Omega HHM17 User Manual

OMEGA
HHM17
Digital Multimeter
OMEGAnetSM On-Line Service
Internet e-mail
info@omega.com
Servicing North America:
USA: ISO 9001 Certified Canada:
One Omega Drive, Box 4047 976 Bergar
Stamford, CT 06907-0047 Laval (Quebec) H7L5A1
Tel: (203) 359-1660 Tel: (514) 856-6928
FAX: (203)359-7700 FAX: (514) 856-6886
e-mail: info@omega.com e-mail: info@omega.com
omega.com
TM
For immediate technical or application assistance:
USA and Canada: Mexico and Latin America:
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
SM
Tel: (95) 800-TC-OMEGA
SM
Customer Service: 1-800-622-2378 / 1-800-622-BEST
SM
FAX: (95) 203-359-7807
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
SM
En Español: (203) 359-7803
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA e-mail: espanol@omega.com
Servicing Europe:
Benelux: Germany/Austria:
Postbus 8034, 1180 LA Amstelveen, Daimlerstrasse 26, D-75392
The Netherlands Deckenpfronn, Germany
Tel: (31) 20 6418405 FAX: (31) 20 6434643 Tel: 49 (07056) 3017 FAX: 49 (07056) 8540
Toll Free in Benelux: 06 0993344 Toll Free in Germany: 0800 82 66342
e-mail: nl@omega.com e-mail: germany@omega.com
Czech Republic: United Kingdom: ISO 9002 Certified
ul. Rude armady 1868, 733 01 Karvina- One Omega Drive
Hranice, Czech Repubic Riverbend Technology Centre Northbank, Irlam,
Tel: 420 (69) 6311627 FAX: 420 (69) 6311114 Manchester, M44 5EX, England
e-mail: czech@omega.com Tel: 44 (161) 777-6611 FAX: 44 (161) 777-6622
France:
9, rue Denis Papin, 78190 Trappes Toll Free in England: 0800-488-488
Tel: (33) 130-621-400 FAX: (33)130-699-120 e-mail: sales@omega.com.uk
Toll Free in France: 0800-4-06342
e-mail: france@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing cer tification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected application.
1
SAFETY INFORMATION
The following safety information must be observed to insure maximum
personal safety during the operation at this meter:
1. Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly.
2. This meter is not recommended for high voltage industrial use; for example, not for measurements of 440 VAC or 600 VAC industrial power mains. The unit is intended for use with low energy circuits to 600VDC or AC or high energy circuit to 250 VAC or DC. Accidental misuse by connection across a high voltage, high energy power source when the meter is set up for mA measurement may be very hazardous.
3. Turn off power to the circuit under test before cutting, unsoldering, or breaking the circuit. Small amounts of current can be dangerous.
4. Use caution when working above 60V dc or 30V ac rms. Such voltages pose a shock hazard.
5. When Using the probes, keep your fingers behind the finger guards on the probes.
6. Measuring voltage which exceeds the limits of the multimeter may damage the meter and expose the operator to a shock hazard. Always recognize the meter voltage limits as stated on the front of the meter.
7. If the equipment is used in a manner not specified by the manufacturer, the protection provided the equipment may be impaired.
2
SPECIFICATIONS
Display: 3¾ digit (4000 counts), 9999counts (Frequency mode), 40 segments
analog bar graph and function units sign annunciators.
Polarity: Automatic, positive implied, negative polarity indication. Overrange: "4000"or"-4000" Most Significant Digit blinks. Low battery indication: the " " is displayed when the battery voltage drops
below the operating level.
Measurement rate: 2/sec,nominal. 1/sec, Capacitance and Frequency mode.
20/sec, Analog display.
Operating Environment: 0°C to 40°C at < 70% relative humidity. Storage Temperature: -20°C to 60°C, 0 to 80% R.H. with battery removed
from meter.
Accuracy: Stated accuracy at 23°C ± 5°C, <75% relative humidity. Safety: According to EN61010-1 protection class II overvoltage category
(CAT II 600V) pollution degree 2.
Auto Power off: 30minutes after rotary switch or mode changes. Power: single standard 9-volt battery. Battery life: 200 hours typical. Dimensions: 192mm (H) x 91mm (W) x 52.5mm (D). Weight: Approx. 365g including battery. Accessories: One pair test leads, One spare fuse installed, 9V battery and
Operating Instructions.
