Omega LDX-3A, LDX-3 User guide

Page 1
omega.com
For latest product manuals:
omegamanual.info
User’s Guide
LDX-3A
Shop online at
Page 2
USA: One Omega Drive, Box 4047
ISO 9001 Certified
Stamford CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: [email protected]
Canada: 976 Bergar
Laval (Quebec) H7L 5A1, Canada Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: [email protected]
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378 / 1-800-622-BEST
®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
®
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico: En Espan˜ol: (001) 203-359-7803 e-mail: [email protected]
FAX: (001) 203-359-7807 [email protected]
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: [email protected]
Czech Republic: Frystatska 184, 733 01 Karvina´, Czech Republic
Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: [email protected]
France: 11, rue Jacques Cartier, 78280 Guyancourt, France
Tel: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427 Toll Free in France: 0800 466 342 e-mail: [email protected]
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: [email protected]
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5BD United Kingdom Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: [email protected]
OMEGAnet®On-Line Service Internet e-mail
www.omega.com [email protected]
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGAis constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
M-1633/0905
LDX-3A
Page 3
Section Title Page
4.6.2 Current Connections . . . . . . . . . . . . . . . . . 12
4.7 Using an EMC Cable Gland . . . . . . . . . . 13
5.0 Setting up the Transducer and LDX-3A . . . 14
5.1 Option Links Explained . . . . . . . . . . . . . 15
5.2 Basic Procedure . . . . . . . . . . . . . . . . 16
5.3 Sensitivity and the X2, X4,DIV2 Links . . . . 20
6.0 Half-Bridge Version . . . . . . . . . . . . . . . . 21
7.0 Specifications . . . . . . . . . . . . . . . . 22
7.1 Electrical . . . . . . . . . . . . . . . . . . . . . 22
7.2 Mechanical and Connections . . . . . . . . . 24
7.3 Environmental. . . . . . . . . . . . . . . . 24
7.4 Notes . . . . . . . . . . . . . . . . . . . . . .25
Section Title Page
1.0 Index . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.0 Safety Information . . . . . . . . . . . . . . . . 2
3.0 Introduction . . . . . . . . . . . . . . . . . . . . . 5
3.1 LDX-3A Enhancements . . . . . . . . . . . . 5
4.0 Installation . . . . . . . . .. . . . . . . . . . . . 6
4.1 Mounting . . . . . . . . . . . . . . . . . . . . . 6
4.2 Operational Environment . . . . . . . . . . . . . . 7
4.2.1 Residential, Commercial & Light Industrial
Environments . . . . . 7
4.2.2 Industrial Environments . . . . . . . . . . . . 7
4.3 Electrical Connections . . . . . . . . . . . . . . . 8
4.4 Connecting the Transducer . . . . . . . . . . . . . 9
4.5 Mains Power Supply . . . . . . . . . . . . . . 10
4.6 Connecting the Signal Out . . . . . . . . . 11
4.6.1 Voltage Connections . . . . . . . . . . . . . . . . .11
LDX-3A
1.0: Index
1
M-1633/0905
Page 4
Terms in this Manual
WARNING statements identify conditions or practices
that could result in personal injury or loss of life.
CAUTION statements identify conditions or practices that could result in damage to the equipment or other property.
Symbols in this manual
This symbol indicates where applicable cautionary or other information is to be found.
WARNINGS: Do not operate in explosive atmosphere
To avoid explosion, do not operate this equipment in an explosive atmosphere.
Installation Environment
The power supply unit is for indoor use only. It should not be mounted where ingress of fluids is possible.
Safety Critical Environments
This equipment is not intended for use in a safety critical environment.
Power Source
Apply no more than 264VAC between mains supply pins and ground.
LDX-3A
2.0: Safety Information
2
M-1633/0905
Page 5
LDX-3A
2.0: Safety Information
3
M-1633/0905
Service Safety
Safety Compliance Statement
LDX-3A is designed as Safety Class 2 apparatus with functional ground to comply with IEC 61010-1. The power supply unit is marked as compliant with EN60950-1, but when as part of the LDX-3A assembly and installed by qualified personnel according to these user instructions, then IEC 61010-1 compliance can be achieved.
