The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves
the right to alter specifications without notice.
This device is marked with international safety and hazard symbols in
accordance with IEC 1010. It is important to read and follow all precautions and
instructions in this manual before operating or commissioning this device as it
contains important information relating to safety and EMC. Failure to follow all
safety precautions may result in injury and or damage to your calibrator.
IEC symbols Description
OS100E Series
Caution, refer to accompanying
documents
Direct Current
Caution and Safety Information
• If the equipment is used in a manner not specified in this manual, the
protection provided by the equipment may be impaired.
• The installation category is one (1).
• There are no user replaceable fuses in this product
• The output terminals of this product are for use with equipment (digital
meters, chart recorders, etc,) which have no accessible five parts. Such
equipment should comply with all the applicable safety requirements.
• Do not operate the equipment in flammable or explosive environments.
Laser Symbol
Frame or Chassis
• All connections to the thermometer should be made via a shielded
cable, 24 AWG stranded wire with the following ratings: 300V, 105°C
(221°F), PVC insulation.
• Power must be disconnected before making any electrical connections.
• The power supply used to power the thermometer should be VDE or
UL approved with the following ratings: 12 to 24 Vdc @150 mA with
overload protection of 500 mA.
iii
Page 6
NOTES:
OS100E Series
iv
Page 7
Section 1 - Introduction
The low cost OS101E mini-infrared transmitter provides non-contact
temperature measurement for industrial applications. The unit measures
a temperature range of -18 to 538°C (0-1000°F) and provides a linear
analog output of either 4-20 mA, 0-5 Vdc, K type TC, 1 mV/°C, or 1
mV/°F.
The new OS102E mini-infrared transmitter has all the functions of OS101E
plus a built-in LED display that shows the measured temperature in
degrees F or degrees C which is switchable in the field.
The miniature sensor head design 2.5 cm dia. x 6.3 cm Length (1" x 2.5")
is ideal for measuring temperature in confined, and hard to reach places.
The aluminum sensor head as well as the rugged electronic housing (Die
cast Aluminum) are NEMA-4 rated.
The sensor head is connected to the electronic housing via a 1.82 m (6
feet) shielded cable as standard. The unit provides field adjustable alarm
output.
OS100E Series - Introduction
1
Section 2 - Installation
2.1 Unpacking
Remove the packing list and verify that you have received all your
equipment. If you have any questions about the shipment, please call
Customer Service at:
1-800-622-2378 or 203-359-1660. We can also be reached on the internet:
omega.com e-mail: info@omega.com
When you receive the shipment, inspect the container and equipment
for any signs of damage. Note any evidence of rough handling in transit.
Immediately report any damage to the shipping agent.
The carrier will not honor any damage claims unless all the shipping materials are saved for
inspection. After examination and removing contents, save packing material and carton in the
event reshipment is necessary.
The following items are supplied in the box:
• The infrared transmitter including the sensor head and the 1.82 m
(6 feet) shielded cable
NOTE:
• User’s Manual
• Mounting Nut
1-1
Page 8
2
OS100E Series - Installation
The following describes the ordering information:
OS102E or OS101E - MA - **, where
MA - 4/20 mA output
V1 - 0 to 5 Vdc output
K - Thermocouple output, K type
MV - Millivolt output
HT- High
temperature
sensor head
The following optional accessories are available:
Model No. Description
OS100-MBMounting Bracket
OS100-DRDIN Rail Mounting adapter
OS100-WCWater Cooling jacket
OS100-APAir purge collar
OS100-LSLaser Sighting
OS100-CA15FTSensor Head Extension Cable (4.6 m, 15')
OS100-CA25FTSensor Head Extension Cable (7.6 m, 25')
TX8-1008 Conductor stranded Shielded cable (30 m, 100')
PSU-93Unregulated 16-24 Vdc Power Supply
CAL3-IRNIST Traceable Calibration
Here are the Features of OS101 and OS102 infrared transmitters:
Features 101E 102E
Accuracy2% Rdg or 4°F (2.2°C)
Temp Range0-1000°F (-18 to 538°C)
Emissivity0.1 to 1.00 Adj.
