The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves
the right to alter specifications without notice.
Page 3
1. LDC-C1 Series
Large format industrial meters with impulse counter, ratemeter and periodmeter funcons
Large format meters for long distance reading, for industrial
applicaons. Dierent formats available with 4 and 6 digits,
with 60 mm and 100 m digit height. Front keypad to access
the conguraon menu, and oponal remote keypad.
Instruments with 5 impulse counter modes (see secon
1.13.3), 2 ratemeter modes (see secon 1.13.9) and 1
periodmeter mode (see secon 1.13.9).
Highly congurable, accepts all types of sensors (NPN,
PNP, push-pull, Namur, inducve, pick-up, mechanical, TTL,
CMOS, ...) including single and bidireconal encoder signals.
Reading from 999999 to -199999 (or 9999 to -1999 for
4 digit formats) with congurable decimal point, scalable
reading with congurable mulplier (1 to 999999) and divider
factors (1 to 999999). Includes internal pull-up and pull-down
resistors, congurable trigger levels, detecon by rising or
falling edge, excitaon voltage congurable from 5 Vdc to
18 Vdc.
Output and control opons with 1, 2 and 3 relays,
transistor outputs, controls for SSR relays, isolated analog
outputs, communicaons in Modbus RTU, RS-485 ASCII and
RS232.
Sturdy metal housing with full IP65 protecon. Internal
connecons by plug-in screw clamp terminals, and output
through cable glands. Housing prepared for panel, wall and
hanging mount.
• Congurable ‘Fast access’ to selected funcons with key
‘UP’ (5) (see secon 1.13.17)
• ‘On power up’ for system protecon on ‘cold’ start-up and
control of alarm status (see secon 1.13.18)
• ‘FAST’ mode for fast counter applicaons (see secon
1.13.3)
• ‘SLOW’ mode for ratemeters applicaons at low
frequencies (slow applicaons) (see secon 1.13.9)
• Easy conguraon for most usual sensors at the ‘SnSr/
Auto’ menu (see secon 1.13.14)
• alarms with 1 or 2 setpoints, independent acvaon
and deacvaon delays, hysteresis, manual unlocking, ...
(see secon 1.13.15)
Mulple display lters, memory of maximum and minimum
reading, password protecon, 5 brightness levels.
1.1 How to use this manual
If this is the rst me you are conguring an LDB Series
large format meter, below are the steps to follow to install
and congure the instrument. Read all the manual secons
1. Idenfy the instrument format (see secon 1.4)
2. Power and signal connecons
- open the instrument (see secon 1.5)
- connect the power (see secon 1.7)
- connect the signal (see secon 1.8)
- close the instrument (see secon 1.5)
3. Congure the instrument (see secon 1.13)
- select the main funcon, and the decimal point posion
(see secon 1.13.2)
- congure the main funcon selected (see secon 1.13.2)
• counter modes from secon 1.13.3
• ratemeter and periodmeter from secon 1.13.9
- congure the sensor (see secon 1.13.13)
in order to have a full and clear view of the characteriscs
of the instrument. Do not forget to read the installaon
precauons at secon 1.17.
4. Advanced conguraon (oponal)
- congure the instrument alarms (see secon 1.13.15)
- congure the fast access (see secon 1.13.17), ‘on power
up’ (1.13.18), key ‘LE’ (1.13.19) and password (1.13.26)
5. If the instrument includes analog output (AO) or serial
communicaons (RTU, S4, S2)
- to include an opon to an instrument see secon 1.6
- to congure an installed opon, access the opon
conguraon menu (see secon 1.13.30)
- see secon 2 for informaon regarding the output and
control opons available
6. Install the instrument
- mount on panel, wall or hanging (see secon 1.16)
- adjust the brightness level according to your
environmental needs (see secon 1.13.29)
Remote keypad terminal
Slot for opon 1
Slot for opon 2
Power
Size B135 mm
Size C3 mm
Size D55 mm
Size E25 mm
Table 1 - Sizes LDB-24
Cut-out G322 mm (±1)
Cut-out F117 mm (±1)
Table 2 - Panel cut-out LDB-24
Panel cut-out
(see Table 2)
F
G
1.4.2 Format LDB-44
Power
Opon 2Opon 1
A
Size A542 mm
B
Size B166 mm
Size C3 mm
Size D55 mm
Size E25 mm
Table 3 - Sizes LDB-44
Remote keypad
Signal
Cut-out G524 mm (±1)
Cut-out F148 mm (±1)
Table 4 - Panel cut-out LDB-44
CDE
Panel cut-out
(see Table 4)
F
G
4
Page 6
1.4.3 Format LDB-26
A
Size A436 mm
Power
Opon 3 Opon 2
Remote keypad
B
Size B135 mm
Size C3 mm
Size D55 mm
Size E25 mm
Table 5 - Sizes LDB-26
SignalOpon 1
Cut-out G418 mm (±1)
CDE
Cable glands
Input signal terminal
Remote keypad terminal
Slot for opon 1
Slot for opon 2
Slot for opon 3
Power
Cut-out F117 mm (±1)
Table 6 - Panel cut-out LDB-26
Panel cut-out
(see Table 6)
F
G
1.4.4 Format LDB-46
Power
Opon 3Opon 2
A
Size A740 mm
B
Size B166 mm
Size C3 mm
Size D55 mm
Size E25 mm
Remote keypad
Table 7 - Sizes
SignalOpon 1
Cut-out G722 mm (±1)
LDB-46
Cut-out F148 mm (±1)
Table 8 - Panel cut-out LDB-46
CDE
Panel cut-out
(see Table 8)
F
G
5
Page 7
1.5 To access the instrument
To open the housing, remove the screws from the back
cover. With each screw there is a metal washer and a plasc
washer. Once the screws are out, remove the back cover.
The gure below shows the instrument internal structure for
a LDB-26 format. It shows the locaon of the 3 slots for opon
al output and control modules, the power terminal and the
input signal terminal.
Waterght seal
Female turret
Power
Slot for opon 3Back cover
To close the instrument, place the back cover, the screws,
the metal washer and the plasc washer. The plasc washer
is in contact with the back cover. Conrm that the screws are
correctly turning inside the internal female screws.
-
To ensure a correct IP65 protecon ghten the back cover
screws with a strength between 30 and 40 Ncm, with the
help of a dynamometer screwdriver.
Slot for opon 2
Slot for opon 1
Remote keypad terminal
Input signal terminal
Screw
Metal washer
Plasc washer
1.6 Modular system
Large format meters from LDB Series are designed with an
internal modular architecture. The output and control
modules are independent and can be installed by accessing
the internal circuits of the instrument, and connecng the
module to the connecon jumpers of the selected slot.
Output and control module
Slot 3
Tie base
Cable e
Slot 2
(2)
(1)
Module pins
Risk of electric shock. Removing the back
cover will grant access to the internal
circuits of the instrument. Operaon must
be performed by qualied personnel only.
Each module is provided with a cable e to x the
module to the e base. The input signal modules denes the
instrument funcon and are exchangeable, switching a
temperature meter to an impulse counter only by replacing
the input signal module.
See secon 2. for informaon regarding the output and
control opons available
To install an output and control module
(1) insert the ‘module pins’ into the
Slot 1
‘connecon jumpers’ in one of the
free slots
(2) place the ‘cable e’ into the ‘e
base’ and embrace the ‘module’
rmly, unl it is xed
Connecon jumpers
6
Page 8
1.7 Power connecons and protecve earth
1. Unscrew the screws from the back cover and remove the
back cover (see secon 1.5).
2. Pass the power cable through the power cable gland
(see secon 1.4).
3. Prepare the power cables so that the earth wire is 20 cm
longer than the other cables (see Figure 1).
Phase (+)
Neutral (-)
Earth
Figure 1 - Longer earth wire
20 cm
4. Connect the earth wire to the internal xed screw ‘PE’
(see Figure 2) located at the inside of the back cover. The
instrument internally connects the back cover metallic
‘PE’ internal xed screw
Power cable gland
structure with the front metallic structure through an
internal green-yellow cable. (doed cable at Figure 3).
5. Connect phase and neutral (in AC power) or posive and
negave (in DC power) to the internal power terminal.
6. The connecons label aached to the outside of the
instrument has some free space le to write the color or
local code for each cable.
7. To comply with security regulaon 61010-1, add to the
power line a protecon fuse acng as a disconnecon
element, easily accessible to the operator and idened
as a protecon device.
Power ‘H’ 500 mA me-lag fuse
Power ‘L’ 1000 mA me-lag fuse
Power Terminal
PE
(orange)
N
L
Screws
Figure 2 - Locaon of the internal ‘PE’ xed screw and power cable gland
1.8 Input signal connecons
1. Unscrew the screws from the back cover and remove the
back cover (see secon 1.5).
2. Locate the input signal terminal (see secon 1.4).
3. Pass the signal cable through the signal cable gland
(see secon 1.4).
4. Connect the input signal cables (see Figure 4).
5. The connecons label aached to the outside of the
instrument has some free space le to write the color or
local code for each cable.
1.9 Connecons for remote keypad
The 4 pin terminal located beside
the input signal module allows to
replicate a remote version of the
front keypad. Connect 4 cables
for front keys ‘SQ’ (<), ‘UP’ (5)
and ‘LE’ (3) and for the common. Pass these cables through
the ‘remote keypad’ cable gland
(see secon 1.4)
.
GND
SQ
UP
LE
Figure 3 - Power connecons
Reset
Input Signal
Channel A
Channel B
Input ‘A’ for impulses
Input ‘B’ for impulses
ResetReset connecon
VexcExcitaon voltage
0VCommon
Figure 4 - Signal connecons
PE
Channel A
Vexc
ABRst
0
fuse
Vexc.Channel B
0V
V
7
Page 9
1.10 Technical specicaons
Digits
number of digits 4 or 6 (see Table 9)
digit 7 segments
view angle 120º
color red or green
digit height (see Table 9)
Reading
max., min. (see Table 9)
decimal point congurable
overrange / underrange congurable (ash, reset or
preset) (see secon 1.13.23)
display refresh 15 refresh / second
memory yes, recovers the last counter
value aer power loss
Input signal
signals accepted NPN, PNP, Namur, pick-up,
TTL, inducve, mechanical,
quadrature, ...
vdc max. at input ±30 Vdc
input impedance 2.4 K with pull-up or pull-down
470 K without pull resistances
frequency max./min.
for counter modes
(see Table 12)
for ratemeter modes (see Table 13)for periodmeter modes (see
Consumpon (without opons installed)3 W5.25 W3.5 W5.5 W
Consumpon (with opons installed)5 W6.75 W5.5 W7 W
Weight2200 gr.2500 gr.3500 gr.4500 gr.
Table 9 - Technical specicaons associated to format
8
Page 10
1.11 Funcons included
CountersModeFrequencySecon
‘FAST’ mode
max. 250 KHz
Counter
normal mode
Counter + inhibion
Counter + control Add / Subtract
max. 9 KHz
max. 9 KHz
max. 9 KHz
1.13.3
Dierenal counter
mode x1
Quadrature
mode x2
counter
mode x4
Table 12 - Maximum frequency for counter modes
max. 9 KHz
max. 17 KHz
max. 16 KHz
max. 11 KHz
RatemeterModeFrequencySecon
normal mode
Ratemeter
‘SLOW’ mode
mode x1
Quadrature
mode x2
Ratemeter
mode x4
Table 13 - Maximum and minimum frequency for ratemeter
modes
max
. 500 KHz
max. 200 Hz
min. 1 mHz
max. 17 KHz
max. 16 KHz
max. 11 KHz
1.13.9
PeriodmeterModeFrequencySecon
normal mode
Periodmeter
‘SLOW’ mode
max. 500 KHz
max. 200 Hz
min. 1 mHz
1.13.9
(1000 sec.)
Table 14 - Maximum and minimum frequency for periodmeter
modes
Funcons includedSecon
Fast access menuyes, congurable1.13.17
‘SLOW’ modeyes, for slow frequencies1.13.9
‘FAST’ modeyes, for fast counng1.13.3
Mulplier and divider from 1 to 999999
Reset
congurable : front, external
and linked to alarm
1.13.3
1.13.9
1.13.3
1.13.15
1.13.19
Presetyes1.13.3
Trigger levelcongurable1.13.13
Sensor seleconby menu1.13.13
Cycle counteryes1.13.15
Retenon memoryyes, recovers with power1.10
‘On Power Up’yes
1.13.18
Excitaon voltagecongurable1.13.13
Average lterrecursive
1.13.3
1.13.9
Memorymax., min., cycles1.13.17
Passwordconguraon locked1.13.26
simple or double setpoint
acvaon delays
Alarms
deacvaon delays
hysteresis
1.13.15
inverted relays
locked alarms
Brightness
Table 10 - Funcons included
congurable, 5 levels1.13.29
1.12 Messages and errors
Error messages related to the local instrument are shown on
display, in ash mode (see Table 11). Examples given are for
instrument with 6 digit formats.
