Omega Products CL-300 Installation Manual

Page 1
TEMPERATURE
MU
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
MU
Wire: Thermocouple, RTD & Thermistor
MU
MU
Recorders, Controllers & Process Monitors
MU
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
M
U
Transducers & Strain Gauges
MU
Load Cells & Pressure Gauges
MU
Displacement Transducers
MU
Instrumentation & Accessories
FLOW/LEVEL
MU
Rotameters, Gas Mass Flowmeters & Flow Computers
MU
Air Velocity Indicators
MU
Turbine/Paddlewheel Systems
MU
Totalizers & Batch Controllers
pH/CONDUCTIVITY
M
U
pH Electrodes, Testers & Accessories
MU
Benchtop/Laboratory Meters
MU
Controllers, Calibrators, Simulators & Pumps
MU
Industrial pH & Conductivity Equipment
DATA ACQUISITION
MU
Data Acquisition & Engineering Software
MU
Communications-Based Acquisition Systems
MU
Plug-in Cards for Apple, IBM & Compatibles
MU
Datalogging Systems
MU
Recorders, Printers & Plotters
HEATERS
MU
Heating Cable
MU
Cartridge & Strip Heaters
MU
Immersion & Band Heaters
MU
Flexible Heaters
MU
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
MU
Metering & Control Instrumentation
MU
Refractometers
MU
Pumps & Tubing
MU
Air, Soil & Water Monitors
MU
Industrial Water & Wastewater Treatment
MU
pH, Conductivity & Dissolved Oxygen Instruments
Where Do I Find Everything I Need for Process
Measurement and Control? OMEGA…Of Course!
M0144/0297
Page 2
CL-300, CL-301, CL-302, CL-303
Portable Simulators
http://www.omega.com
e-mail: info@omega.com
User’s Guide
Page 3
OMEGAnetSMOn-Line Service Internet e-mail
http://www.omega.com info@omega.com
Benelux:
Postbus 8034, 1180 LA Amstelveen, The Netherlands Tel: (31) 20 6418405 FAX: (31) 20 6434643 Toll Free in Benelux: 06 0993344 e-mail: nl@omega.com
Czech Republic:
Ostravska 767, 733 01 Karvina Tel: 420 (69) 6311627 FAX: 420 (69) 6311114 e-mail: czech@omega.com
France:
9, rue Denis Papin, 78190 Trappes Tel: (33) 130-621-400 FAX: (33) 130-699-120 Toll Free in France: 0800-4-06342
Servicing Europe:
USA and Canada:
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
SM
Customer Service: 1-800-622-2378 / 1-800-622-BEST
SM
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
SM
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
USA: ISO 9001 Cer
tified
One Omega Drive, Box 4047 Stamford, CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Servicing North America:
For immediate technical or application assistance:
Mexico and Latin America:
Tel: (95) 800-TC-OMEGA
SM
FAX: (95) 203-359-7807 En Espan~ol: (203) 359-1660 ext: 2203 e-mail: espanol@omega.com
Germany/Austria:
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 Toll Free in Germany: 0130 11 21 66 e-mail: germany@omega.com
United Kingdom: ISO 9002 Cer
tified
•25 Swannington Road, Broughton Astley, Leicestershire, LE9 6TU, England Tel: 44 (1455) 285520 FAX: 44 (1455) 283912
•P.O. Box 7, Omega Drive, Irlam, Manchester, M44 5EX, England Tel: 44 (161) 777-6611 FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
Canada:
976 Bergar Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada@omega.com
Page 4
WARRANTY
OMEGA warrants this unit to be free of defects in materials and workmanship and to give satisfactory service for a period of 37 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal three (3) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product. If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. However, this WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear or which are damaged by misuse are not warranted. This includes contact points, fuses, and triacs.
OMEGA is glad to offer suggestions on the use of its various products. Nevertheless, OMEGA only warrants that the parts manufactured by it will be as specified and free of defects.
OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTA TIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclu­sive and the total liability of OMEGA with respect to this order, whether based on contract warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
Every precaution for accuracy has been taken in the preparation of this manual; however, OMEGA ENGINEERING, INC. neither assumes responsibility for any omissions or errors that may appear nor assumes liability for any damages that result from the use of the products in accordance with the information contained in the manual.
