Omega Products CL121-1 User Manual

Block Calibrators
CL120/CL134 Series
Starts at
$
2600
Standard
Capable of Simultaneous
Heat and Cool Sourcing in 1 Unit
Universal Jaw is
Ideal for Accepting Irregularly Shaped Probes and Sensors
Portable for Field or
Laboratory Use
CE ApprovedNIST-Traceable
Certificate of Calibration
115V Power, 230V Optional
CL122-1, $5750.
CL120 Series Calibration COOL/HEAT Source
The CL120 Series is capable of simultaneous heat and cool sourcing in 1 unit and is available with either 1 or 2 modules. You select the combination of modules: COOL source only, HEAT source only, COOL source and HEAT source, or 2 HEAT sources. Both the COOL and HEAT sources are available with the universal jaw or a “five hole block.”
A dual module system with simultaneous cooling and heating is a significant advantage, because temperature sensors are rarely tested at only 1 temperature. The most common procedure is to test at 0°C (32°F) and at the maximum temperature, thereby characterizing the offset and gain errors. Two HEAT source systems may also be ordered.
K-29
For Com
o
me
plete
D
e
t
ai
g
a
.
c
l
s
V
i
sit
o
m/
c
l
1
2
0
CL134 Series Temperature Source/Measure
The CL134 Series has the same features as the CL120 but includes measurement capabilities. This additional feature simplifies the testing of temperature sensors by conveniently housing both the source and measure in one package. The CL134 Series has 3 distinct capabilities:
1. A COOL source (either universal jaw or “5 hole block”) with range down to -30°C (-22°F)
2. A HEAT source with range to 500°C (932°F) with universal jaw or range to 700°C (1292°F) with a “5 hole block;” the CL134 Series may be ordered with 2 HEAT sources and no COOL source
3. Measures thermocouples, RTDs, current and voltage
5-hole block diameters: 1.6, 3.2, 4.8, 6.4, and 9.5 mm (1⁄
Extended Warranty
Program
S
M
General COOL Source COOL Source Heat Source Heat Source
pecifications Universal Jaw 5 Hole Block Universal Jaw 5 Hole Block
S
6
1
, 1⁄8, 3⁄
, 1⁄4, and 3⁄8" ).
6
1
Range °C Ambient to -30° 30° to 500° 30° to 700° Range °F Ambient to -22° 86° to 932° 86° to 1292°
range is 50°C (90°F) below ambient
Accuracy
1
±0.1°C (±0.2°F) ±(0.05% of setpoint + 0.1°C [0.2°F])
for probes 6.4 mm (0.25") and smaller for probes 6.4 mm (0.25") and smaller
±0.2°C (0.4°F) ±(0.1% of setpoint + 0.1°C [0.2°F])
for probes >6.4 to 12.7 mm (0.25 to 0.5") for probes >6.4 to 12.7 mm (0.25 to 0.5")
Resolution 0.1°C or °F 0.1°C or °F 0.1° to 537°C (999°F)
1° above
537°C (999°F)
Stability Repeatability
2
3
±0.05°C (0.1°F) ±0.05°C (0.1°F) ±0.05°C (0.1°F) ±0.05°C (0.1°F)
Uniformity X-Y Axis
(Across Flats) <±0.03°C (0.05°F) <±0.03°C (0.05°F) <0.1°C (0.2°F) Z Axis (Depth) <0.1°C (0.2°F) from <0.1°C (0.2°F) from 4" <0.1% of setpoint
102 mm (4") deep to (102 mm) deep to jaws from 102 mm
51 mm (2") deep (4") deep to
51 mm (2") deep
Cool/Heat Time 5 minutes to COOL and stabilize 10 minutes to HEAT 12 minutes to HEAT
from ambient to 0°C (32°F) and stabilize @ 300°C and stabilize @ 600°C
Metal Block Aluminum Aluminum Aluminum Bronze Alloy
Jaw accepts up to Drilled holes Jaw accepts up to Drilled holes
12.7 mm (0.5") 1.6, 3.2, 4.8, 6.4, 9.5 mm 12.7 mm (0.5") 1.6, 3.2, 4.8, drilled holes (
3.2, 4.8, 6.4 mm 102 mm (4") deep; 3.2, 4.8, 6.4 mm (
1
8, 3⁄16, 1⁄4") other holes available (1⁄8, 3⁄16, 1⁄4"); 102 mm (4")
(
1
16, 1⁄8, 3⁄16, 1⁄4, 3⁄8") drilled holes 6.4, 9.5 mm
1
16, 1⁄8, 3⁄16, 1⁄4, 3⁄8")
102 mm 102 mm (4") deep deep; other holes
(4") deep available
Cooling method: Peltier elements.
