Omega Products DP9602 User Manual

High Precision Digital Thermometer/Data Logger
For Thermocouple and RTD Sensors with USB Port
DP9602 Series
U Reference Standard Thermometer for
Pt100 and Thermocouple Sensors
U High Accuracy ±0.01ºC Pt100,
±0.18ºC Thermocouple
U ±0.025ºC System Accuracy with Pt100
(Based on DP9600-SYS-CAL)
U Alpha Numeric Display U User Selectable Resolution,
0.001/0.01ºC for Pt100 Inputs
U Data Logging U Measures Pt100 and Thermocouples
J, K, T, E, R, S, B and N
U Readout Directly in ºC, ºF, K, Ω or mV U Channel A, B or A-B Reading Including
Dissimilar Sensor Types
U USB Serial Communications U USB Flash Drive Port U Supplied with PC Software U Programmable Analog
Retransmission Output
U Rechargeable Lithium Ion Battery/
Mains Powered
The DP9602 provides precise platinum resistance thermometer temperature measurement and logging for exacting applications in both laboratory and site locations. It is an exceptionally accurate and stable
DP9602 shown smaller than actual size.
digital thermometer particularly suitable for use with thermal calibration baths as a reference standard.
Applications include medical, pharmaceutical, food, environmental testing, R&D and educational establishments.
Displayed values and user information are indicated on a bright clear LED screen with diffused backlighting. Data can be displayed in °C, °F, K, Ω as required; nine front panel push keys, the only user controls, are used in conjunction with the display screen. It is this arrangement which makes for very simple and “friendly” operation. There are four input ports, two for thermocouples and two for 3- or 4-wire Pt100 sensors; the instrument automatically recognises 3- or 4-wire configurations. The DP9602 can accept two Pt100 sensors and/or two thermocouples of similar or different types. The measured temperature can be displayed directly from one of the inputs or differentially between similar or different inputs. Differential temperature and the two individual channel temperatures are displayed simultaneously.
All four inputs can be scanned and values logged according to parameters set up by the user in the PC software; the instrument incorporates a real-time clock and on-board memory. A flash drive USB port is provided to allow data to be stored and/or exported. PC software running in Windows® is provided as standard; it allows programming of custom calibration, remote control and measure and logging functions.
K-1
K
OMEGACARESM extended warranty program is available
SM
for models shown on this page. Ask your sales representative for full details
Extended Warranty
Program
when placing an order. OMEGACARESM covers parts, labor and equivalent loaners.
Specifications
Inputs/Ranges/Sensors Type: Pt100 to IEC 751 (ITS 90 refers) -200 to 850ºC
(-328 to 1562ºF) Ro = 100Ω 3- or 4-wire connection with automatic recognition (with manual override) thermocouples to IEC 584 No isolation exists between inputs
Types: S: -50 to 1768ºC (-58 to 3214ºF) R: -42 to 1766ºC (-43.6 to 3211ºF) B: 254 to 1818ºC (489 to 3304ºF) J: -200 to 1200ºC (-328 to 2192ºF) K: -195 to 1370ºC (-319 to 2498ºF) E: -190 to 995ºC (-310 to 1823ºF) N: -160 to 1300ºC (-256 to 2372ºF) T: -193 to 399ºC (-315 to 750ºF)
Overall: Pt100 Accuracy: ±0.01ºC ±0.0005% of span Thermocouples J, K, T, E, N: ±0.2ºC from -50 to 700ºC
(-58 to 1292ºF), ±0.4ºC from -200 to -50ºC (-328 to -58ºF), ±0.4ºC from 700ºC (1292ºF) to full scale
Thermocouples R, S, B: ± 0.3ºC from 850ºC (1562ºF) to full scale, ±0.4ºC below 850ºC (1562ºF), ±0.7ºC below 0ºC (32ºF)
Note: Type B not characterized below 200ºC (392ºF).
Linearisation Conformity: Pt100: Better than ±0.01ºC
Thermocouples: ±0.05ºC base metal (-200ºC to span) ±0.1ºC noble metal above 200ºC
Note: Performance of ADC (0.5μV resolution ) is superior to published thermocouple table best resolution of 1μV.
