Omega FTB890 User guide

Page 1
User’s Guide
Shop online at
omega.com
For latest product manuals:
omegamanual.info
Page 2
OMEGAnet®Online Service Internet e-mail
U.S.A.: Omega Engineering, Inc., One Omega Drive, P.O. Box 4047
ISO 9001 Certified
Stamford, CT 06907-0047 Toll-Free: 1-800-826-6342 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: [email protected]
Canada: 976 Bergar
Laval (Quebec), Canada H7L 5A1 Toll-Free: 1-800-826-6342 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: [email protected]
For immediate technical or application assistance:
U.S.A. and Canada: Sales Service: 1-800-826-6342/1-800-TC-OMEGA
Customer Service: 1-800-622-2378/1-800-622-BEST Engineering Service: 1-800-872-9436/1-800-USA-WHEN
®
®
®
Mexico: En Español: 001 (203) 359-7803 FAX: (001) 203-359-7807
Servicing Europe:
Benelux: Managed by the United Kingdom Office
Czech Republic: Frystatska 184
France: Managed by the United Kingdom Office
Germany/Austria: Daimlerstrasse 26
United Kingdom: OMEGA Engineering Ltd.
ISO 9001 Certified
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
Toll-Free: 0800 099 3344 TEL: +31 20 347 21 21 FAX: +31 20 643 46 43 e-mail: [email protected]
733 01 Karviná, Czech Republic Toll-Free: 0800-1-66342 TEL: +420-59-6311899 FAX: +420-59-6311114 e-mail: [email protected]
Toll-Free: 0800 466 342 TEL: +33 (0) 161 37 29 00 FAX: +33 (0) 130 57 54 27 e-mail: [email protected]
D-75392 Deckenpfronn, Germany Toll-Free: 0 800 6397678 TEL: +49 (0) 7059 9398-0 FAX: +49 (0) 7056 9398-29 e-mail: [email protected]
One Omega Drive, River Bend Technology Centre, Northbank Irlam, Manchester M44 5BD England Toll-Free: 0800-488-488 TEL: +44 (0)161 777-6611 FAX: +44 (0)161 777-6622 e-mail: [email protected]
Page 3
TABLE OF CONTENTS
General Information ........................1
Installation ......................................3
Operation ........................................4
Calibration ......................................5
Maintenance ...................................6
Troubleshooting ..............................8
Specications .................................9
Model Numbers ............................12
Parts & Accessories......................12
Warranty/Disclaimer .....................13
Return Requests/Inquiries ............13
Series Turbine Meter System prior to installation and use. If you need assistance, contact the Omega Flow Application Department.
SAFETY INSTRUCTIONS
This symbol is used throughout
the manual to call your attention to safety messages.
WARNING
WARNINGS
you to the poten-
alert
tial for personal injury.
GENERAL INFORMATION
This manual will assist you in installing and maintaining your FTB890 Series turbine meter and associated display electronics. The turbine meters can be purchased with or without the display electronics. (See Figure 1)
Figure 1
Computer Electronics
Rotor Assembly
Turbine Housing
When purchased with the display electronics, optional accessory mod-
ules are available for eld installation.
Information on these accessories is contained in separate manuals. Calibration details using the display electronics are given in this manual.
When purchased without the display
(Sufx - ND) an optional eld installable
Pulse Output Module (FLSC790-P­ND) is available and described in a separate manual.
For best results, take the time to fully acquaint yourself with all information about all components of your FTB890
CAUTION
CAUTIONS call
your attention to practices or pro­cedures which may damage your equipment.
NOTES give information that can
improve efciency of operations.
It is your responsibility to make sure that all operators have access to adequate instructions about safe op­erating and maintenance procedures.
Read Me!
For your safety, review the major warnings and cautions below before operating your equipment.
WARNING
The apparatus enclosure may contain aluminum and is considered to con­stitute a potential risk of ignition by impact or friction. Care must be taken into account during installation and use to prevent impact or friction.
WARNING
Part of the enclosure is constructed from plastic. To prevent the risk of electrostatic sparking the plastic surface should only be cleaned with a damp cloth.
1
Page 4
1. This equipment is approved to handle only uids that are compat­ible with all materials of construc­tion.
2. When measuring ammable liquids, observe precautions against re or explosion.
3. When handling hazardous liquids, always follow the liquid manufac­turer’s safety precautions.
4. When working in hazardous
environments, always exercise
appropriate safety precautions.
5. Always dispose of used cleaning solvents in a safe manner accord­ing to the solvent manufacturer’s instructions.
6. During turbine removal, liquid may spill. Follow the liquid manufac­turer’s safety precautions for clean up of minor spills.
