Thermal Response Time:
SD Model: Typical <10 ms @
4.2 K, 100 ms @ 77 K,
200 ms @ 305 K
BR Model: 1 ms @ 4.2 K,
13 ms @ 77 K, 20 ms @ 305 K
Use in Radiation: Recommended for
use only in low level radiation
Use in Magnetic Field: Not
recommended for use in magnetic field
applications below 60 K; low magnetic
field dependence when used in fields
up to 5 tesla above 60 K
Reproducibility(*): ±10 mK @ 4.2 K
(*) Short-term reproducibility data is obtained
by subjecting sensor to repeated thermal
shocks from 305 to 4.2 K.
Dimensions: mm (in)
Range Limit Limit
of Use Min Max
CY670-SD 1.4 K 500 K
CY670E-BR 1.4 K 500 K
Typical Long Term
Temperature Accuracy Accuracy (*)
1.4 K ±12 mK —
4.2 K ±12 mK 10 mK
10 K ±12 mK —
77 K ±22 mK 40 mK
300 K ±32 mK 25 mK
500 K ±50 mK —
Temperature Tolerance at Temperature Range
Band 2 to 30 K 30 to 100 K 100 to 305 K 305 to 500 K
A (1) ±0.25 K ±0.25 K ±0.50 K ±0.50 K
B (2) ±0.50 K ±0.50 K ±0.50 K ±0.33% of T
(1.01 to 1.65 K)
C (3) ±1.0 K ±1.0 K ±1.0 K ±0.5% of T
(1.53 to 2.50 K)
D (4) ±1.5 K ±0.25 K ±0.30 K ±0.1% of T
(PRT Band) (0.305 to 0.500 K)
E ±1.0 K ±0.25 K ±0.25% of T ±0.25% of T
(Bare Chip Band) (0.25 to 0.76 K) (0.76 to 1.25 K)
Calibrated Accuracy
Temperature Volts dV/dT (mV/K)
1.4 K 1.64 -12.5
4.2 K 1.58 - 31.6
10 K 1.38 -26.8
77 K 1.03 -1.73
305 K 0.560 -2.30
Temperature Response Data Table
(Typical for CY670)
Tolerance Bands for the CY670 Series Diode Thermometer
(*) Long term stability data is obtained by subjecting sensor
to 200 thermal shocks from 305 to 77 K.