Omega OSP, OSP-500, OSP-800, OSP-1000, OSP-1300 User Manual

...
http://www.omega.com
e-mail: info@omega.com
OSP SERIES
Portable IR thermometers
2
OMEGAnetSM On-Line Service http://www.omega.com
Internet e-mail
info@omega.com
Servicing North America:
USA:
ISO 9001 Certified
One Omega Drive, Box 4047 Stamford, CT 06907-0047 Tel: (203) 359-1660 e-mail: info@omega.com
FAX: (203) 359-7700
Canada:
976 Bergar Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 e-mail: info@omega.ca
FAX: (514) 856-6886
For immediate tech nic al or ap pl ic at i on as sis t a nce :
Usa and Canada:
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
SM
Customer Service: 1-800-622-2378 / 1-800-622-BEST
SM
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
SM
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico and Latin America:
Tel: (95) 800-826-6342 En Español: (95) 203-359-7803 e-mail: espanol@omega.com
FAX: (95) 203-359-7807
Servicing Europe:
Benelux:
Postbus 8034, 1180 LA Amstelveen, The Netherlands Tel: (31) 20 6418405 Toll Free in B e nelux: 0800 0993344 e-mail: nl@omega.com
FAX: (31) 20 6434643
Czech Republic:
ul. Rude armady 1868, 733 01 Karvina-Hranice, Czech Republic Tel: 420 (69) 6311899 Toll free: 0800-1-66342 e-mail: czech@omega.com
FAX: 420 (69) 6311114
France:
9, rue Denis Papin, 78190 Trappes Tel: (33) 130-621-400 Toll Free in France: 0800-4-06342 e-mail: france@omega.com
FAX: (33) 130-699-120
Germany/Austria:
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Tel: 49 (07056) 3017 Toll Free in Germany: 0130 11 21 66 e-mail: info@omega.de
FAX: 49 (07056) 8540
United Kingdom:
ISO 9002 Certified
One Omega Drive , River Bend Technology Centre Northbank, Irlam, Manchester M44 5EX, England Tel: 44 (161) 777-6611 Toll Free in United Kingdom: 0800-488-488 e-mail: info@omega.co.uk
FAX: 44 (161) 777-6622
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be corrected but OMEGA Engineering Inc. accepts no liability for any errors it
contains, and reserves the right to alter specifications without notice.
WARNING
: These products are not designed for use in, and should not be used for, patient connected applications.
3
INTRODUCTORY NOTE
This user’s guide contains operating instructions, as well as a description of the principles of operation, of the OSP Series portable IR thermometers. This information covers all models of the instrument, including the basic equipment and its options and accessories. This manual is a complete “USER GUID E”, providing ste p-by-step i nstructions to operate th e instrument in each of its designed functions.
OMEGA has used care and effort in preparing this guide and believes the information in this publication to be accurate. OMEGA products are subjected to continuous improvement, in order to pur sue the technol ogical leadership; these improvements could require changes to the information of this guide. OMEGA reserves the right to change such information without notice. No part of this document may be stored in a retrieval system, or transmitted in any form, electronic or mechanical, without prior written permission of OMEGA Engineering Inc.
OSP portable IR thermometers use sophisticated analogic and digital technologies. Any maintenance operation must be carried out by qualified personne l
ONLY
. We recommend you c ontact our technicians f or any support requirements. When Ni-MH rechargea ble batteries ar e ordered, th e unit can be pow ered als o by 11 5V
±
10% 50/60Hz line supply using the special power supply module provided. OSP is fully tested in conformity with the directive n°89/336/CEE Electromagnetic Compatibility. OMEGA shall not be liable in any event, for technical or publishing error or omissions, or for any incidental and consequential damages, in connection with, or arising out of the use of this guide.
IMPORTANT :
B
EFORE USING THE
OSP
FOR THE FIR ST TIME, GO OVER THESE OPERATING INSTRUCTIONS CAREFULLY AND KEEP
THEM AVAIABLE FOR FUTURE REFERENCE
.
Danger and Certification Labels
Labels Location – refer to section 8.2 Warnings and Cautions – refer to section 8.1
4
TABLE OF CONTENTS
1 GENERAL DESCRIPTION...................................................................................................................6
1.1 Instrument codes..............................................................................................................................................7
1.2 Specifications ...................................................................................................................................................8
2 GENERAL FEATURES ........................................................................................................................9
2.1 General.............................................................................................................................................................9
2.2 Optical System .................................................................................................................................................9
2.3 Target pinpointing.............................................................................................................................................9
2.4 Taking measurements......................................................................................................................................9
2.5 Keyboard ........................................................................................................................................................10
2.6 Display............................................................................................................................................................10
2.7 Digital Interface...............................................................................................................................................10
2.8 Self calibration................................................................................................................................................10
2.9 Thermocouple Input........................................................................................................................................10
2.10 Analog Output ................................................................................................................................................10
2.11 Logging Mode.................................................................................................................................................11
2.12 Calculated Measurements..............................................................................................................................11
2.13 Case ...............................................................................................................................................................11
2.14 Logging Data Manager...................................................................................................................................11
3 PHYSICAL DESCRIPTION ................................................................................................................12
4 FUNCTIONAL DESCRIPTION ...........................................................................................................13
4.1 Power supply..................................................................................................................................................13
4.2 Keyboard ........................................................................................................................................................13
4.3 Microcontroller................................................................................................................................................14
4.4 Firmware ........................................................................................................................................................14
4.5 Display............................................................................................................................................................14
4.6 Battery charger...............................................................................................................................................14
4.7 Digital interface...............................................................................................................................................14
5 UNPACKING ......................................................................................................................................15
6 PRE-OPERATIONAL CHECK ...........................................................................................................16
7 POWER SUPPLY...............................................................................................................................17
7.1 Rechargeable batteries ..................................................................................................................................17
7.1.1 How to maximize the life span of the battery .............................................................................................17
7.2 Power supply with alkaline batteries...............................................................................................................17
7.3 Power supply from main line AC.....................................................................................................................17
8 WARNINGS & CAUTIONS.................................................................................................................18
8.1 Laser Sight .....................................................................................................................................................18
8.2 Analogue input................................................................................................................................................18
8.3 Danger and Certification Labels .....................................................................................................................18
9 OPERATIONS ....................................................................................................................................20
9.1 Quick Start......................................................................................................................................................20
9.2 How to Operate the instrument.......................................................................................................................20
9.3 Hold ................................................................................................................................................................21
9.4 Laser sighting .................................................................................................................................................21
9.5 Unstable temperature measurement ..............................................................................................................22
9.6 Alarm settings.................................................................................................................................................23
9.7 Thermocouple input........................................................................................................................................24
9.8 Minimum, maximum, average and difference indication.................................................................................25
9.9 Automatic Emissivity Setting ..........................................................................................................................26
10 CONFIGURATION..............................................................................................................................27
10.1 Technical Unit selection..................................................................................................................................27
10.2 Thermocouple input enabled..........................................................................................................................27
10.3 Temperature Scale selection..........................................................................................................................27
10.4 Date & Real Time clock setting ......................................................................................................................28
10.5 Acquisition settings.........................................................................................................................................28
10.6 Buzzer On/Off.................................................................................................................................................29
10.7 OSP 500/800 decimal point selection.............................................................................................................29
10.8 Ambient Temperature Compensation (TAM)..................................................................................................29
10.9 Firmware revision number..............................................................................................................................30
10.10 Instrument serial number...........................................................................................................................30
10.11 Battery level of charge ...............................................................................................................................31
5
11 DATA ACQUISITION OPERATIVE MODE........................................................................................ 32
11.1 How to operate in data acquisition mode........................................................................................................32
11.2 Recall stored data...........................................................................................................................................34
12 LOGGING DATA MANAGER ............................................................................................................35
12.1 Installation ......................................................................................................................................................35
12.2 Program Architecture.................................................................................................................................35
12.2.1 Toolbars.....................................................................................................................................................37
12.3 Quick starting .................................................................................................................................................38
13 OPTIONS & ACCESSORIES .............................................................................................................42
13.1 Printer operations ...........................................................................................................................................42
13.2 Sighting telescope system..............................................................................................................................42
13.3 Red Point pinpointing system .........................................................................................................................43
14 APPLICATION NOTES ......................................................................................................................45
14.1 Infrared energy ...............................................................................................................................................45
14.2 Applications ....................................................................................................................................................45
14.3 Emissivity...................................................................................................................................................45
14.4 Reflected energy compensation.....................................................................................................................45
15 DIGITAL INTERFACE........................................................................................................................46
15.1 Digital output wiring practice...........................................................................................................................46
15.2 TTL to RS 232 adapter...................................................................................................................................46
15.3 Communication protocol.................................................................................................................................47
15.3.1 Computer data request from OSP to PC ...................................................................................................47
15.3.2 Computer data setting from PC to OSP.....................................................................................................49
15.3.3 Communication programs..........................................................................................................................50
16 MAINTENANCE..................................................................................................................................52
16.1 Faulty operating conditions.............................................................................................................................52
16.2 Storage...........................................................................................................................................................52
APPENDIX.......................................................................................................................................................53
A1 How to determine an object emissivity ...........................................................................................................54
A1.1 Typical Emissivity Values ..........................................................................................................................54
A1.2 Metals - Typical Emissivity Values.............................................................................................................55
A1.3 Non-metals - Typical Emissivity Values .....................................................................................................56
6
1 GENERAL DESCRIPTION
Portable infrared thermometers measure surface temperature without touching the surface. They collect the infrared energy radiated by a target and compute its surface temperature.
OMEGA OSP Series
are portable infrared temperature devices designed to ease maintenance operations. They also enable the monitoring of operating temperatures of mechanical and electrical plants or production equipment without removing the equipment from the service. Any temperature variation noticeable in whatever kind of system device should show critical overheating conditions which may cause malfunctioning to the device itself. They are also useful to measure product temperatures during manufacturing, to spot problems before they reduce quality or cause production downtime. Their principle of operation is very simple, the instrument determines the temperature of an object by measuring the amount of radiant energy emitted by it. The detector, installed on each
OSP
unit, responds by producing a voltage signal which is directly proportional to the amount of energy received, and therefore is a function of the temperature of the target. By sampling and manipulating the output of the detector, the microcontroller-based electronics can display the temperature and the related computed values such as maximum, minimum, average, and difference temperatures seen during the measurement. The same information can also be stored and processed via Personal Computer (using electronic spreadsheets e.g. Lotus, Excel, Paradox, etc.). The displayed value can be printed out directly into an external printer supplied on request. Thanks to its advanced optical system and electronics,
OSP
can be very useful for objects 5mm and larger and can operate in critical ambient conditions. To take measurements by using
OSP
thermometers just pull the trigger and then point at the target to be measured. The trigger can be locked on if desired. For more distant targets, hold the unit at arm length and use the sighting notch and post to aim. This will give a parallel and offset sighting. Also Laser beam is available for sighting target. Temperature information as well as auxiliary parameter pieces of information are shown on the high contrast liquid crystal display (LCD). The case, made in shock-resistant ABS + polycarbonate, is ergonomically designed for an easy practical use. The instrument is powered by a group of four alkaline or Ni-MH rechargeable batteries (AA type); an external battery charger module is supplied as a standard accessory.
OSP
portable IR thermometers have been developed using the most advanced microcontroller technique to provide high accuracy on extended ranges and a powerful operating flexibility. The calibration uses computerized procedures and the relevant calibration data are memory stored to ensure high accuracy.
General features OSP
non contact infrared thermometers are designed to simplify temperature measurements at distance from a target and to identify hot spots, which normally means anomalous operative conditions avoiding costly downtime or processing problems that lead to rejected products. There is no need to focus as required by photographic equipment. Accurate measurements result if
OSP
field of view diameter is smaller than the target.
