OMEGA OM-21 Manual

Page 1
Low bat. symb o l :Red LED blinks every 2
rep o r t.
OM-21
Cold Chain PDF Logger
INSTRUCTION
SHEET
Shop online at omega.co For latest product manuals: www.omegamanual.info
Made in China
Introduct i o n
Congratulations on your purchase of this
singl
e use cold chain PDF datalogger
This datalogger is specifically designed for
monitoring the transportation of products
Temperature readings is saved throughout
the entire duration of the measurement
program.
M5491/0715
m
.
SM
e-mail: info@omega.com
Produc t D e s c r i p t ion
4
6 5
3
2 7
seconds when battery power is too low to
maintain an accurate logging.
Accurate NTC thermistor for temperature
measuring.
Pre-installed CR2032 battery
START/STOP key:
After the logger is programmed, press
"START" key for 2 seconds to start the
logging. To stop it, press "STOP" key for 2
seconds.
MARK key:
1
To place a bookmark manually during the
delivery transition, press MARK key. You will
see time stamps on generated PDF report.
IP65 zipped plastic bag
Open to program the logger and well seal it
before leaving the logger in shipping
container.
3. T h e pr o g ramm i n g ex e c ute i s
des i g ned a s *.e x e fo r m a t.
Ste p 1 C o n figu r ing t he d a ta l o g ger
Anytime before the logging is started, the
logger can be programmed several times. If
the logger is locked by password after first
configuration, password will also be required
to program the logger again.
- Connect the data logger to a PC via the USB
port.
- The logger status LEDs blink for two runs.
- Window Automatic playback is displayed.
- Click on Open folder to view files.
- Open the file "PDF Logger Configuration
Tool. exe".
- The default language is English, you may
change it to German, French, Italian or
This datalogger is equipped with flexible
programming function. Measurement report
output is implemented via PDF file, no
software and no USB driver are required to be
installed.
Read through the instruction manual befor
using this logger. The logger is calibrated
before shipment.
e
USB2.0 plug & play connector. USB driver
is not required.
LED indicator:
Low / H igh : Red LED blinks when the
measuring set limit value is exceeded.
OK: Green LED blinks triple every 5 seconds
when logger is in standby mode but not yet
be started to log readings. When logging
function is started, green LED blinks every 2
seconds
Operatio n
NOT E :
1. A d obe R e ader s oftw a r e is requ i r ed.
2. P l ease p rogr a m the logg e r and
gen e r ate t h e PD F repo r t un d e r ro o m
tem p e ratu r e . Fo r exam p l e: I f the l ogge r
is t a ken o u t f r o m -3 0
lea v e it i n ai r for f e w mi n u tes b e fore
plu g g ing i n to P C to g e nera t e PDF
fr eeze r , ple a s e
Spanish per y
report will be identical to the language you
choose.
- To review the instruction manual, you may
click "Manual" to open the manual PDF file.
- To check the calibration report of the logger,
you may click "Certificate" to open the
corresponding file.
-The definition of each programmable
parameters are:
our need. The generated PDF
Page 2
to activate the logging function is 5 mins
To summarize, to
activate a LED flashing when 30- min
average value of hazardous alarm delay
period is higher than 8
, please program
the alarm delay as 30 mins, alarm type as
cumulative and range as 2~8
the alarm threshold.
25°C
the logging. You may also directly plug
after you pressing the START key so the first
data is logged 10 mins after pressing START
key.
Temp. Unit
Select the temperature displayed unit you
want to see on PDF report. It can be Celsius
or Fahrenheit.
Password
The password security function is default as
off. You may enable it and input at most 16
characters (alphabets or numbers) to
prevent an unauthorized re-programming.
Company Name
Your company name will be displayed on
PDF report as title.(Max. 20 characters)
Alarm types
No matter what alarm types is picked up,
the
LED will not stop flashing once it is
triggered as ON even if the measured
value later returnes to normal range.
