Omega FPDM1000 User Manual

FPDM1000 SERIES
Positive Displacement
Mechanical Flowmeters
To the owner:
Thank you for purchasing an OMEGA FPDM Series Flowmeter. Please take a few minutes to read through the manual before installing and operating your meter. If you have any problems with the meter, refer to the Maintenance and Troubleshooting sec­tions of the manual.
This manual contains connection and op­erating instructions for the OMEGA FPDM Series meters with mechanical displays. This includes the following models:
FPDM1004 FPDM1006 FPDM1005 FPDM1007
The OMEGA FPDM Series Flowmeter has incorporated the oval rotor principal into its design. This has proven to be a reliable and highly accurate method of measuring flow. Exceptional repeatability and high ac­curacy over a wide range of fluid viscosi­ties and flowrates are features of the OMEGA FPDM Series Flowmeter design. The low pressure drop and high pressure rating means the OMEGA FPDM Series Flowmeter is suitable for both gravity and pump (in-line) applications.
The OMEGA FPDM Series Flowmeters are available in either Aluminum or 316 Stain­less Steel. Standard rotors are made from PPS.
The OMEGA FPDM Series mechanical dis­plays have a resettable batch totalizer and non-resettable accumulative totalizer.
IMPORTANT INFORMATION
Please read this information
!
carefully before use!
Before use, confirm the fluid to be used is compatible with the meter or consult with OMEGA for advice.
NOTE: When a strainer is installed it should
be regularly inspected and cleaned. Failure to keep the strainer clean will dramatically effect flowmeter perfor­mance.
To prevent damage to the meter, slowly fill the system with fluid. This will prevent dam­age caused by air purge.
NOTE: Failure to do this could damage the
meter.
For pump applications, turn off the pump at the end of each day.
1
INSTALLATION
1. OMEGA recommends that when set­ting up pipework for meter installations, a bypass line be included in the de­sign. This provides the facility for a meter to be removed for maintenance without interrupting production. (See Figure 1)
Figure 1
Bypass Line
Flow Inlet
Figure 2
R I G H T
R I G H T
Strainer
Flow Outlet
2. Use thread sealant on all pipe threads.
3. For pump applications, ensure pipe work has the appropriate working pres­sure rating to match the pressure out­put of the pump.
4. Install a wire mesh strainer (Y or bas­ket type 60 mesh) as close as possible to the inlet side of the meter.
5. Ensure that the meter is installed so that the flow of the liquid is in the di­rection of the arrows embossed on the meter body.
6. The meter can be installed in any ori­entation as long as the meter shafts are in a horizontal plane. (See Figure 2) The register assembly may be oriented to suit the individual installation.
W R O N G
NOTE: Incorrect installation can cause pre-
mature wear of meter components
7. Do not overtighten meter connections.
8. It is important that after initial installa­tion you fill the line slowly, high speed air purge could cause damage to the rotors.
9. Test the system for leaks.
10. Check the strainer for swarf or foreign material. After the first 200 litres, check periodically – particularly if the flowrate decreases.
2
OPERATION SERVICE INSTRUCTIONS
When fluid passes through the meter, the rotors turn. The gear located on top of one of the rotors drives the mechanical regis­ters gear train which provides an accurate readout.
Disassembly:
Ensure that the fluid supply to the meter is disconnected, and the line pressure is re­leased before disassembly. Refer to the exploded parts diagram on page 6.
1. Remove the four screws located on the face of the register. Then remove the faceplate cover including register as­sembly.
2. Remove the four register mounting screws. Then remove the lower half of the register housing.
3. Remove the six coverplate screws and remove the coverplate.
4. Remove the four meter cap screws and remove the meter cap.
5. Remove rotors.
Reassembly:
1. Clean all components before reassem­bly.
2. Before reassembly, check the condition of the rotors. Replace if necessary.
3. Replace the rotor (with the gear) on the short shaft in the housing. Then place the second rotor onto the shaft so as the rotors are at 90 degrees to each other. (See Figure 3)
Figure 3
Rotor #1
Rotor #2
Rotors must be at 90° to each other
3
Check rotor operation by turning either of the rotors. If the rotors are not in mesh correctly or do not move freely, remove one of the rotors and replace it correctly at 90 degrees to the other rotor. Recheck the operation of the rotors.
4. Inspect the gears in the meter cap for wear. Replace if required - refer to spare parts on page 6.
5. Replace the O-ring into the groove in the meter cap. If the O-ring has been distorted or is damaged in any way, re­place it with a new part.
6. Replace the meter cap. Make sure the meter cap and the gear on the rotor meshes correctly with the gear in the meter cap. Insert the allen screws and tighten in the sequence 1, 4, 2 and 3.
7. Inspect the bevel gear, O-ring and out­put gear for wear or damage. Replace faulty components if necessary.
8. Replace the output shaft, bush and seal. Disassembly of output shaft:
a. Remove the bevel gear. b. Remove the circlip and push out the
output shaft assembly, including the
washer. c. Remove the seal. d. Carefully press out the output shaft
bush (if required). Assembly of output shaft:
a. Carefully press the new output shaft
bush into place. (Use Loctite Primer
7471, as per instructions, followed
by sealant Loctite 262.) b. Insert a new seal into the groove of
the output shaft bush. c. Replace the output gear and washer
and replace the circlip to lock the
output gear shaft into place. d. Replace the bevel gear and tighten
the grub screw onto the flat face of
the shaft.
9. Place the O-ring into the groove in the meter cap. Replace the O-ring seal if required.
10. Place the coverplate onto the meter. Replace the coverplate screws and tighten the six cap head screws firmly.
11. Place the lower coverplate of the regis­ter into position. Replace the four screws and tighten.
12. Position the register correctly on top of the lower register cover. Replace the four screws and tighten.
13. Test the meter by turning the rotors with a finger or by applying low air pressure (no more than a good breath) to one end of the meter, before returning meter to the line.
4
Loading...
+ 14 hidden pages