Omega FTB790 User guide

Page 1
Page 2
Page 3
TABLE OF CONTENTS
General Information .................................... 1
Product Description .................................... 2
Installation ...................................................2
Connections .................................................3
Operations ................................................... 4
Calibration ...................................................5
Maintenance ................................................ 8
Replacing the Batteries ............................... 9
Troubleshooting ........................................ 10
Specications ............................................ 11
Model Numbers ........................................ 16
Parts & Accessories ................................... 18
Warranty/Disclaimer ................................. 19
Return Requests/Inquiries ......................... 19
GENERAL INFORMATION
This manual will assist you in installing and maintaining your FTB790 Series turbine meter and associated display electronics. The turbine meters can be purchased with or without the display electronics. (See Figure 1)
Figure 1
Turbine Housing
Display Electronics
When purchased with the display electronics, optional accessory modules are available for eld installation. Information on these acces­sories is contained in separate manuals. Cali­bration details using the display electronics are given in this manual.
When purchased without the display (Sufx ­ND) an optional eld installable Pulse Output
Module (FLSC790-P-ND) is available and described in a separate manual.
For best results, take the time to fully ac­quaint yourself with all information about all components of your FTB790 Series Turbine Meter System prior to installation and use. If you need assistance, contact the Omega Flow Application Department.
This symbol is used throughout the
!
manual to call your attention to safety
▲
messages.
Warnings alert you to the potential for per-
sonal injury.
Cautions call your attention to practices or
procedures which may damage your equipment.
Notes give information that can improve
efciency of operations.
It is your responsibility to make sure that all operators have access to adequate instruc­tions about safe operating and maintenance procedures.
Read Me!
For your safety, review the major warnings and cautions below before operating your equipment.
1. This equipment is approved to handle only
uids which are compatible with all wetted
materials.
2. When measuring ammable liquids, ob­ serve precautions against re or explosion.
3. When handling hazardous liquids, always follow the liquid manufacturer ’s safety precautions.
4. When working in hazardous environments, always exercise appropriate safety precau­tions.
5. Always dispose of used cleaning solvents in a safe manner according to the solvent manufacturer’s instructions.
6. During turbine removal, liquid may spill. Follow the liquid manufacturer’s safety precautions for clean up of minor spills.
7. Do not blow compressed air through the turbine.
8. Do not allow liquids to dry inside the turbine.
1
Page 4
9. Handle the rotor carefully. Even small scratches or nicks can affect accuracy.
10. When tightening the turbine, use a wrench
only on the wrench ats.
11. For best results, always verify accuracy before use.
PRODUCT DESCRIPTION
internal diameter of the inlet and outlet.
FTB791 – 1/2 inch (Mid Flow) FTB792 – 3/4 inch (Mid Flow) FTB793 – 1 inch (Mid Flow) FTB794 – 1-1/2 inch (High Flow) FTB795 – 2 inch (High Flow)
Each of these turbines are designed to work with on-board display electronics and/or with one of several accessory modules that can interface to a wide variety of reporting and collecting devices.
The CMOS microprocessor-based electronics
have extremely low power requirements and
data retention capabilities in both RAM and ROM. Information is clearly displayed on a large 6-digit LCD readout with two-point
oating decimal for totals from .01 to 999,999.
All operations are easily accessed with the two buttons on the front panel.
Liquids flow through the turbine housing causing an internal rotor to spin. As the rotor spins, an electrical signal is generated in the pickup coil. The electrical signal provides the output necessary to operate the on-board display electronics for local indication directly on the turbine or one of several accessory
modules that transmit the signal to external
equipment.
Upon receipt, examine your meter for visible
damage. The turbine is a precision measuring instrument and should be handled as such. Remove the protective plugs and caps for a thorough inspection. If any items are dam­aged or missing, contact OMEGA Customer Service.
Make sure the turbine model meets your spe-
cic needs. Refer to the Specications Section and conrm the following:
1. The owrate is within the limits of your
model.
2. The liquid is compatible with the turbine’s wetted components.
3. The system’s pressure does not exceed the turbine’s maximum pressure rating.
Information specic to your particular turbine,
including serial number, model number, manufacturing date, and K-factor is etched into the surface of the turbine. On Models FTB791, FTB792, and FTB793 find the etched information on the side of the turbine. On Models FTB794 and FTB795 the informa-
tion is etched on a wrench at.
SN = Serial Number, a 7-8 digit number that
identies this particular turbine.
MOD = Model number.
KF = K‑Factor given in pulses per gallon
(PPG).