This equipment has been designed and tested to meet the requirements of the Low Voltage Directive (73/23/EEC 1997) and has been supplied in a safe condition. This manual contains information and warnings that must be followed by the user to ensure safe operation and to maintain the apparatus in a safe condition.
Notes Ground Connection
This equipment is supplied with a mains ground connection. This is a functional ground only. Loss of ground connection may result in an AC voltage between accessible parts of the LDX-3A and reference earth. The voltage, current or charge is at a safe level.
The power supply unit contains no user serviceable parts
This equipment must be returned to your dealer for all servicing and repair.
Page 6
LDX-3A
2.0: Safety Information
4
M-1633/0905
CAUTION: Electrostatic Discharge
This equipment is susceptible to ESD (Electrostatic Discharge) when being installed or adjusted, or whenever the case cover is removed. To prevent ESD related damage, handle the conditioning electronics by its case and do not touch the connector pins. During installation, follow the guidelines below.
- Ensure all power supplies are turned off.
- If possible, wear an ESD strap connected to ground. If this is not possible, discharge yourself by touching a metal part of the equipment into which the conditioning electronics is being installed.
- Connect the transducer and power supplies with the power switched off.
- Ensure any tools used are discharged by contacting them against a metal part of the equipment into which the conditioning electronics is being installed.
- During setting up of the conditioning electronics, make link configuration changes with the power supply turned off.
Avoid touching any other components.
- Make the final gain and offset potentiometer adjustments, with power applied, using an appropriate potentiometer adjustment tool or a small insulated screwdriver.
CAUTION: Fuse Rating
The LDX-3A power supply unit does not need a fuse for safe operation. However, wiring to the LDX-3Apower supply unit may need to be fused. If the mains lead supplied with your LDX-3A has a fuse incorporated into the plug, then a fuse of the same type and value should replace it. Otherwise, fusing and wiring to the LDX-3A must be performed in accordance with local safety regulations. Refer wiring and fuse replacement to qualified personnel.
Page 7
The LDX-3A is a development of the original LDX-3A. It is a compact conditioning module powered from an external power supply unit. Adjustable gain and zero controls are provided for use with the complete range of Omega LVDT and half-bridge transducers.* The unit is of robust construction, housed in a die cast metal box providing a substantial degree of mechanical protection. The LDX-3A uses a universal mains input power supply and provides an output of up to ±10V and ±20mA.
The LDX-3A has the following features.
- Wide gain range, allowing ALL Omega transducers to be connected without the need for attenuation resistors.
- Up to ±10V dc and ±20mA are available for all gain settings.
- Fixed and variable offsets make setting of uni-polar output easier.
- Gain and offset adjustment are fully independent.
- Selectable transducer excitation frequency.
- Selectable transducer load resistances.
- Selection between forward and reverse connection.
- Fully CE compliant.
- Half-bridge transducers can be accommodated with simple plug wiring changes.*
*For half-bridge only variant see section 6.0.
LDX-3A
3.0: Introduction
5
M-1633/0905
3.1: LDX-3A Enhancements
Page 8
LDX-3A
4.0: Installation
6
M-1633/0905
The LDX-3A may be mounted in a variety of ways and in any attitude. Ensure that there is enough space for the cover to be removed to allow for internal adjustments. Space should also be allowed for the transducer connector, EMC glands (if fitted) and cabling. It is recommended that the LDX-3A case be connected to earth or chassis. This earth connection is not a safety earth, but is part of the overall electrical screening scheme.
The power supply may be mounted in any position. The power supply is for indoor use only and must be protected from fluids. Ensure that the power supply can be disconnected from the mains. This can be by removing the plug from the power supply itself, or by disconnection from the mains source. Also ensure that the mains lead cannot be accidentally disconnected.
