Omega Products CDTX-90 Installation Manual

Conductivity/Resistivity
ISO 9002
CERTIFIED
CORPORATE QUALITY
MANCHESTER, UK
User’s Guide
62.50 uS/cm
25.0° C
ENTER
www.omega.com
e-mail: info@omega.com
ISO 9001
CERTIFIED
CORPORATE QUALITY
STAMFORD, CT
CDTX-90 Series
Conductivity/Resistivity Transmitters
omega.com
omega.com
TM
®
OMEGAnet® Online Service Internet e-mail
www.omega.com info@omega.com
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Certi ed Stamford CT 06907-0047
TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: 2222 Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
®
®
®
Mexico: En Español: (001) 203-359-7803 e-mail: espanol@omega.com
FAX: (001) 203-359-7807 info@omega.com.mx
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
TEL: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@omegaeng.nl
Czech Republic: Rudé arm
TEL: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
France: 11, rue Jacques Cartier, 78280 Guyancourt, France
TEL: +33 (0)1 61 37 29 00 FAX: +33 (0)1 30 57 54 27 Toll Free in France: 0800 466 342 e-mail: sales@omega.fr
Germany/Austria: 22 Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
United Kingdom: 222One Omega Drive, River Bend Technology Centre
ISO 9002 Certi ed Northbank, Irlam, Manchester
M44 5BD United Kingdom TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-48 e-mail: sales@omega.co.uk
.
dy 1868, 733 01 Karvin. 8
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certi cation of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certi cation.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter speci cations without notice. WARNING: These products are not designed for use in, and should not be used for, patient-connected applications.
Omega CDTX-90 Series Conductivity/Resistivity Transmitter
Conductivity/Resistivity
62.50 uS/cm
25.0° C
ENTER
CAUTION!
• Remove power to unit before wiring input and output connections.
• Follow instructions carefully to avoid personal injury.
Contents
1. Installation
2. Specifi cations
3. Electrical Connections
4. Menu Functions
1. Installation
CDTX-90 Series transmitters are available in two styles: panel mount and fi eld mount. The panel mount is supplied with the necessary hardware to install the transmitter. This manual includes complete panel mounting instructions. Field mounting requires a separate mounting kit. The FP90UM Universal kit enables the transmitter to be installed virtually anywhere. Detailed instructions for fi eld installation options are included with the FP90UM Universal kit.
1.1 Panel Installation
1. The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual panel cutout, an adhesive template is provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch.
2. Place gasket on instrument, and install in panel.
3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.
4. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE.
Press quick-clips outward and remove.
Optional Rear Cover
56 mm
41 mm
(2.2 in.)
(1.6 in.
)
97 mm (3.8 in.)
Panel Mount with
NEMA 4X rear cover
92 mm (3.6 in.)
96 mm
(3.8 in.)
FRONT VIEW
Field &
Panel Mount
96 mm (3.8 in.)
106 mm (4.18 in.)
SIDE VIEW
Field Mount with
Universal kit
82 mm (3.23 in.)
42 mm (1.7 in.)
107 mm (4.2 in.)
SIDE VIEW
Field Mount with
Integral kit
65 mm (2.5 in.)
96 mm
(3.8 in.)
2. Specifi cations
General
Compatible electrodes: Omega CDCE-90 Series Standard and
Accuracy: ±2% of reading Enclosure:
• Rating: NEMA 4X/IP65 front
• Case: PBT
• Panel case gasket: Neoprene
• Window: Polyurethane coated polycarbonate
• Keypad: Sealed 4-key silicone rubber
• Weight: Approx. 325g (12 oz.) Display:
• Alphanumeric 2 x 16 LCD
• Contrast: User selected, 5 levels
• Update rate: 1.8 seconds
Electrical
• Power: 12 to 24 VDC ±10%, regulated, 80 mA max.
Sensor input range:
• Conductivity: 0.055 to 400,000 μS
• Resistivity: 10 KΩ to 18.2 MΩ ( Measurements from 10 MΩ to 18 MΩ (0.055 μS to 0.1 μS) must be performed in solution temperatures from 20 °C to 100 °C. )
• TDS: 0.023 to 200,000 ppm
• Temperature: PT 1000, -25 to 120°C (-13 to 248°F) Current output:
• 4 to 20 mA, isolated, fully adjustable and reversible
• Max loop impedance: 50 Ω max. @ 12 V 325 Ω max. @ 18 V 600 Ω max. @ 24 V
Certifi ed Series Conductivity/Resistivity Electrodes
• Update rate: 200 mS
• Accuracy: ±0.03 mA @ 25°C, 24 V Relay outputs (CDTX-90-2 includes 2 mechanical SPDT contacts):
• Maximum voltage rating: 5 A @ 30 VDC, or 5 A @ 250 VAC, resistive load
• Programmable for:
• High or Low setpoint with adjustable hysteresis
• Pulse operation (max. rate 400 pulses/min)
Open-collector output, optically isolated:
• 50 mA max. sink, 30 VDC maximum pull-up voltage.