MFD = Manufacturing Date indicating the
week and year of manufacture.
For your future reference, it might be useful to record this information in the manual in case it becomes unreadable on the turbine.
INSTALLATION
All Series FTB790 turbines are designed to measure ow in only one direction. The direc­tion is indicated by the arrow cast-molded in the turbine outlet. If the opposite direction is desired, and you are using on-board display electronics, rotate the display electronics 180 degrees prior to installation.
Flow altering devices such as elbows, valves, and reducers can affect accuracy. The fol­lowing recommended guidelines are given to
enhance accuracy and maximize performance.
Distances given here are minimum require­ments; double them for desired straight pipe lengths.
Upstream from the turbine, allow a minimum straight pipe length at least 10 times the in-
ternal diameter of the turbine. For example,
with the FTB793 turbine, there should be 10 inches (25.4cm) of straight pipe immediately upstream. The desired upstream straight pipe length is 20 inches (50.8cm).
2
Page 5
Downstream from the turbine, allow a mini­mum straight pipe length at least 5 times the internal diameter of your turbine. For ex­ample, with the FTB793 turbine, there should be 5 inches (12.7cm) of straight pipe imme­diately downstream. The desired downstream distance is 10 inches (25.4cm).
A typical back pressure of 5 to 50 PSI (0.34 to
3.4 bar) will prevent cavitation. Create back pressure by installing a control valve on the downstream side of the meter at the proper distance detailed above.
Foreign material in the liquid being measured can clog the turbine’s rotor and adversely af­fect accuracy. If this problem is anticipated or
experienced, install screens to lter impurities
from incoming liquids.
Models FTB791, FTB792, FTB793
Maximum Particulate Size Inches: 0.005 Microns: 125 Mesh: 55 Standard Sieve: 125 µm Alternative Sieve: No. 120
Models FTB794 and FTB795
Maximum Particulate Size Inches: 0.018 Microns: 500 Mesh: 28 Standard Sieve: 500 µm Alternative Sieve: No. 35
All Series FTB790 turbines are Factory Mu­tual Approved and carry a Class 1, Division 1 Approval for hazardous environments. They are tested and calibrated at the factory using state-of-the-art calibration procedures and test equipment.
To ensure accurate measurement, remove all air from the system before use. To purge the
system of air:
1. Ensure some back pressure on the turbine.
2. Open the discharge valve or nozzle and
allow uid to completely ll the system.
Make sure the stream is full and steady.
3. Close the discharge valve or nozzle.
4. Start normal operations.
Each turbine contains a removable back cov­erplate. Leave the coverplate installed unless accessory modules specify removal.
CONNECTIONS
1. To protect against leakage, seal all threads with an appropriate sealing compound. Make sure the sealing compound does
not intrude into the ow path.
2. Make sure the arrow on the outlet is
pointed in the direction of the ow.
3. Tighten the turbine onto the ttings. Use a wrench only on wrench ats.
NOTE: If connecting to new male threads,
burrs and curls can adversely effect accuracy. Correct the problem prior to turbine installation.
It is strongly recommended that accuracy be
veried prior to use. To do this, remove all air from the system, measure an exact known
volume into an accurate container, and verify the volume against the readout or recording equipment. If necessary, use a correction
factor to gure nal volume. For best results, accuracy should be veried periodically as
part of a routine maintenance schedule.
The display electronics is normally installed on the turbine housing at the factory unless ordered without it.
If for any reason the display electronics need to be mounted on your turbine, simply mount the display on the turbine with the four screws at the corners of the faceplate. Make sure the seal is fully seated before tightening the screws.
If you ordered the FTB790 Series turbine with an accessory module, please review and thor­oughly understand all installation instructions before proceeding.
Avoid electronically “noisy” environments. Install at least 6 inches (15.2cm) away from motors, relays, or transformers.
3
Page 6
OPERATIONS
All operations are reected in the LCD read­out. The top line identies the calibration curve. The middle line reects ow informa-
tion. The bottom line shows information from
information.
The display electronics is powered by eld
replaceable batteries. When the readout be­comes dim or faded, the batteries need to be replaced. Reference the Maintenance Section for details.
NOTE: Operations can be practiced prior to
installation. To simulate ow conditions,
blow gently through the turbine.
Turn On
The meter is on when any display is present.
It turns on automatically when liquid ows
through the meter. It can be turned on manu­ally by pressing and releasing the DISPLAY button.
Turn Off
The meter turns off automatically approxi­mately four minutes after ow stops. When
the meter is off, the readout is blank.