3
DC VOLTS
Ranges: 400mV,4V,40V,400V,600V Resolution: 100µV Accuracy: ±(0.25%rdg + 1dgt) on 400mV to 400V ranges
±(0.25%rdg + 3dgts) on 600V range
Input impedance: >10MW Overload protection: 600VDC or AC rms
AC VOLTS
Ranges: 400mV,4V,40V,400V,600V (400mV only Manual @50Hz-100Hz) Resolution: 100µV Accuracy:
Input impedance: >10MW Overload protection: 600VDC or AC rms
Range 100Hz 1KHz 10KHz 20KHz 400mV
4V 40V 400V 600V
±(0.75%rdg+5dgts)
±(0.75%rdg+5dgts)
N/A
±(1.5%rdg+4dgts)
±(2.0%rdg+30dgts)
N/A
4
DC CURRENT
Ranges: 4mA,40mA,400mA,10A Accuracy: ±(0.5%rdg + 1dgt) on mA ranges
±(2.0%rdg + 1dgt) on 10A range
Input protection: 0.5A / 250V fast blow ceramic fuse
10A / 600V fast blow ceramic fuse
AC CURRENT (50Hz - 500Hz)
Ranges: 4mA,40mA,400mA,10A Accuracy: ±(1.0%rdg + 4dgts) on mA ranges
±(3.0%rdg + 4dgts) on 10A range
Input protection: 0.5A / 250V fast blow ceramic fuse
10A / 600V fast blow ceramic fuse
RESISTANCE
Ranges: 400W ,4KW ,40KW ,400KW,4MW ,40MW Accuracy: ±(0.3%rdg + 4dgts) on 400W range
±(0.3%rdg + 1dgt) on 4KW to 4MW ranges ±(1.0%rdg + 4dgts) on 40MW range
Open circuit volts: 0.4Vdc Overload protection: 500VDC or AC rms
CONTINUITY
Audible indication: less than 40W±20W Overload protection: 500VDC or AC rms
5
DIODE TEST
Test current: 1.0mA±0.6mA Accuracy: ±(3.0%rdg + 3dgts) Open circuit volts: 3.0Vdc typical Overload protection: 500VDC or AC rms
CAPACITANCE Ranges: 4nF,40nF,400nF,4µF,40µF Accuracy: ±(2.0%rdg+20dgts) on 4n Frange(use 0ADJ)
±(2.0%rdg + 4dgts) on 40nF to 40µF ranges ±(5.0%rdg + 4dgts) above 40µF
Overload protection: 500VDC or AC rms
FREQUENCY (Autoranging)
Ranges: 100Hz,1KHz,10KHz,100KHz,700KHz Resolution: 0.01Hz Accuracy: ±(0.05%rdg + 2dgts) Sensitivity: 1.0V rms min Overload protection: 500VDC or AC rms
TRANSISTOR hFE
Ranges: 0 - 1000 Base current: 10µAdc approx. (Vce=3.0Vdc)
6
OPERATION
Before taking any measurements, read the Safety Information Section.
Always examine the instrument for damage, contamination (excessive dirt, grease, etc.) and defects. Examine the test leads for cracked or frayed insulation. If any abnormal conditions exist do not attempt to make any measurements.
PON Button
When the meter is automatic power-off, press the button to turns meter back on.
Note: Disable Automatic Power-off
If you press and hold down the (PON) button while turning the meter from OFF to on and select a function, the automatic power-off feature is disabled.
MEM Button
When (MEM) button is pressed, the "MEM" annunciator is displayed and the last reading is stored on the meter. If the meter power down automatically and the power back on by pressing (PON) button, these stored readings will remain in memory.
READ Button
Press (READ) Button to recall the data you stored in memory, the readings will be displayed on the LCD, the "HOLD"annunciator turn on,and the "MEM" annunciator will be displayed with a blink. The automatic power-off feature is
disabled. Press (HOLD) button to exit the READ mode.
7
REL
DD
DD
D Button
Press (REL) button to enter the Relative mode, the "RELD" annunciator turn on, zero the display, and store the displayed reading as a reference value. Press and hold down the (REL) button for 2 seconds to exit the relative mode.
