Omega FP6500 User guide

Page 1
Shop online at
omega.com
For latest product manuals:
omegamanual.info
FP6501/6502
FP6511/6512
FP6500 SERIES
Insertion Pa d d l ewheel Flow Sensor
Page 2
OMEGAnet®Online Service Internet e-mail
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
omega.com i n f o @ o m e g a . c o m
Servicing North America:
U.S.A.: One Omega Drive, P.O. Box 4047
ISO 9001 Certified
Stamford, CT 06907-0047 TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: [email protected]
Canada: 976 Bergar
Laval (Quebec) H7L 5A1, Canada TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: [email protected]
For immediate technical or application assistance:
U.S.A. and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
Customer Service: 1-800-622-2378 / 1-800-622-BEST Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico: En Espan˜ ol: (001) 203-359-7803 e-mail: [email protected]
FAX: (001) 203-359-7807 [email protected]
®
®
®
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
TEL: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: [email protected]
Czech Republic: Frystatska 184, 733 01 Karviná, Czech Republic
TEL: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: [email protected]
France: 11, rue Jacques Cartier, 78280 Guyancourt, France
TEL: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427 Toll Free in France: 0800 466 342 e-mail: [email protected]
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: [email protected]
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified
Northbank, Irlam, Manchester M44 5BD United Kingdom TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: [email protected]
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
Page 3
TABLE OF CONTENTS
General Information
General Information, Features, Specications .................................................................................................Page 1
Installation
Piping, Positioning the Meter, Immersion, Depth Setting ................................................................................ Page 2
Installation .......................................................................................................................................................... Page 3
Dimension C, Pipe Wall Thickness .................................................................................................................... Page 4
Flow Rate, K-Factors .......................................................................................................................................... Page 5
Straight Pipe Recommendations ......................................................................................................................
Full Pipe Recommendations ..............................................................................................................................Page 7
Connection, Operation & Repair
Calibration, Theory of Operation, Flow Range, Rotor Replacement ...............................................................Page 8
Repair & Parts
Signal, Parts Explosion, Parts List .................................................................................................................... Page 9
Troubleshooting
Problem, Probable Cause, To Try....................................................................................................................... Page 10
TABLES AND DIAGRAMS
Page 6
Features .............................................................................................................................................................. Page 1
Specications Table ...........................................................................................................................................Page 1
Piping .................................................................................................................................................................. Page 2
Positioning the Meter ......................................................................................................................................... Page 2
Depth Setting ..................................................................................................................................................... Page 2
Meter Installation ............................................................................................................................................... Page 3
Table 1 (Dimension 'C') ...................................................................................................................................... Page 4
Table 2 (Pipe Wall Thinkness) ........................................................................................................................... Page 4
Table 3 (Flow Rate) ............................................................................................................................................ Page 5
Table 4 (K Factors) ............................................................................................................................................. Page 5
