Omega FMA1800A, FMA1700A User Manual

Page 1
FMA1700A & FMA1800A
Mass Flow Meter
TM
e-mail: info@omega.com
For latest product manuals:
www.omegamanual.info
Shop online at
User’s Guide
NORWALK, CT
Page 2
omega.com info@omega.com
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
Servicing North America:
U.S.A. Omega Engineering, Inc. Headquarters: Toll-Free: 1-800-826-6342 (USA & Canada only)
Customer Service: 1-800-622-2378 (USA & Canada only) Engineering Service: 1-800-872-9436 (USA & Canada only) Tel: (203) 359-1660 Fax: (203) 359-7700 e-mail: info@omega.com
For Other Locations Visit omega.com/worldwide
Page 3
TABLE OF CONTENTS
1. UNPACKING THE FMA1700A/1800A MASS FLOW METER.....................
1.1 Inspect Package for External Damage.................................................
1.2 Unpack the Mass Flow Meter...............................................................
1.3 Returning Merchandise for Repair.......................................................
2. INSTALLATION........................................................................................
2.1 Primary Gas Connections.................................................................
2.2 Electrical Connections......................................................................
2.3.1 Remote LCD Readouts.....................................................................
2.3.2 Panel Mounting Readouts................................................................
3. PRINCIPLE OF OPERATION...................................................................
4. SPECIFICATIONS...................................................................................
4.1 CE Compliance.................................................................................
5. OPERATING INSTRUCTIONS..................................................................
5.1 Preparation and Warm Up..................................................................
5.2 Flow Signal Output Readings..........................................................
5.3 Swamping Condition.......................................................................
6. MAINTENANCE.........................................................................................
6.1 Introduction......................................................................................
6.2 Flow Path Cleaning...........................................................................
6.2.1 Restrictor Flow Element (RFE)........................................................
6.2.2 FMA1700A/1800A Series Max. Flow 15 L/min...................................
7. CALIBRATION PROCEDURES.................................................................
7.1 Flow Calibration...............................................................................
7.2 Calibration of FMA1700A/1800A Mass Flow Meters........................
7.2.1 Connections and Initial Warm Up.....................................................
7.2.2 ZERO Adjustment.............................................................................
7.2.3 SPAN Adjustment............................................................................
7.3 Linearity Adjustment..........................................................................
7.3.1 Connections and Initial Warm Up....................................................
7.3.2 ZERO Adjustment.............................................................................
7.3.3 25% Flow Adjustment (using R33 potentiometer)...........................
7.3.4 10% Flow Adjustment......................................................................
7.3.5 25% Flow Adjustment (using R52 potentiometer)..........................
1 1 1 1
1 1 2 4 4
5
6 7
9 9 9
10
10 10 10 10 11
11 11 13 13 13 13 14 14 14 14 15 15
Page 4
7.3.6 50% Flow Adjustment....................................................................
7.3.7 75% Flow Adjustment....................................................................
7.3.8 100% Flow Adjustment....................................................................
7.4 LCD Display Scaling.........................................................................
7.4.1 Access LCD Display Circuit..............................................................
7.4.2 Adjust Scaling..................................................................................
7.4.3 Change Decimal Point......................................................................
8. TROUBLESHOOTING.............................................................................
8.1 Common Conditions........................................................................
8.2 Troubleshooting Guide.....................................................................
8.3 Technical Assistance.......................................................................
9. CALIBRATION CONVERSIONS FROM REFERENCE GASES.......................
APPENDIX 1 COMPONENT DIAGRAM.........................................................
APPENDIX 2 GAS FACTOR TABLE (“K” FACTORS)........................................
APPENDIX 3 DIMENSIONAL DRAWINGS......................................................
PARTS OF THE FLOW METER..............................................
15 15 15 15 16 16 16
16 16 17 21
22
23
24
28 29
Page 5
1
1. UNPACKING THE FMA1700A/1800A MASS FLOW METER
1.1 Inspect Package for External Damage
Remove the Packing List and verify that you have received all equipment. If you have any questions about the shipment, please call the OMEGA7 Customer
Service Department at 1-800-622-2378 or (203) 359-1660.
Your FMA1700A/1800A Mass Flow Meter was carefully packed in a sturdy card­board carton, with anti-static cushioning materials to withstand shipping shock. Upon receipt, inspect the package for possible external damage. In case of exter­nal damage to the package contact the shipping company immediately.
1.2 Unpack the Mass Flow Meter
Open the carton carefully from the top and inspect for any sign of concealed ship­ping damage. In addition to contacting the shipping carrier please forward a copy
of any damage report to OMEGA7 directly.
When unpacking the instrument please make sure that you have all the items indicated on the Packing List. Please report any shortages promptly.
1.3 Returning Merchandise for Repair
Please contact an OMEGA7 customer service representative and request a
Return Authorization Number (AR).
It is mandatory that any equipment returned for servicing be purged and neutral­ized of any dangerous contents including but not limited to toxic, bacterially infec­tious, corrosive or radioactive substances. No work shall be performed on a returned product unless the customer submits a fully executed, signed SAFETY CERTIFICATE. Please request form from the Service Manager.
2. INSTALLATION
2.1 Primary Gas Connections
Please note that the FMA1700A/1800A Mass Flow Meter will not operate with liq­uids. Only clean gases are allowed to be introduced into the instrument. If gases are contaminated they must be filtered to prevent the introduction of impediments into the sensor.
CAUTION: Some of the IC devices used in the FMA1700A & FMA1800A are Electro Static Discharge (ESD) sensitive and may be damaged by improper handling. When wiring interface connector, adjusting or servicing the meter, use of a grounded ESD protection wrist strap is required to prevent inadvertent damage to the CMOS integral solid state circuitry. When 9 pin Interface D-connector is not used do not remove factory installed ESD protection cover.
Page 6
2
CAUTION: IT IS THE USERS RESPONSIBILITY TO DETERMINE IF THE INSTRUMENT IS APPROPRIATE FOR THEIR OXYGEN APPLICATION, AND FOR SPECIFYING O2 CLEANING SERVICE IF REQUIRED. OMEGA IS NOT LIABLE FOR ANY DAMAGE OR PERSONAL INJURY, WHATSOEVER, RESULTING FROM THE USE OF THIS INSTRUMENT FOR OXYGEN GAS.
Attitude sensitivity of the Mass Flow Meter is ±15. This means that the gas flow path of the Flow Meter must be horizontal within those stated limits. Should there be need for a different orientation of the meter, re-calibration may be necessary. It is also preferable to install the FMA1700A/1800A transducer in a stable environ­ment, free of frequent and sudden temperature changes, high moisture, and drafts.
Prior to connecting gas lines inspect all parts of the piping system including fer­rules and fittings for dust or other contaminant’s.
Be sure to observe the direction of gas flow as indicated by the arrow on the front of the meter when connecting the gas system to be monitored.
Insert tubing into the compression fittings (except FMA1700A/1800A Series Max Flow 1000 L/min) until the ends of the properly sized tubings home flush against the shoulders of the fittings. Compression fittings are to be tightened according to the manufacturer's instructions to one and one quarter turns. Avoid over tighten­ing which will seriously damage the Restrictor Flow Elements (RFE's)!
Using a Helium Leak Detector or other equivalent method perform a thorough leak test of the entire system
. (All FMA1700A/1800A are checked prior to ship-
ment for leakage within stated limits. See specifications in this manual.)
2.2 Electrical Connections
CAUTION: Some of the IC devices used in the FMA1700A & FMA1800A are Electro Static Discharge (ESD) sensitive and may be damaged by improper handling. When wiring interface connector, adjusting or servicing the meter, use of a grounded ESD protection wrist strap is required to prevent inadver tent damage to the CMOS integral solid state circuitry. When 9 pin interface D-connector is not used do not remove factory installed ESD protection cover.
