Omega Products DWT1305DA-2000 Installation Manual

Page 1
®
DWT1305D Series
Deadweight Tester
DWT1327D
Portable Pressure Test Set
Operator’s Manual
Deadweight Tester
®
®
®
®
®
®
®
®
®
®
®
®
®
®
An OMEGA Technologies Company
®
Page 2
Servicing North America:
USA: One Omega Drive, Box 4047
ISO 9001 Certified Stamford CT 06907-0047
Tel: (203) 359-1660 FAX: (203) 359-7700
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378 / 1-800-622-BEST
®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
®
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico: Tel: (95) TC-OMEGA
SM
FAX: (95) 203-359-7807
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: (31) 20 6418405 FAX: (31) 20 6434643 Toll Free in Benelux: 06 0993344
Czech Republic: Ostravska 767, 73301 Karviná
Tel: 42 (69) 6311899 FAX: 42 (69) 6311114
France: 9, rue Denis Papin, 78190 Trappes
Tel:: 33 (1) 30.62.14.00 FAX: 33 (1) 30.69.91.20 Toll Free in France: 05-4-OMEGA
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 Toll Free in Germany: 0130-112166
United Kingdom: 25 Swannington Road, Broughton Astley, Leicestershire,
ISO 9002 Certified LE9 6TU, England
Te l: 44 (1455) 285520 FAX: 44 (1455) 283912 Toll Free in the United Kingdom: 0800-488-488
An OMEGA Technologies Company
OMEGAnetSMOn-Line Service Internet E-Mail
http://www.omega.com info@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
CJS0296FSBA
®
Page 3
1.0 PURPOSE AND SCOPE OF MANUAL
This manual is provided to guide users of Model DWT1305D and DWT1327D pressure testing devices in: (1) installing the equipment (2) Routine operations
The instructions in this manual are designed to be per­formed by qualified instrument maintenance personnel.
Omega Engineering does not recommend troubleshooting or repairs beyond the scope of this manual. Problems that cannot be remedied by following the instructions in this manual should be referred to Omega. Immediate assis­tance can often be supplied by telephone. Defective com­ponents will be repaired or replaced as necessary. Returned goods should be accompanied by information requested in Section 6.
1.1 Safety Precautions
Pressure Testing Equipment must be selected and used in accordance with recognized industry codes and safety practices to avoid the possibility of misuse or misapplica­tion which could result in personal injury or property dam­age. Personnel responsible for selection and installation should also be familiar with the safety recommendations of ASME B40.1 that apply to elastic pressure elements and their application in general and specific services.
1. Pressure – Select a range so that the maximum applied pressure will never exceed the upper range limit.
2. Vibration – Excessive vibration could cause a loosen­ing of components and abnormal wear resulting in loss of instrument accuracy or failure to provide valid data.
3. Temperature – Operation of the instrument in an envi­ronment where temperatures are in excess of design rat­ings may result in loss of accuracy and failure.
4. Process – Pressure boundary materials must be resis­tant to the process media. Failure to assure compatibility may result in pressure boundary deterioration or failure. Instruments operated at high pressure or with potentially hazardous service, such as oxygen, should be carefully selected in accordance with recognized industry codes and the recommendations of ASME B40.1.
2.0 PRODUCT DESCRIPTION – THEORY OF OPERATION – CONSTRUCTION
2.1.2 Product Description
2.1.3 Construction
The Omega Type DWT1305D Dual Range Deadweight testers are precision built primary pressure standards, used for testing, setting, calibrating or repairing pressure measuring devices within the test points 15 psi (100kPa) to 10,000 psi (70,000kPa).
The deadweight tester consists of a two stage hydraulic pump containing a manifold which is pressurized during operation. Integral to the pump is a shuttle valve that allows the operator to regulate the speed of pressure increase. One connection to the manifold includes a cylin­der and a free-floating precision machined piston with a plate for holding calibrated weights. A second connection to the manifold accommodates a gauge or other pressure measuring device to be calibrated or checked. Incorporated into the manifold is a hand operated dis­placement valve that allows small adjustments in fluid vol­ume to be made without further operation of the pump handle or release valve.
The tester is dual range having two interchangeable piston
Omega
®
Installation and Maintenance Manual
for the Omega
®
Type DWT1305D Deadweight Tester
and Type DWT1327D Portable Pump
INDEX Section Description
1.
Scope of Manual Safety Precautions
2. Product Description DWT1305 Theory of Operation Construction Specifications
Product Description DWT1327 Theory of Operation Construction Specifications
3. Inspection Installation DWT1305 Operation DWT1305 Installation DWT1327 Operation DWT1327
4. Factors Affecting Operation of Deadweight Testers
5. Maintenance Instructions
Shipping Instructions
List of Illustrations
Figure 1-1 Figure 1-2 Figure 2-1 Figure 5-1
Page 4
and cylinder assemblies. One is a low pressure piston having an effective area five times larger than that of the high pressure piston. The low pressure piston is used for making measurements below 2,000 psi (14,000 kPa). The high pressure piston, with an area 1/5 that of the low pres­sure piston, is used to measure pressure through 10,000 psi (70,000 kPa). The weight masses are pre-measured and identified with the pressure values they produce when operated with the interchangeable piston and cylinder assemblies.
Pressure calibration points produced by the deadweight tester are accurate to within ± 0.1% of the reading certified traceable to the National Institute of Standards and Technoogy (NIST). The tester provides consistent, repeat­able accuracy, maintaining its pressure for an appreciable length of time regardless of temperature changes, slight leaks in the pressure system, or changes in volume of the pressurized system due to movement of a Bourdon tube or other device.
A hand jack set, three wrenches, spare O-rings, and a special adapter for making connections to pressure outlets that do not accommodate cone pipe seating, are included with each unit.
All deadweight testers are supplied with lower and back
connection offset pipe assemblies, with pipe adapters for
1
4 NPT or
1
2 NPT connections.
An all metal, double-latched, top handle carrying case is supplied with the complete tester for all fittings and attach­ments. Deadweights are packed in metal, double-latched storage-carrying boxes.
2.1.2 Theory of Operation – DWT1305D
The theory behind a deadweight tester can be expressed as simply as force acting upon a known area. Pressure produced by the pump is distributed by the manifold, to the base of a precision machined piston and to a device being calibrated or checked. Pre-selected weights loaded onto the piston platform are acted upon by gravity and develop a force that is to be equally opposed by the fluid pressure from the pump. When equilibrium is achieved, the pres­sure value is known, it being a direct result of the sum of the forces from the weights, piston platform and the piston divided by the effective area of the piston and cylinder assembly.
With the DWT1305D two piston and cylinder assemblies are supplied, one having an effective area 1/5 of the other. When using the smaller piston and cylinder assembly, five times more pressure is required to oppose the force of a constant mass being acted upon by gravity. For this rea-
Figure 1-1 Omega Portable Dual Range Deadweight Tester – Type DWT1305
Page 5
son the masses supplied with the tester are stamped with two pressure values, the value being contingent on the effective area of the piston and cylinder assembly selected.
2.1.3 Specifications – DWT1305D
Accuracy:
Combined tolerance of weights and piston and cylinder assemblies within 0.1% of reading. Weight toler­ance within 0.05% of mass. Piston and cylinder is within
