Omega FHG1000, FHG User guide

Page 1
omega.com
For latest product manuals:
omegamanual.info
Shop online at
User’s Guide
FHG FLOW METERS
Page 2
Servicing North America:
ISO 9001 Certified Stamford, CT 06907-0047 USA
Toll Free: 1-800-826-6342 TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: [email protected]
Canada: 976 Bergar
Laval (Quebec), Canada H7L 5A1 Toll-Free: 1-800-826-6342 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: [email protected]
For immediate technical or application assistance:
U.S.A. and Canada: Sales Service: 1-800-826-6342/1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378/1-800-622-BEST
®
Engineering Service: 1-800-872-9436/1-800-USA-WHEN
®
Mexico: En Español: 001 (203) 359-7803 FAX: (001) 203-359-7807
Servicing Europe:
Benelux: Managed by the United Kingdom Office
Toll-Free: 0800 099 3344 TEL: +31 20 347 21 21 FAX: +31 20 643 46 43 e-mail: [email protected]
Czech Republic: Frystatska 184
733 01 Karviná, Czech Republic Toll-Free: 0800-1-66342 TEL: +420-59-6311899 FAX: +420-59-6311114 e-mail: [email protected]
France: Managed by the United Kingdom Office
Toll-Free: 0800 466 342 TEL: +33 (0) 161 37 29 00 FAX: +33 (0) 130 57 54 27 e-mail: [email protected]
Germany/Austria: Daimlerstrasse 26
D-75392 Deckenpfronn, Germany Toll-Free: 0 800 6397678 TEL: +49 (0) 7059 9398-0 FAX: +49 (0) 7056 9398-29 e-mail: [email protected]
United Kingdom: OMEGA Engineering Ltd.
ISO 9001 Certified One Omega Drive, River Bend Technology Centre, Northbank
Irlam, Manchester M44 5BD England Toll-Free: 0800-488-488 TEL: +44 (0)161 777-6611 FAX: +44 (0)161 777-6622 e-mail: [email protected]
OMEGAnet®Online Service Internet e-mail
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, human applications.
Page 3
OPERATING INSTRUCTIONS For FHG Series Flow Meters
Page 4
Table of Contents
Page
Important Basic Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1. Function Description of the FHG Flow Meters .................................... 3
2. General Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3. FHG Flow Meter Selection ................................................... 4
4. Declaration of Conformity .................................................... 4
5. General Operating Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
6. Maximum Operating Pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
7. Information about the EU Pressure Equipment Directive 97/23/EC ................... 5
8. Flow Rate Measuring Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
9. Mounting the Flow Meter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
10. Cleaning and Flushing of Pipeline before Initial Start-Up ..........................6
11. Fluid Filtering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
12. Sensor Electronics Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
13. Pulse Filtering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
14. Programming the Preamplier Electronics .....................................11
15. Signaling LEDs ...........................................................12
16. Operating Mode Messages ................................................12
17. Alarm and Warning Messages ..............................................13
18. Preamplier Technical Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
19. Preamplier Pin Assignment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
20. Maintenance ............................................................14
21. Returning for Repairs and Sample Devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
22. FHG Flow Meter Technical Data ............................................15
23. FHG Flow Meter Flow Characteristics ........................................16
24. FHG Flow Meter Dimensions ...............................................17
25. Pin Assignment ...........................................................18
26. Connection Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
The current publication of this operating instruction supersedes all information from previous publications. Omega reserves the right to make changes and substitutions. Omega is not liable for any printing errors. Reproduction, including excerpts, is permitted only after written appro ­val by Omega. Omega reserves the right to modify technical data at any time. Last revised: 12/2012
2
Page 5
Important basic information
These installation and operating instructions should provide you with the information you need to properly install and commission
the ow meter. Installation, commissioning and testing are to be performed by trained and qualied personnel only. These operating
instructions must be read and applied carefully to ensure proper, trouble-free and safe operation of the ow meter. Omega is not
liable for any damage incurred resulting from not complying with the instructions in this operating instruction. It is not permitted in
any case to open the device.
1. Function Description of the FHG Flow Meters
FHG ow meters measure the ow rate based on the screw pump principle. A pair of rotors tted precisely into the housing constitutes the measuring element. An integrated gear and non-contact signal pick-up system detects the rotations of the measuring element and converts them to digital pulses.
Together with the housing walls, the rotor edges form closed measuring chambers in which the uid is transported from the inlet to the outlet side.
The uid volume put through within one main rotor rotation is the rotation volume, which is divided by the sensing gear and digitised, processed and output in the sensor module.
Advantages
• High degree of precision that is mostly independent of viscosity
• Pulsation-free measurement
• Lowest pressure losses
•
Short response time due to innovative rotor prole and
• Highest functionality due to intelligent sensor technology
• Gentle uid measurements
Sensor System Explanation
The non-contact pick-up system consists of two GMR bridges (sin /cos), which are located in a sensor unit in cartridge design. It detects the movement of the sensing gear and routes the sin/cos signals to the preamplier electronics.
