Omega HHM63C User Manual

OMEGA
HHM63C
Digital Multimeter
OMEGAnetSM On-Line Service
Internet e-mail
info@omega.com
For immediate technical or application assistance:
USA and Canada: Mexico and Latin America:
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
SM
Tel: (95) 800-TC-OMEGA
SM
Customer Service: 1-800-622-2378 / 1-800-622-BEST
SM
FAX: (95) 203-359-7807
Engineering Service: 1-800-872-9436 / 1-800-USA-WHENSMEn Español: (203) 359-7803
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
e-mail: espanol@omega.com
Servicing North America:
USA: ISO 9001 Certified Canada:
One Omega Drive, Box 4047 976 Bergar
Stamford, CT 06907-0047 Laval (Quebec) H7L5A1
Tel: (203) 359-1660 Tel: (514) 856-6928
FAX: (203)359-7700 FAX: (514) 856-6886
e-mail: info@omega.com e-mail: info@omega.com
Servicing Europe:
Benelux:
Postbus 8034, 1180 LA Amstelveen,
The Netherlands
Tel: (31) 20 6418405 FAX: (31) 20 6434643
Toll Free in Benelux: 06 0993344
e-mail: nl@omega.com
Czech Republic:
ul. Rude armady 1868, 733 01 Karvina-
Hranice, Czech Repubic
Tel: 420 (69) 6311627 FAX: 420 (69)
6311114
e-mail: czech@omega.com
France:
9, rue Denis Papin, 78190 Trappes
Tel: (33) 130-621-400 FAX: (33)130-699-120
Toll Free in France: 0800-4-06342
e-mail: france@omega.com
Germany/Austria:
Daimlerstrasse 26, D-75392
Deckenpfronn, Germany
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540
Toll Free in Germany: 0800 82 66342
e-mail: germany@omega.com
United Kingdom: ISO 9002 Certified
One Omega Drive
Riverbend Technology Centre Northbank, Irlam,
Manchester, M44 5EX, England
Tel: 44 (161) 777-6611 FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
e-mail: sales@omega.com.uk
omega.com
TM
SAFETY INFORMATION
The following safety information must be observed to insure maximum
personal safety during the operation at this meter:
1. Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly.
2. This meter is not recommended for high voltage industrial use; for example, not for measurements of 440 VAC or 600 VAC industrial power mains. The unit is intended for use with low energy circuits to 600V AC/DC or high energy circuit to 250 VAC or DC.
3. Use caution when working above 60V dc or 30V ac rms. Such voltages pose a shock hazard.
4. When using the probes, keep your fingers behind the finger guards on the probes.
5. Measuring voltage which exceeds the limits of the meter may cause damage and expose the operator to a shock hazard. Always respect the voltage limits as stated on the meter.
6. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
1
SPECIFICATIONS
Display: Liquid crystal display (LCD) with a maximum reading of 2500. Polarity: Automatic, positive implied, negative polarity indication. Overrange: "OL"or "-OL" is displayed. Low battery indication: The " " is displayed when the battery voltage drops
below the operating level.
Measurement rate: 2.5 times per second, nominal. Operating Environment: 0°C to 40°C at < 70% relative humidity. Storage Temperature: -20°C to 60°C, 0 to 80% R.H. with battery removed from
meter.
Accuracy: Stated accuracy at 23°C ± 5°C, <75% relative humidity. Auto Power off: 30minutes after moving switch and push button no changes.turn
the meter off then on to resume operation.
Safety: According to EN61010-1 protection class II overvoltage category (CAT
II 600V) pollution degree 2.
Power: 4 pcs 1.5V (AAA size). Battery life: 500 hours typical. Dimensions: 170mm (H) x 44mm (W) x 40mm (D). Weight: 140g including batteries.
