Conductivity Transmitters Using 4-Ring Technology
CDTX1201 Series
U Virtually Immune to Contamination
U Built-In Temperature Sensor on the
Probe and the Transmitter’s ATC
Circuitry with a ß of 2%
U Direct Connection of the Probe to
the Transmitter Ensures a Positive
Electrical Connection with No Signal
Loss over Long Distances
U 2 Versions are Available: CDTX1200
(Without an LCD), and the
CDTX1200-D [With LCD (Allows
Easy Verification and Monitoring of
Measured Values, and is Easier to
Calibrate and Maintain)]
U Should be Used in Conjunction
with the CDE-1201 In-Line Probe
or CDE-1202 Tank Probe with
External Threads
CONDUCTIVITY, 4-RING
TECHNOLOGY
The CDTX1200 Series conductivity
transmitters use top-of-the-line, 4-ring
potentiometric probes. The 4-ring
potentiometric method provides much
higher accuracy and repeatability
than the 2-ring amperometric method.
When measuring liquids that have a
high conductivity, the 2-ring system is
susceptible to polarization. This condition
makes it difficult to obtain accurate
measurements. Polarization is directly
related to the electrode’s current load, and
will cause a considerable non-linear drop
in the voltage. As a result, the solution
around the electrode simulates a
low-conductivity condition.
4-ring electrodes eliminate the polarization
effect by splitting the 4 rings into 2 current
electrodes and 2 voltage electrodes. When
placed in a conductive liquid, the 2 current
electrodes take the alternating voltage
and create a current. This alternating
current produces a buffer field from which
polarization is absent. The voltage is then
measured in this field, which ensures
accurate readings.
Voltage electrodes carry very little
current. This allows the electrodes to be
arranged in such a way that the current
electrodes, which are the most susceptible
to polarization, are placed where they
cannot affect the voltage electrodes.
Consequently, the buffer field is large
enough for the voltage electrodes to take
accurate readings without being affected by
polarization.
CDTX1201-D.
CDTX1201.
The wide range of OMEGA
probes includes models for industrial
applications. Flow-thru and DIP versions
are available. The latest additions to this
family are probes with 1⁄2" and 3⁄4" external
threads for mounting on commercial
pipe tees.
The 4-pin CDE-1201 is a molded
conductivity probe with pipe threads
at both ends. The housing is fiberreinforced polypropylene and can work
with pressures up to 5 bar (72.5 psi) and
temperatures as high as 80°C (176°F).
®
conductivity
All products
shown smaller
than actual size.
These probes combine the proven 4-ring
potentiometric method with platinum
sensors and a stainless steel external
thread. The removable protective sleeve
is made of Ultem
harmful effects of most chemicals, and
can be unscrewed for quick and easy
maintenance. The new probes withstand
temperatures up to 120°C (248°F) and
pressures of up to 5 bar (72.5 psi).
CDE-1202 incorporates an NTC sensor. A
10 m (33') cable is an available option.
®
, which resists the
E-27
E-27
For Sales
& Service
U.S.A. and Canada
Specifications
Range:
CDTX1201: 0.0 to 199.9 mS/cm
CDTX1202: 0.00 to 19.99 mS/cm
CDTX1203: 0 to 1999 µS/cm
CDTX1204: 0.0 to 199.9 µS/cm
Resolution:
CDTX1201: 0.1 mS/cm
CDTX1202: 0.01 mS/cm
CDTX1203: 1 µS/cm
CDTX1204: 0.1 µS/cm
Accuracy: ±2% FS
Calibration: 4 mA (zero) ±0.8 mA
Slope: ±1.6 mA (90 to 110%)
Temperature Compensation:
Automatic from 0 to 50°C (32 to 122°F) with a ß of 2%
Output: 4 to 20 mA non-isolated; max 500 Ω sinking signal
Power:
Without LCD: 15 to 30 Vdc
With LCD: 20 to 36 Vdc
Probe: CDE-1201 or CDE-1202 (not included)
Cable Length: 3 m (10')
Protection: NEMA 4 (IP65)
Operating Environment: 0 to 50°C (32 to 122°F)
0 to 95% RH (non-condensing)
Dimensions: 165 L x 110 W x 90 mm H (6.5 x 4.3 x 3.5")
Weight: 1 kg (2.2 lb)
Probe Specifications
Temp Compensation:
CDE1201 and CDE1202:
ATC 0 to 50°C (32 to 122°F)
Body Material:
CDE-1201: Polypropylene
CDE-1202: Ultem
Working Temperature:
CDE-1201: 0 to 80°C (32 to 176°F)
CDE-1202: 0 to 120°C (32 to 248°F)
Max Pressure [@25°C(77°F)]:
CDE-1201 and CDE-1202: 5 bar (72.5 psi)
Cable Length: 3 m (10')
®
CDTX1201-D,
shown smaller
than actual size.
CDTX1201, shown
smaller than actual size.
Dimensions: mm (in)
90
(3.54)
Order Online
Over 100,000 Products Available!
24 (0.94)
45
(1.77)
4 x 4.3
(4 x 0.17)
165 (6.50)
126 (5.00)
58 (2.2)
100
(3.94)
7.01
155 (6.10)
®
110
(4.33)
E-28
E-28