Straight Pipe Recommendations ......................................................................................................................
Full Pipe Recommendations ..............................................................................................................................Page 7
Connection Diagram .......................................................................................................................................... Page 8
Page 6
Rotor Replacement ............................................................................................................................................Page 8
Parts Explosion ................................................................................................................................................... Page 9
Parts List ............................................................................................................................................................. Page 9
Troubleshooting: Problem, Probable Cause, To Try .......................................................................................... Page 10
Page 4
GENERAL INFORMATION
GENERAL INFORMATION
The FP6500 Series are adjustable depth insertion paddlewheels that come in brass or 316 stainless models to t 3” to 40” pipe. Installation ttings are standard 1-1/2" or 2” NPT. Fittings such
as saddles and weldolets may be purchased either locally or from Omega.
Ruby bearings and a non-drag Hall-effect sensor give these meters the widest ow range of any of the paddlewheel types. A sensor de­tects the passage of miniature magnets in the six rotor blades.
feet over unshielded cable without a transmitter and connected directly to many PLC’s and other controls without any additional electronics. If desired, a modular system of electronics can be installed
directly on the ow sensor or mounted remotely.
The “hot-tap” models (FP6511/6512) can be installed or serviced without shutting down the line by means of a 2” full-port isolation
valve that comes with a nipple for installation on the pipe tting.
In most circumstances, no special tool is required.
The resulting square-wave signal can be sent for hundreds of
FEATURES
3/4” diameter tubing for low insertion force
2” Adapter removes to mount hot-tap machine
Full-port 2” ball valve for sensor removal
Adapter tting with
2” NPT threads
FP6511/6512
(Add -SS for stainless)
Locking collar
Modular electronics (optional)
• rate/total/pulse/4-20 mA
• blind 4-20 transmitter
• pulse divider
18 Foot Cable Rugged cast aluminum housing
Compression nut for easy adjustment, secure locking
Adapter tting
with 1-1/2” male NPT threads
FP6501/6502
(Add -SS for stainless)
SPECIFICATIONS*
Pipe Size
Sensor
Materials Housing
Sensor Body
Rotor
Shaft
Bearings
Isolation Valve
Fitting Size
Flow Range
Accuracy
Maximum Temperature
Maximum Pressure
Insertion Force
Power
Cable
Regulatory
Page 1
Rotor housing
Removable jewel bearings for
exceptional low ow performance
Rotor
FP6501/6511 FP6502/6512
3” to 12” (50 - 300mm) 12” to 40” (300 - 1000mm)
Hall Effect sensor, 12 Vdc current sinking pulse
Cast aluminum
Brass or 316 SS
PVDF
Nickel-bound tungsten carbide (zirconia ceramic optional)
Ruby jewel
FP6501/6502 FP6511/6512
None Bronze (316SS optional)
1.5” NPT 2” NPT
0.3 - 30 feet/sec (0.1 - 9 meter/sec)
+/- 1.5% of full scale
200˚ F (93˚ C)
200 psi (14 bar)
0.44 x pressure in pipe
5-24 Vdc, 1.5 mA
#22 AWG 3-con, 18’ (6m); 2,000’ (650m) maximum cable run
Mark (Standard Power Only)
(Add -SS for stainless)
(Add -SS for stainless)
Note: For larger pipe sizes contact factory
*Specications subject to change.
Page 5
INSTALLATION
"D"
These ow sensors are not recommended for
installation downstream of the boiler feedwa-
ter pump where installation fault may expose the meter to boiler pressure and temperature. Maximum recommended temperature is
200ºF.
Piping. For best results, the FP6500 should be installed with at least ten diameters of straight pipe upstream and ve down-
stream. Certain extreme situations such as partially-opened
valves are particularly difcult and may require fteen diameters
upstream (See Straight Pipe Recommendations).
FLOW
10
X Dia.
5
X Dia.
Immersion. The FP6500-Series standard sensors are not
designed for continuous underwater operation.
Depth Setting. It is important for accuracy that the sensor be inserted to the correct depth into the pipe.
1. In Table 1, nd Dimension C for your sensor model
and pipe size.
2. Subtract wall thickness of your pipe (Table 2) to
nd Dimension D.
Horizontal is the preferred installation orientation, since it im-
proves low-ow performance slightly and avoids problems with
trapped air. Bottom, top, and vertical pipe installations are all acceptable if required by the piping layout. See Full Pipe Recom­mendations.
POSITIONING THE METER
Fair (unacceptable if
air is present)
Best
3. Measuring from the outside of the pipe to the joint in the housing, as shown in the diagram, adjust the sensor to Dimension D and hand tighten compression nut.
