Omega OM-DAQPRO-5300 User guide

Page 1
Page 2
Page 3
Page 4
Page 5
Contents
Introduction ...................................................................................................................................7
Chapter 1 OM-DAQPRO-5300 .........................................................................................................9
1.1. General....................................................................................................................... 9
1.1.1. OM-DAQPRO-5300: system contents............................................................................. 9
1.1.2. External connections..................................................................................................... 10
1.1.3. Sensor types and connections ......................................................................................11
1.1.4. User defined sensors ....................................................................................................13
1.1.5. Alarms...........................................................................................................................13
1.1.6. Alarm output.................................................................................................................. 14
1.1.7. Sensor calibration .........................................................................................................15
1.1.8. Serial number and comment .........................................................................................15
1.1.9. Battery...........................................................................................................................15
1.1.10. Mains Adaptor ............................................................................................................... 15
1.2. Stand-Alone Operation............................................................................................. 16
1.2.1. Front Panel Layout........................................................................................................16
1.2.2. Quick-Start .................................................................................................................... 17
1.2.3. Working with the OM-DAQPRO-5300 Menus ...............................................................21
1. Main menu display............................................................................................. 21
2. Menu buttons..................................................................................................... 22
3. Menu Icons and commands...............................................................................22
1.2.4. Graphic Display.............................................................................................................24
2. The Cursor......................................................................................................... 25
3. Zooming.............................................................................................................25
1.2.5. Load the Last Setup ...................................................................................................... 25
1.2.6. Configure Your OM-DAQPRO-5300 .............................................................................26
1. Input Filter..........................................................................................................26
2. Averaging points................................................................................................ 27
3. Temperature units .............................................................................................27
4. Clear memory .................................................................................................... 27
5. Screen contrast .................................................................................................27
1.2.7. Internal Clock and Calendar.......................................................................................... 27
1.2.8. Clear the Memory.......................................................................................................... 27
1.2.9. Choose the Right Setup. ...............................................................................................28
1.2.10. Programming Rules and Limitations .............................................................................30
Page 6
Chapter 2
2.1. Install the Software ................................................................................................... 31
2.1.1. System Requirements................................................................................................... 31
2.1.2. Installation.....................................................................................................................31
2.2. Overview................................................................................................................... 33
2.2.1. DaqLab On-screen Layout ............................................................................................ 33
2.2.2. Working with Projects.................................................................................................... 33
2.2.3. DaqLab Window Layout ................................................................................................34
2.3. Getting Started ......................................................................................................... 35
2.3.1. Set up a Recording Session.......................................................................................... 35
1. Prepare OM-DAQPRO-5300 .............................................................................35
2. Setup the OM-DAQPRO-5300........................................................................... 35
3. Start Recording.................................................................................................. 35
2.3.2. Data recording options .................................................................................................. 36
1. Single measurement.......................................................................................... 36
2. Replace .............................................................................................................36
3. Add .................................................................................................................... 36
2.3.3. Download Data.............................................................................................................. 36
2.3.4. Save Data ..................................................................................................................... 37
2.3.5. Open a File ...................................................................................................................37
2.3.6. Create a New Project .................................................................................................... 38
2.3.7. Import data....................................................................................................................38
2.3.8. Date format settings ...................................................................................................... 39
2.3.9. Print ..............................................................................................................................40
1. Print a graph ...................................................................................................... 40
2. Print a table .......................................................................................................41
2.4. View the Data ........................................................................................................... 42
2.4.1. Display Options ............................................................................................................. 42
2.4.2. Graph Display ...............................................................................................................42
1. Split graph view ................................................................................................. 42
2. The Cursor......................................................................................................... 43
3. Zooming.............................................................................................................43
4. Panning .............................................................................................................45
5. Edit the Graph ...................................................................................................45
6. Display alarm levels........................................................................................... 46
7. Format the graph ............................................................................................... 46
8. Change the graph’s units and its number format ............................................... 47
9. Add a graph to the project .................................................................................47
Page 7
The Table Display .........................................................................................................48
2.4.3.
Formatting the table............................................................................................... 49
2.4.4. Meters ........................................................................................................................... 49
2.4.5. Data Map ......................................................................................................................50
1. Control the display with the Data Map ...............................................................50
2. Understanding Data Map icons .........................................................................51
2.4.6. Export Data to Excel .....................................................................................................52
Export file settings ................................................................................................. 52
2.4.7. Copy the Graph as a Picture......................................................................................... 52
2.5. Program OM-DAQPRO-5300................................................................................... 53
2.5.1. Setup.............................................................................................................................53
1. Quick setup........................................................................................................53
2. Alarm setup .......................................................................................................55
3. Low pass filter.................................................................................................... 56
4. Averaging ..........................................................................................................57
5. Battery level....................................................................................................... 57
6. Presetting the display ........................................................................................57
7. Preset the graph’s X-axis................................................................................... 58
8. Triggering ..........................................................................................................58
2.5.2. Start Recording .............................................................................................................60
2.5.3. Stop Recording .............................................................................................................60
2.5.4. Clear OM-DAQPRO-5300’s Memory ............................................................................60
2.5.5. Adding comment to OM-DAQPRO-5300....................................................................... 60
2.5.6. Edit OM-DAQPRO-5300’s Notes ..................................................................................61
2.5.7. Temperature units ......................................................................................................... 61
2.5.8. Calibrating the sensors .................................................................................................62
1. Calibration options............................................................................................. 62
2. Calibration password .........................................................................................63
3. Calibration procedure ........................................................................................63
2.5.9. Define a Custom Sensor ...............................................................................................69
2.5.10. Communication Setup................................................................................................... 70
2.6. Analyze the data....................................................................................................... 72
2.6.1. Reading Data Point Coordinates................................................................................... 72
2.6.2. Reading the Difference Between two Coordinate Values..............................................72
2.6.3. Working with the Analysis Tools.................................................................................... 72
2.6.4. Smoothing..................................................................................................................... 72
2.6.5. Statistics........................................................................................................................ 73
2.6.6. Most Common Analysis Functions ................................................................................ 73
1. Linear fit............................................................................................................. 73
Page 8
2. Derivative...........................................................................................................74
3. Integral ..............................................................................................................74
2.6.7. The Analysis Wizard .....................................................................................................74
1. Using the Analysis Wizard................................................................................. 74
2. Curve fit ............................................................................................................. 75
3. Averaging ..........................................................................................................76
4. Functions........................................................................................................... 77
2.6.8. Available Analysis Tools................................................................................................78
1. Curve fit ............................................................................................................. 78
2. Averaging ..........................................................................................................78
3. Functions........................................................................................................... 79
2.7. Special Tools ............................................................................................................ 83
2.7.1. Crop Tool ......................................................................................................................83
1. To trim all data up to a point .............................................................................. 83
2. To trim all data outside a selected range ........................................................... 83
2.8. Toolbar Buttons ........................................................................................................ 84
2.8.1. Main (upper) Toolbar..................................................................................................... 84
2.8.2. Graph Toolbar ............................................................................................................... 85
Chapter 3 Troubleshooting Guide...................................................................................................87
Chapter 4 Specifications.................................................................................................................89
Appendix: Figures .................................................................................................................................92
Appendix: Simplified measurement circuits ............................................................................................. 93
Index .................................................................................................................................95
Page 9
Introduction
OM-DAQPRO-5300 is an eight-channel portable data acquisition and logging system with graphic display and built-in analysis functions
OM-DAQPRO-5300 is battery operated and is capable of sampling, processing and displaying measurements without connecting to a computer. OM-DAQPRO-5300 is a professional, cost-effective, compact and stand-alone data logging system that can be used with a wide variety of applications. This 16-bit, high-resolution, eight-channel data logger offers the pros graphic displays and analysis functions for measuring voltage, current and temperature in real-time.
With its high resolution and fast Analog to Digital converter (ADC), OM-DAQPRO-5300 meets the majority of data logging requirements in most industrial applications. Its unique ability to display measured values and analyze them in real-time on a graphical interface minimizes the need to download collected data to a computer for further analysis.
Every OM-DAQPRO-5300 unit is embedded with a unique serial number and can be loaded with a descriptive comment for safe identification.
OM-DAQPRO-5300 5300 includes eight input channels for measuring voltage, current, temperature and pulses. Selectable ranges for each input are 0-24mA, 0-50mV, 0-10V, a large variety of NTC thermistors, PT-100 RTDs and thermocouple temperature sensors including internal temperature, pulse counter, frequency meter and up to 20 user defined sensors. The inputs use pluggable screw terminal blocks for easy connection.
An internal clock and calendar keep tracks of the time and date of every sample measured.
OM-DAQPRO-5300 can automatically activate external alarm events when data is outside a specified range.
OM-DAQPRO-5300 is very easy to use because all its functions are broken down into an 8-icon menu, its four buttons can browse every menu and execute any of the commands.
A rechargeable battery powers the data logger, which shuts off automatically after 15 minutes have passed since the time of the last data recording, the time the last button was pressed, or the time the last communication was made with the PC.
The OM-DAQPRO-5300 system also comes with the powerful DaqLab software. When the OM­DAQPRO-5300 is connected to a PC, live displays can be viewed at rates of up to 100/s, and automatic downloads can be carried out at higher rates. The WINDOWS based software can display the data in graphs, tables or meters, can analyze data with various mathematical tools, or export data to a spreadsheet.
DaqLab also enables you to setup OM-DAQPRO-5300 and to send advanced commands such as alarm settings, triggering conditions and text notes.
Introduction 7
Page 10
This manual is divided into three sections:
• The first section is dedicated to the data logger itself. Topics include: Connecting
sensors, configuration through the data logger keypad, and using the LCD-graphic display to take measurements when working offline.
• The second section gives a comprehensive overview of the DaqLab software. Topics
include: How to download data from the data logger to a PC, analyzing the data both graphically and mathematically and using the DaqLab software to program the data logger when working online.
• The third and last section contains hardware specifications and a comprehensive
troubleshooting guide that gives answers to common questions.
8 Introduction
Page 11
Chapter 1
OM-DAQPRO-5300
This section will focus on the OM-DAQPRO-5300’s data collection device – the data logger; and includes:
• How to operate the OM-DAQPRO-5300’s keypad,
• How to set up OM-DAQPRO-5300
• How to connect sensors to OM-DAQPRO-5300
• How to connect OM-DAQPRO-5300 to your PC.
• How to conduct a logging session
1.1. General
1.1.1. OM-DAQPRO-5300: system contents
1. The OM-DAQPRO-5300 Data Logger
2. USB communication cable
3. An AC-DC adaptor
4. DaqLab software installation CD
5. User Guide
6. Carrying case
Chapter 1 OM-DAQPRO-5300 9
Page 12
p
1.1.2. External connections
1. Sensor
Inputs/alarm output
2. PC USB
connection socket
8th
input/output
1st
ut
in
3. Power input
(DC 9 – 12V)
Figure 1: OM-DAQPRO-5300 external connections
10 Chapter 1 OM-DAQPRO-5300
Page 13
1. Sensor inputs/alarm output – pluggable screw terminal block (marked Input -1 to Input – 8 from right to left). All eight inputs can be used simultaneously.
If you are using one sensor, connect it to input 1. If you are using two sensors connect them to inputs 1 and 2, and so on…
I/O–8 (Input/Output–8) serves either as an input or as alarm output.
To connect a sensor to the OM-DAQPRO-5300 unplug the screw terminal, connect the sensor’s wires to the terminals, and then plug the terminal back to the corresponding socket on the input block.
2. Computer USB communication socket – use this socket to connect OM-DAQPRO-5300 to a computer
Connect the USB Type B plug (square plug) of the supplied communication cable to the OM­DAQPRO-5300 and the USB Type A plug (flat plug) to the computer (see page 32 for USB driver installation).