The preamplier electronics digitise and amplify the sensor signals and mul­tiply them by a high-resolution interpolator using adjustable settings. The square wave signals are bidirectional and can be utilised by any evaluating instrument as well as computers and PLC controls.
The resolution is selectable in steps from factor 1 to 128.
In case of a 1-channel evaluation, a separate directional signal is available.
An adjustable pulse lter can offset and suppress negative ows (e.g. gene­rated by vibrations) while still in the device.
The frequency of the output signals is proportional to the ow (volume ow) and depends on the respective ow meter size. The frequency range is from 0 to 100 kHz. The preamplier is protected against reverse polarity and incorrect connection. It is designed for media temperatures of -30°C to +120°C and is mounted directly on the FHG ow meter.
3
Page 6
2. General Description
Please follow all instructions in this manual to ensure the trouble-free ope­ration of the FHG ow meters. Omega does not assume responsibility or liability for damages resulting from noncompliance with these instructions.
3. FHG Flow Meter Selection
The device may only be opened within the warranty period after consulta­tion and approval by Omega.
For the trouble-free, safe, and reliable operation of the ow meters, se­lecting the correct type and size is critical. Because of the wide variety of applications and ow meter designs, the technical data in the Omega cata­log are general in nature. Certain properties of the devices are dependent
4. Declaration of Conformity
Flow meters of the ”FHG“ series have been tested for their electromagnetic compatibility and interference emissions as outlined by the EMC Directive and are in compliance with the applicable statutory EMC Directives.
They cannot be operated independently, are connected by cable to a po­wer source, and provide digital electrical signals for electronic evaluation. All ow meters have a declaration of conformity, which can be requested if necessary.
5. General Operating Requirements
Before assembly, commissioning or operation, check and verify the following properties & aspects of the respective circumstances of your system to ensure operation is trouble-free, safe, and reliable.
on type, size and measuring range, as well as the liquid to be measured. Please contact Omega or one of our sales and service representatives for detailed information about the appropriate ow meter for your particular application.
Since the electromagnetic compatibility of the entire measuring sys­tem is also dependent on the installation of the cables, the correct connection of the shield, and each individual connected device, all components must comply with the EMC Directive, and the electroma­gnetic compatibility of the entire system, machine, or system must be ensured as well.
All ow meters have been tested in accordance with the applicable statuto­ry EMC Directives of EN 61000-6 and are CE certied. The EC conformity marking is the CE mark afxed to all ow meters.
1. The Fluid to be Processed
Are the flow meter seals and materials compatible with for the fluid and any cleaning agents that will be used?  Is the uid viscous or abrasive? Have you properly sized the meter and selected appropriate bearing materials?  Is the uid dirty or does it contain contaminants/pollutants and solid particles that may require ltration?  Does the uid have llers or other additives? What grain sizes do these solids have, and could they block the measuring element?  Is it necessary to install an upstream hydraulic lter (see ltering requirements in section 22 of this document)?  Are tubes and pipes clean and free of assembly residues such as chips, weld spatter?  Is the tank clean and is it impossible for impurities or foreign substances to reach the pipeline or tubing system from the tank?  Is a different uid used frequently and is the system sufciently ushed and rinsed in between?  Are pipelines/tubes and the entire system completely deaerated (the system should be slowly lled with uid before operation at full ow
to avoid hydraulic shock on the mechanical components)?
2. Hydraulic Properties of the System
 Is the max. operating pressure of the system less than the max. permissible operating pressure of the ow meter?  Is the max. pressure drop ∆p (at ow meter) below the max. permissible pressure drop?  Is the pressure drop ∆p not excessive with max. ow (e.g. high viscosity)?  Does the ow range of the ow meter (dependent on the viscosity) correspond with the present ow?  Please note that the ow range is less with higher viscosity!  Does the temperature range of the ow meter correspond with the present max. temperature of the uid?  Is the cross-section of the pipelines/tubes large enough and are there no overly large pressure drops in the system?  Is the hydraulic connection (inlet/outlet) connected corrected and sealed properly?  Note: A blocked ow meter can stop t he entire ow. Does the sy stem feature an overpr essure / by pass valve? T his valve must be chec ked and
maintained at regular intervals.
4
Page 7
3. Electronic Evaluation and Electrical Safety
Does the supply voltage of the ow meter match the available power supply?  Is the supply voltage to the power supply adapter or the evaluating device sufciently ltered?  Does the output of the supply voltage correspond with the required output?  Is the electrical connection established based on the enclosed wiring plan (see page 19)?  Is the cable shield correctly grounded on both sides to a clean common ground (PE)?  Is there a potential difference between the ground on the ow meter and at the evaluating device?  Is the ow meter permanently grounded (PE) (e.g. via the pipelines)?  If the measuring element of the flow meter is insulated from ground (PE), the meter must be grounded with a cable!!  Is the 4-pin to 5-pin round pin plug of the connection cable rmly attached to the plug of the ow meter?  Are the wires at the evaluating device connected correctly?  Does the entire system meet the legal rules and regulations concerning electromagnetic compatibility (EMC)?  Is compliance with all local rule and regulations, applicable rules, guidelines and basic conditions of the EMC ensured?  Syste ms where a malf unction or fa ilure may lead to p ersonal inju ries must be equ ipped with suitable safety mechanisms. The funct ion of these
safety mechanisms must be checked at regular intervals.