Field of View6 to 1
Alarm Output Standard
1mV/Deg. Output Standard
4-20 mA Output Standard
0/5 Vdc Output Standard
K Type T/C Standard
RS232 PC Interface Standard
LED Display –– Standard
Main Housing 65.5 x 30.5 x 115.3 mm
(2.58 x 1.2 x 4.54")
2-1
2.2 Electrical Connection
Sensor Head Cable - The Sensor head is pre-wired to a 1.8 m (6 feet)
shielded cable. Plug & lock-in the male connector to the mating female
connector on the aluminum housing.
Power & Output Connection - Open the cover of the main aluminum
housing. Slide the cable through the strain relief and connect the wires to
the terminal block on the board as shown in Fig. 2-1. For Alarm output
connection, refer to Fig. 2-2.
Page 9
OS100E Series - Operation
3
1
2
3
4
5
6
7
8
9
POWER
SUPPLY
12-24 VDC
VmVmA
+
–
ANALOG OUTPUT
CHASSIS
Ty pe-K4-20 mAmV/Dgree0-5/0-10 Vdc
+
T/C
–
Figure 2-1. Power Supply & Analog Output Connections
1
2
POWER
SUPPLY
5-24 VDC
+
--
ALARM
RELAY
*IN4004 DIODE
OR
EQUIVOLENT
3
4
5
6
7
8
9
Section 3 - Operation
3-1 Main Board
The Main Board is shown in Fig. 3-1. Here are the important components
on the board:
(1) - Terminal Block for Power & Output connections
(2) - Single Turn Potentiometer to adjust Emissivity in tenths (0.x_)
(3) - Single Turn Potentiometer to adjust Emissivity in hundreds (0._x)
(4) - Slide switch to select between real time (Normal Operation) and alarm
set point
(5) - Alarm set point adjust, P3
(6) - °F to °C Jumper setting
Figure 2-2. Alarm Output Connection
3-1
Page 10
3
OS100E Series - Operation
1
3.2 Ambient Temperature
The Sensing head can operate in an ambient temperature of 0 to 70°C
(32 to 158°F). The Sensing head in the high temperature model (-HT) can
operate in an ambient temperature of 0 to 85°C (32 to 185°F) without any
cooling required. The Sensing head can operate up to 200°C (392°F) using
the water cool jacket accessory OS100-WC (See Fig. 3-6).
2
3
4
1
DISPLAY BOARD
INSTALLED ON
MODEL OS102E ONLY
95
6
78
DS2 DS3 DS4DS1
888
8
5
4
6
3
C
J3
2
P3
F
6
5
8
9
0
4
2
Figure 3-1. Main PC Board
J4
5
4
6
3
77
2
8
11
9
0
3
There is a warm up period of 3 minutes after power up. After the warm
up period, temperature measurement can be made.
When the ambient temperature around the sensor head changes abruptly,
the sensor head goes through thermal shock. It takes a certain amount of
time for the sensor head to stabilize to the new ambient temperature. For
example, it takes about 30 minutes for the sensor head to stabilize going
from 25°C to 50°C (77 to 122°F) ambient temperature.
The sensor head dimensions are shown in Fig. 3-2.
63.5
(2.5)
20
(0.8)
25
(1.0)
Dimensions mm (in.)
3/4-16 UNF
3-2
Figure 3-2. Sensor Housing
Page 11
3.3 Atmospheric Quality
Environments with smoke, dust, and fumes dirty up the optical lens, and
cause erroneous temperature readings. To keep the surface of the optical
lens clean, the air purge collar accessory is recommended, OS100-AP, See
Fig. 3-7.
3.4 Measuring Temperature
Before starting to measure temperature, make sure that the following
check list is met:
U The power and analog output connections are made (Fig. 2-1).
U The sensor head is connected to the main unit.
U The slide switch (SW1) on the main board is set to real time (Fig. 3-1).
U
The target is larger than the optical field of view of the sensor head
(Fig. 3- 3 ).
U The emissivity adjustment on the main board is set properly (Fig. 3-1).
OS100E Series - Operation
3
U The output load is within the product specification.
On 102E transmitters, follow these additional steps:
U The temperature display is set to °F or °C (Fig. 3-1)
U
For 4-20mA output models, make sure an output load is added, ie. 250
ohms.