Figure 2
To change between totals, press and release DISPLAY.
NOTE: Generally, readout displays change
when buttons are released.
Factory and Field Calibration Curves
All calibration information is visible to the user as words in the upper part of the display, above the numeric digits.
All units will be congured with a “factory”
calibration curve, for which units of gallons or litres may be selected by the user (“GAL” or “LTR” will be visible). This curve is NOT
user adjustable: the word “PRESET” is
displayed to show this. (See Figure 3) The factory calibration is stored permanently in the computer’s memory.
Figure 3
Batch and Cumulative Totals
Total ags are displayed on the bottom line. A Batch Total indicates ow during a single use.
It is labeled with TOTAL 2. To zero a batch total, make sure it is displayed and hold down DISPLAY for three seconds until the display changes to zeros.
The Cumulative Total is the total of all liquid measured since the meter’s power was con-
nected. At your rst use, the Cumulative Total
is not zero because of calibration at the factory. The Cumulative Total is labeled as TOTAL LOCKED indicating it cannot be manually zeroed. (See Figure 2) The Cumulative Total is zeroed only when batteries are removed or go dead or when the Cumulative Total reaches
the maximum value of 999,999.
4
The “eld” calibration curve(s) may be set by the user, and can be changed or modied at any
time using the calibration procedure described below in the CALIBRATION section. Totals
or owrate derived from the eld calibration are visible when the eld calibration setting
is selected “CAL B”.
Page 7
Selecting a Different Calibration Setting
You can switch between GAL and LTR modes at will without “corrupting” totalizer contents.
For example, the computer can totalize 10.00
gallons. If the user switches to LTR mode, the display will immediately change to “37.85” (the same amount in units of litres). GAL/LTR switching also works in FLOWRATE mode.
To select a different calibration setting, rst
press and hold the CALIBRATE button.
Continue to hold it while also briey pressing
and releasing the DISPLAY button (you may then also release the CALIBRATE button).
display will change to show the newly selected calibration setting. Calibration settings change
in this order: GAL, LTR, CAL B, GAL (etc.). While uid is owing only the GAL and LTR
selections may be made, however, when NO
uid ow is occurring, any setting may be
selected.
Flowrate
When this feature is activated, the word FLOWRATE displays to the left on the bot-
tom line. (See Figure 4) When this ag is dis­played, the numbers on the middle line reect the rate of ow. To activate this feature, press
and release DISPLAY until FLOWRATE ap­pears to the left on the bottom line.
Figure 4
Propeller
A small propeller displays to indicate liquid
is owing through the meter.
CALIBRATION
Factory Calibration settings are programmed
into each owmeter during production, and are correct for light uids such as water, gasoline,
or diesel fuel. Factory Calibration is completed with stoddard test solvent solvent (on 1" and
smaller owmeters) or water (on 1-1/2" and larger owmeters) at 70°F (21°C). Readings
using the standard factory calibration curves may not be accurate in some situations – for
example, if the unit measures a “heavy” uid such as motor oil, especially under extreme
temperature conditions.
For improved accuracy under such conditions, the ow computer allows for “eld” calibra­tion, that is, user entry of custom calibration parameters. A “single point” calibration may yield acceptable accuracy with light liquids,
however, heavy liquids may require ve or
more calibration points to achieve a high level of accuracy. Up to 15 custom calibration points can be entered.
NOTE: A Field Calibration below the mini-
mum flo wrat e can adve rsely eff ect accuracy.
The use of a uniformly dependable, accurate calibration container is highly recommended for the most accurate results. Due to high flowrate, it is strongly recommended that Field Calibration of 1-1/2" and 2" meters be completed with a combination of volume and
weight using ne resolution scales.
For the most accurate results, dispense at a
owrate which best simulates your actual
operating conditions. Avoid “dribbling” more
uid or repeatedly starting and stopping the ow – these actions will result in less accurate
calibrations.
Make sure you meet the meter’s minimum
owrate requirements.
1/2 inch meters - 1 GPM (3.8 LPM) 3/4 inch meters - 2 GPM (7.6 LPM) 1 inch meters - 5 GPM (19 LPM) 1-1/2 inch meters - 10 GPM (38 LPM) 2 inch meters - 20 GPM (76 LPM)
For best results, the meter should be installed and purged of air prior to Field Calibration.
5
Page 8
Dispense-Display Field Calibration Procedures
Your Actions Notes
1. Hold down CALIBRATE while pressing and Remember that Field Calibration releasing DISPLAY until the Field Calibration curves are not preset. curve appears (“CAL B” or “CAL C” message will be displayed). Release both buttons.