SPECIAL CONDITION: Should this equipment be used in or with any nuclear installation or activity, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the equipment in such a manner.
Page 5
Notes
Page 6
OMEGA’s policy is to make running changes, not model changes, whenever an improve­ment is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC
.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR W
ARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. P.O. number under which the product
was PURCHASED,
2. Model and serial number of the product
under warranty, and
3. Repair instructions and/or specific prob-
lems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST of the
repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific prob-
lems relative to the product.
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
Page 7
Notes
Page 8
Notes
Page 9
42
Schematic–OMEGA Model CL-303
Page 10
Table of Contents Section Page
Model CL-300 Portable
Thermocouple Source..........................................11
Model CL-301 Portable RTD Simulator.................12
Model CL-302 Portable
Transmitter Simulator..........................................21
Model CL-303 Portable
mA Source/2-Wire Simulator..............................30
i
Page 11
Table of Contents Model CL-300 Portable Thermocouple Source Section Page
Section 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . .1
1.1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Section 2 Installation . . . . . . . . . . . . . . . . . . . . . . . . .3
2.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Section 3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . .4
3.1 Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Section 4 Service Information . . . . . . . . . . . . . . . . . . .5
4.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4.2 Equipment Required for Calibration . . . . . . . . . . . . . . . . . . 5
4.3 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Section 5 Specifications . . . . . . . . . . . . . . . . . . . . . . .9
ii
Page 12
Section 5 Specifications CL-303
Internal Batteries: Three 9 Volt Alkaline, MN1604 or
equivalent
Battery Life: Source; 25-hours at 20 mAcontinuous
output, 2-wire Simulator infinite
Battery Status Indicator: LED pulse at turn-on, in source mode Calibrated Accuracy: ±0.1% Resolution: 0.02% Temperature Effect: 0.01%/˚C Output Drive Capability: 1200 ohms with fresh batteries, or
with optional ac adaptor; 800 ohms at battery low limit
Power Supply Effect: 0.005%/Volt Operating Ambient
Temperature: -10˚ to +140˚ F (-25 to +60˚C) Storage Temperature Limits: -40˚ to +160˚F (-40˚ to +70˚C). Remove
batteries if stored for extended period.
Warm-up Time: 2-seconds to rated accuracy Dimensions: H: 4" (101 mm) x W: 2
1
/8" (54 mm)
x D: 2 1/4" (57 mm)
Weight: 11 oz. (0.3 kg)
41
Page 13
40
Figure 4-3. 2-Wire Simulator Hookup
Page 14
Section 1 Introduction
1.1 Description
The OMEGA®Model CL-300 Thermocouple Source provides 22 precise temperature equivalent millivolt signals to transmitters, recorders, controllers, alarms, data acquisition, and computer systems. Model CL-300 is cold-junction compensated for ambient temperature variations and provides a thermocouple connector output. Cables are available in the desired length and terminal types to exactly duplicate the thermocouples in the system being tested. Conformity to the particular thermocouple vs millivolt curve is in accordance with the latest ASTM and IPTS standards for exact temperature simulation. Linear millivolt models are also available.
Thermocouple types E, J, K, and T are available in both ˚F and ˚C. T able 1-1 lists the standard ranges. Resolution is 25˚, 50˚, or 100˚ corresponding to full scale output of 500˚, 1000˚, and 2100˚ respectively.
Dual ranges, with an individual “ON” position for each range, allow quick, easy settings for any output. Calibrated accuracy is ±0.1% of span ±1 degree. For negative temperatur es, add ±2 degrees.
T wo internal AAcells provide power for approximately one year of everyday use. Afront panel LED pulses every time the unit is turned on to indicate proper battery voltage.