• Programmable for:
• High or Low setpoint with adjustable hysteresis
• Pulse operation (max. rate 400 pulses/min)
Environmental
• Operating temperature: -10 to 70°C (14 to 158°F)
• Storage temperature: -15 to 80°C (5 to 176°F)
• Relative humidity: 0 to 95%, non-condensing
• Maximum altitude: 2000 m (6562 ft)
• Insulation category: II
• Pollution degree: 2
Standards and Approvals
• CE
• Immunity: EN50082-2
• Emissions: EN55011
• Safety: EN61010
• Manufactured under ISO 9001 and ISO 14001
• U.S.A. Patent # 5,708,363
page 1
(18 MΩ)
(10 kΩ)
CDCE-90-001
CDCE-90-20
Conductivity Range (µS)
CDCE-90 Series Sensor Ranges
Sensor Models
Ultra-
Pure
Rinse
Pure
CDCE-90-1
CDCE-90-10
CDCE-90-01
0.055 1 10 100 200 1,000 10,000 200,000 400,000
3. Electrical Connections
Sensor Selection
Select the sensor with a range closest to your process minimum and maximum values.
• CDCE-90-001 (0.01 cell): 0.055 to 100 μS (10 kΩ to 18 MΩ) Use the CDCE-90-001 for all resistivity requirements
• CDCE-90-01 (0.1 cell): 1 to 1000 μS
• CDCE-90-1 (1.0 cell): 10 to 10,000 μS
• CDCE-90-10 (10.0 cell): 100 to 200,000 μS
• CDCE-90-20 (20.0 cell): 200 to 400,000 μS
Wiring Tips:
• Do not route sensor cable in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.
• Routing sensor cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Only one wire should be inserted into a terminal. Splice double wires outside the terminal.
• Sensor cable may not be extended over 100 ft.
Caution: Failure to fully open terminal jaws before removing wire may permanently damage instrument.
CDTX-90
Conductivity/Resistivity
62.50 uS/cm
25.0 °C
ENTER
≤ 100 ft. (30 m)
Wiring Procedure
1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end.
2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
3. Insert exposed (non-insulated) wire end in terminal hole until it bottoms out.
4. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.
Wiring Removal Procedure
1. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
2. When fully open, remove wire from terminal.
3.1 Sensor Connections
CDTX-90-1 Terminals
Shield (Sensor Gnd)
10
Black (Iso Gnd)
9
White (Temp In)
8
Red (Signal In)
7
CDTX-90-2 Terminals
Shield (Sensor Gnd)
14
Black (Iso Gnd)
13
White (Temp In)
12
Red (Signal In)
11
CDCE-90 Series Standard and Certified Cells
2
1
CDTX-90-3 Terminals
page 2
Shield (Sensor Gnd)
14
Black (Iso Gnd)
13
White (Temp In)
12
Red (Signal In)
11
Stand-alone application, no current loop used
Transmitter
4 3 2 1
Sys. Pwr. Loop -
Sys. Pwr. Loop +
AUX Power -
AUX Power +
Power Supply DC 12 - 24 V
Power Supply
Power Supply
Connection to a PLC with built-in power supply
PLC
Power Supply Ground
Power Supply
Loop Input 4-20 mA
Loop Input 4-20 mA
Internal PLC Connection
Connection to a PLC/Recorder, separate supply
PLC or Recorder
Loop Input 4-20 mA in
Loop Input 4-20 mA in
DC 12 - 24 V
Power Supply
Power Supply
Example: Two transmitters connected to PLC/Recorder
with separate power supply
Transmitter 2
4 3 2 1
DC 12 - 24 V
Power Supply
Power Supply
Transmitter 1
4 3 2 1
Transmitter
4 3 2 1
Sys. Pwr. Loop -
Sys. Pwr. Loop +
AUX Power -
AUX Power +
Transmitter
4 3 2 1
Sys. Pwr. Loop -
Sys. Pwr. Loop +
AUX Power -
AUX Power +
Sys. Pwr. Loop -
Sys. Pwr. Loop +
AUX Power -
AUX Power +
Sys. Pwr. Loop -
Sys. Pwr. Loop +
AUX Power -
AUX Power +
PLC or Recorder
Channel 1 4-20 mA in
Channel 1 4-20 mA in
Channel 2 4-20 mA in
Channel 2 4-20 mA in
Terminals
Terminals
Terminals
Terminals
-
+
-
+
-
+
-
+
-
+
-
+
-
+
AUX power required for all CDTX-90 systems.
CDTX-90-1 Terminal connections
Output -
Output +
System Pwr Loop -
System Pwr Loop +
4
3
2
1
6
5
Sensr Gnd (SHIELD)
Iso. Gnd (BLACK)
Temp. IN (WHITE)
Signal IN (RED)
10
9
8
7
AUX Power -
AUX Power +
Internal open-collector
output circuit
Output --
Output +
Isolation
15W
S
D
Terminals 3 and 4: Loop Power
12-24 VDC ±10% system power and current loop output. Max. loop impedance: 50 Ω max. @ 12 V 325 Ω max. @ 18 V 600 Ω max. @ 24 V
Terminals 1-2: Auxiliary power
Provides DC power to measurement circuit. Required for all CDTX-90 systems.
