Omega OM88128, OM8829, OM8828, OM8800D User Manual

OM8829,AND INTERFACE
Temperature/Humidity Data Logger
Index
IndexIndex
Introduction...............................................................................
Material supplied.......................................................................
Features....................................................................................
LCD display (Model OM8828 / Om8829).......................................................
Battery replacement..................................................................
Software Installation .................................................................
Main screen & introduction.......................................................
Retrieve File..............................................................................
Save File...................................................................................
Logger setting...........................................................................
A) Sample point setup............................................................................
B) Sample rate........................................................................................
C) Start mode set up...............................................................................
D) Logger's clock setting.........................................................................
E) Alarm setting......................................................................................
F) Unit setting ,Logger ID setting............................................................
G) Sleep Mode setting & Data Table......................................................
Data Table : Edit (Go to ,Find ,Sort ) ,Copy .............................
Statistics....................................................................................
Print Graph................................................................................
Group Files...............................................................................
Previous setting,Temperature Chart,Cursor status...................
Blank logo,Message..................................................................
Download commands................................................................
Display control..........................................................................
General Specification................................................................
F.A.Q. Warranty and Return Authorization................................
1 2
3 4 5 6
7
8-10
11-12
13 14
15 15 16 16
17 18
19
20 21-24 25-28
29 30-31
32 33 34
34-37 38-39
40
Index
INTRODUCTION
Thank you for purchasing this Temperature Data Logger! This unit has been developed to meet your maximum satisfaction with its user-friendly design. Review the entire manual for a complete overview of the operation of this new data logger. You may also review the manual from Autoplay Menu, double click "Run software, Manual (PDF)"and peruse every procedure and function step by step .The logger is very easy-to-use.
The software is used to readout the stored data and see as a graph , showing the history with real-time clock, the tabular data can be viewed or exported to a spreadsheet for various analysis.
One interface can be used with multiple loggers, designed as a saving cost unit. There are 3 color of interfaces for choice: White , Dark Blue , Burgundy .
IDEAL FOR APPLICATION:
Monitoring Ambient condition in Greenhouses, Warehouses, Food transport, Aircraft cabins, refrigerate truck,
containers, railway ....Art galleries and
Museum, incubation process, Hatcheries, HVAC field (Heat, Ventilation, Air Conditioning), collecting data
for QC.....
1
Before starting the logger setting ,please read throughly the whole operation manual , there are some FAQ listed in some pages for troubleshooting reference.
MATERIAL SUPPLIED
Check for damaged or missing parts in your data logger before
starting:
The Data Logger Set should contain:
1. Either one of the following :
Model: OM8800D ,OM88128 or OM8828 or OM8829 The Data Logger
(supplied with one CR2 or ER3, 3.0V lithium battery).
Model: OM8800R(White) ,OM8828P(Dark blue), OM8829P (Burgundy)
The Interface cable & software.
Model:OM 8818 ,OM88128S or OM8828S or OM8829S
The complete set including data logger with lithium battery ,
interface cable and software.
2. Operation manual
R
E G G
O
L % H
9 R .
2 T
8 8
OM88128/8828/8829
H/M/S
C E R
88888
o
C
5
M
8
%
L
~ 0
A
100
-4
:
0~
.
p
: m e T
H R
OM8818
OM8800D
%
1
R
E G G
O
L % H
R . T
9
2 8 8
8
8
C
E
8
R
88
M
o
L
C
A
5 8 ~
% 0
40
0
­1
: .
0~
mp
:
e
T
H
R
%
OM8800R
2
OM8800R/8828P/8829P
2
3
OM88128S/8828S/8829S
FEATURES
Up to 8000 sample readings (OM8800D ) with real-time clock
memory.
Up to 16,000 sample readings (OM88128 / 8828 / 8829) with
real-time clock memory.
LCD display of data ( Model # OM8828/8829 ).
Programmable sample interval from 1 second to 2 hours.
Temperature unit: C / F selectable.
User-defined maximum and minimum alarm.
Built-in real-time clock.
Schedule start / stop time setting or Magnetic start setting.
Programming with RS-232 interface.
Data retained when battery is low or has been removed.
Waterproof housing design.(Model OM8800D / 88128 / 8828).
Water resistant design ( Model OM8829).
Hanger at the rear upper for hanging on the wall to prevent
losing or falling down .
One Interface cable and software can be used with several
loggers to save cost.
Download data to other program for further analysis.
All data is saved in text format and can be easily transformed
into Microsoft Office applications.
Easy to use Software provides :Retrieve file , Save file , Logger
setting , Comport setting , Group file , Print graph , Data table
list , Statistic .......