Innovative design
This new line of portable instruments represents the most versatile and powerful infrared temperature system. The most appropriate aiming system at the application can be selected with a twin laser pinpointing or with a combination of twin lasers and telescope or red point for true universal applications. An on board data logger is available to store up to 500 data points, an analogue output can drive directly a recorder or a controller, The real-time temperature can be displayed together with the Maximum, Minimum, Difference and Average values. Each model incorporates the following :
high quality optical system
high sensitivity infrared radiation detector
auxiliary input for thermocouple type K and S
microcontroller circuit
high contrast LCD display with backlight device
extended life traditional microswitch keyboard
four alkaline or rechargeable Ni-MH type AA batteries
external module for battery charge and/or line operation
analogue output
digital interface
7
1.1 Instrument codes
To Order
Model No. Temperature
Range
Spectral
Response
Target Spot Size
@ Distance **
D/S Optical Ratio
OSP500-(*)
–30 °C to +930 °C
-20°F to +1706°F
8 to 14 µm
10mm @ 600mm
0.39 @ 24”
60:1
OSP800-(*)
–30 °C to +930 °C
-20°F to +1610°F
8 to 14 µm
10mm @ 600mm
0.39 @ 24”
60:1
OSP1000-(*)
–30 °C to +1000 °C
-20°F to +1830°F
8 to 14 µm
10mm @ 1000mm
0.39 @ 39.4”
100:1
OSP1300-(*)
+300 °C to +1300 °C
+570°F to 2370°F
1.6 µm
6mm @ 1000mm
0.236 @ 39.4”
167:1
OSP1600-(*)
+600 °C to +1600 °C
+1120°F to 2900°F
1 µm
5mm @ 1000mm
0.20 @ 39.4”
200:1
OSP2000-(*)
+600 °C to +2000 °C
+1120°F to +3630°F
1 µm
5mm @ 1000mm
0.20 @ 39.4”
200:1
* Insert suffix codes B and C and D from options chart above ** See Target spot size diagram
Note
: all accessories and options must be specified at time of order.
Each Pyrometer supplied with
: Alkaline battery, laser, operators manual, and calibration report
Options Suffix Code Description OSP500 OSP800 OSP1000 OSP1300 OSP1600 OSP2000
-B0
Alkaline battery std std std std std std
-B1
Rechargeable Battery Opt Opt Opt Opt Opt Opt
-C1
Single Laser Std No No No No No
-C2
Twin crossed Lasers No Std Std Std Std Std
-C3
Twin normal Lasers No Std Std std Std Std
-C4
Single Laser with Circle Std No No No No No
-D1
Vinyl Case Std Std No std Std No
-D2
Aluminum Case Opt Opt Std Opt Opt Std
-D3
Memory+Software+RS232 Cable Opt Opt Std Opt Opt Std
-D6
Sighting telescope No No Opt Opt Opt Opt
-D7
Red Point No No No Opt Opt Opt
-D8
Alarms Std Std Std std std Std
-D9
Analog Output Opt Std std std std Std
-DA
Aux Input for Tc K & S Opt std std std std Std Report of Calibration std std std std std Std
No = Not available Std = Standard at no extra cost
Single laser
: General application, target center
Single laser with circle
: General application, target center plus target dimension with circle (short distance)
Twin crossed lasers
: General application, target dimension
Twin normal lasers
(parallel): Long distance application over 5m (16’)
Red Point
: High temperature over 1250°C 1x
Sighting Telescope
: High temperature over 1250°C 2x tele
Software
(D3): Includes memory up to 500 values. You can divide them by Tag (20 different). Each value includes
temperature value and date&time (built in real time clock). Tag name with 7 characters on instrument display. Datalogging manual or automatic with sampling time from 1 to 999 sec. Windows 95 software to download data to PC and cable. You can view, save, print, export to Excel and graph data.
U.S. and international Patents and Patents Pending.
8
1.2 Specifications
Target dimensions at different distances
Nominal target diameter at 95% energy
• Temperature measuring ranges:
-30 to 2000°C (-22 to 3632°F)
Thermocouple measuring ranges:
type K :
-100°C to +1370°C (-148 to 2500°F) 0.1°C resolution
type S :
0°C to +1760°C (-32 to 3200°F) 0.1°C resolution
Spectral response:
OSP 500, 800 & 1000 :
8 - 14 µm
OSP 1300 :
1.6 µm
OSP 1600 & 2000
0.9 µm
Accuracy:
OSP 500, 800 & 1000:
±(1% of the reading +1°C/2°F)
OSP 1300, 1600 & 2000:
±(0.5% of the reading +1°C/2°F)
• Repeatability IR:
OSP 500, P800 & P1000 :
±0.5% of the reading
OSP 1300, P1600 & P2000:
±0.25% of the reading
• Temperature stability IR:
for the band exceeding +18°C to +28°C:
±0.01% of f.s./°C
Display:
High contrast custom LCD with backlight device
Display resolution:
1°C / °F / K (0.1°C/0.1°F in AVG mode up to 200°C)
Alarms:
Acoustic and visible
Measurement sampling time:
<300 ms
• Emissivity:
Adjustable from 0.10 to 1.00 in 0.01 steps
Analog output:
1 mV/ °C/ °F/ K
Digital interface:
full bi-directional TTL level port Optional TTL to RS232 adapter
• Calculated functions:
Hold, average, max, min, ∆T
Data memory:
500 input data structured by tag
Calibration:
self learning technique with automatic procedure
• Laser:
Wavelength
: 650nm
Maximum optical power
: <1mW
FDA Classification
: Class II, Complies with 21CFR Chapter 1,
Subchapter J
Safety classification
: Class 2
Beam diameter
: 3mm
Beam divergence
: <0.5mrad
Laser indicator
: asterisk on display
Power supply:
alkaline or rechargeable type AA battery
Battery life:
16 h (with backlight off)
• Battery low level of charge:
symbol on the LCD display
• Line operation:
100, 115, 230 Vac through the external charger
Charger transformer insulation:
2500 V
Battery recharging time:
8 h at 90% (instrument switched off)
• Operating environment temperature range:
from -10°C to +55°C (-14 to 131°F)
• Storage temperature range:
from -30 °C to +60 °C (-22 to 140°F) without battery
• Case:
Injection moulded ABS+ policarbonate
• Dimensions:
200 x 180 x 80 mm (7.87 x 7.09 x 3.15”)
Weights:
net 0.8 kg (1.76 lb); gross 1.5 kg (3.3 lb)
9
2 GENERAL FEATURES
2.1 General
Temperature measurements of a liquid or gaseous compound have been successfully made with thermoelectric or expansion thermometers thanks to the good thermal exchange of the sensor with the fluid. With solid bodies a good thermal exchange is difficult to be obtained and an additional measuring error should be considered. A direct contact measurement is impractical when the object being measured is moving, it cannot be touched with a thermoelectric sensor because of electrical hazard or of other reasons. A non-contact IR temperature measurement is the best solution to the above application problems. Also other applications benefit because non-contact thermometers do not add or remove heat or disturb the process in any way, and there is no wear and tare on the instrument.
2.2 Optical System
The optical system is equipped with a main objective to focus the infrared energy into the infrared detector through a single lens, a filter and a field stop which defines the visual cone and therefore the target dimensions. As the detector is placed in the focal point of the measurement portion of the optical system no focusing operations are required. A secondary, but extremely important objective, is the correct pinpointing of the target as described below. The optical diagrams are shown in the previous specification section.
2.3 Target pinpointing
A correct aiming at the target is an important factor of a non-contact thermometer. The immediate evaluation of the smallest measurable target area is also a positive key factor in many applications. The following types of pinpointing are available :
A “V” groove on the up side of the instrument can be used stretching the arm.
Twin laser pointers to define, at distance, the measurable target dimensions. Two versions are available with crossing
or parallel twin lasers. The twin crossed lasers are suggested for target distance upper than 5 meters.
Combination of twin lasers and direct telescope viewing for universal applications mainly required for a high distance
targets (more than 10 meters) and high temperature targets (more than 1000 °C) where laser spots are not visible .
Combination of twin lasers and a red point pinpointing for universal applications mainly required for high temperature
target where laser spots are not visible
With a laser pinpointing the operation of the instrument is extremely simplified. You simply aim at the target and read the temperature. The installation procedures for telescope and red point pinpointing, please refers to the dedicated chapters.
2.4 Taking measurements
OSP
portable thermometers are accurate, rugged and compact. Its modern and practical design allows an easy handling and aiming, either at arm length or using a tripod mount, to obtain current temperature value, average, minimum, maximum and difference temperature measurement values. Its analogue to digital outputs allows a continuous documentation on a continuous trace recorder or on a serial printer.
Using the keyboard it is also possible to enable the instrument to measure one of the computed values in addition to the actual temperature value measurement:
MAX maximum temperature recorded during current measurement MIN minimum temperature recorded during current measurement DIF difference between MAX and MlN AVG average temperature for entire measurement Tamb ambient temperature
10
2.5 Keyboard
Traditional metal-click switches, with a working life of one million operations, are used to enter the operator’s instructions. The contact closure of the keyboard keys is acknowledged, as a coded signal, directly by the microcontroller.
2.6 Display
The high contrast customised alphanumeric LCD display indicates the measured temperature value. It is also used for operators’ messages, instrument configuration set-up, special operative modes, etc. It is equipped with a backlight device to allow easy readings even in poor light conditions.
2.7 Digital Interface
A digital interface with TTL logic levels is available for communication with external units. A serial data port provides communication capability at a logic level of 0-5 V (four wires: Tx, Rx, GND, Vcc). A special software set enables the transfer of all the recorded data on a Personal Computer for further statistical analysis.
2.8 Self calibration
The hardware-firmware design allows the automatic calibration of the instrument. The calibration procedure is protected by a security code.
2.9 Thermocouple Input
To extend the operative capability the instrument is equipped with an auxiliary input for thermocouple. This feature, when used with a calibrated thermocouple, can be used to obtain accurate temperature measurements eliminating the problems of emissivity and the interfering light. The thermocouple can be used to obtain an accurate temperature reading of the target material and then these data are used to determine the compensation value for the actual emissivity, including the interfering light, within a range from
0.10 to 1.00.
2.10 Analog Output
An analog output (1 mV/°C/°F/K) is available to drive an external analog input device (eg. a potentiometric or hybrid recorder) for long term trend profiling and tests.
11
In this case the instrument can be powered directly from mains using the external power supply module supplied with the thermometer.
2.11 Logging Mode
The instrument can be equipped with an internal memory to store up to 500 input data. Two types of data acquisitions can be selected by the user.
Continuous acquisition
The operator can select the interval time between each acquisition and store progressively the input data in the Tag file or can manually, step by step, give the acquisition instruction.
Acquisition by dedicated tag
Standard Agencies and Quality Auditors require the collection, organization and availability of traceability documents. A supporting software is available to transfer a selection of plant inspections from a PC to the internal memory of the instrument in order to simplify field check and select the appropriate tag number. Dedicated input data are memory stored and downloaded into a PC to document the inspection activity. Data can be saved on disks, viewed and printed in a numeric or graphic mode.
2.12 Calculated Measurements
For the measurement of unstable input signals by a progressive averaging with a programmable average weight. In addition the operator can select average, minimum, maximum, Tmax -Tmin, differential.
2.13 Case
The case of the instrument is ergonomically designed for an easy hand held operation and transport. The body is injection molded, shock-resistant, flame proof ABS+ polycarbonate. A vinyl case with shoulder strap and an aluminum case for instrument + printer + accessories are available on request.
2.14 Logging Data Manager
LogMan software allows the OSP series users to set and prepare the infrared thermometers to acquire the data organized by ‘Tag’. This software also allows to download data from Instrument on document (table) with date, time and value. You can view table, save table, print table, export table in excel-txt-html and obtain graph. You can also manage data coping, moving on other tables.
12
3 PHYSICAL DESCRIPTION
OSP
portable IR thermometers consist of a rugged, compact and self-extinguishing case ergonomically designed for an easy practical use. The instrument can be supplied either with a vinyl protective case with shoulder strap or with an aluminum case to assure better protection against mechanical knocks or scratches. The battery container is located on the lower part of the handle, and is accessible through a cover fastened by a metal screw.