2
DC VOLTS
Ranges 250mV,2.5V,25V,250V,600V Resolution 100µV Accuracy ±(0.25%rdg + 5dgts) on 250mV range
±(0.25%rdg + 1dgt) on other ranges
Input impedance >10MW Overload & protection 600VDC or AC rms
AC VOLTS (50Hz - 500Hz)
Ranges 2.5V,25V,250V,600V Resolution 1mV Accuracy ±(0.75% of reading + 4dgts) on 2.5V to 600V Input impedance >10MW Overload & protection 600VDC or AC rms
RESISTANCE
Ranges 250W,2.5KW ,25KW ,250KW ,2.5MW ,25MW Resolution 100mW
3
4
Accuracy ±(0.3%rdg + 3dgts) on 250W range
±(0.3%rdg + 1dgt) on 2.5KW to 2.5MW ranges ±(3.5%rdg + 4dgts) on 25MW range
Open circuit volts 0.4Vdc Overload protection 500VDC or AC rms
CONTINUITY
Audible indication <100W Overload protection 500VDC or AC rms
DIODE TEST
Accuracy ±(3.0%rdg + 3dgts) Resolution 1mV Test current 0.25±0.2mA Test voltage <1.6V
CAPACITANCE
Ranges 250nF,2.5µF,25µF Accuracy ±(5.0%rdg + 10dgts) Overload protection 500VDC or AC rms
5
FREQUENCY (Autoranging)
Ranges 5.000Hz,50.00Hz,500.0Hz,5.000KHz Resolution 0.001Hz Accuracy ±(0.05%rdg + 2dgts) Sensitivity 1.0V rms min
Alternating Pulse or SINE wave signal on all range
Overload protection 500VDC or AC rms
DUTY CYCLE (2Hz to 1kHz)
Ranges 10% to 90% Resolution 0.1% Accuracy ±5dgts @ 2Vrms min Alternating Pulse signal Pulse Width 50µs Overload protection 500VDC or AC rms
TEMPERATURE (Autoranging)
Ranges -20°C to 1370°C Resolution 0.1°C,1°C Accuracy ±(2.0%rdg + 3°C) on -20°C to 500°C
±(3.0%rdg + 2°C) on 500°C to 1370°C
6
OPERATION
Before taking any measurements, read the Safety Information Section.
Always examine the instrument for damage, contamination (excessive dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed insulation. If any abnormal conditions exist do not attempt to make any measurements.
Auto Power-down mode
If unused for about 30 minutes, the tester will power-down automatically. If you should diseble Auto Power-down mode, press SELECT buttom when you turn on the meter.
Back-Light and Data-Hold Switch ( >2sec),( H ):
Press this button briefly to activate DATA-HOLD mode. The " H " annunciator is displayed. Press this button for 2 seconds to turn the Back-Light on. As this also activates the DATA-HOLD mode, briefly press the button to return to normal display. To turn the Back-Light off press again for 2 seconds.
7
SELECT Function Button (DC/AC),(
WW
WW
W/ / )
The SELECT Function button is Yellow in color. Press it to toggle to the alternate function (AC, Audible continuity, Diode and capacitance) shown in Yellow on the meter face.
Voltage Measurements
1. Connect the red test lead
to the "VW" jack and the black test lead to the
"COM" jack.
2. Set the Function/Range switch to the desired voltage function.
3. Connect the test leads to the device or circuit being measured.
4. For dc, a (-) sign is displayed for negative polarity; positive polarity is implied.
Frequency Measurements
1. Set the Function/Range switch to the ACV position, and then press Hz-% button toggle the voltage/frequency/duty cycle mode.
2. Connect the red test lead to the "VW" jack and the black test lead to the "COM" jack.
3. Connect the test leads to the point of measurement and read the frequency from the display.
8
Duty Cycle Measurements
1. Set the Function/Range switch to the ACV position, and then press Hz-% button toggle the voltage/frequency/duty cycle mode.
2. Connect the red test lead to the "VW " jack and the black test lead to the "COM" jack.
3. Connect the test leads to the point of measurement. The display will indicate 10% to 90% of the frequency duty cycle.
Resistance Measurements
1. Set the Function/Range switch to
WW
WW
W/ / position.
2. Remove power from the equipment under test.
3. Connect the red test lead to the "VW " jack and the black test lead to the "COM" jack.
4. Touch the probes to the test points. In ohms, the value indicated in the display is the measured value of resistance.
WARNING
The accuracy of the functions might be slightly affected, when exposed to a radiated electromagnetic field environment, eg, radio, telephone or similar.
9
Diode Tests
1. Set the Function/Range switch to
WW
WW
W/ / position.
2. Remove power from the equipment under test.
3. To toggle the W/continity/diode/capacitance modes, press SELECT switch.
4. Touch probes to the diode. A forward-voltage drop is about 0.6V (typical for a silicon diode).
5. Reverse probes. If the diode is good, "OL" is displayed. If the diode is shorted, ".000" or another number is displayed.
6. If the diode is open, "OL" is displayed, in both directions.
7. If the junction is measured in a circuit and a low reading is obtained with both lead connections, the junction may be shunted by a resistance of less than 1kW . In this case the diode must be disconnected from the circuit for accurate testing.
