HH806AW
Internet e-mail
info@omega.com
WARNING
To avoid electrical shock, do not use this instrument
when working voltages at the measurement surface
over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature
measurement in microwave ovens.
CAUTION
Repeated sharp flexing can break the thermocouple
leads. To prolong lead life, avoid sharp bends in the
leads, especially near the connector.
FEDERAL COMMUNICATIONS
COMMISSION
This device complies with Part 15 of the FCC Rules.
Operatio
n is subject to the following two condi-
tions:(1
) this device may not cause harmful interfer-
ence
, and (2) this device must accept any interference
received
, including interference that may cause unde-
sired operation.
NOTE
This equipment has been tested and found to comply
wit
h the limits for a Class B digital device, pursuant to
Par
t 15 of the FCC Rules. These limits are designed to
provid
e reasonable protection. This equipment gener-
ates
, uses and can radiate radio frequency energy and,
if not installed and used in accordance with the instructions
, may cause harmful interference to radio
communications
. However, there is no guarantee that
interferenc
e will not occur in a particular installation If
thi
s equipment does cause harmful interference to radio or television reception, which can be determined
by turning the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and
receiver
.
- Connect the equipment into an outlet on a circuit dif-
ferent from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV tech-
nician for help.
Shielde
d interface cables must be used in order to
Shop online at: omega.com e-mail: info@omega.com
For latest product manuals: omegamanual.info
MULTILOGGER THERMOMETER
OMEGAnet On-Line Service
omega.com
Servicing North America:
U.S.A: One Omega Drive, Box 4047
ISO 9001
Certified e-mail: info@omega.com
Canada: 976 Bergar
U.S.A Sales Service: 1-800-826-6342/1-800-TC-OMEGA
and Customer Service: 1-800-622-2378/1-800-622-BEST
Canada:
Mexico: En Espan ol: (001) 203-359-7803
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
Czech Frystatska 184, 733 01 Karviná, Czech Republic
Republic:
France:
Germany/ Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Austria:
United One Omega Drive, River Bend Technology Centre
Kingdom: Northbank, Irlam, Manchester
ISO 9002 M44 5BD United Kingdom
Certified Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622
It is the policy of OMEGA Engineering, Inc. to comply with all
worldwide safety and EMC/EMI regulations that apply. OMEGA
is constantly pursuing certification of its products to the
European New Approach Directives. OMEGA will add the CE
mark to every appropriate device upon certification.
The information contained in this document is believed to be
correct, but OMEGA accepts no liability for any errors it
contains, and reserves the right to alter specifications without
notice.
should not be used for, human applications.
Stamford, CT 06907-0047
Tel: (203) 359-1660 FAX: (203) 359-7700
Laval (Quebec) H7L 5A1, Canada
Tel: (514) 856-6928 FAX: (514) 856-6886
e-mail: info@omega.ca
For immediate technical or application assistance:
e-mail: espanol@omega.com
FAX: (001) 203-359-7807 info@omega.com.mx
Servicing Europe:
Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643
Toll Free in Benelux: 0800 0993344
e-mail: sales@omegaeng.nl
Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114
Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
11, rue Jacques Cartier, 78280 Guyancourt, France
Tel: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427
Toll Free in France: 0800 466 342
e-mail: sales@omega.fr
Tel: + 49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29
Toll Free in Germany: 0800 639 7678
e-mail: info@omega.de
Toll Free in United Kingdom: 0800-488-488
e-mail: sales@omega.co.uk
These products are not designed for use in, and
®
®
INTRODUCTION
The instrument is a portable digital thermometer that
s external thermocouples of type K, J, R, S, T, E,
measure
N. The thermocouples types comply with the N.I.S.T. –
S 90 standard reference temperature/voltage tables.
IT
e thermometer features a dual thermocouple input, an
Th
e T/C offset and an USB interface for uploading
adjustabl
data to a PC using optional software and cable.
SAFETY INFORMATION
comply with emission limits.
s or modifications not expressly approved by
Change
e party responsible for compliance could void the
th
user‘s authority to operate the equipment.