4.1: Mounting
Underside Mounting Top-Side Mounting
Note: If replacing LDX-3A MKI a retrofit mounting kit is available if repositioning holes is not possible.
Page 9
This section discusses the type of installation required depending on the electrical environment.
Typically, this will be an office, laboratory or industrial environment where there is no equipment likely to produce high levels of electrical interference, such as welders or machine tools.
Connections may be made using twisted, unscreened wire. This is a cost effective option and will give good performance in this environment.
Standard equipment wire such as 7/0.2 (24AWG) can be twisted together as required. Standard data cable such as generic CAT5 UTP will also give good performance.
Typically, this will be an industrial environment where there is equipment likely to produce high levels of electrical interference, such as welders, machine tools, cutting and stamping machines.
Connections should be made using screened cable. Braided or foil screened cables may be used. The cable screen should be connected to the LDX-3A case at cable entry point. The case of the LDX-3A should be connected to a local ground. An EMC cable gland is recommended. This is supplied with the LDX-3A.
When selecting the type or wire or cable to be used, consider the following parameters:
- Screening.
- Conductor size (resistance).
- Mechanical aspects such as flexibility and robustness.
This is not a complete list. Installations may require other special cable characteristics.
M-1633/0905
4.2: Operational Environment
LDX-3A
7
4.0: Installation (continued)
4.2.1: Residential, Commercial and Light Industrial Environments
4.2.2: Industrial Environments
Page 10
The LDX-3A requires three connections.
1. Transducer
2. Mains power supply.
3. Output signal, voltage or current. A small hole should be made in the grommet prior to passing the wires through. If a screened cable is to be used, an
EMC cable gland is recommended. For best performance in electrically noisy environments, the case of the LDX-3A should be connected to a local earth.
This can be achieved via the mounting bracket. This earth connection is not a safety earth, but is part of the overall electrical screening scheme.
The wiring layout arrangements are similar for LDX-3A fitted with EMC glands and screened cable.
0V +VE
V OUT 0V I OUT
COARSE GAIN
FINE GAIN
COARSE OFFSET
-VE +VE
5V
10V
FINE OFFSET
10K
2K
- -
¦
R
¦
|F|
- -
DIV 2
X2
X4
G BK W R B
OUTPUT
POWER
1 2 3 4 5 6 (7)
FREQ
LO
(ON)
HI
(OFF)
NULL
4.3: Electrical Connections
4.0: Installation (continued)
M-1633/0905
LDX-3A
Page 11
LDX-3A
4.0: Installation (continued)
9
M-1633/0905
4.4: Connecting the Transducer
Transducers fitted with a 5-pin DIN plug are simply screwed into the case mounted socket. Transducers not fitted with a plug should be wired to the plug supplied. For the half-bridge only variant see section 6.0.
LVDT Electrical Connections Red and Blue Energising Green and White Signal Yellow Secondary Centre Tap
Red and White
In Phase for Inward Displacement
Black Transducer Body Ground
Half-Bridge Electrical Connections Red and Blue Energising Yellow Signal
Red and Yellow
In Phase for Inward Displacement
Black Transducer Body Ground
CORE
Blue
Yellow
Red
V
V
A
B
Centre Tap
V
O
CORE
Primary
Red
Blue
V
e
Green
Yellow
White
V
V
A
B
Centre Tap
V
O
Measured
Output
Red
Blue
Yellow
+
-
Case
Half-Bridge
Cable Screen
Black
Case
Plug view
into pins
Standard Half-Bridge Gauging Probe Plug Connections
Note 1: + indicates inward movement of the tip. Note 2: The transducer body may be disconnected from the cable screen by cutting the black wire inside the connector
4 2
1
3
5
LINK
Red
Blue
Yellow
+
-
Case
LVDT
Cable Screen
Black
Case
Plug view
into pins
Standard LVDT Gauging Probe Plug Connections
Note 1: + indicates inward movement of the tip. Note 2: The transducer body may be disconnected from the cable screen by cutting the black wire inside the connector
3
4
5
2
1
Green
White
Page 12
10
The mains adaptor is pre-wired to the LDX-3A. This is a universal power adaptor, so no adjustments are required for mains frequency or voltage. Please review safety statements and installation section.
