Compact Non-Contact
Temperature Sensor/Transmitter
with Optional Dual Output
OS210, OS150, OS300
and OS800 Series
OS211MT-MV
shown smaller than
Optional
U Temperature Range: -20 to 500°C
(-4 to 932ºF)
U Emissivity Setting: Fixed at 0.95
U 2-Wire 4 to 20 mA Output or 4-Wire
Voltage/Thermocouple Output
U 4-Wire Unit Provides Internal
Sensor Temperature
U Field of View: 2:1, 15:1, 30:1 or
Close Focus
U Fast Response with High Stability
U Stainless Steel Housing, Sealed to IP65
U Quick and Easy Installation
U Optional Air/Water Cooled Housing,
Air Purge Collar, Laser Sighting Tool
and Mounting Brackets
This Series is a range of high quality, low cost,
compact sensors which measure the temperature
of inaccessible or moving objects and materials.
They measure temperatures from -20 to 500°C
(-4 to 932ºF), accurately and consistently, with an
outstanding response time of 240 ms. Infrared (IR)
sensors are available as either 2- or 4-wire units.
Two-wire IR sensors transmit the target temperature
as a 4 to 20 mA output and offer a simple solution
for most non-contact temperature measurement
applications. Four-wire IR sensors transmit the target
temperature as a 0 to 50 mV or thermocouple output
(Type J, K or T) plus the internal sensor temperature
as a 4 to 20 mA output. This second output can be
used to ensure that the sensor is being operated within
the correct ambient temperature limits and prevent
damage caused by overheating or overcooling. It can
also be used to give an approximate indication of
the air temperature surrounding the sensor. IR sensors
are fitted with precision germanium lenses
for accurate optics.
Specifications
Accuracy: ±1% of reading or ±1ºC whichever is greater
Repeatability: ±0.5% of reading or ±0.5ºC whichever
is greater
Emissivity: Fixed at 0.95
Response Time, t90: 240 ms (90% response)
Spectral Range: 8 to 14 μm
Supply Voltage: 24 Vdc (28 Vdc maximum)
Min Sensor Voltage: 6 Vdc
Max Loop Impedance: 900 Ω (4 to 20 mA output)
Output Impedance: 56 Ω (voltage/thermocouple output)
Mechanical
Construction: Stainless steel
Dimensions: 103 L x 18 mm D (4.06 x 0.71")
Thread Mounting: M16 x 1 mm pitch
Cable Length: 1 m (3.3') (included) longer lengths
available to order
Weight with Cable: 95 g (3.35 oz)
Environmental
Rating: IP65
Ambient Temperature Range: 0 to 70ºC (32 to 158ºF)
Relative Humidity: 95% max non-condensing
Compact stainless steel
IP65 housing.
actual size.
J
J-w

OS210 Series with 2:1 Field of View
Distance: Sensor to object (in) Distance: Sensor to object (in)
Distance: Sensor to object (mm)
Distance: Sensor to object (mm)
11.9
0.5
0
0
11.9
0.5
0
0
5.0
0.20
100
3.9
12.5
0.49
200
7.9
28.6
45.2
1.1
1.8
19.7 39.4
500 1000
D:S 30:1
OS300 Series
Spot
Dia.
(in)
Spot
Dia.
(in)
Spot
Dia.
(mm)
Spot
Dia.
(mm)
45.2
78.6
1.8
3.1
19.7 39.4
Distance: Sensor to object (in) Distance: Sensor to object (in) Distance: Sensor to object (in)
Distance: Sensor to object (mm) Distance: Sensor to object (mm)
Distance: Sensor to object (mm)
11.9
0.5
0
0
11.9
0.5
0
0
11.9
0.5
0
0
OS800 Series
OS150 Series
28.6
45.2
1.1
1.8
19.7 39.4
500 1000
D:S 30:1
OS300 Series
Spot
Dia.
(in)
Spot
Dia.
(in)
Spot
Dia.
(in)
Spot
Dia.
(mm)
Spot
Dia.
(mm)
Spot
Dia.
(mm)
Distance: Sensor to object (in)
Distance: Sensor to object (mm)
11.9
0.5
0
0
5.0
0.20
100
3.9
12.5
0.49
200
7.9
Spot
Dia.
(in)
Spot
Dia.
(mm)
To Order
Model No. Temperature Range Output
OS211-LT -20 to 100ºC (-4 to 212ºF) 4 to 20 mA
OS211-MT 0 to 250ºC (32 to 482ºF) 4 to 20 mA
OS211-LT-MV -20 to 100ºC (-4 to 212ºF) 0 to 50 mV
OS211-LT-T -20 to 100ºC (-4 to 212ºF) Type T thermocouple
OS211-LT-J -20 to 100ºC (-4 to 212ºF) Type J thermocouple
OS211-LT-K -20 to 100ºC (-4 to 212ºF) Type K thermocouple
OS211-MT-MV 0 to 250ºC (32 to 482ºF) 0 to 50 mV
OS211-MT-T 0 to 250ºC (32 to 482ºF) Type T thermocouple
OS211-MT-J 0 to 250ºC (32 to 482ºF) Type J thermocouple
OS211-MT-K 0 to 250ºC (32 to 482ºF) Type K thermocouple
Ordering Example: OS211-MT, 2:1 field of view, 0 to 250ºC (32 to 482ºF) temperature range,
4 to 20 mA output.
