Omega DPG5600 User Manual

DPG3500, DPG3600, DPG5500, DPG5600 Series Digital Pressure Gauges M4094/0112
DPG3500 Battery Powered, Min/Max/Zero, NEMA 4X DPG3600 Battery Powered, Min/Max/Zero, NEMA 4X, Display Backlighting DPG5500 Battery Powered, Min/Max/Zero DPG5600 Battery Powered, Min/Max/Zero, Display Backlighting
Ranges and Resolution
See table below for ranges and user selectable units. Units are user selectable via internal buttons Resolution is fixed for each engineering unit
Accuracy
Includes linearity, hysteresis, repeatability ±0.25% of full scale ±1 least significant digit
Display
Batteries, Battery Life, Low Battery Indication
2 AA alkaline included. Approx. 2000 hours battery life DPG3600, DPG5600: 150-1500 hrs. depending on backlight usage
Low battery symbol on display
Controls & Functions
Front button turns gauge on or off, zeros gauge reference gauges, and cycles through min/max functions Internal push buttons for selection of engineering units and auto shutoff times. Pass code protected setup and calibration. DPG3600, DPG5600: Front button activates backlighting for 1 minute
Min/Max Functions
Minimum and maximum readings stored 3 times per second Front button cycles through min display, max display, clear Configurable for min only, max only, min/max, or none, clear min/max at power off or retain min/max at power off
Calibration
Pass code protected calibration Non-interactive zero, span, and linearity, ±10% of range
Auto Shutoff
User selectable 1 minute to 8 hours or front button on/off Factory default 5 minutes
Weight
Gauge: 9 ounces (approximately) Shipping: 1 pound (approximately)
Housing
Extruded aluminum case, epoxy powder coated, ABS/ polycarbonate bezel, front and rear gaskets, polycarbonate label
DPG3500, DPG3600: ABS/polycarbonate NEMA 4X case, rear gasket, polycarbonate label
Connection, Material, Media Compatibility
1/4" NPT male fitting, 316L stainless steel All wetted parts are 316L stainless steel Compatible with most liquids and gases
Range
add after model number
03G
85INH20G
05G
140INH20G
15A
760TORRA
30INHGVAC
30V15G
15G
1BARG
30A 30G 60G
30V100G
100G
7BARG
30V200G
200G
14BARG
300G 500G
1KG 2KG 3KG 5KG
Default Units
0-3.000 psig 0-85.0 inches of water 0-5.000 psig 0-140.0 inches of water
15.00-0 psi absolute
760.0-0 torr absolute 0-30.00 inHg vac vacuum: psig, inHg, inH –30.00 inHg vac to 15.00 psig ±psig, ±inHg, ±inH 0-15.00 psig 0-1.000 bar
30.00-0 psi absolute absolute: psia, inHg, inH 0-30.00 psig psig, inHg, inH 0-60.00 psig psig, inHg, inH –30.0 inHg vac to 100.0 psig ±psig, ±inHg, ±inH 0-100.0 psig 0-7.000 bar –30.0 inHg vac to 200.0 psig ±psig, ±inHg, ±inH 0-200.0 psig 0-14.00 bar 0-300.0 psig psig, inHg, oz/in 0-500.0 psig psig, inHg, ftH 0-1000 psig psig, inHg, ftH 0-2000 psig psig, inHg, ftH 0-3000 psig psig, inHg, ftH 0-5000 psig psig, MPa, bar, kg/cm
¼" NPT
Overpressure
2 X pressure range for 3 psi to 2000 psi sensors 5000 psig for ranges using 3000 psig sensor 7500 psig for ranges using 5000 psig sensor Over-range display 112.5% FS: I – – – or I –.–.–.–
Vacuum service: all 15 psi, 30 psia, all 100 psi, all 200 psi sensors Under-range display (non-vacuum sensors): –Err
Burst Pressure
4 X sensor pressure rating or 10,000 psi, whichever is less
Environmental
Storage temperature: –40 to 203°F (–40 to 95°C) Operating temperature: –4 to 185°F (–20 to 85°C) Compensated temperature: 32 to 158°F (0 to 70°C)
DPG5500 DPG5600
DPG3500 NEMA 4X DPG3600 NEMA 4X
Low battery indicator
2
XXXXX
8
8
8
8
8
Omega Engineering / Made in USA