Messages and errors on display
‘Err.1’incorrect password.
‘Err.2’at ‘oPt.X’ menu entry. Installed module is not
recognized.
‘Err.W’‘Watchdog’ error
‘999999’+ ashing mode. Reading is in overrange.
‘-199999’ + ashing mode. Reading is in underrange.
Table 11 - Messages and error codes for local instrument
9
Page 11
1.13 Conguraon
1.13.1 How to operate the menus
The instrument has two menus accessible to the user :
‘Conguraon menu’ (key ‘SQ’) (<)
‘Fast access’ menu (key ‘UP’) (5)
Conguraon menu
The ‘conguraon menu’ modies the conguraon
parameters to adapt the instrument to the applicaon
needs. To access the ‘conguraon menu’ press for 1
second the ‘SQ’ (<) key. This access can be blocked by
acvang the ‘Password’ (‘PASS’) funcon. While operang the
‘conguraon menu’, the alarm status is ‘hold’ to the
status it had before accessing the menu, and the output and
control modules remain in ‘error’ state. When leaving the
‘conguraon menu’, the instrument applies a system
reset, followed by a brief disconnecon of the alarms and the
output and control modules. Funconality is then recovered.
For a detailed explanaon on the ‘conguraon menu’
see the following secons, and for a full view of the
‘conguraon menu’ see secon 1.14.
Key ‘LE’ (3) - press the ‘LE’ (3) key to acvate the
congured special funcons associated to this key. Inside
the menu, the ‘LE’ (5) acts as an ‘ESCAPE’. It leaves the
selected menu level and eventually, by leaving all menu
levels, it leaves from the conguraon menu. Then changes
are applied and the instrument is back to normal funcon.
When entering a numerical value, it selects the acve digit,
and the value is then modied by key ‘UP’ (5).
‘Rollback’
Aer 30 seconds without interacon from the operator, the
instrument will rollback and leave the ‘conguraon menu’
or the ‘fast access’ menu. All changes will be discarded.
Instruments with 4 and 6 digits
The conguraon menus included in this document show
values for a 6 digit instrument. In case of 4 digit instruments,
note that maximum reading values should be 9999 instead
of 999999 to 9999 and minimum reading values should be
-1999 instead of -199999.
‘Fast access’ menu
The ‘fast access’ menu is an operator congurable menu,
providing fast and direct access to the most usual funcons
of the instrument with a single key pad stroke. Press key ‘UP’
(5) to access this menu.
See secon 1.13.17 for a list of selectable funcons for the
‘fast access’ menu in this instrument. The ‘Password’ (‘PASS’)
funcon does not block access to this menu. Accessing and
modifying parameters in the ‘fast access’ menu does not
interfere with the normal funconality of the instrument,
and it does not generate any system reset when validang
the changes.
Operang with the front keypad inside the menus
Key ‘SQ’ (<) - press the ‘SQ’ (<) key for 1 second to
access the ‘conguraon menu’. Inside the menu, the ‘SQ’
(<) key acts as an ‘ENTER’. It enters into the menu opon se
-
lected, and when entering a numerical value, it validates the
number.
Key ‘UP’ (5) - press the ‘UP’ (5) key to access the ‘fast
access’ menu. Inside the menu,the ‘UP’ (5) key sequen-
ally moves through the available parameters and menu entries. When entering a numerical value, it modies the digit
selected by increasing its value to 0, 1, 2, 3, 4, 5, 6, 7, 8, 9.
Example of operaon inside
(1)
(6)
(6)
(3)
(2)
(5)
(3)
(5)
(3)
(5)
(3)
(5)
(3)
the ‘conguraon menu’.
(4)
1. The (<) key enters into the
‘conguraon menu’.
(4)
2. The (<) key enters into the
(4)
‘InP’ menu.
3. The (5) key moves through
(4)
the menu opons.
4. The (<) key selects the
desired range and returns
(3)
(3)
to the ‘InP’ menu.
5. The (3) key leaves the
actual menu level and
moves to the previous
menu level.
6. The (3) key leaves the
‘conguraon menu’.
Changes are applied and
Figure 5 - Example of operaon inside the ‘conguraon menu’
saved at this moment.
10
Page 12
1.13.2 Inial set-up
Press ‘SQ’ (<) for 1 second to access the ‘conguraon menu’. For a descripon on how to operate inside the menus
see secon 1.13.1. For a full vision of the ‘conguraon menu’
structure see secon 1.14.
Main funcon
Counter
Quadrature counter
Counter + inhibion
Counter + control add / subtract
Dierenal counter
Ratemeter
Quadrature Ratemeter
Periodmeter
To congure the inial set up, select the main funcon for
the instrument, the decimal point posion, congure the
main funcon selected and congure the sensor.
Enter the ‘Main funcon’ (‘Func’) menu and select the
desired funcon, from the 5 counng modes, 2 ratemeter
modes and the periodmeter mode available.
• select ‘Counter’ (‘cn. 1’) for a standard impulse counter.
Impulses are received at channel A. Channel B is disabled.
• select ‘Counter quadrature’ (‘cnq.2
’) for a
quadrature counter. Impulses are received at channel A and B, in
quadrature format (typical for bidireconal encoders).
• select ‘Counter + inhibion’ (‘cnI.3’) for a counter with
an external control to inhibit the counng. Impulses are
received at channel A. The state of channel B controls de
inhibion funcon.
• select ‘Counter + control add /subtract’ (‘cnc.4’) for
a counter with an external control to add or subtract
impulses received. Impulses received at channel A. The
state of channel B controls de add or subtract funcon.
Decimal point
The next menu entry is the conguraon parameters for the ‘Main funcon’ (‘Func’) selected. Conguraon parameters are slightly
dierent for each ‘main funcon’. All possible conguraon menus are explained, ‘cnF.1’ to ‘cnF.7’, one for each ‘main funcon’. Only
the conguraon menu for the ‘main funcon’ selected is visible
on the instrument.
• select ‘Counter dierenal’ (‘cnd.5’) for a counter
where impulses received at channel A add and impulses
received at channel B subtract.
• select ‘Ratemeter’ (‘rt.6’) for a standard ratemeter.
Impulses are received at channel A. Channel B is disabled.
• select ‘Ratemeter quadrature’ (‘rtq.7’) for a quadrature
ratemeter. Impulses are received at channel A and B, in
quadrature format (typical for bidireconal encoders.
• select ‘Periodmeter’ (‘Prd.8’) for a standard periodmeter.
Impulses are received at channel A. Channel B is disabled.
At the ‘Decimal point’ (‘dP’) parameter, select the decimal
point posion. Move the decimal point with the ‘LE’ (3) key.
Congure the funcon mode selected (‘cnF.2’ to ‘cnF.8’) at
the next menu entry (‘cnF.1’ to ‘cnF.8’). See secons 1.13.3
to 1.13.12.
Congure the sensor at the ‘SnSr’ menu. See secon 1.13.13.
11
Page 13
1.13.3 Counter modes descripon
The instrument oers 5 selectable impulse counter modes.
Each mode has 2 independent input channels ‘A’ and ‘B’. Each
impulse counter mode has a specic funcon assigned to
channel ‘B’.
• Standard counter (‘cn.1’) (see secon 1.13.4) counts
impulses received at channel ‘A’. This counter has an
oponal ‘FAST’ mode to count high frequencies up to
250 KHz. The ‘FAST’ mode detects impulses on the
rising edge of the impulse. The rst edge received (rising or
falling) aer the instrument start up (aer power loss or
conguraon change) will not be counted as a valid
impulse, as it is needed for internal inializaon.
• Quadrature counter (‘cnq.2’) (see secon 1.13.5) counts
quadrature impulses received at channels ‘A’ and ‘B’, (for example from a bidireconal encoder). The counter
increases or decreases depending on the sense of turn of
the encoder.
• Counter with inhibit (‘cnI.3’) (see secon 1.13.6) counts
impulses received at channel ‘A’ if channel ‘B’ is inacve.
Acvate channel ‘B’ to inhibit the counng of impulses
received at channel ‘A’.
• Counter with add / subtract control (‘cnc.4’) (see secon
1.13.7) increases the counter with impulses received at
channel ‘A’ if channel ‘B’ is acve. Deacvate channel ‘B’ to
decrease the counter with impulses received at channel ‘A’.
• Dierenal counter (‘cnd.5’) (see secon 1.13.8)
increases the counter with impulses receive at channel ‘A’
and decreases the counter with impulses received at
channel ‘B’.
All counter modes have scalable reading through
mulplier (1 to 999999) and divider (1 to 999999)
parameters, congurable preset value (preset value loads on
display when ‘reset’ funcon acvates), congurable reset
funcon and accessible from external terminal, front keypad
or at alarm acvaon. Alarms with independent acvaon
and deacvaon delays and funcons to load ‘preset’ or ‘0’
to generate cycles of counng from ‘preset’ to ‘alarm
setpoint’ and back. The number of cycles is accessible.
In case of power loss, the instrument recovers the last
conguraon and last counted value.
1.13.4 Standard counter ‘cn.1’ conguraon menu
Conguraon menu for mode ‘counter’ (‘cn.1’). Total
impulses received are mulplied by the value of the
‘mulplier’ (‘MuLt’) parameter and divided by the ‘divider’
(‘dIV’) parameter. Result is shown on the display.
• set the ‘Mulplier’ (‘MuLt’) parameter from 1 to 999999.
• set the ‘Divider’ (‘dIV’) parameter from 1 to 999999.
• set the ‘Preset’ (‘PrSt’) parameter from -199999 to 999999.
Acvate the reset to load the preset value on display.
• at the (‘ModE’) parameter select ‘uP’ to count upwards
(impulses received add) or select ‘doWn’ to count
downwards (impulses received subtract).
• at the ‘FAST’ (‘FASt’) parameter select ‘on’ to acvate the
fast mode. See secon 1.13.3 for more informaon.
Counter
conguraon
Mulplier
Divider
Preset
Mode
‘FAST’ mode
Mulplier
1 to 999999
Divider
1 to 999999
Preset value
-199999 to 999999
Increasing
Decreasing
1.13.5 Quadrature counter ‘cnq.2’ conguraon menu
Conguraon menu for mode ‘quadrature counter’ (‘cnq.2’).
Total impulses received are mulplied by the value of the
‘mulplier’ (‘MuLt’) parameter and divided by the ‘divider’
(‘dIV’) parameter. Result is shown on the display..
• set the ‘Mulplier’ (‘MuLt’) parameter from 1 to 999999.
• set the ‘Divider’ (‘dIV’) parameter from 1 to 999999.
• set the ‘Preset’ (‘PrSt’) parameter from -199999 to 999999.
Acvate the reset to load the preset value on display.
• at the ‘Quadrature edges’ (‘q.124’) parameter select the
number of edges to consider. Select ‘1--1’ for 1 impulse
per quadrature cycle, ‘1--2’ for 2 impulses per quadrature
cycle, ‘1--4’ for 4 impulses per quadrature cycle.
Quadrature
counter conf.
Mulplier
Divider
Preset
Quadrature
edges
Mulplier
1 to 999999
Divider
1 to 999999
Preset value
-199999 to 999999
1 imp. per cycle
2 imp. per cycle
4 imp. per cycle
12
Page 14
1.13.6 Counter + inhibion ‘cn.3’ conguraon menu
Conguraon menu for mode ‘counter + control inhibion’
(‘cnI.3’). Total impulses received are mulplied by the value
of the ‘mulplier’ (‘MuLt’) parameter and divided by the
‘divider’ (‘dIV’) parameter. Result is shown on the display.
• set the ‘Mulplier’ (‘MuLt’) parameter from 1 to 999999.
• set the ‘Divider’ (‘dIV’) parameter from 1 to 999999.
• set the ‘Preset’ (‘PrSt’) parameter from -199999 to 999999.
Acvate the reset to load the preset value on display.
• at the ‘Mode’ (‘ModE’) parameter select ‘uP’ to count
upwards (impulses increase the counter) or select ‘doWn’ to
count downwards (impulses decrease the counter).
• at the ‘inhibion’ (‘Inh’) parameter select ‘on_h’ to inhibit
the counter when channel ‘B’ is acve (logical state ‘1’) or
select ‘on_0’ to inhibit the counter when channel ‘B’ is
Counter +
inhibion conf.