0.05% of rated mean effective area. Deadweight: Non-magnetic die cast zinc alloy. Total weight to produce maximum pressure of 10,000 psi (70,000kPa) is 125 lbs. (56.7 kg). Piston & Cylinder Assemblies: High strength stainless steel piston and cylinder with brass collar and aluminum weight platform. Pump: Tw o stage, lever operated generates 10,000 psi (70,000 kPa) with 28 pounds (12.7 kg) of force on lever handle. Pump Body: Aluminum, corrosion inhibited and coated with baked blue epoxy finish. Shuttle Valve: Stainless steel bypass valve that controls rate of pressure increase and reduces operator effort when working at high pressure. Displacement Valve: A fine pitched threaded valve rod permitting vernier adjustments to fluid volume and provide
precise pressure changes or adjustment of piston travel. Limit stops prevent rod removal during normal operation. Mounting: Four bench mounting holes located in base for positive mounting to any level surface. Instrument Connections: Tw o coned pipe assemblies provide vertical calibration capability for back and lower connected gauges. Standard
1
4 inch internal NPT and 1⁄2 inch inter NPT fitting adapters are supplied. Operating Fluid: DWT1305D – Light grade machine oils, automotive petroleum base SAE 20 oils or other equiva­lent fluids suitable for use with Buna N O-ring materials.
1.5 pints required (.7 liters).
DWT1305DH – Most hydraulic oils of phosphate ester base, brake fluids, skydrol, pydraul etc., suitable for use with Butyl or Ethylene-Propylene O-ring materials. 1.5 pints required (.7 liters).
2.1.4 Certification: Standard
– Certificate of NIST traceability (accuracy/traceability statement only). Optional (CAL-3) – Certification document includes actual (as left) weight values for each weight and piston, piston diameter values, environmental data and NIST test numbers. Set includes numbered weights.
2.2 Product Description DWT1327D
Page 6
Model Piston Assembly Piston Net
Number Pressure Range Piston Area Value Number of Weights by Value Weight
L-5 L-10 L-20 L-40 L-100
psi Type Low High Low High Low
High
H-25 H-50 H-100 H-200 H-500
lbs. kg.
DWT1305D-10 DWT1305DH-10 15/200 75/1000 5 25 1323--6027 DWT1305D-20 DWT1305DH-20 15/400 75/2000 5 25 132327032 DWT1305D-30 .
0625 in.2.0125 in.
2
DWT1305DH-30 15/600 75/3000 (.4032) (.0806) 5 25 132348539 DWT1305D-50 DWT1305DH-50 15/1000 75/5000 5 25 1323810548 DWT1305D-100 DWT1305DH-100 15/2000 75/10000 5 25 13231817580
Model Piston Assembly Piston Net
Number Pressure Range Piston Area Value Number of Weights by Value Weight
L-.5 L-1 L-2 L-4 L-5
kg/cm2Type Low High Low High Low High
H-2.5 H-5 H-10 H-20 H-25
lbs. kg.
DWT1305DM-20 DWT1305DMH-20 1/20 -- .5 2.5 112126329 DWT1305DM-100 DWT1305DMH-100 1/20 5/100 .5 2.5 112126530 DWT1305DM-150 .
0625 in.2.0125 in.
2
DWT1305DMH-150 1/30 5/150 (.4032) (.0806) .5 2.5 112147534 DWT1305DM-350 DWT1305DMH-350 1/70 5/350 .5 2.5 11211210548 DWT1305DM-700 DWT1305DMH-700 1/140 5/700 .5 2.5 21212617580
Model Piston Assembly Piston Net
Number Pressure Range Piston Area Value Number of Weights by Value Weight
L-.5 L-1 L-2 L-4 L-5
bar Type Low High Low High Low High
H-2.5 H-5 H-10 H-20 H-25
lbs. kg.
DWT1305DB-20 DWT1305DBH-20 1/20 -- .5 2.5 112126329 DWT1305DB-100 DWT1305DBH-100 1/20 5/100 .5 2.5 112126530 DWT1305DB-150 .
0613 in.2.0123 in.
2
DWT1305DBH-150 1/30 5/150 (.3954) (.0791) .5 2.5 112147534 DWT1305DB-350 DWT1305DBH-350 1/70 5/350 .5 2.5 11211210548 DWT1305DB-700 DWT1305DBH-700 1/140 5/700 .5 2.5 21212617580
Model Piston Assembly Piston Net
Number Pressure Range Piston Area Value Number of Weights by Value Weight
L-50 L-100 L-200 L-400 L-500
Pascal Type Low High Low High Low High
H-250 H-500 H-1000 H-2000 H-2500
lbs. kg
.
DWT1305DA-2000 DWT1305DAH-2000
100/2000 -- 50 250 112126329 DWT1305DA-10000 DWT1305DAH-10000
100/2000 500/10000 50 250 112126530 DWT1305DA-15000
.
0613 in.2.0123 in.
2
DWT1305-DAH-15000
100/3000 500/15000 (.3954) (.0791) 50 250 112147534 DWT1305DA-35000 DWT1305DAH-35000
100/700 500/35000 50 250 11211210548 DWT1305DA-70000 DWT1305DAH-70000
100/14000 500/70000 50 250 21212617580
Dimensions in ( ) are square centimeters
TABLE 2-1
Page 7
2.2.1 Construction
The Omega Type DWT1327D Portable Test Pumps are rugged, versatile pressure transfer standards, used for testing, setting, calibrating or repairing pressure measur­ing devices with ranges up to 10,000 psi (70,000kPa). A selection of high accuracy test gauges, with a precision of ±0.25% of span, are available as the standard to which the device under test is compared.
The main component to the tester is a two stage hydraulic pump containing a manifold which is pressurized during operation. Integral to the pump is a shuttle valve that allows the operator to regulate the speed of pressure increase. One connection to the manifold has a straight pipe with a precision test gauge attached serving as the reference standard. A second connection to the manifold accommodates a gauge or other pressure measuring device to be calibrated or checked. Incorporated into the
manifold is a hand operated displacement valve that allows small adjustments in fluid volume to be made with­out further operation of the pump handle or release valve.
Pressure values produced by the pump are able to be clearly read to within ±0.25% accuracy of the span of the precision test gauge selected for testing. A hand jack set, three wrenches, spare O-rings, and special adapter for making connections to pressure outlets that do not accom­modate cone pipe seating, are included with each unit.
Test pump accessories include lower and back connection offset pipe assemblies, with pipe adapters for
1
4 and
1
2
NPT connections.
An all metal, double-latched, top handle carrying case is supplied with the complete tester to hold all fittings and attachments.
2.2.2 Theory of Operation – DWT1327D
Figure 1-2 Omega Portable Test Pump – Type DWT1327
Note: (97) Gauges are not provided with the set and may be ordered separately.
Page 8
The operating principle behind the DWT1327D Test Pump is a simple form of comparison. Pressure produced by the pump is equally distributed by the manifold to a test gauge and to a device being calibrated or checked. The reading of the test gauge serves as the reference to which other device readings are compared.
Selection of the test gauge range is the determining factor in establishing the precision to which a comparison check is to be made. Test gauge accuracy is expressed as a per­centage of its span.
Proper selection of the test gauge range must be made to minimize the amount of the unit error. If a 5,000 psi (35,000kPa) range test gauge were used to test devices whose span values were less than a fourth of the test gauge range, the unit error of 12.5 psi (87kPa) would be greater than 1% of the test pressure. Therefore, it is impor­tant to select a test gauge that has a full scale range equal to or only slightly in excess of the pressure value to be measured.
2.2.3 Specifications – DWT1327D Accuracy:
1
4% of 1% of test gauge span.
Test Gauges: Omega PGT45L 41⁄
2˝ test gauges are
recommended.
Pump: Tw o stage, lever operated generates 10,000 psi (70,000kPa) with 28 pounds (12.7kg) of force on lever handle.
Pump Body: Aluminum, corrosion inhibited and coated
with baked blue epoxy finish.
Shuttle Valve: Stainless steel bypass valve that controls rate of pressure increase and reduces operator effort when working at high pressure.
Displacement Valve: A fine pitched threaded valve road permitting vernier adjustments to fluid volume to provide precise changes. Limit stops prevent road removal during normal operation.
Mounting: Four bench mounting holes located in base for positive mounting to any level surface.
Instrument Connections: Tw o coned pipe assemblies provide vertical calibration capability for back and lower connected gauges. Standard
1
4 inch internal NPT and
1
2
inch internal NPT fitting adapters are supplied.
Operating Fluid: DWT1327D – Light grade machine oils, automotive petroleum base SAE 20 oils or other equiva­lent fluids suitable for use with Buna N O-ring materials.
1.5 pints required (.7 liters).
DWT1327DH – Most hydraulic oils of phosphate ester base, brake fluids, Skydrol, Pydraul etc., suitable for use with Butyl or Ethylene-Propylene O-ring materials. 1.5 pints required (.7 liters).