2. To calibrate, press and hold the CALIBRATE This step puts the unit in dispense-
button. While continuing to hold CALIBRATE, display eld calibration mode
also press and hold the DISPLAY button. Hold (“dd-CAL”). both buttons for about 3 seconds until you see a blinking “dd-CAL” message. Once the “dd-CAL” message appears, release both
buttons. You are now in eld calibration mode.
3. Once the buttons have been released from The computer is waiting for you to
Step 2, the display will show the blinking make a decision to either exit from message “run 01.” eld calibration mode or to begin a dispense run. If you want to exit
the calibration now, go to Step 11.
4. If you want to continue with the calibration,
but have not dispensed any uid yet, make your nal preparations to your pumping
system, but don’t start pumping yet.
5. Start your pumping system so that uid ows When the computer displays a non-
through the meter. The display will stop blinking “run 01” message, it is
blinking and show the “run 01” message. sensing uid ow. For the most Dispense into a container that allows you to accurate results, dispense at a ow judge the amount of uid pumped. When you rate which best simulates your have pumped the desired amount (for example, actual operating conditions. Avoid 10 gallons), stop the uid ow quickly. “dribbling” more uid or repeat­ edly starting and stopping the ow -
these actions will result in less accurate calibrations.
6
Page 9
Your Actions Notes
6. Once the ow has stopped, briey press and When the display shows “0000.0”
release both buttons. At this point the computer the computer has stopped “watch-
display will change to “0000.0” with the left- ing” for uid ow and is now
hand digit blinking. waiting for you to enter some numbers.
7. Enter the volume (amount) of uid that you dispensed (for example, if your 10-gallon con-
tainer is full, enter “10.0” for gallons or “37.5” for litres). To enter numbers use the CALI­ BRATE button to change the value of the digit that is blinking and use the DISPLAY
button to shift the “blink” to the next digit.
8. Once the correct number has been entered, brief- You have installed the new cal­ ly press and release both buttons. The display curve point. You are ready to end will now change to a blinking “run 02” message. calibration (Step 10) or enter another new calibration point (Step 9).
9. To enter another calibration point, go back It is possible to set up to 15 cal-curve and repeat Steps 3 through 8. points, and the “run ##” message will increment each time you repeat the calibration process (run 01, run 02, run 03, etc., up to run 15).
10. To end calibration, press and hold both After you release the buttons, the buttons for about 3 seconds until you see the computer will resume normal “CAL End” message. operations with the new cal point(s) active.
7
Page 10
Dispense-Display Field Calibration Procedures - cont’d.
Your Actions Notes
11. If you HAVE NOT dispensed any uid, you After you release the buttons, the can exit calibration without changing the cal computer will resume normal
curve. If the message “run 01” is showing operation and the old curve (if you
and you have not dispensed any uid, hold have entered one in the past) is
both buttons for about 3 seconds until you still intact. see a “CAL End” message.
MAINTENANCE
Verify Accuracy
Before use, check the turbine’s accuracy and verify calibration.
1. Make sure there is no air in the system.
2. Measure an exact known volume into an
accurate container.
3. Verify the volume against the readout or recording equipment.
NOTE: If necessary, use a correction factor to
gure nal volume.
For best results, accuracy should be veried
periodically as part of a routine maintenance schedule.
Remove the Turbine
! ! ! WARNING ! ! !
!
▲
▲
During turbine removal, liquid may spill. Follow the liquid manufacturer’s safety precautions for clean up of minor spills.
1. Ensure all liquid is drained from the tur-
bine. Wear protective clothing as necessary.
2. Loosen both ends of the turbine. Use a
wrench only on the turbine’s wrench ats.
3. If the turbine is not immediately installed
again, cap lines as necessary.
Replace Internal Parts
1. Remove the turbine from the system as detailed above.
NOTE: Carefully notice the orientation of all
internal parts as they are removed, espe­cially the orientation of the rotor to the
ow direction arrow. (See Figure 5)
Figure 5
Support
2. Carefully clean residue off all parts. Re-
3. Using a small tool such as a screw driver
4. Carefully remove the rotor.
Rotor
Retaining Ring
move internal parts as detailed above. Note orientation carefully for correct assembly. Internal parts can be soaked for 10-15 minutes in compatible cleaning solutions. Use a soft brush or small probe to carefully remove residue from the rotor.
or awl, gently pry one retaining ring from its groove. Remove the support. If neces­sary, use needle nose pliers. Little or no force should be required.