MIN / MAX button
Press (MIN / MAX) button to enter the MIN MAX Recording mode. The minimum, maximum values are then reset to the present input, the readings are stored in memory, and the "HOLD" annunciator turns on. Push the button to cycle through the minimum (MIN), maximum (MAX), and present readings. The MIN or MAX annunciator turns on to indicate what value is being displayed. In the MIN MAX Recording mode, press (HOLD) button to stop the recording of readings, press again to restart recording. If recording is stopped, the minimum, maximum, or present values and analog diaplay are frozen. In the MIN MAX Recording mode, when a new minimum value is exceed the actual minimum readings or a new maximum value is overload, the minimum or maximum value will held on the display, but the analog display continues to be active.
8
HOLD Button
Press (HOLD) button to toggle in and out of the Data Hold mode, except if you are already in the MIN MAX Recording mode. In the Data Hold mode, the "HOLD" annunciator is displayed and the last reading is held on the display, the beeper emits a tone, and the automatic power-off feature is disable. Pressing (MIN / MAX) button when you are in the Data Hold mode causes you to exit Data Hold and enter the MIN MAX Recording mode. In the MIN MAX Recording mode, press (HOLD) button to stop the recording of readings, press (HOLD) again to resume recording.
RANGE Button
Press (RANGE) button to select the Manual Range mode and turn off the "AUTO" annunciator. (The meter remains in the range it was in when manual ranging was selected). In the Manual Range mode. each time you press (RANGE) button, the range (and the input range annunciator) increments, and a new value is displayed. To exit the Manual Range mode and return to autoranging, press and hold down (RANGE) button for 2 seconds. The "AUTO" annunciator turns back on.
9
Alternate Function Button
The Alternate Function button is Blue in color. Press it to toggle to the alternate function (AC, Audible continuity and Capacitance) shown in Blue on the meter face.
Voltage Measurements
1. Connect the red test lead to the "VW" jack and the black test lead to the "COM" jack.
2. Set the Function/Range switch to the desired voltage range and press the Blue toggle button to select AC or DC. The meter will automatically select the best voltage range.
3. Connect the test leads to the device or circuit being measured.
4. For dc, a (-) sign is displayed for negative polarity; positive polarity is implied.
Current Measurements
1. Set the Function/Range switch to the desired current range and press the Blue toggle button to select AC or DC.
2. For current measurements less than 400mA, connect the red test lead to the mA jack and the black test lead to the COM jack.
3. For current measurements of 400mA or greater, connect the red test lead to the 10A jack and the black test lead to the COM jack.
4. Remove power from the circuit under test and open the normal circuit path where the measurement is to be taken. Connect the meter in series with the circuit.
5. Use caution when measuring 10 amps on 10A range for 60s, please waiting for 10 minutes for next measurement of 10 amps for safety reason.
10
Resistance and Continuity Measurements
1. Set the Function/Range switch to the desired resistance range or continuity position.
2. Remove power from the equipment under test.
3. Connect the red test lead to the "VW" jack and the black test lead to the "COM" jack.
4. Touch the probes to the test points. In ohms, the value indicated in the display is the measured value of resistance.
5. Press the Blue Alternate Function button to select Audible Continuity.In continuity test, the beeper sounds continuously, if the resistance is less than 40W.
WARNING
The accuracy of the functions might be slightly affected, when exposed to a radiated electromagnetic field environment, eg, radio, telephone or similar.
Diode Tests
1. Connect the red test lead to the "VW" jack and the black test lead to the "COM" jack.
2. Set the Function/Range switch to the "
" position.
3. Turn off power to the circuit under test.
4. Touch probes to the diode. A forward-voltage drop is about 0.6V (typical for a silicon diode).
5. Reverse probes. If the diode is good, display will be between 2.800V and
3.200V. If the diode is shorted, ".000" or another number is displayed.
6. If the diode is open, display will be between 2.800V and 3.200V, in both directions.
7. If the junction is measured in a circuit and a low reading is obtained with both lead connections, the junction may be shunted by a resistance of less than 1kW. In this case the diode must be disconnected from the circuit for accurate testing.