7. Do not blow compressed air through the turbine.
8. Do not allow liquids to dry inside the turbine.
9. Handle the rotor carefully. Even small scratches or nicks can affect accuracy.
10. When tightening the turbine, do not use a wrench or pliers to tighten the turbine. Hand tighten only.
11. For best results, always verify ac­curacy before use.
Product Description
Omega Industrial Meter Turbines are
identied by the internal diameter of
the inlet and outlet.
FTB891 - 1/2 inch FTB893 - 1 inch Each turbine is designed to work with on-
board computer electronics and/ or with one of several accessory mod­ules that can interface to a wide vari­ety of reporting and collecting devices.
The CMOS microprocessor-based
electronics have extremely low power
requirements and data retention capabilities in both RAM and ROM. Information is clearly displayed on a large 6-digit LCD readout with two-
point oating decimal for totals from
.01 to 999,999. All operations are easily accessed with the two buttons on the front panel.
Liquid ows through the turbine hous­ing causing an internal rotor to spin. As the rotor spins, an electrical signal is generated in the pickup coil. The electrical signal provides the output necessary to operate the on-board computer electronics for local indication directly on the turbine or one of several accessory modules that transmit the
signal to external equipment.
Upon receipt, examine your meter
for visible damage. The turbine is a precision measuring instrument and should be handled as such. Remove the protective plugs and caps for a thorough inspection. If any items are damaged or missing, contact your distributor.
Make sure the turbine model meets
your specic needs. Refer to the Specications Section and conrm
the following:
1. The owrate is within the limits of
your model.
2. The liquid is compatible with the turbine’s wetted components.
3. The system’s pressure does not
exceed the turbine’s maximum
pressure rating.
Each Omega Turbine has a unique
identication number that includes
the Serial Number, Model Number, Manufacturing Date, and K-Factor.
This identication number is etched
into the surface of the turbine. Record
this identication number in the back
of the Owner’s Manual and keep for future reference.
2
Page 5
SN = Serial Number, a 7-8 dig-
it number that identifies this particular turbine.
MOD = Model number
KF = K-Factor given in pulses per
gallon (PPG).
MFD = Manufacturing Date indicating
the week and year of manufac­ture.
INSTALLATION
All Omega turbines are designed to
measure ow in only one direction.
The direction is indicated by the arrow cast-molded in the turbine outlet. If the opposite direction is desired, and you are using on-board computer electron­ics, rotate the computer electronics 180 degrees prior to installation.
Flow altering devices such as elbows, valves, and reducers can affect ac­curacy. The following recommended guidelines are given to enhance ac-
curacy and maximize performance.
Distances given here are minimum requirements; double them for desired straight pipe lengths.
Upstream from the turbine, allow a minimum straight pipe length at least 10 times the internal diameter of the
turbine. For example, with the FTB893
turbine, there should be 10 inches (25.4 cm) of straight pipe immediately up­stream. The desired upstream straight pipe length is 20 inches (50.8 cm).
Downstream from the turbine, allow a minimum straight pipe length at least 5 times the internal diameter of your
turbine. For example, with the FTB893
turbine, there should be 5 inches (12.7 cm) of straight pipe immediately downstream. The desired downstream distance is 10 inches (25.4 cm).
A typical back pressure of 5 to 50 PSI (0.34 to 3.4 bar) will prevent cavitation. Create back pressure by installing a control valve on the downstream side
of the meter at the proper distance detailed above.
Foreign material in the liquid being measured can clog the turbine’s rotor and adversely affect accuracy. If this
problem is anticipated or experienced, install screens to lter impurities from
incoming liquids.
FTB891
Maximum Particulate Size
Inches: 0.005 Microns: 125 Mesh: 55 Standard Sieve: 125 µm Alternative Sieve: No. 120
FTB893
Maximum Particulate Size
Inches: 0.018 Microns: 500 Mesh: 28 Standard Sieve: 500 µm Alternative Sieve: No. 35
The PVDF Series FTB890 turbines are Factory Mutual Approved and carry a Class 1, Division 1 Approval for hazardous environments. Omega Meters are tested and calibrated at the factory using state-of-the-art calibra­tion test equipment.
To ensure accurate measurement, remove all air from the system before use. To purge the system of air:
1. Ensure some back pressure on the turbine.
2. Open the discharge valve or nozzle
