High Precision Digital  
Thermometer/Data Logger 
For Thermocouple and RTD Sensors with USB Port
DP9602 Series
U Reference Standard Thermometer for 
Pt100 and Thermocouple Sensors
U High Accuracy ±0.01ºC Pt100, 
±0.18ºC Thermocouple
U ±0.025ºC System Accuracy with Pt100 
(Based on DP9600-SYS-CAL)
U Alpha Numeric Display 
U User Selectable Resolution, 
0.001/0.01ºC for Pt100 Inputs
U Data Logging 
U  Measures Pt100 and Thermocouples 
J, K, T, E, R, S, B and N
U Readout Directly in ºC, ºF, K, Ω or mV 
U  Channel A, B or A-B Reading Including 
Dissimilar Sensor Types
U USB Serial Communications 
U USB Flash Drive Port 
U Supplied with PC Software 
U Programmable Analog 
Retransmission Output
U Rechargeable Lithium Ion Battery/
Mains Powered
The DP9602 provides precise platinum resistance 
thermometer temperature measurement and logging 
for exacting applications in both laboratory and site 
locations. It is an exceptionally accurate and stable 
DP9602 shown smaller  
than actual size.
digital thermometer particularly suitable for use with 
thermal calibration baths as a reference standard.
The DP9602 is fully characterised for Pt100 sensors 
and all major thermocouples J, K, N, T, E, R, S and B. 
Exceptionally stable automatic cold junction  
compensation with a rejection of 100:1 is incorporated 
for thermocouple ranges. Alternatively, external or 
manual referencing can be selected as required.
Applications include medical, pharmaceutical, food,  
environmental testing, R&D and educational  
establishments.
Displayed values and user information are indicated  
on a bright clear LED screen with diffused backlighting. 
Data can be displayed in °C, °F, K, Ω as required;  
nine front panel push keys, the only user controls, are 
used in conjunction with the display screen. It is this  
arrangement which makes for very simple and  
“friendly” operation. There are four input ports, two for 
thermocouples and two for 3- or 4-wire Pt100 sensors; 
the instrument automatically recognises 3- or 4-wire  
configurations. The DP9602 can accept two Pt100  
sensors and/or two thermocouples of similar or  
different types. The measured temperature can be  
displayed directly from one of the inputs or  
differentially between similar or different inputs.  
Differential temperature and the two individual channel  
temperatures are displayed simultaneously.
All four inputs can be scanned and values logged  
according to parameters set up by the user in the  
PC software; the instrument incorporates a real-time 
clock and on-board memory. A flash drive USB port is 
provided to allow data to be stored and/or exported.  
PC software running in Windows® is provided as 
standard; it allows programming of custom calibration, 
remote control and measure and logging functions.
K-1
K

OMEGACARESM extended  
warranty program is available 
SM
for models shown on this 
page. Ask your sales  
representative for full details 
Extended Warranty
Program
when placing an order.  
OMEGACARESM covers parts, 
labor and equivalent loaners.
Specifications
Inputs/Ranges/Sensors 
Type: Pt100 to IEC 751 (ITS 90 refers) -200 to 850ºC 
(-328 to 1562ºF) Ro = 100Ω 3- or 4-wire connection  
with automatic recognition (with manual override)  
thermocouples to IEC 584 No isolation exists  
between inputs
Types: 
 S: -50 to 1768ºC (-58 to 3214ºF) 
 R: -42 to 1766ºC (-43.6 to 3211ºF)  
 B: 254 to 1818ºC (489 to 3304ºF) 
 J: -200 to 1200ºC (-328 to 2192ºF) 
 K: -195 to 1370ºC (-319 to 2498ºF) 
 E: -190 to 995ºC (-310 to 1823ºF) 
 N: -160 to 1300ºC (-256 to 2372ºF) 
 T: -193 to 399ºC (-315 to 750ºF)
Overall: Pt100 
Accuracy: ±0.01ºC ±0.0005% of span 
Thermocouples J, K, T, E, N: ±0.2ºC from -50 to 700ºC 
(-58 to 1292ºF), ±0.4ºC from -200 to -50ºC (-328 to -58ºF), 
±0.4ºC from 700ºC (1292ºF) to full scale
Thermocouples R, S, B: ± 0.3ºC from 850ºC (1562ºF) to  
full scale, ±0.4ºC below 850ºC (1562ºF), ±0.7ºC below  
0ºC (32ºF)
Note: Type B not characterized below 200ºC (392ºF).
Linearisation Conformity: 
 Pt100: Better than ±0.01ºC 
 Thermocouples: ±0.05ºC base metal (-200ºC to span)  
 ±0.1ºC noble metal above 200ºC
Note: Performance of ADC (0.5μV resolution ) is superior to  
published thermocouple table best resolution of 1μV.