CAUTION: WIRING THE FMA1700A & FMA1800A METER WITH THE POWER ON MAY RESULT IN INTERNAL DAMAGE! PLEASE MAKE ALL WIRING CONNECTIONS BEFORE SWITCHING ON THE POWER.
All FMA1700A & FMA1800A models require power supplies with nominal current rating of at least 200 mA. The operating power input is supplied via the DC power jack or the 9-pin “D” connector located at the side of the flow transducer enclosure. On FMA1700A & FMA1800A's purchased without an LCD readout, a readout panel meter, digital multimeter, or other equivalent device is required to observe the flow signal.
Page 7
3
PIN FUNCTION
1 Remote LCD display signal 2 0 to 5 VDC output indication 3 0 to 5 VDC common 4 Power supply, positive 5 Power supply, common 6 Remote LCD display reference 7 (unassigned) 8 4 to 20 mA output indication
self-powered (sourcing type, non-isolated)
9 4 to 20 mA common (return)
FIGURE 2.a - 9-PIN “D” CONNECTOR PINOUTS FOR FMA1700A/1800A TRANSDUCER
The FMA1700A & FMA1800A flow meters have universal power input and can be used with any power supply voltage between +12 and +26 Vdc.
CAUTION: BEFORE CONNECTING POWER SUPPLY CHECK YOUR METER SERIAL NUMBER AND POWER SUPPLY REQUREMENTS LABEL LOCATED ON THE FLOW METER BACK COVER. DO NOT CONNECT POWER SUPPLY WITH VOLTAGE ABOVE 26 VDC. EXCEEDING SPECIFIED MAXIMUM POWER SUPPLY VOLTAGE LIMIT MAY RESULT IN DEVICE PERMANENT DAMAGE.
IMPORTANT NOTES:
In general, “D” Connector numbering patterns are standardized. There are, how­ever, some connectors with nonconforming patterns and the numbering sequence on your mating connector may or may not coincide with the numbering sequence shown in our pin configuration table above. It is imperative that you match the appropriate wires in accordance with the correct sequence regardless of the par­ticular numbers displayed on your mating connector.
Make sure power is OFF when connecting or disconnecting any cables in the system.
When connecting power to the FMA1700A/1800A mass flow meter via the DC power jack, do not connect any power supply to the 9-pin “D” Connector. The DC power jack has a center positive polarity.
When battery use is required to power the FMA1700A/1800A, use only the optional battery and accompanying charger available from OMEGA7.
Page 8
4
CAUTION: Resettable fuse will not protect meter if power supply voltage exceeds maximum voltage specified for particular model.
In order to operate flow meter within the specification the cable length for analog 0-5 Vdc output and remote LCD display may not exceed 9.5 feet (3 meters).
Use of the FMA1700A/1800A flow transducer in a manner other than that speci­fied in this manual or in writing from OMEGA7, may impair the protection provid-
ed by the equipment.
2.3.1 Remote LCD Readouts
FMA1700A/1800A Mass Flow Meters are available with optional remote reading LCD displays supplied with a three foot long wire to accommodate most applica­tions. This configuration includes the upper block element which serves as the LCD readout mounting. Special lengths of remote extension wiring (up to 9.5 feet [3 meters]) are available on request.
2.3.2 Panel Mounting Readouts
Another option for the FMA1700A/1800A Mass Flow Meter is the Panel Mounting Remote Readout. In this configuration the LCD readout is supplied with a three foot long extension wire, and no aluminum housing around the LCD. The LCD readout for panel mounting includes a bezel with two plastic screws which conve­niently fit into a rectangular cut-out for panel mounting (see Figure 2.b).
The power input is protected by 300 mA M (medium time-lag) resettable fuse. If a shorting condition or polarity reversal occurs, the fuse will cut power to the flow transducer circuit. Disconnect the power to the unit, remove the faulty condition, and reconnect the power. The fuse will reset once the faulty condition has been removed.
FIGURE 2.b - CUTOUT DIMENSIONS FOR LCD PANEL MOUNTING
Page 9
3. PRINCIPLE OF OPERATION
The stream of gas entering the Mass Flow transducer is split by shunting a small portion of the flow through a capillary stainless steel sensor tube. The remainder of the gas flows through the primary flow conduit. The geometry of the primary con­duit and the sensor tube are designed to ensure laminar flow in each branch. According to principles of fluid dynamics the flow rates of a gas in the two laminar flow conduits are proportional to one another. Therefore, the flow rates measured in the sensor tube are directly proportional to the total flow through the transducer.
In order to sense the flow in the sensor tube, heat flux is introduced at two sec­tions of the sensor tube by means of precision wound heater-sensor coils. Heat is transferred through the thin wall of the sensor tube to the gas flowing inside. As gas flow takes place heat is carried by the gas stream from the upstream coil to the downstream coil windings. The resultant temperature dependent resistance differential is detected by the electronic control circuit. The measured gradient at the sensor windings is linearly proportional to the instantaneous rate of flow tak­ing place.
An output signal is generated that is a function of the amount of heat carried by the gases to indicate mass-molecular based flow rates.
5
Page 10
4. SPECIFICATIONS
FLOW MEDIUM: Please note that FMA1700A/1800A Series Max Flow 15, 50, 100, 200, 500 and 1000 L/min Mass Flow Meters are designed to work with clean gases only. Never try to meter or control flow rates of liquids with any FMA1700A/1800A.
CALIBRATIONS: Performed at standard conditions [14.7 psia (1.01 bars) and 70
F
F
(21.1
F
C)] unless otherwise requested or stated.
ENVIRONMENTAL (per IEC 664): Installation Level II; Pollution Degree II.
ACCURACY: FMA1700A/1800A Series Max Flow 15, 50 and 100 L/min ±1.0% F.S.
FMA1700A/1800A Series Max Flow 200, 500 and 1000 L/min. ±1.5% F.S.
REPEATABILITY: ±0.25% of full scale.
TEMPERATURE COEFFICIENT: 0.15% of full scale/
F
C.
PRESSURE COEFFICIENT: 0.01% of full scale/psi (0.07 bar).
RESPONSE TIME: 800 ms time constant; approximately 2 seconds to within ±2% of set
flow rate for 25% to 100% of full scale flow rate.
GAS PRESSURE: 1000 psig (69 bars) FMA1700A/1800A Series Max Flow 15, 50 and 100; 500 psig (34.5 bars) FMA1700A/1800A Series Max Flow 200, 500 and 1000 L/min. Optimum pressure is 20 psig (1.4 bars).
GAS AND AMBIENT TEMPERATURE: 32
F
F to 122 FF (0 FC to 50 FC).
14
F
F to 122 FF (-10 FC to 50 FC) - Dry gases only.
RELATIVE GAS HUMIDITY: Up to 70%.
LEAK INTEGRITY: 1 x 10
- 7
sccs He max. to the outside environment.
ATTITUDE SENSITIVITY: Incremental deviation of up to 1% full scale from stated accuracy, after re-zeroing.
OUTPUT SIGNALS: Linear 0-5 VDC (1000 Ω minimum load impedance) and 4-20 mA self- powered sourcing type, non-isolated, (0-500 Ω loop resistance); 20 mV peak to peak max noise.
Contact Omega© for optional RS232 or RS485 interfaces.
6
TRANSDUCER INPUT POWER:
FMA1700A & FMA1800A models with universal power input between +12 and +25 VDC, 200 mA maximum;
Power input is protected by a 300 mA M (medium time-lag) resettable fuse, and a rectifier diode for polarity protection.