Stability (vs Ambient Temperature): Pt100: Better than 0.0025ºC per 1ºC ambient change Thermocouples:
Zero: Negligible drift excluding CJC effect Scale: < ±0.1ºC in 15ºC
Warm-Up: Negligible under normal ambient conditions; allow 5 to 10 minutes for full stability unless stored at low temperature, then 30 minutes minimum
Cold Junction Compensation: Automatic on all thermocouple ranges; better than ±1ºC max shift for ±10ºC ambient variation (100:1) or external via Pt100 on channel A or B or manual input of CJ temperature value
Pt100 Sensor Current: 0.5 mA Resolution of Data Display: Pt100 0.01/0.001 user select
for ºC, K and Ω Variable Filter: Sampling rate selectable between
4 and 64 (measurements averaged per reading)
Thermocouples: 0.1ºC all ranges Measurement Units: ºC, ºF, K, Ω or mV as applicable
to sensor Measurement Modes: A, B or A-B; any combination of
sensor types Custom Calibration: Up to 10 calibration values can be
allocated to Pt100 input A & B and to thermocouples input A & B; values are retained in non-volatile memory until replaced by user; alternatively, for Pt100 inputs ITS 90 or IPTS 68 coefficients can be used for custom calibration
Smart Sensor Connection: Pt100 inputs only; correction values stored in connector
Null Function: Corrects differential temperature readout between two sensors to zero
Sensor Lead: Pt100: Resistance 5 Ω each lead maximum
Thermocouples: 300 Ω loop total Logging: 8000 readings Supply: Internal lithium Ion rechargeable batteries; mains
90 to 260V 50/60 Hz universal adaptor included; battery charge life up to 12 hours with full charge
Series Mode Rejection: 60 dB @ 50 Hz (50 mV RMS applied)
Common Mode Rejection: 30V RMS applied between input and earth produces no measurable effect
Display: 128 x 64 pixel OLED with diffused backlighting; wide viewing angle, high contrast (adjustable)
Front Panel Controls: 9 membrane push-keys to control all instrument functions
Mechanical/Case: Metal bench top case/adjustable tilt Dimensions Overall: 190 W x 70 H x 250 mm D
(7.5 x 2.75 x 9.85")
Weight: 3 kg (6.61 lb) approx Input Connections: 2 x Pt100 via D type connectors
2 x thermocouple via standard sockets USB Serial Communications: Isolated, 9600 Baud, 8 data,
no parity, 1 stop bit; remote control and measure PC Software (Standard): Supplied as standard on CD ROM;
remote control and measure; log readings to file/download to PC/programming corrections
Analog Output (Standard): User programmable, 0 to 5 Vdc Precision Pt100 Probes: 304 stainless steel probes,
6 mm Dia. with 2 m screened PTFE lead and Pt100 probes and connector
DP9600-L250: 250 mm (9.85") L, -50 to 250°C (-58 to 482°F) DP9600-H450: 350 mm (13.75") L, -50 to 450°C (-58 to 842°F)
To Order
Visit omega.com/dp9602 for Pricing and Details
Model No. Description DP9602 High precision digital RTD/thermocouple
thermometer/data logger with USB port
Accessories
Model No. Description
DP9600-L250 Precision Pt100 RTD probe,
-50 to 250ºC, 250 mm
DP9600-H450 Precision Pt100 RTD probe,
-50 to 450ºC, 350 mm
DP9600-INSTRUMENT-CAL
Calibration for instrument alone
(no NIST, UKAS calibration)
DP9600-SYS-CAL Calibration of instrument and
sensor together at five points (UKAS calibration)
DP9600-SYS-COR-CAL
Calibration of instrument and
sensor together at five points,
after initial calibration of sensor only and programming of corrections (UKAS calibration)
DP9600-TBLK4 Terminal block for connection
of 3- or 4-wire Pt100s
Comes complete with software, power adaptor, 9-pin connector and operator’s manual.
Ordering Example: DP9602, high precision digital RTD/ thermocouple thermometer, DP9600-SYS-CAL, UKAS system calibration, DP9600-H450 probe.
OCW-2, OMEGACARESM extends standard 1-year warranty to a total of 3 years.
K-2