Stability (vs Ambient Temperature): 
 Pt100: Better than 0.0025ºC per 1ºC ambient change 
Thermocouples: 
 Zero: Negligible drift excluding CJC effect 
 Scale: < ±0.1ºC in 15ºC
Warm-Up: Negligible under normal ambient conditions;  
allow 5 to 10 minutes for full stability unless stored at low 
temperature, then 30 minutes minimum
Cold Junction Compensation: Automatic on all  
thermocouple ranges; better than ±1ºC max shift for ±10ºC 
ambient variation (100:1) or external via Pt100 on channel  
A or B or manual input of CJ temperature value
Pt100 Sensor Current: 0.5 mA 
Resolution of Data Display: Pt100 0.01/0.001 user select 
for ºC, K and Ω 
Variable Filter: Sampling rate selectable between 
4 and 64 (measurements averaged per reading)
Thermocouples: 0.1ºC all ranges 
Measurement Units: ºC, ºF, K, Ω or mV as applicable 
to sensor 
Measurement Modes: A, B or A-B; any combination of 
sensor types 
Custom Calibration: Up to 10 calibration values can be 
allocated to Pt100 input A & B and to thermocouples input  
A & B; values are retained in non-volatile memory until  
replaced by user; alternatively, for Pt100 inputs ITS 90 or 
IPTS 68 coefficients can be used for custom calibration
Smart Sensor Connection: Pt100 inputs only; correction 
values stored in connector
Null Function: Corrects differential temperature readout  
between two sensors to zero
Sensor Lead: 
 Pt100: Resistance 5 Ω each lead maximum 
Thermocouples: 300 Ω loop total 
Logging: 8000 readings 
Supply: Internal lithium Ion rechargeable batteries; mains 
90 to 260V 50/60 Hz universal adaptor included; battery  
charge life up to 12 hours with full charge
Series Mode Rejection: 60 dB @ 50 Hz  
(50 mV RMS applied)
Common Mode Rejection: 30V RMS applied between  
input and earth produces no measurable effect
Display: 128 x 64 pixel OLED with diffused backlighting;  
wide viewing angle, high contrast (adjustable)
Front Panel Controls: 9 membrane push-keys to control  
all instrument functions
Mechanical/Case: Metal bench top case/adjustable tilt 
Dimensions Overall: 190 W x 70 H x 250 mm D 
(7.5 x 2.75 x 9.85")
Weight: 3 kg (6.61 lb) approx 
Input Connections: 2 x Pt100 via D type connectors 
2 x thermocouple via standard sockets 
USB Serial Communications: Isolated, 9600 Baud, 8 data, 
no parity, 1 stop bit; remote control and measure 
PC Software (Standard): Supplied as standard on CD ROM; 
remote control and measure; log readings to file/download to 
PC/programming corrections
Analog Output (Standard): User programmable, 0 to 5 Vdc 
Precision Pt100 Probes: 304 stainless steel probes, 
6 mm Dia. with 2 m screened PTFE lead and Pt100 probes  
and connector
DP9600-L250: 250 mm (9.85") L, -50 to 250°C (-58 to 482°F) 
DP9600-H450: 350 mm (13.75") L, -50 to 450°C (-58 to 842°F)
 To Order 
Visit omega.com/dp9602 for Pricing and Details
 Model No. Description 
 DP9602   High precision digital RTD/thermocouple 
thermometer/data logger with USB port
Accessories
 Model No.  Description
 DP9600-L250   Precision Pt100 RTD probe, 
-50 to 250ºC, 250 mm
 DP9600-H450   Precision Pt100 RTD probe, 
-50 to 450ºC, 350 mm 
DP9600-INSTRUMENT-CAL
 Calibration for instrument alone 
(no NIST, UKAS calibration)
 DP9600-SYS-CAL   Calibration of instrument and 
sensor together at five points  
(UKAS calibration)
DP9600-SYS-COR-CAL
 Calibration of instrument and 
   sensor together at five points, 
  after initial calibration of sensor  
  only and programming of  
  corrections (UKAS calibration)
 DP9600-TBLK4   Terminal block for connection 
of 3- or 4-wire Pt100s
Comes complete with software, power adaptor, 9-pin connector and 
operator’s manual. 
Ordering Example: DP9602, high precision digital RTD/ 
thermocouple thermometer, DP9600-SYS-CAL, UKAS system 
calibration, DP9600-H450 probe. 
OCW-2, OMEGACARESM extends standard 1-year warranty to a  
total of 3 years.
K-2