1
Page 15
Table 1-1
Model CL-300 Standard Ranges
2
Page 16
39
Figure 4-2. Adjustment of Span Potentiometer , Model CL-303
Page 17
38
Figure 4-1. Measuring mA Output
Page 18
3
1.2 Features
• Calibrate and T roubleshoot T emperature Systems
• Simulates Thermocouples J, K, E, T, R, S, or mV Sour ce
• Choice of Cable Length and T ype
• T wenty-two Precise T est Points
• Portable, Long Battery Life
• Simple to Use, Inexpensive to Own
Section 2 Installation
2.1 Unpacking
Remove the Packing List to check off actual equipment received. If you have any questions about the shipment, please call OMEGA’s Customer Service Department.
Upon receipt of shipment, inspect the container for any signs of damage in transit. T ake particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
NOTE
The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event shipment is necessary.
Page 19
Section 3 Operation
3.1 Operating Instructions
Select the Model CL-300 for the correct thermocouple type, the desired temperature scale (F or C), and range, the ordered lead type(s).
Connect the Thermocouple Source to the thermocouple input terminals of the instrument or system to be tested.
Cold-junction compensation and instantaneous automatic standardization and built into the instrument.
Set the temperature selection switch to the desired temperature and slide the “ON” switch left or right to the selected scale. The battery check LED will pulse once indicating proper battery voltage. If no pulse is seen, the batteries should be replaced with two AAcells. Alkaline cells are preferred for longest life and widest operating temperature range.
End point calibration temperatures of the instrument are then selected on the Model CL-300 and any required adjustments are made. Intermediate points may then be selected to verify instrument linearity or check critical points.
4
Page 20
37
2. The span potentiometer (Figure 4-2) should now be adjusted so that the meter reads 5,000 volts.
3. The right-hand switch is now moved to the 00.0% position.
4. The zero potentiometer should now be adjusted so that the meter reads 1.000 volts.
5. Check and readjust the span and zero controls until no further adjustments are necessary.
6 Move the right-hand switch to the Dial position. The meter
should read between 1.996 V and 2.004 V when the dial is set at
25.0 (Note: decimal points are not shown on the dial). Other readings are 2.996 V to 3.004 V at a setting of 50.0% and 3.98 V to
4.004 V at a setting of 75.0%. The dial has been mechanically set at the factory and should need no adjustment under normal operating conditions.
4.3.2.1 2-Wire Simulator Mode Check
To check the 2-wire Simulator of the CL-303:
1. Reconnect the test set-up to that shown in Figure 4-3.
2. Move the left-hand switch to the “2-wire” position.
3. Place a 24-30 V dc power source in series with the Model CL-303 and the 250 ohm resistor . The 100%, 00.0% and dial readings should all be within ±0.1% F .S. ±1˚ of those found in the source mode.
Page 21
36
4.3 Calibration Procedure
Both the mA output and the span and zero adjustments must be calibrated, as specified below .
4.3.1 mA Output, Source Mode Calibration
Proceed as follows:
1. Before any adjustments to the Model CL-303 are made, fresh alkaline batteries MN1604, should be placed in the unit.
2. Switch the left-hand switch to position mA OUT while observing the LED. The LED should flash once, indicating proper battery voltage. If the LED does not flash, check each battery for proper voltage and for correct connection.
3. Set the meter to a range which covers one to five V dc with sufficient resolution.
4. Place the 250 ohm (±0.05%) resistor in series with the Model CL-303 output. The meter will read 1 to 5 volts corresponding to 4-20 mA through the resistor. See Figure 4-1.
4.3.2 Span and Zero Adjustments
Make the necessary calibration adjustments as indicated below.
1. Set the left-hand switch to position mA OUT and the right-hand switch to position 100%.
Page 22
Section 4 Service Information
4.1 Maintenance
The OMEGA CL-300 Thermocouple Source requires little servicing. Should measurement inaccuracies be discovered, perform the adjustment and calibration procedure to recalibrate the instrument.
4.2 Equipment Required for Calibration
Digital thermometer, ±0.5˚ accuracy with 0.1˚ r esolution or equivalent. Digital voltmeter, 0.05% accuracy, 10-microvolt resolution or
equivalent. The recommended technique for calibrating a Model CL-300 requires
two meters. The digital thermometer is used to calibrate the output. The digital voltmeter is needed to calibrate the built-in cold-junction compensation.