4. Align the conduit housing with the centerline of the pipe, as shown below. Be sure the arrow on the housing
points in the direction of ow.
strain relief
FLOW
5. Check Dimension D one more time.
6. Tighten the compression nut fully.
RECORD YOUR SETTINGS
Once you have the meter set up and operational, it is im­portant to record your meter setttings and save them for future reference.
Fair (unacceptable if uid
contains sediment)
K-Factor Insertion Depth (Dim. D)
Page 2
Page 6
INSTALLATION
FP6501/6502 INSTALLATION
(Add -SS for stainless)
Fitting Installation. FP6501/6502 sensors come with a 1-1/2” male
NPT pipe thread adapter tting. Any tting that provides the matching
NPT female thread may be used. Installation procedure compensates
for tting height differences. Cut a minimum 1-3/4” hole in the pipe. If
possible, measure the wall thickness and write it down for use in depth
setting. Then install the threaded tting (saddle, weldolet, etc.) on the
pipe.
Meter Installation. Loosen the compression nut so that the adapter
slides freely. Pull the meter fully upward and nger-tighten the compres­sion nut. Using a thread sealant, install the adapter in the pipe tting. Do
not overtighten. Now loosen the compression nut, lower the meter to the appropriate depth setting (see diagram and instructions, preceding page).
Caution: Do not allow the meter to fall into the pipe uncontrolled, as this may damag the meter. Be sure ow is in the direction of the arrow
on the housing. Tighten compression nut fully.
FP6511/6512 INSTALLATION
(Add -SS for stainless)
‘Hot tap’ FP6500 Series meters are designed to be installed and serviced
without depressurizing the pipe.
FP6501/6502
Compression nut
Adapter tting
with
standard NPT
threads
FP6511/6512
Compression nut
Fitting Installation. The sensors have a 2” NPT thread for compatibility
with the 2” isolation valve. Any tting that provides matching NPT
female thread may be used. The installation procedure compensates
for differences in tting height.
If initial installation is performed on an unpressurized pipe, cut a minimum 1-3/4” hole in the pipe. If possible, measure the wall thickness
and write it down for use in depth setting. Then install the threaded tting
(saddle, weldolet, etc.) on the pipe.
If it is necessary to do the initial installation under pressure, any standard hot tap drilling machine with 2” NPT adapter, such as a Transmate or a Mueller, can be used. Ordinarily, it is not necessary to use an installation tool, since the small-diameter tube can be controlled by hand at all but the highest pressures.
Meter Installation. Remove the sensor unit from the valve assembly.
Using a thread sealant, install the valve assembly on the pipe tting. If
the initial installation is a pressure (“hot”) tap, remove the 1-1/2” x 2” adapter bushing at the back of the valve. Thread the tapping machine on, open the valve, and tap using a minimum of 1-3/4” or maximum 1-7/8” cutter. After retracting the machine and closing the valve, reinstall the
ow sensor. When the sensor is secure, open the valve and adjust depth setting (see diagram and instructions, preceding page). Be sure ow is
in the direction of the arrow on the housing. Tighten locking collar and compression nut fully.
Page 3
Locking collar
2” adapter removes
to mount hot-tap
machine
Full-port 2” ball
valve allows sensor
removal
Standard 2” NPT threads
FP6511/6512
Sensor Removal
Page 7
INSTALLATION
Table 1: Dimension "C"
3" 4" 6" 8" 10" 12" 16" 24" 30" 36"
FP6501
FP6502
FP6511
FP6512
9.20 9.03 8.69 8.35 8.01 7.67 6.99 - - -
14.20 14.03 13.69 13.35 13.01 12.67 11.99 10.63 9.61 9.59
16.21 16.01 15.71 15.31 15.01 14.7 14.01 - - -
- 20.06 19.66 19.36 19.06 18.66 17.96 16.66 15.66 14.56
Table 2: Pipe Wall Thickness
Nominal Pipe Size
3" 4" 6" 8" 10" 12" 16" 24" 30" 36"
PVC/Steel
Sch. 40
0.216 0.237 0.280 0.322 0.365 0.406 0.500 0.687 - -
Nominal Pipe Size
PVC/Steel
Sch. 80
Stainless
Steel (10S)
Stainless
Steel (40S)
Copper Tubing
(Type L)
Copper Tubing
(Type K)
Brass Pipe
Duct. Iron
(Class 52)
0.300 0.337 0.432 0.500 0.593 0.687 0.843 1.218 - -
0.120 0.120 0.134 0.148 0.165 0.180 - - - -
0.216 0.237 0.280 0.322 0.365 0.375 0.375 0.375 0.375 -
0.090 0.100 0.140 0.200 0.250 0.280 - - - -
0.109 0.134 0.192 0.271 0.338 0.405 - - - -
0.219 0.250 0.250 0.313 0.365 0.375 - - - -
0.280 0.290 0.310 0.330 0.350 0.370 0.400 0.440 0.470 0.530
Page 4
Page 8
INSTALLATION
Feet / sec.
5
Table 3: Flow (GPM) (Sched. 40 pipe)
Nominal pipe size
3" 4" 6" 8" 10" 12" 16" 24" 30" 36"
(0.3)
(0.5)
(1.0)
(2.0)
(5.0)
(10.0)
(20.0)
(30.0)
3.1 4.5 27 46.8 73.7 105 165 376 660 874
11.5 19.8 45 78 123 174 275 627 1100 1460
23 39.7 90 156 246 349 551 1250 2200 2910
46.1 79.4 180 312 492 698 1100 2510 4401 5830
115 198 450 780 1230 1740 2750 6270 11002 14570
230 397 900 1560 2460 3490 5510 12530 22005 29140
461 794 1800 3120 4920 6980 11020 25060 44010 58270
691 1190 2700 4680 7370 10470 16520 37600 66015 87410
Table 4: FP6500 Series K-factors
Nominal pipe size
3" 4" 6" 8" 10" 12" 16" 24" 30" 36"