13
4 FUNCTIONAL DESCRIPTION
OSP series infrared portable thermometers block diagram is shown below:
Measuring
ob
j
ect
Emitted IR energy
Optical system
Filter
Detector
Amplifier
A/D
converter
Micro-
controller
LCD
display
Tc
Auxiliary
input
Keyboard Digital
interface
Analog Output
Optical system
Filter
Detector
Amplifier
A/D converter
Microcontroller
LCD display
Digital interface
Keyboard
Tc auxiliary input (optional)
Analog output (optional)
The radiated infrared energy is focused by the optical system on an infrared detector sensitive to the required spectral band. The infrared detector generates a signal proportional to the energy received corresponding to the temperature and emissivity of the target. See Appendix A for emissivity values. The output signal of the sensor is then conditioned, converted from analog to digital and transmitted to the microcontroller. The procedures used to process all the operating functions as well as measure and calculation routines are stored into the microcontroller memory. Any operator’s instruction, through the trigger and the keyboard is directly recognized, as a coded signal, by the microcontroller. The actual temperature value and the active operative mode are indicated on the LCD display.
4.1 Power supply
The instrument is powered, if not otherwise specified with the order, by four internal alkaline batteries or rechargeable Ni­MH type AA (nominal voltage 1.25 V) that can be recharged through an external charger module supplied as a standard accessory.
4.2 Keyboard
The operative keyboard is designed with a traditional single microswitch per button for long life and high reliability. The contact closure of the keys is acknowledged as a coded signal by the microprocessor that recognizes the operator’s instructions.
14
4.3 Microcontroller
The microcontroller handles all the logic functions of the instrument, performs the linearization for non linear transducers, compensates for the reference junction temperature, drives the digital display and acknowledges all the operator’s instructions. The core of the circuit is a single-chip microcomputer that utilizes HCMOS technology to provide the low power characteristics and high noise immunity of CMOS plus the high speed operation of HMOS. The microcomputer provides highly sophisticated, on- chip peripheral functions including: 256 bytes of static RAM, an 8 channel analog to digital (A/D) converter (used to read the Rj value, the setting of the input comparator, the battery package voltage and the value of the auxiliary input), a serial communication interface (SCI) subsystem, and a serial peripheral interface (SPI) subsystem. The microprocessor works with an 8-bit communication bus to EPROM and EEPROM memories and is interfaced with a decoder, a latch of address and an inverter-driver.
4.4 Firmware
The operating system firmware handles all the logic instructions to the internal peripheral circuits and performs the computation of the linearization equations. The application system firmware is resident on the non-volatile memory (EEPROM) of the microprocessor chip. It is used to store the installation parameters (autocalibration data, program data, etc.)
4.5 Display
The custom display, placed on an auxiliary board, uses high contrast LCD technologies (STN liquid).
OSP
thermometers
are standard equipped with a backlight device for easy readings in poor light conditions.
4.6 Battery charger
The auxiliary module, supplied as a standard accessory, allows operations from 115 Vac 50/60 Hz.
OSP
, if needed, can be operated directly from a line source through the charger. The plastic case of the battery charger incorporates the line voltage plug and a cable with a connector for interconnections to the instrument. The charger circuit is designed with an insulating transformer and a voltage stabilizer circuit. The step-down transformer reduces the power line 115 Vac to a value of 10 Vac. The above voltage is full wave rectified , filtered and stabilized. The output voltage of 5.6 Vdc is the ideal value to recharge the internal Ni-MH batteries
4.7 Digital interface
The digital interface circuit is essentially based on the serial communication interface subsystem (SCI) on the chip of the microprocessor at 0 / +5V level. Adapters to convert TTL to RS 232 voltage levels are available from OMEGA.
15
5 UNPACKING
Remove the instrument from its packing case and remove any shipping ties, clamps, or packing materials. Carefully follow any instruction given on any attached tags. Inspect the instrument from scratches, dents, damages to case corners etc. which may have occurred during shipment. If any mechanical damage is noted, report the damage to the shipping carrier and then notify OMEGA directly or its nearest agent, and retain the damaged packaging for inspections. A label indicates the serial number of the instrument. Refer to this number for any inquiry for service, spare parts supply or application and technical support requirements.
OMEGA
will keep a data base with all information regarding your instrument.
16
6 PRE-OPERATIONAL CHECK
OSP series
portable thermometers are powered either by four alkaline or by Ni-MH rechargeable batteries (optional). The external battery charger, supplied as a standard , is set for 115Vac power source. Before using the instrument carefully verify the nominal voltage value of the charger with the available mains power line. The instrument should be used in environments where the temperature does not exceed the specified limits (from -5°C to +50°C) and where the relative humidity is lower than 95%.
In case of “low” battery conditions (voltage lower than 4.6 V) the display will show the appropriate symbol. A battery symbol means that the battery package has enough energy for about 20 minutes operation. In this condition the instrument batteries must be recharged.
WARNING
U
SE
OSP IR
PORTABLE THERMOMETERS TO MEASURE TEMPERATURE
: 0.9 - 1.1
µµµµ
M
(OSP1600, OSP2000), 1.6
µµµµ
M
(OSP1300)
AND
8-14
µµµµ
M
(OSP800, OSP1000
AND
OSP500)
SPECTRAL BANDS
.
I
N CASE
IR
THERMOMETER SHOULD MEASURE DIRECT OR REFLECTED HIGH INTENSITY RADIATIONS IN THE ABOVE SPECTRAL BAND
,
THEY CAN DAMAGE PERMANENTLY THE
IR
SENSORS INSIDE THE THERMOMETERS
.
T
HIS KIND OF RADIATION CAN BE PRODUCED BY EITHER THE LASER TYPE ND
:YAG (
λλλλ
EM
= 1.06
µµµµ
M) OR BY THE LASER TYPE
CO
2
(
λλλλ
EM
= 10.6
µµµµ
M
).
17
7 POWER SUPPLY
The instrument may be powered by alkaline or rechargeable batteries or directly from the main line.
OSP
uses 4
batteries type AA that are located inside the base of the handle.
7.1 Rechargeable batteries
The rechargeable batteries are shipped with an average level of charge. After unpacking, a full charge of the batteries is recommended. Connect the instrument to the charger module (“OFF” condition) for a period of 10 hours minimum. The Ni-MH rechargeable batteries do not suffer when used in cyclic operations. The cyclic operation is understood as a method of operation by which the battery is continually charged and discharged. Avoid leaving the instrument, with batteries totally or partially discharged, for a long time without recharging them. To charge the batteries use only the original supplied charging module. The module incorporates protection and current limiting devices not normally found in other commercial chargers. The external battery charger is configured, before shipment, for a supply voltage of 115 Vac, upon order specification. The nominal voltage value is indicated on the front label of the charger.
7.1.1 How to maximize the life span of the battery
Disconnect the ac mains supply when the battery is charged. Use the battery until it is completely discharged. Leaving the ac mains charger plugged in during operations will decrease the life of the Ni-MH batteries. Keeping the battery terminal clean will help maximize the operating time. Periodically wipe the positive and negative terminals with a dry cloth. Removing and replacing the batteries will ensure electrical contact. This should be done when using a battery that has not been used for a long time. Note that the operating time decreases at low temperatures. A Ni-MH battery can be recharged about 500 times when used with the recommended instructions. When replacing the Ni-MH batteries with a new set always replace simultaneously the four pieces.
7.2 Power supply with alkaline batteries
Power supply with alkaline batteries must be specified with the order.
7.3 Power supply from main line AC
The battery charger module can be used to power the instrument for continuous operations from main line AC. Rechargeable batteries or alkaline batteries are not required to be removed with AC power supply.
18
8 WARNINGS & CAUTIONS
8.1 Laser Sight
You may receive harmful laser radiation exposure if you do not adhere to the warnings listed below:
USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF PROCEDURES OTHER THAN THOSE SPECIFIED HERE MAY RESULT IN HAZARDOUS RADIATION EXPOSURE.
DO NOT LOOK AT THE LASER BEAM COMING OUT OF THE LENS OR VIEW DIRECTLY WITH OPTICAL
INSTRUMENTS – EYE DAMAGE CAN RESULT.
USE EXTREME CAUTION WHEN OPERATING THE LASER.
NEVER POINT THE LASER BEAM AT A PERSON.
KEEP OUT OF REACH OF ALL CHILDREN.
CAUTION - LASER SAFETY
LASER RADIATION - DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT CONFORMS TO IEC 823/93
CLASS II LASER PRODUCT COMPLIES WITH 21 CFR CHAPTER 1, SUBCHAPTER J.
W
AVELENGTH
: 630-670
NM
– MAX.O
UTPUT
: <1MW
WARNING
DO NOT ATTEMPT TO OPEN THE LASER SIGHT MODULE.
(T
HERE ARE NO USER-SERVICEABLE PARTS IN THE MODULE
).
8.2 Analogue input
Thermocouple input is optional for OSP series thermometers. The sensors are normally linked to electrical potentials equal or near to the ground potential. However, in some applications, there may be present a common mode voltage to ground. Check for voltage between input terminals and the ground, as this voltage can be transmitted to other devices connected to the OSP series temperature indicator.
8.3 Danger and Certification Labels
The laser sight is standard in your thermometer. You can choose between different laser systems as described in section
1.1. Laser provides a visual indication of the field of view of the thermometer. The following figures show the parts and the labels locations of the Laser sight module.
19
20
9 OPERATIONS
OSP series
portable infrared thermometers are factory calibrated before shipment. During the start-up the operator should only select and load, if required, the pertinent application parameter as described in the following paragraphs. Before entering the procedure to operate the instrument it is useful the understanding of the messages that can be present in the display. The following figure shows the overall layout of the display and an explanation of each symbol or message :
e
+-
A
uto
K
Log Rcl Set
°F °C
Lock
Comp
HAL
LAL
1
2 3
4 7
5
6
8
9
10
11
15
13
14
12
1 - Indicates logging mode operative 2 - Enables recall of memory stored data 3 - Indicates the setting mode enabled during configuration and logging mode set up 4 - Low battery indication 5 - Temperature in degree Fahrenheit 6 - Temperature in degree Celsius 7 - Temperature in Kelvin 8 - Actual measured temperature value 9 - Indicates Low Alarm 10 - Indicates High Alarm 11 - Auxiliary measurement data or setting (AVG, LAL, HAL, MIN, MAX, DIF) or thermocouple input indication 12 - Operative emissivity value 13 - Ambient temperature compensation 14 - Continuous measurement mode 15 - Automatic acquisition in logging mode
The contact closure of the keys is acknowledged as a coded signal by the microprocessor that recognizes the operator’s instructions.
<Trigger>
Keep pressed for operation. Press twice, in sequence, to hold the instrument -ON-
<
▲▲▲▲
> <
▼▼▼▼
>
Value setting or data selection during Configuration or Set-up procedures <*> Laser sighting on/off or data acquisition mode set-up <
MEM
> Data acquisition operation and logging manual step
<
SEL
> Selects the operative mode
<
ENTER
> Memory load key
9.1 Quick Start
To use the
OSP
infrared thermometer right away follow this simple steps : Point the instrument at the object you want to measure and pull the trigger (the actual temperature value, the pre­programmed emissivity value and the auxiliary indication will be displayed); a “LOCK” symbol will appear on display. Set emissivity value using the <> and <> keys. Refer to appendix A to define emissivity and for emissivity tables. Press and release the trigger to switch the unit off.
9.2 How to Operate the instrument
To switch the instrument “On” keep the trigger pressed; when the trigger is released the instrument will be switched “Off”. When continuous operations are required press twice, in sequence, the trigger to lock the operative mode. The first display indication will show the type of instrument and the temperature technical unit enabled.
21
°C
e
°C
e
e
°C
e
°C
e
°C
e
°C
followed, after few seconds, by one of the main operative display pages. The displayed page will be the same as when the instrument was switched Off
e
°C
e
°C
e
°C
Lock
e
°C
Lock
ee
°C
e
°C
Lock
e
°C
Lock
e
°C
Lock
Press the <
SEL
> key to select the required page.
Press the <▲> and <> keys to adjust the emissivity value as required by the application.
9.3 Hold
To hold the measurement, press the <
MEM
> key; the “X” symbol will be displayed on lower display.
ee
°C
To run in continuous measurement, press the <
MEM
> key another time.
9.4 Laser sighting
C
AUTION: BEFORE USING THE LASER SIGHT READ THE RECOMMENDATIONS ON PARAGRAPH
15.2.2.
22
In measuring mode, press the <*> key to enable (disable) the laser spot. After one second the message “LAS EN” (“LAS DIS”) will be displayed. When enabled, the laser can be turn on by pressing the <
Trigger
> button for more than 1 second; a flashing emission
indicator symbol “*” will appear on the lower display.
ee
°C
The laser will go out automatically releasing the <
Trigger
> button.