DISTANCE: SENSOR TO OBJECT (IN)
0"
SPOT DIA.* (IN)
0.4"@0"
1cm@0
SPOT DIA.* (CM)
0
1.0"
3.3
20
12"6"
2.0"
6.6
40
24"
4.0"
D:S=6:1
10.0
60
13.0
80
36"
6.0"
16.0
100
48"
8.0"
20.0
122
DISTANCE: SENSOR TO OBJECT (CM)
* SPOT DIAMETER MEASURED AT 90% ENERGY
Figure 3-3. Optical Field Of View
3-3
Page 12
3
OS100E Series - Operation
COVER
2
3
4
1
NOTE: OPEN THE
COVER OF THE
MAIN HOUSING & SET THE JUMPER SW1 TO F OR C.
95
6
78
DS2DS3 DS4DS1
888
8
4
J3
3
C
2
P3
F
J4
5
5
4
6
6
3
77
828
11
0990
Figure 3-4. Setting the Temperature Engineering Unit
3.5 Alarm Setting
The unit provides 0-100% alarm set point adjustment. Here is an example
of an alarm setting.
• An OS101E-MA (4/20 mA output), the alarm is to be set at 400°F
temperature.
• Connect the alarm output as shown in Fig. 2-2.
• Set the slide switch (SW1) on the main board to the Alarm position.
• Measure the analog output, and set the Potentiometer P3 until the
output reads 10.4 mA which is 40% (400°F) of the temperature range.
40 x (20-4)
[10.4mA =
100
+ 4]
• Set the slide switch (SW1) back to the Real Time position.
• If the temperature reading is below the alarm set point, the alarm
output stays high, otherwise it goes low.
On the OS102E, you can set the alarm set point directly based on the
temperature display.
3.6 Adding Extension Cable
You can add extension cable between the Sensor Head and the main
electronic housing up to 15.2 m (50 feet). After adding the extension cable,
follow the steps in section to adjust the cable offset.
The following figures show the mounting bracket (OS100-MB), Water
cooling jacket (OS100-WC), Air purge collar (OS100-AP), DIN Rail
Mounting adapter (OS-100-DR), and the main aluminum enclosure. The
DIN Rail Mounting adapter (OS100-DR) is mounted to the bottom of the
main aluminum enclosure using two 4-40 screws.
A typical water cool jacket assembly is shown in Fig. 3-7, on the following page.
1. Mounting Nut
2. Mounting Bracket
3. Water Cool Jacket
4. Sensor Head
3-4
Page 13
3.7 RS-232
OS100E Series - Operation
3
1
2
3
4
5
6
7
8
9
Tx
Rx
1
PC SERIAL PORT
GND
5
Rx
Tx
Figure 3-5. RS232 PC Interface Connection
OS101E/OS102E RS232 Command Reference:
Note: All commands sent to the OS100E series must be terminated by
a carriage return or carriage return and linefeed, and is shown in this
document as “CR” and “LF”. Any parameters shown in square brackets
“[ ]” are optional. Parameters are separated from commands and from
each other by a space “ “. Spaces within commands are shown in this
document as “SP”. All commands are case insensitive, meaning any
combination of upper/lower case characters are accepted.
In order to communicate with the OS101E/OS102E, the terminal (or
PC running a terminal program such as HyperTerminal) must have the
following serial communication settings:
Description: - Transmit Target temperature in Deg C (one time).
Syntax: C CR [LF]
Parts: None. There are no parameters for this command.
Remarks: None.
Example:
Assuming target temperature is 125 °C, the following is shown on the
PC (terminal) screen after typing “C” and “Enter”:
C
125
3-5
Page 14
3
OS100E Series - Operation
F Command:
Description: - Transmit Target temperature in Deg F (one time).
Syntax:
F CR [LF]
Parts: None. There are no parameters for this command.
Remarks: None.
Example:
Assuming target temperature is 257 °F, the following is shown on the
PC (terminal) screen after typing “F” and “Enter”:
F
257
ALARM Command:
Description: Transmit alarm setpoint.
C
Syntax: ALARM
L
[
]
R
F
Parts: None. There are no parameters for this command.
Remarks:
The purpose of this command is to transmit the current alarm setpoint
only. The only method of changing setpoint is by opening the OS101E/
OS102E case, and adjusting the alarm setpoint potentiometer.
Example:
The following is shown on the PC (terminal) screen after typing
“ALARM” and “Enter”:
ALARM
537 C
999 F
>
3-6
Page 15
OS100E Series - Operation
VERSION Command:
Description: - Transmit model number, firmware version, and output
mode.
3
Syntax:
Parts:
VERSION CR [LF]
None. There are no parameters for this command.
Remarks:
Units with a display attached will transmit “OS102E” as the model
number. The example below shows a unit without a display attached.