Page 5
PVC/Steel
Sch. 40
PVC/Steel
Sch. 80
Stainless Steel
(10S)
Stainless Steel
(40S)
Copper Tubing
(Type K)
Copper Tubing
(Type L)
Brass Pipe
Duct. Iron
(Class 52)
34.246 19.221 7.830 4.611 2.883 1.859 1.224 .577 - -
38.329 21.844 8.779 4.980 3.213 2.097 1.304 .656 - -
30.331 16.581 7.045 4.270 2.621 1.703 1.160 .519 .332 -
34.246 19.221 7.830 4.611 2.883 1.836 - - - -
38.145 21.420 8.846 5.027 3.302 2.126 - - - -
37.167 20.774 8.505 4.873 3.167 2.011 - - - -
34.381 19.545 7.658 4.593 2.883 1.836 - - - -
28.137 16.994 7.238 4.213 2.633 1.681 1.065 .461 .297 .207
Page 9
INSTALLATION
STRAIGHT PIPE RECOMMENDATIONS
(X = diameter)
Reduced Pipe
Two Elbows In Plane
Two Elbows, Out Of Plane
10X
20X
5X
5X10X
5X
Expanded Pipe
Spiral Flow
Swirling Flow
Propeller Meter
Partially Open
Buttery Valve
20X
5X
30X
50X
Page 6
Page 10
INSTALLATION
FULL PIPE RECOMMENDATIONS
Allows air pockets to form at sensor
RECOMMENDEDNOT RECOMMENDED
Ensures full pipe
RECOMMENDEDNOT RECOMMENDED
Post-valve cavitation can create air pocket Keeps pipe full at sensor
RECOMMENDEDNOT RECOMMENDED
Air can be trapped
Allows air to bleed off
Caution: These ow sensors are not recommended for installation down- stream of the boiler feedwater pump where installation fault may expose the
ow sensor to boiler pressure and temperature. Maximum recommended temperature is 200°F.
Page 7
Page 11
CONNECTION, OPERATION & REPAIR
RED (+) 6-24 Vdc
WHITE (signal)
BLACK (-) Power
18' cable standard
CONNECTION
Sensors are supplied with 18 ft. of cable. See diagram for color coding of connections.
Calibration (“K-Factor”). In order to properly process pulses from the ow sensor, a number must be entered into the con­trol to which the sensor is connected. This number, called the K-factor, is the number of pulses the sensor puts out per unit
of uid passing through the pipe. It is provided in pulses per
gallon, and is given on the chart “K-factors for Various Pipe Sizes.” These numbers are based on extensive testing, which
has shown close agreement between different FP6500-Series
sensors in the same installation. Typically, most K-factor error can be attributed to installation variables, such as depth setting
and tting conguration.
It is occasionally possible to eld calibrate a sensor by catching the uid in a measured container and comparing with the number
of pulses recorded. (To record individual pulses, set the K-factor on the control to 1.00.) This is especially desirable if the instal­lation has less than the recommended length of straight pipe upstream of the sensor.
OPERATION Theory. In principle, an insertion ow sensor measures the veloc-
ity of ow at one point in the pipe, and ow rate and total can be
inferred from this one point. Accuracy is decreased by any factor
which makes the ow at the measured point unrepresentative of the entire ow stream. This includes distorted ow patterns caused by upstream ttings too close to the sensor. The worst of­fenders are ttings that increase the ow on one side of the pipe, such as partially-opened gate or buttery valves. Fluid moving in a pipe does not all ow at the same velocity. Toward the center of the pipe, uid moves faster than at the wall, and the relation­ship between the two changes as overall ow rate increases. This change in the “velocity prole” can result in non-linearity, which means that the K-factor that is correct for one ow rate
may be incorrect for another. The recommended depth settings have been carefully chosen to minimize this source of error, and should be followed carefully, especially in the smaller pipe sizes.
Flow Range. These sensors are designed to operate at
ow velocities of 0.3 to 30 feet per second. (See chart for
conversion to gallons per minute.) If erratic readings are
encountered at low ows, check the chart to see if ow is
below minimum for the pipe size. The standard shaft and
bearings should have a long life at continuous high ow.
REPAIR
CAUTION! Never attempt to remove a
ow sensor when there is pressure in
the pipe. Loosen the compression nut slowly to release any trapped pressure.
If uid sprays out when removing the
sensor, stop turning and depressurize the pipe. Failure to do so could result in the sensor being thrown from the pipe, resulting in damage or serious injury.
Rotor Replacement. Rotors are easily eld-replaced. Shaft
and rotor are a single unit, and are not replaced separately. If replacement is due only to normal shaft wear, bearing re­placement is probably not necessary. If the rotor has been damaged by impact, the bearings should also be replaced. Rotor and bearings can be ordered as a kit, FP15827. Fol­low these steps:
1. Unscrew the threaded bearing housings to expose the shaft ends. If bearings are being replaced, back them completely out.
2. Remove the rotor. Put the new rotor in its place.
3. Thread in one bearing housing part way, then the other. Take care to start the end of the shaft into the bearing hole before tightening further.
4. Screw in bearing housings until they bottom. Note: Do not use excessive force.