Mulplier
Divider
Preset
Mode
Inhibion
Mulplier
1 to 999999
Divider
1 to 999999
Preset value
-199999 to 999999
Increasing
Decreasing
Inhibits if channel
‘B’ to high
Inhibits if channel
‘B’ to low
inacve (logical state ‘0’).
1.13.7 Counter + control add / subtract ‘cnc.4’ conguraon menu
Conguraon menu for mode ‘counter + control add/subtract’ (‘cnc.4’). Total impulses received are mulplied by
the value of the ‘mulplier’ (‘MuLt’) parameter and divided
by the ‘divider’ (‘dIV’) parameter. Result is shown on the
display.
• set the ‘Mulplier’ (‘MuLt’) parameter from 1 to 999999.
• set the ‘Divider’ (‘dIV’) parameter from 1 to 999999.
• set the ‘Preset’ (‘PrSt’) parameter from -199999 to 999999.
Acvate the reset to load the preset value on display.
• at the ‘Control A/S’ (‘Add’) parameter select ‘on_h’ increase
the counter with impulses received at channel ‘A’ when
channel ‘B’ is acve (logical state ‘1’) or select ‘on_0’ to
decrease the counter with impulses received at channel ‘A’
when channel ‘B’ is inacve (logical state ‘0’).
Counter + control
add / subtract conf.
Mulplier
Divider
Preset
Control A/S
Mulplier
1 to 999999
Divider
1 to 999999
Preset value
-199999 to 999999
Adds if channel ‘B’
to high
Subtracts if channel ‘B’ to low
1.13.8 Dierenal counter ‘cnd.5’ conguraon menu
Conguraon menu for mode ‘dierenal counter’ (‘cnd.5’).
Total impulses received are mulplied by the value of the
‘mulplier’ (‘MuLt’) parameter and divided by the ‘divider’
(‘dIV’) parameter. Result is shown on the display.
• set the ‘Mulplier’ (‘MuLt’) parameter from 1 to 999999.
• set the ‘Divider’ (‘dIV’) parameter from 1 to 999999.
• set the ‘Preset’ (‘PrSt’) parameter from -199999 to 999999.
Acvate the reset to load the preset value on display.
Impulses received on channel ‘A’ add to the counter.
Impulses received on channel ‘B’ subtract from the counter.
Dierenal
counter conf.
Mulplier
Divider
Preset
Mulplier
1 to 999999
Divider
1 to 999999
Preset value
-199999 to 999999
13
Page 15
1.13.9 Ratemeter and periodmeter modes descripon
The instrument oers 2 selectable ratemeter modes and
1 periodmeter mode. Ratemeters provide a reading
proporonal to the frequency measured, while reading at
periodmeters is proporonal to the me between impulses
• Standard ratemeter (‘rt.6’) (see secon 1.13.10) to read
speed values from impulse frequency signals.
• Quadrature ratemeter (‘rtq.7’) (see secon 1.13.11)
to read speed values and the turning sense of the axis,
from two quadrature frequency signals, such as those
provided by a bidireconal encoder. Speed is posive
when the quadrature turns clockwise and negave when
turns counterclockwise.
• Standard periodmeter (‘Prd.8’) (see secon 1.13.12)
to read me between impulses. For applicaons with
long periods (long me between impulses) the ‘SLOW’
mode oers the best possible response me for each
applicaon.
All modes have scalable reading through mulplier (1
to 999999) and divider (1 to 999999) parameters, and a
congurable me window (‘GAtE’) to adjust the measure
refresh me.
• ‘SLOW’ mode
The ‘SLOW’ mode is an oponal mode for very slow
applicaons. Applies to ratemeter and periodmeter modes.
The ‘SLOW’ mode accepts measures frequencies down to
1 mHz (0,001 Hz or 1000 seconds between impulses), and is
funconal up to 200 Hz.
The ‘SLOW’ mode oers the fastest response me for any
given applicaon, calculang the frequency and period
values each me a new impulse is received.
At the The ‘Max. waing me’ parameter set a value
between 1 and 1000 seconds. Select ‘0’ to disable the
‘SLOW’ mode. If me between impulses is higher than the
congured value, the instrument assumes that the signal has
stopped and forces the reading to ‘0’ (both in ratemeters
and periodmeters). The ‘GATE’ parameter has no eect if the
‘SLOW’ mode is acve.
At the The ‘Number of pulses’ parameter set a value
between 1 and 32. This paremeter denes the number of
pulses that will be taken to calculate the period.
In ‘Quadrature ratemeter’ (‘rtq.7’) mode, the ‘SLOW’
mode calculates the frequency between two consecuve
impulses received at channel ‘A’, and calculates the turning
direcon by comparing impulses at channel ‘A’ with the state of
channel ‘B’. The ‘Quadrature edges’ parameter is xed to
‘1--1’.
Applicaon: to measure the speed of the propeller on
ships, using two inducve sensors in quadrature, at low
revoluons per minute.
1.13.10 Ratemeter ‘rt.6’ conguraon menu
Mulplier
Ratemeter
conf.
Mulplier
Divider
Time window
‘SLOW’ mode
Average lter
Filter strength
(0 = disabled)
1 to 999999
Divider
1 to 999999
Seconds
Max. waing me
Number of pulses
Conguraon menu for mode ‘ratemeter’ (‘rt.6’).
Measured frequency is mulplied by the value of the
‘mulplier’ (‘MuLt’) parameter and divided by the
‘divider’ (‘dIV’) parameter. Result is shown on the display. The
measure is updated on display as congured on the ‘GAtE’
parameter.
• set the ‘Mulplier’ (‘MuLt’) parameter from 1 to 999999.
• set the ‘Divider’ (‘dIV’) parameter from 1 to 999999.
• select the ‘Time window’ (‘GAtE’) parameter at 0.5, 1.0, 2.0,
4.0, 8.0 or 16.0 seconds. This parameter denes how oen
the measure will be refreshed on display. This parameter
has no eect if ‘SLOW’ mode is acve.
• for slow frequencies acvate the ‘SLoW’ parameter
conguring the ‘tIME’ parameter between 1 and 1000
seconds. See 1.13.9 for more informaon. Congure the
‘nuMb’ parameter between 1 and 32 impulses.
• if reading is unstable, set the ‘Average lter’ (‘AVr’)
parameter to ‘on’ to acvate a recursive lter on the display,
and congure the lter strength from 0.0 to 99.9. The lter
is stronger for higher values. Strong lters make readings
more stable and changes slower to update. Set ‘0’ to disable
the lter.
14
Page 16
1.13.11 Quadrature ratemeter ‘rtq.7’ conguraon menu
Quadrature
ratemeter conf.
Mulplier
Divider
Time window
Quadrature
edges
Mulplier
1 to 999999
Divider
1 to 999999
Seconds
1 imp. per cycle
2 imp. per cycle
4 imp. per cycle
Conguraon menu for mode ‘quadrature ratemeter’
(‘rtq.7’). Measured frequency is mulplied by the value
of the ‘mulplier’ (‘MuLt’) parameter and divided by the
‘divider’ (‘dIV’) parameter. Result is shown on the display.
The measure is updated on display as congured on the
‘GAtE’ parameter.
• set the ‘Mulplier’ (‘MuLt’) parameter from 1 to 999999.
• set the ‘Divider’ (‘dIV’) parameter from 1 to 999999.
• select the ‘Time window’ (‘GAtE’) parameter at 0.5, 1.0, 2.0,
4.0, 8.0 or 16.0 seconds. This parameter denes how oen
the measure will be refreshed on display. This parameter
has no eect if ‘SLOW’ mode is acve.
• at the ‘Quadrature edges’ (‘q.124’) parameter select the
number of edges to consider. Select ‘1--1’ for 1 impulse
per quadrature cycle, ‘1--2’ for 2 impulses per quadrature
cycle, ‘1--4’ for 4 impulses per quadrature cycle.
• for slow frequencies acvate the ‘SLoW’ parameter
conguring the ‘tIME’ parameter between 1 and 1000
seconds. See 1.13.9 for more informaon. Congure the
‘nuMb’ parameter between 1 and 32 impulses.
‘SLOW’ mode
Average lter
Filter strength
(0 = disabled)
Max. waing me
Number of pulses
1.13.12 Periodmeter ‘Prd.8’ conguraon menu
Conguraon menu for mode ‘periodmeter’ (‘Prd.8’).
Measured period is mulplied by the value of the ‘mulplier’ (‘MuLt’) parameter and divided by the ‘divider’ (‘dIV’)
parameter. Result is shown on the display. The measure is
updated on display as congured on the ‘GAtE’ parameter.
Periodmeter
conf.
Mulplier
Divider
Time window
Mulplier
1 to 999999
Divider
1 to 999999
Seconds
• if reading is unstable, set the ‘Average lter’ (‘AVr’)
parameter to ‘on’ to acvate a recursive lter on the display,
and congure the lter strength from 0.0 to 99.9. The lter is
stronger for higher values. Strong lters make readings more
stable and changes slower to update. Set ‘0’ to disable the
lter.
• set the ‘Mulplier’ (‘MuLt’) parameter from 1 to 999999.
• set the ‘Divider’ (‘dIV’) parameter from 1 to 999999.
• select the ‘Time window’ (‘GAtE’) parameter at 0.5, 1.0, 2.0,
4.0, 8.0 or 16.0 seconds. This parameter denes how oen
the measure will be refreshed on display. This parameter
has no eect if ‘SLOW’ mode is acve.
• for slow frequencies acvate the ‘SLoW’ parameter
conguring the ‘tIME’ parameter between 1 and 1000
seconds. See 1.13.9 for more informaon. Congure the
‘nuMb’ parameter between 1 and 32 impulses.
‘SLOW’ mode
Average lter
Filter strength
(0 = disabled)
Max. waing me
Number of pulses
• if reading is unstable, set the ‘Average lter’ (‘AVr’)
parameter to ‘on’ to acvate a recursive lter on the display,
and congure the lter strength from 0.0 to 99.9. The lter is
stronger for higher values. Strong lters make readings more
stable and changes slower to update. Set ‘0’ to disable the
lter.
15
Page 17
1.13.13 Accepted sensors and signals
The instrument accepts the usual sensors and impulse
signals, and provides a list for the operator to choose his
sensor. It also allows to congure a wide range of parameters
to adapt the reading to other non usual sensors and signals.
The directly selectable sensors are:
• Mechanical contact (free potenal contact)
• Namur
• NPN and PNP, 2 or 3 wires
• Push-pull
• TTL and CMOS
• Pickup
• AC voltage signals up to 30 Vp (inducve)
The congurable parameters are:
• Pull-up / pull-down resistors can be enabled or disabled
independently for channel ‘A’, channel ‘B’ and the reset
channel.
• The trigger level can be manually congured to any value
between 0.0 V and 3.9 V. While modifying the trigger level
parameter, the two segments to the le show the actual
state ‘1’ or ‘0’ for channels ‘A’ and ‘B’. This informaon
helps to easily idenfy the real trigger level. When the
le segments switch from ‘high’ to ‘low’ means that the
trigger level for channels ‘A’ and ‘B’ has been reached. The
same trigger level applies to channels ‘A’, ‘B’ and reset.
• Acvaon by rising or falling edges can be congured.
Channels ‘A’ and ‘B’ share the same conguraon. Reset
has its own independent conguraon.
• Excitaon voltage can be congured to 5 V, 9 V, 15 V o 18 V,
or even power o the excitaon voltage.
• An anrrebound lter is congurable, by seng a me
between 0 and 1000 mSeconds. When an impulse is
received,the instrument inhibits the counng of new
impulses for the me congured.
See Table 15 below for a list of directly selectable sensors,
the associated conguraon parameters for each one and
connecons. Parameters can be later on modied through
the conguraon menu.
For signal connecons and reset connecons, see secon
1.8.
SensorConnecons
(0 signal Vexc)
Mechanical
0 V ‘A’pull-upno100 mSeg. 2,5 Vdc
PullsVexc.
