Omega Products DFG60-1 Installation Manual

www.omega.com
e-mail: info@omega.com
User’s Guide
DFG60 SERIES
Digital Force Gauge
Servicing North America:
USA: One Omega Drive, Box 4047
ISO 9001 Certified Stamford, CT 06907-0047
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada@omega.com
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378 / 1-800-622-BEST
®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
®
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico: TEL: (001) 800-826-6342 FAX: (001) 203-359-7807
En Espan˜ ol: (001) 203-359-7803 e-mail: espanol@omega.com
info@omega.com.mx
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
TEL: +31 (0)20 6418405 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: nl@omega.com
Czech Republic: Rudé armády 1868, 733 01 Karviná 8
TEL: +420 (0)69 6311899 FAX: +420 (0)69 6311114 Toll Free in Czech Republic: 0800-1-66342 e-mail: czech@omega.com
France: 9, rue Denis Papin, 78190 Trappes
TEL: +33 (0)130 621 400 FAX: +33 (0)130 699 120 Toll Free in France: 0800-4-06342 e-mail: france@omega.com
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 (0)7056 3017 FAX: +49 (0)7056 8540 Toll Free in Germany: 0800 TC-OMEGA
SM
e-mail: germany@omega.com
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5EX, England TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in England: 0800 488 488 e-mail: sales@omega.co.uk
OMEGAnetSMOn-Line Service Internet e-mail
http://www.omega.com info@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
GENERAL OPERATION
1 Press . The capacity of the gauge is displayed and the gauge will automatically enter into the
measuring mode. Press and hold for at least four (4) seconds each time you wish to select between pounds (ounces), kilograms (grams) and Newtons.
2 Hand tighten (no tool!) selected attachment to the measuring shaft. 3 When the gauge is turned on, it will go directly to its real time measuring mode. Press to mea-
sure peak forces.“Peak icon” on the display indicates peak measuring mode, whereas the peak read­ing will not change until a higher value is measured.To delete the last peak reading,press .To end peak reading mode and go to real time measuring mode, press again.
real time mode display peak mode display
4 If necessary,press to tare the weight of the attachment and shaft orientation. Pressing will
also clear the peak reading.
5 Make sure to apply tension and compression (-) forces to the gauge in line
with the measuring shaft. DO NOT attempt to measure forces at an angle to the measuring shaft - damage to load cell and/or shaft may result.
PRECAUTIONS
1 WARNING!! REGARDLESS of whether the unit is ON or OFF, do not exceed the capacity of the
gauge.At 105% of the rated capacity, the display will flash. Never exceed 200% of the rated capacity, or the load cell will be damaged.
2 When mounting, use M4 mounting screws with a maximum insertion depth of 5 mm into the gauge. 3 Measure in line tension and compression forces only. Do not attempt to measure forces at an angle
to the measuring shaft - damage to load cell and/or shaft may result.
4 Hand tighten attachments only. Do not use tools. 5 Make sure this gauge and all peripherals are powered down before attaching any cables. 6 Do not remove the warranty seal or disassemble the gauge.Disassembly voids warranty.
RECHARGING NI-CAD BATTERY
1 To maximize the life of the battery, power will automatically shut off after 10 minutes of non-use.
This automatic shut off feature can be bypassed and the gauge may be used continuously when the AC adapter/charger is used.
2 “BATT”icon will flash when gauge needs to be recharged.To maximize battery
life, do not recharge until “BATT” icon flashes.With proper recharging, battery can be recharged 500 times.
3 Push to turn off power.
IMPORTANT! Use the provided adapter/charger exclusively and plug into the correct AC output. It
takes 8 hours to charge fully.
4 When the gauge is turned off, make sure the AC adapter/charger is disconnected to avoid overcharging.