WIRELESS NOTE
Wireless receiver must keep a distance of at least 30cm
from the meter and meter to meter distance must be at
least 20cm.
SPECIFICATIONS
ELECTRICAL
Measurement Range:
K-TYPE (0.1°) -200°C to 1372°C (-328°F to 2501°F)
J-TYPE (0.1°) -210°C to 1200°C ( -346°F to 2192°F)
T-TYPE (0.1°) -200°C to 400°C (-328°F to 752°F)
E-TYPE (0.1°) -210°C to 1000°C (-346°F to 1832°F)
R-TYPE (1°) 0°C to 1767°C (32°F to 3212°F)
S-TYPE (1°) 0°C to 1767°C ( 32°F to 3212°F)
N-TYPE (0.1°) -50°C to 1300°C ( -58°F to 2372°F)
*Based on the ITS-90 temperature standard.
According to temperature standard ITS-90.
Accuracy:
K/J/T/E-TYPE
±(0.05% rdg + 0.3°C) on -50°C to 1372°C
±(0.05% rdg + 0.7°C) on -50°C to -210°C
±(0.05% rdg + 0.6°F) on -58°F to 2501°F
±(0.05% rdg + 1.4°F) on -58°F to -346°F
N-TYPE
±(0.05% rdg + 0.8°C) on -50°C to 0°C
±(0.05% rdg + 0.4°C) on 0°C to 1300°C
±(0.05% rdg + 1.6°F) on -58°F to 32°F
±(0.05% rdg + 0.8°F) on 32°F to 2372°F
R/S-TYPE
±(0.05% rdg + 2°C) on 0°C to 1767°C
±(0.05% rdg + 4°F) on 32°F to 3212°F
Temperature Coefficient:
0.1 times the applicable accuracy specification per °C
from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and
82°F to122°F).
Input Protection:
24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Reading Rate: 2.5 time per second.
Wireless Features:
Frequency range: 910~920MHz
Low current consumption less than 1mA
e transmitting distance can reach 25M without
Th
netic interference.
External Connections:
1. USB Port
2. DC power JACK(12V)
ENVIRONMENTAL
Ambient Operating Ranges:
Celsius or Fahrenheit user-selectable
0°C to 50°C (32°F to 122°F) <80% R.H.
Storage Temperature:
-20°C to 60°C (-4°F to 140°F) <70% R.H.
GENERAL
Display:
There are three display areas on the HH806 series LCD
(liquid crystal display). The Main and Second displays
are 4 _ digits with a maximum reading of 19999. These
are used for displaying the value of T1 or T2. The Third
display is for the date, time, or the differential value of
T1 to T2.
Overload: “----.-” or “OL” is display.
Battery: 1.5V x 4 PCS (SIZE AAA) UM-4 R03.
Battery Life: 120 hours typical with carbon zinc battery.
Dimensions: 160mm(H) x 83mm(W) x 38mm(D)
.
Weight: Approx. 365g including batteries.
Supplied Thermocouple:
1 meter (40”) type K insulated beaded wire thermocouple
. Maximum insulation temperature is 482°C (900°F).
Thermocoupl
e accuracy is ±1.1°C or 0.4% of reading
(whichever is greater) from 0°C to 1250°C.
1. OPERATIONAL MODE
There are three operation modes-Normal, Shift, and
Setup Mode.
NORMAL MODE:
Thi
s is the default mode, the operating functions for
th
e normal mode are printed on the top of each button
in white.
SHIFT MODE:
Th
e operating functions for the shift mode are printed
in gray on the buttons. While in the normal mode,
pres
s the SHIFT button to switch to shift mode. At the
lower-right corner of the display, the word “Shift” is
displaye
d to indicate shift mode. To switch back to
normal mode, press the SHIFT button again.
SETUP MODE:
Pres
s the set[ ] button in normal mode to switch to
setu
p mode, the indicator “SET” is shown on the left
sid
e of the display. To switch back to normal mode,
press SET[ ] button again.
WIRELESS MODE:
Pres
s the “ ” key for more than two seconds to start
wireles
s function. Press “
”
fo
r another two seconds
to stop wireless function. The wireless mode will shut
down if there is no wireless signal for two minutes.