Anrrebound
lter
Trigger
level
contact
Namur
NPN 2 wires
NPN 3 wires
PNP 2 wires
PNP 3 wires
Push-pull
TTL
CMOS
‘A’ Vexcpull-down9 Vdcno3,0 Vdc
0 V ‘A’pull-up18 Vdcno2,5 Vdc
0 V ‘A’ Vexcpull-up18 Vdcno2,5 Vdc
0 V ‘A’pull-down18 Vdcno2,5 Vdc
0 V ‘A’ Vexcpull-down 18 Vdcno2,5 Vdc
0 V ‘A’ Vexcno18 Vdcno2,5 Vdc
0 V ‘A’no5 Vdcno2,5 Vdc
Pick-up
AC<30 Vp
Inducve
Table 15 - Parameters congured and connecons for listed sensors. Channel ‘B’ applies the same connecons as indicated for channel ‘A’
0 V ‘A’nonono0 Vdc
channel ‘A’
channel ‘B’
Level ‘1’
Level ‘0’
‘Trigger Sense’ leds
Vdc
‘1’
‘0’
Figure 6 - ‘Trigger sense’ for detecon of trigger level
‘trigger’ level
‘Trigger’ level
at 1.8 Vdc
t
t
16
Page 18
1.13.14 Sensor conguraon menu
Mechanical contact
Sensor
Automac
conguraon
Channel A pulls
Channel B pulls
Reset pulls
Trigger level
Channel A
acvaon
Reset acva-
on
Namur
NPN 2 wires
NPN 3 wires
PNP 2 wires
PNP 3 wires
Push pull
TTL
CMOS
Pick-up
Inducve
Vac <30 V
No pulls
Pull up
Pull down
No pulls
Pull up
Pull down
No pulls
Pull up
Pull down
0.0 Vdc to 3.9 Vdc
rising edge
falling edge
rising edge
falling edge
The ‘Sensor’ (‘SnSr’) conguraon menu contains all
parameters related to the detecon of the input signal,
excitaon voltage and trigger levels.
• enter the ‘Automac conguraon’ (‘Auto’) menu to
select a standard sensor from the list. The instrument
will congure the appropriate parameters for the sensor
selected, as indicated at Table 15. If the instrument does
not detect the signal with this conguraon, the following
parameters can be manually recongured.
• at ‘Channel A pulls’ (‘PuL.A’) select ‘P.uP’ to acvate the
internal pull-up resistors needed for NPN sensors, select
‘P.dn’ to acvate the internal pull-down resistors needed
for PNP sensors, or select ‘no’ to disable the pull resistors.
Selecng pull-up or pull-down resistors sets the trigger
level to 2,5 Vdc.
• at ‘Channel B pulls’ (‘PuL.b’) applies the same as previous
entry but for channel B.
• ‘Reset pulls’ (‘PuL.r’) - applies the same as previous entry
but for the reset channel.
• at ‘Trigger level’ (‘trIG’) congure the trigger level to
detect the impulses. Signals levels above the trigger level
are ‘1’ signals, and signal levels below trigger level are
‘0’ signals. Trigger level is selectable between 0,0 and
3,9 Vdc. Channels ‘A’ and ‘B’ share the same trigger level.
Trigger level for reset channel is xed at 2.5 Vdc. Ver
cal leds to the le are part of the ‘trigger sense’ ulity to
help locate the real trigger level for the actual signal. See
secon 1.13.13 for more informaon.
• at ‘Channel A acvaon’ (‘Act.A’) congure the
acvaon of channel ‘A’ by rising edge (‘on_h’) or falling
edge (‘on_0’)
• at ‘Reset acvaon’ (‘Act.r’) congure the acvaon of
reset by rising edge (‘on_h’) or falling edge (‘on_0’)
• at ‘Excitaon voltage’ (‘V.EXc’) congure the value for the
excitaon voltage to 5 Vdc, 9 Vdc, 15 Vdc or 18 Vdc. Select
‘no’ to disable the excitaon voltage.
• at ‘Anrrebound’ (‘rbnd’) congure the lter that prevents
mechanical rebounds to be accepted as real impulses.
Congure a value between 0 and 1000 mSeconds. When an
impulse is received,the instrument inhibits the counng of
new impulses for the me congured. When me is over,
the next impulse is accepted and the lter acvates again.
Recommended value is 100 mSeconds for a mechanical
contact.
-
Excitaon
voltage
0 to 1000 mSec.
Anrrebound
17
Page 19
1.13.15 Alarms
The instrument manages 3 independent internal alarms,
each one controlling the acvaon of an oponal relay,
transistor or control SSR output.
Oponal modules (see secon 2) are installed at the free
slots inside the instrument (see secon 1.4). LDB-24 and
LDB-44 formats have 2 free slots for output and control
modules, while LDB-26 and LDB-46 formats have 3 free
slots for output and control modules.
The instrument has 3 front leds that reect the state of
the 3 internal alarms. These leds are only for local help
during installaon, as they are not appropriate for long
distance reading.
Each alarm controls the acvaon of the relay,
transistor or control SSR installed on its associated slot,
and the front led.
• Congurable parameters
Each alarm has several parameters for conguraon, starng with the usual setpoint, hysteresis and
maximum (alarm acve when reading is higher than
setpoint) or minimum (alarm acve when reading is
lower than minimum) alarm types (see Figure 7).
• Acvaon and deacvaon delays
Each alarm can congure independent acvaon and
deacvaon delays. These delays aect the alarm as a
whole, and the delay will aect the front led and the
associated relay.
• Second setpoint
Conguring a second setpoint creates ‘windowed
alarms’. The windowed alarm controls with a single
relay output if the reading is inside or outside the values
dened (see Figure 8).
• Inverted relay
Acvate the ‘inverted relay’ funcon to invert the
acvaon logic of the associated relay.
• ‘Locked alarms’
Acvate the ‘locked alarms’ funcon will force the
operator to interact with the instrument when an alarm
has been acvated. Once acvated, the alarm will
remain locked at acve state, even is the reading returns
to a value below setpoint, unl the operator manually
unlocks the alarms pressing the front key ‘LE’ (or the
remote key ‘LE’, see secon 3.1).
Reading
on
o
on
o
on
o
acvaon
delay
histéresis
Alarm as maximum, no
hysteresis, no delays
Alarm as maximum,
hysteresis and delays
deacvaon
delay
Alarm as minimum, no
hysteresis, no delays
setpoint
• ‘On alarm’ funcons
The ‘on alarm’ funcons allow to associate a funcon
to the alarm acvaon event. Funcons available are
reset to ‘0’, load the preset value, or do nothing.
Funcons reset and preset create counng cycles (from
0, then to setpoint, then to 0 again, ...). The number
of cycles performed can be accessed through the fast
access menu
t
Reading
t
t
on
t
o
(see secon 1.13.17)
Setpoint 2
Setpoint 1
.
t
Alarm as minimum
double setpoint, no hysteresis, no delays
t
, with
Figure 7 - Example for alarm with 2 setpoint
18
Figure 8 - Example for alarm with 2 setpoints
Page 20
1.13.16 Alarms conguraon menu
Alarms
To congure the alarm, access the alarm menu (‘ALr1’,
‘ALr2’ or ‘ALr3’) and congure the following parameters :
• at the ‘Acve’ (‘Act’) parameter select ‘on’
Alarm 1
Acve
Type of alarm
Setpoint
Hysteresis
Acvaon
delay
Deacvaon
delay
Setpoint 2
Inverted relay
Locked alarm
On alarm
Connue
to ‘0’
to preset
• at the ‘Type of alarm’ (‘TypE’) parameter select ‘MAX’
for maximum alarm (acvates when reading is higher
than setpoint), or ‘MIn’ for minimum alarm (acvates
when reading is lower than setpoint).
• at the ‘Setpoint’ (‘SEt’) parameter congure the alarm
acvaon point. Parameter value is accessible through
‘fast access’
(see secon 1.13.17)
.
• at the ‘Hysteresis’ (‘hySt’) parameter select the
hysteresis value. Hysteresis applies to the alarm
deacvaon. Alarm deacvates once the reading is
beyond the setpoint plus the hysteresis value.
Hysteresis prevents relay switching in case of signal
uctuaons close to the setpoint value.
• at the ‘Acvaon delay’ (‘dEL.0’) parameter
congure the delay to apply before the alarm is acvated.
Delay starts to count once the setpoint is reached.
Value from 0.0 to 99.9 seconds.
• at the ‘Deacvaon delay’ (‘dEL.1’) parameter
congure the delay to apply before the alarm is
deacvated. Delay starts to count once the setpoint
is reached plus the hysteresis value. Value from 0.0 to
99.9 seconds.
• to work with ‘windowed alarms’
(see Figure 8)
acvate
‘Setpoint2’ (‘SEt2’) to ‘on’ and then congure the
desired second setpoint value. Second setpoint must
always be higher in value than the rst setpoint.
• at the ‘Inverted relay’ (‘r.Inv’) parameter select ‘on’ to
invert the acvaon logic of the relay. Relay is inacve
when alarm is acve, and relay is acve when alarm is
inacve.
• at the ‘Locked alarm’ (‘A.Lck’) parameter select ‘on’
to block the automac alarm deacvaon. Alarm
deacvaon must be performed manually, by pressing
the ‘LE’ front buon
(see secon 1.13.19)
.
• at the ‘On alarm’ (‘on.AL’) parameter congure the
acon to acvate when the alarm acvates. Select
‘cont’ to do nothing and connue counng, select
‘to_0’ to load a ‘0’ on display, or select ‘to_p’ to load
the preset value on display. Selecng ‘to_0’ or ‘to_p’
congures ‘dEL.1’ to 1 second.
19
Page 21
1.13.17 Fast access
The ‘fast access’ is an operator congurable menu. The
operator can access this menu with a single press of the front
key ‘UP’ (5). The congured menu entries will be accessible.
Eligible parameters to be accessed by this menu are:
• access to the alarm setpoints through the ‘UP’ (5) key
allows to read and modify the values.
• access to the maximum and minimum alarms through
the ‘UP’ (5) key allows to read and reset the values. To
reset the memory values: visualize the value on display,
press the ‘UP’ (5) key, when the ‘rSt’ message appears,
press ‘SQ’ (<) . The instrument will return to the memory
visualizaon. Press the ‘LE’ (3) key to exit his menu.
• access to the preset value to read and modify the value.
1.13.18 ‘On power up’ funcon
The ‘On Power Up’ (‘on.Pu’) funcons allows to dene a
series of acons to acvate when the instrument restarts
aer a power loss. Funcons available are a delay so the
instrument waits a dened me before starng to measure
and control, and an automac reset of the counter. The
funcons will apply only aer a restart due to powerloss, they will not apply aer a restart due to changes in
conguraon.
The ‘fast access’ menu is not aected by the password
funcon. This means that the conguraon menu can be
password blocked, while some congured funcons or
parameters can sll be accessible to the operator through the
‘fast access’ menu.
• Super fast access
If only a single funcon is selected for the ‘fast access’ menu,
pressing the the ‘UP’ (5) key will shortly display the funcon
name and then automacally jump to the funcon value.
While on delay mode, the instrument shows all decimal
points lightened and ashing, all alarms are deacvated, and
there is no signal acquision or communicaons control.
When the delay me is over, the instrument starts its normal
funconing.
Delaying the measure and control funcons gives addional
me to elements of the system who are slower, so they can
start completely before the instrument begins to acquire
signal and control the outputs.
1.13.19 Key ‘LE’
The ‘LE’ (3) key at the front of the instrument can be
congured to acvate several funcons. Only one funcon
can be assigned to the ‘LE’ (3) key. Eligible funcons are
reset of the counter and the alarm unlock funcon (see
secon 1.13.15).
Example: an impulse counter acvates alarm 1 when reading
reaches 153.000. Automacally the instrument acvates a
reset, reading goes to ‘0’, and relay 1 is acvated to inform
that the required level has been reached. Counter remains
at 0 receives several addional impulses, related to the
system not stopping immediately. When the operator
arrives, reloads the system and presses key ‘LE’ to return the
reading to ‘0’, unlocks the alarm and restarts the system.
20
Page 22
1.13.20 ‘Fast access’ conguraon menu
Tools
Key UP
(‘Fast access’)
Setpoint 1
Setpoint 2
Setpoint 3
Memory of
maximum
Memory of
minimum
Memory of
cycles
Preset value
At the ‘Key UP (‘fast access’)’ (‘K.uP’) menu congure which funcons and parameters will be accessible
through the ‘fast access’ menu. Select ‘on’ to acvate each
funcon. For more informaon see secon 1.13.17.
• the ‘Setpoint 1’ (‘ALr1’) funcon allows to visual-
ize and modify the alarm 1 setpoint through the ‘fast
access’ menu.
• the ‘Setpoint 2’ (‘ALr2’) funcon allows to visual-
ize and modify the alarm 2 setpoint through the ‘fast
access’ menu.
• the ‘Setpoint 3’ (‘ALr3’) funcon allows to visual-
ize and modify the alarm 3 setpoint through the ‘fast
access’ menu.
• the ‘Memory of maximum’ (‘MAX’)or ‘Memory
of minimum’ (‘MIn’) funcons allow to visualize the
maximum or minimum reading value stored in memory.