6. Maximum Operating Pressure
Before installing the ow meter, you must check whether the max. operating pressure of the system does not exceed the max. permissible operating
pressure of 450 bar of the ow meter. Make sure to keep in mind that peak pressures may occur when operating the system.
Important: Please contact Omega with all operating pressures > 450 bar and in case of special models.
7. Information about the EU Pressure Equipment Directive 97/23/EC
Omega ow meters of the ”FHG“ series qualify as “pressure equipment“ as dened by Section 1, Paragraph 2.1.4. of the directive listed above and as such are affected by the regulations of this directive.
Omega ow meters must therefore meet the technical requirements speci­ed in Section 3, Paragraph 1.4 of the directive. The uids to be measured are for the most part Group 2 uids acc. to Section 9, Paragraph 2.2. Omega ow meters do not reach the limit values specied by Section 3, Paragraph 1.1. The technical requirements for Omega ow meters are
therefore conned to the criteria specied in Section 3, Paragraph 3. This means that the devices must be designed and manufactured in accordance with the provisions of good engineering practice applicable in a member state. This is hereby conrmed. The section also stipulates that such pres­sure equipment and components or accessories are not allowed to bear the CE marking in accordance with the Pressure Equipment Directive. This means that a declaration of conformity is not issued for Omega ow meters and the devices are not provided with the CE mark as pertaining to Direc­tive 97/23/EC.
8. Flow Rate Measuring Range
The ow rate measuring range specied in the data sheet (Q of the ow meter refers to the test uid ‚hydraulic oil‘ with a viscosity of 21 m m2/s at a temperature of 20°C. For this measuring range, Omega species accuracy up to 0.3% of the measured value and a repeatability of 0.05%.
min
– Q
)
In uids with low viscosity (< 21 mm²/s), the measurement accuracy de-
max
grades while it may improve with uid with a high viscosity (> 21 mm²/s). Note also that the ow measuring range is limited at higher viscosity (see data sheet of the ow meter). The characteristic pressure loss curves are listed in Section 23.
Important: Verify that the specied maximum permissible operating pressure of the ow meter can never be exceeded in any operating mode of the system. Also pay attention to the ow measuring range, which is dependent on the viscosity of the uid to be measured.
5
Page 8
9. Mounting the Flow Meter
The ow meter should be mounted in an easily accessible location so that disassembly to clean the measuring elements is easy. Since ow meters operate in any installation position and ow direction, you can mount it anywhere in your system. When installing the ow meter, make sure that liquid remains in the ow meter even at standstill of the system and that the ow meter can never run dry. The outlet of the ow meter should always have a certain backpressure since this xes the measuring element of the ow meter in the liquid column (the measuring element uses to support itself on the liquid column) and the pipeline cannot empty itself. In critical cases, or if the pipeline can run empty in standstill or standby mode, it is always advisable to install an additional non-return check valve in the outlet line.
Fig. 1: Flow meter with backpressure
Flow meter
Important: Make sure that the ow meter measuring elements are always completely lled both in inow and outow and that the outow has a little backpressure. This prevents the measuring elements from being damaged by a sudden and steep increase of ow and at the same time improves measurement accuracy.
Non return valve
Tank
Flow meters of the ”FHG“ series can be installed in the pipeline. Always select large cross-sections (if possible) for the hydraulic inlet and outlet or the entire pipeline system. This reduces the pressure drop and the ow rate throughout the system.
Installation Notes
Installation Position
Any, note arrow indicating preferred direction if necessary (calibration ar­row).Mount the device in such a way that the preamplier is turned away from any potential heat source. Straight pipe sections are not required in inlet/outlet.
Connecting Units
If the connecting units (mounting anges) are to be installed on-site, compli­ance with the specied torque is required.
Pipe Thread
Please comply with the screw-in depths and sealing systems. PTFE tape or liquid sealants such as adhesives are not permitted!
Fastening
The devices must be installed stress-free into the pipeline. This is accomplis­hed with fastening screws located at the face sides in the connecting units. For stress-free assembly, the compressive strength may be limited!
Table 1: Starting torque of the connection units
FHG Flow Meter Size Torque
FH G 1xx2 70 Nm
FH G 1xx4 120 Nm
FH G 1xx5 240 Nm
FH G 1xx7 160 Nm
10. Cleaning and Flushing of Pipeline before Initial Start-Up
Before initial start-up of the ow meter, you must ush and clean the whole system to prevent contaminants from reaching the measuring elements duri­ng the assembly and installation. Foreign matter or contaminants may block the ow meter or severely damage it so that the ow meter readings are no longer valid and the device must be returned for repairs. After completion of the installation or piping, you must rst ush the entire pipeline system and carefully clean and ush the tank. This requires that the ow sensor is removed from the uid circuit to ush out all foreign matter or contaminants (e.g. chips, metal parts) without problems. Use a rinsing uid that is compa­tible with the subsequent used uid and will not cause adverse reactions. Such information can be obtained from the supplier or manufacturer of the uid or from Omega.