DWT1327DO – Distilled water with compatible corrosion inhibitor. A .1% solution of sodium dichromate is a suitable inhibitor. 1.5 pints required (.7 liters).
2.2.4 Certification:
Exclusively dependent upon certification provided with test instrument.
3.0 SETTING UP THE TESTER
Page 9
3.1 Inspection
3.1.1 Component Check DWT1305D
The deadweight tester, as shipped from the factory, con­tains all of the components necessary for operation except pump fluid and the test instrument(s). Figure 1-1 provides an illustrated breakout of these items for quick verification.
3.1.2 Component Check DWT1327D
The portable test pump, as shipped from the factory, con­tains all of the components necessary for operation except pump fluid. Figure 1-2 provides an illustrated break-out of these items for quick verification.
3.1.3 Claim for Missing or Damaged Goods
In the event it is determined that an item is missing or damaged, contact Omega Engineering immediately. Disposition of damaged or missing goods will be provided at that time.
3.2 Installation Type DWT1305D
3.2.1 Set-up Procedure (See Figures 1-1 and 2-1)
To remove the tester from the case, release the spring clip (93), depressing the clip by hand, grasp the pump reser­voir and lift up the back end of the pump assembly.
Having removed the pump assembly from its carrying case (91), remove the two plastic shipping plugs (57).
Select a lower connection (64) or back connection (66) off­set pipe and connect to inner pump body outlet as shown in Fig. 2-1.
Remove the reservoir filler plug (35) and fill reservoir with operating fluid.
3.2.2 Bench Space Required (See Figure 1-1)
Allow a space: 10 inches (25.4 cm) width 27 inches (68.6 cm) depth 22 inches (55.9 cm) height – allows for maximum load of deadweights
3.2.3 Storage Space Required One Tester Box:
101⁄
2 inch (2.6 cm); 22
1
2 inch (57 cm) depth; 10 inch (25
cm) height
One Weight Box:
10 inch (25 cm) width; 10 inch (25 cm); 10 inch (25 cm) height
Storage space dimensions are larger than actual size of tester carrying case and deadweight carrying case by approximately _ inch, (1.3 cm), in order to facilitate ease of handling in storage.
Note: For catalog numbers DWT1305D, DH-100,
DWT1305DM, DMH-150, DWT1305DB, DBH-700 and DWT1305DA, DAH-700 allow storage space for two weight boxes.
Page 10
Figure 2-1 Omega Portable Deadweight Tester DWT1305D
Page 11
3.3 Operation DWT1305D
3.3.1 Piston and Cylinder Selection
The DWT1305D Tester is a dual range device in that most models utilize two interchangeable piston and cylinder assemblies for measurement of pressure throughout the entire range. Each piston and cylinder assembly has a minimum and maximum operating range. First, determine what pressure points need to be measured and select the appropriate assembly. The pressure range of the assem­blies is shown in Table 2-1. When priming the pump the high pressure piston and cylinder assembly should be used.
3.3.2 Priming the Pump
The offset pipe (64 or 66 Figure 2-1) should already be secured in place and the reservoir should contain the operating fluid (refer to Installation Section 3.2).
Close the release valve (37) by turning it clockwise and open the air vent by loosening the filler plug (35) a few turns. The shuttle valve (51) should be pulled outward from the pump body, the end of the knob approximately even with the edge of drip pan. In this position the high volume displacement mode is selected and the pump is self-priming. Operate the pump handle several times using full strokes, until you see fluid appear in the outer pump body outlet.
The cone seat (82) is located in the base of the outer­pump body outlet and serves as the sealing surface between the bottom of the piston and cylinder assembly and the pump body. Confirm the coned end of the seat is facing up and the cylindrical end of it is facing down. Thread the high pressure piston and cylinder assembly (79) or (67) into the outer body outlet as shown, using the wrenches provided.
Operate the pump handle a few more times until fluid is observed at the end of the offset pipe. Connect the gauge or other device to be tested to the offset pipe. For purpos­es of priming, the device must be designed to withstand the full operating pressure of 10,000 psi (70,000kPa). If a lower range device must be used, extreme care must be exercised to avoid overpressuring its pressure element. Seal the connection to the device by tightening nut adapter (62) and adapter (60) until the coned end of the pipe is forced into its inlet. If necessary, rotate the device for view­ing by loosening nut adapter (62), set it to the proper posi­tion and retighten nut adapter.
Again, operate the pump handle a few more times. After a few strokes positive pressure will develop and the pump handle will begin to resist pumping action. Raise the pump handle to its uppermost travel position. Loosen the bleed screw (30) a half turn counter clockwise and slowly oper­ate the pump handle through a downward stroke until fluid flows steadily from the bleed vent. Close the bleed screw just prior to completing the downward handle motion. Repeat this action until no air bubbles are observed in the fluid flow. (Note – the bleed screw must be closed when the handle is being raised). Push the shuttle valve in towards the pump body and continue pumping to the desired test pressure. Open the release valve (37) to vent the pressure. To check operation pull the shuttle valve out­ward, close the release valve and operate the pump han­dle several times. When handle resistance is felt, push the shuttle valve inward and continue pumping within the range of the test device.
Entrapped air will prevent the pump from operating in the high pressure valve position or cause it to achieve only
partial pressure. Repeat the above bleed procedure as necessary to assure all air is removed. Once air is removed the pump will continue to operate without further attention providing the reservoir level is maintained.
3.3.3 Weights
The weight set consists of a selection of various masses that will produce desired pressure increments when oper­ated with the appropriate piston and cylinder assembly. Each weight is stamped with two pressure values. When applied to the high pressure piston and cylinder assembly the equivalent pressure value is indicated next to the letter “H”. Conversely, when applied to the low pressure piston and cylinder assembly the equivalent pressure value is indicated next to the letter “L”. The piston and piston plat­form also contribute to the total mass. Their equivalent pressure value is stamped on the top of the platform.
During normal operation, selected weights are added to the plate and piston assembly to equal the desired pres­sure value.
3.3.4 Levelness
The deadweight tester must be level to function properly. The unit may be leveled by placing a bubble type level on top of the piston plate and revolving it slowly. Shims may be used between drip pan and bench to level the piston plate. The unit is level when the position of the bubble does not change within the glass as it is rotated on top of the piston assembly.
3.3.5 Making the Test
Add weight (99 Figure 1-1) to the piston plate to give desired calibrating pressure.
Pull the shuttle valve (51) outward from the body and close the pressure release valve (37). Operate the pump handle until the fluid pressure forces the piston to raise the weights. When pumping, the weights should be rotated slowly to decrease cylinder wall friction. If pump handle resistance is difficult and the weights have not risen, push the shuttle valve inward and continue pumping. With the shuttle valve pushed in, the fluid displacement of each stroke is reduced, thereby requiring less effort to continue pumping.
The piston assembly has a maximum lift of
3
4 inch (1.9
cm). It is recommended that readings be taken at mid­point, or
3
8 inch (1 cm) lift. Small adjustments to the piston
lift can be made with the pump handle or the displacement valve (41).
Improper readings will result if the piston plate is so low that it rests on the bushing, or so high that the internal stop on the piston assembly is touching the underside of the bushing.