WARNING:
Ensure that the local safety precautions are observed for installation of mains equipment.
4.0: Installation (continued)
4.5: Mains Power Supply
1.8m lead
Gktpodfg ggfdgjs dfofkg f sdreyot
Gktpodfg ggfdgjs dfofkg f sdreyot
LDX-3A
M-1633/0905
Page 13
11
The output signal may be voltage or current.
4.0: Installation (continued)
4.6: Connecting the Signal Out
0VVout
Iout
Process Monitor
+
-
Voltage
Voltage Monitor
LDX-3A
M-1633/0905
4.6.1: Voltage Connections
Voltage is most easily be monitored using a variety of instrumentation such as voltmeters. Voltage drops along wires contribute to measurement errors, so care must be taken when using long cable lengths (100m for example). High impedance instruments are more prone to interference.
Page 14
12
Current output requires the use of purposely designed current input instrumentation. Current output is more suitable for transmitting over longer distances because current is not lost due to wiring resistance. Additionally, with a low impedance, a current loop is less likely to pick up noise.
The total loop resistance (resistance of measuring equipment and wiring) must not exceed specification. Note: The LDX-3A is not loop powered. A power supply must not be used in-line with the current output.
4.0: Installation (continued)
LDX-3A
M-1633/0905
4.6.2: Current Connections
Iout
Vout
0
V
Process Monitor
Current
+
-
Current Monitor
Page 15
LDX-3A
4.0: Installation (continued)
13
M-1633/0905
To ensure the cable screen is properly connected to the LDX-3A case, an EMC cable gland should be used. This is supplied as an extra. The diagrams below assume a single 4-way cable is being used. Remove the grommet and fit the cable gland as shown below.
- Prepare cable as shown.
- Slide the Dome nut, sealing ring and plastic insert over cable.
- Fold and roll the screen back over itself to form a lump.
- Push cable into gland body followed by the plastic insert (ensure anti-rotation slots engage), sealing ring and Dome nut.
- Ensure all components are properly seated before tightening
the dome nut.
10-15 mm
Screen
30 mm
125 mm
Rolled Back Screen
Dome
Nut
Sealing
Ring
Plastic
Insert
Gland
Body
Case Wall
Rotate Gland Body
to Tighten
4.7: Using an EMC Cable Gland
Page 16
LDX-3A
5.0: Setting up the Transducer and LDX-3A
14
M-1633/0905
The LDX-3A may be set-up with output signals anywhere within a ±10 Vdc or ±20 mA range. Typical outputs are ±10 Vdc, ±5 Vdc, 0-10 Vdc and 4-20 mA. These procedures apply to voltage and current output.
Voltage and current output are available at the same time, although they cannot be individually adjusted. Either voltage or current should be chosen as the calibration reference.
All outputs use 0V signal as the signal reference. A list of standard link settings is available, see section 5.2.
CAUTION:
During installation and adjustment, the top of the enclosure has to be removed for access to user adjustments. At this time, standard ESD handling precautions for electronic equipment should be followed.
Page 17
LDX-3A
5.0: Setting up the Transducer and LDX-3A (continued)
15
M-1633/0905
The table below and subsequent diagrams explain the link functions and show the factory setting.