• the ‘Memory of cycles’ (‘cYcL’) funcon allows to
visualize the number of cycles value stored in memory.
The cycles value increases ‘+1’ with each reset or preset
associated to the alarm acvaon or resets associated
to ‘overrange’/‘underrange’.
• the ‘Preset value’ (‘PrSt’) funcon allows to visualize and
modify the preset value through the ‘fast access’ menu.
1.13.21 ‘On power up’ conguraon menu
On power-up
DelaySeconds
Reset
1.13.22 ‘Key LE’ conguraon menu
No funcon
Key LE
Front reset
Alarm unlock
Reset and alarm unlock
The ‘On Power Up’ (‘on.Pu’) menu assigns funcons to be
applied when the instrument starts aer a power loss.
For more
informaon see secon 1.13.18.
• at the ‘Delay’ (‘dLAy’)parameter congure the me the
instrument will wait before starng normal funconality.
Time between 0 and 200 seconds.
• at the ‘Reset’ (‘rSt’)parameter set to ‘on’ to acvate a
reset when restarng aer a power loss
The ‘LE’ (3) key at the front of the instrument can be
congured to acvate several funcons. For more
informaon see secon
1.13.19
.
• the ‘No funcon’ (‘nonE’)value assigns no funcon.
• the ‘Front reset’ (‘F.rSt’) value assigns the reset
funcon.
• the ‘Alarm unlock’ (‘A.Lck’) value assigns the manual
alarm unlocking, when the ‘Locked alarms’ (‘A.Lck’)
funcon is acve.
• the ‘Reset and alarm unlock’ (‘Fr.AL’) value assigns
both funcons at the same key..
21
Page 23
1.13.23 ‘Overrange / underrange’ funcon
Counter
overrange
Counter
underrange
1.13.24 Le zeros funcon
With le zeros.
Le zeros
Without le zeros.
Flash
To zero
To preset
Flash
To zero
To preset
The ‘Counter overrange’ (‘c.orG’) and ‘Counter
underrange’ (‘c.urG’) parameters congure the behavior
of the instrument when reading is higher than ‘999999’
(overrange) or lower than ‘-199999’ (underrange). Select
‘FLSH’ to enter reading into ash mode. Select ‘to_0’ to
apply a reset to ‘0’. Select ‘to_P’ to apply a reset to preset
value.
The ‘Le zeros’ (‘L.ZEr’) parameter controls the le zeros on
or o.
1.13.25 Excitaon voltage funcon
Vexc with error control
Vexc. control
Vexc without error control
At the ‘Vexc control’ (‘V.ctr’) parameter select ‘on’ to
acvate the ‘Err.8’ message, when consumpon
requested to the excitaon voltage is higher than the current the
instrument can provide.
22
Page 24
1.13.26 ‘Password’ funcon
Password
1.13.27 Default factory conguraon
Factory
conguraon
The password funcon blocks access to the conguraon
menu. The ‘fast access’ menu is not aected by the password
funcon. This means that the conguraon menu can be
password blocked, while some congured funcons or
parameters (setpoint values, preset value, ...) can sll be
accessible to the operator through the ‘fast access’ menu.
To acve the ‘Password’ funcon select ‘on’ and introduce
the 6 digits code. The code will be requested when trying to
access the ‘conguraon menu’ (front key ‘SQ’ (<)).
At the ‘FActory conguraon’ (‘FAct’) menu select ‘yes’ to
acvate the default factory conguraon. See secon 1.15
for a list of default parameters.
1.13.28 Firmware version
The ‘Version’ (‘VEr’) menu informs about the rmware
Version
version installed on the instrument.
1.13.29 Brightness conguraon
Minimum
Brightness
Standard
Maximum
At the ‘Brightness’ (‘LIGh’) menu select the intensity level
for the display . Use this funcon to adapt the brightness to
match other instruments in the vicinity or to the darkness or
clarity of your environment.
1.13.30 Access to the opons conguraon menu
The output and control opons are oponal modules that
can be installed at the instrument. Formats LDB-24 and LDB44 have 2 free slots for output and control opons, while
formats LDB-26 and LDB-46 have 3 free slots (see secon
1.4).
Several of these oponal modules have their own
conguraon menu embedded.
The ‘OPt.1’, ‘OPt.2’ and ‘OPt.3’ menu entries give access to
the conguraon menu of the opon installed.
Opon1
Opon
Opon3
Access to the oponal module installed at slot 1
Access to the oponal module installed at slot 2
2
Access to the oponal module installed at slot 3
See secon 2 for a list of available output and control
modules
23
Page 25
1.14 Full conguraon menu
Press ‘SQ’ (<) for 1 second to access the ‘Conguraon menu’.
Pulls on channel A no pull resistor
Pulls on channel B no pull resistor
Pulls on reset pull-up
Trigger 2,5 Vdc
Acvaon for channel A on rising edge (‘on_h’)
Excitaon voltage 5 Vdc
Anrrebound lter 0 mSeconds
Tools
Fast access (Key UP) o
‘On Power Up’
Delay 0 seconds
Key ‘LE’ reset funcon
Memory of maximum -199999
Memory of minimum 999999
Memory of cycles 0
Counter overrange ash
Counter underrange ash
Le zeros o
Vexc. control o
Password o
Brightness 3
Factory conguraon for Ratemeter (‘cnF.6) and periodmeter (‘cnF.8) modes.
Mulplier x1
Divider /1
Time windows 0.5
‘SLOW’ mode
tIME 0 (o)
nuMb 1
Recursive lter 0 (o)
27
Page 29
1.16 Mounng
The instrument xaons are designed to allow panel mount,
wall mount, or hanging mount. For each type of mounng,
• Panel mount. Apply the cut-out to the panel as seen on
secon 1.4. Remove the side xaons. Introduce the
instrument into the panel cut-out. Mount the side xaons
as shown (see Figure 9). Slightly loosen the xaon screw
of one side and press the instrument against the panel.
Tighten the xaon screw so it presses the panel and
maintains the xaon. Repeat with the opposite side
xaon. For IP65 protecon at the panel juncon, see the
IPB accessories at secon 3.
Fixaon screws
Side xaons
see the posion of the xaons at the images below.
• Wall mount. Mount the side xaons against the wall,
as shown (see Figure 11). Each xaon has 2 holes with
4,5 mm diameter and a separaon between hole centers
of 30 mm. Once the side xaons are secured against the
wall, place the instrument and press the xaon screws
slightly. Tilt the instrument to the desired viewing angle
and rmly screw the xaon screws.
Diameter 4,5 mm
30 mm between
hole centers
Fixaon screws
Side xaons
Figure 9 - Panel mount
• Hanging mount. Mount the side xaons as shown (see
Figure 10). Each xaon has 2 holes with 4,5 mm diam
eter and a separaon between hole centers of 30 mm.
Instrument can be hanged using cable, threaded rod, ....
Diameter 4,5 mm
30 mm between
hole centers
Fixaon screws
Side xaons
Figure 11 - Wall mount
-
Figure 10 - Hanging mount
28
Page 30
1.17 Installaon precauons
1.19 CE declaraon of conformity
Risk of electrical shock. Instrument terminals can be
connected to dangerous voltage.
Instrument conforms to CE rules and regulaons.
This instrument has been designed and veried conforming to
the 61010-1 CE security regulaon, for industrial applicaons.
Installaon of this instrument must be performed by
qualied personnel only. This manual contains the appropriate
informaon for the installaon. Using the instrument in ways
not specied by the manufacturer may lead to a reducon of
the specied protecon level. Disconnect the instrument from
power before starng any maintenance and / or installaon
acon.
The instrument does not have a general switch and will start
operaon as soon as power is connected. The instrument
does not have protecon fuse, the fuse must be added during
installaon.
An appropriate venlaon of the instrument must be assured.
Do not expose the instrument to excess of humidity. Maintain
clean by using a humid rag and do NOT use abrasive products
such as alcohols, solvents, etc.
General recommendaons for electrical installaons
apply, and for proper funconality we recommend : if possible,
install the instrument far from electrical noise or magnec eld
generators such as power relays, electrical motors, speed
variators, ... If possible, do not install along the same conduits
power cables (power, motor controllers, electrovalves, ...)
together with signal and/or control cables.
Before proceeding to the power connecon, verify that the
voltage level available matches the power levels indicated in
the label on the instrument.
In case of re, disconnect the instrument from the power
line, re alarm according to local rules, disconnect the air
condioning, aack re with carbonic snow, never with water.
1.18 Warranty
Supplier: Omega Engineering
Products LDB-C1
The manufacturer declares that the instruments indicated
comply with the direcves and rules indicated below.
Between AC power lines
Between AC power lines and earth
Between DC power lines
Between DC power lines and earth
Between signal lines and earth
±4 KV
±8 KV
: ±2 KV
: ±2 KV
: ±1 KV
±1 KV
±1 KV
±2 KV Criteria
±2 KV
±1 KV
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
Criteria
B
B
A
B
B
B
B
B
B
B
B
A
A
A
A
B
B
A
Please see the last page for Omega’s warrenty disclaimer
According to direcve 2012/19/EU, electronic
equipment must be recycled in a selecve and
controlled way at the end of its useful life.
29
Page 31
2. Output and control modules
2.1 Module R1
The R1 module provides 1 relay output to install in large
format industrial meters from LDB series. Formats LDB-26 and
LDB-46 accept up to 3 relays, and formats LDB-24 and LDB-44
accept up to 2 relays.
Conguraon is performed from the front keypad of the
instrument, by seng the alarm parameters. Check the alarm
menu parameters at the instrument user’s manual for full
informaon.
Modules R1 can be provided factory installed into a
LDB series, or standalone for delayed installaon. No
soldering or special conguraon is required. See secon 1.6
on how to install output and control modules.
‘com’ (‘A’)
2.2 Module T1
The T1 module provides 1 transistor output to install in large
format industrial meters from LDB series. Formats LDB-26 and
LDB-46 accept up to 3 transistor outputs, and formats LDB-24
and LDB-44 accept up to 2 transistor outputs.
Conguraon is performed from the front keypad of the
instrument, by seng the alarm parameters. Check the alarm
menu parameters at the instrument user’s manual for full
informaon.
Modules T1 can be provided factory installed into a
LDB series, or standalone for delayed installaon. No
soldering or special conguraon is required. See secon 1.6
on how to install output and control modules.
‘B’
‘NC’ (‘C’)
‘NO’ (‘B’)
Figure 12 - Module ‘R1’ and internal schemac
Type of relay 3 contacts (Com, NO, NC)
Max. current 3 A (resisve load)
Voltage 250 Vac connuous
Isolaon 3500 Ve
Terminal plug-in screw clamp, pitch 5.08 mm
Installaon allowed at
slot 1, slot 2, slot 3
NOpen
NClosedCommon
CBA
Module R1
‘A’
Figure 14 - Module ‘T1’ and internal schemac
Type of output transistor
Max. voltage 35 Vdc
Max. current 50 mA
Isolaon 3500 Ve, optoisolated
Terminal plug-in screw clamp, pitch 5.08 mm
Installaon allowed at
slot 1, slot 2, slot 3
B
A
CBA
Module T1
Not connected
ACommon
BNO (Normally Open)
CNC (Normally Closed)
Figure 13 - Connecons for ‘R1’ relay output module
30
AEmier
BCollector
CNot connected
Figure 15 - Connecons for ‘T1’ transistor output module
Page 32
2.3 Module SSR
2.4 Module AO
The SSR module provides 1 output for SSR relay control, to
install in large format industrial meters from LDB series.
Formats LDB-26 and LDB-46 accept up to 3 SSR control
outputs, and formats LDB-24 and LDB-44 accept up to 2 SSR
control outputs.
Conguraon is performed from the front keypad of the
instrument, by seng the alarm parameters. Check the alarm
menu parameters at the instrument user’s manual for full
informaon.
Modules SSR can be provided factory installed into a
LDB series, or standalone for delayed installaon. No
soldering or special conguraon is required. See secon 1.6
on how to install output and control modules.
+15 Vdc
‘C’
Relé SSR
The AO module provides 1 analog output, congurable
for 4/20 mA or 0/10 Vdc signal, to install in large format
industrial meters from LDB series. Formats LDB-26 and LDB-46
accept up to 3 analog outputs, and formats LDB-24 and LDB44 accept up to 2 analog outputs.
Output signal is fully scalable, both with posive and negave
slopes, and is proporonal to the reading. The mA output can
be congured for acve loops (the instrument provides the
power to the mA loop) or passive loops (the loop power is
external to the instrument).
Conguraon is performed from the front keypad of the
instrument, by accessing the menu entries ‘Opt.1’, ‘Opt.2’ or
‘Opt.3’, according to the slot where the module is installed.