4.3 Calibration Procedure
It is suggested that each time a unit is recalibrated, fresh AAbatteries (MN-1500-2) should be substituted for used batteries.
1 Switch the unit to each “ON” positions, making sure that the
battery check LED flashes.
2 Connect the digital thermometer, using the pr oper type thermo-
couple wire, directly to the exposed ends of the Model CL-300’s output wires.
5
Page 23
3. Connect the DVM to calibrate the cold-junction compensation as follows: The meter’s positive input is connected to the negative output of the Model CL-300 and the negative lead of the meter is connected to the outermost lead. Resistor R17 at the opposite end of the board. See Figure 4-1.
4. Before the cold-junction compensation can be adjusted, the ambient temperature of the Model CL-300 must be obtained. For highest accuracy, this is done by taking a reading on the meter that is already connected to the thermocouple output leads. With the range/power switch in the center “OFF” position, the meter will read the temperature at the internal Model CL-300 terminals.
5. To set the cold-junction compensation, move the range/power switch to the right-hand “ON” position, and the temperature selector to the highest output position.
6. The compensation potentiometer is then adjusted to the value shown in T able 4-1. The span (highest mV output) of the unit is now adjusted by means of the span potentiometer for the highest temperature reading of the unit being calibrated. See figure 4-1.
7. The cold-junction compensation should again be checked, and if readjusted, the span should be rechecked.
8. There are two separate zero (lowest mV output) adjustments. With the range/power switch still in the right-hand “ON” position, move the temperature selector switch to the lowest output position. The right-hand zero is then adjusted by means of the right-hand zero potentiometer. See Figur e 4-1.
6
Page 24
Section 4 Service Information
4.1 Maintenance
The CL-303 mA Source/2-wire Simulator requires little servicing. Should measurement inaccuracies be discovered, perform the adjustment and calibration procedures to recalibrate the instrument.
4.2 Equipment Required for Calibration
Digital voltmeter, 4 1/2" digit, 0.05% or better accuracy, with a precision (±0.5%) 250 ohm resistor . As a less accurate alternative, a dc millimeter may be used directly–which eliminates the 250 ohm resistor.
35
Figure 3-2. As a Milliamp Source
Page 25
3.1.2 Directions for Source Mode
Disconnect any existing input wires from the device to be checked or calibrated. T urn the mode selector switch (left-side) to the mAout position. The LED will pulse brightly once if the battery voltage is sufficient. If the LED does not pulse, turn it off and check the batteries.
Connect the red positive (+) lead to the positive input of the device to be checked or calibrated. Connect the black negative (-) lead to the negative terminal. See Figure 3-2. The LED will glow steadily when current is flowing. If no glow appears, check the hookup and connections.
The 00.0% (4.00 mA) and 100% (20.00 mA) Dial adjustments are selected with the right-hand switch. Fast zero and full scale checks can be made by means of the 00.0% and the 100% positions. Athird quick check position may be established by adjusting the digital dial to the desired percent value and locking it in place. The continuous adjustment dial reads directly in percent of the 4-20 mAsignal. For example, a dial setting of 500 is 50% of span or 12mA. Complete the adjustments and turn the instrument off. Reconnect any wires that were removed for the check out.
34
Figure 3-1. As a 2-Wire Simulator with Self Powered Unit
Page 26
19. To calibrate the left-hand zero, select the left-hand “ON” position.
For the highest accuracy, the left-hand zero should be calibrated at the second lowest setting (+25˚ on 500˚ units; +50˚ on 1000˚ units, +200˚ on 2100˚ units). The left-hand zero is then adjusted by means of the left-hand zero potentiometer . (See Figure 4-1).
10. Recheck the span at the highest output. When the span and both
zeros have been calibrated, check the other output positions to assure that the Model CL-300 is operating properly.
The allowable variation in 2100˚ and 3100˚ units is ±3˚. For 1000˚ units, it is ±2˚. For the 500˚ units, the allowable tolerance is ±1.5˚. Add 1˚ to tolerances for negative temperatures.