OS150 Series with 15:1 Field of View
Model No. Temperature Range Output
OS151-LT -20 to 100ºC (-4 to 212ºF) 4 to 20 mA
OS151-MT 0 to 250ºC (32 to 482ºF) 4 to 20 mA
OS151-HT 0 to 500ºC (32 to 932ºF) 4 to 20 mA
OS151-LT-MV -20 to 100ºC (-4 to 212ºF) 0 to 50 mV
OS151-LT-T -20 to 100ºC (-4 to 212ºF) Type T thermocouple
OS151-LT-J -20 to 100ºC (-4 to 212ºF) Type J thermocouple
OS151-LT-K -20 to 100ºC (-4 to 212ºF) Type K thermocouple
OS151-MT-MV 0 to 250ºC (32 to 482ºF) 0 to 50 mV
OS151-MT-T 0 to 250ºC (32 to 482ºF) Type T thermocouple
OS151-MT-J 0 to 250ºC (32 to 482ºF) Type J thermocouple
OS151-MT-K 0 to 250ºC (32 to 482ºF) Type K thermocouple
OS151-HT-MV 0 to 500ºC (32 to 482ºF) 0 to 50 mV
OS151-HT-J 0 to 500ºC (32 to 932ºF) Type J thermocouple
OS151-HT-K 0 to 500ºC (32 to 932ºF) Type K thermocouple
Ordering Example: OS151-MT-K, 15:1 field of view, 0 to 250ºC (32 to 482ºF) Type K
thermocouple output.
Distance: Sensor to object (in)
0
Spot
Dia.
(in)
0.5
11.9
Spot
Dia.
(mm)
0
Distance: Sensor to object (mm)
Diameter of target spot measured vs
distance from sensing head
Distance: Sensor to object (in)
0
Spot
Dia.
(in)
0.5
11.9
Spot
Dia.
(mm)
0
Distance: Sensor to object (mm)
Diameter of target spot measured vs
distance from sensing head
4
2.4
D:S 2:1
61.9
100
OS210 Series
19.7 39.4
1.8
D:S 15:1
45.2
500 1000
OS150 Series
8
4.4
111.9
200
3.1
78.6
OS300 Series with 30:1 Field of View
Model No. Temperature Range Output
OS301-LT -20 to 100ºC (-4 to 212ºF) 4 to 20 mA
OS301-MT 0 to 250ºC (32 to 482ºF) 4 to 20 mA
OS301-HT 0 to 500ºC (32 to 932ºF) 4 to 20 mA
OS301-LT-MV -20 to 100ºC (-4 to 212ºF) 0 to 50 mV
OS301-LT-T -20 to 100ºC (-4 to 212ºF) Type T thermocouple
OS301-LT-J -20 to 100ºC (-4 to 212ºF) Type J thermocouple
OS301-LT-K -20 to 100ºC (-4 to 212ºF) Type K thermocouple
OS301-MT-MV 0 to 250ºC (32 to 482ºF) 0 to 50 mV
OS301-MT-T 0 to 250ºC (32 to 482ºF) Type T thermocouple
OS301-MT-J 0 to 250ºC (32 to 482ºF) Type J thermocouple
OS301-MT-K 0 to 250ºC (32 to 482ºF) Type K thermocouple
OS301-HT-MV 0 to 500ºC (32 to 932ºF) 0 to 50 mV
OS301-HT-J 0 to 500ºC (32 to 932ºF) Type J thermocouple
OS301-HT-K 0 to 500ºC (32 to 932ºF) Type K thermocouple
Ordering Example: OS301-HT-K, 30:1 field of view, 0 to 500ºC (32 to 932ºF) temperature
range, Type K thermocouple output. OCW-3, OMEGACARESM extends standard 2-year
warranty to a total of 5 years.
J-x
Distance: Sensor to object (in)
0
Spot
Dia.
(in)
0.5
11.9
Spot
Dia.