and allow uid to completely ll the
system. Make sure the stream is full and steady.
3. Close the discharge valve or nozzle.
4. Start normal operations.
Each turbine contains a removable back coverplate. Leave the coverplate installed unless accessory modules specify removal.
3
Page 6
Connections
1. To protect against leakage, seal all threads with an appropriate sealing compound. Make sure the sealing compound does not
intrude into the ow path.
2. Make sure the ow direction arrow
on housing back is pointed in the
direction of the ow.
3. Install union ring over pipe end
prior to installing pipe tting.
4. Install pipe ttings on pipe ends,
and tighten.
5. Tighten union ring to the turbine. Make sure O-ring is positioned in housing ends. Do not use a wrench or pliers. Hand tighten only.
NOTE: If connecting to new male
threads, burrs and curls can adversely affect accuracy. Cor­rect the problem prior to turbine installation.
It is strongly recommended that ac-
curacy be veried prior to use. To do
this, remove all air from the system,
measure an exact known volume into
an accurate container, and verify the volume against the readout or record­ing equipment. If necessary, use a
correction factor to gure nal volume.
For best results, accuracy should be
veried periodically as part of a routine
maintenance schedule. The procedure is found in the Maintenance Section.
The display electronics is normally installed on the turbine housing at the factory unless ordered without it.
If for any reason the display electronics need to be mounted on your turbine, simply mount the display on the turbine with the four screws at the corners of the faceplate. Make sure the seal is fully seated before tightening the screws.
If you ordered the FTB890 Series turbine with an accessory module, please review and thoroughly under-
stand all installation instructions before proceeding.
Avoid electronically “noisy” environ­ments. Install at least 6 inches (15.2 cm) away from motors, relays, or transformers.
OPERATION
Computer Display
All operations are reected in the
LCD readout. The large center digits indicate amounts, where smaller words or “icons” located above and below
indicate spcic information regarding totals, ow, calibration and units of
measure.
Activate the Meter
Computer is on continuously and al­ways ready to perform. The computer is powered by eld replaceable batter­ies. When display becomes dim, faded or the low battery message appears (see below), the batteries need to be replaced. Reference the Maintenance Section for details.
Batch and Cumulative Totals
The computer maintains two totals. The Cumulative Total provides con­tinuous measurement and cannot be manually reset. The Batch Total can be
reset to measure ow during a single
use. The Cumulative Total is labeled TOTAL 1, Batch Total is labeled TOTAL 2 BATCH.
When the Cumulative Total reaches a display reading of 999,999 the com­puter will highlight an X10 icon. This indicates to the operator that a zero must be added to the 6 digits shown.
When the next rollover occurs, the
computer will highlight an X100 icon. This indicates to the operator that two zeroes must be added to the 6 digits shown.
4
Page 7
Press the DISPLAY button briey to
switch between the TOTAL 1, TOTAL 2 BATCH and FLOWRATE. Press
DISPLAY briey to display the TOTAL 2
BATCH. Hold the DISPLAY button for 3 seconds to reset the Batch Total to zero.
When uid is owing through the meter,
a small propeller icon is highlighted.
The “eld” calibration may be set by the user, and can be changed or modied
at any time using the calibration proce­dure described below in the Calibration
Section. Totals or owrate derived from the eld calibration are invoked when
the FAC icon is no longer visible on the top line of the display.
Flowrate Feature
To use this feature, press and release DISPLAY until FLOWRATE icon ap­pears. The factory set time base will be highlighted to the right of FLOWRATE (M = minutes, H = hours, D = days). When FLOWRATE is invoked, the
display will be indicating rate of ow.
Factory and Field Calibration
All calibration information is visible to the user as icons on the top line of the display, above the numeric digits.
All units are congured with a “factory”
calibration. Both gallons and litres are available (“GL” or “LT” will be dis­played). While holding the CALIBRATE
button, briey press DISPLAY to toggle
between gallons and litres. This fac­tory calibration (indicated with FAC) is permanently programmed into the computer and is not user adjustable.
NOTE: Your computer may have other
units of measure programmed into it. If so, holding the CALIBRATE button and momentarily pressing the DISPLAY button will toggle through all factory set units. Other possible units are: IGL (imperial gallon), QT (quart), CF (cubic feet), CM (cubic meter), BL (42 gal. barrel), CC (cubic centimeter) or OZ (ounce).
