It is recommended that this publication be read in its entirety before performing any operation. Failure to understand and follow these
irWUCtiOnS
could result in serious personal injury and/or damage to the equipment.
SECTION 1 INTRODUCTION
1.1
GENERAL DESCRIPTIONThe OMEGA@ Series FL-1600 Rotameters are variable
area flowmeters designed for general in-line and by-passmetering applications where operating conditions arewithin the limitations of glass metering tubes.
A float is free to travel up and down within a taperedborosilicate glass tube. As the float rises a greater flowarea exists: as it descends, flow area decreases. Fluidflowing through the flowmeter (from bottom to top)causes the float to rise to a point of dynamic balance,which is a true indication of the flow.
1.2
SPECIFICATIONS
CAUTiOq
1
Do not operate this instrument in excess of specifica-tions listed below.
PRESSURE RATINGS:
FL-1604FL-1605100
FL-1606
L
FL-1607
NOTES:
1.
Pressure ratings are based on
2.
Above pressure
3. Maximum safe working pressures for glass tubes above 200 degrees F can becalculated by Caution: Do not use meters in excess of these pressures.
Water hammer and surges can be damaging to anyflowmeter and must
always be avoided.
Water hammer occurs when a liquid flow is suddenly
stopped as with quick closing and solenoid operatedvalves. Surges occur when flow is suddenlv begun, aswhen a pump is turned on at full power or a valve is quick-ly opened.
Liquid surges are particularly damaging to flowmeters if the
surqes.
pipe is originally empty. To avoid damaging
fluidlines should remain full (if possible) and pumps should bebrought up to power slowly and valves opened slowly. Inaddition, to avoid both water hammer and surges, a surgechamber should be installed.
I
2.1
RECEIPT OF EQUIPMENTWhen the equipment is received, the outside packingcase should be checked for any damage incurred duringshipment. If the packing case is damaged, the local car-rier should be notified at once regarding his liability.
Remove the envelope containing the shipping list.Carefully remove the equipment from the packing caseand inspect for damaged or missing parts. Be sure thefloat is not accidentally discarded with the packing, assome larger meters are shipped with the iioat
remcved
from the tube to prevent breakage
2.2
RETURN SHIPMENT.Do not return any merchandise without an Authorized
Return (AR) number. Call OMEGA Customer ServiceDepartment at (203) 359-1660.
2.3
I
INSTALLATION OF FLOATSome flowmeters are shipped with the float installed inthe metering tube; others are shipped with the float
_.
I
.:
packed separately to prevent possible damage duringshipment and handling. In either case, remove the wrap-
pings and/or other protective material. Take great care toavoid damaging the float. The accuracy of the meterdepends upon the float remaining dimensionally andgravimetrically stable.
!.4
It is advisable to insert the float before the flowmeter is
installed in the line Where the vertical inlet and outlet areused for piping connections this is absolutely necessary.
Use reasonable care in inserting the float. Don’t
justdropit in the glass metering tube. A sheet of paper slightlycoiled, forms a convenient chute so the float can be slidinto the metering tube avoiding high impact.
To install free or ribbed tube floats, remove the bottomcleanout
plug (or flanged cover) and take out the bottomfloat stop. Place the float in the tube with a paper chute,as described above, and replace the float stop andcleanout
plug.
With rod guided floats, take great care not to bend theguide rod. Should the guide rod become bent, or its
polished surface become marred by burrs, scratches oraccumulation of foreign matter, it is advisable to replaceit with a new one. Never attempt to straighten a bentguide rod.
Rod guided floats are inserted at OMEGA and are shippedwith a wooden dowel protector preventing undue move-ment of the float while the rotameter is in transit. Be sureto remove the wooden dowel before installing therotameter.
To install rod-guided floats, remove both top and bottomcleanout
plugs. Remove the top nuts on the guide rod andremove the guide rod and guide rod cartridge through thebottom opening of the meter. Place the float on the guiderod and put it back in the tube. Guide rod nuts should betightened until the guide rod is tensioned. If the guide isnot sufficiently tightened, there may be a tendency of thefloat to whip under flow conditions. Replace the
cleanout
plugs (or flanged cover).
LOCATION OF CONTROL VALVE
Location of the control valve depends on many variables.On most applications, the flowmeter will operate properlywith the control valve on either the upstream ordownstream side.
If the inlet and outlet valves adjacent to the flowmeterare to be closed for any reason, the flowmeter must becompletely drained. Failure to do so may result in ther-mal expansion of the liquid which can cause rupture ofthe meter and possible personal injury.