AO modules can be provided factory installed into a
LDB series, or standalone for delayed installaon. No
soldering or special conguraon is required. See secon 1.6
on how to install output and control modules.
‘B’
‘A’
Figure 16 - Module ‘SSR’ and internal schemac
Type of output for SSR relay control
Output voltage +15 Vdc
Max. current 45 mA
Isolaon 1000 Vdc
Terminal plug-in screw clamp, pitch 5.08 mm
Installaon allowed at
Not connected
slot 1, slot 2, slot 3
Collector
+15 Vdc
CBA
Figure 18 - Module ‘AO’
Signal output 4/20mA, 0/10Vdc (acve and passive)
Accuracy 0.1% FS
Isolaon 1000 Vdc
Terminal plug-in screw clamp, pitch 5.08 mm
Installaon allowed at
Jumpers MV for
mA or Vdc output selecon
slot 1, slot 2, slot 3
mA or Vdc
M
V
CBA
Module AO
CommonV exc.
Module SSR
ANot connected
BCollector (-)
C+15 Vdc (+)
Figure 17 - Connecons for ‘SSR’ control module
AExcitaon voltage
BSignal in mA or Vdc
CCommon
Jumper MJumper closed for mA output
Jumper VJumper closed for Vdc output
Figure 19 - Connecons for ‘AO’ analog output module
31
Page 33
2.5 Module RTU
The RTU module provides an isolated Modbus RTU
communicaons port, to install in large format industrial
meters from LDB series.
2.6 Module S4
The S4 module provides an isolated RS-485 ASCII
communicaons port, to install in large format industrial
meters from LDB series.
The RTU module implements funcon ‘4’ (‘Read Input
Registers’) of the Modbus RTU protocol, to access the
instrument registers (reading value, alarm status, memory of
maximum and minimum, ...).
Conguraon is performed from the front keypad of the
instrument, by accessing the menu entries ‘Opt.1’, ‘Opt.2’ or
‘Opt.3’, according to the slot where the module is installed.
Modules RTU can be provided factory installed into a
LDB series, or standalone for delayed installaon. No
soldering or special conguraon is required. See secon 1.6
on how to install output and control modules.
Figure 20 - Communicaons module ‘RTU’
The S4 module implements a MASTER / SLAVE protocol, with
up to 31 addressable slaves. In SLAVE mode allows access
to reading values, alarm status, memory of maximum and
minimum, ...
Conguraon is performed from the front keypad of the
instrument, by accessing the menu entries ‘Opt.1’, ‘Opt.2’ or
‘Opt.3’, according to the slot where the module is installed.
Modules S4 can be provided factory installed into a
LDB series, or standalone for delayed installaon. No
soldering or special conguraon is required. See secon 1.6
on how to install output and control modules.
Figure 22 - Communicaons module ‘S4’
Protocol Modbus RTU
Bus RS-485, up to 57.6 Kbps
Isolaon 1000 Vdc
Terminal plug-in screw clamp, pitch 5.08 mm
Installaon allowed at
slot 1, slot 2, slot 3
A wire
GNDB wire
GAB
Module RTU
ABus signal A
BBus signal B
GGND
Protocol ASCII
Bus RS-485, up to 57.6 Kbps
Isolaon 1000 Vdc
Terminal plug-in screw clamp, pitch 5.08 mm
Installaon allowed at
slot 1, slot 2, slot 3
A wire
GNDB wire
GAB
Module S4
ABus signal A
BBus signal B
GGND
Figure 21 - Connecons for Modbus ‘RTU’ communicaons module
32
Figure 23 - Connecons for RS-485 ‘S4’ communicaons module
Page 34
2.7 Module S2
The S2 module provides an isolated RS-232 ASCII
communicaons port, to install in large format industrial
meters from LDB series.
The S2 module implements a MASTER / SLAVE protocol, with
up to 31 addressable slaves, with ‘daisy-chain’ connecon.
In SLAVE mode allows access to reading values, alarm status,
memory of maximum and minimum, ...
Conguraon is performed from the front keypad of the
instrument, by accessing the menu entries ‘Opt.1’, ‘Opt.2’ or
‘Opt.3’, according to the slot where the module is installed.
Modules S2 can be provided factory installed into a
LDB series, or standalone for delayed installaon. No
soldering or special conguraon is required. See secon 1.6
on how to install output and control modules.
Figure 24 - Communicaons module Module ‘S2’
Protocol ASCII
Bus RS-232, up to 57.6 Kbps
Isolaon 1000 Vdc
Terminal plug-in screw clamp, pitch 5.08 mm
Installaon allowed at
A
B
C
D
slot 1, slot 2, slot 3
Tx1
Rx1Rx2
Tx2
D
CBA
Module S2
GND
E
‘Daisy chain’ Tx data transmission
‘Daisy chain’ Rx data recepon
Tx data transmission
Rx data recepon
EGND
Figure 25 - Connecons for RS-232 ‘S2’ communicaons module
All modules menoned in this document are compable with
large format meters from LDB series has 4 formats, and each
format dier in the number of digits, the digit height and the
number of output and control opons they can accept.
Large displays from the LDB series are designed following
a modular architecture that allows the operator to install
any of the output and control modules menoned in this
document. Each module is supplied with 1 cable e, 1
square self adhesive e base and 1 female connector.
FormatDigitsDigit heightOpons
LDB-24460 mm2
LDB-444100 mm2
LDB-26660 mm3
LDB-466100 mm3
•
inside the programming menus, when a 6 digits value is shown,
it is assumed that only 4 digits apply to formats LDB-24 and LDB-44
• when this document explains that a maximum of 3 output
and control modules are installable, it is assumed that the
maximum is 2 modules for formats LDB-24 and LDB-44
The output and control modules menoned in this document,
are covered by the warranty of the instrument where they are
installed. Check the user’s manual of the instrument for more
informaon related to warranty.
The user’s manual of the instrument where the module is
installed, has important informaon related to installaon
that applies also to the output and control modules
menoned in this document. Check the user’s manu-
!
al of the instrument for more informaon related to
installaon precauons.
The output and control modules menoned in this document
are covered by the ‘CE declaraon of conformity’ of
the instrument where they are installed. Check the
user’s manual of the instrument for more informaon
related to the CE declaraon of conformity.
1.3 Installaon and start-up
To install an oponal output and control module into a large
display:
1. remove the rear cover of the instrument (see secon 1.4)
2. install the module at one of the free slots (see secon 1.5)
3. place the squared e base at the free slot selected.
Locaon to place the e base is clearly indicated on the PCB
(see secon 1.5).
4. pass the cable e through the e base (see secon 1.5)
5. place the output and control module at the slot connecon
jumpers (see secon 1.5)
6. use the cable e to rmly x the module (see secon 1.5)
7. if needed, congure the appropriate jumpers at the output
and control module
8. pass the connecon wires through the housing cable gland
9. connect the signal wires to the terminals of the output and
control module
10. place and close the rear cover of the instrument (see
secon 1.4)
11. congure the parameters at the ‘Conguraon menu’.
• modules R1, T1 and SSR are congured from the alarms
menu of the instrument
• other modules are congured from from menu entries
‘Opt.1’, ‘Opt.2’ or ‘Opt.3’, depending on the slot where the
module has been installed.
2
Page 36
1.4 To access the instrument
To open the housing, remove the screws from the back cover.
With each screw there is a metal washer and a plasc washer.
Once the screws are out, remove the back cover.
The gure below shows the instrument internal structure for a
LDB-26 format. It shows the locaon of the 3 slots for oponal
output and control modules, the power terminal and the input
signal terminal.
Waterght seal
Female turret
Power
Slot for opon 3Back cover
To close the instrument, place the back cover, the screws, the
metal washer and the plasc washer. The plasc washer is in
contact with the back cover. Conrm that the screws are correctly
turning inside the internal female screws.
To ensure a correct IP65 protecon ghten the back cover screws
with a strength between 30 and 40 Ncm, with the help of a
dynamometer screwdriver.
Slot for opon 2
Slot for opon 1
Remote keypad terminal
Input signal terminal
Screw
Metal washer
Plasc washer
1.5 Modular system
Large format meters are designed with an
internal modular architecture. The output and control
modules are independent and can be installed by accessing the internal circuits of the instrument, and connecng
Output and control module
Slot 3
Tie base
Cable e
Slot 2
(2)
(1)
Module pins
Risk of electric shock. Removing the back
cover will grant access to the internal
circuits of the instrument. Operaon must
be performed by qualied personnel only.
the module to the connecon jumpers of the selected slot.
Each module is provided with a cable e to x the module to
the e base. A cable gland to install at the back cover is also
provided, in order to enable an output for the connecon
wires.
To install an output and control module
(1) insert the ‘module pins’ into the ‘con-
necon jumpers’ in one of the free slots
(2) place the ‘cable e’ into the ‘e base’
and embrace the ‘module’ rmly, unl
it is xed
Slot 1
(3) an addional white cable e is provided
to x as indicated below. Only needed in
case of vibraons or heavy transportaon.
Connecon jumpers
3
Page 37
1. Opons R1, T1 and SSR
The R1, T1 and SSR modules provide 1 digital ‘on/o’
output. The output is congured from the instrument
alarms menu (‘ALr.1’, ‘ALr.2’ o ‘ALr.3’).
The menu allows to congure the setpoint, hysteresis,
independent acvaon and deacvaon delays, and a
second setpoint to create windowed alarms.
The R1, T1 and SSR output modules are isolated between
them and between all other circuits of the instrument.
1.1 Module R11.2 Module T1
‘com’ (‘A’)
‘NC’ (‘C’)
‘NO’ (‘B’)
Figure 1 - Detail for the ‘R1’ module and internal schemac
Figure 3 - Detail for the ‘T1’ module and internal schemac
‘B’
‘A’
Opon R1
Type of output relay
Type of relay 3 contacts (Com, NO, NC)
Max. current 3 A (resisve load)
Voltage 250 Vac connuous
(max. 150Vac if switching power network with Overvoltage category III)
Isolaon 3500 Ve
Type of terminal plug-in screw clamp
pitch 5.08 mm
Installaon allowed at slot 1, slot 2, slot 3
NOpen
NClosedCommon
CBA
Opon T1
Type of output transistor
Max voltage 35 Vdc
Max. current 50 mA
Isolaon 3500 Ve, optoisolated
Type of terminal plug-in screw clamp
pitch 5.08 mm
Installaon allowed at slot 1, slot 2, slot 3
B
A
Not connected
CBA
Module R1
ACommon
BNO (Normally Open)
CNC (Normally Closed)
Figure 2 - Connecons for the ‘R1’ relay output module
4
Module
AEmier
BCollector
CNot connected
Figure 4 - Connecons for the ‘T1’ transistor output module
T1
Page 38
1.3 Module SSR
+15 Vdc
‘C’
Relé SSR
‘B’
‘A’
Figure 5 - Detail for the ‘SSR’ module and internal schemac
Opon SSR
Type of output to control SSR relay
Output voltage +15 Vdc
Max. current 45 mA
Isolaon 1000 Vdc
Type of terminal plug-in screw clamp
pitch 5.08 mm
Installaon allowed at slot 1, slot 2, slot 3
Collector
+15 Vdc
Opt.2
A B C
Opt.3
A B C
Not connected
Opt.1
A B C
CBA
Module SSR
ANot connected
BCollector (-)
Signal
Power
C+15 Vdc (+)
Figure 6 - Connecons for the SSR control output module
5
Page 39
2. Opon AO
The AO modules provide 1 analog output, congurable for
4/20 mA or 0/10 Vdc signal. The analog output is congured
from the opons menu entry (‘Opt.1’, ‘Opt.2’ or ‘Opt.3’) of
the instrument.
Opon AO
Type of output analog output
Signal output 4/20mA acve
4/20mA passive
0/10Vdc
Max. signal 22mA, 10.5Vdc
Min. signal 0mA, -50mVdc
Scaling proporonal to the reading
posive or negave slopes
Vexc (terminal A) +13.8Vdc ± 0.4 Vdc (max. 25 mA)
protecon against shortcircuit
Load impedances ≤350Ohm (for 4/20mA acve)
≤800Ohm (for 4/20mA passive)
(for 24 Vdc external Vexc) (maximum
voltage 27 Vdc between ‘B’ and ‘C’)
≥10KOhm (en 0/10Vdc)
Accuracy (at 25ºC) <0.1% FS
Thermal stability 60ppm/ºC in mA
50 ppm/ºC in Vdc
Step response <75mSeconds + step response of the
(0% to 99% of the signal)
reading
Isolaon 1000 Vdc
Warm up 15 minutes
Type of terminal plug-in screw clamp
pitch 5.08 mm
Factory conguraon ‘Mode mA’
‘Scaling 0/9999 = 4/20mA’
‘On error ‘to_h’
Installaon allowed at slot 1, slot 2, slot 3
The output signal is proporonal to the reading, and it is
scalable both in posive or negave slopes. The mA output
can be congured for acve loops (the instrument provides
the power to the mA loop) or passive loops (the loop power
is external to the instrument.
