Omega FTB-400A, FTB400 User guide

Page 1
Page 2
Servicing North America:
U.S.A. Omega Engineering, Inc. Headquarters: Toll-Free: 1-800-826-6342 (USA & Canada only)
Customer Service: 1-800-622-2378 (USA & Canada only) Engineering Service: 1-800-872-9436 (USA & Canada only) Tel: (203) 359-1660 Fax: (203) 359-7700 e-mail: [email protected]
For Other Locations Visit omega.com/worldwide
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
Page 3
CONTENTS
1. Introduction.................................................................................... 1
1.1 Specification .......................................................................... 1
1.2 Model Number....................................................................... 2
2. Operation ....................................................................................... 2
2.1 Principle.................................................................................3
2.2 Precautions............................................................................. 4
3. Installation ..................................................................................... 5
3.1 General Piping ....................................................................... 5
3.2 Strainers/Filters......................................................................7
3.3 Flow Straighteners and Installation Kits................................7
4. Maintenance................................................................................... 8
4.1 General................................................................................... 8
4.2 Cleaning Procedures .............................................................. 8
4.2.1 Chemical Cleaning......................................................... 8
4.2.2 Steam Cleaning .............................................................. 8
4.3 Pickup Coil Testing .............................................................10
4.4 Trouble Shooting ................................................................. 11
Appendix A: Drawing
Typical Assembly Sanitary Flowmeter with Installation Kit........... 12
Page 4
1. Introduction
The following information is provided for the proper installation and maintenance of your instrument.
1.1 Specifications
Linearity: 0.5% of reading over linear flow range
Repeatability: 0.1% of reading
Flow Range: 0.35 to 650 gpm
Temperature Range:
Signal Output:
Materials of Construction:
Bearings:
Surface Finish:
End Fitting:
-450 °F to +450 °F w/ Standard MAG coil
10 mVrms or greater into 10K ohm load at minimum flow rate.
316/316L Dual Rated Stainless Steel Rotor 17.4 PH SS Retaining Ring 15.7 MO PH SS
Hard Carbon Composite
32 micro inch finish
Tri-Grip® Sanitary Type
-1-
Page 5
1.2 Model Number
Linear Flow
Range
Model No. Fittings
FTB-401A 11/2″ TRI
FTB-402A 11/2″ TRI
FTB-403A 11/2″ TRI
FTB-404A 11/2″ TRI
FTB-405A 11/2″ TRI
FTB-406A 11/2″ TRI
FTB-407A 11/2″ TRI
FTB-408A 11/2″ TRI
FTB-409A 2″ TRI
FTB-410A 3″ TRI
FTB-411A 4″ TRI
Water
LPM GPM
1.32 to
0.35 to
13.2
3.5
2.84 to
0.75 to
28.4
7.5
4.73 to
1.25 to
36
9.5
6.62 to
1.75 to
61
16
9.5 to
2.5 to
110
29
15 to
4 to 60 5.1
227
23 to
6 to 93 4.3
352
30 to
8 to 130 3.0
492
57 to
15 to
852
225
95 to
25 to
1514
400
151 to
40 to
2460
650
Note: Operating pressure is limited by the TRI fitting.
Max
Pressure
Drop (psid)
3.0
5.0
5.2
3.0
5.0
3.3
4.0
4.0
Length
mm (in)
90.4
(3.56)
90.4
(3.56)
90.4
(3.56)
90.4
(3.56)
82.5
(3.25)
90.4
(3.56)
116.5 (4.59)
116.5 (4.59)
154
(6.06)
254 (10)
254 (10)
Nominal K-factor
Pulses/Gal
15,000 3 lbs
8,900 3 lbs
5,800 3 lbs
5,200 3 lbs
2,200 3 lbs
840 3 lbs
400 4 lbs
230 4 lbs
120 7 lbs
97 8 lbs
45 12 lbs
Weight
kg
(lb)
-2-
Page 6
2. Operation
2.1 Principle
The turbine flow sensor consists of a rotor assembly which is supported on a shaft held in place by triple tube clusters and secured with locking nuts within the flowmeter housing.
