Omega Products CDE-45T1 Installation Manual

Page 1
T
PRODUCT INSTRUCTION SHEET
CDE-45T1 and CDE-45T3 Toroidal Conductivity Sensors
Introduction
®
®
Modern conductivity measurement devices include a class of sensors which use an electrodeless system to determine the conductivity of a fluid. Electrodeless probes function by first exciting a toroid with an AC voltage to produce a magnetic flux in the core. This flux then produces an alternating current flow in the fluid, which passes as a loop through the center of the core. A second toroid then acts as a sensor by using the fluid’s current to produce a magnetic flux in its core which, in turn creates current flow in its windings. The current flow in the fluid and subsequently the detector toroid’s windings, depends on the conductivity of the solution, thereby giving an accurate measurement without contacting the fluid directly.
Your Electrodeless Conductivity Sensor provides a very linear voltage output signal versus the conductivity of the solution within which it is placed. The overall output is proportional to the excitation signal driven into the driver toroid, the amount of gain provided by the output signal amplifier, and the conductivity of the medium being measured. Thus, calibra­tion must be done with a known conductivity solution and the electronics.
Mechanical Installation
Your toroidal sensor can be installed in 2 ways:
FIG. 1
FIG. 1
FIG. 2
O-RING SEAL
FLOW DIRECTION ARROWS x 2
OWARD SENSOR
ALIGNMENT TAB)
(POINT
SENSOR ALIGNMENT NOTCH
SENSOR ALIGNMENT TAB
1. Submersion mount: Mount sensor into rigid or flexible conduit using a 3/4" NPT coupling and attaching to the 3/4" NPT threads near the sensor's cable. Be sure to seal conduit to avoid fluid build up in conduit.
2. In-Line mount: a. Plumb flow cell CDTX-45T-PT into line. Flow cell is provided
FIG. 3
with 2" slip fittings. For threaded connection install a 2" slip x NPT female adapter (CPVC SCH 80).
b. Install o-ring into CDTX-45T-PT as shown in FIG. 1
NUT
c. Install CD-45T1 toroidal sensor into CDTX-45T-PT. Match notch on flange of sensor with tab of flow cell as show in FIG. 2.
d. Install nut onto threads of flow cell and turn clockwise. DO NOT USE TEFLON TAPE OR SEALANT ON THREADS!
CDE-45 Page 1 of 4
Page 2
PRODUCT INSTRUCTION SHEET
Sensor Wiring
®
The probe has 9 total leads (see FIG. 4).
The driver toroid is connected to the red and black leads with the red shrink tubing around them. When using the probe with various controllers, it is important that the red lead be connected to an AC DRIVER or VOLTAGE IN position while the black lead is connected to a GROUND or DRIVER RETURN line.
The detector toroid is connected to the white and black leads with the white shrink tubing around them. The white lead is typically connected to the SIGNAL IN location and the black lead is connected to a GROUND or SIGNAL RETURN line.
The temperature element is connected to the green and black leads with the green shrink tubing around them. Connect these to the TEMPERATURE INPUT locations. The polarity is not important.
The 3 bare leads, which supply individual pair shielding throughout the cable, should typically be connected to a ground or drain line. Note: The shield leads are not connected to one another, they are all individual shields for the three bundles of wires.
Refer to your controller manual for specific wiring details.
FIG. 4
CDE-45T1
Cable Considerations
The cable uses PVC to protect the wires during use. If the cable comes in contact with the working fluid, the temperature and pressure ratings must be adjusted to allow for the lower tem­perature limits of the cable.
C (DRY) but should only be subjected to 70 deg C when im­mersed in a fluid.
by liquid (raising the possibility of shorting) and all attempts should be made to keep the cable out of the fluid environ­ment. Do not run cable in the same conduit with any other A.C. power wiring nor routed close to any high current demanding equipment. Seal conduit to avoid build-up of moisture. Do not cut cable. Shorter or longer cables can be provided.
The jacket becomes significantly weakened
The cable can withstand 105 deg
CDE-45 Page 2 of 4
Page 3
)
PRODUCT INSTRUCTION SHEET
Sensor Calibration
®
After the sensor is properly connected, the system should be calibrated. The typical calibration procedure uses a “low” and a “high” known standard conductivity solution. The "low" solution is often DI water or air and is used to calibrate the zero point of the controller. Plotting these two points will cre­ate a straight line, which can be used to find the conductivity
valueof any solution in the range. Make sure probe is im­mersed in the calibration fluid such that the toroids are totally submerged. A sample plot is provided in FIG. 5 to show the approximate values, which will be encountered during calibra­tion.