NOTE: T
O PREVENT ANY COLLATERAL RADIATION IN EXCESS OF THE LIMITS DURING THE OPERATION OR MAINTENANCE, THE LASER
BEAM WILL AUTOMATICALLY SWITCHED OFF WHEN TRIGGER IS RELEASED
(
ALSO IN LOCK OPERATION MODE
).
9.5 Unstable temperature measurement
When measuring a very unstable temperature process, is necessary a slow and stable reading.
OSP
is equipped with an
advanced average mode that works with the following parameters :
!
Working band
!
Weight
The working band is used to define the maximum difference between the actual and previous measurement values where average mode has to be operative. It means that if you are measuring ambient temperature ( for example 23°C ) and you point the instrument to the process ( for example 600°C ) you don’t need a slow increasing due to average ( The difference is 600 – 23 = 577 °C ). If working band is set at 100°C instrument don’t apply any average and it will pass directly to temperature of the process ( 600°C ) otherwise with standard average you will see the temperature slowly growing from ambient to the target point. When instrument is measuring the process the difference due to unstable temperature is inside working band so it will works in average mode. More unstable is the process higher has to be the weight of the average. Higher is the weight ,slower will be the reading.
Indicated values with weighted average = 0
Weighted average effect
Input temperature fluctuations
Indicated values with weighted average = 2
Indicated values with weighted av erage= 32
Temperature fluctuation
Measured value with weighted average=32 and working band
excluded
Measured value with weighted average=32 and working band
included
Working band effect
To enable/disable the average mode follows the procedure below:
23
" In measuring mode, press the <
Sel
> key until the following page will appear:
ee
°C
The display indicates the actual temperature value together with the technical unit symbol, the programmed emissivity value and the averaged temperature value.
" At this stage press <
ENTER
> key to enable the average mode. On the display will appear the following page:
e
°C
The average mode is now enabled; the “X” symbol will appear. It will remain enabled after the instrument switching off. To disable the average mode, repeat the above procedure. The “X” symbol will disappear. Average mode will remain enabled also in data acquisition mode and in peak mode operations. The standard programmed values on OSP are:
working band
=100°C and
average weight
=32 that are parameters valid in most of your process applications. To change the parameters setting, use the optional
Average
software for personal computer and TTL/RS232 cable.
9.6 Alarm settings
Low and High Alarm levels can be programmed and used in “hot spots” for fast scanning or long term surveillance. To select these operative modes press, in sequence, the <
SEL
> key to obtain the following display pages :
e
°C
Lock
To change the Low and/or High Alarm settings follow the procedure indicated below :
Press the <
ENTER
> key to enter the setting procedure. One digit of the setting line will blink. The blinking digit allows
a new adjustment.
Press the <
SEL
> key to select a different digit or the polarity (+/-)
Press <> and <> keys to modify the value of the blinking digit
Repeat the last two steps to set the new alarm threshold
Press the <
ENTER
> key to return to the normal operative mode with the new Low Alarm setting
If required the Low Alarm can be excluded setting the low alarm level above the full scale temperature value where the message -Off- will appear.
24
e
°C
Press the <
SEL
> key to obtain the following page that enables the High Alarm setting
e
°C
HAL
Repeat the setting procedure used for the low alarm setting
If required the High Alarm can be excluded setting the high alarm level above the full scale temperature value where the message -Off- will appear.
e
°C
An alarm status, as shown in the above figure, will be announced to the operator by the -
HAL
- or -
LAL
-message on the
display and, simultaneously, by an acoustic signal.
The acoustic signal can be acknowledged with the <
MEM
> key.
9.7 Thermocouple input
To select these operative modes press, in sequence, the <
SEL
> key to obtain the following display pages :
e
°C
Lock
Press <> and <> keys to select the required type of thermocouple (type “S” or type “K”). This auxiliary input can
be used for contact temperature measurements or for automatic emissivity adjustments.
25
e
°C
Lock
e
°C
Lock
When selected the required type of thermocouple press the <
SEL
> key to obtain the following page showing
simultaneously both the actual IR temperature and the temperature measured with the thermocouple.
e
°C
Lock
If the measuring thermocouple is not connected or is broken, the display will indicate "Tc OTC".
e
°C
Lock
If the auxiliary input is disable, the display will indicate "Tc OFF".
e
°C
Lock
9.8 Minimum, maximum, average and difference indication
" Press the <
SEL
> key to obtain one of the following indications showing, in the auxiliary display, respectively the Minimum or Maximum value recorded in the interval time from the instrument switch -On- and the calculated Difference value between the identified Maximum and Minimum values.
e
°C
Lock
e
°C
Lock
e
°C
Lock
e
°C
Lock
When positioned on the max or min function, use <
ENTER
> key to reset the stored value.
26
9.9 Automatic Emissivity Setting
This function is used to make an automatic setting of emissivity using an external thermocouple connected with the auxiliary input.
If necessary, enable the thermocouple input (see par. 9.2) and select the thermocouple type (see par. 8.6)
Press the <> or <> key until the thermocouple temperature measure is displayed
e
°C
Lock
Press <
ENTER
> key for setting the emissivity value with reference to the temperature values read from thermocouple
and pyrometer. The emissivity will change automatically to the correct value by the OSP.
e
°C
Lock
An error message appear if the
OSP
measure a temperature highest of the thermocouple. If the error happen, verify
the contact between the thermocouple and the object.
e
°C
Lock
27
10 CONFIGURATION
The operative configuration of the unit can be changed using the below indicated procedure.
Switch the instrument -
Off
-
Keep the <
ENTER
> key pressed and press the <
Trigger
> push button until the instrument displays the
indication relevant with the technical unit selection.
10.1 Technical Unit selection
the display will indicate one of the following technical units (Celsius, Fahrenheit and Kelvin):
e
Set
Lock
e
Set
Lock
e
Set
Lock
Press the <> or <> key if you require to select a new technical unit
Press the <
SEL
> key to select the new configuration step or press the <
Trigger
> to switch the instrument -Off-. The
new selection will be automatically memory stored.
10.2 Thermocouple input enabled
Press the <
SEL
> key until the following indication is displayed :
e
Set
°C
Lock
or
e
Set
°C
Lock
Press the <> or <> key to enable (or to disable) the thermocouple auxiliary input operation.
Press the <
SEL
> key to select the new configuration step or press the <
Trigger>
to switch the instrument -Off-. The
new selection will be automatically memory stored.
10.3 Temperature Scale selection
Press the <
SEL
> key until one of the following indication is displayed :
e
Set
°C
Lock
or
e
Set
°C
Lock
Press the <> or <> key to select the required International Temperature Scale.
28
Press the <
SEL
> key to select the new configuration step or press the <
Trigger>
key to switch the instrument -Off-.
The new selection will be automatically memory stored.
10.4 Date & Real Time clock setting
The two following pages allow respectively date and real time updating .
Press the <
SEL
> key to display the date page
e
Set
°C
Lock
Press the <
ENTER
> key to enable the adjustment
Press the <
SEL
> key to select the parameter to be modified (day, month, year)
Press the <> or <> key to modify the selected parameter (blinking)
Press the <
ENTER
> key to acknowledge the new setting
Press the <
SEL
> key to enter the real time clock page
e
Set
°C
Lock
Press the <
ENTER
> key to enable the adjustment
Press the <
SEL
> key to select the parameter to be modified (hour, minute, second)
Press the <> or <> key to modify the selected parameter
Press the <
ENTER
> key to acknowledge the new setting
Press the <
SEL
> key to select the new configuration step or press the <
Trigger
> to switch the instrument -Off-. The
new selections will be automatically memory stored.
10.5 Acquisition settings
Press the <
SEL
> key until the following indication is displayed :
e
Set
°C
Press <
ENTER
> for setting the acquisition mode;
Press the <▲> or <> key to select the required setting: MAN, MIN, MAX or interval time for automatic acquisition ;
N
OTE
: MAN
UAL MODE DISPLAYS THE REAL TIME TEMPERATURE VALUES AND PERMITS TO STORE THE VALUE PRESSING THE
<MEM>
KEY
. MAX (MIN)
ACQUISITION MODE DISPLAYS ONLY THE MAXIMUM (MINIMUM) VALUE AND PERMITS TO STORE THE VALUE
PRESSING THE
<MEM>
KEY
.
To change the interval time for automatic acquisition, press <
SEL
> to change decade and the <> or <> key to
change its value.
Press <
ENTER
> to store the new acquisition mode. See Cap. 11 for appropriate data acquisition operation.
29
10.6 Buzzer On/Off
If the Alarms option is installed, it is possible to switch the buzzer on or off. This page allows the set-up of the acoustic signal.
e
Set
°C
Lock
e
Set
°C
Lock
When the “Buzzer” is switched -Off- the alarm conditions will be announced only with the relevant -
HAL
- or -
LAL
-
message on the display.
10.7 OSP 500/800 decimal point selection
N
OTE:THIS FEATURE WORKS ON
OSP 500
OR
OSP800
WITH FIRMWARE FROM VERSION
4.001.
It is possible to display the temperature measured with a decimal digit. The following procedure is used to enable the operating mode.
" Switch the instrument off. " Keep pressed <*>+<
SEL
> keys and switch the unit on by pressing the <
Trigger
> key. The temperature
measurements will be displayed with the decimal digit.
Repeat the above procedure to disable the operating mode.
10.8 Ambient Temperature Compensation (TAM)
This function is used to compensate the measure from the ambient temperature.
Targets which have low emissivity will reflect energy from their surrounding environments, which is going to add to their own emitted energy. If the temperature of the surrounding environment is the same value as the internal temperature of the unit, there is no need to consider this value as it is automatically and continually measured. However, in some industrial situations the surrounding environment (machines, furnaces, or other heat sources) has a much higher value than the one of the internal of the unit temperature. The table below "
Effect of TAM on Accuracy
" underlines these effects
on measurement accuracy, with and without the use of the Ambient Temperature Compensation mode.
The Ambient Temperature Compensation mode should be used whenever the stated following points are true :
1. The target has a low emissivity.
2. Temperatures of objects facing targets are much hotter than the temperature of the unit.
3. An increased measurement accuracy is required
Actual Target
Temp.
Emissivity Ambient Temp. Indicated Temp.
w/o TAM
Indicated Temp.
w/ TAM
°C °F °C °F °C °F °C °F 100 200 0.8 25 80 100 200 100 200 100 200 0.8 40 100 102 203 100 200 100 200 0.8 150 300 120 250 100 200 100 200 0.6 150 300 160 321 100 200
Note
: The internal temperature of IR thermometer is assumed to be 25 °C (80 °F)
Press the <
SEL
> key until one of the following indications is displayed :
30
e
Set
°C
e
Set
°C
Press the <▲> or <> key to enable (or to disable) the thermocouple auxiliary input operation.
e
Set
°C
With the compensation function activated, press <
ENTER
> to begin the measurements in order to calculate the
ambient temperature average value. Point
OSP
towards objects and surfaces surrounding the target, making a scan over those items and acquiring corresponding temperature values. A counter (from 1 to 255) indicates the number of measurements considered (
OSP
displays the first value read during this procedure). These acquisitions have to be
done with emissivity ε=1, hence when acquisition starts the emissivity value has set automatically to one.
Press <
ENTER
> to stop acquisition and to confirm the read ambient temperature (at this moment
OSP
will display
the last acquired temperature value).
Press the <
SEL
> key to select the new configuration step or press <
Trigger
> to switch the instrument off. The new
selections will be automatically memory stored.
To execute the temperature measurements with TAM, switch
OSP
on and set the emissivity to the correct target value.
10.9 Firmware revision number
Press the <
SEL
> key until the following page is displayed :
e
Set
°C
Lock
The above page indicates the version of the firmware installed in your instrument. This information is extremely important in all inquires for technical assistance support.
10.10 Instrument serial number
Press the <
SEL
> key until the following page is displayed :
e
Set
°C
Lock
The above page indicates the serial number of the instrument. Such number is extremely important for any request of technical assistance support.