Sleep mode and Non-Sleep mode to be selectable to monitor
current condition of Temperature or Humidity (LCD models)
oo
R
3
FRONT VIEW
REC
- Stands for RECORD
For model without LCD : OM8800D or OM88128 "REC" is flashing while logging the temperature records
"REC" LED will stop flashing after complete recorded or battery is out of power . REC green light will flash every second ( OM8800D ) . REC green light will flash every 5 seconds when sample interval is set over 5 seconds (OM88128) . For model with LCD : OM8828 or OM8829 Both "REC" green light and "REC" indicator will active and display on the screen per every set sample interval . For example , if you set sample interval less than 5 seconds , the REC light will flash per set second , otherwise, will flash every 5 seconds .
ALM
- Stands for ALARM
"ALM" is flashing while recorded value is higher than set HI or less than LO setting , the logger alarm designed is not audible , but for reminding or warning the user to do necessary action .
For model without LCD : OM8800D or OM88128 "ALM" "ALM" LED will stop flashing after downloading to a PC
or battery is out of power . ALM red light will flash every second ( OM8800D ) . ALM red light will flash every 5 seconds when sample interval is set over 5 seconds (OM88128) . For model with LCD : OM8828 or OM8829 Both "ALM" red light and "ALM" indicator will active and display on the screen per every set sample interval . For example , if you set sample interval less than 5 seconds , the ALM light will flash per set second , otherwise, will flash every 5 seconds .
4
LCD DISPLAY
Model: OM8828 / OM8829
RECHILOW
COMM
RH%
o
C
o
F
REC - "REC" is displayed while logging the temperature or humidity. REC will not appear when in the Non-sleep mode. When sample interval set over 5 seconds , REC LED will every 5 seconds , set less than 5 seconds , REC green LED will flash per sample interval , such as 1,2,3, or 4.
HI - "HI" is displayed and alarm LED is flashing if a temperature/or humidity is higher than set High value in the logger unit . LOW - "LOW" is displayed while logged temperature is lower than set Low value in the logger unit . RH% - Display Relative Humidity %.(Though you might have seen RH% when replacing with a new battery , which is not available for the model 8828 but model 8829 Temperature/ Humidity datalogger .)
o
C - Temperature displays reading in Celsius
o
F - Temperature displays reading in Fahrenheit
COMM - When communicating with computer ,"COMM" is shown
at the top of the screen.
5
BATTERY REPLACEMENT
The dataloggers are designed with a waterproof housing, and allow
the user to replace battery when it has expired .You will see "Lo"
appeared on the display when battery is weak for model OM8828 and
OM8829..
Follow the steps to remove and replace battery:
1. Unscrew the datalogger from the rear side .
2. Do not remove o-ring but make sure the o-ring is in its place (groove).
3. Remove the expired battery
4. Insert a new battery CR3 or ER2, make sure the battery is inserted with correct polarity.
5. Re-screw the rear cover with screwdriver .
Model: OM8818
Model: OM88128 OM8828 OM8829
6
+CR2/ER3-
+CR2/ER3-
SOFTWARE INSTALLATION
Microsoft Office is a registered
trademark of Microsoft Corporation.
Installation procedure :
1. Slide the logger unit onto the RS232 interface.
2. Connect the socket of RS232 to the COM1 or COM2 .
3. Insert CD-ROM to the computer for starting software set up:
4. See the diagram as an example of 8818 Temp. logger :
a).Run Software. To run software from CD-Rom.(Or read manual) b).Install Software. Run the set up installation program. c).Brow CD-ROM. Browse the CD-Rom using Windows. d).Web Site. Visit our web site in your browser. e).Exit. Close this menu system.
OM8818
OM88128
R
OM8828
OM8829
7
MAIN SCREEN
The data logger program is easy to operate from its main screen.
Retrieve File
Previous
Setting
Temperature
Chart
Cursor
Status
Com. Set
Save
File
[MM-DD-YY hh:mm:ss
<<<WARNING:DataLOGGER ISNOT plugin interface modele!>>>
TEMPERATURE CHART
# 0T=0.0Date (M-D-Y)=01-01-2000-Time =00:00:00
# 0T=0.0Date (M-D-Y)=01-01-2000-Time =00:00:00
Message
Data
Table
Logger
Set
Cursor 1
(red)
Record No:
Record Number
Time
Statistics
Commands
Print
Graph
Group
Files
Cursor 2
(green)
Time
Download
Maximize
Minimize
MM-DD-YY hh:mm:ss]
Diagram A: Main Screen
Keyboard Actions for Cursor
Make sure you have selected a cursor , when you click the red vertical line , it turns bold red line , the same for green line. You will see the numbers at the left lower part changed while pressing arrow left or down arrow key for last record or pressing right or up arrow key for next record.
8
Close
Program
Logo
100.0
90.0
80.0
70.0
60.0 %Rh
Humidity
50.0
Chart
40.0
30.0
20.0
10.0
0.1
Display Control
Left arrow key To the previous point on the current plot. Right arrow key To the next point on the current plot. <Shift>-left arrow key Back 10 points on the current plot. <Shift>-right arrow key Forward 10 points on the current plot. <Home> To the first Visible point on the current plot. <End> To the last visible point on the current plot.