3. External DC power supply socket – Plug in an AC/DC 9 - 12V adaptor whenever you want to save battery power, or to charge the battery when necessary. Connecting external power to the OM-DAQPRO-5300 automatically charges the internal battery. The adaptor should meet the required specifications (see section 1.1.10 on page 15).
1.1.3. Sensor types and connections
Each of the 8 input channels of OM-DAQPRO-5300 is multipurpose and can be individually configured to any of the following types and ranges:
Sensor Type Range
Current 0 – 24mA
Frequency (input 1 only) 20 to 4000Hz
Pulse counter (input 1 only) 0 to 65,000 pulses
Temperature Internal
Temperature
Temperature
Temperature PT-100 2 wire
Temperature PT-100 3 wire
Temperature Thermocouple J
Temperature Thermocouple K
Temperature Thermocouple T
Voltage 0 – 10V
Voltage 0 - 50mV
Chapter 1 OM-DAQPRO-5300 11
NTC 10KΩ -25 to 150°C
NTC 2252Ω -25 to 150°C
-25 to 70°C
-200 to 400°C
-200 to 400°C
-200 to 1200°C
-250 to 1200°C
-200 to 400°C
Page 14
User defined Up to 20 types 0 – 10V or 0 – 24mA
Connect the sensor to the terminal block at the top of OM-DAQPRO-5300:
Figure 2: OM-DAQPRO-5300’s inputs block terminal
Sensors must be added successively, starting with input–1. If a single sensor is used it must be connected to Input–1. If two sensors are used, they must be connected to Input-1 and Input-2 and so on…
Alarm output
I/O–8 (Input/Output–8) serves either as an input or as alarm output.
Polarity
Current, voltage, thermocouples and user defined sensors have distinct polarity. Be careful to connect them in the right polarity.
Frequency/Pulse counter
Connect the signal wires to I/O–8 screw terminals, and select Frequency or Pulse counter for input 1 from the Setup menu. Inputs 2 to 7 are still available for other sensors.
The Frequency/pulse counter is optically isolated from the internal circuitry and can simultaneously measure a signal source, together with another input.
3 wire PT 100
You have to use two inputs to connect a 3 wire PT 100 You can connect one 3 wire PT 100 to input–1 and input–5, and/or inputs 2 and 6, and/or inputs 3 and 7 and/or inputs 4 and 8.
Connect the single wire to the plus (+) terminal of the first input and the common end wires to the minus (-) terminals of both the inputs. See Figure 3 for wiring configuration of a 3 wire PT 100 connected to input–1 and input–5:
12 Chapter 1 OM-DAQPRO-5300
Page 15
Figure 3: connecting 3 wire PT 100
When working with a 3 wire PT 100 inputs 5 to 8 are not available and you can connect up to 4 sensors.
Sensor mismatch
After logging begins OM-DAQPRO-5300 checks if the sensor type assigned to each channel matches the attached sensor. If there is a mismatch, OM-DAQPRO-5300 disconnects the corresponding input and displays a warning message: “Illegal sensor” (see ‘Troubleshooting Guide’ on page 86 for details)
1.1.4. User defined sensors
OM-DAQPRO-5300 provides a simple and straightforward tool for defining up to 20 custom sensors. Almost any 0 – 10V and 0 – 24mA sensor or transducer is accepted by MultiLogPRO and its electrical units are automatically scaled to meaningful user-defined engineering units.
The sensor definitions are stored in OM-DAQPRO-5300’s memory and are added to the sensors list. The sensor’s readings are displayed in the user defined units both on OM-DAQPRO-5300’s LCD screen and DaqLab software.
To learn how to define custom sensors see section 2.5.9 on page 68
For further manipulating sensors readings use the Analysis wizard (see page 73)
1.1.5. Alarms
Users can define minimum and maximum alarm levels for each input individually.
Chapter 1 OM-DAQPRO-5300 13
Page 16
OM-DAQPRO-5300 places a small alarm icon next to the corresponding input readings and can switch alarm output if either level is breached.
To display alarm warnings in real-time OM-DAQPRO-5300 must be in numeric display mode (see page
20).
To learn how to enter alarm levels and to activate alarm output, see section 2.5.1.2 on page 54.
1.1.6. Alarm output
OM-DAQPRO-5300 can trigger an external event (e.g. sound alarm, warning light or oven).
Connect the external current loop to the screw terminals of I/O–8. Be careful to connect the external power supply in the right polarity (see Figure 4 below).
Figure 4: connecting external alarm device
The alarm output is analogous to electrical switch. In OFF position the terminals of I/O–8 are disconnected. In ON position they are shortened.
If an alarm output is selected this input/output is set to OFF position. When any active alarm level is exceeded the output is set to ON. All active alarms must be false to reset the output to OFF position.
The maximum switch load is 50mA, 5V. The output is protected by 50mA reset-able fuse. For higher loads use a relay.
To learn how to enter alarm levels and to activate alarm output, see section 2.5.1.2 on page 54
14 Chapter 1 OM-DAQPRO-5300
Page 17
1.1.7. Sensor calibration
OM-DAQPRO-5300 ships fully calibrated. However, further calibration can be applied via DaqLab. The calibration parameters are sent to OM-DAQPRO-5300 and stored in its memory. See page 61 for calibration instructions.
1.1.8. Serial number and comment
Every OM-DAQPRO-5300 unit is embedded with a unique serial number and can be loaded with a descriptive comment to identify its task and location.
To add or edit the comment connect OM-DAQPRO-5300 to the computer and use DaqLab software (see page 59).
Every time data is transferred to the computer it is labeled both with OM-DAQPRO-5300’s serial number and comment and then displayed in the graph title.
The serial number is marked on the back of the product. To view its comment, select System information from OM-DAQPRO-5300’s main menu.
1.1.9. Battery
OM-DAQPRO-5300 is equipped with a 7.2V Ni-MH rechargeable battery. Before you first start working with OM-DAQPRO-5300, charge the unit for 10 to 12 hours while it is turned off. Battery life is approximately 40 hours between charges.
If the data logger’s main battery runs out, the internal 3V Lithium battery backs up the memory, so no data will be lost. The Lithium battery also keeps the internal clock and calendar running.
Note: Before storing the data logger make sure you have unplugged all the sensors and
pressed the OFF key.
1.1.10. Mains Adaptor
The Mains adaptor (AC/DC adaptor) converts mains power (from a wall outlet) to a voltage suitable to OM-DAQPRO-5300.
• Output: Capacitor filtered 9 to 12 VDC, 400mA.
• Female plug, center Negative.
Chapter 1 OM-DAQPRO-5300 15
Page 18
1.2. Stand-Alone Operation
One way to program the OM-DAQPRO-5300 is to use its keypad and screen (The other way is to use the DaqLab software – see page 52). The keypad allows us to set all the parameters for data collection, while the LCD screen displays the setting values.
1.2.1. Front Panel Layout
LCD Display
Backward Button
Enter / Run Button
On / Off
Buttons
Figure 5: OM-DAQPRO-5300 front panel
16 Chapter 1 OM-DAQPRO-5300
Forward Button
Escape / Stop Button
Page 19
1.2.2. Quick-Start
Before you first use OM-DAQPRO-5300, charge the unit for 10 to 12 hours while it is turned off.
1. Turn on OM-DAQPRO-5300
Press the on button. You will see the initialization screen. OM-DAQPRO-5300 performs a brief self­check and display its status including battery level and operating mode. At this point you can load the last setup you used (see page 25). If you need a new setup wait until you see the Main Menu screen:
fn
?
2. Connect the sensors
Start with the first input on the right.
Note: Sensors must be added successively, starting with input 1. If a single sensor is
used it must be connected to Input 1. If two sensors are used, they must be connected to Input 1 and Input 2.
See Sensors types and connections on page 11 for more details
3. Identify the Sensors
You must tell OM-DAQPRO-5300 what type of sensor is connected to each input
Chapter 1 OM-DAQPRO-5300 17
Page 20
1. In the Main Menu screen, use the Forward arrow buttons to
select the setup menu icon
2. Press the Enter button
→ Input – 1 Voltage 0–10V Input – 2 Empty Input – 3 Empty Input – 4 Empty
RATE = Every sec SAMPLES = 500 DISPLAY = numeric
3. Use the Forward
select the sensor in input 1 and then press the Enter button. The arrow indicator will move to the second input.
4. Repeat this procedure with all the sensors you plugged in.
5. After the fourth sensor has been selected, the screen will list the next 4 sensors:
→ Input – 5 Empty Input – 6 Empty Input – 7 Empty Input – 8 Empty
RATE = Every sec SAMPLES = 500 DISPLAY = numeric
to enter the setup menu:
and Backward arrow buttons to
.
18 Chapter 1 OM-DAQPRO-5300
Page 21
You can press the Enter button in the last input if you want to go back to the first input.
6. When you’ve finished selecting the sensors press the Escape button.
The arrow indicator will point to the Rate command.
4. Select Rate
Use the Forward
Enter button
OM-DAQPRO-5300 automatically switches to the next step in the setup process and the arrow moves to the “samples” row.
5.Select total number of samples
Use the Forward
press the Enter button.
and Backward arrow buttons to select the desired rate, then press the
.
Input – 1 Curent 0–24mA Input – 2 NTC 10K Input – 3 Input – 4
RATE = Every 10 sec → SAMPLES = 500 DISPLAY = numeric
and Backward arrow buttons to select the number of samples, then
Chapter 1 OM-DAQPRO-5300 19
Page 22
Input – 1 Microphone Input – 2 Light Input – 3 Input – 4
RATE = Every sec SAMPLES = 10,000 → DISPLAY = numeric
6. Choose display
Use the Forward
the Enter button
Note: When you turn the OM-DAQPRO-5300 off it will save the setup for the next
session.
7. Start recording
Press the Forward arrow button
and Backward arrow buttons to select the type of display, then press
.
Input – 1 Microphone Input – 2 Light Input – 3 Input – 4
SAMPLES = 500 DISPLAY = graphic → START = ( > ).
to start recording.
20 Chapter 1 OM-DAQPRO-5300
Page 23
Or press the Enter button if you want to go back to the first item (Rate).
You can stop recording at any time by pressing the Escape button .
1.2.3. Working with the OM-DAQPRO-5300 Menus
Turning OM-DAQPRO-5300 on and off
Note: Pressing OFF will not erase the sample memory. The data stored in the memory
1.Main menu display
When turned on, OM-DAQPRO-5300 opens with a system information window and then displays the main menu:
On
Off
will be kept for up to 5 years.
Turn OM-DAQPRO-5300 on
Turn OM-DAQPRO-5300 off
fn
?
Chapter 1 OM-DAQPRO-5300 21
Page 24
OM-DAQPRO-5300 has 9 menus. Use the Forward or Backward Arrow buttons to highlight a menu and press the Enter button to select it. Then use the Arrow buttons to scan the options. Press the Enter button to select an option. The OM-DAQPRO-5300 automatically executes the command.
2.Menu buttons
Forward
3.Menu Icons and commands
Backward
Enter (Start)
Escape (Stop)
Start
Press the Enter button to start recording
Setup
Rate – select recording rate
Move to the next menu or to the next menu options
Move to the previous menu or menu options
Enter the selected menu or select the current menu option and move to the next menu command or start recording
Return to the main menu or stop recording
Start recording
Setup OM-DAQPRO-5300 in 4 steps:
22 Chapter 1 OM-DAQPRO-5300
Samples – select the total number of recording points
Page 25
Display – select the way OM-DAQPRO-5300 will display the data (at a rate of
up to 1 sample per second):
1) Numeric – displays the sensor values and the sample number.
2) Meter – displays all active sensors in a bar meter display along with their values (without decimals).
3) Table – displays the last 6 values of all the active sensors in a table (without decimals).
4) Graphic – displays a graphic representation of the sampled sensors.
At rates higher then 1/s the OM-DAQPRO-5300 will display the data in a graph at the end of the logging period.