31
10.11 Battery level of charge
Press the <
SEL
> key until the following page is displayed :
e
Set
°C
Lock
e
+-
Set
°C
Lock
e
Set
°C
Lock
The above pages indicate the status of power supply and the level of charge of the battery. When a voltage lower than 4.6V is present a low battery symbol is also displayed. If the instrument is connected to the main line for battery charge or long term continuous operation a symbol “AC” will be displayed instead of the battery voltage value.
32
11 DATA ACQUISITION OPERATIVE MODE
When the optional accessory of the instrument is specified (Internal data memory + RS232 adapter + LogMan software) the instrument is equipped with an internal memory to store up to 500 input data records. Two types of data acquisitions can be selected by the user.
Continuous acquisition
The operator can manually, step by step, store his data through the keyboard or can select the interval time between each acquisition and store progressively the input data in the “Tag 0” file.
Acquisition by dedicated Tags
Standard Agencies and Quality Auditors require the collection, organization and availability of traceability documents. A supporting software is available to transfer a selection of plant inspections from a PC to the internal memory of the instrument in order to simplify field check and select the appropriate Tag number. Dedicated input data are memory stored and downloaded into a PC to document the inspection activity. Data can be saved on disks, viewed and printed in a numeric or graphic mode.
A full description of this acquisition mode is available with the dedicated software. To operate in continuous acquisition mode, refer to the following procedure.
11.1 How to operate in data acquisition mode
Start this procedure with the instrument switched -Off-
Keep the <
MEM
> key pressed and, simultaneously, press the <
Trigger
> until the following indication is displayed :
e
Log
°C
Lock
Press the <▲> or <> key to select the required Tag identification code (Tag 0 or one out of the 20 available Tags
from "A" to "S"). Parameters (emissivity, low and high alarm settings) can be set through keyboard only in Tag 0. Other Tags are reserved to operation through the dedicated PC software.
e
Log
°C
Lock
e
Log
°C
Lock
Press the <
SEL
> key to display the number of memory locations still available (500 indicates a full empty memory).
e
Log
°C
Lock
Data can be recorded in the selected Tag in manual or automatic mode. Manual acquisition mode can be selected between normal (MAN), MAX or MIN. About acquisition mode and setting of time acquisition, refer to par. 9.5.
33
Press the <
MEM
> key to start the acquisition.
If automatic acquisition mode is set,
OSP
acquires automatically with the programmed time interval.
If MANual mode is set, the real time temperature measurement is displayed; press the <
MEM
> key to store the displayed value. If MAX or MIN mode is set, the maximum (minimum) temperature measurement is displayed; press the <
MEM
> key
to trigger acquisition and the <
ENTER
> key to reset the value for the next acquisition.
Press the <
SEL
> key to obtain the page that allows the clearing of the memory stored data :
e
Log
°C
Lock
Press the <
ENTER
> key to clear the memory content and to obtain the following message that require confirmation :
e
Log
°C
Lock
Press the <
ENTER
> key to confirm the cancellation request and to obtain the following message :
e
Log
°C
Lock
Press the <
SEL
> key to obtain the following page indicating the identification code (or name) assigned to the active
Tag (zzzzzzz = Furn686 = Trap888). The identification code is selectable only through software.
e
Log
°C
Lock
Press the <
SEL
> key to review the pages relevant with -
HAL
- and -
LAL
- alarm settings. To modify the alarm level in
the data acquisition mode see the paragraph 10.2.
34
e
Log
°C
Lock
e
Log
°C
Lock
e
Log
°C
Lock
11.2 Recall stored data
Switch off the
OSP
thermometer.
Keeping pressed the <
SEL
> key, press the <
Trigger
> key to obtain the following indication :
e
Log
°C
Lock
Press the <▲> or <> key to select the required Tag identification code (Tag 0 or one out of the 20 available Tags
from "A" to "S").
e
Log
°C
Lock
e
Log
°C
Lock
Press the <
SEL
> key to display the value of the first memory location in the selected Tag.
e
Log
°C
Lock
Press the <> or <> key to display the subsequent memory locations.
35
12 LOGGING DATA MANAGER
LogMan software allows
OSP series
users to set and prepare infrared thermometers to acquire the data organized by ‘Tag’. This software also allows to download data from Instrument on document (table) with date, time and value. You can view table, save table, print table, export table in excel-txt-html and obtain graphs. You can also manage data coping, moving on other tables. The software menages two different files :
" Extension .LMM are models. It means you will save on disk only preference setting. " Extension .LMD are models and contains also data. It means you will save on disk current model with all data
on table.
When you open a new file a DEFAULT.LMM model will be automatically loaded.
If you prefer different preferences (for example header and footer with your company name) load model DEFAULT.LMD change data and save it.
12.1 Installation
LogMan
runs on IBM PC under WINDOWS95 or NT. Minimum requirements are a 486 CPU with 8 Mb Ram and 2Mb on
hard disk, color monitor and a Microsoft mouse or a compatible one.
In order to install
LogMan
for Windows, follow the below procedure:
Place the
LogMan
disk in a 3.5" diskette drive;
From the Windows Program Manager's File Menu, select
<RUN>
;
Enter the filename A:setup.exe (substitute the letter A for the disk drive that contains
LogMan
diskette)
Follow on screen instructions making sure to provide your
LogMan
directory, when prompted, with the correct path.
Once installed, in “Start” menu, the
LogMan
icon will appear it will be possible to boot it by select it as usual.
12.2 Program Architecture
LogMan
has a typical Windows structure with a series of menu and toolbars that can be recalled by clicking on the
name, or on the icon which represents the action the user wants to carry out.
The main menu program is structured as it follows:
36
These menu are subdivided in:
File
New, Open, Save, Print menu.
New
Creates a new document.
Open
Opens an existing document.
Options
Set Communication port, text language and synchronize clock
Exit
Exits from Logman.
Help
Help menu.
Index
Offers you an index about topics on which you can get help.
About
Displays the version number of this application.
When a document is loaded or created, the new main menu is:
File
New, Open, Save, Print menu.
New
Creates a new document.
Open
Opens an existing document.
Save
Saves an opened document using the same file name.
Save As
Saves an opened document to a specified file name.
Export
Exports opened documents as Excel, txt or HTML file
Close
Closes an opened document.
Print setup
Selects a printer and printer connection.
Print
Prints a document.
Options
Set Communication port, text language and synchronize clock
Language
Selects language text
Port settings
Select communication port COM
Synchronize instrument
Sets IR thermometer internal clock equal to Computer
Exit
Exits from Logman.
Data
Instrument operation menu (Menu accessible only if ‘Data Table Window’ is active).
AutoLoad tag
Loads automatically from the instrument.
Load tag
Loads from the instrument.
Save tag
Saves preferences on the instrument.
Save as tag 0
Saves preferences on the instrument Tag 0.
Delete tag
Deletes the Tag on the instrument.
Clear all tags
Deletes all Tag on the instrument.
PC Log
Logs data from the instrument.
Sample now
Logs one datum only from the instrument.
Preferences
Set preferences.
Edit
Data operation menu (Menu accessible only if ‘Data Table Window’ is active).
Copy into new file
Copies a selection in to a new file.
Copy into open file
Copies a selection in to an existing file.
Move into new file
Moves a selection in a new file.
Move into open file
Moves a selection in an existing file.
Delete
Deletes selections.
Subtract
Subtracts two selections in a new file.
Window
Windows operation menu.
Tile
Tiles all active windows.
Cascade
Cascades all active windows.
Arrange Icons
Arranges all active icons.
Help
Help menu.
Index
Offers you an index to topics on which you can get help.
About
Displays the version number of this application.
37
12.2.1 Toolbars
Additional tool bars menu appear when the data view window or the graph view window is selected.
Fig. 11.2a: Data view toolbar
Fig. 11.2b: Graph view toolbar
The following buttons are displayed on the toolbars.
View graph on actual data window
Pressing this button a new window, containing the graph of the active data table, is displayed.
Preference dialog
On preferences menu command you can set :
Instrument
- Tag name ( max 7 characters ).
- Emissivity ( from 0.10 to 1.00 ).
- Temperature scale ( IPTS 68 or ITS90 )
- Units ( °C, °F or K ).
- Decimal point ( 1 or 0.1 ).
- Alarms value and check box ( on/off ).
- Sampling time and total time of PC log.
Header
Lines ( max 10 ).
Footer
Lines ( max 10 ).
Print
- Top, header, left, right, bottom and footer distance in inches.
- Enable/Disable Grid and color.
- Enable/Disable Center table horizontally
Graph
- Autoscale or Lower and Upper Limits.
Select decimal point
(0 or 1)
Select engineering units
( °C, °F or K ).
Enable/Disable alarms
Copy into a new file
Use this command to copy selected data onto a new file. This command is unavailable if there are no data currently selected.
38
Copy into an open file
Use this command to copy selected data onto an opened file. It will appear a list of current opened files to be selected.
This command is unavailable if there are no data currently selected.
Move into a new file
Use this command to move selected data onto a new file. This command is unavailable if there are no data currently selected.
Move into an open file
Use this command to move selected data onto an opened file. It will appear a list of current opened files to be selected.
This command is unavailable if there are no data currently selected.
Set instrument Tag 0
with current preferences
Sample now
Connect the unit to communication port of the computer. Press the <
Sample now
> command to log one datum on actual document.
12.3 Quick starting
This software allows to download data from Instrument on document ( table ) with date, time and value. You can view table, save table, print table, export table in excel-txt-html and obtain graph.
For example if you want to measure the temperature of the transformer model AGC 134.
" Run the command <
File New>
from the menu. The following windows will appear:
" Select the command <
Data
> from the menu and then the option <
Preference
>. Fill in preferences windows the
parameters you need: For example:
TAG name : AGC 134 Emissivity : 0.98 Low Alarm Level: +15 °C (you don’t need it so you can set it very low) High Alarm Level: +30 °C.
39
" Select <OK> button to save the preference setting in the PC Memory. " From the preference menu you can also set the header and footer line as you need for later printed report. " Save model (Save as …) with extension .LMM. " Now connect the instrument with the communication cable and turn it on. " Open the file AGC.LMM (if previously closed ). " Select the <
Data
> command from the menu and then the option <
Save Tag
>. Select for example Tag A and press
the <Ok> button.
" Disconnect the unit and go on the field to take same measurements (see also cap.10 to operate in an acquisition
mode).
" Start this procedure with the
OSP
switched -Off-
" Keep the <
MEM
> key pressed and, simultaneously, press the <
Trigger
> until the following indication is displayed :
e
Log
°C
Lock
Press the <▲> or <> key to select the Tag A identification code.
e
Lo
g
°C
Lock
Press the <
SEL
> key to display the number of memory locations still available.
e
Log
°C
Lock
Data can be recorded in the selected Tag in manual or automatic mode. About the acquisition mode and setting of time acquisition, refer to par. 9.5.
If necessary, press the <
SEL
> key to obtain the page that allows the clearing of the previous memory stored data in
the Tag A:
40
e
Log
°C
Lock
Press the <
ENTER
> key to clear the memory content and to obtain the following message that require confirmation :
e
Log
°C
Lock
Press the <
ENTER
> key to confirm the cancellation request and to obtain the following message :
e
Log
°C
Lock
Press the <
SEL
> key to obtain the following page indicating the identification code (or name) assigned to the active
Tag (i.e. ACG 134).
e
Log
°C
Lock
134
Press the <
SEL
> key to review the pages relevant with -
HAL
- and -
LAL
- alarm settings. To modify the alarm level in
the data acquisition mode see the paragraph 10.2.
e
Log
°C
Lock
e
Log
°C
Lock
" Press the <
MEM
> key to acquire one measurement value.
" Once the acquisition completed, you can switch OSP off and go back to the computer. Reconnect the unit to the PC
and open the file AGC.LMM (if previously closed).
" Run the command <
Autolo ad Tag
> From the <
Data
> menu (the program will search inside the instrument tag AGC
134 and will download data).
" Press the button to display the data graph window containing the graph of the active data table.
41
Now you can save data on disk using name with extension LMD (for example TODAY.LMM).