The firmware version is a six-digit number following the “FW:” shown
below. Depending on the model suffix (the first characters immediately
following OS101-/OS102-), the “OUTPUT:” will be transmitted as
follows:
Model Suffix OUTPUT:
K TC-K
MV-C mV/Deg C
MV-F mV/Deg F
V1 0-5V
V2 0-10V
MA 4-20mA
Example:
The following is shown on the PC (terminal) screen after typing
“VERSION” and “Enter”:
>VERSION
OMEGA ENGINEERING
OS101E FW: 061013
OUTPUT: mV/Deg F
>
Specifies whether the alarm will activate at the setpoint with rising
temperature, or falling temperature. The valid range for this parameter is
0 to 1. A value of 1 will cause the alarm to activate on rising temperature,
and 0 will cause activation on falling temperature. If this parameter is not
specified, the mode currently set will be transmitted.
Remarks:
The default mode for alarm activation is “rising”. The only way to change
the activation mode is through the RS232 “AMODE” command.
3-7
Page 16
3
OS100E Series - Operation
Example 1:
The following is shown on the PC (terminal) screen after typing
“AMODE” and “Enter”. The activation mode is not changed, but it is
reported.
AMODE
ALARM MODE SET FOR
RISING TEMP.
>
Example 2:
The following is shown on the PC (terminal) screen after typing
“AMODE 0” and “Enter”. The activation mode is changed to “falling”
and reported as such.
AMODE 0
ALARM MODE SET FOR
FALLING TEMP.
>
Example 3:
The following is shown on the PC (terminal) screen after typing
“AMODE 1” and “Enter”. The activation mode is changed to “rising”
and reported as such.
AMODE 1
ALARM MODE SET FOR
RISING TEMP.
>
EMS Command:
Description: - Transmit emissivity setting.
Syntax: EMS CR [LF]
Parts: None. There are no parameters for this command.
Remarks:
The purpose of this command is to transmit the current emissivity setting
only. The only method of changing the emissivity setting is by opening
the OS101E/OS102E case, and adjusting the emissivity potentiometers.
Example:
3-8
The following is shown on the PC (terminal) screen after typing “EMS”
Specifies the total length (in feet) of cable attached between the sensor
head and the transmitter box. The valid range for this parameter is 6
to 56. If this parameter is not specified, the length currently set will be
transmitted.
Remarks:
The default length is 6 Feet. The only way to change the cable length
compensation is through the RS232 “CABLE” command. Additional
extension cables can be purchased and added to the system, up to a
maximum total length of 56 Feet. As total cable length is changed, and
error is introduced. This command is used to compensate for that error.
When adding extension cable(s), ensure you specify the total length (i.e.,
25 Ft extension + original 6 Ft = 31 Ft) with the CABLE command.
Example 1:
The following is shown on the PC (terminal) screen after typing
“CABLE” and “Enter”. The current cable length compensation value is
not changed, but it is reported.
CABLE
CABLE LENGTH = 6 FT
>
Example 2:
The following is shown on the PC (terminal) screen after typing
“CABLE 31” and “Enter”. The cable length compensation value is
changed to “31” and reported as such.
CABLE 31
CABLE LENGTH = 31 FT
>
Example 3:
The following is shown on the PC (terminal) screen after typing
“CABLE 200” and “Enter”. The cable length compensation value is
changed to “56” because the value 200 is above the maximum length of
56.
CABLE 200
CABLE LENGTH = 56 FT
>
3-9
Page 18
3
OS100E Series - Operation
25.4
(1)
12.7
(0.5)
90
(1/8)
23.3
(.919)
31.8
(1.25)
42.9
(1.69)
(1/8)
3.2
3.2
Ø 4
(0.156)
50.0
(1.97)
Dimensions mm (in.)
Ø = Diameter
Figure 3-6. Mounting Bracket OS100-MB
1/8 NPT COMPRESSION
FITTING , 2 PLCS.
12.7
(0.5)
Ø 19
(0.755)
8
(.312)
Dimensions mm (in.)
41.3
(1.625)
INLET
3/4-16 UNF-2B3/4-16 UNF
60.3
(2.375)
OUTLET
Figure 3-7. Water Cooling Jacket, OS100-WC
4321
INLET
UP TO 200C (392F) AMBIENT TEMPERATURE
TO MAIN HOUSING
OUTLET
3-10
Figure 3-8. Typical Water Cool Jacket Assembly
Page 19
OS100E Series - Operation
17.5
(0.69)
28.6
(1.125)
1/16 NPT
3/4-16 UNF 2B
THRU.