Sin g le: a LED alarm is triggered
immediately when the measured value
exceeds the alarm threshold.
Dis a ble: No LED alarm function during the
logging process
Alarm delay
The preset alarm delay time for s i n gle alarm
type is always zero.
The adjustable alarm delay time for
cum u l ati v e type is 5 min to 2 hours.
Alarm Limits
Select the alarm threshold value. For
example, if choosing 2~8
measured value below 2
means the
or above 8
trigger the LED indicator.
Time zone
The logger will be auto synchronized to your
PC time when you press "Save" to confirm
the settings. Any time zone changes during
the measurement are not taken into account.
will
Password:disable
Company Name: blank
Language: English
Once all the programming is done, press
"Save" to confirm the setting and then you
may close the se
tup window and remove the
logger from PC USB port.
Ste p 2 S t a rt l o ggin g
-When logger is successfully programmed,
the green LED blink triple every 5 seconds to
indicate the logger is in standby mode.
-Please seal the zipped bag well and then
press "START" key for two seconds when
you want to start the logging.
-Red LEDs will flash three runs to indicate
the logging is activated.
-During the logging, the green LED will blink
every two seconds. If battery power is too
low to maintain a normal operation, red LED
blinks.
- To place a bookmark manually during the
logger into PC USB port to stop the logging
and generating PDF file.
-It takes just few seconds to generate a PDF
report. The status LEDs blink while logger is
generating PDF report.
Techni c a l d a t a
Model PDF data logger, single use Temp. By NTC thermistor, measurement -30~70 Temp. resolution 0.1 Temp. accuracy +/-0.5 Logging Type Single Use Sampling points 8192 readings Shelf Life 12 months shelf time Operating temp. -30~70
Operating RH% Humidity < 80% Storage temp. -40~85 Storage RH% Humidity <90% Weight ~10g Battery 1PC 3.0V CR2032 (installed
Sampling interval 30 seconds, 5, 10, 30, 60, 90,
Start delay 0, 5, 30, 45, 60, 90,120 minutes Alarm range -20~ -10
Alarm delay 0, 5, 30, 45, 60, 90, 120 minutes Alarm type Single, Cumulative, Disable Operation keys 2 Keys, Start/Stop & Mark LED indicator REC, Low Bat., High /Low
Protection class IP65
o
C (-22~158oF)
o
C (0.1oF)
o
C
o
C (Logging status);
room temp.(PC status)
o
C
before shipment)
120 minutes
o
C, -10~0 oC, 2~8 oC,
o
C, 0~25 oC, 15~25 oC or
0~15 other values
alarm
Cum u l ati v e : a LED alarm is not triggered
when the measured value exceeds the alarm
threshold, but only once the overall average
value during alarm delay duration exceeds
The default value of above parameters are:
Sampling rate: 5 min Alarm Type: Single
Start delay: 0 min Alarm delay: 0 min
Temp. Unit: °C Alarm Limits: 15 –
delivery transition, press MARK key.
Ste p 3 D o w
nlo a d da t a
-Press "STOP" key for two seconds to stop
Directives EN12830, FDA21 CFR Part 11
Operating System Windows only
Page 3
omega.com info@omega.com
Servicing North America:
U.S.A.: Omega Engineering, Inc., One Omega Drive, P.O. Box 4047
Stamford, CT 06907-0047 USA
Toll-Free: 1-800-826-6342 (USA & Canada only) Customer Service: 1-800-622-2378 (USA & Canada only) Engineering Service: 1-800-872-9436 (USA & Canada only) Tel: (203) 359-1660 Fax: (203) 359-7700 e-mail: info@omega.com
For Other Locations Visit omega.com/worldwide
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
any Product(s) be used in or
OMEGA’s policy is to make running changes, not model changes, whenever an i mprovement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2015 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
RETURN REQUESTS/INQUIRIES
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, P URCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the
COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.