Note: The values in the plot are arbitrary, each probe will
require unique linear equation values.
If the sensor is to be used in a submersion application, cali­brate the sensor in a large glass or plastic beaker with all sides of the sensor at least one inch away from the wall (FIG. 6). If the production installation is a pipe (plastic or metal), calibration should be performed in a similar pipe arrangement. All Electrodeless (Toroidal) sensors have a wall effect, which must be taken into account during calibration. If the non-con­ductive (plastic) wall is within 1 inch of the sensor, the sensor’s reading will be reduced due to the insulator interaction with the current path. If the sensor is within 1 inch of a conductive (metal) wall, the sensor’s reading will be increased due to the shorting effect of the conducting wall. These wall effects can be calibrated out of the system by simulating the application’s mounting configuration. A plot showing the effects of insulat­ing and conducting walls on the output can be seen in FIG.7. During calibration and production installation (especially in a submersion environment), it is important to dislodge any air bubbles, paying special attention to the center hole of the toroids. Also make sure toroids are totally covered with fluid when calibrating (FIG. 6).
Plot of Sensorex DC Output Versus Conventional
FIG. 5
DC Output vs. Conventional Conductivity Meter
20
15
10
(mS)
5
0
Conductivity Meter Output
Conductivity Meter Output
123
Sensorex Output (V)
FIG. 6
> 1”
> 1”
Series1 Linear (Series1
y = 6.0498x - 5.435
2
= 1
R
Liquid must totally cover toroids (be above this line)
Please refer to your controller's manual for specific calibration instructions.
Sensor Maintenance
The major advantage of the Electrodeless (Toroidal) Sensor is almost no maintenance is required. The only maintenance required during normal operational life of the sensor is to prevent the toroidal opening from being plugged with debris. Use a soft brush or rag to remove any debris in the core open­ing. If that does not work, try a mild detergent or weak acid (5-10% HCl).
Plot of Normalized Output as a Function of Probe
FIG. 7
2.55
2.5
2.45
2.4
2.35
2.3
2.25
2.2
Normalized Output (V/V)
2.15 024681012
Distance from Wall
Wall Distance (In)
CDE-45 Page 3 of 4
Conducting wall Insulating Wall
Page 4
PRODUCT INSTRUCTION SHEET
®
Sensor Troubleshooting
If your sensor is not reading as expected check resistance of leads as shown in FIG. 8. Check temperature leads as shown in FiG 9.
Sensor Specifications
Installation Type: Submersion via 3/4" NPT or in-line via CDTX-45T-PT flow cell
Conductivity Range: 500-2,000,000 uS
Maximum Temperature: Electrode:: 105 deg C Flow cell: 100C Cable:: 105 deg C (Dry) 70 deg C (wet)
Maximum Pressure: 100 psig
Wetted Materials: Sensor: Polypropylene (black) Flow cell: CPVC O-ring: Viton
Temperature Compensation: via Pt1000 RTD
FIG. 8
SHRINK COLOR WIRE COLOR RESISTANCE
RED RED 1 TO 2 MEGOHMS RED BLACK >20 MEGOHMS (OPEN) RED CLEAR
WHITE WHITE 1 TO 2 MEGOHMS WHITE BLACK > 20 MEGOHMS (OPEN) WHITE CLEAR
GREEN GREEN 109 or 1090 OHMS** GREEN BLACK < 2 OHMS GREEN CLEAR
0.5
0.5
>20 MEGOHMS (OPEN)
FIG. 9
RTD TYPE TEMP. (deg C/deg F) RESISTANCE(Ohms)
1000Ohm RTD 18/64.4 1069 1000 Ohm RTD 19/66.2 1073 1000Ohm RTD 20/68 1077 1000 Ohm RTD 21/69.8 1081 1000 Ohm RTD 22/71.6 1084 1000 Ohm RTD 23/73.4 1089 1000 Ohm RTD 24/75.2 1092 1000 Ohm RTD 25/77.0 1096 1000 Ohm RTD 26/78.8 1100 1000 Ohm RTD 27/80.6 1104 * * *
Note: There is approximately 3.9Ohm/ Deg C change for Pt1000RTD.
CDE-45 Page 4 of 4
Loading...