The AO analog output modules are isolated between them
and between all other circuits of the instrument.
Figure 7 - Detail for the ‘AO’ module
mA or Vdc
CommonV exc.
Jumpers MV to
select mA or Vdc
output
MV
CBA
Module AO
AExcitaon voltage
BSignal in mA or Vdc
CCommon
Jumper MJumper closed for mA output
Jumper VJumper closed for Vdc output
Figure 8 - Connecons for ‘AO’ analog output module
2.1 Connecon examples
mA
V exc.
MV
CBA
Module AO
Jumper MJumper closed
Jumper VJumper open
Figure 9 - Connecons for acve 4/20 mA. The current loop is
internally powered from the ‘AO’ module
6
+ mA
- mA
MV
CBA
Module AO
Jumper MJumper closed
Jumper VJumper open
Figure 10 - Connecons for passive 4/20 mA. The current loop
is externally powered.
Page 40
2.2 Conguraon menu
At the ‘Mode’ (‘ModE’) menu congure the type of output
‘4/20mA’ (‘mA’) or ‘0/10 Vdc‘ (‘Vdc’). Posion for jumpers
‘V’ and ‘M’ must be according to the range selected.
At the ‘Scaling’ (‘ScAL’) menu enter the values that dene
the two points of the slope:
• the lower point, dened by the ‘Low Display’ (‘d.Lo’)
and ‘Low Output’ (‘Ao.Lo’)
• the upper point, dened by the ‘High Display’ (‘d.hI’)
and ‘High Output’ (‘Ao.hI’)
Analog output values are shown with ‘XX.XX’ format.
acceptable values are ‘0.00’ to ‘10.00’ Vdc for voltage, and
‘0.00’ to ‘20.00’ mA for current.
Reading
100.0
Example
associated to a reding of -50.0 to 100.0
- 4/20mA, analog output
‘d.hI’=‘100.0’
‘Ao.hI’=‘20.00’
Mode
Scaling
‘On error’
Factory
conguraon
Mode 4/20mA
Mode 0/10Vdc
Display low
Output low
Display high
Output high
in case of error, ‘to_h’
to drive output to high
level, ‘to_L’ to drive
output to low level
select ‘yES’ to reload
the default factory conguraon
-50.0
4 mA
‘Ao.Lo’=‘4.00’
‘d.Lo’=‘-50.0’
signal Vdc
MV
CBA
Module AO
20 mA
common
Version
Analog
output
2.3 Error codes
‘Er.34’ output signal congured to value lower than 0Vdc or 0 mA
‘Er.35’ output signal congured to a value higher than 10Vdc or
20 mA
‘Er.36’ congured slope points are not acceptable, such as :
‘d.Hi’=’d.Lo’
‘Ao.Hi’=’Ao.Lo’
(‘Ao.Hi’-’Ao.Lo’)>(’d.Hi’-’d.Lo’)
Jumper MJumper open
Jumper VJumper closed
Figure 11 - Connecons for 0/10 Vdc.
7
Page 41
3. Opon
RTU
The RTU modules provide 1 port for communicaons
in Modbus RTU protocol. Use funcon ‘4’ (‘Read Input Registers’) of the Modbus RTU protocol, to access the
instrument registers (reading value, alarm status,
memory of maximum and minimum, setpoint values, ...).
Opon RTU
Type of output Modbus RTU communicaon
Funcon implemented 4 (Read_Input_Registers)
Addresses 01 to 247
Excepon codes see secon 3.3
Registers* see secon 3.1
*available registers can vary for dierent instruments
Bus RS-485
Speed 57.6 Kbps to 600 bps
Data format 8e1 (standard), 8o1, 8n2
Bus terminator not included
Isolaon 1000 Vdc
Temperature operaon from 0 to 50 ºC
storage from -20 to +70 ºC
The communicaon parameters are congured from the
opons menu entry (‘Opt.1’, ‘Opt.2’ or ‘Opt.3’) of the
instrument.
The RTU modules are isolated between them and
between all other circuits of the instrument.
bit 0...7 alarm status
bit 8...16 instrument status
4 Digit Models
(LDB-24 y LDB-44)
0 to 4
Not accessible
Table 1 - Registers accessible through MODBUS-RTU. Registers codied as binary numbers. Negave values codied in two’s complement. Available
registers can vary for dierent instruments. Register 11 is not accessible for instruments with formats LDB-24 and LDB-44 ( slot 3 is not available).
8
Page 42
3.2 Conguraon menu
AddressConguraon
Speed
(kbps)
1 to 247
57.6 Kbps
...
...
to 600bps
At the ‘Conguraon’ (‘rtu’) menu, congure the ‘Address’
(‘Addr’) parameter with the address value between ‘1’ and
‘247’, at the ‘Speed’ (‘bAud’) parameter select the bus speed
(in Kbps) and at the ‘Format’ (‘bItS’) parameters select the
data format.
Inside the ‘Tools’ (‘TooL’) menu, special tools and funcons
are grouped.
• the ‘Decimal point’ (‘dP’) menu is provided for
compability with ancient hardware that does not
support decimal point retransmission. By default, select
‘Automac’ (‘Auto’). If your instrument does nos transmit the
decimal point posion, select ‘Manual’ (‘MAnL’) and x
the posion of the decimal point manually.
• at the ‘Factory reset’ (‘FAct’) menu, select ‘yes’ to load the
default factory conguraon for the instrument.
the ‘Version’ (‘VEr’) menu informs of the current rmware
version installed in the module.
3.3 Excepon codes
Tools
Format
Factory
conguraon
Version
AutomacDecimal point
Manual
8 bits, even parity, 1 stop
8 bits, odd parity, 1 stop
8 bits, no parity, 2 stop
Move with LE
The Modbus RTU protocol denes the following scenarios
when a ‘Master’ is sending a frame to a ‘Slave’:
• the ‘Slave’ device receives the frame correctly and replies
with the requested data
• the ‘Slave’ devices detects a CRC error, parity error, or other.
and discards the frame without generang a reply frame. The
‘Master’ will detect a ‘TIMEOUT’ condion due to the absence
of reply.
• the ‘Slave’ device receives the frame correctly, but replies
with an ‘EXCEPTION_CODE’ as it can not process the funcon
or register requested.
The ‘EXCEPTION_CODES’ congured in the RTU module are :
Table 3 - Firmware versions compable with the indicated registers
Firmware
version
Formats
LDB-24, LDB-44
Firmware
version
9
Page 43
3.5 Descripon and example of registers
Registers R0 and R1 (DISPLAY1_L y DISPLAY1_H)
Contains the display value of the instrument, codied in
two registers of 16 bits each. Possible values are from
999999 to -199999. Decimal point posion is codied at
register R2.
Example R0=FBF1 (hex) and R1=0009 (hex)
Register value = 0009 FBF1 (hex)
Reading value = 654321
Register R2 (DECIMALS1)
Contains the number of decimals of the display, codied
in a single register of 16 bits. Possible values are from
0 to 6.
Example R2=0002 (hex)
Number of decimals = 2 = 6543.21
Register R3 and R4 (MAXMEM_L and MAXMEM_H)
Contains the memory of maximum reading of the
instrument, codied in two registers of 16 bits each.
Possible values are from 999999 to -199999. Decimal
point posion is codied on register R2.
Example - same example as in R0 and R1 but accessing
to R3 and R4.
Registers R5 and R6 (MINMEM_L and MINMEM_H)
Contains the memory of minimum reading of the
instrument, codied in two registers of 16 bits each.
Possible values are from 999999 to -199999. Decimal
point posion is codied on register R2.
Example - same example as in R0 and R1 but accessing
to R5 and R6.
Registers R9 and R10 (SETPOINT2_L and SETPOINT2_H)
Contains the setpoint value of alarm 2, codied in two
registers of 16 bits each. Possible values are from 999999
to -199999. Decimal point posion is codied on register
R2.
Example - same example as in R0 and R1 but accessing
to R9 and R10.
Registers R11 and R12 (SETPOINT3_L and SETPOINT3_H)
Contains the setpoint value of alarm 3, codied in two
registers of 16 bits each. Possible values are from 999999
to -199999. Decimal point posion is codied on register
R2.
Example - same example as in R0 and R1 but accessing
to R11 and R12.
Register R13 (STATUS)
Informaon bit-by-bit, for the alarm status (on / o) and
instrument status. See below for a descripon.
Bit 0 Alarm 1 status (0 = inacve, 1 = acve)
Bit 1 Alarm 2 status (0 = inacve, 1 = acve)
Bit 2 Alarm 3 status (0 = inacve, 1 = acve)
Bit
3 to 7
Reserved
Bit 8 Display overrange
Bit 9 Display underrange
Bit 10 Lost communicaon with the main processor
Bit 11 to 15 Reserved
Registers R14, R15 and R16
Reserved
Registers R7 and R8 (SETPOINT1_L and SETPOINT1_H)
Contains the setpoint value of alarm 1, codied in two
registers of 16 bits each. Possible values are from 999999
to -199999. Decimal point posion is codied on register
R2.
Example - same example as in R0 and R1 but accessing
to R7 and R8.
10
Page 44
4. Opon
S4
The S4 modules provide 1 port for communicaons RS485
ASCII protocol. Protocol with ‘master’ - ‘slave’ architecture,
addressable up to 31 modules. Frames codied in representable ASCII characters (codes 32 to 255), which are visible
using ‘hyperterminal’ or similar programs. Instrument
Opon S4
Type of output RS-485 ASCII communicaon
Bus RS-485
Speed 57.6 Kbps to 600 bps
Data format 8n1 (standard), 8o1, 8n2, 8e1
Bus terminator not included
Protocol ASCII
Architecture ‘master - slave’
Addresses 01 to 31
‘Broadcast’ address 128
Registers* see secon 4.1
*available registers can vary for dierent instruments
Isolaon 1000 Vdc
Temperature operaon from 0 to 50 ºC
storage from -20 to +70 ºC
Factory conguraon ‘Mode Slave’
‘Address 1’
‘Speed 19.2 Kbps’
‘Format 8n1’
‘Decimal point Auto’
Conguraon ‘Master’ ‘Desnaon address 31’
‘Frequency 0.5 sec.’
Tools ‘Decimal point Auto’
‘Legacy O’
‘Answer delay 0 mSec.’
Installaon allowed at ‘Opt.1’, ‘Opt.2’, ‘Opt.3’
registers are accessible through the RS-485 ASCII port
(reading value, alarm status, memory of maximum and
minimum, setpoint values, ...). The communicaon
parameters are congured from the opons menu
entry (‘Opt.1’, ‘Opt.2’ or ‘Opt.3’) of the instrument. The S4
modules are isolated between them and between all other
circuits of the instrument.
Figure 14 - Detail for the ‘S4’ module
A wire
B wire
GAB
Module S4
ABus signal A
BBus signal B
GGND
Figure 15 - Connecons for ‘S4’ module
GND
4.1 Accessible registers
Display values (DISPLAY1, MAXMEM, MINMEM, AL1, AL2,
AL3) are codied with a minimum of 6 digits (le zeros
are added if necessary), polarity and decimal point.
RegisterNameDescripon
0DISPLAY1Display1 value
1MAXMEMMemory of maximum
2MINMEMMemory of minimum
3AL1Setpoint 1 value
4AL2Setpoint 2 value
5AL3Setpoint 3 value
6STATUSAlarm status
Table 4 - Accessible registers for ASCII protocol.
Register 0 - DISPLAY1
Contains the display value of the instrument, in ASCII code,
including polarity (posive / negave) and decimal point.
Example 1 - R0=’+’ ‘0’ ’6’ ‘5’ ‘4’ ‘3’ ‘.’ ‘2’ Display value = 6543.2
Example 2 - R0=’-’ ‘0’ ‘0’ ‘0’ ‘4’ ‘.’ ‘5’ ‘2’ Display value = -4.52
Register 1 - MAXMEM
Contains the value for memory of maximum, in ASCII code,
including polarity (posive / negave) and decimal point.
Register 2 - MINMEM
Contains the value for memory of minimum, in ASCII code,
including polarity (posive / negave) and decimal point.