Table 4-1
Cold Junction Compensation Set Points
Degrees C
20 21 22 23 24 25 26 27 28 29 30
T ype E
18.151
18.211
18.272
18.333
18.394
18.455
18.516
18.577
18.638
18.699
18.760
T ype J
15.409
15.460
15.512
15.564
15.615
15.667
15.719
15.770
15.822
15.874
15.926
T ype K
12.071
12.111
12.152
12.192
12.233
12.273
12.314
12.354
12.395
12.435
12.476
T ype T
12.130
12.170
12.211
12.252
12.292
12.333
12.374
12.414
12.455
12.496
12.537
T ype R& S
1.788
1.794
1.800
1.806
1.812
1.818
1.824
1.830
1.836
1.842
1.848
7
Page 27
8
Figure 4-1. Calibration Hook-Up, Model CL-300
Page 28
Section 3 Operation
3.1 Operating Instructions
Outlined below is the operating protocol for the two modes of operation.
3.1.1 Directions for 2-Wire Simulator Mode
Disconnect any existing 2-wire transmitter from the loop to be checked or calibrated. (Only one of the two wires need be disconnected). Leave the power source and the receiver in place.
Place the mode selector switch into the 2-wire position. Connect the red clip lead of the Model CL-303 to the positive (+) terminal of the field connections. Connect the black lead to the negative (-) terminal. See Figure 3-1.
The LED will glow steadily when current flows. If no glow appears, recheck all external connections. Internal batteries are not used in this mode.
The 00.0% (4.00 mA) and 100% (20.00 mA) Dial adjustments are selected with the right-hand switch. Fast zero and full scale checks can be made by means of the 00.0% and the 100% positions. Athird quick check position may be established by adjusting the digital dial to the desired percent value and locking it in place. The continuous adjustment dial reads directly in percent of the 4-20 mAsignal. For example, a dial setting of 750 is 75% of span or 16mA. Complete the adjustments and turn the instrument off.
33
Page 29
32
1.2 Features
• Calibrate or T est 4-20 mAdc Systems
• Precise 4-20 mAdc Source
• Simulates Loop Powered Transmitter
• Continuous 0 to 100% settings
• Quick Check 0% to 100% switch
Section 2 Installation
2.1 Unpacking
Remove the Packing List to check off actual equipment received. If you have any questions about the shipment, please call OMEGA’s Customer Service Department.
Upon receipt of shipment, inspect the container for any signs of damage in transit. T ake particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
NOTE
The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event shipment is necessary.
Page 30
Section 5 Specifications CL-300
Batteries: Two “AA” alkaline cells provide
approximately one year of use
Battery Indicator: LED light pulse at turn-on in either
range
Accuracy: ±0.1% of span ±1˚. For negative
temperatures, add ±2˚.
Cold-junction Compensation: Built-in for specified thermocouple
type
Cold-junction Temperature Effect: Within ±0.25 degrees at 75˚F (20˚C)
±0.025 degrees/degree change in ambient
Operating Ambient Temperature: -10˚ to +130˚F (-25˚ to +55˚C)
Ambient Temperature Effect: Zero included in cold-junction effect.
Span: ±0.01% of span/degree
Storage Temperature Limits: -40˚ to +160˚F (-40˚ to +70˚C) Output Impedance: Fixed, 50 ohms nominal Dimensions: H: 4" (101 mm) x W: 2
1
/8" (54 mm)
x D: 2 1/4" (57 mm)
Weight: 6 oz. (0.15 kg)
9
Page 31
10
Schematic–OMEGA Model CL-300
Page 32
Table 1-1
CL-303 Dial Settings and Output
Dial Setting Output Current
% Milliamps
00.0 4.00
05.0 4.80
10.0 5.60
15.0 6.40
20.0 7.20
25.0 8.00
30.0 8.80
35.0 9.60
40.0 10.4
45.0 11.2
50.0 12.0
55.0 12.8
60.0 13.6
65.0 14.4
70.0 15.2
75.0 16.0
80.0 16.8
85.0 17.6
90.0 18.4
95.0 19.2
100.0 20.0
NOTE: Decimal point not shown on instrument’s dial.