Flow meters are sensors manufactured with a high degree of precision. They have mechanical measuring elements consisting of two rotors tted into the housing with narrow gaps. Even the smallest damage to the rotors causes a measuring error. Always make sure that foreign matter or conta­minants cannot reach the measuring elements and that the uid owing through the ow meter is always free of pollutants and particles. Once the system is thoroughly ushed and no extraneous material is in the piping sys­tem, you can mount the ow meter into the uid circuit and start the actual initial startup process.
6
Page 9
11. Fluid Filtering
Heavily contaminated uids or foreign matter in the uid can block, da­mage, and even destroy the ow meter. In these cases, always install a sufciently large lter in front of the ow meter so that foreign particles and
Table 2: Upstream lters
Flow meter size Filter size for
ball bearing
FH G 1xx2 250 µm
FH G 1xx4 250 µm
FH G 1xx5 500 µm
FH G 1xx7 500 µm
solids are prevented from entering the measuring elements, thus preventing damage to the ow meter. The required ltering depends on the size, bea­ring, and design of the ow meter.
The lter size for ow meters with slide bearings, in special designs, or with specially adapted measuring element tolerances can be obtained from Omega upon request.
Important: A blocked ow meter is capable of stopping the entire ow. An overpressure / bypass valve must be installed in the system side.
12. Sensor Electronics Function
The liquid to be measured ows through the rotor chambers in axial direc­tion, resulting in an even rotation of the screw spindles.
This is done especially gentle and with very low resistance for the uid to be measured as well as pulsation-free and almost free of leaks due to the specially designed uidic prole geometry.
A magnet wheel permanently afxed to the rotors is scanned without contact with a sensor module. The non-contact pick-up system consists of two GMR bridges (sin /cos), which are located in a sensor unit in cartridge design. It detects every movement of the sensing gear and routes the sin/ cos signals to the preamplier electronics. The preamplier electronics di­gitise and amplify the sensor signals and multiply them by a high-resolution interpolator using adjustable settings. The square wave signals phase-shif­ted by 90° are bidirectional and can be utilised by any evaluating device as well as computers and PLC controls.
The ow is proportional to the edges/pulse count and the ow rate is pro­portional to the frequency. The adjustable interpolator can be used to ad­just the resolution explicitly to the downstream connected evaluating unit for obtaining highly precise measuring results of the entire system. This ap­plies to the following application cases, for example:
• Measuring, controlling, and regulating high viscosity uids
• Measuring, controlling, and regulating in lower ow ranges
• Measuring, controlling, and regulating when passing through zero
• Measuring, controlling, and regulating in both ow directions
• Measuring, controlling, metering, and lling of small volumes
The resolution is selectable in steps from factor 1 to 128. The frequency range is from 0 to 100 kHz.
In case of a 1-channel evaluation, a separate directional signal is availa­ble.
The preamplier is protected against reverse polarity and incorrect con­nection. It is designed for uid temperatures of -30°C up to +120°C and is mounted directly on the FHG ow meter.
The uid volume passed through by one gear division of the sensing wheel within the measuring element is divided by the set interpolation fac­tor. This forms the measurement volume per pulse (Vm) with the dened unit [c m ³/pulse]. The frequency of the output signals can be calculated as follows:
Formula 1: Calculation of the output frequency with Q in l/min
Q x 1000
f= Vm 60
Table 2, Formula 2, and the subsequent diagrams can be used to determine the corresponding resolution or the corresponding IPF for the respective application.
Adjustable interpolation factors IPF: 1, 2, 5, 10, 25, 32, 50, 64, 100, 128
7
Page 10
Table 3: Measurement volumes and K-factors
FHG 100
Interpolation factor (IPF)
Switch position S3
Measurement
volume V
m
[cm3/Imp]
K-Factor
[Imp/l]
K-Factor
[Imp/
gal.]
1 0 0, 5815 172 0 6510
2 1 0,29075 3439 130 20
5 2 0,11630 8598 32549
10 3 0, 05815 17197 65098
25 4 0,02326 42992 162 74 5
32 5 0,01817 55036 208335
50 6 0, 01163 85985 325489
64 7 0,00909 110 0 11 416 4 4 0
100 8 0,00582 171821 650419
128 9 0,00454 220264 833797
FHG 800
Interpolation factor (IPF)
Switch position S3
Measurement
volume V
m
[cm3/Imp]
K-Factor
[Imp/l]
K-Factor
[Imp/
gal.]
1 0 10,00000 100 379
2 1 5,00000 200 757
5 2 2,00000 500 1893
10 3 1,00000 1000 3785
25 4 0,40000 2500 9464
32 5 0,31200 3200 1 2113
50 6 0,20000 5000 18927
64 7 0,15625 6400 24227
100 8 0,10000 10000 37854
128 9 0,07813 12 799 484 51
FHG 400
Interpolation factor (IPF)
Switch position S3
Measurement
volume V
m
[cm3/Imp]
K-Factor
[Imp/l]
K-Factor
[Imp/
gal.]