SE
T CH/ID to 00,00 by pressing “T1-T2” key and
“ ” power key for more than six seconds when the
mete
r with the meter powered down. The second dis-
pla
y will show 00, which means that the channel and
ID has been set to 00.
To check the channel and ID of the meter:
Whe
n the meter is off press “ ” key and “ ” for
5 seconds. The LCD’s main display will show channel
number, the second display will show ID number.
2. NORMAL MODE
The following functions can only be used in the normal
mode.
(1) “ ” Power Button
The “ ” button turns the thermometer on or off.
Whe
n the meter is in MAX/MIN record mode, the
power off function is disabled.
(2) “[LIMITS]” Button (only Main display)
The limits function will alert the user when a measuremen
t exceeds a specified limit. To set the limit val-
ues
, refer to limits function in the setup mode. Press
th
e [LIMITS] button to activate the limits function; the
word “LIMIT” should be displayed on the LCD.
Whe
n the value of the main display exceeds the Hi
limit
, the word “Hi” will be displayed and the ther-
momete
r will beep in a pulsed tone. If the value of the
mai
n display is lower then the Lo limit, the word “Lo”
wil
l be displayed and the thermometer will beep in a
continuou
s tone To exit the limits function, press the
[LIMITS] button.
(3) “ ” Button
The backlight function is represented by this button
“
”. Pressing the button will turn on or off the LCD
backlight. The backlight will turn off automatically
after.
(4) “SAVE/READ” Button
The read data function works in conjunction with the
save function in the shift mode, it is used for reading
d data. The save function can be activated in shift
save
. Press the SAVE/READ button to read saved
mode
; the word “READ” should be displayed on the
data
. To navigate around the save data table, press the
LCD
y “SECOND” button until the “#” sign is dis-
overla
d on the second display. The location of the read
playe
r within the saved data table will be displayed.
pointe
e arrow buttons on the overlay are used for scrolling
Th
h the saved data. Press the smaller arrows “ ”
throug
or “ ” to step through the data one at a time. Press the
large
r arrows “ ” or “ ” to step through the data
n at a time. Pressing the overlay “ESC” button deac-
te
tivates the read data function.
(5) “LOG/READ” Button
The read log function works in conjunction with the log
. It is used for reading logged data. The log
function
n can be activated in the shift mode. Press the
functio
LOG/REA
th
overla
displa
play
ren
th
button
th
“ ” to step through the logged data or groups one at a
time
throug
th
D button to activate the log read function;
e word “READ” is displayed on the LCD. Press the
y SECOND button to rotate through following
y menus: T1, T2, GRP, and #. T1 and T2: Dis-
s the T1 or T2 saved data. GRP: Displays the cur-
t group number. #: Displays the current location of
e read pointer within a selected group. The arrow
s on the overlay are used for scrolling through
e data and groups. Press the smaller arrows “ ” or
. Press the larger arrows“” or “ ” to step
h the data or groups ten at a time. To navigate
e logg ed data and groups , press the overla y
SECON
D button until GRP appears in the second dis-
y panel. Then select the group using the arrows.
pla
s the SECOND button again until the “#” sign is
Pres
displayed
lected group will be displayed. Use the arrows to
scrol
button deactivates the read data function
. The location of the read pointer in the se-
l through the data. Pressing the overlay “ESC”
(7) HOLD Mode (only Main display)
When HOLD mode is selected, the thermometer holds
e present readings and stops all further measure-
th
. To activate the data hold mode, press the
ments
D button, and “HOLD” is displayed on the LCD.
HOL
g the HOLD button again cancels the function,
Pressin
d the instrument will automatically resume meas-
an
urements.
(8 ) M IN/M AX wit h time record Mode
(only Main display)
The MIN/MAX function records the highest and low-
t value recorded, and it calculates the average read-
es
, and the differences of MAX to MIN. Press
ing
MIN/MA
mode
or maximum measurement is recorded. Press the
MIN/MA
readings
MIN
Th
ment. AVG: The average values of the measurements.