Switching between different units will not corrupt the Total’s contents. For
example, in GL mode, the computer
totalizes 10.00 gallons, if the user switches to LT mode, the display will read 37.85 litres (the same volume, different unit).
CALIBRATION
Verify Accuracy Before Beginning Field Calibration
For the most accurate results, dis­pense at a owrate which best simu­lates your actual operating conditions.
Avoid “dribbling” more uid or repeat­edly starting and stopping the ow. This
can result in less accurate calibrations.
Make sure you meet the meter’s mini-
mum owrate requirements:
1/2 inch meter 0.6 GPM (2.2 LPM) 1 inch meter 2.5 GPM (9.5 LPM)
The use of a uniformly dependable, accurate calibration container is recommended for the most accurate results. For best results, the meter should be installed and purged of air
before eld calibration.
Field Calibration with Computer Display
Field Calibration and Factory Cali-
bration are dened in the Operation
Section. Factory calibration settings are programmed into each computer during manufacturing, using Stoddard test solvent at 70° F (21° C). Settings are correct for light liquids such as water, gasoline or diesel. Readings using the Factory Calibration (FAC) may not be accurate in some situations,
for example, “heavy” liquids such as motor oil under extreme temperature
conditions, non-standard plumbing
congurations or with uids other than
those mentioned above.
5
Page 8
For improved accuracy under such conditions, the computer allows for
“eld” calibration, that is, user entry
of custom calibration parameters. A “single point” calibration may yield acceptable accuracy when used in a non-standard application.
Field Calibration Procedures (Dispense/Display Method)
1. To eld calibrate, press and hold
the CALIBRATE and DISPLAY buttons for about 3 seconds until you see FLdCAL. Release both buttons and you will see dd000.0.
You are now in the eld calibration
mode.
2. Dispense a known amount of uid at a owrate representative of the
application. Any amount between .1 and 999.9 units can be used.
Display will count up while uid is owing through the meter.
3. The DISPLAY button can then be pushed to select the digit location and the CALIBRATE button can be pushed to scroll the desired value at the blinking position. Edit the amount shown with the value that was dispensed above. Values from
000.1 to 999.9 can be entered.
4. When satised with the value,
press both CALIBRATE and DISPLAY buttons simultaneously. CALEnd will be displayed and unit will go back to normal operation, less the FAC (factory calibration) icon.
5. The meter will now be operating with a custom calibration number unique to the dispense procedure. No unit of measure (gallon, litre, etc.) icon will be highlighted.
NOTE: To return to factory calibra-
tion (FAC), press and hold both CALIBRATION and DISPLAY buttons for about 3 seconds, until FAcCAL is displayed. Then release buttons. Unit should re-
turn to normal operation and FAC icon is visible.
NOTE: If the eld calibration mode is
entered and NO uid is dispensed,
then upon leaving, the computer will use data from the last success-
ful eld calibration.
MAINTENANCE
Verify Accuracy
Before use, check the turbine’s ac­curacy and verify calibration.
1. Make sure there is no air in the system.
2. Measure an exact known volume
into an accurate container.
3. Verify the volume against the readout or recording equipment.
NOTE: If necessary, use a correction
factor to gure nal volume.
For best results, accuracy should be
veried periodically as part of a routine
maintenance schedule.
Remove the Turbine
WARNING
During turbine removal, liquid may spill. Follow the liquid manufacturer’s safety precautions for clean up of minor spills.
1. Ensure all liquid is drained from the turbine. Wear protective clothing as necessary.
2. Loosen both union rings at the ends of the turbine.
3. If the turbine is not immediately installed again, cap lines as nec­essary.
Replace Internal Parts
1. Remove the turbine from the sys­tem. See Figure 5.
6
Page 9
Computer Electronics
Rotor Assembly
Figure 5
parts. Remove internal parts as detailed above. Note orientation carefully for correct assembly. Internal parts can be soaked for 10 to 15 minutes in compatible cleaning solutions. Use a soft brush or small probe to carefully remove residue from the rotor.
Turbine Housing
2. Use your ngers to gently re­move the rotor assembly from the groove. Do not use force to remove the rotor assembly.
CAUTION
Handle the rotor carefully. Even small scratches or nicks can affect accuracy.
3. Use the procedure below to clean the turbine.
4. Install the rotor assembly into the turbine housing. Make sure the pointed end of the rotor assembly is