5.1: Option Links - Explained
Link ON Link PARKED
Link OFF
Link
Description
Options Standard Setting
Course Gain Sets the basic gain 1 link on Positron 1 to 6 Link ON Position 1 Fine Gain Adjustment between course gain ranges Potentiometer Adjustment Mid Position Course Offset Shifts the output by a fixed amount Link ON -VE or +VE and Link ON 5V or 10V
No offset - Link Parked
No offset - Links Parked
Fine Offset Fine trim around any fixed offset Potentiometer Adjustment Mid Position (7) Null Used during set-up to null output ON, OFF OFF Freq. Selects transducer primary frequency Lo - ON, Hi - Parked Hi Freq. - Link Parked Input Resistance Sets transducer secondary load 100kΩ Parked, 10kΩ - ON, 2kΩ - ON 100KΩ - Link Parked Polarity (FR) Enables output signal direction change 2 Links across Forward or Reverse F Position - 2 Link ON Input Gain Input Gain of x1, x2, x4 or divide 2 X1 - Parked, X2 - ON, X4 - ON, DIV2 - ON Link parked on X2
0V +VE
V OUT 0V I OUT
COARSE GAIN
FINE GAIN
COARSE OFFSET
-VE +VE
5V
10V
FINE OFFSET
10K
2K
- -
¦
R
¦
|F|
- -
DIV 2
X2
X4
G BK W R B
OUTPUT
POWER
1 2 3 4 5 6 (7)
FREQ
LO
(ON)
HI
(OFF)
NULL
Page 18
LDX-3A
5.0: Setting up the Transducer and LDX-3A (continued)
16
M-1633/0905
5.2: Basic Procedure
To set-up the LDX-3A, some basic steps should be followed. The following steps describe a typical setting procedure and applies to most applications. Other procedures may be
used as appropriate.
Zero
-5V +5V
Null
electronics
transducer
Null
Zero
transducer
electronics
Null
+5V0V +10V
electronics
transducer
Shift zero
For a bi-polar output i.e. ±10 Vdc or ±20 mA, follow steps 1 to 3. For a uni-polar output i.e. 0-10 V, 0-20 mA or 4-20mA, follow steps 1 to 4.
In either case, step 5 (final checks) should be followed on the completion of the set-up.
Step 1 Step 2 Step 3 Step 4 Step 5
V/V
±V Hz
K
ΩΩ
Set-up LDX-
3A links
Align LDX-3A and
Transducer Null
Set LDX-3A and
Transducer Range
Final Checks
Add Offset if Required
Page 19
LDX-3A
5.0: Setting up the Transducer and LDX-3A (continued)
17
M-1633/0905
5.2: Basic Procedure
STEP 1 - Set-up LDX-3A Links
If the transducer characteristics are known, set the frequency and input resistance links as required. If the transducer characteristics are not known, the standard link settings should be used.
If your transducer is known to be outside of the standard sensitivity range, the X2 or DIV2 links will have to be used. See section 5.3.
STEP 2 - Align LDX-3A Null and Transducer Null
Any electrical offset in the LDX-3A is removed. The transducer position is adjusted so that transducer and LDX-3A nulls are aligned.
Null the LDX-3A.
1. Put the Gain link on position (7) as shown. This allows any electronics offset in the
output stage to be removed.
2. Adjust the Fine Offset control to give as near to zero output as practical.
Null the transducer.
1. Replace the Gain link to the original position.
2. Adjust the position of the transducer to give as near to zero output as practical.
This is the centre of the mechanical range.
If the transducer cannot be centered for practical reasons an offset will remain within the system. There may be noticeable interaction between Gain and Offset adjustment. This does not prevent the LDX-3A being set-up, however several iterations may be required when adjusting Gain and Offset. Please contact your supplier if guidance is required.
1 2 3 4 5 6 (7)
FREQ
LO
(ON)
HI
(OFF)
NULL
Page 20
LDX-3A
5.0: Setting up the Transducer and LDX-3A (continued)
18
M-1633/0905
STEP 3 - Setting Bi-Polar Full-scale Output
1. Move the transducer to the position where maximum LDX-3A output is required.
2. If the polarity of the output is wrong, move the Polarity (FR) links to the R position (see link diagram).
3. Move the RANGE link between 1 and 6 until the LDX-3A output is near the required value.
4. Adjust the Fine Gain control to give the required output.
5. Abi-polar output has now been set, proceed to step 5. If a uni-polar output is required, proceed to step 4.
Example: ±10Vdc is required from a ±1mm transducer. Set the transducer to the +1mm position and set the output to 10V.