Start – press the Forward arrow to start recording
fn
Function
1) Minimum – the minimum graph value.
2) Maximum – the maximum graph value.
3) Average – the graph average.
Use the Enter button to browse the different sensors
Open
Use the Forward and Backward Arrow buttons to browse the stored files,
press the Enter button to open a file
Notes
Use the Arrow buttons to browse the notes.
You can use the DaqLab software to edit the notes or to write new notes
Display statistics of the current data
Open a stored data in graphic display
Display user information
Chapter 1 OM-DAQPRO-5300 23
Page 26
System information
1) Comment (use DaqLab to enter a comment)
2) Number of files stored in OM-DAQPRO-5300’s internal memory
?
3) Memory usage
4) Battery level (unplug any sensor and external power supply)
5) Ambient temperature (the readings of the internal temperature sensor)
6) Current date and time
Help
Configuration
Filter – turn input filter on and off
Average – select number of averaging points
Temperature units – select between Fahrenheit and Celsius
Clear memory– delete the stored data files
Display system information:
On-line help and specifications
Configure the OM-DAQPRO-5300:
Contrast – use the arrow buttons to tune the screen contrast
1.2.4. Graphic Display
OM-DAQPRO-5300 will automatically use a graphic display in three cases:
1. If the recording rate is every second or less and you selected Graphic Display in the setup menu
2. Once logging has ended for a recording rate of over one per second.
3. When opening a stored recording.
The graphic display is available for recordings with up to 4 sensors simultaneously.
24 Chapter 1 OM-DAQPRO-5300
Page 27
2.The Cursor
Use the Cursor in Graphic Display mode to read data values or to zoom in to the area around a selected point. The cursor is displayed automatically after logging has ended, or when opening a stored recording.
1. Use the Forward move the cursor.
2. OM-DAQPRO-5300 displays the point coordinates at the bottom of the graph.
3. If there is more then one sensor, press the Enter button move the cursor to another plot on the graph.
3. Zooming
1. Position the cursor in the area you want to zoom into.
2. Press the two arrow buttons, will zoom in around the cursor in a 2:1 ratio.
3. Press the Escape button to zoom out.
and Backward arrow buttons to
to
and , simultaneously. You
1.2.5. Load the Last Setup
When you turn OM-DAQPRO-5300 on, once the self testing has been completed, it will display:
_ _ _ INITIALIZATION _ _ _
OM-DAQPRO-5300 Ver 1.0 Init. Display Init. RAM Batt. 99%
ENTER – load setup
Chapter 1 OM-DAQPRO-5300 25
Page 28
Press the Enter button
to load the last setup.
1.2.6. Configure Your OM-DAQPRO-5300
Use the System configuration menu to turn input filter on and off, to select number of averaging points, to set temperature units, to clear the OM-DAQPRO-5300’s memory or to change the screen contrast.
In the MAIN MENU screen, select the System configuration icon to display the configuration screen:
_ _ _ CONFIGURATION _ _ _
→ Input filter Off Average 4 samples Temperature in ° C Clear memory ( > ) Contrast ( < ) ( > )
Use the Forward
Enter button
configuration menu at any time, saving the new changes you made. Press Enter button in the last item (Contrast) if you want to go back to the first item (Input Filter).
The new configuration will be saved until the next time you change it.
and Backward arrow buttons to select the mode and then press the
to move to the next item. You can press the Escape button to leave the
1. Input Filter
To filter out 50/60Hz line noises use the low pass RC input filter.
Use the Forward
When turned on, the low pass filter is applied to all inputs except for 24mA current measurements
26 Chapter 1 OM-DAQPRO-5300
and Backward arrow buttons to turn input filter on and off
Page 29
2.Averaging points
Use this option to reduce random noises. OM-DAQPRO-5300 replaces every data sample with the average of the last preset number of samples.
Use the Forward
3.Temperature units
Use the Arrow buttons to select between Fahrenheit (°F) and Celsius (°C) temperature units.
4.Clear memory
Press the Forward arrow button DAQPRO-5300.
5.Screen contrast
Use the arrow buttons to adjust the LCD screen contrast.
Any contrast adjustment will be saved until the next time you change it.
and Backward arrow buttons to select the number averaging samples.
if you want to delete all previous data files from the OM-
1.2.7. Internal Clock and Calendar
The internal clock is set the first time you use the Setup command from the DaqLab software to program the OM-DAQPRO-5300, and is automatically updated to the PC’s time and date each time you connect your OM-DAQPRO-5300 to a PC.
The internal clock and calendar is kept updated independent of the 7.2V battery condition, even when the OM-DAQPRO-5300 is turned off.
1.2.8. Clear the Memory
If you want to start recording and the OM-DAQPRO-5300’s internal memory is full you will see this message at the bottom of the display:
Chapter 1 OM-DAQPRO-5300 27
Page 30
Input – 1 Voltage 0–10V Input – 2 Empty Input – 3 Empty Input – 4 Empty
SAMPLES = 200 DISPLAY = graphic Mem full, clear = ( > )
Press the Forward arrow button
In order to clear the OM-DAQPRO-5300’s memory when it is not full, use the Memory clear command from the Configuration menu (see page 26), or clear the memory from the Logger menu in the DaqLab software (see page 59).
to clear the memory
1.2.9. Choose the Right Setup.
1. Sampling rate - The sampling rate should be determined by the rate of change of the phenomenon being sampled. If the phenomenon is periodic, sample at a rate of at least twice the expected frequency. Changes in temperature can be measured at slower rates such as once per second or even slower, depending on the speed of the expected changes. Usually it’s a good practice to sample at a rate 10 times higher than the expected frequency but for extremely smooth graphs, the sampling rate should be about 20 times the expected frequency.
Note: Sampling at a rate slower than the expected rate can cause "frequency aliasing".
In such a case, the graph will show a frequency much lower than expected. In Figure 6 below, the higher frequency sine wave was sampled at 1/3 of its frequency. Connecting the sampled points yielded a graph with a lower, incorrect frequency.
28 Chapter 1 OM-DAQPRO-5300
Page 31
Figure 6: Frequency Aliasing
Manual sampling - use this mode for:
• Recordings or measurements that are not related to time.
• Situations in which you have to stop recording data after each sample
obtained, in order to change your location, or any other logging parameter (Note: During recording NO CHANGES can be made to the OM-DAQPRO­5300’s configuration).
To start a recording using manual data logging, set the RATE to “manual” and press the Enter
button collect a sample.
2. Sampling Points - After you have chosen the sampling rate, choosing the number of points will determine the logging period: Samples / Rate = Logging time. You can also choose the duration of a run first, and then calculate the number of samples: Samples = Logging time × Rate.
Continuous
In the Continuous mode, OM-DAQPRO-5300 must be connected to the PC and the DaqLab software must be running. In this mode OM-DAQPRO-5300 can continue logging indefinitely. The data is automatically downloaded to the computer, displayed in a real time graph and saved every 10 minutes.
OM-DAQPRO-5300 stores the data until its memory is full. You can download this data only if you stopped logging before this point
To operate in Continuous mode select RATE equal to or less than 1/s and SAMPLES = Continuous.
You can also select Continuous mode directly from the DaqLab software.
Note: OM-DAQPRO-5300 must be set to a display mode other than graphic to enable
Chapter 1 OM-DAQPRO-5300 29
once to start the data recording, then press the Enter button each time you want to
the Continuous mode.
Page 32
1.2.10. Programming Rules and Limitations
The following are some rules and limitations you must take into account when programming the OM­DAQPRO-5300, as OM-DAQPRO-5300 integrates all programming limitations automatically. OM­DAQPRO-5300 will only allow the programming of settings that comply with the rules below.
1.Sampling rate:
The number of sensors in use limits the maximum sampling rate:
1. Maximum sampling rate with one sensor is 4000/s
2. Maximum Temperature sampling rate is 1/s
3. Maximum sampling rate with two sensors or more is 1/s
4. Maximum sampling rate with two or more PT 100 sensors or more is every 10 seconds
5. Maximum sampling rate with eight PT 100 sensors or more is every minute
6. OM-DAQPRO-5300 displays readings in real time at rates up to 1/s
7. DaqLab displays online readings at rates up to 100/s
2.Sampling points:
Increasing the number of active inputs limits the number of sampling points one can choose. The following condition must be always satisfied:
Number of samples × Active Inputs ≤ Memory
OM-DAQPRO-5300’s memory is sufficient for 512,000 samples.
The table below lists the maximum number of sampling point according to the number of sensors:
Number of sensors Maximum number of sampling points
1 512,000 (exact number: 522,752)
2 256,000 (exact number: 260,608)
3 128,000 (exact number: 129,536)
4 128,000 (exact number: 129,536)
5 64,000
6 64,000
7 64,000
8 64,000
3.Continuous sampling
• Continuous sampling is possible up to a maximum sampling rate of 1/s.
• The data must be presented in a display mode other than graphic.
30 Chapter 1 OM-DAQPRO-5300
Page 33
Chapter 2
Working with DaqLab
2.1. Install the Software
2.1.1. System Requirements
To work with DaqLab, your system should be equipped with the following:
1.Software
Windows 98 or later
Internet Explorer 5.0 or later (you can install Internet Explorer 5 when you install
DaqLab, since it ships with the product)
2.Hardware
Pentium 200MHz or higher
32 MB RAM (64 MB recommended)
10 MB available disk space for the DaqLab application (50 MB to install the supporting
applications)
2.1.2. Installation
1. Close all programs
2. Insert the CD labeled DaqLab into your CD-ROM drive
Installation will begin automatically. Simply follow the on-screen instructions to continue
In case auto run is not working, open My Computer and click on the CD drive folder (d: drive in most cases) and double-click on the setup icon, then follow the on-screen instructions.
To uninstall the software: From the Start menu select Settings and click on Control Panel, then use the Add/Remove programs function to remove the DaqLab application.
When updating the software, always remove the old version before starting a new installation.
Chapter 2 Working with DaqLab 31
Page 34
To install the USB driver:
1. Insert the CD into your CD drive. If Installation begins automatically (and you have already installed DaqLab), click Cancel to stop installation
2. Connect the OM-DAQPRO-5300 to a USB port on your PC and turn the OM-DAQPRO-5300 on. Windows will automatically detect the new device and open the Add New Hardware Wizard
3. Select Specify the location of the driver, then click Next
4. Select Search for the best driver for your device, then check the Removable Media checkbox, and then click Next
Windows will automatically detect and install the necessary software.
Note to the Windows XP user – if the found New Hardware wizard prompts you to a Hardware Installation dialog:
Just click Continue Anyway to proceed.
32 Chapter 2 Working with DaqLab
Page 35
2.2. Overview
2.2.1. DaqLab On-screen Layout
DaqLab is a comprehensive program that provides you with everything you need in order to collect data from the OM-DAQPRO-5300 display the data in graphs, meters and tables and analyze it with sophisticated analysis tools.
The program includes three windows: A graph window, table window, and a navigation window called the Data Map. You can display all three windows simultaneously or in any combination.
The most commonly used tools and commands are displayed on two toolbars. Tools that relate to all aspects of the program and tools that control the OM-DAQPRO-5300 are located in the main (upper) toolbar. Tools specific to the graphs are located on the graph (lower) toolbar.
2.2.2. Working with Projects
Every time you start a new recording, DaqLab automatically creates a new project file. All the information you collect and process for a given session is stored in a single project file. Each of these files contain all the data sets you collect with the OM-DAQPRO-5300, the analysis functions you’ve processed, specific graphs you’ve created, and the DaqLab settings for the recording.
Note: all data sets in a single project must be with the same sampling rate.