Terminals 7-10: Sensor input 7 is conductivity input 8 is temperature input 9 is the isolated signal ground 10 is the sensor earth ground
Terminals 5-6: Open-collector Output
• A transistor output, programmable (see CALIBRATE menu) as High or Low setpoint with adjustable hysteresis and proportional pulse (to 400 pulses per minute maximum).
• May be disabled (Off) if not used.
CDTX-90-1 System Power/Loop Connections
page 3
Stand-alone application, no current loop used
Transmitter
Terminals
4 3 2 1
Sys. Pwr. Loop ­Sys. Pwr. Loop + AUX Power ­AUX Power +
Power Supply DC 12 - 24 V
Power Supply Power Supply
Connection to a PLC with built-in power supply
PLC
Terminals
Power Supply Ground Power Supply Loop Input 4-20 mA Loop Input 4-20 mA
Internal PLC Connection
Connection to a PLC/Recorder, separate supply
PLC or Recorder
Loop Input 4-20 mA in Loop Input 4-20 mA in
DC 12 - 24 V
Power Supply Power Supply
Example: Two transmitters connected to PLC/Recorder
with separate power supply
Transmitter 2
4 3 2 1
DC 12 - 24 V
Power Supply Power Supply
Transmitter 1
4 3 2 1
Transmitter
Terminals
4 3 2 1
Sys. Pwr. Loop ­Sys. Pwr. Loop + AUX Power ­AUX Power +
Transmitter
Terminals
4 3 2 1
Sys. Pwr. Loop ­Sys. Pwr. Loop + AUX Power ­AUX Power +
Sys. Pwr. Loop ­Sys. Pwr. Loop + AUX Power ­AUX Power +
Sys. Pwr. Loop ­Sys. Pwr. Loop + AUX Power ­AUX Power +
PLC or Recorder
Channel 1 4-20 mA in Channel 1 4-20 mA in
Channel 2 4-20 mA in Channel 2 4-20 mA in
-
+
-
+
-
+
-
+
-
+
-
+
-
+
-
+
AUX required for all CDTX-90 systems.
System Pwr Loop -
System Pwr Loop +
AUX Power -
AUX Power +
4 3 2 1
Sensr Gnd (SHIELD)
Iso. Gnd (BLACK)
Temp. IN (WHITE)
Signal IN (RED)
14 13 12 11
Relay 2 (NO)
Relay 2 (COM)
Relay 2 (NC)
Relay 1 (NO)
Relay 1 (COM)
Relay 1 (NC)
10
9 8 7 6 5
CDTX-90-2 Terminal Connections
Terminals 3 and 4: Loop Power
12-24 VDC ±10% system power and current loop output. Max. loop impedance: 50 Ω max. @ 12 V 325 Ω max. @ 18 V 600 Ω max. @ 24 V
Terminals 11-14: Sensor input 11 is conductivity input 12 is temperature input 13 is the isolated signal ground 14 is the sensor earth ground
Terminals 1-2: Auxiliary power
Provides DC power to sensor. Required for all CDTX-90 systems.
CDTX-90-2 System Power/Loop Connections
Terminals 5-10: Relay Outputs
• Two SPDT mechanical relays, programmable as:
• High or Low setpoint with adjustable hysteresis.
• proportional Pulse (400 pulses per minute maximum).
• May be disabled (Off) if not used.
page 4
1
Internal open-collector
output circuit
Isolation
15W
S
D
_
CDTX-90-3 Terminal Connections
Terminals 3-6: Loop Power
12-24 VDC ±10% system power and current loop output. Max. loop impedance: 50 Ω max. @ 12 V 325 Ω max. @ 18 V 600 Ω max. @ 24 V
Loop 2-
6
Loop 2+
5
10
Output 2-
Terminals 11-14: Sensor input 11 is conductivity input 12 is temperature input 13 is the isolated signal ground 14 is the sensor earth ground
System Pwr
4
Loop -
System Pwr
3
Loop +
AUX
2
Power -
AUX
1
Power +
Terminals 7-10: Open-collector Outputs
•Two transistor outputs, programmable
Terminals 1-2: Auxiliary power
Provides DC power to measurement circuit. Required for all CDTX-90 systems.