31
Page 33
Section 1 Introduction
1.1 Description
The dual function OMEGA®CL303 Simulator combines a self­contained 4-20 mA source with a 2-wire Simulator in one pocket­sized instrument.
Internal batteries are used in the source mode to provide continuously adjustable 4-20 mA into any load from 0-800 ohms. An optional ac adaptor allows full-time bench use.
The 2-wire Simulator mode utilizes external loop power to pass precisely adjustable 4-20 mA. Loop power may vary from 6 to 45 V dc.
In addition to the 0.1% accurate digital dial adjustments, switch selected signals are provided at 4.00 mA(00.0%) and 20.00 mA(100%) for “quick check” zero and full scale settings in both modes (refer to T able 1-1).
Small size is made possible through state-of-the-art design which includes a precision reference, an ultra-stable amplifier, and high-gain power Darlington output. Accuracy, stability and low power consumption are combined through he use of the latest microcircuitry.
Continuous adjustability with memory lock assures fast, precise settings of current trips, recorders, data loggers, controllers, computers, and final control elements. Long life batteries allow complete portability for checkout and calibration of all milliamp input devices.
30
Page 34
Table of Contents Model CL-301 Portable RTD Simulator Section Page
Section 1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . 12
1.1 Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
1.2 Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Section 2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Section 3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1 Operating Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Section 4 Service Information. . . . . . . . . . . . . . . . . . . 16
4.1 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2 Equipment Required for Calibration. . . . . . . . . . . . . . . . . . 16
4.3 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Section 5 Specifications . . . . . . . . . . . . . . . . . . . . . . . 19
11
Page 35
Section 1 Introduction
1.1 Description
The OMEGA®Model CL-301 Platinum RTD Simulator provides eleven precise temperatures for inputs to transmitters, recorders, contr ollers, alarms, data acquisition, and computer systems. Model CL-301 allows 2, 3, or 4-wire connections. Conformity to the particular RTD curve is in accordance with the latest DIN 43760 standard.
Calibrations are available in both ˚F and ˚C and standard ranges are listed in T able 1-1. Other ranges and curves are available as options. Resolution is 25˚, 50˚, or 100˚, corresponding to the full scale output of 250˚, 500˚, and 1000˚ respectively.
Positive switch selection allows quick, easy settings for any output. Calibrated accuracy is ±0.05% of span ±0.25˚.
1.2 Features
• Calibrate and T roubleshoot 2, 3, or 4-wire RTD systems
• Simulates 100 ohm RTD, alpha = .00385
* Choice of Output T erminal
• Eleven Precise Test Points
• Portable, No Battery or Power Required
• Simple to Use, Inexpensive to Own
12
Page 36
Table of Contents Model CL-303 Portable mA Source/
2-Wire Simulator Section Page
Section 1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . 30
1.1 Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
1.2 Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Section 2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Section 3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.1 Operating Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.1.1 Directions for 2-Wire Simulator Mode . . . . . . . . . . . . . . . . 33
3.1.2 Directions for Source Mode. . . . . . . . . . . . . . . . . . . . . . . . . . 34
Section 4 Service Information. . . . . . . . . . . . . . . . . . . 35
4.1 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.2 Equipment Required for Calibration. . . . . . . . . . . . . . . . . . 35
4.3 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.3.1 mA Output, Source Mode Calibration . . . . . . . . . . . . . . . . 36
4.3.2 Span and Zero Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.3.2.1 2-Wire Simulator Mode Check . . . . . . . . . . . . . . . . . . . . . . . 37
Section 5 Specifications . . . . . . . . . . . . . . . . . . . . . . . 41
29
Page 37
28
Schematic–OMEGA Model CL-302
Page 38
Table 1-1
Model CL-301 Standard Ranges
Section 2 Installation
2.1 Unpacking
Remove the Packing List to check off actual equipment received. If you have any questions about the shipment, please call OMEGA’s Customer Service Department.
Upon receipt of shipment, inspect the container for any signs of damage in transit. T ake particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
13
Page 39
14
Section 3 Operation
3.1 Operating Instructions
Select the particular OMEGAModel CL-301 for the desired temperature scale (F or C), and appropriate temperature range. Then choose the cable type (2, 3, or 4-wire) that is equivalent to the RTD you are simulating . See Figure 3-1.