Spin the weights by hand and take readings only when the weights are spinning. Speed of rotation is unimportant, although a slow speed is more convenient and recommended.
To release pressure, turn valve (37) counterclockwise slowly.
3.3.6 To Dismantle (See Figure 1-1)
To replace the tester in its case, remove the gauge offset pipe assembly (64 or 66) and the piston and cylinder assembly (67 or 79). Install shipping plugs (57) in body outlet holes. Screw in the displacement valve (41) until it stops. Close the vent plug (35). Open the release valve (37) approximately _ turn. Replace the tester in its case, reservoir last. Engage the spring clip catch.
Page 12
3.3.7 General Precautions
It is important that the deadweight tester be connected to a leak tight system.
The deadweight tester should be set-up so that the axis of the located piston is vertical, the weights carefully cen­tered on that axis, and the piston rotated during use. The purpose of the rotation is to spread the lubricant over the entire surface between piston and cylinder, so that there will be no metal-to-metal contact. If rotation is not main­tained, the lubricant film will not cover the surface properly, and readings will be in error.
The high and low pressure piston assemblies have an internal overload stop, which prevents the piston and cylin­der from being forced apart if weights are accidentally removed.
3.3.8 Operating Fluids Standard Tester:
Any medium weight oil may be used (including automotive oils S.A.E. 10, 20 or 30) S.A.E. 20-W recommended.
Hydraulic Tester:
Any hydraulic fluid (silicate or phosphate base). Skydrol or Pydraul is suitable.
For normal operation, it is not necessary to change the weight of oil for various pressures. A lighter oil may be used where low pressures are being checked, and a heav­ier oil where higher pressures are being tested.
The reservoir can be refilled while the tester is operating under pressure.
3.3.9 Caution
Standard testers designed for oil service may not be used with water for oxygen service.
3.3.10
Hydraulic service testers should not be filled with water or any oil other than hydraulic fluid. Serious pump failure may occur due to O-ring damage.
3.4. Installation Type DWT1327
3.4.1 Set up Procedure (See Figures 1-2 and 2-2)
To remove the test pump from the case, release the spring clip (93), depressing the clip by hand, grasp the pump reservoir, and lift up the back end of the pump assembly.
Remove the two plastic shipping plugs (57) from the outlet connections. Remove the threaded filler plug (35) from reservoir filling hole. Remove the offset pipe (64) or (66) from the carrying case and connect to the inner pump body outlet as shown in Figure 2-2. Remove the straight pipe extension (58), and assemble to outer vertical pump body outlet as shown in Figure 2-2.
3.4.2 Connecting The Gauge (See Figure 2-2)
Select a test gauge that is adequate for the pressure range desired. Assemble the test gauge to the straight pipe extension, using adapter nut (62), collar (61), adapter (60) and reducer bushing (59).
Seal the connection by tightening nut adapter (62) and adapter (60) until the coned end of the pipe is forced into the gauge socket. Rotate the gauge for viewing by loosen­ing nut adapter (62), setting the gauge to the desired posi­tion, and re-tightening nut adapter.
Repeat the above procedure to connect the gauge being tested to the offset pipe assembly (64) or (66).
3.4.3 Bench Space Required
Allow a space: 10 inch (2.5 cm) width 23 inch (5.8 cm) depth 22 inch (5.6 cm) height allows for testing gauges up to 8_ inch (250 mm) size
3.4.4 Storage Space Required
101⁄
2 inch (27 cm) depth
25 inch (63.5 cm) width 10 inch (25 cm) height
Storage space dimensions are larger than actual size of test pump carrying case by approximately _ inch (1.3 cm), in order to facilitate ease of handling in storage.
Page 13
Figure 2-2 Omega Portable Test Pump – Type DWT1327D
Page 14
3.5 Operation DWT1327D
3.5.1 Test Gauge Selection
The portable test pump is used for calibrating instruments such as pressure gauges, pressure switches or other pressure devices rated up to 10,000 psi (70,000kPa).
First, select the proper test gauge, depending upon the pressure range desired. When priming the pump a test gauge rated to 10,000 psi (70,000kPa) should be used.
3.5.2 Priming The Pump
The offset pipe (64 or 66 Figure 2-2) and straight pipe extension (58) should already be secured in place and the reservoir should contain the operating fluid (refer to instal­lation Section 3.4).
Close the release valve (37) by turning it clockwise and open the air vent by loosening the filler plug (35) a few turns. The shuttle valve (30) should be pulled outward from the pump body, the end of the knob approximately even with the edge of the drip pan. In this position the high volume displacement mode is selected and the pump is self-priming. Operate the pump handle several times using full strokes, until you see fluid appear at the top of the straight pipe extension. Connect the test gauge to it (refer to Installation Section 3.4 for proper gauge connection).
Operate the pump handle a few more times until fluid is observed at the end of the offset pipe. Connect the gauge or other device to be tested to the offset pipe. For purpos­es of priming, the device must be designed to withstand the full operating pressure of 10,000 psi (70,000kPa). If a lower range device must be used, extreme care must be exercised to avoid overpressuring its pressure element. Seal the connection to the device by tightening nut adapter (62) and adapter (60) until the coned end of the pipe is forced into its inlet. If necessary, rotate the device for view­ing by loosening nut adapter (62), set it to the proper posi­tion and retighten nut adapter.
Again, operate the pump handle a few more times. After a few strokes positive pressure will develop and the pump handle will begin to resist pumping action. Raise the pump handle to its uppermost travel position. Loosen the bleed screw (51) a half turn counter clockwise and slowly oper­ate the pump handle through a downward stroke until fluid flows steadily from the bleed vent. Close the bleed screw just prior to completing the downward handle motion. Repeat this action until no air bubbles are observed in the fluid flow. (Note – the bleed screw must be closed when the handle is being raised). Push the shuttle valve in towards the pump body and continue pumping to the desired test pressure. Open the release valve (37) to vent the pressure. To check operation pull the shuttle valve out­ward, close the release valve and operate the pump han­dle several times. When handle resistance is felt, push the shuttle valve inward and continue pumping within the range of the test device.
Entrapped air will prevent the pump from operating in the high pressure valve position or cause it to achieve only partial pressure. Repeat the above bleeding procedure as necessary to assure all air is removed. Once air is removed the pump will continue to operate without further attention providing the reservoir level is maintained.
3.5.3 Making The Test
Tu rn displacement piston handle (41) to mid-position, so that it may be used for setting an exact pressure on the gauge. The displacement piston has internal stops which prevent unscrewing or accidental loss of pressure.
Clockwise rotation of displacement piston will produce an increase in pressure; counterclockwise will decrease pressure.
Prior to taking readings, both gauges should be finger­tapped lightly at the center of the gauge face, to eliminate any movement friction. Note the pressure readings on the test gauge and the gauge under test. If the pressure indi­cated on the gauge under test is not equal (within the tol­erance of the gauge) to the pressures of the master test gauge, the gauge being tested requires calibration.
CAUTION:
Do not pump handle to pressures greater than the pres­sure range of the gauges connected to the test pump, as this may damage the gauges.
3.5.4 To Release Pressure