inserted rst. (see Figure 5) Guide
the assembly into place using a smooth motion, little or no force is required.
5. Reinstall the turbine, purge the system of air, and verify accuracy before use.
Clean the Turbine
During use, the turbine should be kept full of liquid to ensure that drying does not occur inside the turbine. If drying or caking should occur, the rotor will stick or drag, affecting accuracy. To determine if the rotor is stuck or drag­ging, remove rotor from housing and physically turn rotor.
CAUTION
Never blow compressed air through the meter. It could damage the rotor.
1. Remove the turbine from the sys­tem following the directions above.
2. Carefully clean residue off all
WARNING
Follow the liquid manufacturer’s instructions for the disposal of contaminated cleaning solvents.
3. When the rotor turns freely, assemble and install it again following the instruc­tions above.
Display Electronics
The display electronics are powered by lithium batteries. If the meter’s readout should become dim, blank or the low battery message appears (see below), the batteries should be replaced. Replacement batteries can be ordered from the factory. See details in the Parts Section.
When batteries are disconnected or fail, the Batch and Cumulative Totals values will remain. Factory and Field Calibration Curves are retained in the meter’s computer when power is lost.
It is strongly recommended that battery check and terminal cleaning be a part of a routine maintenance schedule. Battery terminals should be cleaned annually. Batteries can be replaced without removing the meter from the piping system.
Replacing the Batteries
1. Remove the corner screws from the meter face and lift the display electronics from the turbine.
2. Remove the batteries.
3. Check the battery terminals and remove any corrosion.
7
Page 10
4. Install the new batteries and make sure the positive posts are posi­tioned correctly. When the batteries are installed correctly, the computer
5. Make sure the seal is fully seated before placing the computer elec­tronics on the turbine. Tighten the four screws.
powers on automatically and the readout displays information.
TROUBLESHOOTING
SYMPTOM PROBABLE CAUSE CORRECTIVE ACTION
METER IS NOT ACCURATE
READOUT FADED OR BLANK
NORMAL FLOWRATE BUT METER DOES NOT COUNT (Meter comes on when DISPLAY button pushed.)
REDUCED FLOWRATE & METER DOESN'T COUNT (Meter comes on when DISPLAY button pushed.)
1. Field Calibration not performed properly
2. Factory Calibration not suitable for liquid being measured
3. Meter operated below
minimum owrate
4. Meter partially clogged with dried liquid
5. Turbine bearings partially clogged with dried liquid
6. Sealant material wrapped around rotor
7. Installed too close to ttings Install correctly.
8. Installed too close to motors or electrically “noisy” environment
9. Improper connections to recording device
10. Accuracy needs verication Complete normal accuracy verication
1. Batteries weak, dead, or not connected
2. Display electronics defective Contact the factory.
1. Field Calibration not performed correctly
2. Rotor stuck or damaged Remove meter.
3. Sealant material wrapped around rotor
4. Display electronics defective Contact the factory.
1. Meter clogged with dried liquids
2. Below minimum owrate Increase ow.
Field calibrate again or select Factory Calibration.
Perform a Field Calibration according to Calibration Section.
Increase owrate.
Remove meter. Clean carefully. Make sure rotor spins freely.
Remove meter. Clean carefully. Make sure rotor spins freely.
Remove meter. Make sure rotor spins freely.
Install correctly.
Check all electrical connections. Reference appropriate installation instructions.
procedures. Repeat periodically.
Remove display electronics. Check and replace batteries if necessary.
Field Calibrate again or select Factory Calibration.
Make sure rotor spins freely
Remove meter. Make sure rotor spins freely
Remove meter. Clean carefully. Make sure rotor spins freely.
8
Page 11
SPECIFICATIONS
All data on FTB891 and FTB893 determined with 1 centipoise stoddard solvent test
uid at 70° F (21° C).
Models Size
Linear Flow Range
Gallons/minute (GPM) Liters/minute (LPM)
Extended Flow Range
Gallons/minute (GPM) Liters/minute (LPM)
Maximum Flow
Gallons/minute (GPM) Liters/minute (LPM)
Fluid Velocity in Extended Range
Feet/second Meters/second
Maximum Pressure Drop in 10:1 Range
PSIG bar
Frequency Range in Linear Flow Range 45-450 Hz 45-475 Hz
Weight*
Pounds Kilograms
Ship Weight*
Pounds Kilograms
FTB891
1/2 in.
1.2-12
4.54-45.4
0.6-12
2.2-45.4
15
56.8
0.5-10.6
0.2-3.2
10.0
0.68
.75 lbs. .340 kg
1.13 lbs. .535 kg
FTB893
1 in.
5-50
18.9-190
2.5-50
9.5-190
75
284
0.93-18.6
0.28-5.7
6.0
0.40
1.28 lbs. .580 kg
1.70 lbs. .770 kg