Chapter 2 Working with DaqLab 33
Page 36
2.2.3. DaqLab Window Layout
Main toolbar
Data map
Graph window
Table window
Information bar
Graph toolbar
34 Chapter 2 Working with DaqLab
Figure 7: DaqLab window layout
Page 37
2.3. Getting Started
2.3.1. Set up a Recording Session
1. Prepare OM-DAQPRO-5300
1. Connect OM-DAQPRO-5300 to the PC (see page 10)
2. Turn on OM-DAQPRO-5300
3. Plug in any external sensors
4. Open the DaqLab software
You can open DaqLab by double clicking its icon
2. Setup the OM-DAQPRO-5300
1. Click Setup Wizard
2. Follow the instructions in the Setup Wizard (see page 52)
3. Start Recording
Click Run
If the recording rate is 100 measurements per second or less, DaqLab automatically opens a graph window displaying the data in real time, plotting it on the graph as it is being recorded. If the recording rate is higher than 100/s, the data will be downloaded and displayed automatically, once the data recording is finished.
You can stop recording anytime by clicking Stop
on the toolbar to start recording.
on the main toolbar
on the desktop or from the start menu
on the toolbar.
Chapter 2 Working with DaqLab 35
Page 38
2.3.2. Data recording options
To set the behavior of the data display when you start a new recording session, click on the down arrow
next to the Run button
1.Single measurement
DaqLab will open a new project file every time you start a new recording session.
2. Replace
DaqLab will display the new data set in place of the old set. The project’s old data sets will still be available in the same project file. They will be listed in the Data Map and you can add them to the display at any time
3. Add
DaqLab will add the new data set to the graph in addition to the old ones.
Note: A maximum of 8 data sets can be displayed on the graph at the same time.
, and select one of the following:
2.3.3. Download Data
Whenever data is received from the OM-DAQPRO-5300, it is accumulated and displayed automatically by DaqLab. There are two modes of communication: online and offline.
Online communication
When OM-DAQPRO-5300 is connected to the PC and programmed to run at sampling rates of up to 100/s, OM-DAQPRO-5300 transmits each data sample immediately, as it is recorded, to the PC. The software thus displays the data in real-time in both the graph window and the table window.
When OM-DAQPRO-5300 is connected to the PC and programmed to run at a sampling rate of 500/s or higher, data is accumulated in OM-DAQPRO-5300‘s internal memory. This data is not transmitted to the PC until the recording period has ended, when the data is automatically downloaded to the PC and displayed.
Offline data logging
• To download data that was recorded offline, or while OM-DAQPRO-5300 was not
connected to a PC, connect OM-DAQPRO-5300 to the computer, run the DaqLab
36 Chapter 2 Working with DaqLab
Page 39
program and click Download on the main toolbar. This will initiate the Post­Recording Data Transfer communication mode. Once the transfer is complete, the data will be displayed automatically in the graph window and in the table window. If there are several files stored in the OM-DAQPRO-5300, the first download will bring up the most recent file; the second download will bring up the second most recent file, and so on.
• If you are logging offline at rates of up to 1/s, you can connect OM-DAQPRO-5300 to
the computer and download the accumulated data at anytime. Apart from during the transfer period, the data will be downloaded without interrupting the logging process.
• To download a particular file, choose Selective download from the Logger menu,
then select the file’s number in the Download dialog.
• Click Cancel in the Download progress window at any time to stop downloading the
data.
2.3.4. Save Data
Click Save on the main toolbar to save your project. This will save all the data sets and graphs under one project file.
Saving the project will also save any special formatting and scaling you did.
If you made any changes to a previously saved project, click Save to update the saved file or select
Save as… from the file menu to save it under another name.
Note:
To delete a specific data set, a graph or a table from the project, use the Data Map (see page 49)
To remove unwanted data from a specific data set, apply the crop tool (see page 82).
2.3.5. Open a File
1. Click Open on the main toolbar
2. Navigate to the folder where the project is stored
Chapter 2 Working with DaqLab 37
Page 40
3. Double click the file name to open the project
DaqLab opens the project and displays the first graph on the graph list. If the project does not include saved graphs, the file opens with an empty graph window. Use the Data Map (see page 49) to display the desired data set.
2.3.6. Create a New Project
There are three ways to create a new project:
1. Open the DaqLab program, which will open a new file each time
2. When working in Single Measurement mode, a new project is opened every time you click on the Run button to start a new recording
3. Any time you click New button
on the toolbar
2.3.7. Import data
Any file that is in comma separated values text format (CSV) can be imported into DaqLab
To import a CSV file:
1. Click File on the menu bar, then click Import CSV file
2. In the dialog that opens, next to Look in, navigate to the drive and folder that contains the CSV file
3. Select the file
4. Click Open
Tips:
To create a text file in a spreadsheet:
1. Open a new spreadsheet
2. Enter your data according to the following rules:
a) The first row should contain headers. Each header includes the name
of the data set and units in brackets, e.g. Distance (m)
38 Chapter 2 Working with DaqLab
Page 41
b) The first column should be the time. The time interval between
successive rows must match the time intervals accepted by DaqLab. You can export DaqLab files to Excel to learn about these time formats
See for example the table below:
3. On the File menu, click Save As
4. In the File name box, type a name for the workbook
5. In the Save as type list, click the CSV format
6. Click Save
To import files that were previously exported from DaqLab open DaqLab and import the file as described above as they are already in CSV format.
2.3.8. Date format settings
To set the way data will be displayed on screen click Date format settings to open the Date format settings dialog:
Chapter 2 Working with DaqLab 39
Page 42
Click the desired option, then click OK
2.3.9. Print
1. Print a graph
1. Click Print
2. Select the Graph 1 option (when in split graph mode you can choose between Graph 1 and Graph 2)
3. Click Print to open the print dialog
4. Click OK
DaqLab will print exactly what you see in the graph display.
40 Chapter 2 Working with DaqLab
on the main toolbar
Page 43
2. Print a table
1. Click Print
2. Select the Table option
3. If you want to print only a specific range, uncheck the Print all data check box and type the desired row numbers into the To and From edit boxes
4. Click Print to open the print dialog
5. Click OK
DaqLab will print exactly what you see in the table display as well as the OM-DAQPRO-5300 comment, serial number and the alarm level setup. Data that exceeds any of the alarm levels will be highlighted by arrows.
on the main toolbar
Chapter 2 Working with DaqLab 41
Page 44
2.4. View the Data
2.4.1. Display Options
The DaqLab program’s screen consists of three parts: the graph window, table window and Data Map window. You can display all three parts simultaneously (the default view) or any combination of the three.
The graph window is the main window by default and is and displayed in the center of the application window.
In addition to these sections, you have the option to display an on-screen meter for each of the sensors (see page 48).
2.4.2. Graph Display
Click Graph to display or hide the graph. The default graph display is the data set or sets plotted vs. time, but you can change the X-axis to represent any of the individual data sets (see page 44).
The graph usually displays all the data sets of a given recording, but you can use the Data Map to remove one or more of the sets from the graph (see page 49).
In order to keep the graph clear and simple, only two Y-axes are shown on the graph at once. If there are three curves in the graph, one of the Y-axes is hidden. In order to make this axis visible, select the corresponding plot with the cursor (see section 2 below).
You can identify the Y-axis by its color, which matches the plot color.
1. Split graph view
DaqLab enables you to display your data in two separate graphs within the graph window.
1. Click Split graph into two separate graphs
2. Click Edit graph dialog
3. Choose which data sets to display on each of the graphs (or use the Data Map to do so – see page 49)
42 Chapter 2 Working with DaqLab
on the graph toolbar to split the graph window
on the graph toolbar to open the Edit graph
Page 45
4. To return to the single graph display, click Split graph a second time
2. The Cursor
You can display up to two cursors on the graph simultaneously.
Use the first cursor to display individual data recording values, to select a curve or to reveal the hidden Y-axis.
Use two cursors to display the difference between two coordinate values, to display the frequency of periodic data or to select a range of data points.
To display the first cursor:
Double click on an individual data point or click Cursor
on the graph toolbar. You can drag the cursor with the mouse onto any other point on the plot, or onto a different plot. For finer cursor movements use the forward and backward keys on the keyboard.
The coordinate values of the selected point will appear in the information bar at the bottom of the graph window.
To display the second cursor:
Double click again anywhere on the graph area or click 2
nd
Cursor .
The information bar will now display the difference between the two coordinate values.
To remove the cursors:
Double click anywhere on the graph area, or click 1
st
Cursor a second time.
To remove the 2nd cursor:
Click 2nd Cursor a second time.
To display the cursors in split graph mode:
To display the cursors on the upper graph, use the same method as for single graph mode.
To display the cursors on the lower graph, you must first remove them from the upper graph and then double click anywhere on the lower graph to display the first cursor. Double click a second time to display the second cursor, and double click a third time to remove the cursors.
3. Zooming
1. To zoom in to the center of the graph
1. Click Zoom in
on the graph toolbar
Chapter 2 Working with DaqLab 43
Page 46
2. To reverse the operation, click Zoom out on the graph toolbar
2. To zoom in to a specific data point
1. Select the point with the cursor (see above)
2. Click Zoom in
3. To reverse the operation click Zoom out on the graph toolbar.
3. To zoom in to a range
1. Select the range with both cursors
2. Click Zoom in on the graph toolbar
3. To reverse the operation click Zoom out on the graph toolbar
4. To zoom in to a specific area
Click Zoom to selection want to magnify. Release the mouse button to zoom in to the selected area.
Click Zoom to selection a second time to disable the zoom tool.
5. Autoscale
Click Auto scale auto scale that axis alone.
6. Manual scaling
on the graph toolbar to view the full data display, or double click on an axis to
on the graph toolbar
on the graph toolbar and drag the cursor diagonally to select the area you
1. Click Graph properties properties dialog
2. Select the Scale tab, and choose the axis you want to scale in the Select axis drop list
44 Chapter 2 Working with DaqLab
on the graph toolbar to open the Graph
Page 47
3. Uncheck the Autoscale check box and enter the new values in the edit box
4. Click OK
To manually scale a specific axis, right click on the axis to open its Properties dialog.
To restore auto scaling click Autoscale
7. The stretch/compress axis tool
Move the cursor onto one of the graph axes. The cursor icon changes to the double arrow symbol ( indicating that you can stretch or compress the axis scale. Drag the cursor to the desired location.
Repeat the procedure for the other axis if necessary.
Double click on the axis to restore auto scaling.
4. Panning
Use the pan tool after zooming in to see any part of the graph that is outside the zoomed area.
To do this, click Pan view another area.
Click Pan a second time to disable the Pan tool.
5. Edit the Graph
Use the Edit graph dialog to select which data sets to display on the graph’s Y-axis and to change the X-axis from time, to one of the data sets.
1. Click Edit graph dialog:
on the graph toolbar, then click anywhere on the graph and drag the mouse to
.
↔),
on the graph toolbar to open the Edit graph
Chapter 2 Working with DaqLab 45
Page 48
Figure 8: Edit graph dialog
2. To select a data set to display on the Y-axis, click on the data set’s name in the Y-axis list. To display more than one curve, click on the data sets you want
3. A list entry that begins with a OM-DAQPRO-5300 comment denotes a recorded data set. A list entry that begins with an input number denotes the next recording and will be displayed on the graph the next time you start a recording
4. To deselect a data set, click on it a second time
5. To select a data set for display on the X-axis, click on the data set’s name in the X-axis list. You can only select one data set at a time for the X-axis
6. Click OK
6. Display alarm levels
1. Click Display alarm level
2. Select the sensor you wish to display from the select sensor drop list
To display alarm levels on graph 2 in split graph mode click the down arrow next to the button and select graph 2.
7. Format the graph
46 Chapter 2 Working with DaqLab
.
Page 49
You can change the data line’s color, style and width. You can also add markers that represent the data points on the graph and format their style and color.
The color of the Y-axis matches the corresponding plot’s color and will automatically change with any change made to the color of the corresponding plot.