Open release valve slowly (37), until pressure returns to zero. Do not loosen any connections until pressure in the gauge tester has reached zero, as indicated on the test gauge.
If additional gauges are to be tested, close the release valve when the pressure reaches zero. This will prevent complete drainage of oil in the tester back to the reservoir.
Unseal the gauge that has been tested by unscrewing adapter nut (62) from connector (60) several turns. Unscrew the gauge from bushing (59) if used, or connec­tor (60), and remove.
Remove the test gauge in the same manner as the gauge under test.
3.5.5 To Dismantle
To replace test pump in case, remove gauges and both tube assemblies. Put shipping plugs (57) into pump and tighten. Close filler plug (35).
3.5.6 Operating Fluids Standard Test Pump:
Any medium weight oil may be used (including automotive oils S.A.E 10, 20 or 30) S.A.E. 20-W recommended.
Hydraulic Test Pump:
Any hydraulic fluid (silicate or phosphate base). Skydrol or Pydraul is suitable.
Oxygen Test Pump:
Distilled or demineralized water with a compatible corro­sion inhibitor added. A .1% solution of sodium dichromate is a suitable inhibitor.
For normal operation it is not necessary to change the weight of oil for various pressures. A lighter oil may be used where low pressures are being checked, a heavier oil where higher pressures are being tested.
The reservoir can be refilled while the test pump is operat­ing under pressure.
3.5.7 Caution
Standard testers designed for oil service may not be used with water for oxygen service.
3.5.8
Hydraulic service testers should not be filled with water or any oil other than hydraulic fluid. Serious pump failure may occur due to O-ring damage.
4.0 FACTORS AFFECTING OPERATION OF DEADWEIGHT TESTERS
4.1 Deadweight Tester Accuracy
Excellent accuracy is possible using the deadweight tester through analysis and control of certain factors. If the tester’s rated accuracy of 1/10th of 1% is adequate, then
Page 15
the nominal pressure (sum of the denominations of the weights loading the piston) may be assumed to be correct. The pressure normally developed is determined by this formula: Deadweight Tester Pressure =
Mass of weights plus piston mass
Effective area of piston and cylinder
P = (M-ph) g
A gs P=pressure M=mass of the load on the piston A=effective area of the piston in sq. inches, or sq. cm g/gs = ratio of the value (g) of gravity at the point of use to
the standard value of gravity (gs) P=density of liquid used in test h=difference in level between gauge being tested, and
the bottom of the piston
The deadweight tester is capable of measuring pressures to proper corrections after analyzing these factors.
4.2 Gravity, Calibrated Weights
Weights furnished are calibrated at standard gravity of
980.665 gals. If precise accuracy is required, the error introduced by change of weight due to change in gravity at the locality should be calculated and included in results.
The mechanism of an Omega Pressure Gauge includes a Bourdon tube and geared movement, which is unaffected by variations in gravity. Conversely, the pressure devel­oped by a deadweight tester is proportional to the value of gravity. Readings of this type pressure gauge will corre­spond to those of a deadweight tester when the tester is subjected to standard gravity (gs = 980.665 gals. in the International System). In southern sections of the United States, the value of gravity may be several thousandths less than the standard value, if the latitude (Ø) and the ele­vation above sea level (a) for an area are known, the approximate value of (g) in gals. may be calculated from this formula:
g = 980.632-2.586 COS 2Ø
+.003 COS 4Ø - .000094a
Ø=Latitude (Degrees) a=Elevation above sea level (ft)
4.3 Effective Measured Area: Chamber
The effective area of the deadweight tester may be deter­mined by the average of the cross sectional area of the piston and the area of the cylinder bore. This effective area is affected somewhat by temperature, and by the elastic distortion of the piston and cylinder when pressure is being applied. The effective area of a stainless steel piston and cylinder increases approximately .068% with a 50 degree F (28 degree C) change in temperature. The pressure will, therefore be less than indicated at high temperatures.
4.4 Mass, Height, and Buoyancy
The density of air at room temperature and sea level pres­sure is about 0.0012 grams/c.c., and the mass of the pis­ton assembly and weights under these conditions will be reduced by about one part in 7,000 or .014%. When the submerged part of the piston has a uniform cross section, as with the DWT1305D Deadweight Tester, a buoyancy correction is not necessary. In other designs, the piston is sometimes enlarged to provide a stop for its upward mo­tion or for increased strength. If these enlargements are submerged in liquid, a buoyancy correction is necessary.
4.5 Absence of Friction
By rotating the weights and piston, friction effects are greatly reduced.
4.6 Head of Transmitting Fluid
Oftentimes, the gauge being tested, or the point at which pressure is being measured, is not at the same level as the lower end of the piston. A correction, therefore, should be made for the pressure distance between these points; the height is considered positive when the gauge is above the piston. When oil is used in the gauge tester, the cor­rection will be approximately 0.03 psi (.2 kPa) for each inch (2.54 cm) difference in level.
4.7 Method of Operation
It is important that the piston be kept floating in mid-posi­tion, either spinning or oscillating.
4.8 Levelness
The piston assembly should be vertical to within ±1 degree. A 3 degree tilt to piston axis may cause a .13% of 1% error. The deadweight tester is manufactured and test­ed to 1/10th of 1% accuracy, to this degree of levelness. A tilt piston/cylinder axis causes excessive friction, due to side loading of the piston against the cylinder.
4.9 Cleanliness
The weights have been manufactured and tested to a pre­cision of 0.05%. A buildup of dirt and grease may cause the weight value to exceed its original tolerance and produce erroneous pressure readings.
Periodically clean the weights to assure proper performance.
5.0 MAINTENANCE INSTRUCTION
5.1 General Maintenance
The Deadweight tester and test pump are designed to serve as precise pressure measuring standards. They are precision built units and should be cared for in the same manner as other sensitive laboratory equipment. General maintenance is limited to cleaning and replacement of O­ring packings, which can be done with tools supplied with the equipment and requires only limited disassembly of the gauge tester.
The piston supporting the weight platform has been manu­factured to very close tolerances. It has an area accuracy of 1/20th of 1% and a weight accuracy of 1/20th of 1%. The deadweights have been certified traceable to National Institute of Standards and Technology with the accuracy of the finished weights better than 1/20th of 1%. To maintain deadweight tester accuracy, handle the weights with care and keep the piston and cylinder in clean condition.
The tester should be flushed with a solvent occasionally, preferably every six months, so that operating fluid is always clean. This will prolong component life, and provide protection against possible sticky action between the pis­ton and cylinder. After cleaning, always lubricate parts before assembly.
If a deadweight tester is not used for long periods of time, or if a piston and cylinder assembly is being replaced, the piston should be removed from the cylinder and coated with the hydraulic fluid, so that it never operates in a dry state. When removing or replacing the piston, it should be rotated back and forth.
Should a piston or cylinder wear excessively, the tester will leak oil at a high rate, and will not be operable. A new piston and cylinder assembly should be installed.
Page 16
Piston wear will result from improper or contaminated lubrication, excessive dirt, or from several years of continuous use.
5.2 Cleaning