1
Accuracy for 1 year with ambient between 20° to 26°C.
2
Measured for 24 hours with ambient between 20° to 26°C.
3
Measured with rosemount 162CE as follows: 1. Measure temperature at 100°C. 2. Heat to maximum (500°C for universal jaw, 700°C for 5 hole block). 3. Remeasure at 100°C. The difference between 1. and 3. is <0.05°C.
OMEGACARESMextended warranty program is available for models shown on this page. Ask your sales representative for full details wh en p la cing an orde r. OME GA CARE covers parts, labor and equivalent loaners.
MOST POPULAR MODELS HIGHLIGHTED!
To Order (Specify Model Number)
Model No. Price Description CL121-1 $3250 Block calibrator with single module COOL universal jaw -30°C (-22°F) CL121-2 3050 Block calibrator with single module HEAT universal jaw 500°C (932°F) CL121-3 2800 Block calibrator with single module COOL 5-hole block -30°C (-22°F) CL121-4 2600 Block calibrator with single module HEAT 5-hole block 700°C (1292°F) CL122-1 5750 Block calibrator with dual module COOL universal jaw -30°C (-22°F),
HEAT universal jaw 500°C (932°F)
CL122-2 5550 Block calibrator with dual module HEAT universal jaw 500°C (932°F) CL122-4 4850 Block calibrator with dual module COOL 5-hole block -30°C (-22°F),
HEAT 5-hole block 700°C (1292°F)
CL122-5 4650 Block calibrator with dual module HEAT 5-hole block 700°C (1292°F)
Comes complete with NIST-traceable certificate of calibration, power cord 2.4 m (8'), ºC and ºF stickers and operator’s manual.
Note: 2 COOL sources not available in 1 instrument. Add “-230” suffix for 230 Vac power option and add $75 to price. Ordering Example: CL121-1, block calibrator with single module COOL universal jaw, -30ºC (-22ºF), $3250. OCW-3, OMEGACARE
standard 1-year warranty to a total of 4 years ($350) $3250 + 350 = $3600.
Model No. Price Description
CL134-1 $7550 Block calibrator with COOL universal jaw -30°C (-22°F),
HEAT universal jaw 500°C (932°F), and measurement readout
CL134-2 7350
Block calibrator with dual mode HEAT universal jaw 500°C (932°F), and measurement readout
CL134-4 6650 Block calibrator with COOL 5 hole block -30°C (-22°F)
HEAT 5-hole block 700°C (1292°F), and measurement readout
CL134-5 6450 Block calibrator with HEAT 5 hole block 700°C (1292°F),
and measurement readout (dual heat sources)
Add “-230” suffix for 230 Vac power option and add $75 to price. K, J, T, E, C, S, R, B, and N thermocouple types may be read with 0.1°C or °F resolution. RTDs (resistance temperature detectors) may be
tested in 2-, 3-, or 4-wire modes with alpha values of 0.385 or 0.392. The outputs of temperature transmitters may be read directly by the MEASURE unit in either mA or Vdc.
Ordering Example: CL134-4, block calibrator with cooling and heating source using standard 5-hole block, $6650. OCW-3, OMEGACARE extends standard 1-year warranty to a total of 4 years ($350) $6650 + 350 = $7000.