Capacitance Measurements
1. Set the Function/Range switch to the desired " " range and press the Blue toggle button to select Cx.
2. Never apply an external voltage to the Cx sockets. Damage to the meter may result.
3. Insert the capacitor leads directly into the Cx socket.
4. Read the capacitance directly from the display.
Transistor Gain Measurements
1. Set the Function/Range switch to the desired hFE range (PNP or NPN type transistor).
2. Never apply an external voltage to the hFE sockets. Damage to the meter may result.
3. Plug the transistor directly into the hFE socket. The sockets are labeled E, B and C for emitter, base, and collector.
4. Read the transistor hFE (dc gain) directly from the display.
Frequency Measurements
1. Set the Function/Range switch to the Hz position.
2. Connect the red test lead to the "VW" jack and the black test lead to the "COM" jack.
3. Connect the test leads to the point of measurement and read the frequency from the display.
11
MAINTENANCE
WARNING
Remove test leads before changing battery or fuse
or performing any servicing.
Battery Replacement
Power is supplied by a 9 volt "transistor" battery. (NEDA 1604, IEC 6F22). The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two screws from the back of the meter and lift off the battery case. Remove the battery from battery contacts.
Fuse Replacement
If no current measurements are possible, check for a blown overload protec­tion fuse. There are two fuses; F1 for the "mA" jack and F2 for the "10A" jack. For access to fuses, remove the four screws from the back of the meter and lift off the battery cover and case. Replace F1 only with the original type 0.5A/ 250V, fast acting ceramic fuse. Replace F2 only with the original type 10A/ 600V, fast acting ceramic fuse.
Cleaning
Periodically wipe the case with a damp cloth and detergent, do not use abrasives or solvents.
12
WARRANTY / DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA's customers receive maximum cov erage on each product.If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA's Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defectiv e it will be repaired or replaced at no charge. OMEGA's W ARRANTY does not apply to def ects resulting from any action of the purchaser, including b ut not limited to mishandling, improper interf acing, oper ation outside of design limits, improper repair, or unauthorized modification. This W ARRANTY is V OID if the unit shows e vidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA's control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However , OMEGA neither assumes responsibility for any omissions or err or s nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO O THER W ARRANTIES OR REPRESENTA TIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITA TION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with respect to this order , whether based on contract, warranty , negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold b y OMEGA is not intended to be used, nor shall it be used: (1) as a "Basic Component" under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic W ARRANTY / DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner .
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA'S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR W ARRANTY RETURNS, please have the following
information available BEFORE contacting OMEGA:
1. P.O. number under which the product was PURCHASED.
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the
product.
FOR NON-WARRANTY REPAIRS, consult OMEGA for
current repair charges. Have the following information available BEFORE contacting OMEGA:
1. P.O. number to cover the COST of the repair.
2. Model and serial number of product , and
3. Repair instructions and/or specific problems relative to the product.
OMEGA's policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 1999 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable from, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.
M-2855/0799
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA...Of Course!
TEMPERATURE
þ Thermocouple, RTD & Thermistor
Probes, Connectors, Panels & Assemblies
þ Wire: Thermocouple, RTD & Thermistor
þ Calibrators & Ice Point References
þ Recorders, Controllers & Process Monitors
þ Infrared Pyrometers
PRESSURE/STRAIN AND FORCE
þ Transducers & Strain Gauges
þ Load Cells & Pressure Gauges
þ Displacement Transducers
þ Instrumentation & Accessories
FLOW/LEVEL
þ Rotameters, Gas Mass Flowmeters
& Flow Computers
þ Air Velocity Indicators
þ Turbine/Paddlewheel Systems
þ Totalizers & Batch Controllers
pH/CONDUCTIVITY
þ pH Electrodes, Testers & Accessories
þ Benchtop/Laboratory Meters
þ Controllers, Calibrators, Simulators
& Pumps
þ Industrial pH & Conductivity Equipment
DATA ACQUISITION
þ Data Acquisition &
Engineering Software
þ Communications-Based
Acquisition Systems
þ Plug-in Cards for Apple, IBM
& Compatibles
þ Datalogging Systems
þ Recorders, Printers & Plotters
HEATERS
þ Heating Cable
þ Cartridge & Strip Heaters
þ Immersion & Band Heaters
þ Flexible Heaters
þ Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
þ Metering & Control Instrumentation
þ Refractometers
þ Pumps & Tubing
þ Air, Soil & Water Monitors
þ Industrial Water & Wastewater
Treatment
þ pH, Conductivity & Dissolved Oxygen
Instruments
Loading...