HORIZONTAL
A
-
Inlet Valve
-
Outlet Valve
B
FIGURE 2.1 BY-PASS PIPING
SECTION 3 OPERATION
3.1
3.2
PRE-OPERATIONAL CHECK
Prior to initial start-up and each time the flowmeter is re-assembled, the zero alignment should be checked. Thezero line on the tube should coincide with the zero lineon the indicating scale. If these marks line up, assured that the tube is correctly seated andflowmeter is ready for accurate flow indication.
OPERATING PROCEDURE
WARNING
LINE
VERTICAL LINE
-
By-Pm
C
-
Control Valve
D
Valve
be
cm
it
the
2.5
INSTALLATION OF FLOWMETER
The flowmeter should always be mounted within
.
true vertical with the outlet connection (downstream
connection) at the top of the meter, and the inlet connec-
tion (upstream connection) at the bottom;
highest
scale graduation and largest part of the glass tubeat the top. Use a plumb-bob or other device to assureperfect vertical positioning. Should the rotameter not bemounted in a vertical position, both accuracy and sen-sitivity may be affected.
!.6
BY-PASS PIPINGIt is strongly recommended to install a by-pass piping
rangement around the meter. By-pass piping permits themeter to be isolated from the flow line for servicing or
cleaning.
Where connections are interchangeable (for vertical horizontal connections), it is recommended that the
horizontal connections be used if at all practical in theoverall piping arrangement. Horizontal connections per-mit the plugged vertical openings to serve as convenientcleanoutrotated in
ports. The design permits the end-fitting to be
90°
increments, allowing inlet and outlet con-nections to be made at right side, left side, at the front orat the back of the rotameter. Refer to Fiaure 2.1
that
6O
of
is, with
ar-
or
During operation of this instrument. do not exceedtemperature and pressure ratings as listed inSpecifications.
It is well to by-pass the original flow surge when themeter is first put in the line Refer to Figure 2-l. First, open
(C)
the by-pass valve
fully and the control valve
slightly. Then, open the meter inlet and outlet valves
(D)
(A
and B respectively). After the inlet line has been purged
(C)
of air, close the by-pass valve
slowly. This method
brings the flowmeter on-stream smoothly; allows the
float to start indicating the flow without bouncing orshooting up to the very top of the metering tube.
For best results, the meter inlet and outlet valves (A andB) should remain fully opened,
(D)
valve
to be used for complete flow
permitt+
cc‘itror.
the
c-ntrol
T;IIS
ar-rangement is consistent with good control practice andprovides steadiest float operation.
When it is necessary to clean or repair the meter, open
the by-pass valve outlet valves
(C)
first. Then, close the meter inlet and
(A
and B).
If the by-pass arrangement is not used, the valves shouldbe opened with extreme caution to prevent damage to
the float and metering tube.
SECTION 4 MAINTENANCE
4.1CLEANING When the recommended horizontal connections are us-
ed, the flowmeter may be cleaned in this manner:
1.
Remove top and bottom
2.
Remove the float and float stops. Be careful not to
damage the float.
Swab out the metering tube and end fittings with a
3.
suitable solvent.Replace float, float stops. and
4.
flanges).
4.2
DISASSEMBLY
1.
Remove front and back windows and windowframes.
Completely unscrew gland bolts from top and bot-
2.
tom end fittings.Remove cap screws from the side plates of the top
3.
end fittings.
4.
Loosen the cap screws in the bottom end fitting sothat the side plates may be spread apart to allow thetop and bottom end fittings to be removed.
The metering tube may now be removed.
5.
4.3
REASSEMBLY
1.
Replace the tube, being careful to see that the outlet
(wide) end of the tube fits flush against the tube seat
gasket in the outlet (top) end fitting.
-
cleanout
plugs (or flanges).
cleanout
plugs (or
Replace the windows and window frames. The seal-
5.ing gaskets on the windows should be placed facedown against the meter body. Tighten the shield
bolts on the window frame only until they are“snug,” then tighten one-quarter additional turn.
After the flowmeter has been re-assembled, it is im-
6.portant that it be hydrostatically tested at the liquidpressure rating as shown in the Hydrostatic TestPressure Table.
Hydr&&ic
testing should be performed only bytrained and qualified personnel or seriousdamage could result.
HYDROSTATIC
TABLE 4-l.