Register 3 - AL1
Contains the value for alarm 1 setpoint, in ASCII code,
including polarity (posive / negave) and decimal point.
Register 4 - AL2
Contains the value for alarm 2 setpoint, in ASCII code,
including polarity (posive / negave) and decimal point.
Register 5 - AL3
Contains the value for alarm 3 setpoint, in ASCII code,
including polarity (posive / negave) and decimal point.
Register 6 - STATUS
Contains the alarm status (on/o).
Bit 0 Alarm 1 status (0 = inacve, 1 = acve)
Bit 1 Alarm 2 status (0 = inacve, 1 = acve)
Bit 2 Alarm 3 status (0 = inacve, 1 = acve)
Bit 3 to 15 Reserved
11
Page 45
4.2 Conguraon menu
Conguraon
ASCII
Conguraon
‘Master’
Mode
Address
Speed
(kbps)
Format
Desnaon
address
Frequency
1 a 31
1 to 31
128 for ‘broadcast’
Mode ‘Slave’
Mode ‘Master’
0.1 seconds
0.5 seconds
1 seconds
5 seconds
15 seconds
60seconds
At the ‘Conguraon ASCII’ (‘AScI’) menu, congure
the ‘Mode’ (‘ModE’) parameter to select the ‘slave’ or
the ‘master’ mode, at the ‘Address’ (‘Addr’) parameter
congure the local port address between ‘1’ and ‘31’, at
the ‘Speed’ (‘bAud’) parameter select the bus speed (in
Kbps) and at the ‘Format’ (‘bItS’) parameter select the data
format.
When working as ‘master’, the instrument connuously
transmits the display value data frame. The local module
address is ‘0’. Congure at menu ‘Conguraon Master’
(‘cnF.M’) the ‘Desnaon address’ (‘d.Add’) parameter
from ‘1’ to ‘31’ or use value ‘128’ for a broadcast message.
At parameter ‘Frequency’ (‘FrEq’) select the how oen the
frame with the reading value will be transmied.
Special tools are grouped inside the ‘Tools’ (‘TooL’) menu.
• the ‘Decimal point’ (‘dP’) menu is provided for
compability with ancient hardware that does not
support decimal point retransmission. By default,
select ‘Automac’ (‘Auto’). If your instrument does nos
transmit the decimal point posion, select
‘Manual’ (‘MAnL’) and x the posion of the decimal point
manually.
• the ‘Legacy mode’ (‘LEG’) parameter is provided to
maintain compability with instruments with older
communicaon protocols. Select ‘on’ to acvate this
mode.
• the ‘Answer delay’ (‘AnS.d’) parameter applies only
to ‘Slave’ mode. The local module delays the answer
frame. Congure for applicaons where the ‘Master’
needs addional me to switch between ‘transmit’ and
‘receive’ modes. Enter a numeric value between ‘0’ and
‘1000’ mSeconds.
• at the ‘Factory reset’ (‘FAct’) menu, select ‘yes’ to load
the default factory conguraon for the instrument.
the ‘Version’ (‘VEr’) menu informs of the current
rmware version installed in the module.
12
Tools
Decimal point
Legacy mode
Answer delay
Factory
conguraon
Version
Automac
Manual
use key ‘LE’ to
select
delay for answers, from 0
to 1000 mSec.
4.3 Compable versions
Formats
LDB-26, LDB-46
Instruments with access to registers 0, 1, 2, 6
LDB26-P, LDB46-P
LDB26-C1, LDB46-C1
LDB26-CR, LDB46-CR
Table 5 - Firmware versions compable with the indicated registers
Version
rmware
50.00
27.08
28.02
Formats
LDB-24, LDB-44
LDB24-P, LDB44-P
---
LDB24-T, LDB44-T
LDB24-R, LDB44-R
LDB24-C1, LDB44-C1
LDB24-CR, LDB44-CR
Version
rmware
41.57
44.05
45.05
47.07
48.05
---
Page 46
4.4 Frame types
The ASCII protocol denes the following frames:
• Frame ‘read’ (‘RD’). Id code 36. Request data frame. The
requested register is indicated into the ‘REG’ byte (‘Header’
secon).
• Frame ‘answer’ (‘ANS’). Id code 37. Response frame to a
request data frame. The requested register is indicated
into the ‘REG’ byte’ (‘Header’ secon). Data of the request
ed register is indicated into data bytes ‘D0’ to ‘Dn’ (‘Data’
secon).
• Frame ‘error’ (‘ERR’). Id code 38. Response frame to a
request data frame. Indicates that an error has occurred.
Error code is codied into the ‘REG’ byte (‘Header’ secon).
• Frame ‘ping’ (‘PING’). Id code 32. Used to conrm the
existence of the remote instrument.
-
• Frame ‘pong’ (‘PONG’). Id code 33. Response to a ‘ping’
frame. It conrms the existence of the remote instrument.
4.5 Frame structure
HeaderDataTrail
STXIDRSVFROMTOREGRSVLONGD0D1...DnCRCETX
2x32xxx32n+1[data]x3
0123456789...n+7n+8n+9
Protocol frames have a structure made of ‘Header’, ‘Data’ and ‘Trail’.
Secon ‘Header’
Contains the start byte (‘STX’), the frame idener (‘ID’), the
origin address (‘FROM’) and the desnaon address (‘TO’),
the register id (‘REG’) and the length (‘LONG’) of the ‘Data’
secon.
Secon ‘Data’
Contains data for the requested register (‘REG’).
Secon ‘Trail’
Contains the ‘CRC’ code and the end of frame byte (‘ETX’).
‘Real value’ and ‘Frame value’
To use representable ASCII values, the real values are codied
before being sent into the frame. The following denions apply :
• ‘real value’ is the value of the eld without codicaon
Example - ‘Master’ (address ‘0’) requests the value of register
‘0’ (display value) to the ‘Slave’ at address ‘28’ (‘RD’ frame)
and the ‘Slave’ replies to the ‘Master’ with a reply frame (‘ANS’
frame) containing the requested data (765.43).
*Instruments with 4 digits also send reading values formaed
with 6 digits : value -321.5 is transmied as -00321.5
HeaderTrail
STXIDRSVFROMTOREGRSVLONGCRCETX
236323260323232583
StartRD---0280---0CRCStop
HeaderDataTrail
STXIDRSVFROMTOREGRSVLONGD0D1D2D3D4D5D6D7CRCETX
2373260323232404348555453465251153
StartANS---2800---8+0765.43CRCStop
4.8.2 Frames ‘ERR’ (38)
Example - ‘Slave’ at address ‘11’ replies to the ‘Master’
(address ‘0’) with an error frame (‘ERR’ frame)
indicang that the requested register number is unknown
HeaderTrail
STXIDRSVFROMTOREGRSVLONGCRCETX
238324332333232463
StartERR---1101---0CRCStop
(‘UNKNOWN_REGISTER’, error code ‘1’). The error code
is codied into the ‘REG’ byte. For a list of error code see
secon 4.6.
4.7.1 Frames ‘PING’ (32) and ‘PONG’ (33)
Example - ‘Master’ (address ‘0’) requests conrmaon of
existence to the ‘Slave’ at addrress ‘22’ (‘PING’ frame) and
the ‘Slave’ replies to the ‘Master’ with a ‘PONG’ frame.
HeaderTrail
STXIDRSV
232323254323232523
Start Ping ---0220---0CRCStop
HeaderTrail
STXIDRSV
233325432323232533
Start
Pong
FROM
TOREGRSV
FROM
TOREGRSV
---2200---0CRCStop
LONG
LONG
CRCETX
CRCETX
4.7 CRC calculaon
The ‘frame value’ for the CRC byte is calculated applying a
XOR funcon to the ‘frame value’ (see secon 4.5) of all bytes
in secons ‘Header’ and ‘Data’, from byte ‘0’ (‘STX’) to the last
data byte (‘Dn’).
• if the calculated CRC value is lower than ‘32’, it is normalized
by applying the ‘one’s complement’ funcon .
CRC0=STX ^ ID ^ RSV ^ FROM ^ TO ^ REG ^ RSV ^ LONG ^ D0
^...^ Dn
• if (CRC0<32) -> CRC=!CRC0 (one’s complement funcon)
• if (CRC0>31) -> CRC=CRC0
//example of CRC calculaon in C language
int8 Calculate_CRC(int8 CRC_Posion)
{
int8 i,CRC=0;
for(i=0;c<CRC_Posion;c++)
{
crc=crc ^ frame[i];
}
if(crc<32) CRC=~CRC;
return(CRC);
}
14
Page 48
5. Opon
S2
The S2 modules provide 1 port for communicaons
RS232 ASCII protocol. The S2 modules use the same
protocol as the S4 modules (see secon 4), the only
dierence is the physical layer of the bus, that is RS232
for the S2.
S2 modules allow for point-to-point communicaon over
RS232 and also allow for mulnode communicaon over
Opon S2
Type of output RS-232 ASCII communicaon
Bus RS-232
Speed 57.6 Kbps a 600 bps
Data format 8n1 (standard), 8o1, 8n2, 8e1
Protocol ASCII
Architecture ‘master - slave’
Address 01 to 31
‘Broadcast’ address 128
Registers* see secon for S4 module
RS232 using a ‘Daisy-Chain’ type of connecon.
Terminals RX1 and TX1 are for the main communicaon
with the RS232 bus. Terminals RX2 and TX2 are for the
mulnode connecon, so all frames received at RX1 with
desnaon address dierent from the local address, will
be retransmied through TX2. On the same way, frames
received at RX2 with desnaon address dierent from
the local address, will be retransmied through TX1.
Figure 16 - Detail for the ‘S2’ module
Tx1
Rx1Rx2
Tx2
GND
*available registers can vary for dierent instruments
Isolaon 1000 Vdc
Temperature operaon from 0 to 50 ºC
storage from -20 to +70 ºC
Installaon allowed at ‘Opt.1’, ‘Opt.2’, ‘Opt.3’
D
CBA
E
Module S2
AGND
B
C
D
E
Figure 17 - Connecons for ‘S2’ module
Rx data recepon
Tx data transmission
‘Daisy chain’ Rx data recepon
‘Daisy chain’ Tx data transmission
15
Page 49
1. Remote keypad LDB-RKB
Industrial keypad with 3 push buons to connect to large format meters from LDB series. It allows to replicate the front
keypad of the instrument to a remote locaon.
A RKB remote keypad allows the operator to access the
advanced control funcons from the large format
meters, such as fast access to alarm setpoints, preset value
modicaon, access to maximum and minimum reading
values, signal tare for load applicaons, front reset, manual
alarm unlock, ...
All these features are accessible while maintaining the main
feature of these instruments, which is the installaon in
heights for long distance reading.
The RKB remote keypad is provided with an industrial IP65
protected housing, with cable gland output, aligned with the
technical specicaons of the LDB series. The RKB remote
keypad can be easily installed against wall. The push buons
are 25 mm size for easy use even with protecon gloves.
The RKB remote keypad is provided with labeled push
buons and does not included cable.
Normal buon state open
Recommended wire 0.25 mm2
Protecon IP65
Output by cable gland
Mounng accepts wall mount
Color grey
Material plasc
Weight 200 gr
GND
(<)
SQ
UP
(5)
LE
(3)
Connect the wire to the 4 pole terminal located close to the
input signal module. Connect 4 wires for keys ‘SQ’ (<), ‘UP’
(5), ‘LE’ (3) and common. Pass the wires through the cable
gland idened as ‘remote keypad’ (see Figure 2) and connect
the other end to the internal RKB push buons.
181 mm
75 mm
2
Page 50
Figure 1 - Connecons from RKB to the internal 4 pole terminal
GND
SQ
UP
(5)
LE
(<)
(3)
Power
Remote keypad
Opon 3Opon 2
Figure 2 - LDB-26 instrument front view (top), rear view (middle) and internal view (boom).
SignalOpon 1
3
Page 51
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a
period of 61 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace
period to the normal five (5) year product warranty to cover handling and shipping time. This ensures
that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no
charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser,
including but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for
any damages that result from the use of its products in accordance with information provided
by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the
company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of
OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the
component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR)
NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING
DELAYS). The assigned AR number should then be marked on the outside of the return package and on any
correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR WARRANTY RETURNS, please have the
following information available BEFORE contacting
OMEGA:
1. Purchase Order number under which the product
was PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our
customers the latest in technology and engineering.
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior
written consent of OMEGA ENGINEERING, INC.
FOR NON-WARRANTY REPAIRS, consult
OMEGA for current repair charges. Have
the following information available BEFORE
contacting OMEGA:
1. Purchase Order number to cover the COST
of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems
relative to the product.