42
13 OPTIONS & ACCESSORIES
13.1 Printer operations
" Connect a compatible printer to
OSP
and switch the thermometer on.
e
°C
Lock
" Press the <
MEM
> key to obtain the following indication:
e
°C
Lock
" Press the <
ENT
> key to print the Date, Time and measured value as follows:
Date : 18/02/98 Time : 14:44:02
Measure : 186 °C
13.2 Sighting telescope system
Item Q.ty Part n.° Description
1 2 EE010015 fastening screws for guide 2 1 EE280169 guide 3 2 EE280179 mounting rings 4 1 EE280181 telescope 5 1 EE280198 filter ring 6 1 EE360053 filter HT 7 1 EE370133 O-ring
Installation of the telescope :
1 Remove with a cut screwdriver the nut protection caps placed on the upper part of the instrument. 2 Place the guide on the upper part of the pyrometer, taking care of orientating it so that it leans entirely on the back of
the instrument. 3 Fasten the screws in order to make the guide integral with the pyrometer. 4 Place the mounting rings on the guide, in order to cling them to the side edges. Let each ring slide along the guide
until each catch is housed in the respective slot. Fasten firmly each fastening screw with a coin or a screwdriver. 5 Separate the upper and the lower half of each mounting ring by removing the two screws, using the enclosed
hexagonal wrench. 6 Put the telescope in the supporting framework formed by the lower halves of the mounting rings and place the
extremity with the widest diameter turned towards the target. 7 Rotate the telescope into the mounting rings, in order to bring the elevation adjustment (UP) high. Fasten the
screws, avoiding to make an excessive clamping.
Alignment of the telescope :
1 Make the pyrometer integral with a support.
43
2 Remove the elevation and the windage adjustment caps. 3 Switch on the two pinpointing lasers and aim them to a target placed at a distance equal to the working distance. 4 The aiming must be carried out by keeping the eyepiece at a distance of about 80 cm from the eye. 5 Operating on the two adjustments (elevation and windage), place the center of the reticule in connection with the
center of the axis joining the two spot lasers. The minimum suggested working distance is 5 meters, shorter working
distance can cause alignment problems. 6 In this way the telescope is aligned to operate at the working distance; a variation in the working distance requires a
new alignment of the telescope.
Filter installation
1 Filter use is suggested in viewing targets with temperature higher than 1000°C 2 Install the filter ring on the front side of the telescope lightly pushing the component
CAUTION
NEVER LOOK AT THE SUN THROUGH THE TELESCOPE: THE CONCENTRATION OF STRONG SOLAR RAYS CAN CAUSE SERIOUS EYE
DAMAGE
.
NEVER LOOK AT TARGET WITH TEMPERATURE HIGHER THAN
1000°C
WITHOUT THE TELESCOPE PROTECTION FILTER
.
C
ONCENTRATION OF STRONG VISIBLE
- IR
RADIATION CAN CAUSE EYE DAMAGE
13.3 Red Point pinpointing system
Item Q.ty Part n.° Description
1 2 EE010015 fastening screws for the guide 2 1 EE280169 guide 3 1 EE280180 N°1
red point sight device
N°2 mounting rings N°1 polarize filter
N°1 lens Hood extension 4 1 EE360055 filter for T > 1300°C 5 1 EE360057 filter for T from 1000 to 1300°C 6 1 EE370127 O-ring/O-ring
Installation of the sighting device :
1. Remove with a cut screwdriver the nut protection caps placed on the upper part of the instrument.
2. Place the guide on the upper part of the pyrometer, taking care of orientating it so that it leans entirely on the back of the instrument.
3. Fasten the screws in order to make the guide integral with the pyrometer.
4. Place the mounting rings on the guide, in order to cling them to the side edges. Let each ring slide along the guide until each catch is housed in the respective slot. Fasten firmly each fastening screw with a coin or a screwdriver.
5. Separate the upper and the lower half of each mounting ring by removing the two screws, using the enclosed hexagonal wrench.
6. Put the sighting device in the supporting framework formed by the lower halves of the mounting rings, so that the rheostat hand grip is situated on the left side of the pyrometer, with respect to the aiming direction.
7. Rotate the sighting device into the mounting rings, in order to bring the elevation adjustment (UP) high. Fasten the screws, avoiding to make an excessive clamping.
Alignment of the sighting device :
1. Make the pyrometer integral with a support.
2. Activate the illuminated red point by turning the rheostat hand grip clockwise to increase the brightness or in an counter-clockwise direction to reduce it. The position “1” indicates the minimum brightness and the position “7”, the maximum one.
3. The aiming must be carried out by keeping the eyepiece at a distance of at least 7 cm from the eye.
4. Remove the elevation and the windage adjustment caps.
5. Switch on the two pinpointing lasers and aim them to a target placed at a distance equal to the working distance. The minimum suggested working distance is 5 meters, shorter working distances can cause alignment problems
6. Operating on the two adjustments (elevation and windage), place the illuminated red dot in connection with the centre of the axis joining the two spot lasers.
7. In this way the sighting device is aligned to operate at the working distance ; a variation in the working distance requires a new alignment of the sighting device.
Filter installation
1 Select the filter depending on the working temperature 2 Insert the selected filter in the internal side of the polarized filter mount 3 Place the O-ring over the filter to lock it
44
4 Screw up the polarized filter with the inserted optical filter placed on the front side of the sighting device
CAUTION
IF Y
OUR
OSP
UNIT IS EQUIPPED WITH TELESCOPE OR "RED POINT" PINPOINTING SYSTEM, NEVER AIM
OSP
DIRECTLY AT THE SUN, AS
EYE DAMAGE COULD RESULT
.
45
14 APPLIC ATION NO TES
14.1 Infrared energy
Each body, at temperatures above the absolute zero (-273°C or 0K), emits energy in the form of electromagnetic radiation. As the temperature of the body rises, the intensity of this infrared energy increases. The temperature of this particular body can therefore be determined by measuring the intensity of this infrared energy. An equipment used to measure the temperature with this method is called "infrared thermometer" or a "non-contact thermometer" since the thermometer is not required to be in contact with the body in order to measure its temperature.
14.2 Applications
The temperature measurement of liquids or gases is well accomplished using a thermoelectric sensor thanks to the good thermal exchange between the sensor and the fluid. When solid objects are to be measured it is difficult to obtain a good thermal exchange and the possibility of making an additional error should be kept into consideration. Temperature measurements with direct contact are often impossible to be carried out when the target is moving or is connected to dangerous electrical sources or when, for any other reason, it is impossible or difficult to touch it.
14.3 Emissivity
The infrared energy emitted by a body differs according to the composition of the body and to the physical condition of the surface. Non-contact thermometers are calibrated using a blackbody source (made with material that absorbs energy at all wavelengths) as a reference standard. However, to obtain the reading of the true temperature, it is necessary to compensate the thermometer for the actual emissivity of the object to be measured.
Emissivity = Surface Radiation Blackbody Radiation
All
OSP
infrared thermometers are equipped with an emissivity adjustment from 0.10 to 1.00. The emissivity values relevant to different materials and surface conditions are detailed in “How to determine an object emissivity” in Appendix A2 of this manual. The values from the above tables are reference values only: emissivity can in fact be slightly higher with a higher oxidation of the material .
14.4 Reflected energy compensation
The radiation perceived from the thermometer is the one emitted by the target plus the radiation reflected by the surface of the object itself. To obtain more accurate readings, particularly for measurements of low emissivity temperature objects, the energy reflected from the target should be considered; that energy changes according to the temperature of the surrounding environment. A typical example of that principle of operation comes from re-heating furnaces, where the temperature of the vault is considerably higher than the target temperature.
OSP
thermometers provide an auxiliary function "TAM" to compensate for the ambient temperature. When the average value of the ambient temperature is set, the microprocessor processes a correction algorithm and then displays the real temperature value of the target (see par. 9.7).
As an example, the error occurring when no auxiliary function "TAM", to compensate for the ambient temperature, is enabled can reach 20°C when measuring a target at 100°C with the ambient temperature of 150°C.
46
15 DIGITAL INTERFACE
The
OSP
portable IR thermometer can be equipped with a digital interface. The interface circuit is essentially based on the serial communication interface subsystem (SCI) on the chip of the microcontroller. The output voltage levels are TTL at 0 to +5 V. An optional adapter to convert the voltage level from 0 to +5 V to RS232 standard levels can be supplied. This adapter is required to interface
OSP
with a Personal Computer.
15.1 Digital output wiring practice
The wiring to digital output signals is made through a mini DIN connector mounted on the lower end of the case. The pertinent connections are indicated below :
5 V
Rx IRtec P
Female miniDIN connector (case mounted - external view)
Dig. ground
Anal. ground
Analog Output
Tx IRtec P
For easy interconnections a miniDIN connector with cable (cat. EE420123) can be supplied on request. The color codes of the conductors can change with different suppliers; please check before using.
Front view
1 2
3
4 5 6
7
8
LINDY CINCH
pin 1 : brown black pin 2 : red green pin 3 : green blue pin 4 : gray gray pin 5 : purple yellow pin 6 : blue white pin 7 : orange red pin 8 : yellow brown
15.2 TTL to RS 232 adapter
The cat. BB530001 TTL to RS232 adapter consists of a cable to which are connected a male mini DIN connector (for
OSP
) and a DB25 connector, that contains the electrical circuitry (for the PC).
The basic circuit and connections are as follows:
47
+
IBM
3
2
7
D B 2 5
IBM - PC




















+ 5 V
10 µF, 16 V
10 µF, 16 V
10 µF, 16 V
+
15
8
Rx
Tx
Ground
IBM
IBM
Tx
Rx
6
2 16 10
11
12
14
13
1
3
4
5
+
+
10 µF, 16 V
10 µF, 16 V
ICL 232
TSC 232
Tx
Rx
+ 5 V
Mini-Din
TTL to RS 232 converter
15.3 Communication protocol
The exchange of information when a
OSP
is interconnected with a PC is as it follows.
Serial Interface of PC has to be set to 9600 Baud, 8 bit and No parity.
15.3.1 Computer data request from OSP to PC
Computer OSP
Tx IDNAME Rx IDNAME Proceed if the name is acknowledged Rx IDNAME Tx IDNAME If not, do not answer
Tx Instruction Rx Instruction Rx Instruction Tx Instruction
Tx char Rx char Rx DATA 1 Tx DATA 1
Tx char Rx char Rx DATA 2 Tx DATA 2
Tx char Rx char Rx DATA 3 Tx DATA 3
Tx char Rx char Rx DATA 4 Tx DATA 4
Tx CHKSUM Rx char Rx CHKSUM Tx CHKSUM
IDNAME, Instruction, DATA 1, DATA 2, DATA 3, DATA 4 and CHKSUM are 8-bit values (1 byte) Tx CHKSUM = DATA 1+DATA 2+DATA 3+DATA 4 .AND. FF Rx CHKSUM = DATA 1+DATA 2+DATA 3+DATA 4 .AND. FF
The above is useful to verify correct received data. The minimum time-out of OSP is 3 seconds. IDNAME is set to 1.
