OMEGA OS-1300-161 Manual

Page 1
TM
U
ser’s Guide
Shop online
omega.com
e-mail: info@omega.com
For latest product
manuals:
www.omegamanual.info
OS1300 Series
Performance Handheld Infrared
Thermometer
Page 2
En
gineering Service: 1
-
800-872-9436 (USA & Canada only)
Headquarters:
Tel: (203) 359
-
1660
Fax: (203) 359
-
7700
alter specifications without
notice.
U.S.A.
omega.com info@omega.com
Servicing North America:
Omega Engineering, Inc. 800 Connecticut Ave. Suite 5N01, Norwalk, CT 06854
Toll-Free: 1-800-826-6342 (USA & Canada only) Customer Service: 1-800-622-2378 (USA & Canada only)
e-mail: info@omega.com
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains and reserves the right to
Page 3
Table of Contents
Table of Contents ............................................................................................................................................. 3
1) Introduction .............................................................................................................................................. 4
1.1 How it works ....................................................................................................................................................4
1.2 Caution ............................................................................................................................................................5
2) Specifications ............................................................................................................................................ 6
3) Operating Instructions ............................................................................................................................... 6
3.1 Battery Installation: .........................................................................................................................................6
3.2 Button and Key Layout: ....................................................................................................................................7
3.3 Locating a Hotspot: ..........................................................................................................................................7
4) Maintenance ............................................................................................................................................. 8
Page 4
1) Introduction
The OS1300-Series handheld infrared thermometer is a non-contact temperature sensing thermometer. It has a wide temperature range of -50 to 1300°C (-58 to 2372°F). This thermometer features a backlit display which is automatically illuminated when the trigger is pulled and allows working in poor or zero-lighting conditions. The single laser is also automatically active when the trigger is pulled. The laser sight can be turned on and off with the press of a button. The laser is located above the IR sensor to locate the center of the area being measured. The field of view is 16:1. The thermometer constantly updates the temperature measurement as long as the trigger is held in and will “hold” on the last temperature seen when the trigger is released. The OS1300-Series has adjustable emissivity, min, max, and averaging functions, as well as high and low alarm capability. This safely measures surface temperatures of hot, hazardous, or hard-to-reach objects without contact.
1.1 How it works
This handheld infrared thermometer measures the surface temperature of an object. The unit’s optics sense emitted, reflected, and transmitted energy which is collected and focused onto a detector. The unit’s electronics converts the energy value into a temperature in °C or °F which is displayed on the digital display. A clean line of sight free of dust or mist is needed between the sensor and the object. For increased ease of use and accuracy the laser pointer makes aiming even more precise.
Measurement:
When taking measurement, point the thermometer toward the object to be measured and hold the trigger. The temperature will be displayed and the word “SCAN” will flash on the left-hand side of the LCD screen to indicate that the temperature is continually being updated on the display. When the trigger is released, the last reading seen will be held on the display and the word “HOLD” will be visible on the left-hand side of the LCD screen. The object under test should be larger than the spot size calculated by the field of view diagram below.
Field of View:
Field of view, or distance to spot ratio, is the size of an area that can be measured from a specified distance. As the distance from the object increases, the spot size of the measuring area becomes larger. Make sure the target is larger than the unit’s spot size. When accuracy is critical, ensure that the target is at least twice as large as the spot size. Please see field of view diagram for the OS1300-Series below.
Emissivity:
Most organic materials and painted or oxidized surfaces have an emissivity of 0.95 (pre-set in the unit but adjustable). Inaccurate readings will result from measuring shiny or polished metal surfaces. To compensate, cover the measured surface with masking tape of flat black paint. Measure the tape or painted surface when the tape or painted surface reaches the material temperature. Please see the Emissivity table below for common emissivity values and adjust accordingly.
Page 5
Emissivity Table:
Material
Emissivity
Material
Emissivity
Aluminum
0.30 Iron 0.70 Asbestos
0.95 Lead
0.50 Asphalt
0.95 Limestone
0.98 Basalt
0.70 Oil 0.94 Brass
0.50 Paint
0.93 Brick
0.90 Paper
0.95 Carbon
0.85 Plastic
0.95 Ceramic
0.95 Rubber
0.95 Concrete
0.95 Sand
0.90 Copper
0.95 Skin 0.98 Dirt 0.94 Snow
0.90 Frozen Food
0.90 Steel
0.80 Hot food
0.93 Textiles
0.94 Glass (plate)
0.85 Water
0.93 Ice 0.98 Wood
0.94
1.2 Caution
Infrared thermometer should be protected for the following:
- EMF (electro-magnetic fields) from arc welders and induction heaters.
- Thermal shock (caused by large or abrupt ambient temperature changes allow 30 minutes for unit to stabilize before use).
- Do not leave the unit on or near objects of high temperature.
Warning: Do not point laser directly at eye or indirectly off reflective surfaces (see diagram below)
Page 6
2) Specifications
Temperature Range -50 to 1300°C (-58 to 2372°F) Accuracy 3% of reading + 3°C, -50°C to 0°C (-58 to 32°F)
2% of reading +2°C, above 0°C
Field of View 16:1 Repeatability 1% of reading or 1°C Response Time 500 ms (95% response) Spectral Response 8 – 14 µm Emissivity Adjustable from 0.1 to 1.0 Relative Humidity 10 -95% RH noncondensing Storage Temperature -20 - 60°C (-4 - 140°F) Weight 130 g (4.86 oz) without battery Dimensions 5.74 x 3.14 x 1.49 in (146 x 80 x 38 mm) Power 9V Battery Estimated Battery Life 12 hours (using laser)