FAQ: Why can't I see the movement by pressing above keys ? Ans : Make sure the bottom right corner "Display Control ­ Offset " figures are not in black , if it is yellow button , please click other area to restore to the original black .
Drag a cursor to move it.
The cursor tracks the mouse until you release the mouse button, and then the cursor snaps to the nearest data point.
Actions by Offset Knob
You can operate on graph from the "Offset Knob" after pressed "Offset Knob" in the following ways :
Press the up or down arrow key to increase or decrease one record number (or corresponding to date ) in the graph .
Zooming and Panning on Graphs
To start zooming in on a point, press the <Ctrl> key and click on the left mouse button over the point; you can release the <Ctrl> key after you press the mouse button. The resolution in the viewport is increases dynamically until you release the mouse.
To zoom out, click on the right mouse button, and then press the <Ctrl> key as you do to zoom in.
9
To pan, press the <Ctrl-Shift> keys and click on the left mouse button over a point on the viewport. Then drag the mouse to another point. The graph viewport scrolls so that the original point now appears under the new mouse cursor location. You can drag the mouse anywhere on the screen, even beyond the viewport.
To restore the viewport to its original state after zooming or panning . Press and hold the left mouse to drag the yellow line from " Offset knob " and "Range knob" .
Data record number or Time display selection
Press the switch button and select between Record number and Time ( xx : xx : xx ) the hour , minute , second . The record number and time will display in turn while pressing the switch button .
Warning message
The left bottom corner will show a warning message while you just retrieve an existing file without downloading a new logger.
<<< WARNING : Data Logger is not plug in interface module !>>>
Display Control
After your downloading or retrieving a record , you will see the number on the knobs of "Offset" and "Range" be changed to Offset: 0000/5000/10000/15000 and Range:50/250/500/1K/2K/4K/8K/12K/ 16K. These are different from main screen without any records displayed.
10
RETRIEVE FILE
Click on icon to retrieve and to load a data file into this program. (See Diagram B.) This program is designed to log up to 8,000 / 16,000 sample readings.(Model OM88128/ OM8828/ OM8829 are available for 16,000 sample collecting )
Directory History:
Look in:
(Example)
File name:
File of type:
Cancel
NOTE: Above file names are the examples to help you understanding what screen will be .Don't be alarmed if you couldn't see the
files shows on your software .
Open data file "RoomByF1125.txt " for test file. See Diagram C for an example of data in file :
To aid in reading a graph , a grid can be drawn on the graph, and a dialogue box display at the left bottom corner shows the appearance of this grid.(CONTROL LOCATION-Cursor 1 and Cursor 2).
Diagram B: Retrieve File
¡ö
11
After you downloading from model OM8829 or retrieving a file with
%Rh
Temperature
Humidity
Humidity records, you will find " Temperature " and " Humidity " indicated on the start point of curves . (See Dia.B-1 ) Otherwise, you will only one curve indicated Temperature .
Temperature
Diagram B-1: Retrieve File
Humidity
For model OM8818,OM88128,OM8828,you will not see RH% on the right side of the screen but only model OM8829 .(See Dia.B-2 )
Meantime, you will see the date and time of the first point at the left corner , and see the last point date and time at the right corner . (See Dia. B-1 and Dia.B-2 in blue )
Diagram B-2: Retrieve File
12
Diagram C: Main Screen with Sample Data
PREVIOUS SETTING:
79.0
78.8
78.6
78.4
78.2
78.0
77.8
77.6
77.4
77.2
77.0
[MM-DD-YY hh:mm:ss
CURSOR LOCATION
Retrieve file completed.
TEMPERATURE CHART
Record No:
(Example)
Time
2500
0000
MM-DD-YY hh:mm:ss]
5000
250
50
7500
SAVE FILE
Selecting icon, select folder and name to save the data to be saved. The windows " Save File " dialog box allows you to specify the file format to save the data,the file name to be called , where the file to be saved to .
13
Directory History:
Look in:
(Example)
NOTE: Above file names are the examples to help you understanding what screen will be .Don't be alarmed if you couldn't see the
files shows on your software .
File name:
File of type:
COM PORT. SETUP
Select icon for "Com. Port Setting". (See Diagram D). Set correct COM port, Baud rate, Data bits, Parity and Stop bits. Usually, it's COM1 for most notebook and desktop computers. Select up to COM 4 for special systems. Select OK to accept setting changes, Cancel to abort and to exit Setting menu.
Note:
Always remember to select 4800 baud rate for model OM8818 , or select 9600 baud rate for model OM88128/OM8828/OM8829 .
Cancel
Diagram D:
COM Port Setup
14
LOGGER SETTING
Depress icon to perform Next Logger Setting. (Diagram E.)