48
Reading values
$00 - Pyrometer measure -------------------------------------------------------------- RX
DATA1 STATUS DATA2 ­DATA3 VALUE (H) DATA4 VALUE (L)
IF STATUS=0 MEASURE = SIGNED INTEGER ( VALUE ) / 10 IF STATUS=2 MEASURE = ERROR RJ IF STATUS=10 MEASURE = OFL IF STATUS=11 MEASURE = UFL
$01 - Emissivity---------------------------------------------------------------------------- RX
DATA1 ­DATA2 ­DATA3 ­DATA4 VALUE
EMISSIVITY = VALUE / 100
$03 - Max input value -------------------------------------------------------------------- RX
DATA1 ­DATA2 ­DATA3 VALUE (H) DATA4 VALUE (L)
MAX = SIGNED INTEGER ( VALUE ) / 10
$04 - Min input value--------------------------------------------------------------------- RX
DATA1 ­DATA2 ­DATA3 VALUE (H) DATA4 VALUE (L)
MIN = SIGNED INTEGER ( VALUE ) / 10
$05 - Average input value-------------------------------------------------------------- RX
DATA1 ­DATA2 ­DATA3 VALUE (H) DATA4 VALUE (L)
AVERAGE = SIGNED INTEGER ( VALUE ) / 10
$06 - Tag ID,Eng.unit,Tc type,IPTS ------------------------------------------------- RX
DATA1 Tag ID (0=Tag0, 1=TagA, 2=TagB,.....,19=TagS)
DATA2 Eng.unit (1=°C - 2=K - 3=°F) DATA3 Tc type (0=Tc K - 1=Tc S) DATA4 IPTS (0=68 - 1=90)
$07 - Tc measure-------------------------------------------------------------------------- RX
DATA1 STATUS DATA2 ­DATA3 VALUE (H) DATA4 VALUE (L)
IF STATUS=0 MEASURE = SIGNED INTEGER ( VALUE ) / 10 IF STATUS=1 MEASURE = ERROR RJ IF STATUS=2 MEASURE = TC OFF IF STATUS=10 MEASURE = OPTC
$08 - HAL Setpoint------------------------------------------------------------------------ RX
DATA1 ­DATA2 ­DATA3 VALUE (H) DATA4 VALUE (L)
HAL = SIGNED INTEGER ( VALUE ) IF HAL=3001 THEN HAL=OFF
$09 - LAL Setpoint------------------------------------------------------------------------ RX
DATA1 ­DATA2 ­DATA3 VALUE (H) DATA4 VALUE (L)
LAL = SIGNED INTEGER ( VALUE ) IF LAL=3001 THEN LA L=OFF
$0B - First 4 chars tag description-------------------------------------------------- RX
DATA1 'X' (ASCII CODE) DATA2 'X' (ASCII CODE) DATA3 'X' (ASCII CODE)
49
DATA4 'X' (ASCII CODE)
$0C - Last 4 chars tag description-------------------------------------------------- RX
DATA1 'X' (ASCII CODE) DATA2 'X' (ASCII CODE) DATA3 'X' (ASCII CODE) DATA4 -
$0D - Alarm status------------------------------------------------------------------------ RX
DATA1 LAL Status (0=No alarm - 1=Alarm active) DATA1 HAL Status (0=No alarm - 1=Alarm active) DATA1 ­DATA1 -
$1C - Firmware version ----------------------------------------------------------------- RX
DATA1 'X' (ASCII CODE) DATA2 'X' (ASCII CODE) DATA3 'X' (ASCII CODE) DATA4 'X' (ASCII CODE)
15.3.2 Computer data setting from PC to OSP
Computer OSP
Tx IDNAME Rx IDNAME Proceed if the name is acknowledged Rx IDNAME Tx IDNAME If not, do not answer
Tx Instruction Rx Instruction Rx Instruction Tx Instruction
Tx DATA 1 Rx DATA 1 Rx char Tx char
Tx DATA 2 Rx DATA 2 Rx char Tx char
Tx DATA 3 Rx DATA 3 Rx char Tx char
Tx DATA 4 Rx DATA 4 Rx char Tx char
Tx CHKSUM Rx char Rx char Tx char
IDNAME, Instruction, DATA 1, DATA 2, DATA 3, DATA 4 and CHKSUM are 8-bit values (1 byte). CHKSUM = DATA 1+DATA 2+DATA 3+DATA 4 .AND. FF
The above is useful to verify correct received data. The minimum time-out of
OSP
is 3 seconds.
IDNAME is set to 1.
Writing values
$80 - Emissivity---------------------------------------------------------------------------- TX
DATA1 0 DATA2 0 DATA3 0 DATA4 VALUE
VALUE = EMISSIVITY X 100
$82 - Tc type -------------------------------------------------------------------------------- TX
DATA1 0 DATA2 Tc type (0=Tc K - 1=Tc S) DATA3 0 DATA4 0
$84 - HAL Setpoint------------------------------------------------------------------------ TX
DATA1 VALUE (HH) DATA2 VALUE (H) DATA3 VALUE (L)
50
DATA4 VALUE (LL)
SIGNED LONG VALUE = ALARM HI
$85 - LAL Setpoint------------------------------------------------------------------------ TX
DATA1 VALUE (HH) DATA2 VALUE (H) DATA3 VALUE (L) DATA4 VALUE (LL)
SIGNED LONG VALUE = ALARM LO
$86 - Tag ID selection-------------------------------------------------------------------- TX
DATA1 Tag ID (0=Tag0, 1=TagA, 2=TagB,.....,19=TagS)
DATA2 ­DATA3 ­DATA4 -
$87 - Tag description -------------------------------------------------------------------- TX
DATA1 Char position (0÷7) DATA2 'X' (ASCII CODE) DATA3 ­DATA4 -
15.3.3 Communication programs
In this paragraph are illustrated two examples of communication programs between the
OSP
and an IBM or IBM
compatible PC.
Example A: Data transfer from OSP to PC
Connect
OSP
through adapter BB530001 (TTL-RS232 converter), to personal computer communication port COM1.
You will read temperature value.
Example B: PC instructions to OSP
Connect
OSP
through adapter BB530001 (TTL-RS232 converter), to personal computer communication port COM1.
You will set emissivity to 0,80.
Example A
: 1Ø CHAR = Ø 2Ø IDNAME = 1 3Ø INSTRUCTION = 0 35 OPEN "COM1: 96ØØ, N,8,1,CD,CS,DS,RS" FOR RANDOM AS # 1 4Ø PRINT #1, CHR$ (IDNAME);:REM TRANSMIT IDNAME TO OSP 5Ø WHILE LOC (1) = Ø: WEND: REM WAIT RECEIVING IDNAME FROM OSP 6Ø IDNAME = ASC (INPUT$ (1, 1)): REM READ RECEIVED IDNAME FROM OSP 7Ø PRINT #1, CHR$ (INSTRUCTION); 8Ø WHILE LOC (1) = Ø: WEND 9Ø INSTRUCTION = ASC (INPUT$ (1, 1)) 1ØØ PRINT #1, CHR$ (CHAR); 11Ø WHILE LOC (1) = Ø: WEND 12Ø DATA 1 = ASC (INPUT$ (1, 1)) 13Ø PRINT #1, CHR$ (CHAR); 14Ø WHILE LOC (1) = Ø: WEND 15Ø DATA 2 = ASC (INPUT$ (1, 1)) 16Ø PRINT #1, CHR$ (CHAR); 17Ø WHILE LOC (1) = Ø: WEND 18Ø DATA 3 = ASC (INPUT$ (1, 1)) 19Ø PRINT #1, CHR$ (CHAR); 2ØØ WHILE LOC (1) = Ø: WEND 21Ø DATA 4 = ASC (INPUT$ (1, 1)) 22Ø PRINT #1, CHR$ (CHAR); 231Ø WHILE LOC (1) = Ø: WEND 24Ø CHKSUM = ASC (INPUT$ (1, 1)) 25Ø IF CHKSUM <> ((DATA1 + DATA2 + DATA3 + DATA4) AND &HFF) THEN PRINT "Error": END 26Ø VALUE = DATA3 * 256 + DATA4 27Ø IF VALUE > 32767 THEN VALUE =VALUE - 65536: REM 2'S COMPLEMENT
51
28Ø PRINT "VALUE: " ; VALUE / 1Ø 29Ø END
Example B
: 1Ø CHAR = Ø 2Ø IDNAME = 1 3Ø INSTRUCTION = &H80 4Ø VALUE = 80 5Ø VALUE$ = HEX$ (VALUE/100) 55 WHILE LEN (VALUE$)<4: VALUE$ ="Ø"+VALUE$: WEND 6Ø IF LEN (VALUE$) > 4 THEN VALUE$ =RIGHT$ (VALUE$,4) 65 DATA1 = VAL ("&H" + LEFT$ (VALUE$, 2)) 7Ø DATA2 = VAL ("&H" +RIGHT$ (VALUE$, 2)) 75 DATA3 = Ø 8Ø DATA4 = Ø 9Ø CHKSUM = (DATA1 + DATA2 + DATA3 + DATA4) AND &H7F 1ØØ REM MEMORY RELEASE PREVIOUS 4.xx1 (Example 4.ØØØ, 4.2ØØ) MUST USE 11Ø REM NEXT LINE INSTEAD LINE 9Ø OR PROGRAM DOES'NT WORK 12Ø REM CHKSUM =(DATA1 + DATA2 + DATA3 + DATA4) AND &HFF 13Ø OPEN "COM1: 96ØØ,N,8,1,CD,CS,DS,RS" FOR RANDOM AS #1 14Ø PRINT #1, CHR$ (IDNAME) ; : REM TRANSMIT IDNAME TO OSP 15Ø WHILE LOC (1) = Ø: WEND: REM WAIT RECEIVING IDNAME FROM OSP 16Ø IDNAME = ASC (INPUT$ (1 , 1)): REM READ RECEIVED IDNAME FROM OSP 17Ø PRINT #1, CHR$ (INSTRUCTION) ; 18Ø WHILE LOC (1) = Ø : WEND 19Ø INSTRUCTION = ASC (INPUT$ (1 , 1)) 2ØØ PRINT #1, CHR$ (DATA1) ; 21Ø WHILE LOC (1) = Ø : WEND 22Ø CHAR = ASC (INPUT$ (1 , 1)) 23Ø PRINT #1, CHR$ (DATA2) ; 24Ø WHILE LOC (1) = Ø : WEND 25Ø CHAR = ASC (INPUT$ (1 , 1)) 26Ø PRINT #1, CHR$ (DATA3) ; 27Ø WHILE LOC (1) = Ø : WEND 28Ø CHAR = ASC (INPUT$ (1 , 1)) 29Ø PRINT #1, CHR$ (DATA4) ; 3ØØ WHILE LOC (1) = Ø : WEND 31Ø CHAR = ASC (INPUT$ (1 , 1)) 32Ø PRINT #1, CHR$ (CHKSUM) ; 33Ø WHILE LOC (1) = Ø : WEND 34Ø CHAR = ASC (INPUT$ (1 , 1)) 35Ø PRINT "Transmitted." 39Ø END
52
16 MAINTENANCE
Each
OSP
portable infrared thermometer has been factory tested and calibrated before shipment. The calibration should be verified and re-adjusted if the instrument shows an error exceeding the declared specifications or when a critical active or passive component is replaced (either at the component level or at the board level).
OMEGA
engineers will give prompt support for any request of assistance.
16.1 Faulty operating conditions
During the start-up and measuring modes, faulty conditions of the instrument will be announced with coded messages. If the faulty condition is critical for the application, it is recommended to re-start the pertinent set-up procedure. All errors which cannot be recovered without the user's knowledge, result in some system action to inform the operator via message and where possible to restore the system. Restarting the instrument from -Power ON- may clear the error, but generally such messages are caused by hardware or software faults, which require the user's action.
Display Error Type Solution
Err 1
Internal EEPROM memory error Contact technical support
Err 2
IR sensor reading error Verify ambient temperature
(from -5 to +55°C)
Err 3
Calibration data error Recalibrate unit
OTC
No sensor connected at Tc input Connect sensor
RJer
Cold-junction sensor reading error Verify ambient temperature
(from -5 to +55°C)
16.2 Storage
If the instrument has been left unused for a long time, it is recommended to remove the batteries. Store the instrument in the original package, at a temperature from -30°C to +60°C, with R.H. less than 90%. If the instrument has been unused for a month check the battery voltage, and charge the Ni-MH batteries for at least 12 hours with the unit switched off.
53
APPENDIX
54
A1 How to determine an object emissivity
Emissivity is the measure of an object ability to absorb, transmit, and emit infrared energy. It can have a value from 0 (shiny mirror) to 1.0 (blackbody). If a value of emissivity higher than the actual one is set, the output will read low, provided that the target temperature is above the ambient one. For example, if 0.95 is set in and the actual emissivity is
0.9, the reading will be lower than the true temperature when the target temperature is above the ambient one.
The emissivity can be determined by one of the following methods, in order of preference:
1. Determine the actual temperature of the material using a sensor such as a RTD, thermocouple or another suitable
method. Next, measure the object temperature and adjust the emissivity setting until the correct value is reached. This is the correct emissivity for the measured material.
2. For relatively low temperature objects (up to 260°C or 500°F, place a piece of tape, such as a masking, on the
object. Make sure the tape is large enough to cover the field of view. Next, measure the tape temperature using an emissivity setting of 0.95. Finally, measure an adjacent area on the object and adjust the emissivity setting until the same temperature is reached. This is the correct emissivity for the measured material.