(mm)
0
Distance: Sensor to object (mm)
Diameter of target spot measured vs
OS300 Series
distance from sensing head
19.7 39.4
1.1
28.6
500 1000
1.8
D:S 30:1
45.2

OS800 Series with Close-Focus Field of View
To Order
Model No. Temperature Range Output
OS801-LT -20 to 100ºC (-4 to 212ºF) 4 to 20 mA
OS801-MT 0 to 250ºC (32 to 482ºF) 4 to 20 mA
OS801-HT 0 to 500ºC (32 to 932ºF) 4 to 20 mA
OS801-LT-MV -20 to 100ºC (-4 to 212ºF) 0 to 50 mV
OS801-LT-T -20 to 100ºC (-4 to 212ºF) Type T thermocouple
OS801-LT-J -20 to 100ºC (-4 to 212ºF) Type J thermocouple
OS801-LT-K -20 to 100ºC (-4 to 212ºF) Type K thermocouple
OS801-MT-MV 0 to 250ºC (32 to 482ºF) 0 to 50 mV
OS801-MT-T 0 to 250ºC (32 to 482ºF) Type T thermocouple
OS801-MT-J 0 to 250ºC (32 to 482ºF) Type J thermocouple
OS801-MT-K 0 to 250ºC (32 to 482ºF) Type K thermocouple
OS801-HT-MV 0 to 500ºC (32 to 932ºF) 0 to 50 mV
OS801-HT-J 0 to 500ºC (32 to 932ºF) Type J thermocouple
OS801-HT-K 0 to 500ºC (32 to 932ºF) Type K thermocouple
Comes complete with sensor head, 1 m (3.3') cable and operator’s manual.
Ordering Example: OS801-MT-K, close-focus field of view, 0 to 250°C (32 to 932°F)
Type K thermocouple output.
Distance: Sensor to object (in)
0
Spot
Dia.
(in)
0.5
11.9
Spot
Dia.
(mm)
0
Distance: Sensor to object (mm)
Diameter of target spot measured vs
distance from sensing head
3.9
0.20
5.0
100
OS800 Series
7.9
0.49
12.5
200
PSR-24L, $60, shown
smaller than actual size.
PSR-24S, $60, shown
smaller than actual size.
SM
Extended Warranty
Program
OMEGACARESM extended warranty program
is available for models shown on this page.
Ask your sales representative for full details
when placing an order. OMEGACARESM
covers parts, labor and equivalent loaners.
Power Supplies
Model No. Description
PSR-24S Regulated power supply, US plug, 90 to 264 Vac
input, 24 Vdc output, 400 mA, screw terminals, UL
PSR-24L Regulated power supply, US plug, 90 to 264 Vac
input, 24 Vdc output, 400 mA, stripped leads, UL
PSR-24L-230 Regulated power supply, European plug, 230 Vac
input, 24 Vdc output, 400 mA, stripped leads, CE
PSU-93 Unregulated power supply, 16 to 23 Vdc, 300 mA
maximum, screw terminal
TX8-100 Power/output cable, 30 m (98')
Options (Not Field Installable)
Suffix Description
-WC For air/water cooled jacket (includes air purge collar);
must be ordered with sensor; cannot be retrofitted
-CE(*) Cable extended, for uncooled PC sensor, standard
4 to 20 mA output, 30 m (98') maximum total cable length
-CEV(*) Cable extended, for PC sensor, for 0 to 50 mV output, 30
m (98') maximum total cable length
-CET(*) Cable extended, for PC sensor, for Type T thermocouple
output, 30 m (98') maximum total cable length
-CEJK(*) Cable extended, for PC sensor for Types J and K
thermocouple output, 30 m (98') maximum total cable
length
-CEHT(*) Cable extended, high temperature, for cooled PC
standard 4 to 20 mA output, 30 m (98') maximum total
cable length
1 m (3.3') cable is included as standard on every sensor. Extended cable is added to that length,
*
add suffix to model number.
J
J-y

Accessories
OS210-FBS
Fixed Mounting Bracket
The L-shaped fixed mounting bracket offers a rigid support
for the sensor and allows fine adjustment in a single plane.
OS210-APSN
OS210-ABS
Adjustable Mounting Bracket
The adjustable mounting bracket consists of a fixed mounting
bracket plus another L-shaped bracket. When assembled as
shown the adjustable mounting bracket offers a rigid support for
the sensor and allows fine adjustment in two planes.
Option “WC” air/water cooled
jacket, additional cost.
Air Water Cooled Housing
The air/water cooled housing allows the sensor to withstand
ambient temperatures which exceed the normal 70°C limit. Air or
water (depending on the degree of cooling required) flows into
Air Purge Collar
The air purge collar is used to keep dust, fumes, moisture and
other contaminants away from the lens. Air flows into the fitting
on the side and out of the aperture at the front.
Laser Sighting Tool
The laser sighting tool screws onto the front of the sensor during
installation and indicates precisely where the sensor is aiming.
Once the sensor has been aimed at the centre of the target and
locked in position the laser sighting tool can be removed. The
laser is activated by means of a push button on the front of the
tool which has a latching mechanism.
one of the fittings on the side and out of the other. To prevent
condensation forming on the lens, the air/water cooled housing is
supplied complete with an air purge collar. Please note, the air/
water cooled housing must be ordered with the sensor and cannot
be fitted by the user.
OS210-LSTS
All models shown
smaller than actual size.
Accessories
Model No. Description
OS210-FBS Fixed bracket for compact sensors
OS210-ABS Adjustable bracket for compact sensors
OS210-APSW Air purge collar for compact 2:1 sensors
OS210-APSN Air purge collar for compact 15:1, 30:1 and close-focus sensors
OS210-LSTS Laser sighting tool for compact sensors
CAL-3-IR† NIST-traceable calibration
J-z