5. Check for free spin. Blowing lightly on the rotor should result in it spinning rapidly and coasting to a smooth stop.
Page 8
Page 12
2
1
9
8
15
1
6
12
5
6
4
3
7
14
13
11
10
15
8
REPAIR & PARTS
Signal
The ow sensor has only one moving part, the rotor. If this is
turning properly and there is no signal, the Hall-effect sensor is not operating properly. To check the signal, apply 12 Vdc regu­lated* power to the red (+) and black (-) leads. Set a multimeter to voltage reading. Put the positive multimeter lead on the red wire and the negative lead on the white wire. Slowly turn the
rotor. Voltage reading should swing between +12 Volts and 0
Volts as the rotor turns. If it does not, the Hall effect sensor is not working properly. Checking for continuity is not a useful test of these sensors.
*NOTE: An unregulated power supply can exceed max voltage of micro powered sensor (gray cable) and damage sensor.
FP6501/6502 Parts
1 Upper housing assembly 30475
2 Gasket 26211
3 Lower housing Not Field Replaceable
4 Housing screw (4 req'd) 26229
5 Plug, steel 26073
6 Plug, plastic 26079
Parts Explosion
7 Strain relief 07655
Pickup, Micropower (for FT415)
8
Pickup, Standard (for FT420)
9 Tube Not Field Replaceable
10 Compression nut
11 Compression ferrule 26065
12 Adapter tting
13 Rotor housing O-ring (EPDM) 16454
14 Rotor housing
15 Jewel bearings (2 req) 25901
16 Rotor with shaft
Rotor repair kit
17
(includes of #15 & #16)
29953
26310
14199 (Brass) 15064 (S/S)
30998 (Brass) 30999 (S/S)
14220 (Brass) 15220 (S/S)
11130 (PVDF/Tungsten Carbide) 11132 (PVDF/Ceramic)
25902 (PVDF/Tungsten Carbide) 12131 (PVDF/Ceramic)
Page 9
HOT TAP: FP6511/6512 Parts (not shown)
Adapter tting
Ball valve assembly
Collar, locking
Hex nipple, 2"
14180 (Brass) 15060 (S/S)
14225 (Brass) 15225 (S/S)
14190 (Brass) 15070 (S/S)
14210 (Brass) 15100 (S/S)
Page 13
TROUBLESHOOTING
Problem
No pulse output
Output pulses incorrect
Jumpy reading
Probable Cause Try...
Below minimum ow cutoff
Empty pipe
No power
Incorrect depth setting
Pipe not full
Not enough straight pipe
Fluctuating ow rate
Fluctuating around low ow cutoff
Not enough straight pipe
Check Table for velocity vs. pipe size
Check plumbing
Check connections
Check depth setting from Dimension “C” table
Refer to Installation/diagrams
Refer to Installation/diagrams
Refer to installation diagrams
Check Table for velocity vs. pipe size
Refer to Installation/diagrams
Page 10
Page 14
WARRANTY/DISCLAIMER
FOR NON-WARRANTY REPAIRS,
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months f rom date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be re t u r ned to the factory for evaluation. O M E G A ’s Customer Serv i c e D e p a rtm ent will issue an Authorized Return (AR) number immediately upon phone or written re q u e s t . Upon examination by OMEGA, if the unit is found to be defective, it will be re p a i red or replaced at no c h a rge. O M E G A ’s WARRANTY does not apply to defects resulting from any action of the purc h a s e r , including but not limited to mishandling, improper interfacing, operation outside of desi i m p roper re p a i r, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corro s i o n ; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s c o n t r ol. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to o ffer suggestions on the use of its various p roducts. Howev er, OMEGA neither assumes responsibility for any omissions or e damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR R E P R E S E N T ATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY W AND FITNESS FOR A PA R TICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF L I A B I L I T Y: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this ord e r , whether based on contract, warr a n t y, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the componen t upon which liab ility is based. In no eve n t shall OM EGA be liab le for co
nsequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or a c t i v i t y , medical application, used on humans, or misused in any way, OMEGA assumes no re s p o n s i b i l i t y as set forth in our basic WA R R A N T Y/ D ISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA h a rmless from any liability or damage whatsoever arising out of the use of the P roduct(s) in such a manner.
rrors nor assumes liability for any
ARRANTY OF MERCHANTA B I L I T Y
gn limits,
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR
) NUMB E R F ROM OME G A’ S C U S TOME R S E RV I C E DEPA R TMENT ( I N O R DER TO AV O I D PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
WARRANTY RETURNS, please have the
FOR following information available BEFORE contacting OMEGA:
1 . P u r chase Order number under which the pro d u c t
was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2005 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.
consult OMEGA
Page 15
W h e r e Do I Find Eve rything I Need for
P r ocess Measurement and Control?
OME GA…Of Cours e !
Shop online at omega.com
T E M P E R A T U R E

Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

Wire: Thermocouple, RTD & Thermistor

Calibrators & Ice Point References

Recorders, Controllers & Process Monitors

Infrared Pyrometers
PRESSURE, STRAIN AND FO R C E

Transducers & Strain Gages

Load Cells & Pressure Gages

Displacement Transducers

Instrumentation & Accessories
F L OW / L E V E L

Rotameters, Gas Mass Flowmeters & Flow Computers

Air Velocity Indicators

Turbine/Paddlewheel Systems

Totalizers & Batch Controllers
p H / C O N D U C T I V I T Y

pH Electrodes, Testers & Accessories

Benchtop/Laboratory Meters

Controllers, Calibrators, Simulators & Pumps

Industrial pH & Conductivity Equi
pment
DATA AC Q U I S I T I O N

Data Acquisition & Engineering Software

Communications-Based Acquisition Systems

Plug-in Cards for Apple, IBM & Compatibles

Datalogging Systems

Recorders, Printers & Plotters
H E A T E R S

Heating Cable

Cartridge & Strip Heaters

Immersion & Band Heaters

Flexible Heaters

Laboratory Heaters
E N V I R O N M E N T A L M O N I T ORING AND CONTRO L

Metering & Control Instrumentation

R e f r a c t o m e t e r s

Pumps & Tubing

Air, Soil & Water Monitors

Industrial Water & Wastewater Treatment

pH, Conductivity & Dissolved Oxygen Instruments
PL-OM-10611-051811 05/18/2011
M-1949/0511
Loading...