1. Click Graph properties properties dialog
2. Select the Lines tab, and then select the plot or axis you want to format in the Select plot drop list
3. From here you can format the line’s color, style and width, as well as the markers’ color and style. To remove the line or the marker, uncheck the corresponding Visible check box
4. Click OK
5. To restore the default formatting, click Restore default
8. Change the graph’s units and its number format
1. Click Graph properties properties dialog
2. Select the Units tab, and then select the plot or axis you want to format in the select plot drop list
3. Choose the prefix option you want
4. Select the desired number of decimal places
5. To display numbers in scientific format, check the Scientific check box
on the graph toolbar to open the Graph
on the graph toolbar to open the graph
6. Click OK
9. Add a graph to the project
DaqLab displays new data in the graph window every time you start a new recording. You can always display previous data using the Edit graph dialog or by double-clicking on the data’s icon in the Data
Chapter 2 Working with DaqLab 47
Page 50
Map. If you want to save a graph that you created to your project, or to update a saved graph with changes you made, use the Add to project tool:
Click Add to project on the graph toolbar.
2.4.3. The Table Display
Click Table to display or to remove the table window.
The data in the table always matches the data that is currently displayed on the graph.
When you start a new recording, DaqLab displays the new data in the table as well as on the graph.
48 Chapter 2 Working with DaqLab
Page 51
Formatting the table
1. Changing column width
Drag the boundary on the right side of the column heading until the column is the desired width.
2. Changing row height
Drag the boundary below the row heading until the row is the desired height.
3. Formatting the fonts
1. Click Table on the menu bar, then click Properties
2. Select the Font tab
3. Format the font, as well as the font’s style and size
4. Click OK
4. Changing units and number format
1. Click Table on the menu bar, then click Properties
2. Select the Units tab, and then select the plot you want to format from the select plot drop list
3. Choose the prefix option you want
4. Select the desired number of decimal places
5. To display numbers in scientific format, check the Scientific check box
6. Click OK
2.4.4. Meters
DaqLab enables you to view data in meters format on the screen (one meter for each sensor), with up to four meters showing at once. The meters can display live data while DaqLab is recording, or saved data when a saved file is replayed.
When a cursor is displayed, the meter shows the measured values that correspond to the time of the point at which the cursor is positioned.
There are three meter types: analog, bar and digital. The meter’s scaling automatically matches the graph’s scaling.
Chapter 2 Working with DaqLab 49
Page 52
To set up the meters:
1. Click Meter Setup
2. Select the meter type, and the data set to be displayed
3. A list entry that begins with a graph number denotes a displayed data set. A list entry that begins with an input number denotes the next recording, and will be displayed on the meter the next time you start a recording
4. Repeat this procedure for up to four meters
5. To remove the meters click Meter Setup
on the main toolbar
, and click Remove all
2.4.5. Data Map
Click Data Map to display or remove the Data Map.
The data map is a separate window that displays the list of data sets that were recorded or downloaded in the current session, as well as the lists of all the saved graphs. Use the Data Map to navigate through the available data sets and to keep track of the data that is being displayed in the graph window.
Note: The data in the table always matches the data that is currently displayed on the
graph.
1. Control the display with the Data Map
The items in the Data Map are sorted into two main categories:
• Data sets (including analysis functions)
• Saved graphs
Double click on a category to bring up the full list. Double click a second time to collapse the list. You can also use the plus (+) and minus (-) signs next to the icons to expand or collapse the categories.
The Data sets' list expands to sub-categories of recorded data and functions. To display the complete list of measurements, or the complete list of analysis functions performed on the measurements for any individual data capture, double click the file’s icon or click the plus sign (+) next to it.
To collapse a list under an individual data capture, double click the corresponding icon or click the minus sign (-) next to it.
50 Chapter 2 Working with DaqLab
Page 53
To display a data set or a saved graph double click its icon. Double click a second time to remove it.
You can also use a shortcut menu to display or remove a data set from the graph. Simply right-click an icon, then click display on graph #1 or display on graph #2.
To delete an individual item from the Data Map and from the project, right-click on its icon and click
Delete. To permanently delete the item, click Save
To remove all data sets from the Data Map, right-click the Data sets icon
2. Understanding Data Map icons
Data sets list
Functions – includes all the analysis functions of the project
Individual function – currently not on display
Individual function – currently displayed on graph 1
Individual function – currently displayed on graph 2
Individual function – currently displayed both on graphs 1 and 2
Function
data capture – includes all recorded data of the recording
Individual data set – currently not on display
Individual data set – currently displayed on graph 1
Individual data set – currently displayed on graph 2
Individual data set – currently displayed both on graphs 1 and 2
Raw data
List of saved graphs
Saved graph – data sets vs. time
Saved graph – data sets vs. data set
Graph currently not on display
Graph currently displayed on graph 1
Graph currently displayed on graph 2
Graph currently displayed both on graphs 1 and 2
on the main toolbar.
and click Clear All Data.
Chapter 2 Working with DaqLab 51
Page 54
2.4.6. Export Data to Excel
Click Export to Excel on the graph toolbar to export the data that is currently displayed on the graph to Excel. DaqLab will then open a new Excel workbook displaying the data as well as the OM­DAQPRO-5300 comment, serial number and the alarm level setup. Data that exceeds any of the alarm levels will be highlighted by arrows.
Export file settings
If DaqLab fails to export the data properly, try to change the export file settings:
1. Click File on the menu bar, then click Export file settings
2. Check the Ignore regional settings check box
3. Click OK
2.4.7. Copy the Graph as a Picture
You can copy the graph to the clipboard as a picture and then paste it to other Windows programs such as Word or PowerPoint:
1. In the Graph menu, click Copy graph
2. Open the destination file
3. In the destination file, right click and select Paste
52 Chapter 2 Working with DaqLab
Page 55
2.5. Program OM-DAQPRO-5300
2.5.1. Setup
1. Quick setup
Use the Setup wizard to guide you through the setup process.
1. Click Setup wizard on the main toolbar to open the setup wizard:
Figure 9: Setup wizard step 1 of 3
The first step is to assign sensors to each input. Select the sensors you are using from the corresponding input drop lists.
To set alarm levels and alarm action click Display alarm settings (see page 54 for detailed alarm setup)
2. Click Next to move to the second step:
Chapter 2 Working with DaqLab 53
Page 56
Figure 10: Setup wizard step 2 of 3
3. Select the desired recording rate in the Select rate drop list
4. Select the Scroll option if you want the graph window to scroll as the data recording progresses
5. Select the desired data recording mode in the Select recording mode drop list
6. Click Next to move to the third step:
54 Chapter 2 Working with DaqLab
Page 57
×
Figure 11: Setup wizard step 3 of 3
7. Select the desired length of the recording period in the Select recording time drop list. The recording period is displayed as a unit of time by default. To display the recording period as the number of recording samples, select the By samples option. The number of samples is calculated with the following formula:
Total recording samples
8. To start the data recording only when a specific time or measurement condition has been met click Triggering to open the triggering dialog (see page 57)
9. Click Finish to complete the setup and to send the setup command to the OM-DAQPRO-5300
2. Alarm setup
You can setup alarm levels and output for each sensor input individually in the first step of the Setup wizard:
1. Click Setup Wizard wizard
2. Click Display alarm settings to show the alarm controls:
on the main toolbar to open the setup
=
Recording periodRecording rate
Chapter 2 Working with DaqLab 55
Page 58
Figure 12: alarm settings
3. Check the checkbox or checkboxes near the inputs you want to activate the alarm
4. Type in Low or/and High alarm levels
5. If you want a sensor or sensors to activate the alarm output check the corresponding Output checkbox or checkboxes
6. Click Next to proceed to step 2 of the setup wizard or click Finish to complete the alarm setup
See page 13 to learn more on alarms and alarm output.
You can set DaqLab to display the alarm levels on the graph window (see page 45)
3. Low pass filter
To filter out 50/60Hz line noises use the low pass RC input filter:
Check the Input filter checkbox in step 2 of the Setup wizard
Note: We strongly recommend to use the filter with low output sensors like 50mV voltage measurements
and thermocouples.
The low pass filter is applied to all inputs except for 24mA current measurements
56 Chapter 2 Working with DaqLab
Page 59
4. Averaging
Use this option to reduce random noises. OM-DAQPRO-5300 replaces every data sample with the average of the last pre-set number of samples
To apply averaging:
Check the Averaging checkbox in step 2 of the Setup wizard, and select the number of samples to average from the drop list
5. Battery level
A battery level indicator is displayed in step 3 of the Setup wizard. The indicator shows the remaining working hours as a percent of the battery’s life (approximately 40 hours between charges), before it should be recharged.
6. Presetting the display
You may want to define the graph’s settings, such as formatting and scaling the graph and selecting the axes, and to define the meters’ settings prior to beginning a recording session. You can define the settings of each sensor input individually in the first step of the Setup wizard:
1. Click Setup Wizard wizard
2. Click Properties
3. Click the Display properties tab
4. Select a measurement from the Select measurement drop list
5. Select the graph or graphs in which you would like to display the data, from the Display on graph drop list
6. Select a meter in the Meter type drop list if you want to display the data in a meter as well as on the graph
7. Format the line and markers in the Plot section
8. If you want a specific scale, uncheck the Auto scale check box and enter the desired minimum and maximum values of the axis
9. Click OK
Chapter 2 Working with DaqLab 57
on the main toolbar to open the setup
next to the input you want to set
Page 60
7. Preset the graph’s X-axis
1. Click X-Axis display properties in the first step of the Setup wizard to open the dialog:
Figure 13: X-axis properties dialog
2. Select the graph you wish to modify
3. Select the measurement you’d like to display on the X-axis
4. If you want a specific scale, uncheck the Auto scale check box and enter the minimum and maximum values of the axis
5. Click OK
8.Triggering
To start the data recording only when a specific time or measurement condition has been met, click Triggering in step 3 of the Setup wizard to open the triggering dialog:
58 Chapter 2 Working with DaqLab
Page 61
Figure 14: Triggering dialog
1. Select the triggering sensor in the Based on sensor drop list
2. Choose one of the following from the Type options:
• None – Trigger is disabled
• Above level – Start logging only once the measured value is
the trigger level.
• Below level – Start logging only once the measured value is
trigger level.
• Event Recording – This function enables you to record the exact time and
date at which a certain phenomenon occurred: The trigger level set for this option is actually a threshold setting. Each time the threshold is crossed, OM­DAQPRO-5300 will record the exact time and date of the occurrence, and will continue to do so until the desired number of samples has been obtained.
When logging has ended download the data to the PC (if OM-DAQPRO-5300 is already connected to the PC it will be done automatically). The event list will be displayed in the table window. You can now save the list, export it to Excel or print it.
HIGHER than
LOWER than the
Note: The trigger condition must be fulfilled for at least 300µS.
• Time delay – This trigger type enables you to set a timer that will start the
logging after a predetermined amount of time. After setting the trigger to Timer Delay, click the down arrow on the Level combo-box, and select from the 17
Chapter 2 Working with DaqLab 59
Page 62
different time options. The timer will start its countdown when you click RUN, and the actual recording will start once the countdown has ended.
3. Select the trigger level in the Level edit box
4. Click OK
Note: When you turn off OM-DAQPRO-5300, it will save the setup for the next session.
2.5.2. Start Recording
Click Run on the main toolbar. Click the down arrow to change the recording mode.
2.5.3. Stop Recording
Click Stop on the main toolbar
2.5.4. Clear OM-DAQPRO-5300’s Memory
To erase all data currently stored in the OM-DAQPRO-5300, click Clear memory in the Logger menu (OM-DAQPRO-5300 must first be connected to the PC).
2.5.5. Adding comment to OM-DAQPRO-5300
You can add a comment to each OM-DAQPRO-5300. The comment will serve as a name to identify the specific OM-DAQPRO-5300 in addition to its serial number. Every time data is being downloaded from a OM-DAQPRO-5300, its comment and the serial number will be displayed in the graph title and on the corresponding icon in the Data Map. To display the comment on OM-DAQPRO-5300’s LCD select the System information menu (see page 22)
The comment and the serial number will also be displayed in prints and exported files.