Page 11
7
WETTED MATERIALS:
FMA1700A/1800A Series Max Flow 15, 50, 100, 200, 500 and 1000 L/min: Anodized
aluminum, brass, and 316 stainless steel with FKM O-rings seals; BUNA, EPR or Perfluoroelastomer O-rings are optional.
FMA1700A/1800A Series Max Flow 15, 50, 100, 200, 500 and 1000 L/min: 316 stain­less steel with FKM O-rings seals; BUNA, EPR or Perfluoroelastomer O-rings are optional.
OMEGA7 makes no expressed or implied guarantees of corrosion resistance of mass flow meters as pertains to different flow media reacting with components of meters. It is the customers' sole responsibility to select the model suitable for a particular gas based on the fluid contacting (wetted) materials offered in the different models.
INLET AND OUTLET CONNECTIONS:
FMA1700A/1800A Series Max. Flow 15 and 50 L/min: 1/4" compression fittings.
Optional: 6mm compression, 1/4" VCR
®
, 3/8" or 1/8" compression fittings (Max. Flow 15).
FMA1700A/1800A Series Max. Flow 100 and 200 L/min: 3/8" compression fittings. FMA1700A/1800A Series Max. Flow 500 L/min: 1/2" compression fittings. FMA1700A/1800A Series Max. Flow 1000 L/min: 3/4" FNPT fittings or 3/4"
compression fittings.
LCD DISPLAY: 3½ digit LCD (maximum viewable digits “1999”), 0.5 inch high characters. On FMA1700A/1800A aluminum or stainless steel models the LCD display is built into the upper block element and may be tilted over 90 degrees for optimal viewing comfort. Remote or panel mounting remote reading is optional.
Standard readings are in direct engineering units for the given gas and flow rate (i.e. liters/minute [lpm], standard cubic centimeters/minute [sccm], standard cubic feet/hour [scfh], etc.). 0 to 100% LCD calibration scaling is available upon request at time of order. Contact OMEGA7 when non-standard display settings are desired.
TRANSDUCER INTERFACE CABLE: Optional shielded cable is available mating to the FMA1700A/1800A transducer 9-pin “D” connector. In order to operate flow meter within the specification the cable length for analog 0-5 Vdc output and remote LCD display may not exceed 9.5 feet (3 meters).
4.1 CE Compliance
Any model FMA1700A/1800A bearing a CE marking on it, is in compliance with the below stated test standards currently accepted.
EMC Compliance with 89/336/EEC as amended; Emission Standard: EN 55011:1991, Group 1, Class B Immunity Standard: EN 55082-2:1992.
Page 12
8
FLOW RANGES
TABLE II FMA1700A/1800A
SERIES MAX FLOW 50 L/min
MEDIUM FLOW MASS FLOW METER*
TABLE III FMA1700A/1800A
SERIES MAX FLOW 100 L/min
HIGH FLOW MASS FLOW METER*
* Flow rates are stated for Nitrogen at STP
conditions [i.e. 70
F
F (21.1 FC) at 1 atm].
For other gases use the K factor as a multiplier from APPENDIX 2.
CODE
L/min [N2]
23 15
24
20 26 30
27 40 28 50
CODE
L/min [N2]
40 60 41 80
42 100 43 200 44 500 45 1000
TABLE I FMA1700A/1800A
SERIES MAX FLOW 15 L/min
LOW FLOW MASS FLOW METER*
CODE
mL/min [N2]
02 0 to 10 04 0 to 20
06 0 to 50 08 0 to 100 10 0 to 200 12 0 to 500
CODE
L/min [N2]
14 0 to 1
16 0 to 2
18 0 to 5
20 0 to 10
TABLE IV PRESSURE DROPS
MAXIMUM
FLOW RATE
SERIES
FLOW RATE
[liters/min]
MAXIMUM PRESSURE DROP
[mm H2O]
[psid] [mbar]
10 L/min up to 10 25 0.04 2.5
50 L/min
15 63 0.09 6.4 20 300 0.44 30 30 800 1.18 81 40 1480 2.18 150 50 2200 3.23 223
100 L/min
60 3100 4.56 314
100 5500 8.08 557 200 L/min 200 2720 4.0 280 500 L/min 500 3400 5.0 340
1000 L/min 1000 6120 9.0 620
Page 13
9
5. OPERATING INSTRUCTIONS
5.1 Preparation and Warm Up
It is assumed that the Mass Flow Meter has been correctly installed and thor­oughly leak tested as described in section 2. Make sure the flow source is OFF. Apply power to the unit by plugging the power supply line into the DC power jack (or 9-pin 'D' connector) on the side of the meter. Allow the Mass Flow Meter to warm-up for a minimum of 15 minutes.
SUPPLYING DC POWER TO THE POWER JACK AND THE “D” CONNECTOR AT THE SAME TIME WILL DAMAGE THE METER. DC POWER JACK POLARITY IS CENTER POSITIVE.
During initial powering of the FMA1700A/1800A transducer, the flow output sig­nal will be indicating a higher than usual output. This is indication that the FMA1700A/1800A transducer has not yet attained it's minimum operating tem­perature. This condition will automatically cancel within a few minutes and the transducer should eventually zero. If after the 15 minutes warm-up period, the display still indicates a reading of less than ± 3.0 % of F.S., readjust the ZERO potentiometer [R34] through the access hole. Before zero adjustment it is good practice to temporarily disconnect the gas source, to ensure that no see page or leak occurs in to the meter.
5.2 Flow Signal Output Readings
Initiate a controlled gas flow after warm up. The flow signal output can be viewed either on the LCD display, remote panel meter, digital multimeter, or other display device.
If an LCD display has been supplied with the FMA1700A/1800A, the observed reading is in direct engineering units (0 to 100% indication is optional).
Analog output flow signals of 0 to 5 VDC and 4 to 20 mA are attained at the appro­priate pins of the 9-pin “D” connector (see Figure 2.a) on the side of the FMA1700A/1800A transducer.
Meter signal output is linearly proportional to the mass molecular flow rate of the gas being metered. By default calibration is done against 0 to 5 VDC output sig­nal. If 4-20 mA output signal is used for flow indication on the FMA1700A/1800A, which was calibrated against 0 to 5 VDC, the total uncertainty of the reading may be in the range of +
2.5% of full scale. Optional calibration for 4-20 mA output sig­nal is available upon request at time of order. The full scale range and gas for which your meter has been calibrated are shown on the flow transducers front label.
ADJUSTING ZERO READING MORE THAN ± 3.0% F.S. FROM THE FACTORY SETTINGS MAY AFFECT DEVICE CALIBRATION ACCURACY. IF SUCH ADJUSTMENT IS REQUIRED IT IS RECOMMENDED TO PERFORM METER RECALIBRATION TO PRESERVE DEVICE ACCURACY.
Page 14
10
,
5.3 Swamping Condition
If a flow of more than 10% above the maximum flow rate of the Mass Flow Meter is taking place, a condition known as “swamping” may occur. Readings of a “swamped” meter cannot be assumed to be either accurate or linear. Flow must be restored to below 110% of maximum meter range. Once flow rates are lowered to within calibrated range, the swamping condition will end. Operation of the meter above 110% of maximum calibrated flow may increase recovery time.
6. MAINTENANCE
6.1 Introduction
It is important that the Mass Flow Meter/Controller is used with clean, filtered gases only. Liquids may not be metered. Since the RTD sensor consists, in part, of a small capillary stainless steel tube, it is prone to occlusion due to impedi­ments or gas crystallization. Other flow passages are also easily obstructed. Therefore, great care must be exercised to avoid the introduction of any potential flow impediment. To protect the instrument a 50 micron (FMA1700A/1800A Series Max Flow 15 L/min) or 60 micron (FMA1700A/1800A Series Max Flow 100 and 200 L/min) filter is built into the inlet of the flow transducer. The filter screen and the flow paths may require occasional cleaning as described below. There is no other recommended maintenance required. It is good practice, however, to keep the meter away from vibration, hot or corrosive environments and excessive RF or magnetic interference.