The rotor is free to spin on a self lubricated ceramic ball bearing. A magnetic type pickup coil is attached on the exterior of the flowmeter housing.
A low mass rotor design allows for rapid dynamic response. The deflector cones eliminate downstream thrust on the rotor and allows for dynamic positioning of the rotor between deflector cones.
The dynamic positioning of the low mass rotor provides wider rangeability and longer bearing life than that of conventional turbine flowmeters. Integral flow straightening tubes minimize the effects of upstream flow turbulence.
As the liquid flows through the flowmeter the rotor spins at rate proportional to the volumetric liquid flowrate.
Each rotor blade passing through the pickup coil generates an electrical pulse. The frequency of the pulses is proportional to flowrate. The summation of pulses represents total amount of liquid volume passed through the meter.
The number of pulses generated per unit of volume is called the calibration factor or K-Factor. This calibration factor is used to calculate flowrate and total amount of flow.
Material Selection and Construction
The housing is made of 316 stainless steel. The rotor is made of 17.4 pH stainless steel. Bearings are composite hard carbon, FDA approved.
-3-
Page 7
Flowmeter Calibrations
The standard calibration provided with an Omega turbine flowmeter consists of a 10-point water calibration that is traceable to NIST. Based on this water calibration, we derive an average k-factor for water for the flowmeter.
The uncertainty of this calibration is typically 0.1%.
The K-factor on turbine flowmeters used on liquid service is NOT density dependent.
2.2 Precautions
Do not drop the meter. Dropping the meter may result in damage
to the meter housing and/or internals.
Do not operate the meter at flowrates greater than the maximum
flowrate marked on the meter. Operating at flowrates greater than the maximum flowrate may over-spin the meter. Over-spinning may result in damage to the meter.
CAUTION: Avoid over-spinning the meter. Over-spinning the meter may
result in damage to the meter internals and lead to meter failure.
-4-
Page 8
3. Installation
CAUTION: Turbine meter has to be installed with pickup coil pointing
down (see Appendix A) to ensure proper cleanability.
Inspect all packages for any indications of damage which may have occurred during shipment.
Verify that all meter parts or auxiliary components have arrived with the shipment. Refer to the packing list/invoice for a detailed list of items included in the shipment.
3.1 General Piping
It is required to install meter with a minimum straight run of pipe approximately 10 pipe diameters ahead of the inlet and 5 pipe diameters following the outlet to avoid any effect of fluid swirls.
The meter housing is marked by a flow direction arrow and the inlet is marked ‘IN’ and the outlet is marked ‘OUT’. The meter must be installed in the piping in the correct orientation to ensure the most accurate operation.
Install meter with adequate distance and isolation from electric motors, transformers, welding equipment and solenoids to avoid any electromagnetic interference from ambient electrical field.
When it is expected that flow will be intermittent, the meter should not be mounted at a low point in the piping system. Solids which settle or congeal in the meter may affect meter performance.
-5-
Page 9
A typical flowmeter installation is shown below:
BYPASS RUN
Bypass Valve
Flow Straighener
Typical Flowmeter Installation
METER RUN
Turbine FlowmeterStrainer
Blocking ValveBlocking Valve
Blocking and Bypass valves should be installed if it is necessary to do preventive maintenance on the flowmeter without shutting down the flow system. The Bypass valve can be opened before the Blocking valves are closed allowing the flow to continue while removing the turbine flowmeter for service.
IMPORTANT
: All flow lines should be purged prior to installing the
meter. To prevent possible damage to the meter, install the meter ONLY in flow lines that are clean and free of debris.