1 0 3,138 319 120 6
2 1 1,569 637 2413
5 2 0,6276 1593 6032
10 3 0 , 3138 3187 120 63
25 4 0,12552 7967 30158
32 5 0,09806 10198 38603
50 6 0,06276 15934 60 316
64 7 0,04903 20396 77207
100 8 0,03138 318 67 120632
128 9 0,02452 40783 15 438 2
FHG 2500
Interpolation factor (IPF)
Switch position S3
Measurement
volume V
m
[cm3/Imp]
K-Factor
[Imp/l]
K-Factor
[Imp/
gal.]
1 0 37,00000 27 102
2 1 18,50000 54 204
5 2 7,40000 135 511
10 3 3,70000 270 1022
25 4 1,48000 675 2555
32 5 1,15 62 5 864 3270
50 6 0,74000 1350 511 0
64 7 0, 57813 1728 6540
100 8 0,37000 2700 10220
128 9 0,28906 3456 13081
Formula 2: Calculating the max. IPF
f
IPF ≈
max
Q
x Vm
x 1000
max
IPF1
x 60
The set IPF may not be larger than the calculated IPF!
IPF Interpolation factor f
Max. processable input frequency
max
V
Measurement volume with IPF = 1 (volume of a gear structure of the sensing wheel)
mIPF1
Q
Max. operating ow in l/min
max
8
Page 11
Flow diagrams vs frequency
FH G 1xx2
FLOW vs FREQUENCY
FHG 1xx4
FLOW vs FREQUENCY
9
Page 12
FH G 1xx5
FLOW vs FREQUENCY
FH G 1xx7
FLOW vs FREQUENCY
Example Flow meter: FHG 1xx4 Max processable input frequency of the downstream evaluating unit: 20 kHz Max. operating ow: 140 l/min Path 1: The diagram yields an IPf of 25
Path 2:
f
IPF
max
≈
Q
10
x Vm
x 1000
max
IPF1
x 60
20.000 x 3,138 ml x 60 s
= 140 1000 ml
1
s
26,9 ≈ 25
=
Page 13
13. Pulse Filtering
Oscillations in uid systems manifest themselves through constant forward and backward movements of the liquid column, which is also detected by the rotor sensors and converted into proportional electronic pulses or edge se­quences. Depending on the application, oscillations or vibrations can occur during the ow rest phases or discontinuous ows. The pulses generated du­ring the osciallation phase can be incorrectly interpreted by the downstream evaluating unit or controller, which can be very distracting for the respective operating process.
The signal ltering function of the internal electronics continuously offsets these generated edges during the rapid forward and backward movements
of the rotor measuring unit. The signals at the channel outputs are also sup ­pressed at the same time until the internal offset is equalized or the initial position of the rotor measuring unit has been reached again (see Fig. 3).
The user is able to set the degree of ltering in the form of partial volumes using rotary coding switches.
Adjustable pulse ltering: 0Z, 0.25Z, 0.5Z, 0.75Z, 1.0Z, 1.25Z, 1.5Z, 1.75Z,
2.0Z, 2.25Z, 2.5Z, 2.75, 3.0Z, 3.25Z, 3.5Z, 3.75Z (Z: gear unit)
Fig. 3: Pulse ltering principle
Table 4: Suppressed volume with pulse ltering activation [ml]
Filter position
0 0 0 0 0
1 0.14 5375 0.7845 2.5 9.25
2 0.29075 1.569 5.0 18.50
3 0.436120 2.3535 7. 5 27. 7 5
4 0. 5 815 3.138 10.0 37.0 0
5 0.726875 3.9225 12 . 5 46.25
6 0.87225 4 .707 15.0 55.5
7 1.017625 5.4915 17. 5 64 .75
8 1.163 6. 276 20.0 74 . 0 0
9 1.308375 7.0605 22.5 83.25
10 (A) 1.45375 7.8 4 5 25.0 92.50
11 ( B ) 1.599120 8.6295 2 7. 5 101.75
12 (C) 1. 74 45 9. 414 30.0 111. 0 0
13 ( D) 1.889875 10.1985 32.5 120 .25
14 (E) 2.03525 10.983 35.0 12 9. 50
15 ( F) 2.180625 11. 7675 3 7. 5 138 . 75
FHG 1xx2 FHG 1xx4 FHG 1xx5 F HG 1x x 7
148
14. Programming the Preamplier Electronics
The electronics elements are quickly and easily set. There are two rotary coding switches on the electronics (S3, S4), a jumper (B2), a switch (S2) and a key (S2). With the rotary coding switches, the IPF and the degree of ltering are programmed.
B2
S1
S2
S4
S3
Fig. 4: Preamplier electronics
11
Page 14
During initial startup, the switch S2 must rst be set to the corresponding preferred direction of the ow. The positive ow direction of the FHG ow meter system is specied in the top view of the 5-pin M12 connector. In this case, the switch S1 must be set to ON. For the opposite negative direction, the switch position is to be down and thus set to OFF. This setting ensures that the pulse ltering is activated in the right direction from the very begin­ning after switching on the supply voltage.