* Computer electronics add 0.2 lbs. (0.1 kg) to total weight.
9
Page 12
Performance FTB891 FTB893
Linear Range: 10:1 @ ±2% of reading 10:1 @ ±1.5% of reading
Extended Range: 20:1 @ ±5.0% of reading 20:1 @ ±5.0% of reading
Repeatability: ± 0.3%
Pressure Rating 100 PSIG (6.9 bar)
Wetted Components
Housing: PVDF Journal Bearings: Ceramic (98% Alumina) Shaft: Ceramic (98% Alumina) Rotor and Supports: PVDF O-Ring: FKM (Standard)
Temperature Range These temperatures are for the turbine without computer electronics. Final
operational temperature range is determined by computer electronics or acces­sory modules.
Operating Temperature
-20° to +180° F (-28° to 82° C)
Storage Temperature
-40° to 250° F (-40° to 121° C)
Sample Dimensions for FTB890 Series Model Shown
Length
Width
Height
Dimensions
Models Size
A = Height:
Inches Centimeters
B = Width:
Inches Centimeters
C = Length:
Inches Centimeters
Computer electronics add 0.7 in. (1.8 cm) to height of turbine.
10
FTB891
1/2 in.
3.2 in.
8.1 cm
2.1 in.
5.3 cm
7.3 in.
18.5 cm
FTB893
1 in.
3.3 in.
8.3 cm
2.8 in.
7.1 cm
8.1 in.
20.5 cm
Page 13
Display Specications
Input Pulse Rate:
Minimum Pulse In: DC Minimum Coil Input: 10 Hz
Maximum Raw: 1,000 Hz
K-Factor:
Minimum: .01 pulses/unit
Maximum: 999,999 pulses/
unit
Field Calibration:
Minimum Time: 10 seconds
Readout Totals:
Minimum Display: 0.01
Maximum Display: 999,999
Temperatures:
Operational: 0° to +140° F (-18° to +60° C)
Storage: -40° to +158° F (-40° to +70° C)
If wider operating temperature ranges are desired, reference information on Remote Kits.
Power:
Internal Power Supply: 2 Lithium Batteries at 3 volts each Battery Life: 5 years
Optional External Power: 7-30 VDC
Display Electronics Terminal Connections
J 1
J 10
J-1 Reset
When connected by a jumper wire to Ground (J1-6), this has the same effect as initial power up and zeroes out all totalizers.
J-2 Pulse Signal Output
This supplies a high-level ampli-
ed open collector signal. Output will withstand a maximum open-
circuit voltage of 60 volts DC
and a maximum closed-circuit
of 100 mA.
J-3 Not Used
J-4 Pulse Signal Input
Requires a sine or square wave with open-circuit voltage of 3-30
volts P-P, a maximum rise/fall
rate of 0.01 V/µ second and a
maximum frequency of 750 Hz.
J-5 Power Input
When used with Ground (J1-6), this has reverse polarity protec­tion, but no on-board voltage regulation. Supplied voltage must be 5.75 volts DC ±5%.
J-6 Ground
J-7, 8, 9, 10
Programming interfaces. Not accessable to user-
NOTE: Applicable safety approvals
are void if any external con­nections are made to computer electronics.
11
Page 14
MODEL NUMBERS
Model No.
w/Display
FTB891 FTB891-ND
FTB893 FTB893-ND
* Requires signal output module P/N - FLSC790-P-ND ordered separately.
† Extended low ow range and eld calibration for viscosity available on - ND units w/o display.
Model No.*
w/o Display
Range
GPM (LPM)
1.2-12
(4.54-45.4)
5-50
(18.9-190)
Extended
Range
Low Flow†
0.6-12
(2.2-45.4)
2.5-50
(9.5-190)
FNPT
Size
1/2"
1"
PARTS & ACCESSORIES
Order Replacement Kits with the part numbers given here.
Part Number Description
FTB890-ORING O-Ring (Computer)
FTB891-RK FTB891 (1/2 inch) Rotor Assembly Replacement Kit
FTB893-RK FTB893 (1 inch) Rotor Assembly Replacement Kit
FTB891-ORING O-Ring FKM Union Fitting (1/2 inch)
FTB893-ORING O-Ring FKM Union Fitting (1 inch)
Field Installable Options and Accessories
Model Number Description
FLSC790-MA 4-20 mAdc Output Module
FLSC790-P Pulse Output Module
(open collector output)
FLSC790-P-DC External Power Module for FLSC790-P (9-30Vdc)
FTB790-RK Remote Display Kit Module
FTB790-RK-FM FM Approved Remote Display Kit Module
FLSC790-BATT Replacement Batteries
FLSC790-P-ND Pulse Output for Models without Displays
(-ND sufx)
Except as noted, Options and Accessories are for Display Models only and only one Module may be installed per unit. All models are supplied with 10 ft. cable. Cable may be cut or extended as required. Customer supplied pull-up resistor required, 820 ohm (min).
(open collector output)
12
Page 15
WARRANTY/ DISCLAIMER
OMEGA ENGINEERING, INC. warrants this u nit to be f ree of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moistur e or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
with any nuclear installation or
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
RETURN REQUESTS/INQUIRIES
BEFORE RETURNING ANY PRODUCT(S) TO O MEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2009 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
RETURNS, please have
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the
COST of the repair,
2. Model and serial number of theproduct, and
3. Repair instructions and/or specific problems
relative to the product.
13
Page 16
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
TEMPERATURE

Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

Wire: Thermocouple, RTD & Thermistor

Calibrators & Ice Point References

Recorders, Controllers & Process Monitors

Infrared Pyrometers
PRESSURE, STRAIN AND FORCE

Transducers & Strain Gages

Load Cells & Pressure Gages

Displacement Transducers

Instrumentation & Accessories
FLOW/LEVEL

Rotameters, Gas Mass Flowmeters & Flow Computers

Air Velocity Indicators

Turbine/Paddlewheel Systems

Totalizers & Batch Controllers
pH/CONDUCTIVITY

pH Electrodes, Testers & Accessories

Benchtop/Laboratory Meters

Controllers, Calibrators, Simulators & Pumps

Industrial pH & Conductivity Equipment
SM
DATA ACQUISITION

Data Acquisition & Engineering Software

Communications-Based Acquisition Systems

Plug-in Cards for Apple, IBM & Compatibles

Datalogging Systems

Recorders, Printers & Plotters
HEATERS

Heating Cable

Cartridge & Strip Heaters

Immersion & Band Heaters

Flexible Heaters

Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL

Metering & Control Instrumentation

Refractometers

Pumps & Tubing

Air, Soil & Water Monitors

Industrial Water & Wastewater Treatment

pH, Conductivity & Dissolved Oxygen Instruments
M-4909/0413