If periodic calibrations are required they should be performed by qualified per­sonnel and calibrating instruments, as described in section 7. It is recommended that units are returned to OMEGA7 for repair service and calibration.
CAUTION: TO PROTECT SERVICING PERSONNEL IT IS MANDATORY THAT ANY INSTRUMENT BEING SERVICED IS COMPLETELY PURGED AND NEUTRALIZED OF TOXIC, BACTERIOLOGICALLY INFECTED, CORROSIVE OR RADIOACTIVE CONTENTS.
6.2 Flow Path Cleaning
Before attempting any disassembly of the unit for cleaning, try inspecting the flow paths by looking into the inlet and outlet ends of the meter for any debris that may be clogging the flow through the meter. Remove debris as necessary. If the flow path is not unclogged, then proceed with steps below.
Do not attempt to disassemble the sensor. If blockage of the sensor tube is not alle­viated by flushing through with cleaning fluids, please return meter for servicing.
NOTE: DISASSEMBLY MAY COMPROMISE CURRENT CALIBRATION.
6.2.1 Restrictor Flow Element (RFE)
The Restrictor Flow Element (RFE) is a precision flow divider inside the trans­ducer, which splits the inlet gas flow by a preset amount to the sensor and main flow paths. The particular RFE used in a given Mass Flow Meter depends on the gas and flow range of the instrument.
Page 15
6.2.2 FMA1700A/1800A Series Max Flow 15 L/min models
Unscrew the inlet compression fitting of meter. Note that the Restrictor Flow Element (RFE) is connected to the inlet fitting.
Carefully disassemble the RFE from the inlet connection. The 50 micron filter screen will now become visible. Push the screen out through the inlet fitting. Clean or replace each of the removed parts as necessary. If alcohol is used for clean­ing, allow time for drying.
Inspect the flow path inside the transducer for any visible signs of contaminant. If necessary, flush the flow path through with alcohol. Thoroughly dry the flow paths by flowing clean dry gas through.
Carefully re-install the RFE and inlet fitting, avoiding any twisting and deforming the RFE. Be sure that no dust has collected on the O-ring seal.
NOTE: OVER TIGHTENING WILL DEFORM AND RENDER THE RFE DEFECTIVE.
It is advisable that at least one calibration point be checked after re-installing the inlet fitting-See section (g).
IT IS NOT RECOMMENDED TO ATTEMPT TO DISASSEMBLE, OR REPAIR MAXIMUM FLOW RATE SERIES MODELS 50 L/min, 100 L/min, 200 L/min, 500 L/min or 1000 L/min. DISASSEMBLY NECESSITATES RE-CALIBRATION.
7. CALIBRATION PROCEDURES
NOTE: REMOVAL OF THE FACTORY INSTALLED CALIBRATION SEALS AND/OR ANY ADJUSTMENTS MADE TO THE METER, AS DESCRIBED IN THIS SECTION, WILL VOID ANY CALIBRATION WARRANTY APPLICABLE.
7.1 Flow Calibration
OMEGA7 Engineering Flow Calibration Laboratory offers professional calibration support for Mass Flow Meters, using precision calibrators under strictly controlled conditions. NIST traceable calibrations are available. Calibrations can also be per­formed at customers' site using available standards.
Factory calibrations are performed using NIST traceable precision volumetric calibrators incorporating liquid sealed frictionless actuators. Generally, calibra­tions are performed using dry nitrogen gas. The calibration can then be corrected to the appropriate gas desired based on relative correction [K] factors shown in the gas factor table - see Appendix 2. A reference gas, other than nitrogen, may be used to better approximate the flow characteristics of certain gases.
11
,
NOTE: REMOVAL OF THE FACTORY INSTALLED CALIBRATION SEALS AND/OR ANY ADJUSTMENTS MADE TO THE METER, AS DESCRIBED IN THIS SECTION, WILL VOID ANY CALIBRATION WARRANTY APPLICABLE.
,
Page 16
This practice is recommended when a reference gas is found with thermodynamic properties similar to the actual gas under consideration. The appropriate relative correction factor should be recalculated - see section 9. It is standard practice to
calibrate Mass Flow Meters with dry nitrogen gas at 70
0
F (21.1 FC), 20 psig (1.4 bars) inlet pressure and 0 psig (0 bar) outlet pressure. It is best to calibrate the FMA1700A/1800A transducers to actual operating conditions. Specific gas cali­brations of non-toxic and non-corrosive gases are available at specific conditions.
Please contact your OMEGA7 for a price quotation. It is recommended that a flow calibrator of at least four times better collective accuracy than that of the Mass Flow Meter to be calibrated be used. Equipment required for calibration includes a flow calibration standard and a certified high sensitivity multimeter (which together have a collective accuracy of ±0.25% or better), an insulated (plastic) screwdriver, a flow regulator (example: metering needle valve) installed upstream from the Mass Flow Meter and a pressure regulated source of dry filtered nitro­gen gas (or other suitable reference gas).
The gas and ambient temperature, as well as inlet and outlet pressure conditions should be set up in accordance with actual operating conditions.
12
R52-25%
(IF ENABLED)
R1
J1A (R52)
J1B (R38) J1C (R39)
J1D (R40)
R33-SPAN
(25% or 10%)
R34-ZERO
R38-50%
R39-75% R40-100%
POWER JACK
D-CONNECTOR
MODULAR JACK
INCREASE
DECREASE
FIGURE 7.a - CALIBRATION POTENTIOMETER AND JUMPER LOCATIONS
Page 17
7.2 Calibration of FMA1700A/1800A Mass Flow Meters
All adjustments in this section are made from the outside of the meter, there is no need to disassemble any part of the instrument.
FMA1700A/1800A Mass Flow Meters may be field recalibrated/checked for the same range they were originally factory calibrated for. When linearity adjustment is needed, or flow range changes are being made proceed to step 7.3. Flow range changes may require a different Restrictor Flow Element (RFE). Consult
OMEGA7 for more information.
7.2.1 Connections and Initial Warm Up
At the 9-pin “D” connector of the FMA1700A/1800A transducer, connect the mul­timeter to output pins [2] and [3] for 0-5 VDC (or pins [8] and [9] for 4-20 mA)-(see Figure 2.a).
Power up the Mass Flow Meter for at least 30 minutes prior to commencing the calibration procedure.
7.2.2 ZERO Adjustment
Shut off the flow of gas into the Mass Flow Meter. To ensure that no seepage or leak occurs into the meter, it is good practice to temporarily disconnect the gas source. Using the multimeter and the insulated screwdriver, adjust the ZERO potentiometer
[R34] through the access window for 0 VDC (or 4 mA respectively) at zero flow.
13
CAUTION: The minimum voltage on 0-5 Vdc output can be in the range of 7 to 25 mV. Trying to reduce voltage below this level may increase negative zero shift. This shift may be invisible on devices without LCD display. Stop R34 zero potentiometer adjustment if voltage on 0-5 Vdc output is in the range from 7 to 25 mV and does not decrease any lower.
7.2.3 SPAN Adjustment
Reconnect the gas source. Using the flow regulator, adjust the flow rate to 100% of full scale flow. Check the flow rate indicated against the flow calibrator. If the devia­tion is less than ±10% of full scale reading, correct the SPAN potentiometer [R33] setting by using the insulated screwdriver through the access window, to eliminate any deviation. If the deviation is larger than ±10% of full scale reading, a defective condition may be present.