3.38" Standard
3.5" NEMA 4X
psig, inHg, inH
psig, inHg, inH
absolute: psia, inHga, inH
psig, inHg, inH
psig, inHg, inH
psig, inHg, inH
2.88"
Standard
Turn at
hex
fitting
only
0.75"
1.65"
Selectable Units (models manufactured after 12-23-2012)
2
2O, oz/in
, g/cm2, mmHg, torr, mbar, cmH2O, mmH2O, ftH2O, kPa
2
2O, oz/in
, g/cm2, mmHg, torr, mbar, cmH2O, mmH2O, ftH2O, kPa
2
2Oa, oz/in
a, g/cm2a, mmHga, torra, mbara, cmH2Oa, mmH2Oa, ftH2Oa, kPaa, MPaa, bara, kg/cm2a, atma
2
2O, oz/in
, g/cm2, mmHg, torr, mbar, cmH2O, mmH2O, ftH2O, kPa, MPa, bar, kg/cm2, atm
2
2O, ±oz/in
, ±g/cm2, ±mmHg, ±torr, ±mbar, ±cmH2O, ±mmH2O, ±ftH2O, ±kPa, ±MPa, ±bar, ±kg/cm2, ±atm
2
2O, oz/in
, g/cm2, mmHg, torr, mbar, cmH2O, mmH2O, ftH2O, kPa, MPa, bar, kg/cm2, atm
2
2O, oz/in
, g/cm2, mmHg, torr, mbar, cmH2O, mmH2O, ftH2O, kPa, MPa, bar, kg/cm2, atm
2
2O, oz/in
, g/cm2, mmHg, torr, mbar, cmH2O, mmH2O, ftH2O, kPa, MPa, bar, kg/cm2, atm
2
2O, oz/in
, g/cm2, mmHg, torr, mbar, cmH2O, mmH2O, ftH2O, kPa, MPa, bar, kg/cm2, atm
2
2O, ±oz/in
, ±g/cm2, ±mmHg, ±torr, ±mbar, ±cmH2O, ±mmH2O, ±ftH2O, ±kPa, ±MPa, ±bar, ±kg/cm2, ±atm
2
2O, oz/in
, g/cm2, mmHg, torr, mbar, cmH2O, mmH2O, ftH2O, kPa, MPa, bar, kg/cm2, atm
2
2O, ±oz/in
, ±ftH2O, ±kPa, ±MPa, ±bar, ±kg/cm2, ±atm
2
2O, oz/in
, ftH2O, kPa, MPa, bar, kg/cm2, atm
2
, ftH2O, kPa, MPa, bar, kg/cm2, atm
2O, kPa, MPa, bar, kg/cm 2O, kPa, MPa, bar, kg/cm 2O, MPa, bar, kg/cm 2O, MPa, bar, kg/cm
2
, atm
2
, atm
2
, atm
2
, atm
2
, atm
3.0"
NEMA 4X
Turn at
hex
fitting
only
2.0"
User’s Guide
Shop online at
omega.com
e-mail: info@omega.com
For latest product manuals:
omegamanual.info
MADE IN
2
DPG 3500, 3600, 5500, 5600 Series Digital Pressure Gauges
Installation Precautions
Read these instructions before using the gauge. Configuration may
be easier before installation. Contact the factory for assistance.
These products do not contain user-serviceable parts. Contact us
for repairs, service, or refurbishment.
Gauges must be operated within specified ambient temperature ranges. Outdoor or wash down applications require a NEMA 4X gauge or
installation in a NEMA 4X housing.
Use a pressure or vacuum range appropriate for the application. Use fittings appropriate for the pressure range of the gauge. Due to the hardness of 316 stainless steel, it is recommended that
a thread sealant be used to ensure leak-free operation.
For contaminated media use an appropriate screen or filter to
keep debris out of gauge port.
Remove system pressures before removing or installing gauge. Install or remove gauge using a wrench on the hex fitting only. Do
not attempt to turn gauge by forcing the housing.
Good design practice dictates that positive displacement liquid
pumps include protection devices to prevent sensor damage from pressure spikes, acceleration head, and vacuum extremes.
Avoid permanent sensor damage! Do not apply vacuum to non-
vacuum gauges or hydraulic vacuum to any gauges.
Avoid permanent sensor damage! NEVER insert objects into
gauge port or blow out with compressed air.
Gauges are not for oxygen service. Accidental rupture of sensor dia-
phragm may cause silicone oil inside sensor to react with oxygen.
Types of Gauges
Gauge reference models read zero with the gauge port open. Compound ranges read vacuum in inches of Mercury, pressure in
psig, and zero with the gauge port open. 1000 psi and higher sensors are a sealed reference type. They read
zero with the gauge port open are internally referenced to 14.7 psi. They are functionally similar to gauge reference sensors.
