Block Calibrators
CL120/CL134 Series
Starts at
$
2600
Standard
⻬ Capable of Simultaneous
Heat and Cool Sourcing
in 1 Unit
⻬ Universal Jaw is
Ideal for Accepting
Irregularly Shaped
Probes and Sensors
⻬ Portable for Field or
Laboratory Use
⻬ CE Approved
⻬ NIST-Traceable
Certificate of Calibration
⻬ 115V Power, 230V Optional
CL122-1, $5750.
CL120 Series Calibration
COOL/HEAT Source
The CL120 Series is capable of
simultaneous heat and cool
sourcing in 1 unit and is available
with either 1 or 2 modules.
You select the combination of
modules: COOL source only,
HEAT source only, COOL source
and HEAT source, or 2 HEAT
sources. Both the COOL and
HEAT sources are available
with the universal jaw or a
“five hole block.”
A dual module system with
simultaneous cooling and heating
is a significant advantage, because
temperature sensors are rarely
tested at only 1 temperature.
The most common procedure is
to test at 0°C (32°F) and at the
maximum temperature, thereby
characterizing the offset and gain
errors. Two HEAT source systems
may also be ordered.
K-29
For Com
o
me
plete
D
e
t
ai
g
a
.
c
l
s
V
i
sit
o
m/
c
l
1
2
0
CL134 Series Temperature
Source/Measure
The CL134 Series has the same
features as the CL120 but includes
measurement capabilities. This
additional feature simplifies the
testing of temperature sensors by
conveniently housing both the
source and measure in one
package. The CL134 Series has
3 distinct capabilities:
1. A COOL source (either universal
jaw or “5 hole block”) with
range down to -30°C (-22°F)
2. A HEAT source with range to
500°C (932°F) with universal
jaw or range to 700°C (1292°F)
with a “5 hole block;” the CL134
Series may be ordered with
2 HEAT sources and no
COOL source
3. Measures thermocouples, RTDs,
current and voltage
5-hole block diameters: 1.6, 3.2, 4.8, 6.4, and 9.5 mm (1⁄
Extended Warranty
Program
S
M
General COOL Source COOL Source Heat Source Heat Source
pecifications Universal Jaw 5 Hole Block Universal Jaw 5 Hole Block
S
6
1
, 1⁄8, 3⁄
, 1⁄4, and 3⁄8" ).
6
1
Range °C Ambient to -30° 30° to 500° 30° to 700°
Range °F Ambient to -22° 86° to 932° 86° to 1292°
range is 50°C (90°F) below ambient
Accuracy
1
±0.1°C (±0.2°F) ±(0.05% of setpoint + 0.1°C [0.2°F])
for probes 6.4 mm (0.25") and smaller for probes 6.4 mm (0.25") and smaller
±0.2°C (0.4°F) ±(0.1% of setpoint + 0.1°C [0.2°F])
for probes >6.4 to 12.7 mm (0.25 to 0.5") for probes >6.4 to 12.7 mm (0.25 to 0.5")
Resolution 0.1°C or °F 0.1°C or °F 0.1° to 537°C (999°F)
1° above
537°C (999°F)
Stability
Repeatability
2
3
±0.05°C (0.1°F) ±0.05°C (0.1°F)
±0.05°C (0.1°F) ±0.05°C (0.1°F)
Uniformity
X-Y Axis
(Across Flats) <±0.03°C (0.05°F) <±0.03°C (0.05°F) <0.1°C (0.2°F)
Z Axis (Depth) <0.1°C (0.2°F) from <0.1°C (0.2°F) from 4" <0.1% of setpoint
102 mm (4") deep to (102 mm) deep to jaws from 102 mm
51 mm (2") deep (4") deep to
51 mm (2") deep
Cool/Heat Time 5 minutes to COOL and stabilize 10 minutes to HEAT 12 minutes to HEAT
from ambient to 0°C (32°F) and stabilize @ 300°C and stabilize @ 600°C
Metal Block Aluminum Aluminum Aluminum Bronze Alloy
Jaw accepts up to Drilled holes Jaw accepts up to Drilled holes
12.7 mm (0.5") 1.6, 3.2, 4.8, 6.4, 9.5 mm 12.7 mm (0.5") 1.6, 3.2, 4.8,
drilled holes (
3.2, 4.8, 6.4 mm 102 mm (4") deep; 3.2, 4.8, 6.4 mm (
1
⁄8, 3⁄16, 1⁄4") other holes available (1⁄8, 3⁄16, 1⁄4"); 102 mm (4")
(
1
⁄16, 1⁄8, 3⁄16, 1⁄4, 3⁄8") drilled holes 6.4, 9.5 mm
1
⁄16, 1⁄8, 3⁄16, 1⁄4, 3⁄8")
102 mm 102 mm (4") deep deep; other holes
(4") deep available
Cooling method: Peltier elements.