Pin 5 of the M12 connector is either used for the separate direction signal or an error signal. This is set accordingly with the bridge B2. The gure above depicts the bridge attached to the middle and right pin of the 3-pin row of pins, which routes the separate zero signal to the third output. If the bridge is on the left and middle pin, the error signal in case of a fault is output. A description of the error states is found in the ”Alarm and Warning Messages“ chapter.
Ten different interpolation factors can be set with the coding switch S3. The corresponding interpolation factors for the respective switch positions are listed in Table 3. This setting can be changed at any time while the system is running. Simply use a small screwdriver to adjust the rotary coding switch and then briey press the S2 key for the acknowledgment. The new pulse rate is enabled a once.
The rotary coding switch for the pulse ltering has 16 switch settings. The degree of ltering is determined with quarter gear division increments. The corresponding suppressed partial volumes of the respective size are listed in Table 4. Changes can also be performed during operation and become active after pressing the S2 key.
The electronics is sensitive to electrostatic discharges. People making adjustments to the electronics must rst discharge their electrostatic charges using a grounded object.
Important: People making adjustments to the electronics must rst discharge their electrostatic charges using a grounded object.
15. Signaling LEDs
The signa ling LEDs provide in formation ab out the corresp onding status o f the electronics. These include certain operating and fault states (see Figure 5).
The three LEDs have a different combination of states for each signal. The LEDs signal either operating modes or alarms and warnings. Operating mode messages signal the respective mode that has been set. Alarm and warning signals provide explicit information about overload, conditions that can negatively affect the measurement, or component er­rors of the measuring system.
Fig. 5: Signaling LEDs of the preamplier electronics
16. Operating Mode Messages
Table 5: Operating mode messages
Mode Yellow
LED
Normal operation off on off off
Offset mode off Flashes off off
Green
LED
Red LED
Error output
GreenYellow Red
12
Page 15
17. Alarm and Warning Messages
The electronics of the FHG ow meters can detect ve events that could lead to measurement errors. In case of serious errors, the third output has a ”high“ signal or a ”pulse“ signal if activated with the bridge B1. The different error causes can be determined with the states of the three LEDs.
Table 6: Alarm and warning messages
Warning Yellow
Offset adjustment necessary 1
Alarm Yellow
Interpolator electronics errors 2
Error at pick-up 3 off o n / off on on
Flow overload on off on on
max. frequency range exceeded (>100 kHz) 4
Fluid temperature > 120°C 5
Flashes on off off
Flashes off Flashes Pulse
Flashes on Flashes Pulse
Green
LED
LED
on on Flashes Pulse
LED
Gr een
LED
Red LED
Red
LED
Error output
Error output
The red LED is linked with the error output. Each active state of this LED or the error output signals an event that has negative effects on the measure­ments.
Description of the Error Messages
1. Offset adjustment necessary: The sensor and/or the preamplier elec­ tronics were replaced. A different size was set.
2. Electronics errors: Defective component in interpolator circuit, unable to determine internal conguration values
3. Pick-up errors: The sensor is defective or quit working. The distance bet­ ween the sensor and the magnet wheel has changed = mechanical damage
4. Overload: The maximum permissible ow range was exceeded
5. Frequency errors: The max. output frequency of 100,000 Hz was excee­ ded. The IPF is dimensioned too high for the respective ow
6. Temperature errors: The temperature of the uid is too high (> 120°C) and may result in awed or incorrect measurements
18. Preamplier Technical Data
Scanning sensor 2 x GMR sensors in a bridge circuit (sin/cos)
Adjustment automatic offset adjustment
Resolution programmable 1, 2, 5, 10, 25, 32, 50, 64, 100, 128
Adjustable pulse ltering 0Z, 0.25Z, 0.5Z, 0.75Z, 1.0Z, 1.25Z, 1.5Z, 1.75Z, 2.0Z, 2.25Z, 2.5Z, 2.75, 3.0Z, 3.25Z, 3.5Z, 3.75Z (Z: gear unit)
Frequency up to 100 kHz
Output signals Channel A, channel B, directional signal DIREC (high: positive, low: negative) or error signal ERROR
(high or pulse: error)
Channel A and B two signal outputs for outputting the digital ow sensor signals, a channel offset of 90° between channel A and channel B;
Flow direction Detection of ow direction from the channel offset of the signals from channel A to channel B or via the separate direc- tional signal.
Outputs Three current-limited and short-circuit-proof power ampliers (channel A, channel B, DIREC / ERROR); integrated adjustment to a characteristic impedance of 75 Ω; driver current approx. 300 mA at supply of 24 V; small saturation voltage of up to 30 mA load current, short switching times, integrated freewheeling diodes against Vb and GND, ther­ mal shutdown with hysteresis; in case of error, the outputs are high impedance;
The 24 V line drivers are designed for control applications with cable adjustment
Error messages Electronics fault (e.g. faulty interpolator), sensor errors (e.g. sensor failure), offset adjustment necessary, overload (ow peaks), frequency error (> 100 kHz), temperature error (> 120°C)
Operating voltage Vb = 10 … 28 VDC
Current consumption l
= approx. 40 mA, total current consumption depends on output load
noload
13
Page 16
19. Preamplier Technical Data
Fig. 6 depicts the pin assignment of the preamplier. This plug has ve pins. Two pins are for the power supply (pin 1 and 3), two for the signal output of channel 1, 2 (pins 2 and 4) and a separate output for error or direction detection (pin 5).