Thi
tion
an
no
cordin
of MIN, MAX and AVG data, the MIN/MAX function
ca
th
automati
REL
also disabled.
X button to enter the MIN/MAX recording
. The beeper emits a tone when a new minimum
X button again to rotate through the current
: MAX: The highest measurement recorded.
: The lowest measurement recorded. MAX-MIN:
e difference of the highest and the lowest measure-
s mode works in conjunction with the hold func-
, pressing the HOLD button will stop the recording
d measurements (Previously recorded readings are
t erased). Press HOLD button again to resume re-
g and measurements. To prevent accidental loss
n only be cancelled by pressing and holding down
e MIN MAX key for more then 2 seconds. The
c power-off feature, and the power, °C/°F,
, SET, Hi/Lo Limits, TYPE, T1/T2 buttons are
(10) “T1/T2” Button (Main display)
The input selection button [T1/T2] selects the input for
e main display, T1 thermocouple or T2 thermocou-
th
. Press the T1/T2 button to switch between the two
ple
. When meter is turned on, it is set to the display
inputs
that was last in use.
(11) “T1/T2” Button (Second display)
The input selection button [T1/T2] selects the input for
e second display, T1 thermocouple or T2 thermo-
th
. Press the T1/T2 button to switch between the
couple
o inputs. When meter is turned on, it is set to the
tw
display that was last in use.
(12) “T1-T2/TIME” Button (Third display)
The input selection button [T1-T2] selects the system
e and date, or the differential between the two
tim
thermocouple
T1-T
mete
in use.
s (T1-T2) for the third display. Press the
2 button to switch the display options. When
r is turned on, it is set to the display that was last
3. SHIFT MODE
The following functions can only be used in the shift
mode.
(2) “°C/°F” Button
Press the °C/°F button to select the temperature
s can be displayed in Celsius (°C) or Fahre
reading
t (°F). When the thermometer is turned on, it
hei
to the temperature scale that was last in use.
(4) “SAVE” Button
The save function stores the T1, T2 data in a nonvola
e memory. Press the SAVE button to save the
til
t data, the word SAVE is displayed to indicate
ren
a has been saved. The built in memory can store
dat
to 256 data. The data can be read using the read
tion in the normal mode.
(5) “LOG” Button
The data log function continuously records the
g to a specified time interval. The time
accordin
l can be set using the interval setup function [INVT]
va
in the set up mode. Press the LOG button to activate
e log function; the indicators “LOG” and “MEM”
th
l be displayed on the LCD. There are 16 groups
wil
e used for storing the log data. A maximum
ar
0 data points can be saved in one log session.
16,00
s the LOG button again to exit the data log
Pres
tion.
(6) “CLR ?” Button
The CLR function clears all the saved and logged
in memory. When the CLR button is pressed, indicator
” is displayed and the “CLR” on upper-right
“MEM
D will blink. Pressing the “ENTER” button printed
LC
on the overlay in white will clear all saved and logged
. Press “ESC” button to exit this function without
data
clearing data.
(7) “REL” Button (Main display)
The relative value function can be used for comparing
a saved reference value with other measurements.
s the “REL” button to store the current measur
Pres
t as the reference value, and “REL” should be
men
d on the right part of the LCD. The next
playe
t will display the current value compared
uremen
e value. Press “REL” button again to clear
referenc
e value and deactivate the relative value
referenc
urement function.
(8) “APO” Button
Press the APO button to turn the “Auto power
n on or off. When this function is enabled,
functio
r “APO” is shown at the upper left part
indicato
. When APO (Auto power off) is enabled, it
LCD
automaticall
presse
e default time for APO is 5 minutes). Press power
(th
button to resume operation.
y turn the thermometer off no button
d for a period longer than the set time inte
(10) “TYPE” Button (Main display)
Press this button to change the type of thermocouple
e main display (K/J/T/E/R/S/N). If the inputs
th
n and second display are the same, then pressing
mai
s button will change the thermocouple type for
thi
displays.
(11) “TYPE” Button (Second display)
Press this button to change the type of thermocouple