OM-143
USB RH/Temperature/GPP
Data Logger with LCD Display
Page 2
Servicing North America:
U.S.A.: Omega Engineering, Inc., One Omega Drive, P.O. Box 4047
Stamford, CT 06907-0047 USA
Toll-Free: 1-800-826-6342 (USA & Canada only) Customer Service: 1-800-622-2378 (USA & Canada only) Engineering Service: 1-800-872-9436 (USA & Canada only) Tel: (203) 359-1660 Fax: (203) 359-7700 e-mail: [email protected]
For Other Locations Visit omega.com/worldwide
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
Page 3
TABLE OF CONTENTS
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 –4
Key Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 –5
Safety Instruction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
What’s in the Blister Pack. . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Product Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 –7
Setup Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 –13
Installing/Replacing the Battery. . . . . . . . . . . . . . 7 –8
Install Software and Drivers. . . . . . . . . . . . . . . . . 8 –9
Configure the Unit . . . . . . . . . . . . . . . . . . . . . . 10 –13
Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . 13 –17
Deploy the Unit . . . . . . . . . . . . . . . . . . . . . . . . 13 –14
Stop Logging and Import Logs. . . . . . . . . . . . . 14 –15
Viewing, Printing and Exporting Data Logs. . . . 15 –16
Other Indications. . . . . . . . . . . . . . . . . . . . . . . 16 –17
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Maintenance Tips. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2
Page 4
INTRODUCTION
Thank you for purchasing the OM-143 USB RH/Temperature /GPP Data Logger with LCD Display. Please read this user’s manual carefully and thoroughly before using the instrument.
The OM-143 is a low-cost, compact, battery-powered thermo­hygrometer capable of unattended logging (recording) of the ambient temperature and relative humidity (RH) of an environment for days, weeks or months. In addition, software included with the product can also calculate and display absolute humidity levels in units of grains per pound (GPP) or g/kg. Also called mixing ratio, GPP is a more useful moisture metric than RH to water damage remediators. Using RH alone, a remediator might unknowingly introduce moist air— with a low RH but a high GPP—during a job’s drying phase.
Data logging can be started or stopped by pushing a button on the unit, eliminating the need to bring a laptop to the job site. After a data log has been captured and stored, it can be uploaded as a .txt file to any PC running the Windows
®
Windows
XP operating system after plugging the OM-143 into one of the computer’s USB ports. The OM-143’s internal flash memory is large enough to store 16,300 pairs of RH and temperature readings.
Included with the product—which is slightly larger than a thumb drive—is a mini-disc with the drivers needed to interface the unit to the computer, as well as software that can display a data log file as a table or graph and/or export it to Microsoft Excel. Exporting to Excel is recommended for sophisticated trending and analysis of humidity and temperature data and easy detection of unexpected excursions.
®
7 or
Windows®7 and Windows®XP are registered trademarks of Microsoft Corporation.
3
Page 5
Applications for the OM-143 include:
• Water damage remediation—Because the OM-143 is designed to be deployed and left in place over the durion of the job, it is always in equilibrium with the environment.
• Process control—The OM-143 enables, simplifies and cuts
the cost of environmental monitoring of office buildings, greenhouses, food and equipment storage facilities, wineries, freezers, shipping containers, HVAC/R installations, computer rooms, labs, libraries, museums and saunas.
• Electronics manufacturing—Alerts provided by the low-
humidity alarm function of the OM-143 can mitigate the risk of damaging high-voltage static discharges.
KEY FEATURES
• One button starts/stops data logging without a PC
• Plug-and-play USB 2.0 interface; no need for cables,
cradles or docks
• Large storage capacity: 16,300 pairs of RH and
temperature readings
• 3-1/2-digit LCD displays real-time temperature and humidity readings; also indicates device working status
and “memory full”
• Included software is used to set: Sampling time from 10 seconds to 12 hours; Hi/Lo alarm setpoints for humidity and temperature; Temperature unit (°F or °C)
4
Page 6
• Software also calculates and plots GPP and dew point and wet bulb temperatures and displays data logs as time­based graphs synchronized to your PC’s clock
• One-click exporting of logs to Excel or Word
• Long battery life (up to one year)
• Includes USB extension cable
• One-year warranty
SAFETY INSTRUCTION
Do not use the OM-143 in the presence of flammable or explosive gases.