Diagram E: Logger Setting
Sample point Setup
Sleep Mode Setting
Non-Sleep Sleep
Logger's ID Setting
Identifier
SAMPLE POINTS SETTING -
Select sample point 1,000 / 2,000 / 4,000/ 8,000 or 12,000 / 16,000 (Note: Only OM88128, OM8828 & OM8829 are available for selecting sample point : 12,000 / 16,000 ).
¡ö
15
By scrolling up or down and then click OK to save the selection. (See Dia-E )For example ,select 1,000 sample data,the logger will stop collecting data and power off for saving battery when 1000 data collected completely . Recording inactivated when selected data is full , and yellow LED light will not blink to confirm operation is stopped .
FAQ: If I want to stop logging before completing all sample points , how should I do ?
Ans: Download with computer to stop logging. Note: The download operation before completing all sample points won't work for the logger sample rate to be set as "1" second .
SAMPLE RATE : Select the sampling time ranges from 1 ~ 2
seconds. Scroll up for increasing time seconds or down for decreasing time in seconds. The datalogger defaults sample rate (the length of time between each sample) as "2" though "1" is the minimum sample rate for selecting .
0
7,
START MODES : (Diagram F):
Schedule -
Set up Start Date/Start Time desired to start logging. The logger will start when the setting time clock reaches the time you entered. Make sure the logger date and time are the time you are now . Magnetic - Link with your computer to setup the Magnetic start mode , select Magnetic , make sure the logger date and time are the time you are now , press "OK" , enter desired ID , press"OK" To confirm the Magnetic start setting . Immediate - While pressing "OK" Button , the logger starts recording immediately.
16
0
To start with magnetic , use a strong magnetism one pass over the middle of the bottom , now LED is flashing .Flashing times is depend on the time interval of every sample logging.
Diagram F: Start Mode Setup
Sleep Mode Setting
Non-Sleep Sleep
Logger's ID Setting
Identifier
-40
¡ö
LOGGER'S CLOCK SETTING
Enter your Logger's clock setting allows real-time setting for this data logger unit.See Diagram F.
17
ALARM SETTING
Setting up the High Alarm and the Low Alarm enables to activate
the red LED flash of the Data Logger. (See Dia. G)
-High Alarm temperature defaults at 85 C.-Low Alarm temperature defaults at -40 C.
-High Alarm Relative Humidity defaults at 100% .
-Low Alarm Relative Humidity defaults at 0% .
User simply select desired High value or Low value by scrolling up for increasing or down for decreasing .
NOTE: Hi/Lo humidity setting is only available for model OM8829. Select " oF ( w/ RH% alarm) or oC (w/RH% alarm) for Humidity alarm setting in degree F or degree C . (See Dia.F-1 )
o
o
Diagram F-1: Alarm Setting
ALM LED Flashing activation for model OM8800D . If a Temperature is higher than HI setting or less than LO setting , ALM LED will flash every second until you download to a PC .
For models OM88128, OM8828 ,OM8829 , if a temperature is higher than HI setting or less than LO setting , ALM LED will flash per every sample interval , for example , if set to 3 seconds for an interval , ALM and REC LED will flash every 3 seconds until you download to a PC .
18
UNIT SETTING
Data logger is available in " C" and " F" (See Diagram G) as well as Humidity unit " F (w / Rh% alarm) or C (w / Rh% alarm) "which is
oo
o
only available for model OM8829.
Diagram G:
o
Diagram G: Unit setting
LOGGER ID SETTING
Logger's ID Setting enables user to give unique ID names to each
data logger (up to 8-digits), simplifying data tracking.
Diagram G:
19
SLEEP MODE SETTING
Model OM8828/OM8829 logger (LCD display) will stop displaying on the LCD after completed recording, i.e. The figures will be disappeared (This should be set up from computer before recording ). Easy to click and press the slide switch to the right for SLEEP then click OK . You may select NON-SLEEP and press OK , then the logger will always display current Temperature / Relative Humidity after completed recording , but please be noted that this setting will consume the battery power . (The software default Non-Sleep mode)
Sleep Mode Setting
SleepNon-Sleep
Ok
Diagram G-1: Sleep Mode setting
DATA TABLE
Depress icon to view detailed data table,which includes 200 sample readings.Temperature data shows the curser 1,If you are logging model OM8829 (Temp. & Relative Humidity) , RH% reading shows next to the temperature reading .
Print out the data table to either a printer or a fax /internet application depending on your available facilities.(See Diagram H )
20
FAQ: Why won't data table be updated when removing the cursor ?
Ans: It is when you remove the cursor between the ranges ( 200 sample from cursor 1).For example : If you set number 50 record as cursor 1, the table data shows from 50 to 249 , now you remove cursor 1 within the range, the data won't be updated, unless remove the curser 1 outside the range, i.e less than 50 or more than 249 .