1
Accuracy for 1 year with ambient between 20° to 26°C.
2
Measured for 24 hours with ambient between 20° to 26°C.
3
Measured with rosemount 162CE as follows: 1. Measure temperature
at 100°C. 2. Heat to maximum (500°C for universal jaw, 700°C for
5 hole block). 3. Remeasure at 100°C. The difference between 1.
and 3. is <0.05°C.
OMEGACARESMextended warranty program
is available for models shown on this page.
Ask your sales representative for full details
wh en p la cing an orde r. OME GA CARE
covers parts, labor and equivalent loaners.
MOST POPULAR MODELS HIGHLIGHTED!
To Order (Specify Model Number)
Model No. Price Description
CL121-1 $3250 Block calibrator with single module COOL universal jaw -30°C (-22°F)
CL121-2 3050 Block calibrator with single module HEAT universal jaw 500°C (932°F)
CL121-3 2800 Block calibrator with single module COOL 5-hole block -30°C (-22°F)
CL121-4 2600 Block calibrator with single module HEAT 5-hole block 700°C (1292°F)
CL122-1 5750 Block calibrator with dual module COOL universal jaw -30°C (-22°F),
HEAT universal jaw 500°C (932°F)
CL122-2 5550 Block calibrator with dual module HEAT universal jaw 500°C (932°F)
CL122-4 4850 Block calibrator with dual module COOL 5-hole block -30°C (-22°F),
HEAT 5-hole block 700°C (1292°F)
CL122-5 4650 Block calibrator with dual module HEAT 5-hole block 700°C (1292°F)
Comes complete with NIST-traceable certificate of calibration, power cord 2.4 m (8'), ºC and ºF stickers and operator’s manual.
Note: 2 COOL sources not available in 1 instrument. Add “-230” suffix for 230 Vac power option and add $75 to price.
Ordering Example: CL121-1, block calibrator with single module COOL universal jaw, -30ºC (-22ºF), $3250. OCW-3, OMEGACARE
standard 1-year warranty to a total of 4 years ($350) $3250 + 350 = $3600.
Model No. Price Description
CL134-1 $7550 Block calibrator with COOL universal jaw -30°C (-22°F),
HEAT universal jaw 500°C (932°F), and measurement readout
CL134-2 7350
Block calibrator with dual mode HEAT universal jaw 500°C (932°F), and measurement readout
CL134-4 6650 Block calibrator with COOL 5 hole block -30°C (-22°F)
HEAT 5-hole block 700°C (1292°F), and measurement readout
CL134-5 6450 Block calibrator with HEAT 5 hole block 700°C (1292°F),
and measurement readout (dual heat sources)
Add “-230” suffix for 230 Vac power option and add $75 to price.
K, J, T, E, C, S, R, B, and N thermocouple types may be read with 0.1°C or °F resolution. RTDs (resistance temperature detectors) may be
tested in 2-, 3-, or 4-wire modes with alpha values of 0.385 or 0.392. The outputs of temperature transmitters may be read directly by the
MEASURE unit in either mA or Vdc.
Ordering Example: CL134-4, block calibrator with cooling and heating source using standard 5-hole block, $6650. OCW-3, OMEGACARE
extends standard 1-year warranty to a total of 4 years ($350) $6650 + 350 = $7000.
K-30
SM
extends
SM
K
SM