward.
b. Hold metering tube (1), and rotate seal spindle (4)
to the right until the stop is reached. Remove metering tube.
c. Using tweezers, remove outlet float stop (2).
Remove float from metering tube. Be careful not to drop float. Remove inlet float stop (2). If float is stuck in metering tube, a rod (wooden dowel or pipe cleaner preferred) may be inserted through the inlet end of the tube, and the float pushed through. Care must be taken not to score or scratch the tube.
d. Depress seal spindle retainer tab. Rotate seal
spindle (4) to the left until it is disengaged from the meter. Be careful not to drop the check ball (6) from inside the seal spindle. Remove check ball.
e. Remove control valve (9) from meter body.
4.4 ASSEMBLY PROCEDURE (SEE FIGURE 5.1)
After the meter and the meter components have been cleaned or checked, the following assembly procedures should be followed. A separate assembly procedure for the control valve is given in paragraph 4.6 a. Insert control valve body into meter frame and
finger tighten to secure.
b. Insert check ball (6) into seal spindle (4). Check for
freedom of movement.
c. Depress seal spindle retainer tab. Insert seal
spindle (4) into end fitting, and turn to the right until fully seated. Release seal spindle retainer.
d. Insert inlet float stop (2) through outlet end of
metering tube. Using a rod (wooden dowel or pipe cleaner preferred) push inlet float stop through metering tube until it is approximately
1
/16 to 1/32 inch from end of metering tube.
(See Figure 4.2). e. Insert float into metering tube. f. Insert outlet float stop (2). Push float stop into
metering tube, until it is seated approximately
1
/16 to 1/32 inch from end of tubes. (See Figure 4.2).
Figure 4.2 Float Stop Installation Diagram
Loading...
+ 2 hidden pages