5-DIGIT PRECISION STRAIN GAGE
INDICATORS WITH EXCITATION SUPPLY
Vi
si
t
o
m
e
g
a
.
c
o
m
f
t
s a
o
r
e
se
l
e
c
t
i
o
n
so
r
s a
n
d
va
i
l
a
b
l
e
E
GA
DP80 Series
DP87, $729, shown smaller than actual size, with
LC101, load cell, $305, page F-51, and PX610,
pressure transducer, $515, page B-44. Transducers
sold separately.
t
h
e
c
o
m
p
l
o
t
e
f
p
r
o
c
e
ss se
r
a
n
sd
u
c
e
r
f
r
o
m
OM
n
Starts at
$
729
⻬ High Accuracy: 0.01% FS
⻬ Precise Digital Scaling
on Front-Panel Keypad
⻬ 3-Character
Alphanumeric Display
of Engineering Units
⻬ Plug-In Options Include
Alarms, Math, Analog
Outputs, RS232C/20 mA,
and DC Power
⻬ Up to 1 µV/Count
Sensitivity
⻬ Lockout Switch Prevents
Unauthorized Program
Tampering
⻬ Power Cord Included
The OMEGA®DP80 Series digital
strain gage indicators demonstrate
impressive compatibility with
lower-output, bonded-foil-type
transducers, as well as with
higher-output semiconductor-type
transducers (up to 1000 mV). The
user can select 1 of 4 input ranges.
The front-panel membrane keypad
facilitates accurate digital scaling,
eliminating any DIP switches,
pot adjustments, or reference
standards. This allows for drift-free
scaling to precise engineering units
and decimal location. The display
features 3 alphanumeric LED
digits for engineering units
(e.g., psi, kg, lb), which can be
assigned as dead zeros.
D-75
DP80 Series meters are
available in single and dual
width, and benchtop
configurations.
The DP80 also has primary and
secondary engineering units
displays, both of which can be
selected on the front panel. If alarm
points are set up in pounds (lb) on
the primary display, the instrument
will convert the alarm point to
kilograms (kg) if the secondary
display is active. The system can be
calibrated by using 3-point “live”
load (actual weight on scale) or by
entering the appropriate transducer
sensitivity (mV/V). Programmable
functions include voltmeter
conversion; auto-zero; and rounding
of digits by 1, 2, 5, 10, or 100
counts. Optional features include
alarms, math, analog outputs,
RS232C/20 mA, and DC power.
ACCURATELY MEASURE PRESSURE, STRAIN, FORCE,
Extended Warranty
P
rogram
SM
LOAD, THRUST, AND TORQUE
Zero Offset: Entire input range
OMEGACARESMextended warranty program is
available for models shown on this page. Ask
your sales respresentative for full details when
placing an order. OMEGACARESMcovers parts,
labor and equivalent loaners.
SPECIFICATIONS
Excitation Supplies:
DP87: 5, 10 and 15V; switch
selectable, isolated; drives 2 parallel
350 Ω strain gages @ 15V (90 mA)
DP88: 5, 10, 15 and 20V; switch
selectable, isolated; drives
8 parallel 350 Ω strain gages
@ 15V (350 mA)
Accuracy: 0.01% FS, ±1 digit
Repeatability: ±1 digit
Zero Stability: 0.5 µV/°C;
no measurable drift with time
Span Stability: 0.005% of reading/°C;
0.1% of reading per year
Normal-Mode Noise Rejection:
>60 dB @ 50/60 Hz
ommon-Mode Noise Rejection:
C
>120 dB @ 50/60 Hz input impedance;
>10 MΩ to 500 MΩ depending on range
Display: 8-digit, 14-segment,
alphanumeric LEDs; 14 mm (0.54")
high; one negative (-) LED digit at left
Display Count Rate: ±100,000 active
Reading Rate: 2 readings per second
Overload Protection: 270 Vdc/Vac
RMS across inputs and input to
ground; power lead to ground
1500 Vdc/Vac RMS
Ambient Operating Conditions:
0 to 50°C (32 to 122°F); 0 to 90% RH,
non-condensing
Storage Temperature: -40 to 65°C
(-40 to 149°F)
Input Connections: AC power, quickconnect; DC power and sensors, screw
terminals; RS232C, DB25 connector
Case Construction: Metal, black
anodized, extruded aluminum
Power: 90 to 132 Vac; 48 to 400 Hz
Weight: DP87, 1.8 kg (4 lb);
DP88/DP89, 2.7 kg (6 lb)
without options
Dimensions:
DP87: 67 H x 136 W x 250 mm D
(2.64 x 5.35 x 9.84"), cutout
68 H x 138 mm W (2.67 x 5.43"),
bezel
72 x 142.5 mm (2.83 x 5.61")
DP88/DP89:
67 H x 272 W x 250 mm D
(2.6 x 10.71 x 9.84"), cutout
68 H x 274 mm W (2.67 x 10.79"),
bezel
72 x 284 mm (2.83 x 11.18")
AVAILABLE FOR FAST DELIVERY!
To Order (Specify Model Number)
SELECTABLE FULL SCALE DISPLAY
MODEL NO. DESCRIPTION PRICE INPUT RANGE MIN GAIN MAX GAIN
DP87
*
DP88
*
DP89
*
Comes complete with power cord and operator’s manual.
Single width $729 -15 to 100 mV 0.01 cts/mV 1000 cts/mV
Double width panel mount 899
Double width benchtop 899
-30 to 200 mV 0.05 cts/mV 500 cts/mV
-75 to 500 mV 0.02 cts/mV 200 cts/mV
-150 to 1000 mV 0.01 cts/mV 100 cts/mV
* DP87 can accommodate 1 additional option. DP88 and DP89 can accommodate up to 4 additional options.
OPTIONS
To Order (Specify Model Number)
MODEL NO. PRICE DESCRIPTION
DP80-ALM $169 Alarms, dual limits (up to 2 per unit), selectable hysteresis and alarm delay
DP80-AOV 189 Analog output, 0 to 10 Vdc isolated, linearized and scalable (N/A DP80-AOC)
DP80-AOC 189 Analog output, 4 to 20 mA isolated, linearized and scalable (N/A DP80-AOV)
DP80-MTH 99 Math, min, max, average, rate of change
DP80-SER** 189 RS232C/20 mA serial output
Selectable baud rate from 300 to 9600. Includes DB25 mating male connector.
**
OMEGACARE
placing an order.
Ordering Example: DP89, double-width benchtop strain gage indicator, DP80-ALM, alarm capability, DP80-SER, RS232C communications
output, $899 + 169 + 189 = $1257. OCW-1, OMEGACARE
SM
extended warranty program is available for models shown on this page. Ask your sales representative for full details when
SM
extends standard 1-year warranty to a total of 2 years ($125), $1257 + 125 = $1382.
INSTRUMENTATION
D
Recommended Reference Book: Handbook of Systems Engineering,EE-2530, $255
See Section Y For Additional Books
D-76