Connect the RTD Simulator leads to the input terminals of the instrument or system to be tested. Typically the CL-301 might feed a recording system, a controller, a datalogger, and a computer interface.
Set the temperature selection switch on the CL-301 to the desired temperature. End point calibration temperatures of the instrument are then selected on the CL-301 and any required adjustments are made. Intermediate points on the CL-301 may then be selected to verify instrument linearity or to check critical points.
NOTE
The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event shipment is necessary.
Page 40
Section 5 Specifications CL-302
Calibrated Accuracy: ±0.1% of span Precise Switch Settings: 0, 25, 50, 75, 100% of full scale
equivalent to 4, 8, 12, 16, 20 mA
Temperature Effect: ±0.01%/˚C Loop Voltage Rating: Any between 6 and 45 volts Voltage Drop: 6 V and 20 mA Dimensions: H: 3
1
/4" (82 mm) x W: 2 1/8"
(54 mm) x D: 1 1/8" (29 mm)
Weight: 3 oz. (85 grams)
27
Page 41
26
Figure 4-2. Zero and Span Controls, Model CL-302
Figure 4-1. Calibration Check
Page 42
15
Figure 3-1. Connecting the RTD Simulator to Transmitter
or Receiver
Page 43
16
Section 4 Service Information
4.1 Maintenance
The OMEGA CL-301 RTD Simulator requires little servicing. Should measurement inaccuracies be discovered, perform the adjustment and calibration procedure to recalibrate the instrument.
4.2 Equipment Required for Calibration
Digital ohmeter, 0.025% accuracy with 4-wir e ohm connections (OMEGA Model 881 or equivalent).
4.3 Calibration Procedure
Calibrate the instrument in the following manner: 1 Connect the unit to the meter in a 3-wire hookup.
2. Move the slide switch to the lowest temperature position (position
1) and adjust corresponding potentiometer (Pot 1, Figure 4-1) according to the resistance value obtained form Table 4-1.
3. Move the switch to the next temperature position (2) and adjust corresponding potentiometer (Pot 2).
4. Continue this procedure for all remaining temperature positions in ascending order .
Page 44
25
4.3.2 Span and Zero Adjustment Procedures
1 Set the switch on the Model CL-302 to the 209 mA (100%) position.
2. The span potentiometer is adjusted so that the meter reads 5.000 volts. See Figure 4-2.
3. The switch is now moved to position 4 mA (0%).
4. The zero potentiometer is adjusted until the meter reads 1.000 volts.
5. Check and readjust the span and the zero controls until they are within ±0.1% accuracy (±0.02% typical).
4.3.3 Linearity Test Protocol
1 Move the switch to position 8 mA (25%). The meter should read
between 1.996 V and 2.004 V.
2. With the meter set at 12 mA (50%), the meter should read between
2.996 V and 3.004 V.
3. Setting the meter at 16 mA ((75%), the meter should read between
3.996 V and 3.004 V.
Page 45
24
Section 4 Service Information
4.1 Maintenance
The OMEGA CL-302 Transmitter Simulator is long-lasting and requires little servicing. Should measurement inaccuracies be discovered, perform the adjustment and calibration procedures, specified below , to recalibrate the instrument.
4.2 Equipment Required for Calibration
Digital voltmeter,4 1/2digit, 0.05% (or better) accuracy with a precision (±0.05%) 250-ohm resistor.
4.3 Calibration Procedure
Calibration requires a hookup protocol, span and zero adjustments, and a linearity test.
4.3.1 Calibration Hookup Protocol
1 Set the calibrating meter to a range which covers the 1 to 5 V dc
spectrum with sufficient resolution.
2. Place the 24 to 30 V dc power supply in series with the 250 ohm (±0.05%) resistor and the Simulator which is to be calibrated. The meter will read 1 to 5 volts corresponding to 4-20 mAthrough the resistor . See Figure 4-1. The LED on the Simulator should be lit, indicating that the connections are correct and that the unit is functioning..