3. If a portion of the surface of the object can be coated, use a flat black paint, which will have an emissivity of about
0.98. Next, measure the painted area using an emissivity setting of 0.98. Finally, measure an adjacent area on the object and adjust the emissivity setting until the same temperature is reached. This is the correct emissivity for the measured material.
A1.1 Typical Emissivity Values
The following table provides a brief reference guide to determine emissivity and can be used when one of the above methods is not practical. Emissivity values shown in the table below are only approximate, since several parameters may effect the emissivity of an object. These include the following ones:
1. Temperature
2. Angle of measurement
3. Geometry (plane, concave, convex, etc.)
4. Thickness
5. Surface quality (polished, rough, oxidized, sandblasted)
6. Spectral region of measurement
7. Transmissivity (i.e., thin film plastics)
55
A1.2 Metals - Typical Emissivity Values
1.0 µm 1.6 µm 5.1 µm 8-14 µm
Aluminum
Non-Oxidized 0.1-0.2 0.02-0.2 0.02-0.2 0.02-0.1 Oxidized 0.4 0.4 0.2-0.4 0.2-0.4
Alloy A 3003
Oxidized 0.4 0.4 0.3 Roughened 0.2-0.8 0.2-0.6 0.1-0.4 0.1-0.3 Polished 0.1-0.2 0.02-0.1 0.02-0.1 0.02-0.1
Brass
Polished 0.8-0.95 0.01-0.05 0.01-0.05 0.01-0.05 Burnished 0.3 0.3 Oxidized 0.6 0.6 0.5 0.5
Carbon
Non-oxidized 0.8-0.95 0.8-0.9 0.8-0.9 0.8-0.9
Graphite 0.8-0.9 0.8-0.9 0.7-0.9 0.7-0.8 Chromium 0.4 0.4 0.03-0.3 0.02-0.2 Copper
Polished 0.05 0.03 0.03 0-03
Roughened 0.05-0.2 0.05-0.2 0.05-0.15 0.05-0.1
Oxidized 0.2-0.8 0.2-0.9 0.5-0.8 0.4-0.8 Gold 0.3 0.01-0.1 0.01-0.1 0.01-0.1 Haynes Alloy 0.5-0.9 0.6-0.9 0.3-0.8 0.3-0.8 Inconel
Oxidized 0.4-0.9 0.6-0.9 0.6-0.9 0.7-0.95 Sandblasted 0.3-0.4 0.3-0.6 0.3-0.6 0.3-0.6 Electro-polished 0.2-0.5 0.25 0.15 0.15
Iron
Oxidized 0.4-0.8 0.5-0.9 0.6-0.9 0.5-0.9
Non-oxidized 0.35 0.1-0.3 0.05-0.25 0.05-0.2 Rusted 0.6-0.9 0.5-0.8 0.5-0.7 Molten 0.35 0.4-0.6
Iron Cast
Oxidized 0.7-0.9 0.7-0.9 0.65-0.95 0.6-0.95
Non-oxidized 0.35 0.3 0.25 0.2
Molten 0.35 0.3-0.4 0.2-0.3 0.2-0.3
Iron Wrought
Dull 0.9 0.9 0.9 0.9
Lead
Polished 0.35 0.05-0.2 0.05-0.2 0.05-0.1 Rough 0 65 0.6 0.4 0-4 Oxidized 0.3-0.7 0.2-0.6 0.2-0.6
Magnesium 0.3-0.8 0.05-0.3 0.03-0.15 0.02-0.1 Mercury 0.05-0.15 0.05-0.15 0.05-0.15 Molybdenum
Oxidized 0.5-0.9 0.4-0.9 0.3-0.7 0.2-0.6
Non-oxidized 0.25-0.35 0.1-0.3 0.1-0.15 0.1 Monel (Ni-Cu) 0.3 0.2-0.6 0.1-0.5 0.1-0.14 Nickel
Oxidized 0.8-0.9 0.4-0.7 0.3-0.6 0.2-0.5 Electrolytic 0.2-0.4 0.1-0.3 0.1-0.15 0.05-0.15
Platinum
Black 0.95 0.9 0.9 Silver 0.04 0.02 0.02 0.02 Steel
Cold-Rolled 0.8-0.9 0.8-0.9 0.8-0.9 0.7-0.9 Ground Sheet 0.5-0.7 0.4-0.6 Polished Sheet 0.35 0.25 0.15 0.1 Molten 0.35 0.25-0.4 0.1-0.2 — Oxidized 0.8-0.9 0.8-0.9 0.7-0.9 0.7-0.9
Stainless 0.35 0.2-0.9 0.15-0.8 0.1-0.8 Tin (Non-oxidized) 0.25 0.1-0.3 0.05 0.05 Titanium
Polished 0.5-0.75 0.3-0.5 0.1-0.3 0.05-0.2 Oxidized 0.6-0.8 0.5-0.7 0.5-0.6
56
1.0 µm 1.6 µm 5.1 µm 8-14 µm
Tungsten 0.1-0.6 0.05-0.5 0.03
Polished 0.35-0.4 0.1-0.3 0.05-0.25 0.03-0.1 Zinc
Oxidized 0.6 0.15 0.1 0.1 Polished 0.5 0.05 0.03 0.02
A1.3 Non-metals - Typical Emissivity Values
1.0 µm 1.6 µm 5.1 µm 8-14 µm
Asbestos 0.9 0.9 0.95 0.95 Asphalt 0.95 0.95 0.95 Basalt 0.7 0.7 0.7 Carborundum 0.9 0.9 0.9 Ceramic 0.4 0.8-0.95 0.95 0.95 Clay 0.8-0.95 0.95 0.95 Concrete 0.65 0.9 0.95 0.95 Cloth 0.95 0 95 0.95 Glass
Plate 0.98 0.85 0.85
“Gob” 0.9 — Gravel 0.95 0.95 0.95 Gypsum 0.4-0.97 0.8-0.95 0.8-0.95 Ice 0.98 0.98 Limestone 0.4-0.98 0.98 0.98 Paint — 0.9-0.95 0.9-0.95 Paper (any color) 0.95 0.95 0.95 Plastic (opaque, 0.95 0.95 0.95
over 20 mils) Rubber 0.9 0.9 0 95 Sand 0.9 0.9 0.9 Snow 0.9 0.9 0.9 Soil 0.9-0.98 0.9-0.98 Water 0.93 0.93 Wood, Natural 0.9-0.95 0.9-0.95 0.9-0.95
To optimize surface temperature measurements consider the following guidelines:
1. Determine the object emissivity using the suitable instrument for measurement.
2. Avoid reflections by shielding the object from surrounding high temperature sources.
3. For higher temperature objects use shorter wavelength instruments, whenever any overlap occurs.
4. For semi-transparent materials such as plastic films and glasses, assure that the background is uniform and lower in
temperature than the object.
5. Mount the sensor perpendicularly to the surface whenever the emissivity is less than 0.9. In any case, do not exceed
angles more than 30 degrees from incidence.
57
INDEX
A
Acquisition by dedicated Tags; 32 Acquisition settings; 28 Alarm settings; 23 Ambient Temperature Compensation (TAM); 29 Analog Output; 10 Analogue input Warnings & Cautions; 18 APPLICATION NOTES; 45 Applications; 45 Automatic Emissivity Setting; 26
B
Battery charger; 14 Battery level of charge; 31 Buzzer On/Off; 29
C
Calculated Measurements; 11 Case; 11 Communication programs; 50 Communication protocol; 47 Computer data request from OSP to PC; 47 Computer data setting from PC to OSP; 49 CONFIGURATION; 27 Continuous acquisition; 32
D
Danger and Certification Labels; 18 DATA ACQUISITION OPERATIVE MODE; 32 Date & Real Time clock setting; 28 Decimal point selection; 29 Digital Inteface; 10 Digital interface; 14; 46 Digital output wiring practice; 46 Display; 10; 14
E
Emissivity; 45
F
Faulty operating conditions; 52 Firmware; 14 Firmware revision number; 30 FUNCTIONAL DESCRIPTION; 13
G
General; 9 GENERAL DESCRIPTION; 6 GENERAL FEATURES; 9
H
Hold; 21 How to determine an object emissivity; 54 How to maximize the life span of the battery; 17 How to operate in data acquisition mode; 32 How to Operate the instrument; 20
I
Infrared energy; 45 Instrument codes; 7 Instrument serial number; 30 INTRODUCTORY NOTE; 3
K
Keyboard; 10; 13
L
Laser Sight Warnings & Cautions; 18 Laser Sighting; 21 Logging Data Manager; 11 LOGGING DATA MANAGER; 35 Logging Mode; 11 LogMan
Program architecture; 35
LogMan
Installation; 35 Quick starting; 38 Toolbars; 37
M
MAINTENANCE; 52 Metals - Typical Emissivity Values; 55 Microcontroller; 14 Minimum, maximun and difference indication; 25
N
Non-metals - Typical Emissivity Values; 56
O
OPERATIONS; 20 Optical System; 9 OPTIONS & ACCESSORIES; 42
P
PHYSICAL DESCRIPTION; 12 Power supply; 13 POWER SUPPLY; 17 Power supply from main line; 17 Power supply with alkaline batteries; 17 PRE-OPERATIONAL CHECK; 16 Printer operations; 42
Q
Quick Start; 20
R
Recall stored data; 34 Rechargeable batteries; 17 Red Point pinpointing system; 43 Reflected energy compensation; 45
58
S
S
AFETY WARNINGS
; 18 Self calibration; 10 Sighting telescope system; 42 Specifications; 8 Storage; 52
T
TABLE OF CONTENTS; 4 Taking measurements; 9 Target dimensions at different distances; 8 Target pinpointing; 9 Technical Unit selection; 27 Temperature Scale selection; 27
Thermocouple input; 24 Thermocouple Input; 10 Thermocouple input enabled; 27 TTL to RS 232 adaptor; 46
U
UNPACKING; 15 Unstable temperature measurement; 22
W
Warnings and Cautions; 18
Weight
; 22
Working band
; 22
59
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of
13
months
from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal
one (1)
year product warranty
to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum
coverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WAR­RANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSEO OR IMPUED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTlES INCLUDING ANY WARRANTY OF MERCHANTABIUTY AND RTNESS FOR A PARTlCULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATlON OF LIABILITY: The remedies of purchaser set forth herein ate exclusive and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence. Indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a ”Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR
WARRANTY
RETURNS, please has the following information available BEFORE contacting OMEGA:
1. P.O. number under which the product was
PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
FOR
NON-WARRANTY
REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems relative to
the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. (C) Copyright 1999 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced,
translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.
PATENT NOTICE: U.S. PAT. B1 5,368,392; 5,524,984; 5,727,880; 5,465,838; 5,823,678; 5,823,678; 5,823,679. Other U.S. and
Foreign Patents and Applications Pending. Manufactured in Italy. Licensed by Omega Engineering Inc.
60
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
TEMPERATURE
!!!!
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
!!!!
Wire: Thermocouple, RTD & Thermistor
!!!!
Calibrators & Ice Point References
!!!!
Recorders, Controllers & Process Monitors
!!!!
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
!!!!
Transducers & Strain Gauges
!!!!
Load Cells & Pressure Gauges
!!!!
Displacement Transducers
!!!!
Instrumentation & Accessories
FLOW/LEVEL
!!!!
Rotameters, Gas Mass Flowmeters & Flow Computers
!!!!
Air Velocity Indicators
!!!!
Turbine/Paddlewheel Systems
!!!!
Totalizers & Batch Controllers
pH/CONDUCTIVITY
!!!!
pH Electrodes, Testers & Accessories
!!!!
Benchtop/Laboratory Meters
!!!!
Controllers, Calibrators, Simulators & Pumps
!!!!
Industrial pH & Conductivity Equipment
DATA ACQUISITION
!!!!
Data Acquisition & Engineering Software
!!!!
Communications-Based Acquisition Systems
!!!!
Plug-in Cards for Apple, IBM & Compatibles
!!!!
Datalogging Systems
!!!!
Recorders, Printers & Plotters
HEATERS
!!!!
Heating Cable
!!!!
Cartridge & Strip Heaters
!!!!
Immersion & Band Heaters
!!!!
Flexibie Heaters
!!!!
Laboratory Heaters