Upon initial start-up of the system a spool piece should be installed in place of the flowmeter so that purging of the system can be performed to remove all particle debris which could cause damage to the meter internals. In applications where meter flushing is required after meter service, care should be taken as to not over-spin the meter, as severe meter damage may occur.
CAUTION: Avoid over-spinning the meter. Over-spinning the meter may
result in damage to the meter internals and lead to meter failure.
-6-
Page 10
To maintain an accurate flow measurement it is necessary to maintain a downstream pressure sufficient to prevent flashing/cavitation. Flashing of the liquid will result in an indication of flow significantly higher than the actual flow. In order to eliminate this condition adequate downstream pressure must be maintained. The minimum required downstream pressure may be calculated using the following equation:

Downstream pressure may be maintained by a downstream valve that provides the necessary downstream pressure to prevent flashing/cavitation in the metering run.
essurexVaporessureDropxessureMinimum Pr25.1Pr2Pr
3.2 Strainers/Filters
Turbine flowmeters are designed for use in a clean fluid service. However, the service fluid may carry some particulate material which would need to be removed before reaching the flowmeter. Under these conditions a strainer/filter may be required to reduce the potential hazard of fouling or damage that may be caused by foreign matter.
METER SIZE MESH SIZE PARTICLE SIZE
¼” to ½” 100 .0055
5
/8” to 1¼” 70 .008
1½” to 3” 40 .015
If a strainer/filter is required in the system, it should be located upstream of the flowmeter taking care that the proper minimum distance is kept between the strainer and flowmeter.
(Maximum)
3.3 Flow Straighteners and Installation Kits
Proper application of the Omega FTB-400A Series Flowmeter requires a minimum inlet straight pipe run of 10 pipe diameters and a minimum outlet straight pipe run of 5 pipe diameters.
Installation kits for the Omega FTB-400A Series Flowmeter consist of two lengths of appropriate tubing cut to a length appropriate for the upstream and downstream straight pipe run with appropriate end fittings. Flow straightening sections may be provided within the installation kit.
-7-
Page 11
4. Maintenance
4.1 Cleaning Procedures
The Omega FTB-400A Series flowmeters have been designed to allow for cleaning by commercially accepted practices. These include removing the flowmeter from the line for cleaning in an approved fluid, flushing the line with an approved cleaning solution, and steam cleaning. With all cleaning methods, care must be taken to not over­spin the meter, as severe meter damage may occur.
CAUTION: Avoid over-spinning the meter. Over-spinning the meter may
result in damage to the meter internals and lead to meter failure.
4.2.1 Chemical Cleaning
The flowmeters may be chemically cleaned using an approved cleaning solution by removing the meter from the service line and using a bath or by flushing the meter in place.
The hard carbon composite bearing designs used have been tested and found to be compatible with the following CIP fluids manufactured by Klenzade; Mandate, AC-300, AC-101, Principle, and XY-12.
Following the cleaning operation, the cleaning solution should be flushed from the meter and/or service line with potable water to remove the chemically active cleaning solution.
Care should be taken to ensure that flowrates occurring during chemical cleaning do not exceed the flow capacity of the flowmeter.
4.2.2 Steam Cleaning
Steam cleaning is only recommended for meters with hard carbon composite bearings.
The steam flow velocity during the cleaning should not exceed 1/3 of the maximum liquid flow capacity of the flowmeter.
-8-
Page 12
Steam Cleaning Rates at Various Steam Pressures
Meter
Size
1/4 1.25 1.70 2.25 2.50 1.72 1.05
3/8 2.70 3.67 4.75 5.39 3.68 2.25
1/2 3.50 4.73 6.14 7.00 4.74 2.90
5/8 5.78 7.82 10.20 11.50 5.02 4.80
3/4 10.50 14.20 18.40 20.90 6.32 8.70
1 21.70 29.40 38.10 43.10 7.35 18.00
1 ¼ 33.70 45.70 59.30 67.10 7.32 28.00
1 ½ 47.00 63.60 82.50 93.50 7.08 39.00
2 81.30 110.10 142.80 162.00 6.89 67.50
2 ½ 144.60 196.00 254.00 288.00 7.84 120.00
3 235.00 318.00 412.60 467.00 8.85 195.00
NOTES
1. PPH = Pounds Per Hour
2. The velocity is expressed for a line size equal to the inlet bore of the flowmeter.
3. The apparent GPM is provided since many applications have a flow rate indicator
50 psig
PPH1
which can be used to set a safe flow rate during the steam cleaning cycle.