WHAT’S IN THE BLISTER PACK
The OM-143 comes in a blister pack that also contains
• One “1/2 AA” battery. Replacement batteries (Part No. OM-EL-BATT) are available from Omega
• A disc with software drivers for Windows7 and Windows XP computers, the ELusb V1.21.3 program for capturing data logs and displaying them as curves, and a PDF of this user’s manual
• A USB extension cable
• A hard copy of this user’s manual (inside the fold-over card)
5
Page 7
PRODUCT OVERVIEW
Fig. 1a below shows the main components, controls and connectors of the OM-143. Fig. 1b (on the facing page) shows all possible indications on the OM-143’s display. Familiarize yourself with their names and functions before moving on to the Setup Instructions.
E
F
D
C
B
Fig. 1a. The components, controls and connector
of the OM-143
A. Temperature and humidity sensor window B. Battery compartment (latch is on rear of unit) C. 3-1/2-digit LCD (displays temperature and humidity
readings; also indicates device working status and
“memory full”) D. USB plug E. Protective cap F. MODE button
6
A
Page 8
Fig. 1b . All possible display indications on the OM-143
Display Legend
[-]188.8: Humidity or temperture value or one of the
▼
: indicates reading is higher than high alarm setpoint
for that parameter
: indicates reading is lower than low alarm setpoint
▼
for that parameter
%RH: relative humidity unit
°C: Celsius temperature unit
°F: Fahrenheit temperature unit
following four working status indications
LOG: Unit is logging data
- - -: Unit is in standby mode, ready to log data FUL: Internal memory is full (and logging has been
stopped)
CON. Unit needs to be configured by software
SETUP INSTRUCTIONS
INSTALLING/REPLACING THE BATTERY
LATCH
The OM-143 uses a 3.6VDC “1/2 AA” Lithium-ion battery. To open the battey compartment, use a paper clip or a screwdriver with a fine point to push in the silver flange showing through the small square hole in the rear of the unit
7
Page 9
(left photo on previous page). Push back the cover (middle photo), remove the old battery (right photo) and insert a fresh battery in the correct orientation. Then replace the cover, pushing it forward until you feel and hear a click.
After you replace the battery, the LCD will “wake up” and begin displaying real-time readings. If the MODE button is not pressed within 60 seconds, the LCD will be automatically powered off.
Note: If the battery is removed from a unit before it has been set up from software or while it is recording, the unit must be reconfigured before it can be reactivated. If it is not reconfigured, the unit will remain in standby mode.
INSTALL SOFTWARE AND DRIVERS
1. Place mini-disc in tray of CD/DVD drive, shiny side down. Close tray.
2. When the AutoPlay window appears, click on “Open folder to view files using Windows Explorer.”
3. Windows Explorer will indicate that there are two “Files Currently on the Disc”. One is actually a folder containing a PDF of the OM-143 user’s manual (as well as PDFs of the manuals for the OM-141 and OM-144—versions of the OM-143 without an LCD and with an IR thermometer). Copy the OM-143 manual to your computer’s hard drive for permanent storage. The other file is a Windows Installer Package (.msi file) named “ELusb_V1.23.1.msi”. Double­click its icon.
4. A window titled “Welcome to ELUsb Installation Wizard” will appear. Click Next to continue.
5. Within the Software License Agreement window that appears next, click the I agree (T) box.
8
Page 10
6. The next three windows advise that the ELusb application will be installed in your computer’s C:\Program Files (x86) folder in a new subfolder named <ELusb>, and that shortcuts to the program will be created on your desktop and Quick Launch bar. If you want to install the software elsewhere, click Browse and choose another location Within each window, after selecting either the default location or another location, click Next(N) to continue.
7. When the “Preparing to Install” window detailing the changes to made appears, click Install (I) to continue.
8. Before the installation begins, a User Account Control window may appear and ask, “Do you want to allow [ELusb V1.21.3msi] to make changes to this computer?” Click Yes to continue.