3) Operating Instructions
3.1 Battery Installation:
Use the thumb and forefinger to pull on the finger grips on either side of the trigger of the thermometer (Figure 1). The battery compartment door will swing down (Figure 2). Insert a 9 Volt alkaline battery into the battery holder (Figure 3). Take care not to pinch the battery wires when closing the battery door back to its original position.
Figure 1 Figure 2 Figure 3
Page 7
3.2 Button and key Layout:
8
2 3
4
5 1
7
6
3.3 Operating Steps
1) Hold the thermometer by the handle grip and point it towards the surface to be measured.
2) Pull and hold the trigger to turn the thermometer on. The “SCAN” icon will appear on the top line of the display.
3) The surface temperature being measured will be displayed on the LCD screen.
4) When the trigger is released, the “HOLD” icon will appear and the last reading seen will remain on the display screen.
5) The thermometer will automatically turn off after 7 seconds of inactivity.
3.4 Button function
3.4.1 Mode Selection
1) With trigger held, press the “MODE” button [3] to cycle through the different modes. a. “AVG” displays average temperature reading seen during the session. b. “MAX” displays maximum temperature reading seen during session. c. “MIN” displays minimum temperature reading seen during session. d. “HAL” shows High Alarm value (use UP and DN buttons to adjust up or down). e. “LAL” shows Low Alarm value (use UP and DN buttons to adjust up or down). f. “DIF” shows realtime difference between current reading and original starting value.
3.4.2 Changing Temperature Units
1) With trigger [1] held and “SCAN” flashing on display, press “°C/DN” [4] to select temperature
units in Celsius. Press “F/UP”[5] to select temperature units in Fahrenheit.
3.4.3 Adjusting Laser
1) While holding the trigger and “SCAN” activated on display, press the “Laser” button [8] to
toggle the single laser dot on and off.
3.4.4 Adjusting Emissivity
1) With unit turned “on”, press the “MODE” button [3] a few times to get to the “Є” annunciator which will show the current Emissivity value.
2) Press the Backlight/UP button [2] to increase value and the “T/DN” button [4] to decrease value.
3) When desired value is reached, press “SET” again to lock in the value.
3.4.5 Adjusting High/Low Alarm
1) With trigger held and “SCAN” activated on the display, press the “Mode” button [3] one press a time to get to “LAL” (low alarm) or “HAL” (high alarm) annunciator.
2) Press the Backlight/UP button [2] to increase value and the “T/DN” button [4] to decrease value.
Page 8
3.5 Locating a Hotspot:
To find a hot spot, aim the thermometer outside the area of interest, then scan across with up and down motions until you locate the hot spot. Turning on the laser will make for more precise targeting.
3.3 Locating a Hotspot:
To find a hot spot aim the thermometer outside the area of interest, then scan across with up and down motions until you locate the hot spot. Turning on the laser will make for more precise targeting.
4) Maintenance
1) Lens cleaning : blow off loose particles using clean compressed air. Gently brush remaining debris away with a moist cotton cloth.
2) Case cleaning: clean the case with a damp sponge/cloth and mild soap.
Note: Do not use solvent to clean lens and do not submerge the unit into water or any other liquid.
Page 9
13
grace period to the normal
ensures that OMEGA’s customers receive
Department
Upon examination by
charge. OMEGA’s WARRANTY does
including but not limited to mishandling, improper
, or unauthorized modification. This WARRANTY is
OMEGA neither assumes
indemnification, strict liability or otherwise, shall not exceed
applications or used on
as set forth in our basic WARRANTY/DISCLAIMER
BEFORE RETURNING
AVOID PROCESSING DELAYS). The assigned AR number should
breakage in
3.
our customers the
reproduced, translated, or
prior written consent of OMEGA ENGINEERING,
INC.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of
months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month one (1) year product warranty to cover handling and shipping time. This
maximum coverage on each product. If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service
will issue an Authorized Return (AR) number immediately upon phone or written request. OMEGA, if the unit is found to be defective, it will be repaired or replaced at no not apply to defects resulting from any action of the purchaser, interfacing, operation outside of design limits, improper repair VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
NON-WARRANTY
OMEGA for current repair charges. Have the
REPAIRS, consult
FOR
following information available BEFORE
1. Purchase Order number under which the product was PURCHASED,
2.
Model and serial number of the product under warranty, and
Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords latest in technology and engineering.
OMEGA is a trademark of OMEGA ENGINEERING, INC. © Copyright 2019 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reduced to any electronic medium or machine-readable form, in whole or in part, without the
contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.
Page 10
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
TEMPERATURE
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
Wire: Thermocouple, RTD & Thermistor
 
Calibrators & Ice Point References Recorders, Controllers & Process Monitors
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
 Transducers & Strain Gages  Load Cells & Pressure Gages
Displacement Transducers
 
Instrumentation & Accessories
FLOW/LEVEL
Rotameters, Gas Mass Flowmeters & Flow Computers
Air Velocity Indicators
 
Turbine/Paddlewheel Systems Totalizers & Batch Controllers
pH/CONDUCTIVITY
pH Electrodes, Testers & Accessories
Benchtop/Laboratory Meters
Controllers, Calibrators, Simulators & Pumps
Industrial pH & Conductivity Equipment
DATA ACQUISITION
Communications-Based Acquisition Systems
 
Data Logging Systems Wireless Sensors, Transmitters, & Receivers
Signal Conditioners
Data Acquisition Software
HEATERS
Heating Cable
Cartridge & Strip Heaters
Immersion & Band Heaters
Flexible Heaters
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
Metering & Control Instrumentation
Refractometers
Pumps & Tubing
Air, Soil & Water Monitors
Industrial Water & Wastewater Treatment
pH, Conductivity & Dissolved Oxygen Instruments
UNRESTRICTED UNRESTRICTED
Loading...