Page 46
17
5. Upon completion, move the switch back to position 1. Check and readjust potentiometer (Pot 1), if necessary. In increasing order, check and readjust each position as necessary to obtain specified accuracy.
Each time a position is adjusted, ALL higher number positions must be rechecked.
CAUTION
Figure 4-1. Potentiometer Adjustments, Model CL-301
Page 47
Table 4-1
Adjustment Values
18
Material and Nominal Resistance Alpha (α) R (100˚C)/R (0˚C)
Platinum (PT.)* 1.3850 (DIN Standard 43760) 100 OHM 1.3926 (U.S. Laboratory) 200 OHM 1.3889 (Canadian) 500 OHM 1.3911 (U.S. Industrial)
1.3902 (U.S. Industrial)
Nickel 1.6720 120 OHM 600 OHM
Nickel-Iron 1.5188 604 OHM
Copper 1.4274
9.035 OHM (10 OHMS at 25˚C) 234 OHM
421.2 OHM 1000 OHM 2000 OHM (at 70˚F)
*If not specified DIN standard curve 43760 will be supplied.
Page 48
23
Figure 3-1. Alternative Hookup Methods
Page 49
Upon receipt of shipment, inspect the container for any signs of damage in transit. T ake particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
Section 3 Operation
3.1 Operating Instructions
First disconnect the transmitter form the loop, leaving the power source and receiver in place.
Connect the two clip leads of the Simulator to the positive (+) and the negative (-) field connections of the receiver or controller (to be checked or calibrated) in place of the field 2-wire transmitter. See Figure 3-1. An LED indicates current flowing through the calibrator. The selector switch is then positioned at the 0% point. Exactly 4 mA will pass through the loop. At this time, the zero of the r eceiver is adjusted. Move the switch to allow 100% to pass and make the required full scale adjustment. When both span and zero are corr ect, the 25%, 50% and 75% settings are used to check linearity. Selected settings are accurate within ±0.1% of span.
22
NOTE
The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event shipment is necessary.
Page 50
Section 5 Specifications CL-301
Accuracy: ±0.05% of span ±0.25 degree Operating Ambient
Temperature: -10˚ to +130˚F (-25˚ to +55˚C) Ambient Temperature Effect: ±0.01% degree (% of span) Storage Temperature Limits: -40˚ to +160˚F (-40˚ to +70˚C) Dimensions: H: 4" (101 mm) x W: 2
1
/8" (54 mm)
x D: 2 1/2" (38 mm)
Weight: 3 oz. (0.12 kg)
19
Schematic–OMEGA Model CL-301
Page 51
Table of Contents Model CL-302 Portable Transmitter Simulator Section Page
Section 1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . 21
1.1 Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
1.2 Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Section 2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Section 3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.1 Operating Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Section 4 Service Information. . . . . . . . . . . . . . . . . . . 24
4.1 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.2 Equipment Required for Calibration. . . . . . . . . . . . . . . . . . 24
4.3 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.3.1 Calibration Hookup Procedure . . . . . . . . . . . . . . . . . . . . . . 24
4.3.2 Span and Zero Adjustment Procedures . . . . . . . . . . . . . . . 25
4.3.3 Linearity Test Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Section 5 Specifications . . . . . . . . . . . . . . . . . . . . . . . 27
20
Page 52
Section 1 Introduction
1.1 Description
The OMEGA®Model CL-302 T ransmitter Simulator provides lab accuracy for a wide range of settings. Instant availability of data is assured; no internal batteries or line cords are ever needed. Latest microcircuitry includes a precision refer ence, ultra-stable comparator, and a high-gain power Darlington output. Switch settings of 0%, 25%, 50%, 75%, and 100% allow fast full range check out of the instrument to be calibrated. Low minimum voltage drop enables other devices to remain in the loop while the receiving device is being calibrated.
1.2 Features
• Calibrate or T est, 4-20 mASystems
• Simulates Loop Powered Transmitter
• Five Precise Test Points
• Portable, 0.1% Accuracy
• No Batteries of Power Required
Section 2 Installation
2.1 Unpacking
Remove the Packing List to check off actual equipment received. If you have any questions about the shipment, please call OMEGA’s Customer Service Department.
21
Loading...