Support
Retaining Ring
8
Page 11
CAUTION: Han dle the rotor carefully.
Even small scratches or nicks can affect accuracy.
5. Turn the turbine over and remove the other retaining ring. Remove the other support.
6. Clean, as detailed below, or discard as necessary.
7. Replace one support and retaining ring. Parts should drop easily into place with little or no force.
8. Install the rotor. Make sure the wide end of the rotor’s blades faces the ow direc­tion. (See Figure 6)
Figure 6
Wide End of Rotor Blade
Flow Direction Arrow
9. Turn the turbine over and drop the second
support into place. Put the nal retaining
ring into position.
10. Reinstall the turbine, purge the system of air, and verify accuracy before use.
Clean the Turbine
During use, the turbine should be kept full of liquid to ensure that drying does not occur inside the turbine. If drying or caking should occur, the rotor will stick or drag, affecting accuracy. To determine if the rotor is stuck or dragging, gently blow air through the meter and listen for the quiet whir of the rotor.
CAUTION: Never blow compressed air
through the meter. It could damage the rotor.
1. Remove the turbine from the system fol lowing the directions above.
-
2. When the rotor turns freely, assemble and install it again following the instructions above.
Display Electronics
The display electronics are powered by lithium batteries which provide at least 4,000 hours of actual use. If the meter’s readout should become dim or blank, the batteries should be replaced. Replacement batteries can be ordered from the factory. See details in the Parts Section.
When batteries are disconnected or fail, the Batch and Cumulative Totals return to zero. Factory and Field Calibration Curves are retained in the meter’s computer when power is lost.
It is strongly recommended that battery check and terminal cleaning be a part of a routine maintenance schedule. Battery terminals should be cleaned annually. Batteries can be replaced without removing the meter from the piping system.
Replacing the Batteries
1. Remove the corner screws from the meter face and lift the display electronics from the turbine.
2. Remove the batteries.
3. Check the battery terminals and remove any corrosion.
4. Install the new batteries and make sure the positive posts are positioned correctly. When the batteries are installed correctly, the computer powers on automatically and the readout displays information.
5. Make sure the seal is fully seated before placing the computer electronics on the turbine. Tighten the four screws.
! ! ! WARNING ! ! !
!
▲
▲
Follow the liquid manufacturer’s instruc‑ tions for the disposal of contaminated cleaning solvents.
9
Page 12
TROUBLESHOOTING
Symptom Probable Cause Corrective Action
Meter is not accurate 1. Field Calibration not perform- Field calibrate again or select Factory ed properly Calibration.
2. Factory Calibration not suit- Perform a Field Calibration according to able for liquid being measured Calibration Section.
3. Meter operated below mini- Increase owrate. mum owrate
4. Meter partially clogged with Remove meter. Clean carefully. Make sure dried liquid rotor spins freely.
5. Turbine bearings partially Remove meter. Clean carefully. Make sure clogged with dried liquid rotor spins freely.
6. Sealant material wrapped Remove meter. Make sure rotor spins freely. around rotor
7. Installed too close to ttings Install correctly.
8. Installed too close to motors Install correctly. or electrically “noisy” envi­ ronment
9. Improper connections to Check all electrical connections. Reference recording device appropriate installation instructions.
10. Accuracy needs verication Complete normal accuracy verication
procedures. Repeat periodically.
Readout faded or 1. Batteries weak, dead, or not Remove display electronics. Check and blank connected replace batteries if necessary.
2. Display electronics defective Contact the factory.
Normal owrate 1. Field Calibration not perform- Field Calibrate again or select Factory
but meter does not ed correctly Calibration. count (Meter comes
2. Rotor stuck or damaged Remove meter. Make sure rotor spins freely. on when DISPLAY
button pushed)
3. Sealant material wrapped Remove meter. Make sure rotor spins freely. around rotor
4. Display electronics defective Contact the factory.
Reduced owrate 1. Meter clogged with dried Remove meter. Clean carefully. Make sure
and meter does not liquids rotor spins freely. count (Meter comes
on when DISPLAY 2. Below minimum owrate Increase ow.
button pushed)
Cannot get meter 1. Factory Calibration Hold down CALIBRATE and push and re-
into eld calibration (PRESET) curve active lease DISPLAY until PRESET ag goes off.
Proceed with calibration according to the Calibration Section.