However, please note that the cable shield at the plug side is on the metal housing of the plug. The connection cable shield must be applied on both sides. The shield is used to connect the PE from the evaluation electronics to the preamplier housing and the measuring element of the ow meter. The cable shield should always be continuous to the ow meter and not sepa-
rated by distribution boxes or junction boxes. Route the connection cable as directly as possible from the evaluating device to the ow meter since interruptions are always potential sources of error. The measuring elements of the ow meter must be connected electrically with the protective earth conductor (PE). This is generally ensured with the grounded pipelines.
If potential differences exist between the preamplier housing and the protective conductor (PE) of the evaluating device, you must pro­vide equalization.
Pin 2
Digital signal
Channel 1
Pin 5
Digital signal
ERROR/DIREC
Pin 3
Power supply
GND (-Vb = 0 V)
Plug top view
Figure 6: Flange plug of the preamplier housing
Pin 1 Power supply Vb = 10-28 VDC
Metal housing connected with shield and protective earthing conductor PE
Pin 4 Digital signal Channel 2
Important: Use only well shielded connection cables with a wire cross-section of ≥ 4 to 5 x 0.25 mm². Please note that the housing of the round pin plug is metallic, has a connection for the shield and that the potential of the PE is connected to the cable shield and the housing of the preamplier.
Important: Please make sure that no additional inductors such as contactors, relays, valves, etc. are connected to the power supply of the ow meter. These components are potential sources of interference (especially if the inductors are not provided with adequate protective circuits), produce high interference pulses during the switching, and may disrupt the function of the ow meter even though it complies with the EMC directives.
20. Maintenance
Depending on the operating conditions, the service life and thus the specic characteristics of the equipment are limited due to wear, corrosion, deposits, or aging. The operator is responsible for carrying out periodic inspections, maintenance, and re-calibrations. Each observation of a malfunction or damage makes it necessary to stop operation. We can loan a device for the duration of the overhaul if requested. We recommend an annual inspection and recalibration.
14
Page 17
21. Returning for Repairs and Sample Devices
Repairs on the ow meter and other components can be carried out quickly and efciently only if you include detailed information about the claim or defect when returning the device. In addition, a safety sheet must be enc­losed, clearly indicating what uid has been used with the ow meter and how hazardous this uid is.
Compliance with the laws on occupational safety, such as Workplace Re­gulations (ArbStättV), Accident Prevention Regulations and Regulations on Environmental Protection, Waste Law (AbfG) and Water Act (WHG), re­quire that businesses protect their employees and other people as well as the environment from harmful effects when handling hazardous substances.
22. FHG Flow Meter Technical Data
If additional precautions are required despite careful draining and clea­ning of the ow meter, the associated required information must be inclu­ded when returning the device. Please note that inspection and repair is only performed on ow meters returned to Omega if the safety sheet of the used uid is enclosed and if the ow meter has been completely cleaned and ushed. This is to protect our employees and makes our work easier.
In case of noncompliance with this rule, the devices are returned to the sender without attaching postage to the package.
Overall size Measuring range
(Q
) l/min.
max.
FH G 1xx2 0.50 – 100 (120) 15. 7 0. 5 815 1,720 220,000 450 250
FH G 1xx4 1.00 – 400 (525) 56.5 3.138 318 40,800 450 250
FH G 1xx5 4.00 – 800 (1,000) 180 .0 10 100 12,800 450 500
FH G 1xx7 10.00 – 2,500 (3,000) 666.0 37 27 3, 459 40 500
Frequency range 0 … 100 kHz, adjustable
Measuring accuracy ± 0.3% [0.5%]*, [1%]** of measured value at viscosit y of 21 cSt
*FHG 1xx5, **FHG 1xx7
Repeatability ± 0.05% with same operating conditions
Materials
Gray cast iron version EN -GJS – 400 – 15 (EN 1563) / 100 Cr 6
Stainless steel version Stainless steel 1.4305/1.4112, additional available upon request
Bearing Fluid- dependent as anti-friction bearing or SSIC/wolfram carbide friction bearing
Seals FPM (standard) PTFE, NBR, EPDM upon request
RV
ccm/rev.
VE
ccm/Imp.
K – Factor
Imp./l min.
K – Factor
Imp./l max.