Continuity Measurements
1. Set the Function/Range switch to
WW
WW
W/ / position.
2. Remove power from the equipment under test.
3. To toggle the W/continity/diode/capacitance modes, press SELECT switch.
4. Connect the test leads to the two points at which continuity is to be tested. The buzzer will sound if the resistance is less than approximately 100W .
10
Capacitance Measurements
1. Set the Function/Range switch to
WW
WW
W/ / position.
2. Remove power from the equipment under test.
4. To toggle the W /continity/diode/capacitance modes, press SELECT switch.
5. Dischange capacitors before trying to measure it.
6. Connect the "+" lead to the "VW " jack and the "-" lead to the "COM" jack.
7. Read the capacitance directly from the display.
Temperature Measurements
1. Set the Function/Range switch to the "°C" position, and then press °C button toggle the ohm/°C mode.
2. Connect a type k thermocouple to the jack on the instrument. Place the probe or thermocouple tip on or in the material to be measured and take the temperature reading directly from the display.
WARNING
Remove test leads being measured
11
WARNING
Remove test leads before changing batteries or fuse
or performing any servicing.
Battery Replacement
Power is supplied by four 1.5V (AAA size) batteries. The " " appears on the LCD display when replacement is needed.To replace the batteries, remove the screw from the back of the meter and lift off the battery cover case. Remove the batteries from battery contacts.
Cleaning
Periodically wipe the case with a damp cloth and detergent, do not use abrasives or solvents.
MAINTENANCE
WARRANTY
OMEGA warrants this unit to be free of defects in materials and workmanship and to give satisfactory service for a period of 13 months from date of purchase. OMEGA W arr anty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA's customers receive maximum coverage on each product. If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA's Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon e xamination by OMEGA, if the unit is f ound to be def ective it will be repaired or replaced at no charge. Howev er, this WARRANTY is VOID if the unit shows e vidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA's control. Components which wear or which are damaged by misuse are not warranted. This includes contact points, fuses, and triacs.
OMEGA is glad to offer suggestions on the of use of its various products. Nevertheless, OMEGA only warrants that the parts manufactured by it will be as specified and free of defects
OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are e xclusive and the total liability of OMEGA with respect to this order, whether based on contract warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
Every precaution for accuracy has been taken in the preparation of this manual; however , OMEGA ENGINEERING, INC. neither assumes responsibility for any omissions or errors that ma y appear nor assumes liability for any damages that result from the use of the products in accordance with the information contained in the manual.
SPECIAL CONDITION: Should this equipment be used in or with any nuclear installation or activity, purchaser will indemnity OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the equipment in such a manner.
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected application.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) T O OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA'S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges , freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please hav e the following
information available BEFORE contacting OMEGA:
1. P.O. number under which the product was
PURCHASED.
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems relative
to the product.
FOR NON-WARRANTY REPAIRS, consult OMEGA for
current repair charges. Have the follo wing information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST of the repair.
2. Model and serial number of product , and
3. Repair instructions and/or specific problems relative to the product.
OMEGA's policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 1999 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readab le from, in whole or in part, without prior wr itten consent of OMEGA ENGINEERING, INC.
M-2895/0799
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA...Of Course!
HEATERS
þ Heating Cable
þ Cartridge & Strip Heaters
þ Immersion & Band Heaters
þ Flexible Heaters
þ Laboratory Heaters
PRESSURE/STRAIN AND FORCE
þ Transducers & Strain Gauges
þ Load Cells & Pressure Gauges
þ Displacement Transducers
þ Instrumentation & Accessories
FLOW/LEVEL
þ Rotameters, Gas Mass
Flowmeters & Flow Computers
þ Air Velocity Indicators
þ Turbine/Paddlewheel Systems
þ Totalizers & Batch Controllers
TEMPERATURE
þ
Thermocouple, RTD & Thermistor
Probes, Connectors, Panels &
Assemblies
þ Wire: Thermocouple, RTD &
Thermistor
þ
Calibrators & Ice Point References
þ
Recorders, Controllers & Process
Monitors
þ Infrared Pyrometers
ENVIRONMENTAL MONITORING AND CONTROL
þ
Metering & Control Instrumentation
þ Refractometers
þ Pumps & Tubing
þ Air, Soil & Water Monitors
þ Industrial Water & Wastewater
Treatment
þ pH, Conductivity & Dissolved
Oxygen Instruments
pH/CONDUCTIVITY
þ pH Electrodes, Testers &
Accessories
þ Benchtop/Laboratory Meters