2. Display electronics circuit Replace display electronics. Contact the board defective factory.
3. Button defective Replace display electronics. Contact the factory.
Display electronics 1. Flowrate too low Try again and increase owrate to minimum
blinks “NO” after calibration rate. See Calibration Section.
eld calibration
2. Rotor not spinning freely Remove meter. Clean carefully. Make sure rotor spins freely.
10
Page 13
SPECIFICATIONS - STAINLESS STEEL
All data on Models FTB791, FTB792, and FTB793 determined with 1 centipoise stoddard
solvent test uid at 70°F (21°C). Data on Models FTB794 and FTB795 is determined with water at 70°F (21°C).
Models FTB791 FTB792 FTB793 FTB794 FTB795 Size 1/2 in. 3/4 in. 1 in. 1‑1/2 in. 2 in.
Linear Flow Range Gallons/minute (GPM) 1-10 2-20 5-50 10-100 20-200 Liters/minute (LPM) 3.8-37.9 7.6-75.7 18.9-190 38-380 76-760
Extended Flow Range
Gallons/minute (GPM) 0.5-10 1-20 2.5-50 5-100 10-200 Liters/minute (LPM) 1.9-37.9 3.8-75.7 9.5-190 19-380 38-760
Maximum Flow
Gallons/minute (GPM) 15 30 75 150 300 Liters/minute (LPM) 56.8 113.6 284 568 1,136
Fluid Velocity
in Extended Range
Feet/second 0.5-10.6 0.6-12.1 0.93-18.6 0.8-15.9 0.96-19.1 Meters/second 0.2-3.2 0.2-3.7 0.28-5.7 0.24-4.8 0.29-5.8
Maximum Pressure Drop in 10:1 Range
PSIG 8 7.5 5 4 4 bar 0.55 0.5 0.34 0.28 0.28
Frequency Range in Linear Flow Range 45-450 Hz 37-370 Hz 45-475 Hz 35-350 Hz 33-330 Hz
Connections NPT Female Inlet/Outlet Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.
Wrench Size:
Inch 1-1/16 in. 1-5/16 in. 1-5/8 in. 2-3/8 in. 3 in. Millimeter 27mm 33mm 41mm 60mm 75mm
Weight* Pounds 1.8 lbs. 2.0 lbs. 2.4 lbs. 4.0 lbs. 6.3 lbs. Kilograms 0.8 kg 1.0 kg 1.1 kg 1.8 kg 2.9 kg
* Display electronics add 0.2 lbs. (0.1 kg) to total weight.
11
Page 14
SPECIFICATIONS - STAINLESS STEEL continued
Performance
Linear Range for 1/2 in. and 3/4 in.: 10:1 @ ±2.0% of reading Linear Range for 1 in.: 10:1 @ ±1.5% of reading Linear Range for 1-1/2 in. and 2 in: 10:1 @ ±1.0% of reading Extended Range: 20:1 @ ±5.0% of reading Repeatability: ±0.1%
Pressure Rating 1,500 PSIG (103 bar)
Wetted Components
Housing: 316 Stainless Steel Journal Bearings: Ceramic (96% Alumina) Shaft: Tungsten Carbide Rotor and Supports: PVDF Retaining Rings: 316 Stainless Steel
Temperature Range
Turbine only without computer: -40° to +250°F (-40° to +121°C) Turbine with display electronics: +14° to +140°F (-10° to +60°C) (Display electronics or accessory modules determine nal operational temperature.)
A
B
Dimensions
Models FTB791 FTB792 FTB793 FTB794 FTB795 Size 1/2 in. 3/4 in. 1 in. 1‑1/2 in. 2 in.
A = Height:
Inches 1.8 in. 2.0 in. 2.2 in. 2.8 in. 3.2 in. Centimeters 4.6cm 5.1cm 5.6cm 7.1cm 8.2cm
B = Width
Inches 2.0 in. 2.0 in. 2.0 in. 2.7 in. 3.3 in. Centimeters 5.1cm 5.1cm 5.1cm 6.9cm 8.4cm
C = Length
Inches 4.2 in. 4.3 in. 4.5 in. 5.3 in. 6.3 in. Centimeters 10.7cm 10.9cm 11.4cm 13.5cm 16.0cm
Display electronics add 0.7 in. (1.8cm) to height of turbine.
12
C
Page 15
SPECIFICATIONS - BRASS
All data on FTB791-BRASS, FTB792-BRASS, and FTB793-BRASS determined with
1 centipoise stoddard solvent test uid at 70°F (21°C). Data on FTB794-BRASS and FTB795­BRASS is determined with water at 70°F (21°C).
