Thermocouple
Calibrator
CL130
†
Optional
U Measure and Source Thermocouples
U Large, Easy-to-Read Display with Backlight
U Thermocouple Simulation (Manual or Auto)
U 3 Programmable Step and Ramp Routines
U Auto Step Simulation (0 to 100% or 25%
Increments) with Definable LRV & URV
The CL130 thermocouple calibrator has been designed
using the latest technology to provide reliable accuracy
throughout its entire operating temperature range. The
compact design makes the unit easy to carry and the
intuitive user interface makes it easy to use. Temperature
measurements/simulations are displayed on a large, easyto-read liquid crystal display (LCD). This unit has two
backlights: a green backlight which is user activated to
improve readability in low light conditions, and an automatic
red backlight which indicates an over-range condition.
Twelve keys provide access to the functional modes and
menus, including measure, simulate, setup, run, record,
edit, and backlight activation.
Specifications
mV Range, Accuracy: -15 to 80 mV, 0.005% Rdg ±6 mV
mV Temp. Effect: ±2 µV/°C (Ref. Temp = 23°C)
Cold Junction Effect: ±0.05°C/°C
Cold Junction Accuracy: ±0.1°C at 25°C
Thermocouple Range-Read and Source
Range Accuracy
Type ºC ºF ºC ºF
-200 1372 -328 2501 0.4 0.7
-200 -165 -328 -265 0.3 0.5
-165 1200 -265 2192 0.2 0.4
-200 -168 -328 -270 0.4 0.7
-168 -86 -270 -123 0.3 0.5
-86 346 -123 655 0.2 0.4
-200 -179 -328 -290 0.3 0.5
-179 15 -290 59 0.2 0.4
15 1000 59 1832 0.1 0.2
-200 110 -328 230 0.3 0.5
110 1300 230 2372 0.2 0.4
-50 27 -58 81 2.0 3.6
27 736 81 1357 1.0 1.8
736 1768 1357 3214 0.5 0.9
-50 27 -58 81 2.0 3.6
27 1179 81 2154 1.0 1.8
1179 1768 2154 3214 0.5 0.9
250 295 482 563 3.0 5.4
295 605 563 1121 2.0 3.6
605 1301 1121 2374 1.0 1.8
1301 1820 2374 3308 0.5 0.9
SPHT-K-12
thermocouple
probe sold
separately.
CL130
Resolution:
Thermocouple Types E, J, K, N, T: 0.1°C
Thermocouple Types B, R, S: 1°C
mV: 0.001 mV (1 µV)
Environmental:
Storage: -40 to 85°C (-40 to 185°F)
Operating: -10 to 50°C (14 to 122°F)
Humidity: 10 to 90% RH
Maximum Input Voltage: 30 Vdc
Power Requirement
Batteries: 3 “AA” batteries (included),
field replaceable
Dimensions:
Enclosure: 165 H x 81 W x 28 mm D
(6.5 x 3.2 x 1.1")
Polycarbonate, Permanently Static Dissipative,
ESD Protection Enclosure with Boot:
173 H x 89 W x 33 mm D (6.8 x 3.5 x 1.3")
Extended Warranty
Program
OMEGACARESM
extended warranty
program is available for
models shown on this
page. OMEGACARE
covers parts, labor and
equivalent loaners.
AVAILABLE FOR FAST DELIVERY!
To Order
Model No. Description
CL130 Thermocouple calibrator
CL130-NIST† Thermocouple calibrator with NIST
traceable certificate of calibration
Accessories
Model No. Description
CL130-CASE Carrying case for CL130
CL-300-CABLE-(*)-2 0.6 m (2') thermocouple
CL-300-CABLE-(*)-10
SMP male to spade lugs
Comes complete with operator’s manual, shock resistant holster,
3 “AA” alkaline batteries.
For CL-300-CABLE, insert one thermocouple type: J, K, T, E, R,
*
S, B or N
Ordering Example: CL130-NIST, thermocouple/mV calibrator with
NIST certificate. OCW-2 OMEGACARE
1-year warranty to a total of 3 years.
Visit omega.com/cl130 for Pricing and Details
cable, SMP male to spade lugs
3 m (10') thermocouple cable,
SM
extends standard
SM
K
SM
K-54