Edit Cell
6-29-2001
123 124
6-29-2001
125
6-29-2001
6-29-2001
126
150
6-29-2001
151
6-29-2001
152
6-29-2001
6-29-2001
153 154
6-29-2001
6-29-2001
155
6-29-2001
156
6-29-2001
157
(Example)
Diagram H:
8:28:20
8:28:21
8:28:22
8:28:23
8:28:24
8:28:52
8:28:53
8:28:54
8:28:55
8:28:58
8:28:56
8:28:57
9.3
9.3
9.3
9.3
9.3
9.3
9.3
9.3
9.3
9.3
9.3
9.3
9.3
EXIT
29.7
29.7
29.7
29.7
29.7
29.8
29.8
29.8
29.8
29.8
29.8
29.7
29.7
You switch from selection to edit state by double-Clicking on the cell. If the active cell (in other words, the currently selected cell) can be edited, pressing <F2> also enables the edit state.(Dia. H) edit from the blue background column.
FAQ: How can I restore the original text ? Ans: You have to retrieve the file again to restore to the original text .
21
Data table pop-Up Menu
If you click on the right mouse button anywhere over the table control you activate a menu which contains Goto, Find, and Sort menu items. Goto You can specify a target cell to go to, using its row and column indices. There are 4 columns for model OM8828 : No.-Date-Time-Temp. There are 5 columns for model OM8829 : No.-Date-Time-Temp. And Humidity(%) A total of 200 rows from cursor 1 to cursor 2 . Key in which row and which column you want to go , then press "OK" , the cursor will go to the destination .
Find You can search all numeric and text cells in the entire table, or in a selection range, for a text string. After the initial search, you can press <F3> to find the next occurrence of the same text string, or you can press <Shift-F3> to search backwards for the next occurrence of the text string.
Sort If a range of cells is selected, you can sort the cells in the range using the data in one of the rows or columns as the sort key.
Goto.....
Find.....
Sort.....
22
Select the column range by clicking on title ( Number ,Date , Time , Temp. Humidity (RH%) , the whole column will be selected and dark block , above pop menu will be came out , there are 4 Orientation for selection :
1. Sort by Ascending Row Values.
2. Sort by Descending Row Values.
3. Sort by Ascending Column Values.
4. Sort by Descending Column Values.
Move the mouse to click one of them , and key column will show the number of column , for example , you have selected "Temp;" as sort by Column values, the key column will display "4" , then press "OK" , the data will be sorted as requested .
23
From Find menu dialogue , you may enter the value or number or letters you try to find in the "Find What " block. Select from the following :
1. Case Sensitive
2. While Cell
3. Search Vertically
4. Select Cell Only
5. Wrap
Click any of above , and press "Find Next " button to find , press again to find the next , press "Find Prev." To find the previous found point , or exit by pressing "Cancel" key .
Copying and Pasting
You can copy one or more cells to the cilpboard by clicking left button , then press and hold <Shift> button to select the area . You can then paste the contents of the system clipboard into a table control by pressing <Ctrl-V> after selecting the cell(s) into which you want to paste the data.
24
STATISTICS
Depress icon to view histograms of data based on "Whole Range" and on different "Cursor Range" basis. See (Diagram I)
Whole Range - Covers all sample readings which have been logged. Cursor Range - Covers all sample readings between two cursors.
Click and drag the yellow indicator on the button display in the left
1.
part of Diagram I enables equally divide all sample readings into 4 divisions (up to 130). This can also be done by clicking on / or by keying in the figure directly.
Note: You might not be able to see this diagram, this is an example to show you the graph that select whole range to 40 and cursor range to 40 .
2. As shown in the middle top of Diagram I-1, by changing the cursor position, you will find an example for the statistic screen of temperature records . 1021 samples located at the temperature of 44 C( or 213 samples at the temperature of 62.7 between two cursors' range as shown
o
C
in the middle bottom).
3. If you set all sample readings into 100 divisions, you will find the histogram of whole range changed as well. 4
. Diagram I-2 is another example of setting the
been set into 100 division as well.
5. If you compare Diagram I, I-1,I-2, you will notice the differences of changing setting have also changed the related figures. 6
. For the model OM8829 ( Temperature & Humidity datalogger ) , you will see Diagram J-1 as an example, the upper diagram shows the cursor 1 temperature , the lower diagram shows the Relative Humidity records. Diagram J-2 set division as 130.