Models FTB791‑ FTB792‑ FTB793‑ FTB794‑ FTB795‑ BRASS BRASS BRASS BRASS BRASS Size 1/2 in. 3/4 in. 1 in. 1‑1/2 in. 2 in.
Linear Flow Range Gallons/minute (GPM) 1-10 2-20 5-50 10-100 20-200 Liters/minute (LPM) 3.8-37.9 7.6-75.7 18.9-190 38-380 76-760
Extended Flow Range
Gallons/minute (GPM) 0.5-10 1-20 2.5-50 5-100 10-200 Liters/minute (LPM) 1.9-37.9 3.8-75.7 9.5-190 19-380 38-760
Maximum Flow
Gallons/minute (GPM) 15 30 75 150 300 Liters/minute (LPM) 56.8 113.6 284 568 1,136
Fluid Velocity
in Extended Range
Feet/second 0.5-10.6 0.6-12.1 0.93-18.6 0.8-15.9 0.96-19.1 Meters/second 0.2-3.2 0.2-3.7 0.28-5.7 0.24-4.8 0.29-5.8
Maximum Pressure Drop in 10:1 Range
PSIG 8 7.5 5 4 4 bar 0.55 0.5 0.34 0.28 0.28
Frequency Range in Linear Flow Range 45-450 Hz 37-370 Hz 45-475 Hz 35-350 Hz 33-330 Hz
Connections NPT Female Inlet/Outlet Size Yes Yes Yes Yes Yes Female Yes Yes Yes Yes Yes Inlet/Outlet Size 1/2 in. 3/4 in. 1 in. 1-1/2 in. 2 in.
Wrench Size:
Inch 1-1/16 in. 1-5/16 in. 1-5/8 in. 2-3/8 in. 3 in. Millimeter 27mm 33mm 41mm 60mm 75mm
Weight* Pounds 2.0 lbs. 2.3 lbs. 2.7 lbs. 6.0 lbs. 9.6 lbs. Kilograms 0.9 kg 1.0 kg 1.2 kg 2.7 kg 4.3 kg
* Computer electronics add 0.2 lbs. (0.1 kg) to total weight.
13
Page 16
SPECIFICATIONS - BRASS continued
Performance
Linear Range for 1/2 in. and 3/4 in.: 10:1 @ ±2.0% of reading Linear Range for 1 in.: 10:1 @ ±1.5% of reading Linear Range for 1-1/2 in. and 2 in: 10:1 @ ±1.0% of reading Extended Range: 20:1 @ ±5.0% of reading Repeatability: ±0.1%
Pressure Rating 300 PSIG (21 bar)
Wetted Components
Housing: Brass Journal Bearings: Ceramic (96% Alumina) Shaft: Tungsten Carbide Rotor and Supports: PVDF Retaining Rings: 316 Stainless Steel
Temperature Range
Turbine only without computer: -40° to +250°F (-40° to +121°C) Turbine with display electronics: +14° to +140°F (-10° to +60°C) (Display electronics or accessory modules determine nal operational temperature.)
A
B
Dimensions
Models FTB791‑ FTB792‑ FTB793‑ FTB794‑ FTB795‑ BRASS BRASS BRASS BRASS BRASS Size 1/2 in. 3/4 in. 1 in. 1‑1/2 in. 2 in.
A = Height:
Inches 1.8 in. 2.0 in. 2.2 in. 2.8 in. 3.2 in. Centimeters 4.6cm 5.1cm 5.6cm 7.1cm 8.2cm
B = Width
Inches 2.0 in. 2.0 in. 2.0 in. 2.7 in. 3.3 in. Centimeters 5.1cm 5.1cm 5.1cm 6.9cm 8.4cm
C = Length
Inches 4.2 in. 4.3 in. 4.5 in. 5.3 in. 6.3 in. Centimeters 10.7cm 10.9cm 11.4cm 13.5cm 16.0cm
Computer electronics add 0.7 in. (1.8cm) to height of turbine.
14
C
Page 17
SPECIFICATIONS continued
Display Specifications
Input Pulse Rate:
Minimum Pulse In: DC Minimum Coil Input: 10 Hz Maximum Raw: 1,000 Hz
K‑Factor:
Minimum: 0.1 pulses/unit Maximum: 999,999 pulses/
unit
Field Calibration:
Minimum Time: 10 seconds
Readout Totals:
Minimum Display: 0.01 Maximum Display: 999,999
Temperatures:
Operational: +14° to +140°F (-10° to +60°C) Storage: -40° to +158°F (-40° to +70°C)
If wider operating temperature ranges are desired, reference information on Remote Kits.