as:
• High or Low setpoint with adjustable hysteresis
• proportional pulse (to 400 pulses per minute maximum)
• May be disabled (Off) if not used
CDTX-90-3 System Power/Loop Connections
Stand-alone application, no current loop used
Trans mi tte r
Termina ls
NC
Loop 2 –
Loop 2 +
System Power Loop 1 –
System Power Loop 1 +
AUX Power –
AUX Power +
6
NC
5
4
3
2
1
– +
Power
S upply
12–24 V DC
9
8
7
Output 2+
Output 1-
Output 1+
Sensr Gnd
14
(SHIELD)
Iso. Gnd
13
(BLACK)
Temp. IN
12
(WHITE)
Signal IN (RED)
11
Connection to a PLC with internal power supply
Loop 2 –
Loop 2 +
System Power Loop 1 –
System Power Loop 1 +
AUX Power –
AUX Power +
Transmitter
Termina ls
6
5
4
3
2
1
PLC
+ –
+ –
– +
Internal
Connection
PLC Termina ls
Channel 2
4-20 mA
Channel 1
4-20 mA
Power
Supply
12–24 VDC
Two transmitters connected to a PLC/Recorder with separate power supply
+
Channel 1 4-20 mA in
+
Channel 2 4-20 mA in
Channel 3 4-20 mA in
+
– +
– +
Channel 4 4-20 mA in
Power
Supply
12–24 VDC
Loop 2 –
Loop 2 +
System Power Loop 1 –
System Power Loop 1 +
AUX Power –
AUX Power +
Trans mitter 1
Term inals
6
5
4
3
2
1
Trans mitter 2
Term inals
6
5
4
3
2
1
Loop 2 –
Loop 2 +
System Power Loop 1 –
System Power Loop 1 +
AUX Power –
AUX Power +
Connection to a PLC with internal power supply
Loop 2 –
Loop 2 +
System Power Loop 1 –
System Power Loop 1 +
AUX Power –
AUX Power +
Transmitter
Termina ls
6
5
4
3
2
1
PLC
+ –
+ –
– +
Internal
Connection
PLC Termina ls
Channel 2
4-20 mA
Channel 1
4-20 mA
Power
Supply
12–24 VDC
page 5
Hysteresis
Time
Low Setpoint
Process
Hysteresis
Output active
Output inactive
High Setpoint
Process
10
5
Open collector or Relay Output
PlsRate: 0 to 100 Pulses/min.
endpoint
Pulse rate
0 pulses
100 pulses
Starting point
3.3 Relay and Open Collector Outputs
The Relay and Open Collector outputs can be used as a switch that responds when the process value moves above or below a setpoint, or they can be used to generate a pulsing signal with a rate proportional to the process value.
Low setpoint:
Output triggers when process is less than the setpoint. The output will relax when the process moves above the setpoint plus the hysteresis value.
High setpoint:
Output triggers when process is greater than the setpoint. The output will relax when the process moves below the setpoint plus the hysteresis value.
Proportional Pulsing
The output will generate a 100 mS pulse at the rate defi ned by settings in the CALIBRATE menu.
In the example below:
• The output will be 0 pulses/min. when value is less than 5 μS.
• The output will be 50 pulses/min. when value is 7.5 μS.
• The output will be 100 when value is greater than 10 μS.
The starting point, endpoint and maximum pulse rate are select ­able in the CALIBRATE menu.
VIEW menu
• During normal operation, the transmitter displays the VIEW menu.
• When editing the CALIBRATE or OPTIONS menus, the transmitter will return to the VIEW menu if no activity occurs for 10 minutes.
• To select a VIEW display, press the UP or DOWN arrow keys. The selections will scroll in a continuous loop.
• Changing the VIEW display does not interrupt system operations.
• No key code is necessary to change display selection.
• Output settings cannot be edited from the VIEW menu.
CDTX-90 Series View Menu
Display Description
123.45 μS/cm
+67.89 ºC
The VIEW displays below are temporary. The permanent display will return after 10 minutes
Loop Output:
13.75 mA
Last CAL: 04-20-07
Monitor the Conductivity and Temperature values from the sensor. This is the permanent display.
Monitor the Loop Current output.
Check the Last Calibration date.
Conductivity/Resistivity
62.50 uS/cm
25.0° C
ENTER
page 6
CDTX-90 Series Editing Procedure:
OPTIONS
CALIBRATE
VIEW
2s 5s
Press & hold for access:
ENTER
Step 1. Press and hold ENTER key:
• 2 seconds to select the CALIBRATE menu.
• 5 seconds to select the OPTIONS menu.
Step 2. The Key Code is UP-UP-UP-DOWN keys in sequence.
• After entering the Key Code, the display will show the fi rst item in the selected menu.
Step 3. Scroll menu with UP or DOWN arrow keys. Step 4. Press RIGHT ARROW key to select menu item to be edited.
The fi rst display element will begin fl ashing.
Step 5. Press UP or DOWN keys to edit the fl ashing element.
• RIGHT ARROW key advances the fl ashing element.
Step 6. Press ENTER key to save the new setting and return to Step 3.
Notes on Step 1:
• The View Menu is normally displayed.
• The CALIBRATE and OPTIONS menus require a KEY CODE.
Notes on Step 2:
If no key is pressed for 5 minutes while display is showing "Enter Key Code", the display will return to the VIEW menu.
CALIBRATE: ----
Enter Key Code
Notes on Steps 3 and 4:
• Refer to pages 8-13 for complete listing of menu items and their use.
• From the Step 3 display, pressing the UP and DOWN keys simultaneously will return the display to the VIEW menu.
• If no key is pressed for 10 minutes, display will also return to the VIEW menu.
Step 3: Finished Editing?
Press the UP and DOWN keys simultaneously after saving the last setting to return to normal operation.