75 psig
PPH1
100 psig
PPH1
125 psig
PPH1
Velocity
FPS2
Rate
GPM3
-9-
Page 13
4.3 Pickup Coil Testing
Testing the MAG and MCP (RF) coils consists of measuring the resistance with an ohmmeter. Resistance measurements are to be made only when there is no flow through the meter.
1. Measure the resistance between pin A and pin B. The resistance
should be approximately as listed in the following table common coils.
2. The resistance from any pin to the case should be greater than 1
Mohm.
COIL
DC RESISTANCE
(Ohms)
MC2PAHT
MCP3A
PC13-74G
PC13-74S
PC24-45G
PC24-45S PC28-13G PC28-14G
of some
15.0 10%
11.5 10% 1800 10% 1850 15% 1350 10% 1850 15%
120 20% 180 20%
If either resistance measurement fails, replace the pickup coil. Firmly seat the new coil in the flowmeter and tighten the locking nut.
For specific coils not listed contact the HFC Customer Service Department for the
approximate resistance readings.
-10-
Page 14
4.4 Trouble Shooting
Refer to the following troubleshooting guide for assistance with possible meter malfunctions:
TROUBLE CAUSE REMEDY
Fluid will not flow Meter clogged. Clean meter. through the meter Line to meter
blocked. Reduced flow through the meter
Line to meter
Meter readings inaccurate
Meter drag due to
Meter partially
clogged.
partially blocked.
Fluid flowrate is
not within meter
flow range.
improper
installation
Clean line to meter.
Clean meter.
Clean line to meter.
See “Specifications” for min and max flowrates.
Replace internals.
-11-
Page 15
DRAWING
Typical Assembly
Sanitary Flowmeter with Installation Kit
-12-
Page 16
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
RETURN REQUESTS/INQUIRIES
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2016 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the
COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.
Page 17
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
TEMPERATURE
MU
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
MU
Wire: Thermocouple, RTD & Thermistor
MU
Calibrators & Ice Point References
MU
Recorders, Controllers & Process Monitors
MU
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
MU
Transducers & Strain Gages
MU
Load Cells & Pressure Gages
MU
Displacement Transducers
MU
Instrumentation & Accessories
FLOW/LEVEL
MU
Rotameters, Gas Mass Flowmeters & Flow Computers
MU
Air Velocity Indicators
MU
Turbine/Paddlewheel Systems
MU
Totalizers & Batch Controllers
pH/CONDUCTIVITY
MU
pH Electrodes, Testers & Accessories
MU
Benchtop/Laboratory Meters
MU
Controllers, Calibrators, Simulators & Pumps
MU
Industrial pH & Conductivity Equipment
SM
DATA ACQUISITION
MU Communications-Based Acquisition Systems MU Data Logging Systems MU Wireless Sensors, Transmitters, & Receivers MU Signal Conditioners MU Data Acquisition Software
HEATERS
MU
Heating Cable
MU
Cartridge & Strip Heaters
MU
Immersion & Band Heaters
MU
Flexible Heaters
MU
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
MU
Metering & Control Instrumentation
MU
Refractometers
MU
Pumps & Tubing
MU
Air, Soil & Water Monitors
MU
Industrial Water & Wastewater Treatment
MU
pH, Conductivity & Dissolved Oxygen Instruments
M2124/0618
Loading...