9. The ELUsb Setup Wizard will now install the program on your computer. On the next window that appears, click the Completed (F) box.
10. At the bottom of the Device Driver Installation Wizard that appears next, click Next to continue.
11. The next window to appear is a Windows Security window. Click on “Install this driver software anyway” to continue.
12. The next Device Driver Installation Wizard window will advise you that an OM-141/143/144 driver from Omega or Kecheng Laboratories was successfully installed and that your device is now “Ready to use.” Click Finish to continue.
9
Page 11
CONFIGURE THE UNIT
Note that the software installation has added a icon to your desktop. To facilitate access to the program, right-click on the icon and pin it to both your Start button and the Quick Start area at the left of your taskbar. You can now eject the installation mini-disc and store it with your other important software.
Plug the OM-143 into an available USB port of your computer.
If your computer is a laptop and the only available USB jack does not have enough vertical clearance to accommodate jack the OM-143 without mechanically stressing its USB plug, do not plug the OM-143 directly into the jack. You do not want the plug of the OM-143 to support any of the weight of the laptop. If that is the case, plug the OM-143 into
the jack of the included USB extension cable and insert the plug of the extension cable into the computer’s USB jack.
The first time you plug the OM-143 into your computer, your system tray will notify you that the computer is “Installing Device Driver Software”. When the notification disappears open the ELusb program by selecting it via your Start button, clicking on its icon on your taskbar, or double-clicking its desktop icon.
If the software has installed correctly and you have plugged the OM-143 into a working USB port of your computer, the following screen should appear on your computer’s display.
10
Page 12
Fig. 2. The three options made available by plugging the OM-143 into a USB port
Note: If you have already used your OM-143 to log data and it is still in recording mode, plugging the unit in will cause the following window to appear.
Click Yes to stop recording and import all data recorded to that point in time. Click No to continue to Setup without importing recorded data (meaning it will be lost). Click Cancel to go back and unplug the unit from the USB port.
11
Page 13
You can no w use the software to “configure” your OM-143. Configuring the device means choosing four key parameters:
• Sampling time—How often it makes and stores humidity
and temperature measurements.
• Temperature unit—degrees Fahrenheit or Celsius (°F or °C).
• High and low temperature alarm setpoints
• High and low humidity alarm setpoints
To begin, click on “Set up the USB Data Logger” in the window shown in the figure above. Doing so will cause the following window to appear.
Fig. 3. The OM-143 configuration (Setup) window
Using your mouse, trackball or touchpad, first choose your Temp Scale and Sample Rate (sampling time). With regard to sampling time, using the shortest setting of 10 seconds may result in too much essentially identical data being stored in memory, potentially limiting its capacity to capture important events such as an unexpected excursion of temperature or
12
Page 14
humidity. Conversely, using the longest sampling time setting of 12 hours will likely cause important events to go undetected.
If you are an experienced data logger, you already know how to choose the optimum sampling time for your application. If you are a novice, pick a sampling time closer to the minimum sampling time than the maximum, on the assumption that the OM-143 probably has enough memory (16,300 pairs of RH and temperature readings) to store all the data you ask the meter to log—unless you use it to log data for months.
Be aware that if you do manage to fill up the OM-143’s memory (by logging for a very long time at a very short sampling time, for example), the device will simply stop recording. It will not continue recording by overwriting the oldest saved data with new data.
If you wish, you can also drag the sliders shown in the figure to set high and low alarm setpoints for temperature and RH.
Once you have chosen your data logging parameters, click Setup. A window will appear advising that “Setup completed successfully.” Clicking Setup also synchronizes the OM-143’s internal clock with your computer’s clock, giving all data logs to be captured an absolute time reference.
You can now unplug the OM-143 from your computer’s USB port .The unit is in “standby” mode, ready to begin logging data.
OPERATING INSTRUCTIONS
DEPLOY THE UNIT
The OM-143 is particularly easy to deploy on a job site because the unit does not need to be plugged into a computer to be activated (ready to start logging data). Working together, the MODE button and the LCD serve as the control and indicators needed to: 1) start and stop data logging and 2) verify and change the unit’s working status.