CALIBRATE: *---
Enter Key Code
CALIBRATE: **--
Enter Key Code
CALIBRATE: ***-
Enter Key Code
Cell Constant:
Standard >
Cell Constant:
Standard >
Output Setpnt:
20.0 uS >
First item in CALIBRATE menu
Step 3
Step 4
Notes on Steps 5 and 6:
• All output functions remain active during editing.
• Only the fl ashing element can be edited.
• RIGHT ARROW key advances the fl ashing element in a continuous loop.
• Edited value is effective immediately after pressing ENTER key.
• If no key is pressed for 10 minutes unit will restore the last saved value and return to step 3.
• Step 6 (pressing ENTER key) always returns you to Step 3.
• Repeat steps 3-6 until all editing is completed.
Press the UP and DOWN keys simultaneously while any element is fl ashing. This will recall the last saved value of the item being edited and return you to Step 3.
Step 5: Made an Error?
Output Setpnt:
20.00 uS
Output Setpnt:
10.00 uS
Output Setpnt:
0.00 uS
1
Step 6
Output Setpnt:
19.00 uS >
Step 5
Output Setpnt:
9
.00 uS
1
Output Setpnt:
Saving
ENTER
page 7
CDTX-90-1 Calibrate Menu
Display
(Factory settings shown)
Description
Cell Constant:
Standard >
Cell: Standard
1 >
Cell: Custom
1.0000 >
Cond Units:
uS >
PPM Factor:
2.00 >
Set:
Temperature >
Set:
Conductivity >
Select CUSTOM only if you are connecting a certifi ed conductivity sensor. Select STANDARD for all other systems.
For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0.
For CUSTOM sensors: Enter the precise cell constant from the certifi cate provided with your sensor, or from the information label on the sensor.
Select from the following options: μS, mS, kΩ, MΩ, PPM
If the Units selection is PPM, you can set the ratio of Total Dissolved Solids to μS. See page 15 for additional information. The TDS factor can only be set in PPM.
Adjust the temperature of the system based on an accurate external reference.
For use with STANDARD sensors: Perform this single-point wet calibration for most accurate results. Not required for CUSTOM sensors except for periodic system accuracy confi rmation. Enter all zeroes to restore factory calibration to TEMP and COND settings.
Loop Source:
Cond >
Loop Range: uS
0.0000 → 100.000 >
Output Source:
Cond >
Output Mode:
Low >
Output Setpnt:
10.0000 uS >
Output Hys:
0.5000 uS >
Output Rng: uS
10.0000 → 40.0000 >
Output PlsRate:
120 Pulses/Min >
Last CAL:
04-20-07 >
Select the input source to be associated with the current loop output: Conductivity or Temperature.
Select the minimum and maximum values for the current loop output. Be sure to modify this setting if you change the Cond. Units.
Select the source for the Open Collector output: Conductivity or Temperature
Select the mode of operation for the Open Collector output. Options available are High, Low or proportional Pulse. The signal may be disabled (Off) if not in use.
In Low or High Mode, this Open Collector output will be activated when the process reaches this value. Be sure to modify this setting if you change the Cond. Units.
The Open Collector output will be deactivated at Setpoint ± Hysteresis, depending on High or Low Setpoint selection. (See details in section 3.3.)
In Pulse mode, set the process values where the proportional pulse will start and where it will reach the maximum rate. Be sure to modify this setting if you change the Cond. Units.
In Pulse mode, set the maximum rate for the proportional Pulse output. The CDTX-90 will accept any value from 0 to 400.
Use this “note pad” to record important dates, such as annual recertifi cation or scheduled maintenance.
page 8
CDTX-90-1 Options Menu
Display
(Factory settings shown)
Contrast: 3 >
Cond Decimal: ****.* >
Averaging
Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. In general, select lower contrast if the display is in warmer ambient surroundings.
Set the decimal to the best resolution for your application. The display will auto matically scale down to this restriction. Select *****., ****.*, ***.** **.*** or *.****
OFF provides the most instantaneous response to process changes. Select LOW (4 sec) or HIGH (8 sec) averaging if your process experiences frequent or extreme fl uctuations.
O >
Loop Adjust:
4.00 mA >
Loop Adjust:
20.00 mA >
Adjust the minimum and maximum current output. The display value represents the precise current output. Adjustment limits:
• 3.80 mA < 4.00 mA > 5.00 mA
• 19.00 mA < 20.00 mA > 21.00 ma Use this setting to match the system output to any external device.
Temp Display:
ºC >
Select ºC or ºF
Temperature
Set the percent change in Conductivity caused by a 1ºC change in temperature. May be
Comp %: 2.00 >
from 0.00 to 10.00.
Description
Output Active:
Active HIGH: This setting is used to turn a device (pump, valve) ON at the setpoint.
Low >
Test Loop: >
Test Output:
Active LOW: This setting is used to turn a device OFF at the setpoint.
Press UP and DOWN keys to manually order any output current value from 3.6 mA to
21.00 mA to test the output loop.
Press UP and DOWN keys to manually toggle the Open Collector output state.
