It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification
of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains,
and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected application.
For immediate technical or application assistance:
Display: 3½ digit liquid crystal display (LCD) with a maximum reading of 1999.
Polarity: Automatic, positive implied, negative polarity indication.
Overrange: (OL) or (-OL) is displayed.
Zero: Automatic.
Low battery indication: the "" is displayed when the battery voltage drops
below the operating level.
Measurement rate: 2.5 times per second, nominal.
Operating Environment: 0°C to 40°C at < 70% relative humidity.
Storage Temperature: -20°C to 60°C, 0 to 80% R.H. with battery removed
from meter.
Accuracy: Stated accuracy at 23°C ± 5°C, <75% relative humidity.
Power: single standard 9-volt battery.
Battery life: 30 hours typical.
Dimensions: 200mm (H) x 90mm (W) x 40mm (D).
Weight: Approx. 14 oz. (400g) including battery.
Accessories: One pair test leads, One spare fuse installed, 9V battery and
Operating Instructions.
1
Page 4
RESISTANCE
Ranges: 20W,200W ,2KW,20KW,200KW ,2MW ,20MW
Resolution: 20Wrange 10mW
Accuracy: ±(1.0%rdg + 10dgts) on 20W range
±(0.3%rdg + 3dgts) on 200W range
±(0.3%rdg + 1dgt) on 2KW to 2MW ranges
±(2.0%rdg + 2dgts) on 20MW range
Open circuit volts:
6.5VDC on 20W to 200W Ranges
1.2VDC on other range
Overload protection: All ranges 25VDC or AC rms
Note: in the range 20
WW
, subtract residual offset reading from result.
W
WW
DIODE TEST
éé
Including: , , microwave , zener (<6.8V)
Test current: 3mA (approx)
Open voltage: 8VDC typical
Accuracy: ±(10%rdg +10dgts)
Display: forward junction voltage
Overload protection: 25VDC or AC rms
2
Page 5
CAPACITANCE
Ranges: 200pF,2nF,20nF,200nF,2µF,20µF,200µF,2000µF
Accuracy: ±(1.0%rdg + 3dgts) on 200pF to 200nF ranges
±(2.0%rdg + 3dgts) on 2µF to 200µF ranges
±(3.0%rdg + 3dgts) on 2000µF range: £1000µF
±(5.0%rdg + 10dgts) >1000µF
Test freguency:
1000Hz on 200pF to 2µF range
100Hz on 20µF to 200µF range
10Hz on 2000µF range
Temperature Coefficient:
£0.5µF: 0.1%/°C
>0.5µF: 0.2%/°C
Overload protection: 0.1A/250V fast blow fuse
Note: in lower range 200pF, 2nF subtract residual offset reading from
result with test leads opening.
3
Page 6
INDUCTANCE
Ranges: 200µH,2mH,20mH,200mH,2H,20H,200H
Accuracy: ±(3.0%rdg + 3dgts) on 200µH to 200mH ranges
±(5.0%rdg + 10dgts) on 2H to 200H ranges
Test freguency:
1000Hz on 200µH to 2H ranges
100Hz on 20H to 200H ranges
Temperature Coefficient:
£0.5H: 0.2%/°C
>0.5H: 0.5%/°C
Overload protection: 0.1A/250V fast blow fuse
Note: in lower range 200µH, 2mH subtract residual offset reading from
result with test leads being shorted.
4
Page 7
OPERATION
However, electrical noise or intense electromagnetic fields in the equipment may disturb the measurement circuit. Measuring instruments will also
respond to unwanted signals that may be present within the measurement
circuit. Users should exercise care and take appropriate precautions to avoid
misleading results when making measurements in the presence of electronic interference.
Capacitance
1.Dischange capacitors before trying to measure it.
2.Set the Range to the desired C range.
3.Insert the leads directly in to socket or test leads sockets.
4.Never apply an external voltage to sockets damage to the meter may
resuet.
5.Read the capacitance directly from the display.
Note: in lower range 200pF, 2nF subtract residual offset reading from
result with test leads opening.
5
Page 8
Inductance
1. Set the Ranges to the desired L range.
2. never apply an external voltage to the sockets damage to the meter may
result.
3. Insert the inductor leads derectly into sockets or test leards sockets.
4.Read the inductance directly from the display.
Note:in loues range 200µH, 2mH subtract residual offset reading from
result with test leads being shorted.
Resistance
1. Set the Function/Range switch to the desired resistance range.
2. Remove power from the equipment under test.
3. Connect the red test lead to the
-
" jack.
4. Touch the probes to the test points. In ohms, the value indicated in the
display is the measured value of resistance.
WARNING
The accuracy of the functions might be slightly affected, when exposed
to a radiated electromagnetic field environment, eg, radio, telephone or
similar.
Note:in the range 20
WW
, subtract residual offset reading from result.
W
WW
WW
W " + " jack and the black test lead to the
WW
6
WW
W"
WW
Page 9
Diode Tests and Continuity Measurements
1. Connect the red test lead to the " + " jack and the black test lead to the
" - " jack.
2. Set the Function/Range switch to the , , microwave , zener
éé
position.
3. Turn off power to the circuit under test.
4. Touch probes to diodes. A forward-voltage drop on diode. microwave diode
about 0.6VDC typical, LED about 2VDC typical, zener diode about its shown
voltage .
5. Reverse probes. if the diode is good, diode, microwave diode. LED about
open voltage 8VDC typical, zener diode about 0.7V typical.
6. If the junction is measured in a circuit and a low reading is obtained with both
lead connections, the junction may be shunted by a resistance of less than
1kW. In this case the diode must be disconnected from the circuit for
accurate testing.
7
Page 10
MAINTENANCE
WARNING
Remove test leads before changing battery or fuse
or performing any servicing.
Battery Replacement
Power is supplied by a 9 volt "transistor" battery. (NEDA 1604 IEC 6F22). The
"" appears on the LCD display when replacement is needed. To replace the
battery, remove the two screws from the back of the meter and lift off the
battery case. Remove the battery from battery contacts.
Fuse Replacement
If no capacitance and inductance measurements are possible, check for a
blown overload protection fuse.For access to fuses, remove the two screws
from the back of the meter and lift off the battery case. Replace F1 only with
the original type 0.1A/250V, fast acting fuse.
Cleaning
Periodically wipe the case with a damp cloth and detergent, do not use
abrasives or solvents.
8
Page 11
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months
from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year productwarranty to cover handling and shipping time. This ensures that OMEGA's customers receive maximum cov erage on each product.If
the unit should malfunction, it must be returned to the factory for evaluation. OMEGA's Customer Service Department will issue an
Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to
be defective it will be repaired or replaced at no charge. OMEGA's W ARRANTY does not apply to defects resulting from any action
of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair,
or unauthorized modification. This WARRANTY is V OID if the unit shows e vidence of having been tampered with or shows evidence
of being damaged as a result of excessive corrosion; or current, heat moisture or vibration; improper specification; misapplication;
misuse or other operating conditions outside of OMEGA's control. Components which wear are not warranted, including but not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However , OMEGA neither assumes responsibility
for any omissions or errors nor assumes liability for an y damages that result fr om the use of its products in accordance
with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it
will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND
WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE AND ALL IMPLIED WARRANTIES INCLUDING ANY
WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with
respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall
not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a "Basic Component" under 10
CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any
Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way,
OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and additionally, purchaser will
indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in
such a manner.
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA'S
CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then
be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the f ollowing
information available BEFORE contacting OMEGA:
1. P.O. number under which the product was PURCHASED.
2. Model and serial number of the product under warranty , and
3. Repair instructions and/or specific problems relative to the