25
o
cursor range; it has
Temperature Logger Monitor
File name:d\TRLog\TempRhEx2.txt
ID=
PREVIOUS SETTING:
86.0
88.0
70.0
60.0
50.0
40.0
30.0
20.0
10.0
0.0
-10.0
-20.0
-30.0
-36.0
Temp(C)
DATE(M-D-Y) TIME
11-23-2001
17:12:57
0001 0401 0801 1201 1601 2001 2401 2801 3201 3601 4000 4400 4800 5200 5600 6000 6400 6800 7200 7600 8000
08-03-2001 14:46:05
TEMPERATURE CHART
Record No
Time
08-03-2001 19:12:43
CURSOR LOCATION
Cursor 1: # 2381 T=62.2 Date(M-D-Y)=08-03-2001 Time=16:10:23
Cursor 2: # 6028 T=41.4 Date(M-D-Y)=08-03-2001 Time=17:20:31
Diagram I
STATISTIC:FILE=d:\TRLog\TempEx2.txt
Histogram(Whole Range)
WHOLE RANGE
25
25
WHOLE RANGE
25
25
3773
#Of Intervals
3000
75
50
2000
100
4
130
40
CURSOR
#Of Intervals
75
50
100
4
130
40
STATISTIC:FILE=d:\TRLog\TempEx2.txt
#Of Intervals
75
50
100
4
130
100
CURSOR
#Of Intervals
75
50
100
4
130
100
Diagram I-1
1000
00
-38.5 -26.0 -13.6 1.1 11.3 23.8 36.3 48.7 61.2 73.6 86.1
Histogram(Cursor Range)
614
400
200
00
40.4 45.0 49.5 54.1 58.7 63.2 67.8 72.4 77.0 81.5 86.1
Histogram(Whole Range)
1617
1250
1000
Diagram I-2
750
500
250
00
-38.5 -26.0 -13.6 1.1 11.3 23.8 36.3 48.7 61.2 73.6 86.1
Histogram(Cursor Range)
604
400
200
00
40.4 45.0 49.5 54.1 58.7 63.2 67.8 72.4 77.0 81.5 86.1
44.0 1021
OC
62.7 213
41.9 212
OC
63.0 11 4
%
26
Records
Maximum
85.1
Minmium
-37.5
Mean
9.9
Std.Dev
43.3
C
Records
Maximum
Minmium
Mean
Std.Dev
85.1
41.4
72.0
11.6
Records
Maximum
85.1
Minmium
-37.5
Mean
9.9
Std.Dev
43.3
Records
85.1
Maximum
41.4
Minmium
Mean
72.0
Std.Dev
11.6
7. Diagram I is an Temperature example for the statistic source to the Diagram I-1 and Diagram I-2 . And Diagram J is another Temperature/Humidity example for the statistic source to the Diagram J-1 and Diagram J-2.
8. The right part of Diagram I indicates Maximum, Minimum, Mean and Standard Deviation (Std.Dev) values from the temperature of whole range ( Upper part ) and between 2 cursors' range ( Lower part ).
Maximum
85.1
Minmium
Mean
Std.Dev
Maximum : The greatest value of whole logged records, or the greatest value between two selected cursors , or the greatest value of whole Humidity records. Minimum : The least value from the logged records , or the least value between two selected cursors, or the least value whole logged Humidity records. (Model: OM8829) Mean : Average value from the logged records , or average value between two selected cursors , or average value from whole logged Humidity records. (Model: OM8829). Std. Dev. : This is a very useful feature to see more reliable deviation while recording . Calculate each deviation between each value and Mean value, then get an average figure from totalism deviation. (See Diagram I-1 )
41.4
72.0
11.6
27
Temperature/Humidity Logger Monitor
Time
44.0 909
OC
50.0 137
%
43.8 408
OC
51.0 37
%
DATE(M-D-Y) TIME
Records
Maximum
Minmium
Mean
Std.Dev
Records
Maximum
Minmium
Mean
Std.Dev
Records
Maximum
Minmium
Mean
Std.Dev
Records
Maximum
Minmium
Mean
PREVIOUS SETTING:
ID=
86.0
88.0
70.0
60.0
50.0
Humidity
40.0
30.0
Temperature
Temp(C)
20.0
10.0
0.0
-10.0
-23.0
0001 0401 0801 1201 1601 2001 2401 2801 3201 3601 4000 4400 4800 5200 5600 6000 6400 6800 7200 7600 8000
08-03-2001 14:46:05
CURSOR LOCATION
Cursor 1: # 2530 T=64.7 RH=10.1 Date(M-D-Y)=08-03-2001 Time=16:10:23
Cursor 2: # 4634 T=4.2 RH=21.0 Date(M-D-Y)=08-03-2001 Time=17:20:31
STATISTIC:FILE=d:\TRLog\TempRhEx1.txt
TEMPERATURE
#Of Intervals
50
25
4
HUMIDITY
#Of Intervals
50
25
4
STATISTIC:FILE=d:\TRLog\TempRhEx1.txt
TEMPERATURE
#Of Intervals
50
25
4
HUMIDITY
#Of Intervals
50
25
4
File name:d\TRLog\TempRhEx1.txt
TEMPERATURE/HUMIDITY CHART
Histogram(Temperature)
75
100
130
40
Histogram(Humidity)
1137 1000
75
100
130
40
Histogram(Temperature)
75
100
130
130
Histogram(Humidity)
75
400
100
200
130
130
Record No
1914
1500
Diagram J-1
1000
500
00
-23.3 -12.3 -1.3 9.6 20.6 31.6 42.6 53.6 64.5 75.5 86.5
750
500
250
00
-0.3 8.6 17.6 26.5 35.5 44.4 53.4 62.3 71.3 80.2 89.2
1012
800
Diagram J-2
600
400
200
00
-23.3 -12.3 -1.3 9.6 20.6 31.6 42.6 53.6 64.5 75.5 86.5
552
00
-0.3 8.6 17.6 26.5 35.5 44.4 53.4 62.3 71.3 80.2 89.2
Diagram J
11-23-2001
Std.Dev
17:12:57
100
90
80
70
60
50
40
30
20
10
0
08-03-2001 19:12:43
85.5
-22.3
24.0
41.1
88.2
35.6
28.9
85.5
-22.3
24.0
41.1
88.2
35.6
28.9
%Rh
0.7
0.7
28
PRINT GRAPH
Depress icon to print the graph out to either a printer or a fax/internet program depending on your available facilities. See diagram L.