LIKELY REASONS FOR A MALFUNCTIONING SIGNAL MAY BE:
Occluded or contaminated sensor tube.Leaking condition in the FMA1700A/1800A transducer or the gas line and fittings.For gases other than nitrogen, recheck appropriate “K” factor from the Gas Factor Table.Temperature and/or pressure correction errors.
Page 18
14
See also section 8. TROUBLESHOOTING. If after attempting to remedy the above conditions, a malfunction still persists, return the meter for factory service, see section 1.1.
At this point the calibration is complete. However, it is advisable that several additional points between 0 and 100%, such as 25%, 50%, and 75% flow be checked. If discrepancies are found, proceed to step 7.3 for Linearity Adjustment.
7.3 Linearity Adjustment
7.3.1 Connections and Initial Warm Up
At the 9-pin “D” connector of the FMA1700A & FMA1800A transducer, connect the multimeter to output pins [2] and [3] for 0-5 VDC (or pins [8] and [9] for 4-20 mA)-(see Figure 2.a). If calibration to a new flow range or different gas is being performed, it may be necessary to remove any jumpers at J1.A, J1.B, J1.C, and J1.D before the beginning the linearizing procedure.
Power up the Mass Flow Meter for at least 30 minutes prior to commencing the calibration procedure.
7.3.2 ZERO Adjustment
Shut off the flow of gas into the Mass Flow Meter. To ensure that no seepage or leak occurs into the meter, it is good practice to temporarily disconnect the gas source. Using the multimeter and the insulated screwdriver, adjust the ZERO potentiometer [R34] through the access window for 0 VDC (or 4 mA respectively) at zero flow.
CAUTION: The minimum voltage on 0-5 Vdc output can be in the range of 7 to 25 mV. Trying to reduce voltage below this level may increase negative zero shift. This shift may be invisible on devices without LCD display. Stop R34 zero potentiometer adjustment if voltage on 0-5 Vdc output is in the range from 7 to 25 mV and does not decrease any lower.
7.3.3 25% Flow Adjustment (using R33 potentiometer)
Reconnect the gas source. Using the flow regulator, adjust the flow rate to 25% of full scale flow. Check the flow rate indicated against the flow calibrator.
Adjust the setting for potentiometer [R33] by using the insulated screwdriver through the access window, until the output of the flow meter reads 1.25VDC ±63mV (or 8mA ±0.25mA).
Using the flow regulator, adjust the flow rate until the output of the flow meter reads 0.5 VDC (or 5.6mA). Check the flow rate against the flow calibrator. If the flow rate indicated by the calibrator is within 10% ± 1.5% of F.S. then skip para­graphs 7.3.4, 7.3.5 and proceed directly to paragraph 7.3.6, if not, perform 10% flow adjustment according to paragraph 7.3.4.
Page 19
7.3.4 10% Flow Adjustment
Using the flow regulator, adjust the flow rate to 10% of full scale flow according to the calibrator. Check the flow rate indicated against the flow calibrator. Adjust the setting for potentiometer [R33] by using the insulated screwdriver through the access win­dow, until the output of the flow meter reads 0.5VDC ±63mV (or 5.6mA ±0.25mA).
7.3.5 25% Flow Adjustment (using R52 potentiometer)
Using the flow regulator, adjust the flow rate to 25% of full scale flow according to the calibrator. Check the flow rate indicated against the flow calibrator. The output of the flow meter should read 1.25VDC ±63mV (or 8.0mA ±0.25mA). If the reading is outside of that range, place the jumper at [J1.A] as appropriate to increase or decrease the signal. Adjust the setting for potentiometer [R52] by using the insu­lated screwdriver through the access window, until reading is within specification.
7.3.6 50% Flow Adjustment
Using the flow regulator, increase the flow rate to 50% of full scale flow according to the calibrator. Check the flow rate indicated against the flow calibrator. The out­put of the flow meter should read 2.50VDC ±63mV (or 12mA ±0.25mA). If the read­ing is outside of that range, place the jumper at [J1.B] as appropriate to increase or decrease the signal. Adjust the setting for potentiometer [R38] by using the insu­lated screwdriver through the access window, until reading is within specification.
7.3.7 75% Flow Adjustment
Using the flow regulator, increase the flow rate to 75% of full scale flow according to calibrator. Check the flow rate indicated against the flow calibrator. The output of the flow meter should read 3.75VDC ±63mV (or 16mA ±0.25mA). If the reading is outside of that range, place the jumper at [J1.C] as appropriate to increase or decrease the signal. Adjust the setting for potentiometer [R39] by using the insu­lated screwdriver through the access window, until reading is within specification.
7.3.8 100% Flow Adjustment
Using the flow regulator, increase the flow rate to 100% of full scale flow. Check the flow rate indicated against the flow calibrator. The output of the flow meter should read 5.00VDC ±63mV (or 20mA ±0.25mA).
If the reading is outside of that range, place the jumper at [J3] as appropriate to increase or decrease the signal. Adjust the setting for potentiometer [R40] by using the insulated screwdriver through the access window, until reading is within specification.
Repeat steps 7.3.3 to 7.3.6 at least once more.
7.4 LCD Display Scaling
It may be desirable to re-scale the output reading on the LCD readout supplied with certain model FMA1700A/1800A transducers. Re-calibration for a new flow range or different engineering units are two examples of when this may be necessary.
15
Page 20
16
7.4.2 Adjust Scaling
Using a digital multimeter connected to either the 0 to 5 VDC or 4 to 20 mA sig­nal at the 9-pin “D” connector, set the flow rate on the FMA1700A/1800A to full scale flow (5 VDC or 20mA). Maintain full scale flow, and adjust the potentiome­ter [R3] on the LCD printed circuit board to desired full scale flow reading.
7.4.3 Change Decimal Point
To change the decimal place on the LCD display readout, simply move the jumper to the appropriate location on the 8-pin header block. The numbers are printed to the side of the connections. Do not attempt to place more than one jumper for decimal setting.
8. TROUBLESHOOTING
8.1 Common Conditions
Your Mass Flow Meter was thoroughly checked at numerous quality control points during and after manufacturing and assembly operations. It was calibrated according to your desired flow and pressure conditions for a given gas or a mix­ture of gases.
It was carefully packed to prevent damage during shipment. Should you feel that the instrument is not functioning properly please check for the following common conditions first:
Are all cables connected correctly? Are there any leaks in the installation? Is the power supply correctly selected according to requirements? When several meters are used a power supply with appropriate current rating should be selected.
Were the connector pinouts matched properly? When interchanging with other manufacturers' equipment, cables and connectors must be carefully wired for cor­rect pin configurations. Is the pressure differential across the instrument sufficient?
JUMPER POSITION MAXIMUM SCALABLE DISPLAY READING
“0”
1999
“3” 199.9
“2” 19.99 “1” 1.999
7.4.1 Access LCD Display Circuit
Carefully remove the LCD from the FMA1700A/1800A or panel mounted surface. Remove the aluminum housing on the side of the connection cable. Slide the LCD assembly out of the aluminum housing.
Page 21
17
8.2 Troubleshooting Guide
INDICATION
Lack of reading or output.
Unstable or no zero reading.
Full scale output at “no flow” condition or with valve closed.
Calibration off.
LIKELY REASON
Power supply off
Fuse blown
Filter screen obstructed at inlet.
Occluded sensor tube.
PC board defect.
Power supply problem.
Gas leak.
PC board defective.
Defective sensor.
Gas leak.
Gas metered is not the same as what meter was calibrated for.
Composition of gas changed.