TEST PRESSURES
Max. Working
Model
FLil601,1602FL-1603
FL-1604,FL-1
1605
606
Pressure (PSI)
200°F
at
300
175100
FL1607100
75
’
Hydrostatic Test
Pressure
(PSI)
450260
150
110
150
NOTEBecause both Teflon hat gaskets are purposely tightfitting, they can be more easily assembled from theinlet (smaller) end of the tube The smaller (inlet end)gasket is often difficult to place over the glass tube.A suggested method is to place the hat gasket in thegland ring on a horizontal surface and carefully forcethe end of the tube through the gasket and ring. Thiswill stretch the gasket enough that it can be removedand reassembled on the tube in the proper order.
Im-mersing the gaskets in warm water will further in-crease their elasticity and help prevent damaging
them, particularly in cold weather. If the outletgasket on the larger tubes seems unable to be mov-ed into position when being driven by the packingring, a small amount of silicone lubricant may beused.
2. Since the dowel-pin construction assures perfectalignment of the frame assembly, merely tighten thecap screws to attach the side plates to the endfittings.
3. Insert the packing gland bolts. Tighten the glandfollower on the “seating end fitting” before tighten-
ing the gland follower on the other end fitting. The“seating end fitting” is always the outlet (top) endfifing. If this procedure of tightening the seating endfirst is not followed, the tube may be pulled awayfrom the seat gasket and the zero line on the tubemay not show above the inlet gland follower. In
tightening the packing bolts, tighten them
diagonally.Check the zero line on the metering tube and on the
4.scale for exact alignment.
Type
(rib-guided)
Spherical
Aumb-Bob
Figure 3-1. Float Types
SECTION 5 PARTS LIST
When ordering replacement parts be sure to specify thefollowing: SERIAL NUMBER, Complete model number,part description and quantity. Refer to Figure 5-2.
Glass metering tubesare designed for operation up to
the maximum operating pressures and temperatures as
specified herein. Due to the inherent brittle characteristicof glass and conditions beyond our control, tubebreakage could result below specified operating condi-
tions. Possible glass tube breakage represents a poten-
tial hazard to operating personnel; therefore, operatorprotection should be supplied where operating pressuresmay exceed 50 psig. A safety shield construction ofinch acrylic plastic may be used or the glass tube meter
may be replaced with an all metal (armored) meter.
RS
Type GV
(rod-gurcfed)
Spool
WARNING
GS
Type
(rod-guided)
LJ
Type
%
9
1011
17
._
1
13
1
14
17
IfI
I
Window
I1
19
20
I
1
Scale Blank
21
-
Follower
_..-
-
Ring Rina
___
Seat Gasket
Phi&d
1..
...”
.
....
Outlet
___.
-
Inlet
-
Outlet
lasket
3lass
(
Packing Side PlateScale Side Plate
Glrnd
_._..
GlandGland
Window FrameMillnrl
-----~-
FrameGasket
Meter Gasket
SteelSteel
steot
_...“.
,
-
,
1
11
Steel
Steel
1
AsbestosAsbestosStn. Stl.Stn. Stl.
Safety FeltCork
1I
I
1I Aluminum
Glass
Figure 5-2. Parts Drawing
WARNING
FLOWMETER OPERATION
If the inlet and outlet valves adjacent to the flowmeter are to be closed
”
for any reason, the flowmeter must be completely drained. Failure to doso may result in thermal expansion of the liquid which can cause ruptureof the meter and possible personal injury.
HORIZONTAL LINE
No. 6 Plain
2829
t 30 I Stuffino Box Aina
31
3233
35
3436Cartridge Gasket 373839
Scale Holder Screws
Hd. Self-Tapping
Scale Holder Screws
Hd. Self-Taopina
StuffingGuide RodGuide Rod
I
Cartridae
1
Cartridge
Cartridge Gasket.Teflon Hat GasketTeflon Hat Gasket
80x
-
-
Outlet
Inlet
..
Ring
VERTICAL LINE
_
-
inlet
-
Outlet
Inlet
-
Outlet
-
Inlet
-
Outlet
-
OvalRd.
2Stn. Stl.27
1
I
112
I
1
1
11111
Stn. Stl.Stn. Stl.
I
I
I
I
COPYRIGHT 1987 OMEGA ENGINEERING, INC. PRINTED IN U.S.A.
-
Inlet Valve
A
-
Outlet Valve
B
-
Bypass Valve
C
-
Control Valve
D
TYPICAL INSTALLATION
M379/097
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.