It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
Czech Republic:
Frystatska 184, 733 01 Karvina´, Czech Republic Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
Germany/Austria:
Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Tel: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
Servicing Europe:
U.S.A. and Canada:
Sales Service: 1-800-826-6342/1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378/1-800-622-BEST
®
Engineering Service: 1-800-872-9436/1-800-USA-WHEN
®
U.S.A.: ISO 9001 Certified
One Omega Drive, Box 4047 Stamford, CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Servicing North America:
For immediate technical or application assistance:
Mexico:
En Espan~ol: (001) 203-359-7803 FAX: (001) 203-359-7807 e-mail: espanol@omega.com
info@omega.com.mx
United Kingdom:
ISO 9002 Certified
One Omega Drive River Bend Technology Centre Northbank, Irlam Manchester M44 5BD United Kingdom Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622
Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
Canada:
976 Bergar Laval (Quebec) H7L 5A1, Canada Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
®
TABLE OF CONTENTS
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Key Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operating Instructions . . . . . . . . . . . . . . . . . . . . . 4 – 11
What’s in the case . . . . . . . . . . . . . . . . . . . . 4 – 5
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 – 8
Normal Operation . . . . . . . . . . . . . . . . . . . . . 8 – 9
Holding and Storing Measurements. . . . . . . . . . 9
Automatic vs. Manual Datalogging . . . . . . 9 – 10
Transferring Data from the SD Card
to a Computer . . . . . . . . . . . . . . . . . . . . . . 10 – 11
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Maintenance & Troubleshooting Tips . . . . . . . . . . . . 13
Optional Accessories. . . . . . . . . . . . . . . . . . . . . . . . . 13
2
INTRODUCTION
Thank you for purchasing the HHLM112SD UV Light Meter with Data Logging SD card.
Please read this user’s manual carefully and thoroughly
before using the instrument.
The HHLM112SD is a general-purpose handheld instrument suitable for measuring the intensity of light in commercial, industrial and consumer settings. Applications include photography, videography, construction, engineering and surveillance.
This light meter measures light intensity using either of two units: lux or foot-candles. Th
e lux is the standard international unit of illuminance, or how much luminous flux is spread over a given area. One can think of luminous flux as a measure of the total “amount” of visible light present, and illuminance as a measure of the intensity of illumination on a surface. A given amount of light will illuminate a surface more dimly if it is spread over a larger area, so illuminance is inversely proportional to area.
Th
e standard international unit of luminous flux is the lumen, and one lux is equal to one lumen/m2. For perspective on lux readings from the natural world, consider that a moonless clear night sky has an illuminance of about 0.002 lux, while direct sunlight has an illuminance of 30,000 to 130,000 lux.
The footcandle (or lumen per square f
oot) is a non-standard unit of illuminance.
Like th
e Btu, it is mainly in common use only in the U.S., particularly in construction­related engineering and in building codes. Because lux and footcandles are different units of the same quantity, it is perfectly valid to convert footcandles to lux and vice versa. One footcandle (ft-cd) = 10.764 lux.
The HHLM112SD has the performance and features needed to satis
fy the most
deman
ding aspects of the above applications. It also has the ability to measure surface temperature, using an optional Type K of Type thermocouple. A light probe comes standard with the instrument.
Because it is microprocessor-based, the HHLM112SD can make full use of the portability, reliability and large storage capacities that SD memory cards offer. Measurements of light intensity or temperature c
an be made automatically at any
of 11
sampling rates between one second and one hour. After time-stamping and storing the measurements on an SD card plugged into the instrument (a process called datalogging), the user can remove the card and transfer the measurements to a laptop or desktop computer either directly or via a USB card reader.
The HHLM112SD has a backlit 2-1/2 in. diagonal display and is powered
by six “AA”
Alkal
ine batteries or an optional 9-VDC AC adapter.
3
KEY FEATURES
Includes light probe
Measures light intensity within three automatically switched full-scale
ranges: 2,000 lux, 20,000 lux and 100,000 lux (200 ft-cd, 2,000 ft-cd and 10,000 ft-cd)
Selectable lux or foot-candle unit
Displays maximum and minimum readings
Performs real-time automatic datalogging at sampling time settable from
1 second to 1 hour
Also supports manual logging and chang
ing
of card storage location
Accepts SD memory cards of up to 16 GB capacity
Big (2.5 in. diagonal) front-panel green backlit LCD is easy to read
Auto power off function
Powered by six “AA” batteries or optional 9V AC/DC adapter
OPERATING INSTRUCTIONS
WHAT’S IN THE CASE
The HHLM112SD comes fully assembled in a carrying case along with a light probe with a protective cover, a 2 GB SD memory card and this user’s manual.
Figure 1 shows all of the controls and indicators on the front, back, top, bottom and right side of the HHLM112SD. It also shows (at right) the main components of the supplied light probe. Familiarize yourself with their positions