>
page 9
CDTX-90-2 Calibrate Menu
Display
(Factory settings shown)
Description
Cell Constant:
Standard >
Cell: Standard
1 >
Cell: Custom
1.0000 >
Cond Units:
μS >
PPM Factor:
2.00 >
Set:
Temperature >
Set:
Conductivity >
Select CUSTOM only if you are connecting a certifi ed conductivity sensor. Select STANDARD for all other systems.
For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0.
For CUSTOM sensors: Enter the precise cell constant from the certifi cate provided with your sensor, or from the information label on the sensor.
Select from these options: μS, mS, kΩ, MΩ, PPM
If the Units selection is PPM, set the ratio of Total Dissolved Solids to μS. The maximum allowable is 3 μS = 1 PPM. The TDS factor can only be set in PPM.
Adjust the temperature of the system based on an accurate external reference.
When using a h STANDARD sensor perform this single-point wet calibration for best accuracy. Not required for CUSTOM sensors except for periodic calibration checks.
Loop Source:
Cond >
Loop Range: μS
0.0000 → 100.000 >
Relay1 Source:
Cond >
Relay1 Mode:
Low >
Relay1 Setpnt:
10.0000 μS >
Relay1 Hys: μS
0.5000 >
Relay1 Rng: μS
10.0000 → 40.0000 >
Relay1 PlsRate:
120 Pulses/Min >
Last CAL:
04-20-07 >
Select the input (Conductivity or Temperature) source to be associated with the current loop output:
Select the minimum and maximum values for the current loop output. Be sure to modify this setting if you change the Cond. Units.
Select the source (Conductivity or Temperature) for this relay output:
Select the mode of operation for the relay. Options available are High, Low or proportional Pulse. The relay may be disabled (Off) if not in use.
In Low or High Mode, the relay will be activated when the process reaches this value. Be sure to modify this setting if you change the Cond. Units.
The relay will be deactivated at Setpoint ± Hysteresis, depending on High or Low Setpoint selection.
In Pulse mode, set the process values where the proportional pulse will start and where it will reach the maximum rate. Be sure to modify this setting if you change the Cond. Units.
In Pulse mode, set the maximum rate for the proportional Pulse. The transmitter will accept any value from 0 to 400.
Use this “note pad” to record important dates, such as annual recertifi cation or scheduled maintenance.
page 10
CDTX-90-2 Options Menu
Display
(Factory settings shown)
Contrast:
3 >
Cond Decimal:
****.* >
Averaging:
O >
Loop Adjust:
4.00 mA >
Loop Adjust:
20.00 mA >
Temp Display:
ºC >
Temperature
Comp %: 2.00 >
Description
Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. In general, select lower contrast if the display is in warmer ambient surroundings.
Set the decimal to the best resolution for your application. The display will automatically scale down to this restriction. Select *****., ****.*, ***.** **.*** or *.****
OFF provides the most instantaneous response to process changes. Select LOW (4 sec) or HIGH (8 sec) averaging if your process experiences frequent or extreme fl uctuations.
Adjust the minimum and maximum current output. The display value represents the precise current output. Adjustment limits:
• 3.80 mA < 4.00 mA > 5.00 mA
• 19.00 mA < 20.00 mA > 21.00 ma Use this setting to match the system output to any external device.
Select ºC or ºF.
Set the percent change in Conductivity caused by a 1ºC change in temperature. May be from 0.00 to 10.00.
Test Loop:
>
Test Relay1:
Press UP and DOWN keys to manually order any output current value from 3.6 mA to
21.00 mA to test the output loop.
Press UP and DOWN keys to manually toggle the relay state.
>
page 11
CDTX-90-3 Calibrate Menu
Display
(Factory settings shown)
Description
Cell Constant:
Standard >
Cell: Standard
1 >
Cell: Custom
1.0000 >
Cond Units:
uS >
PPM Factor:
2.00 >
Set:
Temperature >
Set:
Conductivity >
Select CUSTOM only if you are connecting a certifi ed conductivity sensor. Select STANDARD for all other systems.
For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0.
For CUSTOM sensors: Enter the precise cell constant from the certifi cate provided with your sensor, or from the information label on the sensor.
Select from the following options: μS, mS, kΩ, MΩ, PPM
If the Units selection is PPM, you can set the ratio of Total Dissolved Solids to μS. See section 3 of this manual for additional information. The TDS factor can only be set in PPM.
Adjust the temperature of the system based on an accurate external reference.
For use with STANDARD sensors: Perform this single-point wet calibration for most accurate results. Not required for CUSTOM sensors except for periodic system accuracy confi rmation. Enter all zeroes to restore factory calibration to TEMP and COND settings.
Loop1 Source:
Cond >
Loop1 Range: uS
0.0000 → 100.000 >
Output Source:
Cond >
Output1 Mode:
Low >
Output1 Setpnt:
10.0000 uS >
Output1 Hys:
0.5000 uS >
Output1 Rng: uS
10.0000 → 40.0000 >
Select the input source to be associated with the current loop output: Conductivity or Temperature
Select the minimum and maximum values for the current loop output. Be sure to modify this setting if you change the Cond. Units.
Select the source for the Open Collector output: Conductivity or Temperature
Select the mode of operation for the Open Collector output. Options available are High, Low or proportional Pulse. The signal may be disabled (Off) if not in use.