If your transducer is known to be outside of the standard sensitivity range, the X2 or DIV2 links will have to be used. See section 5.3.
5.2: Basic Procedure
10K
2K
- -
¦
R
¦
|F |
- -
DIV 2
X2
X4
10K
2K
- -
¦
R
¦
|F |
- -
DIV 2
X2
X4
Normal Output Polarity Reverse Output Polarity
Page 21
LDX-3A
5.0: Setting up the Transducer and LDX-3A (continued)
19
M-1633/0905
STEP 4 - Setting Uni-polar Full-Scale Output (adding an offset)
1. Move the transducer to the null position. LDX-3A output will be 0V or 0mA.
2. Apply offset using the +VE, -VE, 5V and 10V links and then adjust the Fine Offset control to set the offset precisely.
3. Perform final checks, step 5.
Example: 0-10V is required for a ±1mm transducer. Set the transducer to give ±5V over the full range and then, with the transducer at null, add +5V offset. Adjust the Fine Offset control to give 5V. When the transducer is moved to the +1mm position, the output will be +10V.
Example: 4-20mA is required for a ±1mm transducer. Set the transducer to give ±8mAover the full range and then, with the transducer at null, add +2.5V (approx. 5mA) offset. Adjust the Fine Offset control to give +12mA. When the transducer is moved to the +1mm position, the output will be +20mA.
STEP 5 - Final Checks
Ensure that the calibration is correct by moving the transducer across the required mechanical range, checking calibration points. Fine adjustment can be made if required.
It may only be possible to set the output accurately at the two calibration points. This is due to non-linearity within the transducer.
5.2: Basic Procedure
Page 22
LDX-3A
5.0: Setting up the Transducer and LDX-3A (continued)
20
M-1633/0905
5.3: Transducer Sensitivity and the X2, X4, DIV2 Link
The LDX-3A compensates for changes in primary signal amplitude by producing an internal error signal that is the ratio between the primary and secondary signals. If the transducer output signal is too high or too low, errors may occur that can degrade the performance of the LDX-3A / transducer combination. For these transducers, the X2, X4 or DIV2 input gain link must be used.
Transducer Full Range Output
In general, transducer sensitivity is quoted as mV/V/mm Where: mV is the output of the transducer
V is the primary voltage mm is the mechanical position of the transducer from null (usually mid mechanical range).
To get the transducer Full Range Output, multiply all three together. Example: GP011-1 sensitivity is 210 mV/V/mm
GP911-1 range is ±1 mm LDX-3A primary voltage 3V
Transducer Full Range Output = 210 x 3 x 1 = 630 mV (0.63V)
Set the X2, X4, DIV2 link as shown in the table below.
Transducer Full Range Output Input Gain Link Setting
400mV FR to 2500mV FR Standard Range - Link Parked on X2 2500mV FR to 5000mV FR High Transducer Output - Link ON DIV2 150mV FR to 400mV FR Low Transducer Output - Link ON X2 55mV FR to 150mV FR Very Low Transducer Output - Link ON X4
Page 23
21
6.0: Half-Bridge only Variant
This is a Half-Bridge optimised variant of the standard product. The excitation frequency is higher (see specification) and the transducer input connector is wired to accept half-bridge transducers with standard connections.
6.1: Connecting the transducer
The 5-pin DIN plug is screwed into the case mounted socket. Transducers not fitted with a plug should be wired to the plug supplied. LVDT transducers cannot be connected to this input.
6.2: Setting up the Half-Bridge Transducer
The setting up procedure is the same as LVDT transducers. See section 5. The sensitivity of half-bridge transducers is generally lower than for LVDT types, the x2 and x4 gain position may have to be used.