P max.
bar
Filtering
µm
Fluid temperature -30°C ... +120°C
Viscosity range 1 … 1,000,000 cSt
Installation position Any using selectable connection units, also customer specic
Supply voltage 9 … 28 VDC
Current consumption 65 mA at 24 VDC unloaded
Delay time ≤ 8 mµs
Protection type IP 65
15
Page 18
Baugröße 100
Baugröße 100
Size 1xx2
Flow range 0 up to 120 l/min
Durchussbereich 0 bis 120 l/min
Flow range 0 up to 10 l/min
Durchussbereich 0 bis 10 l/min
Durchusswiderstand ∆p
Flow range pressure drop ∆p
Baugröße 400
Size 1xx4
Flow range 0 up to 500 l/min
Durchussbereich 0 bis 500 l/min
Durchusswiderstand ∆p
Flow range pressure drop ∆pFlow range pressure drop ∆p
Durchusswiderstand ∆p
Flow range pressure drop ∆p
Durchuss Q Durchuss Q
Flow range Q
Durchussbereich 0 bis 50 l/min
Flow range 0 up to 50 l/min
Durchusswiderstand ∆p
Flow range Q
Flow range pressure drop ∆p
Durchuss Q
Durchuss Q
Flow range QFlow range Q
Baugröße 800
Size 1xx5
Durchussbereich 0 bis 1.000 l/min
Flow range 0 up to 1,000 l/min
Durchusswiderstand ∆p
Durchuss Q
Flow range Q
Size 1xx7
Flow range 0 up to 3,000 l/min
Flow range pressure drop ∆p
Flow range Q
Page 19
Vorverstärker
Sensormodul
Messanschluss
Erdung
Ringschraube
Anschlusseinheit
Anschlusseinheit mit Sensormodul
Anschlusseinheit Anschlusseinheit mit Sensormodul
Messanschluss
Anschluss SAE 2
Vorverstärker Sensormodul Messanschluss G1/4 Erdung Ringschraube
Anschlusseinheit AR. 2500-R..
Flanschanschlussmaße DIN EN 1092
Gewicht 120 kg
FHG1xx5
Preamplier
Test Port
Connection unit AR. 800- H..
Sensor module
Test Port
Ring bolt
Connection unit with sensor module AR. 800 -H..
Earthing
Weight 81 kg
deep
FHG1xx7
Connection
Connection unit AR. 800-X..
Preamplier
Connection unit with sensor module AR. 800 -X ..
Sensor module
Test Port G 1/4
Earthing
Weight 81 kg
Ring bolt
Connection unit AR. 2500-R..
18
Flange connection dimension
Weight 120 kg
Page 20
4 x M10-18 tief
Vorverstärker
Sensormodul
Messanschluss
Erdung
Anschlusseinheit AR. 400-F..
Anschlusseinheit mit Sensormodul AR. 400-F..
Anschlusseinheit AR. 400-V..
Anschlusseinheit mit Sensormodul AR. 400-V..
Gewicht 22 kg
Messanschluss
Anschluss SAE 1 1/4
Gewicht 24,8 kg
M8-15 tief
beidseitig
4 x M14-25 tief
M8-15 tief
beidseitig
24. FHG Flow Meter Dimensions
FHG1xx2
Connection unit AR. 100-E..
Connection unit AR. 100-T..
Preamplier
Sensor module
Test Port
Connection unit with sensor module AR. 100-E..
Connection unit with sensor module AR. 100-T..
Test Port
Connection
Earthing
deep
Weight 12 kg
Weight 12.7 kg
FHG1xx4
Connection unit AR. 400- F..
Connection unit with sensor module AR. 400 -F..
Preamplier
Sensor module
Test Port
Test Port
M8-15 deep on both sides
M8-15 deep on both sides
Earthing
Weight 22 kg
Connection unit AR. 400-V..
Connection unit with sensor module AR. 400 -V..
Connection
deep
Weight 24.8 kg
17
Page 21
25. Pin Assignment
26. Connection Diagram
Flow sensor FHG
Channel 1
Power supply 0 voltage
channel 1 (white)
0 voltage (blue)
channel 2 (black)
Error/Direction (grey)
Power supply + voltage
Channel 2
Evaluating unit (e.g. display)
Signal input 1
Signal input 2
control input
Q: 8 l/min V: 11,53 l direction: pos. f: 424,9 Hz
19
Page 22
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY
RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product
was PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST
of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2009 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
Page 23
Where Do I Find Everything I Need for
Shop online at omega.com
SM
Process Measurement and Control?
OMEGA…Of Course!
TEMPERATURE
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies Wire: Thermocouple, RTD & Thermistor Calibrators & Ice Point References Recorders, Controllers & Process Monitors Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
Transducers & Strain Gages Load Cells & Pressure Gages Displacement Transducers Instrumentation & Accessories
FLOW/LEVEL
Rotameters, Gas Mass Flowmeters & Flow Computers Air Velocity Indicators Turbine/Paddlewheel Systems Totalizers & Batch Controllers
pH/CONDUCTIVITY
pH Electrodes, Testers & Accessories Benchtop/Laboratory Meters Controllers, Calibrators, Simulators & Pumps Industrial pH & Conductivity Equipment
DATA ACQUISITION
Data Acquisition & Engineering Software Communications-Based Acquisition Systems Plug-in Cards for Apple, IBM & Compatibles Datalogging Systems Recorders, Printers & Plotters
HEATERS
Heating Cable Cartridge & Strip Heaters Immersion & Band Heaters Flexible Heaters Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
Metering & Control Instrumentation Refractometers Pumps & Tubing Air, Soil & Water Monitors Industrial Water & Wastewater Treatment pH, Conductivity & Dissolved Oxygen Instruments
M-5241/0613
Loading...