Gas leak.
PC board defective.
RFE dirty.
Occluded sensor tube.
Filter screen obstructed at inlet.
Transducer is not mounted properly.
REMEDY
Check connection of power supply.
Disconnect FMA1700A/1800A transducer from power supply; remove the shorting condition or check polarities; fuse resets automatically.
Flush clean or disassemble to remove impediments or replace.
Flush clean or or return to factory for replacement.
Return to factory for replacement.
Check power supply for appropriate output.
Locate and correct.
Return to factory for replacement.
Return to factory for replacement.
Locate and correct.
Use matched calibration.
See K factor tables in APPENDIX 2.
Locate and correct.
Return to factory for replacement.
Flush clean or disassemble to remove impediments.
Flush clean or return to factory for replacement.
Flush clean or disassemble to remove impediments or replace.
Check for any tilt or change in the mounting of the transducer; generally, units are calibrated for horizontal installation (relative to the sensor tube).
Page 22
18
INDICATION
No zero reading after 15 minute warm up time and no flow condition.
No zero reading after 15 minute warm up time and no flow condition. Display reading does not response on zero adjustment.
No zero reading after 15 minute warm up time and no flow condition. Display reading does not response on zero adjustment. But analog output 0-5 Vdc can be adjusted from 10mV up to 0.5 Vdc with zero potentiometer R34.
LCD Display remains blank when unit is powered up. But flow can be observed on analog output 0-5 Vdc (pis 2 and 3 of the D-connector).
LCD Display remains blank when unit is powered up. No response on the flow from analog output 0-5 Vdc (voltage is less than 15 mV).
LIKELY REASON
Embedded temperature has been changed.
Power supply voltage is less than 11.0 Vdc.
Wire is disconnected inside of the LCD Display.
LCD Display connector is not attached to the FMA1700A/1800A or connection is loose.
Wire is disconnected inside of the LCD Display.
Power supply is bad or polarity is reversed.
PC board is defective.
REMEDY
Readjust ZERO potentiometer R34 through the access hole (see page 13 for details).
Measure voltage on pins [4] and [5] of the 9-pin D-connector. If voltage is less than 11.0Vdc replace power supply with new one (regulated 12.0 to 24.0 Vdc, 250 mA minimum is recommended).
Carefully remove the LCD from FMA1700A/1800A. Remove the aluminum housing on the side of the connection cable. Side the LCD assembly out of the aluminum housing. Check connection for all four wires. If any wire is disconnected restore connection and readjust zero potentiometer R34 to get zero reading on the display (ensure gas source is disconnected and no seepage or leak occurs in to the meter).
Check LCD connector (remove and reinstall LCD connector).
See instructions for pos. 3.
Measure voltage on pins [4] and [5] of the 9-pin D-connector. If voltage is less than 11.0Vdc replace power supply with new one (regulated 12.0 to 24.0 Vdc, 250 mA minimum is recommended). If polarity is reversed (reading is negative) make correct connection.
Return FMA1700A/1800A to factory for repair.
Page 23
19
INDICATION
LCD Display reading does not correspond the correct flow range according analog output 0-5 Vdc signal.
LCD Display reading and analog output 0-5Vdc signal are fluctuate in wide range during the flow measurement.
LCD Display reading does correspond the correct flow range, but 0-5 Vdc output signal does not change (always the same reading or around zero).
LCD Display reading and 0-5 Vdc output voltage do correspond the correct flow range, but 4-20 mA output signal does not change (always the same or reading around 4.0 mA).
Calibration is off (no more than 3.0 % F.S.).
REMEDY
Readjust LCD Display scaling for required full scale flow (see 7.4 on page 15).
Check external connections to pins [2] and [3] of the D-connector. Make sure the load resistance is more than 1000 Ohm.
Return FMA1700A/1800A to factory for repair.
Check external connections to pins [8] and [9] of the D-connector. Make sure the loop resistance is less than 500 Ohm.
Return FMA1700A/1800A to factory for repair.
Shut off the flow of gas into the FMA1700A/1800A (ensure gas source is dis­connected and no seepage or leak occurs in to the meter). Wait for 15 minute. with no flow condition and readjust zero potentiometer R34 to get zero reading on the display.
LIKELY REASON
LCD Display is adjusted for wrong flow range or engineering units.
Output 0-5 Vdc signal (pins [2] and [3] of the D-connector) is shorted on the GND or overloaded.
Output 0-5Vdc schematic is burned out or damaged.
External loop resistance is open or more than 500 Ohm.
Output 4-20 mA schematic is burned out or damaged.
FMA1700A/1800A has initial zero shift.
Page 24
20
INDICATION
LCD Display reading is above maximum flow range and output voltage 0-5 Vdc signal is more than 5.5 Vdc when gas flows through the FMA1700A/1800A.
Gas flows through the FMA1700A/1800A, but LCD Display reading and output voltage 0-5 Vdc signal do not responds on the flow.
Gas does not flow through the FMA1700A/1800A with inlet pressure applied to the inlet fitting. LCD Display reading and output voltage 0-5 Vdc signal show zero flow.
Gas flows through the FMA1700A/1800A, but LCD Display reading is negative and output voltage 0-5 Vdc signal do not responds on the flow (reading near 10mV).
Gas flows through the FMA1700A/1800A, but LCD Display reading is negative and does not change according to gas flow. Output voltage 0-5 Vdc signal corresponds correct gas flow.
REMEDY
Lower the flow through FMA1700A/1800A within calibrated range or shut down the flow completely. The swamping condition will end automatically.
Return FMA1700A/1800A to factory for repair.
Check maximum flow range on transducer's front panel and make required flow adjustment.
Unscrew the inlet compres­sion fitting of the meter and reinstall RFE (see 6.2.2 on page 11). NOTE: Calibration accuracy can be affected.
Return FMA1700A/1800A to factory for repair.
Flush clean or disassemble to remove impediments or replace the filter screen (see 6.2 on page 9). NOTE: Calibration accuracy can be affected.
Check the direction of gas flow as indicated by the arrow on the front of the meter and make required reconnection in the installation.
Locate and correct gas leak in the system. If FMA1700A/1800A has internal leak return it to factory for repair.
Carefully remove the LCD from FMA1700A/1800A. Remove the aluminum housing on the side of the connection cable. Side the LCD assembly out of the aluminum housing. Check connection for green wire (5.00 Vdc reference voltage). If any wire is disconnected restore connection.
LIKELY REASON
Sensor under swamping conditions (flow is more than 10% above maximum flow rate for particular FMA1700A/1800A).
PC board is defective.
The gas flow is too low for particular model of FMA1700A/1800A.
FMA1700A/1800A Series Max Flow 15 L/min models: RFE is not connected prop­erly to the inlet fitting.
Sensor or PC board is defective.
Filter screen obstructed at inlet.
Direction of the gas flow is reversed.
FMA1700A/1800A is installed in the installation with back pressure conditions and gas leak exist in the system.
5.00 Vdc reference voltage (green wire) is disconnected inside of the LCD Display.
Page 25
21
INDICATION
FMA1700A/1800A is disconnected from the source of the gas (no flow conditions) but LCD Display reading is fluctuating in wide range. Output voltage 0-5 Vdc signal also fluctuating. The power supply voltage is 12.0 to
24.0 Vdc and stable.
Reading on the LCD Display ten (hundreds) times less or more than actual gas flow. Output voltage 0-5 Vdc signal corresponds correct gas flow.
REMEDY
Return FMA1700A/1800A to factory for repair.
Carefully remove the LCD from FMA1700A/1800A. Remove the aluminum housing on the side of the connection cable. Side the LCD assembly out of the aluminum housing. Reinstall jumper to the appropriate location on the 8-pin header block (see 7.4.3 on page 16).