Red
Blue
Yellow
+
-
Case
Half-Bridge
Cable Screen
Black
Case
Plug view
into pins
Standard Half-Bridge Gauging Probe Plug Connections
Note 1: + indicates inward movement of the tip. Note 2: The transducer body may be disconnected from the cable screen by cutting the black wire inside the connector
4
1
2
5
3
CORE
Blue
Yellow
Red
V
V
A
B
Centre Tap
V
O
Half-Bridge Electrical Connections Red and Blue Energising Yellow Signal Red and Yellow In Phase for Inward Displacement Black Transducer Body Ground
LDX-3A
M-1633/0905
Page 24
7.0: Specifications
LDX-3A
22
M-1633/0905
7.1: Electrical
Specifications are for the LDX-3Aonly. Effects due to transducers are not included. Output is set to ±10Vdc into a 1kΩ load, unless specified differently. FRO means ‘Full Range Output’.
Parameter Value Comments Power Supply
Mains Voltage 90 to 264 Vac AC Input Frequency 47 to 86 Hz AC Input Current 0.25 A(120 Vac), 0.1 A (250 Vac)
Transducer Excitation
Energising Voltage 3Vrms nominal see note 1 Energising Frequency 2.5kHz (Lo) or 5kHz (Hi) nominal
10kHz (Lo) or 13kHz (Hi)
link selectable Half-Bridge version only
Energising Current 30mA max.
Transducer Signal Input
Input Signal Range Standard 400 to 2500mV FR 6 gain ranges (applies to LVDT only)
Special
DIV2 2500 to 500mV FR x2 150 to 400mV FR x4 55 to 150mV FR
see note 2
Input Load Resistance 2, 10, 10kΩ link selectable Options Forward and Reverse
LVDT Input Half-Bridge Input
link selectable standard special plug wiring or half-bridge version only
Page 25
LDX-3A
7.0: Specifications (continued)
23
M-1633/0905
7.1: Electrical
Parameter Parameter Comments Signal Output
Output Voltage Range
Up to ±10Vdc into 1kΩ min
see notes 3 and 4 Output Residual Noise <1 mV rms Output Current
Up to ±20mA into 150Ω load
see note 5
Output Offset Coarse ±5Vdc (approx 10mA) fixed link selectable
±10Vdc (approx 10mA) fixed link selectable
Fine ±2.8Vdc (approx 5.6mA) Variable (adds to fixed offsets)
Temperature Coefficient Gain <0.01% FRO/°C Temperature Coefficient Offset <0.01% FRO/°C Warm-Up 15 minutes recommended Linearity <0.1% FRO Bandwidth (-3dB) 500Hz typical
Protection (see note 6)
Inputs and Outputs
Short circuit protected Transient and ESD Protected
Certification (see note 7)
Immunity BS EN61000-6-2:2001 Immunity for Industrial Environments Emissions BS EN61000-6-3:2001
Emission for Residential, commercial and light-industrial environments
Power Supply Safety Approvals
TUV EN60950, CE, CSA22.2, UL1950
Overall Safety Approval Compatible with IEC 61010-1 See note 8
Page 26
LDX-3A
7.0: Specifications (continued)
24
M-1633/0905
7.2: Mechanical and Connections
7.3: Environmental
Parameter Value Comments
Operating Temperature Range 0 - 60°C see note 9 Storage Temperature Range -20 - 85°C IP Rating IP40
Parameter Value Comments
Transducer 5-pin circular DIN Power Supply Cable Length 1.8m maximum, fixed length Adaptor to LDX-3A case Mains Connector IEC320 C14 C13 Mains lead supplied as required Output Signal Internal terminal block Enclosure Size 120 x 65 x 40 mm excluding connectors (See note 10)
Weight 300g approximately Material Die-cast Zinc Alloy (painted)
Power Supply Size 65 x 49 x 33 mm
Weight 120g approximately
Page 27
LDX-3A
7.0: Specifications (continued)
25
M-1633/0905
7.4: Notes
1. Primary voltage absolute value and drift are not specified. The LDX-3Auses ratiometric techniques to compensate for primary voltage drift.