Dimensions mm (in.)
COMPRESSION
FITTING
Figure 3-9. Air Purge Collar, OS100-AP
4-40 THRD. HOLE, 2 PLCS.
44.5
(1.75)
109.5
(4.3125)
3
DIN RAIL
Dimensions mm (in.)
SNAP FEET
Figure 3-10. DIN Rail Mounting Adapter, OS-100-DR
65.5
(2.578)
53.3
(2.10)
47.7
(1.880)
35.5
(1.400)
5.7 (0.225) DIA.
(4) HOLES
78.0
103.4
(3.07)
(4.07)
115.4
(4.545)
5.0
(0.20)
MAIN HOUSING
Dimensions mm (in.)
Figure 3-11. NEMA-4 Aluminum Enclosure
8.6 (0.340) DIA. x
22.6 (0.890) DEEP
(2) PLACES
COVER
26.4
(1.04)
3-11
Page 20
3
OS100E Series - Operation
NOTES:
3-12
Page 21
Section 4 – Software
4.1 Software Overview
The OS100-Soft application enables you to communicate with your
OS100E Series transmitter on a PC using serial communication. The
application has the following capabilities:
• Real-time display of temperature reading.
• Data acquisition and plotting of real-time data.
• Saving acquired data in comma separated values (csv) or tab format for
easy importation into spreadsheet programs.
• Data analysis function.
• Printing of graph and data.
4.2 Hardware Hookup
Wire the OS100E unit to the PC as shown in section 3.7, and table 4-1.
OS100E Series - Software
4
OS100E, Terminal Block TB1 Computer DB-9 Male Connector
Pos. No. Description Pos. No. Description
5 Signal GND 5 Signal GND
6 Rx 3 Tx
7 Tx 2 Rx
Table 4-1 OS100E Serial Communications Wiring Table
4.3 Software Installation
4.3.1 System Requirements
Minimum System Requirements
The following are the minimum recommended system requirements:
• Windows 98/ME, Windows 2000, Windows XP, Windows Vista,
Operating System.
• Available serial port or USB to serial port adapter.
• 26 MB Ram
• 280 MB available hard drive space.
Recommended System Configuration
The following system configuration is recommended:
4-1
Page 22
4
OS100E Series - Software
• Windows XP or newer operating system.
• At least 512 MB Ram.
• 280 MB available hard drive space.
4.3.2 Software Installation
Insert the OS100-Soft CD into your computer’s CD-Rom drive. The
installation program should start up automatically. If it does not start
automatically locate and run the Setup.exe file on the CD.
Microsoft .NET framework 2.0 is a pre-requisite for this application and
will be automatically installed from the CD if missing this will require
you to accept the license agreement prior to proceeding.
4-2
Figure 4-1. NET Framework 2.0 Eula
Once the .NET framework is installed or if previously installed on your
PC the OS100-Soft installation process begins. The first screen displayed is
the welcome screen click “Next>” button to continue. This is followed by
the Installation Options Screen, check the box if you would like to install
a shortcut for the application on your desktop. Click “Next>” button to
continue.
The wizard then takes you to the Installation Folder Screen. Accept the
default or choose alternate path then click “Next>” button to continue.
Page 23
OS100E Series - Software
4
This is followed by the Confirm Installation Screen. If satisfied with click
“Next>” button to begin installation, otherwise use “<Back” button to
navigate to and change settings as needed.
Figure 4-2. Confirm Installation Screen
A license agreement for the OS100-Soft is next displayed, plead read and
select “I agree” option to proceed with installation.
4-3
Page 24
4
OS100E Series - Software
Figure 4-3. OS100-Soft License Agreement
The installation process then begins and the Installation Complete screen
is displayed when done. Click “Close” button to finish.
4-4
Page 25
OS100E Series - Software
4
Figure 4-4. Installation in Progress
Figure 4-5. Installation Complete
4-5
Page 26
4
OS100E Series - Software
Congratulations! You have successfully installed the OS100-Soft program
and you can now use it to communicate with your OS100E series unit(s).
4.4 Using Software
To run the application click on the desktop shortcut (if present) or access
the program from the Start Menu, Start\Programs\OS100-SOFT\OS100Soft.exe (or select path installed if different).