Both the Tester and Test Pump should be cleaned thor­oughly whenever the operating fluid is seriously contami­nated with dirt, grit, or chemicals. A good practice is to clean the test unit prior to anticipated periods when the unit will not be in use.
5.3 To Disassemble for Cleaning (See Figure 2-1, Figure 2-2 and Figure 5-1)
It is necessary to disassemble only those components which come in contact with the operating fluid. Complete disassembly is rarely necessary.
a. Remove pipe extension assembly (58) or piston and
cylinder assembly(67) and (79).
b. Then remove lower or back connection pipe assembly
(64, 66).
5.4 Removing Pump Handle and Piston
a. Remove four retaining rings (8). b. Slide out two clevis pins (9). c. Remove clevis (10). d. Lift hand assembly (4) with piston pin (7) piston (20)
O-ring packing (12, 19) and back-up ring (11) attached to it, out of piston sleeve (6).
e. Remove back-up washers (18, 11) and O-ring packing
(12, 19). f. Unscrew piston (20) g. Remove piston sleeve (6) and O-ring packing (45, 16)
and back-up washers (46, 17)
5.5 Removing Reservoir and Fill Tube
a. Unscrew filler plug (35) from reservoir. b. Remove gasket (36). c. Remove reservoir (5) by removing three screws (24). d. Twist reservoir to free O-ring packing (21) seat. e. Remove O-ring packing (21) from body (2). f. Remove fill tube assembly (23) and O-ring packing (27). g. Remove plug (22).
5.6 Removing Shuttle Valve, Shuttle Valve Plug and Check Valve
a. Remove set screw (33) and spring (32). b. Slide shuttle valve piston (31) out of body (2). Operate
back and forth as necessary to dislodge pin (31) engagement.
c. Remove shuttle pin (31) by pushing it into the shuttle
valve piston cavity using a plastic shaft less than
1
8
inch (3mm) in diameter.
d. Unscrew shuttle valve plug (43) and remove O-ring
packing (27).
e. Remove check valve spring (14) and check valve (44). f. Remove O-ring packing (15) from check valve (44).
5.7 Removing Bleeder Valve and Check Valve Assembly
a. Unscrew bleeder valve (51) b. Unscrew bleed plug (50) and remove O-ring packing
(27).
c. Remove valve spacer (29) check valve spring (14)
and check valve (44).
d. Remove O-ring (15) from check valve (44).
5.8 Removing High Pressure Check Valve
a. Unscrew plug (28) and remove O-ring packing (27). b. Unscrew adjusting screw (26). c. Remove guide rod (13), check valve spring (14), and
check valve (44).
d. Remove O-ring packing (15) from check valve (44).
5.9 Removing Release Valve
a. Remove release valve assembly (37) and O-ring
packing (27)
b. Disassemble valve assembly by rotating handle
counter-clockwise until valve stem is free from valve
body. c. Remove O-ring packing from valve stem. d. Remove screw (38) and seal (39) by turning counter-
clockwise. e. Remove O-ring packing (40).
5.10 Removing Displacement Valve
a. Remove displacement valve assembly (41). b. Remove O-ring packing (27). c. Remove handle from stem by rotating handle counter-
clockwise. This is not required for normal cleaning. d. Rotate the stem clockwise (using screwdriver in the
slot at the top of the stem) until it is free from body. e. Remove nut back-up washers and ring packing from
stem.
5.11 Stem Body from Drip Pan
a. Take out three bolts (25). This is not required for
normal cleaning. b. Remove body assembly (2) from drip pan (3). This is
not required for normal cleaning.
5.12 Inspection
Visually inspect these parts for wear, damage, chips, cracks and stripped threads:
Body Assembly Threads
Piston
Piston Sleeve (6)
Displacement Stem and Mating Surface in Body
Back-up Washers
Moving O-ring Packings
Dead Weights
Piston and Cylinder Assembly – DWT1305D
Test Gauges – DWT1327D
Back-up washers must fit snugly into piston sleeve and into displacement valve body. Then check static (non-mov­ing) packing for pinch marks, tearing or extrusion. Check all valve seats for scratches and roughness.
Replace all worn or damaged parts. Replace all O-ring packings at each overhaul. Coat O-ring packings with suit­able lubricant before replacing, to prevent sticking and tearing during assembly and tightening of connections.
5.13 Reassembly
When reassembling, use where possible the wrenches supplied with the equipment. These wrenches permit enough leverage to seal all connections. Excessive tight­ening of parts with tools other than those supplied may cause distortion and eventual failure of threaded portions of the tester body assembly casting. When replacing O­ring packings, coat them with suitable lubricant to prevent sticking and tearing during assembly and tightening of connection.
5.14 To Reassemble
Body assembly on drip pans. Replace body assembly (2) on drip pan (3) using bolts (25).
5.15 Replacing Displacement Valve
a. Place O-ring packing between back-up washers. b. Attach back-up washers and O-ring packing to dis
placement valve stem with nut. c. Thread stem into body from the bottom – use screw
driver in the slot at top of stem.
Page 17
d. Screw stem through body until the back-up washers,
and O-ring packing are fully enclosed in the body. CAUTION: Do not damage O-ring packing when
threading stem through body. e. Screw handle onto stem. f. Replace displacement valve assembly (41) with
O-ring packing (27) in body assembly (2).
5.16 Replacing Release Valve
a. Place O-ring packing (40) in body casting (2) hole. b. Insert seat (39) into body casting hole. c. Thread & tighten screw (38) into body casting hole.
CAUTION: Be sure that stem of seat (39) fits into
screw hole. d. Place O-ring packing on valve stem. e. Screw handle onto valve stem. f. Screw valve stem into body until end of stem does not
extend beyond valve body.
CAUTION: Do not damage O-ring packing when
threading valve stem through valve body. g. Replace O-ring packing (27). h. Replace release valve assembly (37) in body
assembly (2).
5.17 Replacing High Pressure Check Valve
a. Place O-ring packing (15) on check valve (44). b. Slide check valve (44), check valve spring (14), and
guide rod (13) into body assembly (2). c. Replace adjusting screw (26) and screw it in until it
stops. d. Then turn (26) back two complete revolutions. e. Replace plug (28) with O-ring packing (27) in body
assembly (2).
5.18 Replacing Bleeder Valve and Check Valve Assembly
a. Replace O-ring packing (15) on check valve (44). b. Slide check valve (44), check valve spring (14) and
valve space (29) into body assembly (2).
c. Replace bleed plug (50) with O-ring packing (27) into
body assembly (2).
5.19 Replacing Shuttle Valve, Shuttle Valve Plug and Check Valve
a. Insert shuttle valve piston (3) into body (2). b. Slide shuttle pin (31) into 1/8 inch (3mm) diameter
opening adjacent to fill tube (23) port. Apply slight inserting pressure to shuttle pin (31) with plastic shaft and operating shuttle valve (30) to insure proper pin engagement.
c. Replace spring (32) directly behind shuttle pin (31)
and secure with set screw (33). Set screw must be
recessed by .050 inch (12mm) minimum. d. Replace O-ring packing (15) on check valve (44). e. Slide check valve (44) and check valve spring (14)
into body assembly (2).
5.20 Replacing Reservoir and Fill Tube
a. Replace plug (22). b. Replace fill tube assembly (21) and O-ring packing (27). c. Replace O-ring packing (21) on body assembly (2). d. Replace reservoir (5) using three screws (24). e. Replace filler plug (35) and gasket (36) on body.
5.21 Replacing Pump Handle and Piston
a. Replace piston sleeve (6) and O-ring packing (45, 16)
and back-up washers (46, 17) in body assembly (2). b. Place O-ring packing (12, 19) between back-up
washers (18, 11) c. Attach back-up washers (18,11) and O-ring packing
(12, 19) to piston (20). d. Lubricate piston and sleeve.
e. Attach assembled piston to handle assembly (4).
CAUTION: Do not damage O-ring packing (12, 19) when sliding assembled piston into piston sleeve (6).
f. Attach clevis (10) to handle assembly and to body
assembly (2), using clevis pins (9) and retaining rings (8).
5.22 Replacing Piston/Cylinder Assemblies
The piston/cylinder assembly for a deadweight tester should be replaced when excessive wear is detected on any component part. Worn piston assemblies are usually noted by:
1. Excessive leakage of operating fluid through piston assembly when under pressure.
2. Seizure of piston in cylinder.