Selecting "Print" button on the menu bar, Windows "Print" dialog box allows you to change the pre-set format ,then print the contents of the currently active window .
Diagram L
(Example)
29
GROUP FILES
Click on its icon to view "Multi-Logger Monitor", which is designed to view and to compare different data files.Diagram M is the screen before you clicking the icon.You may select and change the color of curve to refer to the file you retrieve by clicking the color block. You can print the screen datas by clicking the printer icon at the right top corner.
Diagram M: MULTI-LOGGER GROUP FILE
TEMPERATURE CHART
FAQ: If I want to retrieve another file to replace current file shown on the graph , how should I ?
Ans: Select the file you don't want, then click the color icon and choose the transparent, the old file's graph is disappear , then select the desired file you want to retrieve .
Record No:
30
Click to select the file you want to compare or view with
Group File Monitor
G File
H File
E File
C File
A File
B d:\TRLog\TempRhEx3.txt F d:\TRLog\TempRhEx2.txt
D File
27.0
26.0
F.TEMP
other files , click color icon box to select the color you want , After file retrieved completed , the start point of the file curve indicate with a Capital initial ( A, B, C, D, E, F, G, H ) and Temperature or Humidity, the file name follow the capital shown on below screen .
(Example)
Group File Monitor
A File
B d:\TRLog\TempRhEx3.txt
C File
D File
27.0
F.TEMP
26.0
20.0
15.0
10.0
5.0
0.00.0
F.RH%
Temperature
-5.0
-10.0
-15.0
B.TEMP
-20.0
B.RH%
-22.0 4398 4598 4798 4998 5198 5398 5598 5798 5998 6198 6398 6598 6798 6998 7198 7398 7598 7798 8000
TEMPERATURE CHART
B:Date=07-16-2001 Time=15:26:10 T=-17.0 RH=47.8
F:Date=07-17-2001 Time=14:09:15 T=26.1 RH=67.4
Retrieve B file completed
E File
F d:\TRLog\TempRhEx2.txt
G File
H File
Record No:
Display Control
31
99.9
95.0
90.0
85.0
80.0
%Rh
75.0
70.0
65.0
60.0
55.0
50.0
46.0
To compare with different data files :
Diagram M displays main screen before retrieving any file. Diagram N is what you see after loading the example files:
Indicates different data files loaded and shown in different colors of histogram.
You can always change the offset and range setting desired by moving the indicator on the yellow button. Please see Diagram O , refer to the introduction DISPLAY CONTROL , read throughly for more detail description & examples .
TEMPERATURE CHART
The Temperature Chart displays the diagram after all sample readings are completely downloaded. (Please refer to Diagram A, B)
CURSOR STATUS
The Cursor Status indicates exact figures of Cursor 1 and Cursor 2. (See DiagramQ)
PREVIOUS SETTING:
Previous setting records are listed in this column for reference. ( Please refer to Diagram P. )
Diagram P
(Example)
32
FAQ: If I want to download next logger , how to update the last
#2796 Value=48.1 Date(M-D-Y)=
#5104 Value=26.9 Date(M-D-y)=
one graph without turning off or exiting current screen ?
Ans.: You may have two choices to down load the next logger :
a) Click "PREVIOUS SETTING"to replace the last logger data
with a new one.
b) Close the program and reopen again.
Note : Click "Retrieve File" will not activate with a new logger data
unless you follow above instruction.
BLANK LOGO
This area is designed to insert customers' own logo. The logo will be set as default image and can't be changed by the users. (Please refer to Diagram P-1) If private logo is needed , a clear logo artwork has to be sent to the manufacturer for setting up purpose.
(Example)
Diagram P-1
Diagram Q
Logo
Cursor 1
(red)
Cursor 2
(green)
MESSAGE
It is blank if there is no file retrieved (See Diagram A). After retrieving file, the message will be changed to "Retrieve file completed." (See Diagram Q)
33
DOWNLOAD COMMANDS
Start - To start downloading data to the program.(See below diagram) Stop - To manually stop downloading data to the program; which
means if you don't want to download complete sample readings, you can always stop it at any time you want.