In Low or High Mode, this Open Collector output will be activated when the process reaches this value. Be sure to modify this setting if you change the Cond. Units.
The Open Collector output will be deactivated at Setpoint ± Hysteresis, depending on High or Low Setpoint selection. (See details on page 4.)
In Pulse mode, set the process values where the proportional pulse will start and where it will reach the maximum rate. Be sure to modify this setting if you change the Cond. Units.
Output1 PlsRate:
120 Pulses/Min >
Last CAL:
In Pulse mode, set the maximum rate for the proportional Pulse output. The transmitter will accept any value from 0 to 400.
Use this “note pad” to record important dates, such as annual recertifi cation or scheduled maintenance.
04-20-07 >
Settings will repeat for Loop 2 and Output 2.
page 12
CDTX-90-3 Options Menu
Display
(Factory settings shown)
Contrast: 3 >
Cond Decimal: ****.* >
Averaging
O >
Loop1 Adjust:
4.00 mA >
Loop1 Adjust:
Adjust the LCD contrast for best viewing. A setting of 1 is lower contrast, 5 is higher. In general, select lower contrast if the display is in warmer ambient surroundings.
Set the decimal to the best resolution for your application. The display will automatically scale down to this restriction. Select *****., ****.*, ***.** **.*** or *.****
OFF provides the most instantaneous response to process changes. Select LOW (4 sec) or HIGH (8 sec) averaging if your process experiences frequent or extreme fl uctuations.
Adjust the minimum and maximum current output. The display value represents the precise current output. Adjustment limits:
• 3.80 mA < 4.00 mA > 5.00 mA
• 19.00 mA < 20.00 mA > 21.00 ma Use this setting to match the system output to any external device.
20.00 mA >
Temp Display:
Select ºC or ºF
ºC >
Temperature
Comp %: 2.00 >
Set the percent change in Conductivity caused by a 1ºC change in temperature. May be from 0.00 to 10.00.
Description
Output1 Active:
Active HIGH: This setting is used to turn a device (pump, valve) ON at the setpoint.
Low >
Test Loop 1: >
Test Output 1:
Active LOW: This setting is used to turn a device OFF at the setpoint.
Press UP and DOWN keys to manually order any output current value from
3.6 mA to 21.00 mA to test the output loop.
Press UP and DOWN keys to manually toggle the Open Collector output state.
>
Settings will repeat for Loop 2 and Output 2.
page 13
Calibration Procedure
Iso. Gnd (BLACK)
Temp. IN (WHITE)
Signal IN (RED)
9
8
7
Simulation resistor
TC resistor
1. Requirements
The CDTX-90 Transmitter is factory calibrated using simulated input signals. System calibration will reduce errors caused by sensor wire lengths longer than the standard 15 ft. length. Wire lengths of 100 feet are acceptable; cable shield must be maintained through cable splice. Calibration may be done by known solution value (A), or by resistance simulation (B).
A) Calibration with NIST Traceable Solutions:
When using NIST traceable standards, ensure the sensor and test solution are at the solution temperature specifi ed on the test solution label. Prevent contamination of the calibration solution. Thoroughly rinse the sensor in a small amount of test solution before placing in any test solution for calibration purposes.
B) Optional Verifi cation with Precision Resistors:
The use of precision resistors (±0.1%) connected to the rear "Temp In", "Signal IN", and "Iso Gnd" terminals in place of the sensor, yields quick and accurate electronic instrument calibration. Calibration is completed as follows:
1) Select a standard cell constant based on desired range of operation.
2) Place a 1096 Ω (25°C) resistor between "Temp IN" and "Iso. Gnd" terminals as shown. Note: Temperature simulation errors can adversely effect calibration: 3.85 per °C.
3) Calculate the required simulation resistor that represents a value within the selected cell range. The formula for determining the required simulation resistance is:
Resistance = Sensor Cell Desired conductivity (Siemens*) 0.000020 (Siemens*)
e.g. 0.1 Cell = 5,000 Ω or 5 KΩ
:
Conductivity = Sensor Cell Simulation resistance (Ω) 100,000 (Ω) or 1μS/cm
e.g. 0.1 Cell = 0.000001 Siemens*
:
(*Conversion: 1 μS = 1 X 10-6 Siemens or 0.000001 Siemens)
4) Place the calculated simulation resistance between the "Signal IN" and "Iso Gnd" terminals as shown.
5) Set temperature and set conductivity. (Optional: Reset to factory calibration by entering zero as the "Set Conductivity" value.)
2. Temperature Coeffi cient
Conductivity measurement is highly dependent on temperature. Temperature dependence is expressed as the relative change per °C, commonly known as percent/°C change from 25°C, or slope
temperature. Place sensor in the sample solution allowing several minutes for stabilization. Record displayed temperature and conductivity values in the spaces provided:
of the solution.
Displayed temperature: T2 = _______ °C
Slopes can very signifi cantly depending on process solution type.