13
Page 15
That leaves only two things to be sure of when choosing a place to deploy the OM-143:
• The temperature and humidity sensor window
(Fig. 1a, Callout A) is not covered
• The humidity and temperature values at the unit’s
location are “representative” of the room
The OM-143—with its protective cap on—can be deployed either by laying it on a horizontal surface or by hanging it
vertically within the included plastic mounting bracket. The back of the bracket features both a magnet and a hanger hole.
Once you have deployed the OM-143, you can initiate data logging. To start data logging, press and hold the MODE button for at least 3 seconds. The LCD will display the word
“LOG” to confirm that logging has begun.
STOP LOGGING AND IMPORT LOGS
To stop data logging, press and hold the MODE button for at least 3 seconds. The LCD will display three dashes (- - -)
to confirm that logging has stopped.
That is how you should stop data logging if you plan to initiate another logging session later without uploading logged data to your computer in the interim. If you allow logging to continue and you do not upload your data to your computer soon, you risk filling up the OM-143’s memory.
A simpler and safer way to stop data logging is to plug the OM-143 into your computer’s USB port immediately after completing a logging session and then use the ELUSB
software to stop logging.
Each time you plug the OM-143 into the USB port, the window
shown in Fig. 2 will appear. When you plug in the unit, it
makes no difference whether it is still logging; the procedure
14
Page 16
for uploading (importing) data logs is identical. Just click on the middle selection—Stop the USB Data Logger and download data—to instantly execute both functions. The data will be imported as a .txt file. A “Save As” window will appear, asking you to name the file and choose where you want it to be stored.
VIEWING, PRINTING AND EXPORTING DATA LOGS
Immediately after you close the “Save As” window by clicking on Save, the uploaded data log will automatically be displayed as a set of curves representing all loggable values. Fig. 4 is a typical example. The initial display includes all parameters, including dew point and GPP or g/kg (which the software automatically calculates) and alarm setpoints that you may have entered during Setup. To remove any parameter or horizontal and vertical grid lines from the overall plot, uncheck its box at right.
Fig. 4. A typical initial plot of an imported data log
15
Page 17
There are several features shown in Fig. 4 worth noting:
• To mark the position of every recorded data point on the
curve, check the “Mark Points” box at the top of the sidebar at right.
• To get a higher-resolution view of the data recorded during
a specific period of time, scale up and down the curves by dragging the slider at the bottom of the window to the left and right.
• Hovering the cursor over a particular point of the curve
displays all data related to this point—including the maximum, minimum and average values of the entire curve—in a shaded box at upper left.
• The dates and times shown on the x-axis scale and in the
upper left corner are in sync with your computer’s clock.
• To print the .txt file of any data log, click the Print virtual
button.
• To open a previously saved data log, click the Open button.
This function is also a vailable as the “View previously saved data” option shown in Fig. 2.
• To export a file to Excel for advanced trending and
analysis, click the Export to Excel button.
OTHER INDICATIONS
As mentioned earlier, the MODE button and the LCD can be used together to display real-time temperature and humidity measurements, and to indicate the current working status of the OM-143.
Circuitry inside the OM-143 automatically shuts down the LCD (to conserve battery life) if the MODE button has not been pressed within the past 60 seconds. That is the case even
16
Page 18
when the unit is logging data. In other words, the LCD is normally “asleep”.
To wake up the LCD, briefly press the MODE button. This first press of the button will cause the LCD to resume displaying the same indication that was on-screen when th LCD was last put to sleep.
There are four possible display indications: (1) current temperature in Celsius, (2) current temperature in Fahrenheit, (3) current relative humidity (in %), and (4) current working status. Each press of the MODE button changes the display to the next indication in the sequence; pressing the MODE button when the LCD is displaying the unit’s current working status (4) resets the cycle to (1).