2. The way in which the LDX-3A functions means a special configuration must be used for transducers outside of the
standard range. This is selectable by links. The majority of Omega LVDT transducers are within the standard range. The user manual provides details of the selection criteria.
3. LDX-3A can drive into a 1 kΩ load but this offers no advantage. 10-100 kΩ is recommended.
4. Output voltage range can be adjusted as required anywhere within this range by using a combination of gain and offset, for example, ±10Vdc, ±5Vdc, 0-5Vdc, 0-10Vdc, 4-20mA. See user manual for adjustment method.
5. Current output may be used at the same time as voltage output. Calibration of voltage and current cannot be individually adjusted.
6. Protection applies to the product when fully installed according to the user manual. During installation the top of the
enclosure has to be removed for access to user adjustments. At this time standard ESD handling precautions for electronic equipment should be followed.
7. The LDX-3A complies with the toughest electrical emissions regulations. The power supply unit alone is only compliant
with BS EN61000-6-1:2001 (Immunity for residential, commercial and light-industrial environments). However, when used as part of the LDX-3A assembly, the compliance levels are extended to BS EN61000-6-2:2001 (Immunity for industrial environments). Compliance requires installation according to user manual. The flexibility of LDX-3Ameans it can be installed in a variety of ways according to user requirements. Simple installations with short non-screened cables will meet the lesser light-industrial immunity regulations. Heavy industrial installations, especially with longer cables, will need more careful installation with screened cables.
Page 28
LDX-3A
7.0: Specifications (continued)
26
M-1633/0905
7.4: Notes
8. The power supply unit alone is certified as described. When used as part of the LDX-3A and installed as described in the user manual, then compliance with IEC 61010-1 can be achieved.
9. The LDX-3A itself has a 0 - 60°C specification. The power supply will function over a wider temperature range, but the
safety approval is only valid for 0 - 50°C.
10. The LDX-3A enclosure is smaller than the old model. This means that it will fit into all existing applications.
Mounting holes may need to be re-drilled. Extra mounting bars are available for situations where this is not possible.
Page 29
M-1633/0905
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR W
ARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under
which the product was PURCHASED,
2. Model and serial number of the
product under warranty, and
3. Repair instructions and/or
specific problems relative to the product.
FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the
product, and
3. Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2004 OMEGA ENGINEERING, INC. All rights reserved. This
document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
LDX-3A
Page 30
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at www.omega.com
M-1633/0905
TEMPERATURE
Thermocouple, RTD & Thermistor Probes, Connectors,
Panels & Assemblies
Wire: Thermocouple, RTD & Thermistor
Calibrators & Ice Point References
Recorders, Controllers & Process Monitors
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
Transducers & Strain Gauges
Load Cells & Pressure Gauges
Displacement Transducers
Instrumentation & Accessories
FLOW/LEVEL
Rotameters, Gas Mass Flowmeters & Flow Computers
Air Velocity Indicators
Turbine/Paddlewheel Systems
Totalizers & Batch Controllers
pH/CONDUCTIVITY
pH Electrodes, Testers & Accessories
Benchtop/Laboratory Meters
Controllers, Calibrators, Simulators & Pumps
Industrial pH & Conductivity Equipment
DATA ACQUISITION
Data Acquisition & Engineering Software
Communications-Based Acquisition Systems
Plug-in Cards for Apple, IBM & Compatibles
Datalogging Systems
Recorders, Printers & Plotters
HEATERS
Heating Cable
Cartridge & Strip Heaters
Immersion & Band Heaters
Flexible Heaters
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
Metering & Control Instrumentation
Refractometers
Pumps & Tubing
Air, Soil & Water Monitors
Industrial Water & Wastewater Treatment
pH, Conductivity & Dissolved Oxygen Instruments
LDX-3A
Loading...