Note: the comment can include up to 32 characters.
To add a comment:
1. Connect OM-DAQPRO-5300 to the PC
2. Click Logger on the menu bar, then click Set comment
60 Chapter 2 Working with DaqLab
Page 63
3. Click the Comment edit box and type a comment (e.g. its location)
4. Click OK to upload the edited notes to OM-DAQPRO-5300
2.5.6. Edit OM-DAQPRO-5300’s Notes
1. Connect OM-DAQPRO-5300 to the PC
2. Select Edit notes from the Logger menu to open the Notes dialog with the OM-DAQPRO-5300’s current notes downloaded to the window
Figure 15: Edit notes dialog
3. From this dialog, you can edit, delete, or write new notes
Note: A note is limited to 140 characters and a maximum of 3 notes can be stored in
OM-DAQPRO-5300 at any time.
4. Use the arrow buttons to move to the next note, or to the previous one
5. Click OK to upload the edited notes to OM-DAQPRO-5300
2.5.7. Temperature units
To select between Fahrenheit (°F) and Celsius (°C) temperature units:
Click Logger on the menu bar, point to Temperature units and then select the desired units
Chapter 2 Working with DaqLab 61
Page 64
2.5.8. Calibrating the sensors
1.Calibration options
OM-DAQPRO-5300 ships fully calibrated. However, DaqLab enables you to calibrate any of the OM­DAQPRO-5300 sensors except for the internal temperature sensor, which doesn’t need calibration. The calibration parameters are sent to OM-DAQPRO-5300 and stored in its memory.
Tips:
• Prior to any calibration procedure you must restore factory defaults
• Calibrating a sensor type applies to all inputs
• Calibrating the PT 100 2 wires sensor applies also to PT 100 3 wires except for fine
offset calibration
• Prior to calibrating any of the thermocouple sensor types, calibrate the Voltage 50mV
sensor as this will set the gain (slope) of all thermocouple sensor types
• The internal temperature sensor needs no calibration
DaqLab employs three different calibration methods Two-point calibration, Offset calibration and Zero calibration. When selecting a sensor type to calibrate, DaqLab prompts you to the appropriate calibration procedure:
Sensor Type Calibration method
Current 0 – 24mA Two-point calibration
Temperature
Temperature
Temperature PT-100 2 wire Two-point calibration
Temperature PT-100 3 wire Two-point calibration*
Temperature Thermocouple J Offset calibration
Temperature Thermocouple K Offset calibration
Temperature Thermocouple T Offset calibration
Voltage 0 – 10V Two-point calibration
Voltage 0 - 50mV Two-point calibration
NTC 10KΩ
NTC 2252Ω
Zero calibration
Zero calibration
Two-point calibration
The two-point calibration sets the gain (slope) and offset (intercept) of the sensor's conversion function.
62 Chapter 2 Working with DaqLab
Page 65
Use the two point calibration to calibrate the voltage and current sensors and the PT100 temperature sensors.
* To calibrate the PT100 3 wire, first calibrate the PT100 2 wire as the slopes of their conversion
functions are identical. Then proceed to adjust the PT100 3 wire offset.
In some cases you may need to refine your calibration using the fine offset tuning tool.
Offset calibration
To calibrate the thermocouple temperature sensors: TC-J, TC-K and TC-T first calibrate the 50mV sensor type. That will set the slope for all thermocouple temperature sensors. Then proceed to adjust the offset using the Offset calibration technique.
Zero calibration
The zero calibration sets only the offset (intercept) of the sensor's conversion function.
Use the zero calibration to calibrate the NTC 10kΩ and NTC 2252Ω temperature sensors.
2.Calibration password
To prevent accidental change of the calibration, the calibration procedure is protected by a password.
The default password is: 1234
To change the password:
1. Click Logger on the menu toolbar, and then click Calibration
2. Click Change Password to open the Change Password dialog box
3. Enter the current password in the Current Password edit box
4. Enter the new password in the New Password edit box
5. Enter the new password a second time in the Confirm New Password edit box to confirm your new password
6. Click OK
Note: The password must include at least 4 characters and is case sensitive.
3.Calibration procedure
Two point calibration:
Measure two reference values
Chapter 2 Working with DaqLab 63
Page 66
1. Connect OM-DAQPRO-5300 to the PC
2. Click Logger on the menu bar, then click Calibrate sensors
3. Enter the calibration password, then click OK
4. Choose a sensor from the Select sensor drop list and click OK to open the Sensor calibration dialog:
Figure 16: Two-point calibration dialog
Note: we strongly recommend calibrating using the values that appear in the two
Reference value edit boxes
5. Click Restore defaults, then click OK
6. Measure two known reference values and record both the reference values and the values displayed by DaqLab
64 Chapter 2 Working with DaqLab
Page 67
Enter calibration values
Repeat steps 1 to 4 above
7. Select the Two-point calibration option
8. Enter a distinct reference value in each of the Reference Value edit boxes and the corresponding logger values in each of the Logger Value edit boxes (The logger values are the values displayed by DaqLab when measuring the two reference values)
9. Click OK
Fine offset calibration
If, after the calibration procedure, OM-DAQPRO-5300 still exhibits some offset value use the Fine offset tuning tool to correct it.
The offset value is the difference between the value displayed by OM-DAQPRO-5300 and the reference value. For example if the calibrator is set to 0°C and OM-DAQPRO-5300 reads 1°C the offset value is +1, but if OM-DAQPRO-5300 reads −1°C then the offset value is −1.
1. Connect OM-DAQPRO-5300 to the PC
2. Click Logger on the menu bar, then click Calibrate sensors
3. Enter the calibration password, then click OK
4. Choose a sensor from the Select sensor drop list and click OK to open the Sensor calibration dialog
5. Select the Fine offset calibration option
6. Enter the offset value
7. Click OK
The calibrated sensor parameters will be saved both in OM-DAQPRO-5300 and DaqLab memories, so there is no need to calibrate OM-DAQPRO-5300 every time you run the DaqLab program.
Offset calibration:
Calibrating any of the thermocouple sensors comprises of two steps. First calibrate the 50mV voltage sensor to set the slope of the sensor's conversion function. Then use the calibration procedure described below to calibrate the sensor's offset.
Chapter 2 Working with DaqLab 65
Page 68
Measure two reference values
Before taking measurements you need to disable the temperature compensation both in OM­DAQPRO-5300 and in the calibrator.
To disable the temperature compensation in OM-DAQPRO-5300:
Click Logger on the menu bar, point to Thermocouple compensation, and then click Disable.
Note: don’t forget to enable the temperature compensation prior to normal operation of
OM-DAQPRO-5300. One way to enable the temperature compensation is to turn OM-DAQPRO-5300 off and on again. Another way is via DaqLab: Click Logger on the menu bar, point to Thermocouple compensation, then click Enable
To take two reference measurements:
1. Connect OM-DAQPRO-5300 to the PC
2. Click Logger on the menu bar, then click Calibrate sensors
3. Enter the calibration password, then click OK
4. Choose a sensor from the Select sensor drop list and click OK to open the Sensor calibration dialog:
Figure 17: Offset calibration dialog
66 Chapter 2 Working with DaqLab
Page 69
5. Click Restore defaults, then click OK
6. Measure two known reference values and record both the reference values and the values displayed by DaqLab. Calculate the offset values
The offset value is the difference between the value displayed by OM-DAQPRO-5300 and the reference value. For example if the calibrator is set to 0°C and OM-DAQPRO-5300 reads 1°C the offset value is +1, but if OM-DAQPRO-5300 reads −1°C then the offset value is −1.
Enter calibration values
1. Repeat steps 1 to 4 above
2. Enter the offset value in the Offset edit box (in case the two offset values are not identical enter the average of the two values)
3. Click OK
Zero calibration:
1. Connect OM-DAQPRO-5300 to the PC
2. Click Logger on the menu bar, then click Calibrate sensors
3. Enter the calibration password, then click OK
4. Choose a sensor from the Select sensor drop list and click OK to open the Sensor calibration dialog:
Chapter 2 Working with DaqLab 67
Page 70
Figure 18: Zero calibration dialog
5. Click Restore defaults, then click OK
6. Repeat steps 1 to 4 above
7. Short circuit input 1
8. Click Start
9. Wait until OM-DAQPRO-5300 begins recording
10. Click Get zero
11. Click OK
The calibrated sensor parameters will be saved both in OM-DAQPRO-5300 and DaqLab memories, so there is no need to calibrate OM-DAQPRO-5300 every time you run the DaqLab program.
68 Chapter 2 Working with DaqLab
Page 71
2.5.9. Define a Custom Sensor
DaqLab enables the user to define custom sensors. This is a useful tool for when OM-DAQPRO-5300 is communicating with many sensors from different vendors.
Any additional sensor that you would like to connect to OM-DAQPRO-5300 must comply with the following restrictions:
1) The sensor’s output must be either voltage in the range of 0 to 10V or current in the range of 4 to 24mA.
2) The sensor Transfer Function (sensor output voltage changes vs. the sampled phenomenon changes) must be a linear Transfer Function.
To define a new sensor:
1. Connect OM-DAQPRO-5300 to the PC
2. Click Logger on the menu bar, then click Define new sensor to open the Define new sensor dialog:
Figure 19: Define new sensor dialog
3. Click Add New Sensor
4. Select a Based on option (depending on whether your sensor’s output is 0 to 10V or 4 to 24mA)
5. Type in a sensor name and a sensor unit
Chapter 2 Working with DaqLab 69
Page 72
6. Type in two calibration values (two real values and the corresponding output values of the sensor)
7. Click OK
OM-DAQPRO-5300 will send the definition to OM-DAQPRO-5300 and will add the new sensor to the sensor list.
To use the custom sensor:
1. Plug the custom sensor into OM-DAQPRO-5300
2. Set up OM-DAQPRO-5300 using the Setup Wizard (see page 52) and select the sensor from the sensor list next to the corresponding input
3. Proceed with the setup as you would normally do
2.5.10. Communication Setup
Communication between the PC and OM-DAQPRO-5300 takes place automatically whenever you send a command to OM-DAQPRO-5300. However, the Communication Setup dialog can be used for more advanced communication options.
1. Click Logger on the menu bar, and then click Com setup. DaqLab will display a dialog reporting what COM Ports are already in use, and which port is recommended for use with OM-DAQPRO-5300:
Figure 20: Communication setup dialog
2. Click Try to connect and DaqLab will try to communicate with OM- DAQPRO-5300. If communication is achieved, DaqLab will designate the selected port as OM-DAQPRO-5300’s communication port.
You can override the DaqLab recommendation and choose the COM Port yourself by clicking the check box button to the left of the desired port, and clicking Try to connect.
Note: In case of any communication malfunction, please refer to the troubleshooting
guide on page 86.
70 Chapter 2 Working with DaqLab
Page 73
3. You can click Work offline if you wish to work without OM-DAQPRO-
5300. While working offline, you can still open or import saved files to view and analyze them
Chapter 2 Working with DaqLab 71
Page 74
2.6. Analyze the data
2.6.1. Reading Data Point Coordinates
Position the cursor (see page 42) on a point to display its coordinates on the information bar at the bottom of the graph window.
If more than one curve is displayed, and you want to read the Y coordinates of all of the curves simultaneously, bring up digital meters (see page 48) for each of the curves. When you position a cursor on one curve, the meters will show the corresponding Y-coordinates for the other curves.
2.6.2. Reading the Difference Between two Coordinate Values
Position one cursor on the first point and a second cursor on the second point to display the difference between the two coordinate values on the information bar at the bottom of the graph window
2.6.3. Working with the Analysis Tools
The analysis tools can only be applied to data sets that are displayed in the graph window.