There are four possible working status indications:
• LOG: The unit is currently logging data
• - - -: The unit is in standby mode, ready to log data
• FUL: The unit’s internal memory is full and logging has been
stopped
• CON: The unit needs to be configured by software
When the LCD is displaying a temperature or humidity reading, it also gives an indication if the reading is outside the “acceptable” range of readings for that parameter— assuming that high and/or low alarm setpoints were entered during the Setup procedure. If an “up” arrow appears at the left of the reading (see Fig. 1b on p. 7), the measured value of that parameter is higher than its high alarm setpoint. If a “down” arrow appears at the left of the reading, it is lower than the low alarm setpoint for that parameter.
17
Page 19
SPECIFICATIONS
Temperature Measurement Range -4° to 158°F (-20° to 70°C) Temperature Measurement Accuracy
Temperature Measurement ±0.2°F (0.1°C) Resolution and Repeatability
Relative Humidity (RH) 0 to 100% Measurement Range
RH Measurement Accuracy
RH Measurement Resolution ±0.1% and Repeatability
Internal Storage Capacity 16,300 pairs of RH/Temp
Display 3-1/2-digit LCD with
Temperature Response Time 0.5 seconds Humidity Response Time 5 seconds Interface USB 1.0/2.0 Length of USB Extension Cable 7.9 in. (200mm) Dimensions 4.96 x 1.10 x 2.01 in.
Weight <7 oz. (<200g) Power Source 3.6V non-rechargeable
Current Consumption <0.2mA Battery Life (typical) 1 year
±1.8°F (1°C) from 14° to 104°F (-10° to 40°C); ±3.6°F (2°C) elsewhere in range
±3% from 20 to 80%RH, ±3.5% elsewhere
readings
60 second auto power off
(126 x 28 x 51 mm)
(1/2 AA) 1200mAh Lithium-ion battery
18
Page 20
MAINTENANCE TIPS
• Avoid dropping the unit. Do not subject it to violent shock or
vibration or expose it to strong electromagnetic fields (for example, near arc welders or induction heaters).
• Keep the unit out of direct sunlight.
• Do not use chemicals or abrasive cloth to clean the display
window or housing.
• Before storing the OM-143 for an extended period of time
(several months or more), remove the battery to avoid having it leak and damage the unit.
19
Page 21
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
WARRANTY/ DISCLAIMER
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the
product was PURCHASED,
2. Model and serial number of the product
under warranty, and
3. Repair instructions and/or specific
problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2016 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
RETURN REQUESTS/ INQUIRIES
FOR NON-WARRANTY REPAIRS, OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems
relative to the product.
consult
Page 22
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
SM
TEMPERATURE
MU
Thermocouple, RTD & Thermistor
Probes, Connectors, Panels & Assemblies
MU
Wire: Thermocouple, RTD &
Thermistor
MU
Calibrators & Ice Point References
MU
Recorders, Controllers & Process
Monitors
MU
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
MU
Transducers & Strain Gages
MU
Load Cells & Pressure Gages
MU
Displacement Transducers
MU
Instrumentation & Accessories
FLOW/LEVEL
MU
Rotameters, Gas Mass Flowmeters
& Flow Computers
MU
Air Velocity Indicators
MU
Turbine/Paddlewheel Systems
MU
Totalizers & Batch Controllers
pH/CONDUCTIVITY
MU
pH Electrodes, Testers &
Accessories
MU
Benchtop/Laboratory Meters
MU
Controllers, Calibrators, Simulators
& Pumps
MU
Industrial pH & Conductivity
Equipment
DATA ACQUISITION
MU
Data Acquisition & Engineering
Software
MU
Communications-Based Acquisition
Systems
MU
Plug-in Cards for Apple, IBM &
Compatibles
MU
Data Logging Systems
MU
Recorders, Printers & Plotters
HEATERS
MU
Heating Cable
MU
Cartridge & Strip Heaters
MU
Immersion & Band Heaters
MU
Flexible Heaters
MU
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
MU
Metering & Control
Instrumentation
MU
Refractometers
MU
Pumps & Tubing
MU
Air, Soil & Water Monitors
MU
Industrial Water & Wastewater
Treatment
MU
pH, Conductivity & Dissolved
Oxygen Instruments
M5565/0416
Loading...