Omega Products CA450 User Manual

Process Multimeter
Loop Power, 4 to 20 mA Digital Multimeter Output Function
CA450
U 4 to 20 mA Step,
0 to 100% Toggle, Auto Step, Sweep Output
U Transmitter Simulate
by SINK Function
U HART® Mode Setting
with Loop Power (Adds 250Ω Resistor)
U 24V Loop Power
Supply and mA Measurement Available Simultaneously
U High Accuracy
Measurement of DC mA
0.05%, 30.000 mA
U DCS Power Supply
Crest Value Measurement by Peak Hold Function
U Direct Reading (ACA/
DCA) of Various Signals by Sensor Mode
U Data Transmission by
PC Communication with IR-USB*
CA450 shown smaller than actual size.
The OMEGA® CA450 Process Multimeter is a portable test instrument to combine a digital RMS multimeter and a loop calibrator to give process technicians the same functionality in one tool. The model CA450 provides 24V loop power, reducing the need for carrying a separate power supply when performing transmitter testing. And with its built-in, selectable 250Ω HART resistor, it also eliminates the need for a separate resistor. Process technicians can now do the same job while carrying less equipment.
* For details visit omega.com/ir-usb
Q
CA450 shown calibrating a tunable diode laser during a final inspection test.
Q-1
SM
OMEGACARESM extended warranty program is available for models shown on this page. Ask your sales respresentative
Extended Warranty
Program
for full details when placing an order. OMEGACARESM covers parts, labor and equivalent loaners.
Input Range Accuracy
0.1 to 600.0 mV 0.09% +2
0.001 to 6.000V 0.09% +1
DCV
0.01 to 60.00V 0.09% +1
0.1 to 600.0V 0.09% +1
0 to 1000V 0.1% +1
0.001 to 30.000 mA 0.05% +2
DCA
0.01 to 100.00 mA 0.05% +2
Clamp 0.1 to 180A 1.2% +0.4 mV
50/60 Hz
0.1 to 600.0 mV 0.5% +5
0.001 to 6.000V 0.5% +5
0.01 to 60.00V 0.5% +5
ACV
0.1 to 600.0V 0.5% +5
0 to 1000V 0.5% +5
50/60 Hz 0.5% +5
500 Hz 1.0% +5
1 kHz 1.5% +5
ACA with Clamp Probe
Clamp 1 0.1 to 130.0A 0.5% +DCV accuracy
Clamp 2 0.1 to 500.0A 1.0% +DCV accuracy
ACA
Clamp 3 0.1 to 3000A 1.2% +0.4 mV @ 50/60 Hz
Clamp 1 0.5% +DCV accuracy
Clamp 2 1.0% +DCV accuracy
Clamp 3 1.2% +0.4 mV @ 50/60 Hz
0.1 to 600.0Ω 0.2% +2
0.001 to 6.000 kΩ 0.2% +1
Resistance
0.01 to 60.00 kΩ 0.2% +1
0.1 to 600.0 kΩ 0.2% +1
0.001 to 6.000 MΩ 0.35% +3
0.01 to 60.00 MΩ 1% +2
0.01 to 199.99 Hz 0.005% +1
Frequency
0.1 to 1999.9 Hz 0.005% +1
0.001 to 19.999 KHz 0.005% +1
Diode Test 0.001 to 2V 1% +2
Peak Hold DCV ±100 digit >6 ms
DC mA
SOURCE
0.001 to 25.000 mA 0.05% of range
20 mA SINK 0.001 to 25.000 mA 0.05% of range
24V
LOOP POWER
24V 20 mA @ 24V
Q-2