LIKELY REASON
Sensor or PC board is defective.
Decimal point jumper is installed in wrong position on the LCD Display Circuit.
For best results it is recommended that instruments are returned to the factory for servicing. See section 1.3 for return procedures.
8.3 Technical Assistance
OMEGA7 Engineering will provide technical assistance over the phone to quali­fied repair personnel. Please call our Flow Department at 800-872-9436 Ext.
2298.
Page 26
Q
O2
= Q
a
= Q
r
X K = 1000 X 0.9926 = 992.6 sccm
where K = relative K factor to reference gas (oxygen to nitrogen)
1
d X C
p
where d = gas density (gram/liter) C
p
= coefficient of specific heat (cal/gram)
Q
a
K
a
Q
r
K
r
where Qa= mass flow rate of an actual gas (sccm) Q
r
= mass flow rate of a reference gas (sccm)
K
a
= K factor of an actual gas
K
r
= K factor of a reference gas
22
=
9. CALIBRATION CONVERSIONS FROM
REFERENCE GASES
The calibration conversion incorporates the K factor. The K factor is derived from gas density and coefficient of specific heat. For diatomic gases:
=
K
=
K
gas
Note in the above relationship that d and Cp are usually chosen at the same con­ditions (standard, normal or other).
If the flow range of a Mass Flow Meter remains unchanged, a relative K factor is used to relate the calibration of the actual gas to the reference gas.
For example, if we want to know the flow rate of oxygen and wish to calibrate with nitrogen at 1000 SCCM, the flow rate of oxygen is:
Page 27
23
APPENDIX 1
COMPONENTS DIAGRAM
FMA1700A/1800A METERING PC BOARD
Page 28
24
APPENDIX 2
GAS FACTOR TABLE (“K” FACTORS)
CAUTION: K-Factors at best are only an approximation. K factors should not
be used in applications that require accuracy better than +/- 5 to 10%.
* Flow rates indicated ( ) is the maximum flow range of the Mass Flow meter being used.
ACTUAL GAS
K FACTOR
Relative to N
2
Cp
[Cal/g]
Density
[g/I]
Acetylene C2H
2
.5829 .4036 1.162 Air 1.0000 .240 1.293 Allene (Propadiene) C3H
4
.4346 .352 1.787 Ammonia NH
3
.7310 .492 .760 *Argon Ar
*Argon AR-1 (>10 L/min)
1.4573
1.205
.1244 .1244
1.782
1.782
Arsine AsH
3
.6735 .1167 3.478 Boron Trichloride BCl
3
.4089 .1279 5.227 Boron Trifluoride BF
3
.5082 .1778 3.025 Bromine Br
2
.8083 .0539 7.130 Boron Tribromide Br
3
.38 .0647 11.18 Bromine PentaTrifluoride BrF
5
.26 .1369 7.803 Bromine Trifluoride BrF
3
.3855 .1161 6.108 Bromotrifluoromethane (Freon-13 B1) CBrF
3
.3697 .1113 6.644
1,3-Butadiene C4H
6
.3224 .3514 2.413
Butane C4H
10
.2631 .4007 2.593
1-Butene C4H
8
.2994 .3648 2.503
2-Butene C4H8 CIS
.324 .336 2.503
2-Butene C4H8TRANS
.291 .374 2.503
*Carbon Dioxide CO
2
*Carbon Dioxide CO2-1 (>10 L/min)
.7382 .658
.2016 .2016
1.964
1.964
Carbon Disulfide CS
2
.6026 .1428 3.397
Carbon Monoxide C0
1.00 .2488 1.250
Carbon Tetrachloride CCl
4
.31 .1655 6.860
Carbon Tetrafluoride (Freon-14)CF
4
.42 .1654 3.926
Carbonyl Fluoride COF
2
.5428 .1710 2.945
Carbonyl Sulfide COS
.6606 .1651 2.680
Chlorine Cl
2
.86 .114 3.163
Chlorine Trifluoride ClF
3
.4016 .1650 4.125
Chlorodifluoromethane (Freon-22)CHClF
2
.4589 .1544 3.858
Chloroform CHCl
3
.3912 .1309 5.326
Chloropentafluoroethane(Freon-115)C2ClF
5
.2418 .164 6.892
Chlorotrifluromethane (Freon-13) CClF
3
.3834 .153 4.660
CyanogenC2N
2
.61 .2613 2.322
CyanogenChloride CICN
.6130 .1739 2.742
Cyclopropane C3H
5
.4584 .3177 1.877
Page 29
25
* Flow rates indicated ( ) is the maximum flow range of the Mass Flow meter being used.
ACTUAL GAS
K FACTOR
Relative to N
2
Cp
[Cal/g]
Density
[g/I]
Deuterium D
2
1.00 1.722 1.799
Diborane B2H
6
.4357 .508 1.235
Dibromodifluoromethane CBr2F
2
.1947 .15 9.362
Dichlorodifluoromethane (Freon-12) CCl2F
2
.3538 .1432 5.395
Dichlofluoromethane (Freon-21) CHCl2F
.4252 .140 4.592
Dichloromethylsilane (CH3)2SiCl
2
.2522 .1882 5.758
Dichlorosilane SiH2Cl
2
.4044 .150 4.506
Dichlorotetrafluoroethane (Freon-114) C2Cl2F
4
.2235 .1604 7.626
1,1-Difluoroethylene (Freon-1132A) C2H2F
2
.4271 .224 2.857
Dimethylamine (CH3)2NH
.3714 .366 2.011
Dimethyl Ether (CH3)2O
.3896 .3414 2.055
2,2-Dimethylpropane C3H
12
.2170 .3914 3.219
Ethane C2H
6
.50 .420 1.342
Ethanol C2H6O
.3918 .3395 2.055
Ethyl Acetylene C4H
6
.3225 .3513 2.413
Ethyl Chloride C2H5Cl
.3891 .244 2.879
Ethylene C2H
4
.60 .365 1.251
Ethylene Oxide C2H4O
.5191 .268 1.965
Fluorine F
2
.9784 .1873 1.695
Fluoroform (Freon-23) CHF
3
.4967 .176 3.127
Freon-11 CCl3F
.3287 .1357 6.129
Freon-12 CCl2F
2
.3538 .1432 5.395
Freon-13 CClF
3
.3834 .153 4.660
Freon-13B1 CBrF
3
.3697 .1113 6.644
Freon-14 CF
4
.4210 .1654 3.926
Freon-21 CHCl2F
.4252 .140 4.592
Freon-22 CHClF
2
.4589 .1544 3.858
Freon-113 CCl2FCClF
2
.2031 .161 8.360
Freon-114 C2Cl2F
4
.2240 .160 7.626
Freon-115 C2ClF
5
.2418 .164 6.892
Freon-C318 C4F
8
.1760 .185 8.397
Germane GeH
4
.5696 .1404 3.418
Germanium Tetrachloride GeCl
4
.2668 .1071 9.565
*Helium He *Helium He-1 (>50 L/min) *Helium He-2 (>10-50 L/min)
1.454
2.43
2.05
1.241
1.241
1.241
.1786 .1786 .1786
Hexafluoroethane C2F6(Freon-116)
.2421 .1834 6.157
Hexane C6H
14
.1792 .3968 3.845
*Hydrogen H2-1 *Hydrogen H
2
-2 (>10-100 L)
*Hydrogen H
2
-3 (>100 L)
1.0106
1.35
1.9
3.419
3.419
3.419
.0899 .0899 .0899
Page 30
26
* Flow rates indicated ( ) is the maximum flow range of the Mass Flow meter being used.