Exit - To exit Download Commands.
DISPLAY CONTROL
Offset - To set up from which sample Reading you would like to start.
There are 0000 , 2500 , 5000 , 7500 ranges for selection.
2500
0000
5000
250
7500
Diagram R
34
500
50
1K
2K
4K
8K
For example : Set Offset by clicking and dragging the red indicator to " 0000 ",Range unchanged at " 8K ", you will see the Diagram R-1 The screen shows the number 0001 as starting point of the
total 8,000 records.
Diagram R-1
06-29-2001 17:18:50 TEMPERATURE Record No Time 06-29-2001 23:50:30
5000
2500
250
7500
0000
50
Another example :
Set Offset by clicking and dragging the red indicator to "4000", Range unchanged at " 8K ", you will see the Diagram S below.
The screen shows the number 4000 as starting point of the total 8,000 records.(See diagram S-1 ) Actually there are 4,000 records per the offset you selected but not a total 8,000 records ( Range set on 8K ).
Diagram S
35
Diagram S-1
2500
5000
250
0000
Display Control ( For X-Axis records) Range - To set up the range of sample Reading you would like to
cover in the chart. There are 50,250,500,1K,2K,4K,8K for selection.
For example:
Set Range by clicking & dragging the red indicator as D iagram T to "1K" , Diagram T-1 is shown the data record from 0001 to 1000.
36
50
7500
Diagram T-1
Another example : Set Range to "4K" as Diagram U, then Diagram U-1 is shown the data record from 0001 to 4001.
Diagram U-1
37
GENERAL SPECIFICATION
TEMPERATURE DATALOGGER
SPECIFICATIONS
TEMP. RANGE
SHAPE
POWER BAT.
ACCURACY
SAMPLE DATA
SAMPLE POINTS
HOUSING DIMENSIONS(mm)
LED SIGNAL
RESOLUTION
CABLE LENGTH
OM8800D
-40~85 C
-40~185
o
o
F
OVAL SQUARE
ER3 or CR2 lithium battery x1
oo
+
Temp. 0.6 C(-20~50 C)
-
ooo
+
1.2 C(-40~20 C,51~85 C)
-
Up to 8000 up to 16000
1K/2K/4K/8K
1K/2K/4K/8K/12K/16K
WATERPROOF
124(L)x92(W)x37(T)(Interface)
108(L)x66(W)x22(T)
80(L)x55(W)x22(T)
124(L)x92(W)(Complete set)
Red (HI,LO Alarm) /Yellow(Record)
o
o
0.1 C (0.1 F) for range -40 C~50 C
oo
0.2 C for range 50~70 C
oo
0.3 C for range 70~85 C
oo
0.1 C (0.1 F)
150 cm with DB9 connector
38
OM88128
OM88128
oo
TEMPERATURE / RELATIVE HUMIDITY
DATALOGGER WITH LCD DISPLAY
SPECIFICATIONS
TEMP. RANGE
HUMIDITY RANGE
FRONT PANEL
POWER BAT.
ACCURACY
BAT LOW DISPLAY
SAMPLE DATA
Up to 16000
SAMPLE POINTS
LCD DISPLAY HOUSING DIMENSIONS(mm)
LED SIGNAL
RESOLUTION
CABLE LENGTH
OM8828 OM8829
-40~85 C
-40~185
o
o
F
N/A 0~100% RH
o
C
OM8828
ER3 or CR2 lithium battery x1
oo
+
Temp. 0.6 C(-20~50 C)
-
ooo
+
1.2 C(-40~20 C,51~85 C)
-
+
Humidity: 3%RH
-
"Lo" DISPLAY
1K/2K/4K/8K/12K/16K
SIZE:12 x 25.5 mm
WATERPROOF
WATER RESISTANT
124(L)x92(W)x37(T)(Interface)
80(L)x55(W)x22(T)
124(L)x92(W)(Complete set)
Red (HI,LO Alarm) /Yellow(Record)
oo
0.1 C (0.1 F)
150 cm with DB9 connector
RH%
OM8829
39
F.A.Q.(Frequently Ask Question)
FAQ: If I can't down load and the display won't active, what
should I do ?
Ans: Make sure you have selected correct baud rate , see page select 4800 baud rate for model OM8800D , or select 9600 baud rate for the rest models. This is to ensure you could download quicker by selecting correct baud rate value .
FAQ: If the logger is not flashing , how should I do ? Ans: If you were set up Magnet start function , when a magnet
passes over the middle of rear side , the logger is not flashing , change with a new and stronger magnetism and try it again. Make sure the flashing interval is the same as programmed sampling time . If you were set 5 minutes as sample rate , the LED will just flash every 5 minutes .
40