The software’s online help will guide you through its actual use.
4-6
Page 27
OS100E Series - Laser Sight Accessory
Section 5 - Laser Sight Accessory
5.1 Warning and Cautions
You may receive harmful laser radiation exposure if you do not adhere to the warnings listed
below:
• Use of controls or adjustments or performance of procedures other
than those specified here may result in hazardous radiation exposure.
• Do not look at the laser beam coming out of the lens or view directly
with optical instruments - eye damage can result.
• Use extreme caution when operation the laser sight accessory
• Never point the laser accessory at a person
• Keep out of the reach of all children
5
CAUTION:
WARNING:
Do not attempt to open the laser sight accessory. There are no user serviceable parts inside.
5.2 Operating the Laser Sight Accessory
The laser sight accessory screws onto the front of the sensor head. This
accessory is only used for alignment of the sensor head to the target area.
After the alignment process, the accessory has to be removed from the
front of the sensor head before temperature measurement.
The laser sight accessory is powered from a small compact battery pack
(included with the accessory). Connect the battery pack to the accessory
using the cable provided. Aim at the target, and turn on the battery power
using the slide switch on the battery pack. Adjust the sensor head position
so that the laser beam points to the center of the target area. Turn off the
battery pack, and remove the laser sighting accessory from the sensor
head. See Fig. 4-1 for reference.
5-1
Page 28
5
LED LASER
POWER
INDICATOR
LASER WARNING LABEL
SENSOR HEAD
LASER BEAM
DC POWER JACK
POWER CABLE
SWITCH
ON/OFF
BATTERY PACK
POWER SUPPLY
50.8
(2.0)
19
(.75)
38.1
(1.5 )
Dimensions mm (in.)
OS100E Series - Laser Sight Accessory
Figure 5-1. Laser Sighting Accessory, OS100-LS
LASER RADIATION DO NOT STARE
INTO BEAM OR VIEW DIRECTLY WITH
OPTICAL INSTRUMENTS. CLASS 2
LASER PRODUCT.
LASER RADIATION - DO NOT STARE INTO BEAM
OUTPUT <1 mW, WAVELENGTH 630-670 nm, CLASS II (2) LASER
PRODUCT. COMPLIES WITH FDA 21CFR 1040.10 & EN60825-1/11.2001
Figure 5-2. Laser Warning Label
®
CAUTION
AVOID EXPOSURE. LASER
RADIATION IS EMITTED
FROM THIS APERTURE.
5-2
Page 29
Section 6 - Specifications
6.1 General
Temperature Range: -18 to 538°C (0 to 1000°F)
Accuracy @ 22°C (72°F) ± 2% of Rdg. or 2.2°C (4°F) whichever is
ambient temperature &
emissivity of 0.95 or greater:
Optical Field of View: 6:1 (Distance/Spot Size)
Repeatability: ± 1% of Rdg.
Spectral Response: 5 to 14 microns
Response Time: 100 msec (0 to 63% of final value)
Emissivity Range: 0.1 to 1.00, adjustable
Operating Ambient Temperature
Main Transmitter:
Sensor Head:
Sensor Head (-HT Model):
Sensor Head with OS100-WC
(Water Cooling Jacket):
OS100E Series - Specifications
greater
0 to 50°C (32 to 122°F)
0 to 70°C (32 to 158°F)
0 to 85°C (32 to 185°F)
0 to 200°C (32 to 392°F)
6
Operating Relative Humidity: Less than 95% RH, non-condensing
Water Flow Rate for OS100-WC: 0.25 GPM, room temperature
Thermal Shock: About 30 minutes for 25°C
abrupt ambient temperature change
Warm Up Period: 3 minutes
Air Flow Rate for OS100-AP: 1 CFM (0.5 Liters/sec.)