1. Use the cursors (see page 42) to select the graph and the data range to which you want to apply the analysis.
2. Select the analysis function you wish to use.
The analysis function will be added onto the graph, with the exception of the smoothing (averaging) function, which will replace the original data set.
2.6.4. Smoothing
The smoothing tool is very useful in reducing random "noises", especially if you want to apply any analysis functions to the data. The smoothing process replaces every data point with the average of its neighboring points.
1. Use the cursor to select the function on the graph that you want to smooth
2. Click More smoothing
3. You can repeat the procedure to further smooth the data
72 Chapter 2 Working with DaqLab
on the graph toolbar
Page 75
+
=
4. Click Less smoothing to reduce the amount of smoothing
2.6.5. Statistics
Use the statistics tool to display statistics of a selected data set or a range of data.
The statistics include:
Average – The average of all the numbers in the range
StDev. – The standard deviation
Minimum – The smallest value in the range
Maximum – The largest value in the range
Sum – Adds all the numbers in the range
Area – The area between the graph and the x-axis in the range
Samples – the number of data points in the range
Rate – the recording rate
To display statistics:
1. Use the cursors to select the graph and the data range to which you want to apply the statistics
2. Click Analysis on the menu bar, then click Statistics
DaqLab will open a statistics window and will display the results in it
2.6.6. Most Common Analysis Functions
1. Linear fit
Click Linear Fit
and to display the line’s equation
Chapter 2 Working with DaqLab 73
on the main toolbar to draw a line of linear least square fit
baxy
Page 76
=
Note: If you want the automatic curve fit equation to start at 0
before applying the linear fit (see page 82).
2. Derivative
Click Derivative consecutive points on the source graph
3. Integral
Choose Integral from the Analysis menu to construct a graph in which each point is the integral of all the preceding points on the source graph
on the main toolbar to construct a graph in which each point is the slope of the 3
t , apply the crop tool
2.6.7. The Analysis Wizard
1. Using the Analysis Wizard
The analysis wizard will guide you through the various analysis functions available in the DaqLab program. The analysis functions are grouped into three main categories: curve fit (regression), averaging, and mathematical and trigonometric functions.
To apply an analysis function to a data set:
1. Use the cursors to select the graph or data range to which you would like to apply the analysis
2. Click Analysis Wizard on the main toolbar
3. Click a category tab:
a. Curve fit
b. Averaging
c. Functions
74 Chapter 2 Working with DaqLab
Page 77
2. Curve fit
Figure 21: Analysis wizard – curve fit
1. Select a fit Type by clicking its icon (If you choose polynomial fit, select the order you want)
2. If you use the cursor to select a data set, it will be highlighted in the Based on data set list, but you have the option of selecting a different data set
3. Type a name in the Name box (optional - the default name includes both the function and the data set names)
4. Click the Show equation on graph check box to display the line’s equation on the information bar
5. Click the Show R
2
on graph check box to display the correlation
coefficient on the information bar
6. Click OK
Chapter 2 Working with DaqLab 75
Page 78
=
Note: If you want the automatic curve fit equation to start at 0
before any type of curve fitting (see page 82).
3. Averaging
Figure 22: Analysis wizard – averaging
1. If you use the cursor to select a data set, it will be highlighted in the Based on data set list, but you have the option of selecting a different data set
t , apply the crop tool
2. Select an averaging option (if you choose custom averaging, choose the number of averaging points in the Points box)
3. Enter a name in the Name box (optional - the default name include both the function and the data set names)
4. Select an Open option
5. Click OK
76 Chapter 2 Working with DaqLab
Page 79
4. Functions
Figure 23: Analysis wizard – functions
1. Select a function from the Functions list. DaqLab displays the function’s formula at the bottom of the dialog
2. If you use the cursor to select a data set, it will be highlighted in the G1 drop list, but you have the option of selecting a different data set
3. If the analysis function involves two data sets, select the second data set from the G2 drop list
4. Enter a constant in each of the C1 and C2 boxes (optional)
5. Enter a name in the Name box (optional – the default name includes both the function’s formula and the data set name)
6. Enter a unit (optional)
7. Select an Open option
Chapter 2 Working with DaqLab 77
Page 80
+
=
x
=
=
8. Click OK
2.6.8. Available Analysis Tools
This section includes a brief description of each of DaqLab’s analysis functions
1. Curve fit
a
n
bx
aey =
baxy
−L1
nn
10
axaxay +++=
n
Linear
Polynomial
Power
Exponential
DaqLab displays the curve fit equation and the correlation coefficient (R the graph
2. Averaging
Draws a line of linear least square fit
Draws a line of polynomial least square fit (you must select an order between 1 and
6.)
y =
Draws a line of power least square fit
Draws a line of exponential least square fit
2
) on the information bar below
The average function replaces every point with the average of n neighboring points from both sides of the point
Low averaging:
78 Chapter 2 Working with DaqLab
5=n , Medium averaging: 11
n , High averaging: 41
n
Page 81
=
=
=
=
3. Functions
In the formulas below, G can enter. The constant’s default value is 1
Absolute
Add
Arccosine
Arcsine
Average
and G2 represent selected data sets, and C1 and C2 are constants that you
1
GCCy =
121
Draws a line of the absolute values of a data set
GCGCy +
2211
Draws a line of the addition of two data sets
GCCy
Draws a line of the arccosine values of a data set (in radians). Arccosine is the angle
whose cosine is
Draws a line of the arcsine values of a data set in radians. Arcsine is the angle
whose sine is
·{average of C2 points around G1}
C
1
GC . The argument 12GC must be between –1 and 1.
12
GCCy
121
GC . The argument 12GC must be between –1 and 1.
12
)arccos(
121
)arcsin(
Cosine
Chapter 2 Working with DaqLab 79
Replaces every point with the averaging of C
)cos(
GCCy
121
Draws a line of the cosine values of a data set. The argument
expressed in radians.
neighboring points.
2
GC must be
12
Page 82
−
=
Delta Y
tGGy
11
)0(
=
Derivative
Divide
Envelope (lower)
Envelope (upper)
Draws a line of the difference between the Y-coordinate of every point and Y­coordinate of the first point. Use this function to move the data set along the Y-axis so that the point will intersect the Y- axis at the origin.
yy
−
nn
11
−+
y
=
n
Draws a line of the slopes of every three consecutive points of a data set. For high recording rates and small ∆t, this line may be very noisy, which is why smoothing the data set is recommended before applying the derivative function.
y =
Draws a line of the division of two data sets
Lower envelope of G
Draws a line that connects the minimum values of a data set. The tolerance defines the minimum distance (in sampling points) between two minima, so that the envelope function is able to ignore random "noises".
Upper envelope of G
t
∆
2
GC
11
GC
t
=∆
,
22
with tolerance of C1 points
1
with tolerance of C1 points
1
1
ratesampling
80 Chapter 2 Working with DaqLab
Draws a line that connects the maximum values of a data set. The tolerance defines the minimum distance (in sampling points) between two maxima, so that the envelope function will be able to ignore random "noises".
Page 83
+
=
+
=
=
=
Exp.
eCy =
1
GC
12
Fourier transform
Frequency
Integral
Linear
Ln
Draws a line of
Discrete Fourier transform of G
Draws a line of the amplitudes of the harmonics of Fourier transform vs. frequency.
The frequency of G1 (minimum of C
Draws a line of the frequency of a periodic data set vs. time. The constant C the minimum data points in one cycle.
Draws a line in which each point is the discrete integral of all the preceding points in a data set.
Draws a line of a linear displacement of a data set. This function is useful when you want to change the point of origin of a data set.
e raised to the power of a data set.
.
1
points in one cycle).
1
tGCCy ∆∑
121
GCCy
121
)ln(
GCCy
121
defines
1
Log
Chapter 2 Working with DaqLab 81
Draws a line of the natural logarithm of a data set. The argument
positive.
)(log
GCCy
12101
Draws a line of the logarithm of a data set to base 10. The argument
positive.
GC must be
12
GC must be
12
Page 84
=
=
=
=
Multiply
Reciprocal (1/X)
Sine
Square (X
Square root
Subtract
2
GCGCy ⋅
2211
Draws a line of the multiplication of two data sets
C
1
G
1
)sin(
GCCy
121
GC must be
12
2
)( GCCy =
121
GCCy =
121
GC must be
12
GCGCy −
2211
y =
Draws a line of the reciprocal values of a data set.
Draws a line of the sine values of a data set. The argument
expressed in radians.
)
Draws a line of the squares of a data set.
Draws a line of the square root values of a data set. : The argument
greater than or equal to zero.
Tan
82 Chapter 2 Working with DaqLab
Draws a line the subtraction of two data sets.
)tan(
GCCy
121
Draws a line of the tangent values of a data set. The argument
expressed in radians.
GC must be
12
Page 85
=
=
2.7. Special Tools
2.7.1. Crop Tool
The Crop tool enables you to trim the edges of a data set. Use it to remove unwanted data.
The time scale of the trimmed data is shifted so that it will start at
0
If you want the automatic curve fit equation to start at curve fitting
After applying the crop tool, the trimmed data set replaces the original set on the graph display and a new icon is added to the Data Map under cropped data.
1. To trim all data up to a point
1. Position a cursor (see page 42) on the data point
2. Click on the Graph menu, and then click Crop.
2. To trim all data outside a selected range
t , apply the crop tool before any type of
0
t .
1. Use the cursors to select the range you want to keep
2. Click on the Graph menu, and then click Crop.
Chapter 2 Working with DaqLab 83
Page 86
2.8. Toolbar Buttons
2.8.1. Main (upper) Toolbar
New
Open
Save
Print
Run
Stop
Download
Setup
Start new project
Open saved project
Save project
Select print options
Start a data recording session. Click on the down-arrow to select recording mode
Stop recording
Download the data from the most recent recording session
Open the Setup dialog
Meter setup
Data map
84 Chapter 2 Working with DaqLab
Open the Meter setup dialog
Display the Data Map
Page 87
Graph
Table
Analysis
Linear fit
Derivative
Display alarm
Display the graph window
Display the table window
Open the Analysis wizard
Draw a line of linear fit of the selected data set
Draw a line of the derivative of the selected curve
Displays or removes alarm levels from the graph
2.8.2. Graph Toolbar
Zoom in
Zoom in around a selected point or in to a selected range
Zoom out
Autoscale
Zoom to selection
Pan
More smoothing
Chapter 2 Working with DaqLab 85
Reverse the most recent zoom operation
Display all the data
Zoom in to a selected area
Pan in all directions while in zoom mode
Smooth (average) the selected curve
Page 88
Less smoothing
1st Cursor
nd
Cursor
2
Split graph
Edit graph
Graph properties
Add to project
Export to Excel
Reverse the most recent smoothing operation
Display or remove the first cursor
Display or remove the second cursor
Switch to a split graph display
Select the data to display on the axes
Graph formatting, scaling and units selection
Add the displayed graph to the project
Export the displayed graph to Excel
86 Chapter 2 Working with DaqLab
Page 89
Chapter 3 Troubleshooting Guide
Symptom Cause of problem
I pressed the ON key on OM-DAQPRO­5300 and it did not turn on.
DaqLab was unable to connect to OM­DAQPRO-5300.
OM-DAQPRO-5300 and DaqLab displays a warning message: “Illegal sensor”
I downloaded “unexpected” data.
While running with an active trigger, OM­DAQPRO-5300 does not complete the data logging.
• Try charging the battery, it may be weak.
• You are using an unsuitable AC/DC adaptor (see
page 15).
• The internal 3V disk battery is dead.
• OM-DAQPRO-5300 is turned off.
• The communication cable is not connected
properly.
• The communication cable is not connected to the
correct PC COM Port.
• There is no available Serial COM Port.
• OM-DAQPRO-5300 measured a voltage drop of
over 5V across an input that was selected as one of the following types: 50mV, TCK, TCJ, TCT, PT 100 or NTC.