ACTUAL GAS
K FACTOR
Relative to N
2
Cp
[Cal/g]
Density
[g/I]
Hydrogen Bromide HBr
1.000 .0861 3.610
Hydrogen Chloride HCl
1.000 .1912 1.627
Hydrogen Cyanide HCN
.764 .3171 1.206
Hydrogen Fluoride HF
.9998 .3479 .893
Hydrogen Iodide HI
.9987 .0545 5.707
Hydrogen Selenide H2Se
.7893 .1025 3.613
Hydrogen Sulfide H2S
.80 .2397 1.520
Iodine Pentafluoride IF
5
.2492 .1108 9.90
Isobutane CH(CH3)
3
.27 .3872 3.593
Isobutylene C4H
6
.2951 .3701 2.503
Krypton Kr
1.453 .0593 3.739
*Methane CH
4
*Methane CH4-1 (>10 L/min)
.7175 .75
.5328 .5328
.7175 .7175
Methanol CH
3
.5843 .3274 1.429
Methyl Acetylene C3H
4
.4313 .3547 1.787
Methyl Bromide CH3Br
.5835 .1106 4.236
Methyl Chloride CH3Cl
.6299 .1926 2.253
Methyl Fluoride CH3F
.68 .3221 1.518
Methyl Mercaptan CH3SH
.5180 .2459 2.146
Methyl Trichlorosilane (CH3)SiCl
3
.2499 .164 6.669
Molybdenum Hexafluoride MoF
6
.2126 .1373 9.366
Monoethylamine C2H5NH
2
.3512 .387 2.011
Monomethylamine CH3NH
2
.51 .4343 1.386
Neon NE
1.46 .246 .900
Nitric Oxide NO
.990 .2328 1.339
Nitrogen N
2
1.000 .2485 1.25
Nitrogen Dioxide NO
2
.737 .1933 2.052
Nitrogen Trifluoride NF
3
.4802 .1797 3.168
Nitrosyl Chloride NOCl
.6134 .1632 2.920
Nitrous Oxide N2O
.7128 .2088 1.964
Octafluorocyclobutane (Freon-C318) C4F
8
.176 .185 8.397
Oxygen O
2
.9926 .2193 1.427
Oxygen Difluoride OF
2
.6337 .1917 2.406
Ozone
.446 .195 2.144
Pentaborane B5H
9
.2554 .38 2.816
Pentane C5H
12
.2134 .398 3.219
Perchloryl Fluoride ClO3F
.3950 .1514 4.571
Perfluoropropane C3F
8
.174 .197 8.388
Phosgene COCl
2
.4438 .1394 4.418
Phosphine PH
3
.759 .2374 1.517
Page 31
27
ACTUAL GAS
K FACTOR
Relative to N
2
Cp
[Cal/g]
Density
[g/I]
Phosphorous Oxychloride POCl
3
.36 .1324 6.843
Phosphorous Pentafluoride PH
5
.3021 .1610 5.620
Phosphorous Trichloride PCl
3
.30 .1250 6.127
Propane C3H
8
.35 .399 1.967
Propylene C3H
6
.40 .366 1.877
Silane SiH
4
.5982 .3189 1.433
Silicon Tetrachloride SiCl
4
.284 .1270 7.580
Silicon Tetrafluoride SiF
4
.3482 .1691 4.643
Sulfur Dioxide SO
2
.69 .1488 2.858
Sulfur Hexafluoride SF
6
.2635 .1592 6.516
Sulfuryl Fluoride SO2F
2
.3883 .1543 4.562
Tetrafluoroethane (Forane 134A) CF3CH2F
.5096 .127 4.224
Tetrafluorohydrazine N2F
4
.3237 .182 4.64
Trichlorofluoromethane (Freon-11) CCl3F
.3287 .1357 6.129
Trichlorosilane SiHCl
3
.3278 .1380 6.043
1,1,2-Trichloro-1,2,2 Trifluoroethane (Freon-113) CCl
2
FCClF
2
.2031 .161 8.36
Triisobutyl Aluminum (C4H9)AL
.0608 .508 8.848
Titanium Tetrachloride TiCl
4
.2691 .120 8.465
Trichloro Ethylene C2HCl
3
.32 .163 5.95
Trimethylamine (CH3)3N
.2792 .3710 2.639
Tungsten Hexafluoride WF
6
.2541 .0810 13.28
Vinyl Bromide CH2CHBr
.4616 .1241 4.772
Vinyl Chloride CH2CHCl
.48 .12054 2.788
Xenon Xe
1.44 .0378 5.858
Page 32
28
APPENDIX 3
DIMENSIONAL DRAWINGS
FLOW
2.38
0.95
F
G
J
H
A
B
C
C
*
*
*
E
D
E
D
NO LCD
VERSION
DASHED LINE FOR HIGH FLOW UNITS
NOTES: OMEGA7 reserves the right to change designs and dimensions at its sole discretion at any
time without notice. For certified dimensions please contact OMEGA7.
MAXIMUM
FLOW
SERIES
CONNECTION
Compression Fitting
(Except Model
FMA1700A/1800A
Series Max Flow
1000 L/min)
DIMENSION (INCH)
LCD VERSION
NO
LCD
A B C/*C D/*D E/*E F G H
10 L/min
¹/" Tube O Diameter
5.60 1.00 1.00 3.00 5.02 0.69 2.69 4.50
50 L/min
¹/" Tube O Diameter
5.98 1.37 1.25 4.13 6.15 0.69 2.69 4.88
100 L/min
³/" Tube O Diameter
5.98 1.37 1.25 4.13 6.27 0.69 2.69 4.88
200 L/min
³/" Tube O Diameter
6.60 2.00 1.75 6.69 8.83 0.99 4.69 5.50
500 L/min
¹/" Tube O Diameter
7.60 3.00 3.00 7.25 9.67 2.250 6.750 6.50
1000 L/min
³/" NPT Female
8.60 4.00 4.00 7.30 - 3.000 6.800 7.50
Page 33
29
PARTS OF THE FLOW METER
LEFT AND RIGHT VIEWS
FRONT VIEW
Page 34
NOTES:
30
Page 35
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a trademark of OMEGA ENGINEERING, INC. © Copyright 2018 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the
COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.
Page 36
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
TEMPERATURE
MU
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
MU
Wire: Thermocouple, RTD & Thermistor
MU
Calibrators & Ice Point References
MU
Recorders, Controllers & Process Monitors
MU
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
MU
Transducers & Strain Gages
MU
Load Cells & Pressure Gages
MU
Displacement Transducers
MU
Instrumentation & Accessories
FLOW/LEVEL
MU
Rotameters, Gas Mass Flowmeters & Flow Computers
MU
Air Velocity Indicators
MU
Turbine/Paddlewheel Systems
MU
Totalizers & Batch Controllers
pH/CONDUCTIVITY
MU
pH Electrodes, Testers & Accessories
MU
Benchtop/Laboratory Meters
MU
Controllers, Calibrators, Simulators & Pumps
MU
Industrial pH & Conductivity Equipment
DATA ACQUISITION
MU
Communications-Based Acquisition Systems
MU
Data Logging Systems
MU
Wireless Sensors, Transmitters, & Receivers
MU
Signal Conditioners
MU
Data Acquisition Software
HEATERS
MU
Heating Cable
MU
Cartridge & Strip Heaters
MU
Immersion & Band Heaters
MU
Flexible Heaters
MU
Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
MU
Metering & Control Instrumentation
MU
Refractometers
MU
Pumps & Tubing
MU
Air, Soil & Water Monitors
MU
Industrial Water & Wastewater Treatment
MU
pH, Conductivity & Dissolved Oxygen Instruments
M5283/0918
Loading...