Power: 12 to 24 Vdc @ 150 m
(0-10 Vdc output requires 14 Vdc)
Analog Outputs
MV:
K:
MA:
V1
V2:
Output Load requirements
Min. Load (0 to 5Vdc):
Max. Load (4 to 20 mA):
Transmitter Housing: NEMA-4 & IP65, Die Cast Aluminum
Sensor Head Housing: NEMA-4; Anodized Aluminum
1 mV/°F or 1 mV/°C
K Type TC
4 to 20 mA
: 0 to 5 Vdc
0-10 Vdc
1 K-Ohms
(Supply Power - 4 )/20 mA
Alarm Output: Open Drain, 100 mA
Alarm Set Point: 0 to 100%, Adjustable via P3
Alarm Deadband: 12°C (10°F)
Alarm Indication: Display Flashing
6-1
Page 30
6
OS100E Series - Specifications
6.1 General Con’t.
Dimensions
Sensor Head:
Main Housing:
25.4 OD. x 63.5 mm L (1" OD. x 2.5" L)
65.5 W x 30.5 H x 115.3 mm L
(2.58 W x 1.2 H x 4.54" L)
Weight:272 g (0.6 lb)
6.2 RS232
Baud: 19200
Data Bits:
Parity:
Stop Bits:
Handshaking:
8
None
1
None
6.3 Laser Sight Accessory (OS100-LS)
Wavelength (Color): 630 - 670 nm (Red)
Operating Distance (Laser Dot): Up to 9.1 m (30 ft.)
Max. Output Optical Power: Less than 1 mW at 22°F ambient
temperature.
European Classification: Class 2, EN60825-1/11.2001
Maximum Operating Current: 45 mA at 3 Vdc
FDA Classification: Complies with 21 CFR 1040.10,
Class II Laser Product
Beam Diameter: 5 mm
Beam Divergence: < 2 mrad
Operating Temperature: 0 to 50°C (32 to 122°F)
Operating Relative Humidity: Less than 95% RH, non-condensing
Power Switch: ON / OFF , Slide switch on the Battery
Pack
Power Indicator: Red LED
Power: Battery Pack, 3 Vdc (Consists of two 1.5
Vdc AA size Lithium Batteries)
Laser Warning Label:Located on the head sight circumference
6-2
Identification Label: Located on the head sight circumference
Dimensions: 38 DIA x 50.8 mm L (1.5" DIA x 2" L)
Page 31
Section 7 - Emissivity Table
Material Emissivity (ε)
Aluminum – pure highly polished plate ....................... 0.04 to 0.06
Aluminum – heavily oxidized
Aluminum – commercial sheet
Brass – dull plate
Brass – highly polished, 73.2% Cu, 26.7% Zn
Chromium – polished
Copper – polished
Copper – heated at 600°C (1112°F)
Gold – pure, highly polished or liquid
Iron and steel (excluding stainless) – polished iron
Iron and steel (excluding stainless) – polished cast iron
Iron and steel (excluding stainless) – polished wrought iron
Iron and steel (excluding stainless) – oxidized dull wrought iron
Iron and steel (excluding stainless) – rusted iron plate
Iron and steel (excluding stainless) – polished steel
Iron and steel (excluding stainless) – polished steel oxidized at
600°C (1112°F)
Iron and steel (excluding stainless) – rolled sheet steel
METALS
Iron and steel (excluding stainless) – rough steel plate
Lead – gray and oxidized
Mercury
Molybdenum filament
Nickel – polished
Nickel
Platinum – pure polished plate
Silver – pure and polished
Stainless steel – polished
Stainless steel – Type 301 at 232 to 942°C (450°F to 1725°F)
Tin – bright
Tungsten – filament
Zinc – polished commercial pure
Zinc – galvanized sheet
............................................. 0.09 to 0.12
– oxidized at 649 to 1254°C (1200°F to 2290°F) . . . . . . . . . . 0.59 to 0.86
Paints, lacquers, and varnishes – Black matte shellac
Paints, lacquers, and varnishes – aluminum paints
Paints, lacquers, and varnishes – flat black lacquer
NONMETALS
Paints, lacquers, and varnishes – white enamel varnish
Porcelain – glazed
Quartz – opaque
Roofing Paper
Tape – Masking
Water
............................................... 0.95 to 0.96
............................................ 0.92
....................................... 0.68 to 0.92
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a
period of 25 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month
grace period to the normal two (2) year product warranty to cover handling and shipping time. This
ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no
charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser,
including but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for
any damages that result from the use of its products in accordance with information provided
by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the
company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of
OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the
component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED
RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return
package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR WARRANTY RETURNS, please have the
following information available BEFORE
contacting OMEGA:
1. Purchase Order number under which the product
was PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords
our customers the latest in technology and engineering.
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the
prior written consent of OMEGA ENGINEERING, INC.
FOR NON-WARRANTY REPAIRS,
for current repair charges. Have the following
information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST
of the repair,