• OM-DAQPRO-5300 measured a reversed voltage
drop of over -1V across an input that was selected as one of the following types: 50mV, TCK, TCJ, TCT or 10V.
• There may be up to 100 files stored in the OM-
DAQPRO-5300. Performing a “Download” means loading one of them in sequential order. You have probably loaded an old file accidentally. Try using Selective download from DaqLab (see page 36)
• The trigger condition has not been fulfilled.
• The trigger condition was fulfilled, but the time
period was too short (less than 300µS).
Chapter 3 Troubleshooting 87
Page 90
Symptom Cause of problem
I started a new logging and the data logger stopped immediately afterwards.
The recorded data is “noisy”
OM-DAQPRO-5300’s memory is empty – there are no stored files or notes.
• First check if the sampling rate was so fast that the
recording began and ended in less than a second. If this is not the reason, recharge the 7.2V battery and try again. OM-DAQPRO-5300 monitors the battery power during the logging. If the battery voltage is less then 5V, the logging is stopped automatically.
• With voltage or current sensors, use short
connections between the circuit being tested and the sensor.
• Avoid working near strong electromagnetic fields.
(e.g. engines, fluorescent lights)
• The internal 3V disk battery is dead.
88 Chapter 3 Troubleshooting
Page 91
Chapter 4 Specifications
OM-DAQPRO-5300
Inputs
• 8 channel differential inputs
• Selectable type for each
input: 0-24mA, 0-50mV, 0-10V, NTC, PT-100, Thermocouple, Pulse counter (Input 1 only), frequency (Input 1 only) and user defined sensors
Input types
0 – 24mA
Range 0 to 24mA Resolution 0.47µA Accuracy ±0.5% Loop impedance 21Ω Maximum load 30mA, 5.2V
0 – 50mV
Range 0 to 50mV Resolution 3µV Accuracy ±0.5% Input impedance 50MΩ Maximum voltage 5.2V
0 – 10V
Range 0 to 10V Resolution 200µV Accuracy ±0.5% Input impedance 125KΩ Maximum voltage 25V
Temperature NTC-1
NTC – 10KΩ resistor Range -25 to 150ºC Resolution 0.05ºC Accuracy ±0.5%
Temperature NTC-2
NTC – 2252Ω resistor Range -25 to 150ºC Resolution 0.05ºC Accuracy ±0.5%
Temperature PT100
Range -200ºC to 400ºC Resolution 0.1ºC (7mΩ) Accuracy -200 – -50 ºC ±0.5% 50 – 400 ºC ±0.5%
-50 – 50 ºC ±0.5 ºC Possible to connect up to 8 PT100 2 wires or 4 PT100 3 wires
Temperature thermocouple J
Range -200ºC to 1,200ºC Resolution 0.1ºC (1µV) Accuracy -200 – -50 ºC ±0.5% 50 – 1200 ºC ±0.5%
-50 – 50 ºC ±0.5 ºC Cold junction compensation ±0.3ºC
Temperature thermocouple K
Range -250 to 1200ºC Resolution 0.1ºC (1µV) Accuracy -250 – -50 ºC ±0.5% 50 – 1200 ºC ±0.5%
-50 – 50 ºC ±0.5 ºC Cold junction compensation ±0.3ºC
Temperature thermocouple T
Range -200 to 400ºC Resolution 0.1ºC (1µV) Accuracy -200 – -50 ºC ±0.5% 50 – 400 ºC ±0.5%
-50 – 50 ºC ±0.5 ºC Cold junction compensation ±0.3ºC
Chapter 4 Specifications 89
Page 92
Pulse counter (input 1 only)
Zero crossing detector Opto coupler interface Range 0 to 65,000 counts Resolution 1 count Frequency range 0 to 25Hz Input signal 0 – 5V Input impedance 470Ω
Frequency (input 1 only)
Zero crossing detector Opto coupler interface Range 20 to 4,000Hz Time resolution 6.5µs Input signal 0 – 5V Input impedance 470Ω
Internal Temperature
NTC – 10KΩ resistor Range -25 to 70ºC Resolution 0.1ºC (1µV) Accuracy ±0.3°C
General A to D specifications
Noise 30µV rms Internal linearity error ±0.08% of FSR Offset error 0.1%
Alarm output (output 8)
Open collector Close position resistance 50Ω Max. Load 50mA, 5V DC Overload protection 50mA re-settable fuse
Communication
• USB 1.1 compliant
Sampling
• Capacity: 512KB
• Analog sampling rate:
Variable, 1 sample per hour to 4,000 samples per second, 1 channel
• Analog sampling resolution: 16 bits
• Channel separation 80dB
Man Machine Interface
• Full keyboard operation - enables manual programming of the Logger
• Graphic LCD 64x128 pixels
Power Supply
• Internal rechargeable 7.2V Ni-MH battery
• Built in battery charger
• External 9V to 12V DC input
• Battery life: 40 hours
between charges
Operating Temperature Range
• 0 to 50°C
Casing
• Plastic ABS box
• Dimensions:
182×100×28 mm
• Weight: 450gr
Standards Compliance
• CE, FCC
DaqLab software
Main features
• Graphical, analysis WINDOWS based software for the OM-DAQPRO-5300
• Fast data download from the OM-DAQPRO-5300
• Data displayed in numeric or graphical display forms Graphical display of all inputs
90 Chapter 4 Specifications
Page 93
• On-line retrieval and display of data in real-time
• Setup wizard for the OM-DAQPRO-5300
• Define of up to 20 new sensor in the OM-DAQPRO-5300
Read the defined sensor’s units on the logger’s display
• Full calibration of the OM-DAQPRO-5300 via software
• Documentation and filing
• Alarm levels on graphs
• Export and import to and from spreadsheets
• More than 30 analytical function, for professional analysis of the collected data
System requirements
Software
Windows 98 or later
Internet Explorer 5.0 or later (you can install Internet Explorer 5 when you install
DaqLab, since it ships with the product)
Hardware
Pentium 200MHz or higher
32 MB RAM (64 MB recommended)
10 MB available disk space for the DaqLab application (50 MB to install the supporting
applications
Chapter 4 Specifications 91
Page 94
Appendix: Figures
Figure 1: OM-DAQPRO-5300 external connections................................................................................ 10
Figure 2: OM-DAQPRO-5300’s inputs block terminal ............................................................................. 12
Figure 3: connecting 3 wire PT 100.........................................................................................................13
Figure 4: connecting external alarm device ............................................................................................. 14
Figure 5: OM-DAQPRO-5300 front panel................................................................................................ 16
Figure 6: Frequency Aliasing................................................................................................................... 29
Figure 7: DaqLab window layout .............................................................................................................34
Figure 8: Edit graph dialog ......................................................................................................................45
Figure 9: Setup wizard step 1 of 3........................................................................................................... 52
Figure 10: Setup wizard step 2 of 3......................................................................................................... 53
Figure 11: Setup wizard step 3 of 3......................................................................................................... 54
Figure 12: alarm settings.........................................................................................................................55
Figure 13: X-axis properties dialog..........................................................................................................57
Figure 14: Triggering dialog.....................................................................................................................58
Figure 15: Edit notes dialog..................................................................................................................... 60
Figure 16: Two-point calibration dialog.................................................................................................... 63
Figure 17: Offset calibration dialog..........................................................................................................65
Figure 18: Zero calibration dialog ............................................................................................................ 67
Figure 19: Define new sensor dialog ....................................................................................................... 68
Figure 20: Communication setup dialog ..................................................................................................69
Figure 21: Analysis wizard – curve fit ...................................................................................................... 74
Figure 22: Analysis wizard – averaging ................................................................................................... 75
Figure 23: Analysis wizard – functions.....................................................................................................76
92 Appendix
Page 95
Appendix: Simplified measurement circuits
0 to 10V 0 to 24mA
100K
80R
21R
100K
80R
100R
ADC
VREF
1UF
0-10V
100K
80R
100K
80R
25K
100R
ADC
VREF
1UF
0-24mA
2 wire PT 100 3 wire PT 100
PT100
80R
100K
100K
80R
2 Wire PT100
100uA
10k
80R
PT100
ADC
80R
100K
100K
100uA
ADC
100uA
3 Wire PT100
4.7k
Appendix 93
Page 96
NTC Internal NTC
80R
20uA
20uA
NTC 10K
NTC
Thermocouple
100K
Thermocouple
80R
80R
100K
100K
100K
80R
1UF
1UF
100R
VREF
1UF
10k
ADC
ADC
NTC 10K
100K
100K
ADC
10k
Internal NTC
94 Appendix
Page 97
Index
A
AC/DC adaptor · 15 alarm · 45
level · 13 output · 14, 55 set · 13 setup · 54
analysis · 71
averaging · 75 carve fit · 74 derivative · 73 function · 76 integral · 73 smoothing · 71
wizard · 73 autoscale · 43 averaging · 75
B
battery · 15
level · 56 block terminal · 12
C
cable · 9 calibration · 15, 61 channels · 11 chart · See table clear
data · 50
memory · 59
clock
internal · 27 COM port · 69 comment · 15, 59 communication · 69 configuration · 26 connect · 69 connection
3 wire PT 100 · 12
frequency/pulse counter ·
12 polarity · 12 screw terminal · 12 to a computer · 11
continuous · 29, 30 contrast · 27 coordinates · 25, 42 copy · 51 counter · 12 crop · 82 cursor · 25, 42 curve fit · 72, 74
D
data map · 49 define
custom sensor · 68
delayed start · See
triggering
delete · 50 derivative · 73 display
control the display · 49 data map · 49 graph · 41 meters · 48 options · 41
properties · 56 select · 20 table · 47
download · 36, 37
E
edit
graph · 44
notes · 60 enter · 19, 22 escape · 22 event · 14 event recording · 58 Excel · 51 export · 51 external power · 11
F
filter · 26, 55 format
graph · 46
table · 48 frequency · 12 functions · 76
G
getting started · 35 graph · 24, 41
Index 95
Page 98
I
icon · 22 Illegal sensor · 13 import · 38 input · 11 integral · 73
K
keypad · 16
L
layout
DadPRO · 16 DaqLab · 34
load
last setup · 25 notes · 60
low pass filter · 26, 55
M
mains adaptor · 15 manual
sampling · 29
memory · 30
clear · 27, 59 menu · 21 meter · 23, 48
N
noise · 27, 56 notes · 60 number
format · 46, 48 of samples · 19, 22, 54
O
online
communication · 36 open · 37 output · 14, 55
P
panel · 16 panning · 44 Password · 62 polarity · 12 power · 11 preset · 56 print
graph · 39
taple · 40 program
DaqPRO · 52 programming
rules · 30 project · 33, 38 properties
graph · 46
table · 48 PT 100
3 wire · 12 PT 100 · 12 pulse · 12
R
rate · 19, 22, 28, 30, 53, 87 recharge · 15 recording
data · 21, 35 options · 36 rate · See rate start · 59 stop · 59
time · 19, 22, 54 regression · See curve fit requirements · 31 run · 35, 59
S
sampling Points · 29 sampling rate · See rate save · 37 scale · 43 screen contrast · 27 scroll · 53 sensor · 11, 12, 17
assign to input · 52
calibration · 61
custom · 68
plug in · 17 serial number · 15 setup · 18, 22, 25, 28, 52 slope · See derivative smoothing · 71 specifications · 88 spreadsheet · 51 stand alone · 16 start · 20, 22, 35 stop · 21, 22, 35 system configuration · 26
T
table · 23, 47 temperature
units · 27, 60 terminal · 12 timer delay · 58
96 Index
Page 99
toolbar
graph · 84
main · 83 triggering · 57 trim · See crop troubleshooting · 86
U
units
format · 46, 48
temperature · 27
V
value · 25, 42 view · 41
W
wizard
analysis · 73 setup · 52
X
X-axis · 45, 57
Y
Y-axis · 41, 45
Z
zoom · 25, 42
Index 97
Page 100
Loading...