3. Damaged piston plate, damaged cylinder threads, or damage to any component part that results in inferior performance, or malfunctioning of the unit.
The piston/cylinder assembly is available as a unit only. This assures the user of maximum accuracy in his Omega Deadweight Tester and maintains certified traceability to the National Institute of Standards and Technology.
Page 18
Pressure does not build up when pumping handle.
Insufficient fluid level in reservoir.
Piston O-rings worn or ruptured.
Bleed port check valve or shuttle check valve inoperative.
Fill tube assembly and filter plugged.
Shuttle valve pushed in for small fluid dis­placement when large fluid dis­placement is required.
Add fluid.
Inspect O-rings and replace as necessary. If O-rings are new, verify their fluid compatibility.
Inspect O-rings and replace as necessary.
Remove and clean fill tube assembly. Refill reservoir with clean fluid.
Pull shuttle valve outward from body.
High pressure does not increase when pumping handle while shuttle valve is pushed in.
Air trapped in pump piston.
Rapid pump han­dle strokes which lessen fluid flow through the pump piston.
Reprime pump.
Operate pump handle with smooth moderate action.
Pump handle rises after pumping.
High pressure check valve O-ring worn or ruptured.
Inspect O-ring and replace as necessary.
Piston plate (with deadweights) drops rapidly.*
Unable to hold constant pressure.
Worn piston and cylinder assembly.
Use a heavy oil temporarily. Replace piston & cylinder.
Stacked dead­weights wobble when spinning pis­ton plate.*
Damaged dead­weights.
Check dead­weights for visible damage (bends, dents, nicks, etc.) and alignment.
Piston plate assembly will not spin.*
Too heavy an oil grade being used.
Flush & fill unit with proper grade of oil (SAE 20 or SAE 10). Replace piston & cylinder assembly. Disassemble and flush complete test unit with kerosene or alcohol. Reassemble unit.
Pressure builds up when pumping handle, but decreases when pumping is stopped.
Leakage at outlet or gauge connec­tions(s).
High pressure check valve O-ring worn or ruptured.
Defective pressure release valve.
Inspect connec­tions and tighten as necessary.
Inspect O-ring and replace as neces­sary.
Hand tighten release valve.
Inspect release valve seating sur­faces. Replace as necessary.
Remove release valve seat and inspect O-ring underneath. Replace as necessary.
5.4 Troubleshooting Chart
Corrective
Symptom Case Action
Corrective
Symptom Case Action
*These symptoms apply only to the DWT1305D Deadweight Testers.
Page 19
5.0 RETURN SHIPPING INFORMATION
Company Name
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
Phone Number
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
FAX Number – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
Person to Contact
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
Address
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
email Address
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
Model
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
Serial Number
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
Symptoms
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
– – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – – –
Page 20
4A
Page 21
REPLACEMENT PARTS
PORTABLE DEADWEIGHT TESTER – TYPE DWT1305D
PORTABLE TEST PUMP – TYPE DWT1327D
Qty.
Item Per
No. Asmy. Part No. Description
21603D001-01 Body Only, Pump 31 LB285A Drip Pan 41603X010-01 Handle 4A 1 636X003-01 Grip 51 LF341A Reservoir 61610C001-01 Sleeve, Piston 71 LA810A Pin, Piston 84 BA221L Retaining Ring
92 LB810A Pin, Clevis 10 1 LD278A Clevis 11 1 607A002-02 Backup Ring 12 1 607A001-02 “O” Ring 13 1 LB802 Guide Rod 14 3 LA242 Spring, Valve 15 4 SSSS882 “O” Ring 16 1 607A001-03 “O” Ring 17 1 607A002-03 Backup Ring 18 1 607A002-01 Backup Ring 19 2 607A001-01 “O” Ring 20 1 614C001-01 Piston, Dual Diameter 21 1 BA122R “O” Ring 22 1 604A003-01 Plug 23 1 615B001-01 Fill Tube Assembly 24 3 605A002-06 Screw, Round Head 25 3 SAFH37 Screw, Allen Socket Head 26 1 LK260A Adjusting Screw 27 6 BA122V “O” Ring 28 1 LK696A Plug 29 1 Spacer, Valve, Use Item 50 30 1 614B002-01 Piston, Shuttle 31 1 609A001-01 Pin, Shuttle 32 1 612B001-01 Spring, Shuttle Pin 33 1 605A001-01 Set Screw 34 2 LP696 Plug, Shipping 35 1 604B002-01 Plug, Fill 36 1 607A003-01 Gasket, Fill Plug 37 1 LAH292C Valve Assembly, Release 37a 1 328A105-01 Knob 37b 1 LD870B Valve Stem 37c 1 LAH292A Body, Release Valve 38 1 LD83A Scew, Seat 39 1 LD911A Seat 40 2 BA122U “O” Ring 41 1 LAG292C Valve Assembly, Displacement 41a 1 328A105-02 Knob 41 b 1 LE83 Nut 41c 2 LB122 Backup Washer 41d 1 LB870A Stem 41e 1 LAG292A Body, Displacement Valve 42 1 LA521A Sleeve 43 1 604B001-01 Plug, Shuttle Valve 44 3 609B003-01 Check Valve 45 1 BA122W “O” Ring 46 1 607A002-04 Backup Ring 50 1 604B004-01 Plug, Bleed 51 1 609B004-01 Bleed Valve 52 1 618B001 Nameplate 58 1 LK227 Pipe Extension Assembly
PA RTS IN THE SECTION BELOW ARE USED IN HYDRAULIC SERVICE UNITS. THESE UNITS CAN BE RECOGNIZED BY THE INCLUSION OF THE LETTER “H” IN THE MODEL DESIGNATION.
PA RTS IN THE SECTION BELOW ARE USED IN OXYGEN (DISTILLED WATER) SERVICE UNITS WHICH ARE IDENTI­FIED BY THE INCLUSION OF THE LETTER “O” IN THE MODEL DESIGNATION.
Qty.
Item Per
No. Asmy. Part No. Description
59 2 LD227A Bushing 60 2 LN141 Adapter 61 4 LD186 Collar 62 4 LD117 Nut, Adapter 63 1 LH227A Extension Pipe 64 1 LJ227 Lower Connection Pipe Assembly 65 1 LF227A Lower Connection Pipe 66 1 LG227A Back Connection Pipe 67 Low Pressure Piston Assembly
1LM869 psig 1 LM869D kg/cm
2
1 LM869E bar & Pascal
79 High Pressure Piston Assembly
1 LN869 psig 1 LN869D kg/cm
2
1 LN869E bar & Pascal 82 1 LF217 Disc Seat 83 1 LM141 Adapter
1 617C002-01 “O” Ring Kit, Oil
1 617A003-01 Overhaul Kit, Oil (see note 1)
1 LH244A Wrench (1 x 11⁄8)
1 LH244 Wrench (7⁄8 x 1)
1 LG244 Wrench (
5
8)
12 1 607A004-02 “O” Ring 15 4 BA122H “O” Ring 16 1 607A004-03 “O” Ring 19 2 607A004-01 “O” Ring 21 1 BA122E “O” Ring 27 6 BA122D “O” Ring 37 1 LAH292D Valve Assembly, Release 40 2 BA122G “O” Ring 41 1 LAG292D Valve Assembly, Displacement 45 1 BA122F “O” Ring
1 LGG236A Nameplate, Hydraulic 1 617C002-02 “O” Ring Kit, Hydraulic 1 617A003-02 Overhaul Kit, Hydraulic (see note 1)
47 1 LKK236 Oxygen Warning Label 58 1 LK227A Pipe Extension Assembly 59 3 LD227B Bushing 60 3 LN141A Adapter 61 6 LD186A Collar 62 6 LD117A Nut, Adapter 64 1 LJ227A Lower Connection, Pipe Assembly 83 1 LM141A Adapter
*Piston and bushing can only be replaced as a complete assembly, the
part number for this assembly is for: DWT1305DH – part no. 617B001-01 DWT1305DH – part no. 617B001-02
1 Overhaul kit includes all O-rings required to overhaul pump as
well as piston and bushing assembly.
Page 22
Page 23
WARRANTY/DISCLAIMER
OMEGA warrants this unit to be free of defects in materials and workmanship and to give satisfactory service for a period of
13 months from date of purchase. OMEGA’s Warranty adds an additional one (1)
month grace period to the normal
one (1) year product warranty to cover handling and shipping
time. This ensures that OMEGA’s customers receive maximum coverage on each product. If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. However, this WARRANTY is VOID, if the unit shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear or which are damaged by misuse are not warranted. These include contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. Nevertheless, OMEGA only warrants that the parts manufactured by it will be as specified and free of defects.
OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHAT­SOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA ENGINEERING Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
FOR
WARRANTY RETURNS, please have the
following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS OR CALIBRATION, consult OMEGA for current repair charges. Have the following information available BEFORE con­tacting OMEGA:
1. P. O. number to cover the COST of the repair/recalibration,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
MADE IN
Page 24
M0627/0495
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
TEMPERATURE

Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

Wire: Thermocouple, RTD & Thermistor

Calibrators & Ice Point References

Recorders, Controllers & Process Monitors

Infrared Pyrometers
PRESSURE, STRAIN AND FORCE

Transducers & Strain Gages

Load Cells & Pressure Gages

Displacement Transducers

Instrumentation & Accessories
FLOW/LEVEL

Rotameters, Gas Mass Flowmeters & Flow Computers

Air Velocity Indicators

Turbine/Paddlewheel Systems

Totalizers & Batch Controllers
pH/CONDUCTIVITY

pH Electrodes, Testers & Accessories

Benchtop/Laboratory Meters

Controllers, Calibrators, Simulators & Pumps

Industrial pH & Conductivity Equipment
DATA ACQUISITION

Data Acquisition & Engineering Software

Communications-Based Acquisition Systems

Plug-in Cards for Apple, IBM & Compatibles

Datalogging Systems

Recorders, Printers & Plotters
HEATERS

Heating Cable

Cartridge & Strip Heaters

Immersion & Band Heaters

Flexible Heaters

Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL

Metering & Control Instrumentation

Refractometers

Pumps & Tubing

Air, Soil & Water Monitors

Industrial Water & Wastewater Treatment

pH, Conductivity & Dissolved Oxygen Instruments
Loading...