Power:
Internal Power Supply: 2 Lithium
Batteries at 3 volts each
Minimum Battery Life: 4,000 opera-
tional hours
Optional External Power: 7-30 VDC ±5%
Display Electronics Terminal
Connections
J 1
J 10
J‑1 Reset
When connected by a jumper wire
to Ground (J1-6), this has the same effect as initial power up and zeroes out all totalizers.
J‑2 Pulse Signal Output
This supplies a high-level amplied
open collector signal. Output will
withstand a maximum open-circuit voltage of 60 volts DC and a maxi-
mum closed-circuit of 100 mA.
J‑3 Not Used
J‑4 Pulse Signal Input
Requires a sine or square wave with
open-circuit voltage of 3-30 volts
P-P, a maximum rise/fall rate of
0.01 V/µ second and a maximum
frequency of 750 Hz.
J‑5 Power Input
When used with Ground (J1-6),
this has reverse polarity protection, but no on-board voltage regulation. Supplied voltage must be 5.75 volts
DC ±5%.
J‑6 Ground
J‑7, 8, Programming interfaces. Not access- 9, 10 able to user.
NOTE: Safety approvals are void if any
external connections are made to
computer electronics.
15
Page 18
SPECIFICATIONS continued
MODEL NUMBERS - STAINLESS STEEL
Extended Model No. Model No.* Range Range FNPT w/Display w/o Display GPM (LPM) Low Flow† Size
FTB791 FTB791-ND 1-10 0.5 1/2" (3.8-37.9) (1.9)
FTB792 FTB792-ND 2-20 1 3/4" (7.6-75.9) (3.8)
FTB793 FTB793-ND 5-50 2.5 1" (19-190) (9.5)
FTB794 FTB794-ND 10-100 5 1-1/2" (38-380) (9.0)
FTB795 FTB795-ND 20-200 10 2" (76-760) (38)
* Requires signal output module P/N - FLSC790-P-ND ordered separately.
† Extended low ow range and eld calibration for viscosity available on - ND units w/o display.
16
Page 19
MODEL NUMBERS - BRASS
Extended Model No. Model No.* Range Range FNPT w/Display w/o Display GPM (LPM) Low Flow† Size
FTB791- FTB791-ND- 1-10 0.5 1/2" BRASS BRASS (3.8-37.9) (1.9)
FTB792- FTB792-ND- 2-20 1 3/4" BRASS BRASS (7.6-75.9) (3.8)
FTB793- FTB793-ND- 5-50 2.5 1" BRASS BRASS (19-190) (9.5)
FTB794- FTB794-ND- 10-100 5 1-1/2" BRASS BRASS (38-380) (9.0)
FTB795- FTB795-ND- 20-200 10 2" BRASS BRASS (76-760) (38)
* Requires signal output module P/N - FLSC790-P-ND ordered separately.
† Extended low ow range and eld calibration for viscosity available on - ND units w/o display.
17
Page 20
PARTS & ACCESSORIES
Order Replacement Kits with the part numbers given here.
Part Number Description
FTB890-ORING O-Ring FTB791-RK (1/2 inch) Rotor/Support Replacement Kit FTB792-RK (3/4 inch) Rotor/Support Replacement Kit FTB793-RK (1 inch) Rotor/Support Replacement Kit FTB794-RK (1-1/2 inch) Rotor/Support Replacement Kit FTB795-RK (2 inch) Rotor/Support Replacement Kit FTB791-RETAINING (1/2 inch) Retaining Ring FTB792-RETAINING (3/4 inch) Retaining Ring FTB793-RETAINING (1 inch) Retaining Ring FTB794-RETAINING (1-1/2 inch) Retaining Ring FTB795-RETAINING (2 inch) Retaining Ring
Field Installable Options and Accessories
Model Number Description
FLSC790-MA 4-20 mAdc Output Module FLSC790-P Pulse Output Module
FLSC790-P-DC External Power Module for FLSC790-P (9-30Vdc)
FTB790-RK Remote Display Kit Module FTB790-RK-FM FM Approved Remote Display Kit Module FLSC790-BATT Replacement Batteries
FLSC790-P-ND Pulse Output for Models without Displays (-ND sufx)
Except as noted, Options and Accessories are for Display Models only and only one Module may be installed per unit. All models are supplied with 10 ft. cable. Cable may be cut or extended as required. Customer supplied pull-up resistor required, 820 ohm (min).
(open collector output)
(open collector output)
18
Page 21
Page 22