Displayed conductivity: C2 = _______ μS
The factory default temperature compensation factor is 2.00%/°C. Process solutions may require adjustment for maximum accuracy. To determine the optimum temperature compensation factor for a
(A 10% change in conductivity between steps 2 and 3 is
recommended.)
process:
4. Substitute recorded readings (steps 2 and 3) into the
1. Disable the temperature compensation % factor by entering
following formula:
0.00.
TC Slope = 100 x (C1 - C2)
2. Heat the sample solution close to the maximum process
(C2 x (T1 - 25)) - (C1 x (T2 - 25)) temperature. Place sensor in the sample solution allowing several minutes for stabilization. Access the VIEW menu and record the displayed temperature and conductivity values in the spaces provided:
Example: A sample solution has a conductivity of 205 μS
@ 48°C. After cooling the solution, the conductivity was measured at 150 μS @ 23 °C. (C1 = 205, T1 = 48, C2 = 150, T2 = 23)
Displayed temperature: T1 = _______ °C
The TC is calculated as follows:
Displayed conductivity: C1 = _______ μS
TC Slope = 100 x (205 - 150) = 5500 = 1.42%/°C
(Do not use this procedure for solutions from 0.055 μS to 0.1 μS
(150 x (48 - 25)) - (205 x (23 - 25)) 3860
(10 MΩ to 18 MΩ). An internal pure water curve is used for these ranges. The factory default setting of 2.00%/°C should be used.)
3. Cool the sample solution close to the minimum process
page 14
Parts Per Million (PPM) Factor
This feature is only applicable when PPM display units are selected. The programmable PPM Factor is adjustable from 0.01 to
3.00 (factory default = 2.00). Determine the best PPM Factor for a process solution by calculating the solution's conductivity (μS) and the percent of total dissolved solids (PPM).
PPM Factor = Solution conductivity (μS)/cm Total dissolved solids (PPM)
TDS (PPM) = Solution conductivity (μS)/cm PPM Factor
Example: • Solution conductivity = 400 μS/cm
• TDS = 200 PPM (mg/L)
200 PPM
PPM Factor = 400 μS/cm = 2.00
Troubleshooting
Display Condition Suggested SolutionsPossible Causes
"- - - -"
Value Must be less than 3
Value must be greater than 0
"Value must be 400 or less"
Display erratic, shows "0" or "------" intermittently
Display is over range. This may be a normal condition if your process operates at/near the limits of the sensor range. kΩ or MΩ units selected & sensor is open.
The PPM factor must be a value from
0.00 to 3.00
Custom cell constant cannot be set to 0
Pulse Rate cannot be greater than 400
Auxiliary Power not connected
Check sensor for correct range. Check Decimal setting in OPTIONS menu. Check Calibrate menu settings for incompatible SOURCE and RANGE values.
Set a PPM value less than 3.
Set cell constant to a value greater than 0
set a pulse rate value less than 400
Connect Auxiliary power
"Too much Error Check Sensor"
Defective or dirty/coated sensor Wiring error
Clean or replace sensor
Check and correct wiring Fluid conductivity too low Defective temp element
Replace sensor
Technical notes
If a Current Loop is locked at 3.6 mA, the problem is related to the temperature circuit: This occurs only if the CDTX-90 detects a resistance from the temperature sensor that is less than 250Ω or greater than 2800Ω.
• Check the sensor wiring for open/short or poor connections on white (TEMP IN) and black (ISO GND)wires.
• The PT1000 temperature device in the sensor is defective.
• The transmitter is defective.
Auxiliary Power (terminals 1 and 2) must ALWAYS be connected for the CDTX-90 to operate.
page 15
Notes
page 16
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product
2. Model and serial number of the product under
3. Repair instructions and/or speci c problems relative
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This a ords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2000 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated,
or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
was PURCHASED,
warranty, and
to the product.
FOR NON-WARRANTY REPAIRS, repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or speci c problems relative to the product.
consult OMEGA for current
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at www.omega.com
TEMPERATURE
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
Wire: Thermocouple, RTD & Thermistor
Calibrators & Ice Point References
Recorders, Controllers & Process Monitors
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
Transducers & Strain Gages
Load Cells & Pressure Gages
Displacement Transducers
Instrumentation & Accessories
FLOW/LEVEL
Rotameters, Gas Mass Flowmeters & Flow Computers
Air Velocity Indicators
Turbine/Paddlewheel Systems
Totalizers & Batch Controllers
pH/CONDUCTIVITY
pH Electrodes, Testers & Accessories
Benchtop/Laboratory Meters
Controllers, Calibrators, Simulators & Pumps
Industrial pH & Conductivity Equipment
DATA ACQUISITION
Data Acquisition & Engineering Software
Communications-Based Acquisition Systems
Plug-in Cards for Apple, IBM & Compatibles
Datalogging Systems
Recorders, Printers & Plotters
HEATERS
Heating Cable
Cartridge & Strip Heaters
Immersion & Band Heaters
Flexible Heaters
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
Metering & Control Instrumentation
Refractometers
Pumps & Tubing
Air, Soil & Water Monitors
Industrial Water & Wastewater Treatment
pH, Conductivity & Dissolved Oxygen Instruments
6-8850.090-OM/(D 10/08) M3632/1008
Loading...