Absolute reference gauges read zero at full vacuum and atmo­spheric pressure with the gauge port open. Note that readings of atmospheric pressure vary continuously.
Operation
Function Button Prompt (Release Button)
On Press for 1
Zero
(gauge ref.
Hi reading Press/hold Max > max. reading & units Lo reading Press/hold MIN > min. reading & units
Exit Hi/Lo Press/hold
Clear Hi/Lo Press/hold Max > Clr mx/mn > actual reading
Clear, off Press/hold Max > Clr mx/mn > OFF
Power-Up
Press and hold the front button for approximately 1 second. The display is tested, the default full-scale range is indicated, the
full-scale range in the selected units are indicated, the display is tested again, then the actual pressure and units are displayed.
Power-Up With Zero
This applies to gauge reference models only. Absolute reference gauges do not use the zero feature since they read atmospheric pressure under normal conditions.
Be sure the gauge port is exposed to normal atmospheric pressure and no pressure is applied. The zeroing function is only activated at each power-up and the stored zero correction is erased when the gauge is shut off.
Press and hold the front button. The display is tested and then oooo is displayed. The gauge is now zeroed. Release the button and the full-scale range in the selected units are indicated, the display is tested again, then the actual pressure and units are displayed.
Attempting to zero the gauge with pressure greater than approxi­mately 3% of full-scale pressure or vacuum applied will result in an error condition, and the display will alternately indicate Err 0 and the actual measured pressure. The gauge must be powered down to reset the error condition.
Normal Operation
Following the start-up initialization, the display indicates the pres­sure reading updated approximately 3 times per second. The auto shutoff timer starts when the gauge is powered up or whenever the button is pushed, unless the gauge shutoff time was set to zero for on/off operation.
Press and
only)
Display test > default units > selected
second
units FUll scale > display test > actual reading
Display test > selected units FUll scale >
hold
display test > actual reading

oooo
> actual reading
> default units >
Operation—continued
If excessive vacuum is applied to a pressure-only gauge, the display will indicate -Err until the vacuum is released. Applying vacuum to a gauge designed for pressure may damage the pressure sensor.
If excessive pressure is applied (112.5% over range), an out-of­range indication of 1 – – – or 1.–.–.– will be displayed depending on model.
Display Backlighting (DPG3600, DPG5600 models only)
Display backlighting can be turned on by momentarily pressing the front button whenever the gauge is on. The backlighting will turn on for one minute and then automatically shut off. This also restarts the auto shutoff timer. The display backlighting will not be apparent under bright lighting conditions.
Minimum and Maximum Readings
Gauges are normally configured with minimum and maximum capture functions enabled. One or both can be disabled in the Advanced Configuration mode.
Minimum and maximum readings are continuously stored and updat­ed whenever the gauge is on. The stored readings can be manually cleared if desired. The MAX and MIN memory is also cleared when­ever the gauge is off unless configured to save the readings.
Press and hold the button for about 1 second until MAX is displayed alternating with the units. The maximum reading will be continu­ously updated. The gauge may be left in this mode.
After MAX is displayed, press and hold the button for about 1 sec­ond until MIN is displayed alternating with the units. The minimum reading will be continuously updated. The gauge may be left in this mode. If excessive vacuum is applied to a pressure-only gauge while in this mode, the display will indicate –Err until the MAX/MIN readings are cleared.
After MIN is displayed, press and hold the button again for about 1 second until * * * * is displayed. The MAX and MIN memory is not erased and the gauge returns to normal operation with the display indicating the current reading.
Press and continue to hold the button until the display indicates clr MX/MN (about 3 seconds total) and then release the button. Both maximum and minimum values are cleared and the gauge returns to the normal mode and displays the current pressure.
Shut-Down
To shut off the gauge manually at any time, press and hold the button until the display indicates OFF (about 5 seconds) and then release.
When an auto shutoff timer is used, the display indicates OFF five seconds prior to auto shutoff. The button can be pressed to keep the gauge on. The auto shutoff and backlight (if equipped) timers are reset whenever the button is pressed and released.
If the gauge set up without auto shutoff (on/off operation) it will stay on until manually shut off or until the batteries are depleted. Turn gauge off when not in use to conserve battery life.
Basic Configuration
Engineering Unit Selection
Engineering unit selection is done via internal buttons to help pre­vent accidental or unauthorized changes. The selected engineering unit is stored in non-volatile memory and will be retained even with the gauge off or batteries removed. The available engineering units depend on the sensor range and available display resolution.
Compound (inHg/PSIG) gauges must be changed to display single­unit vacuum/pressure readings in the Advanced Configuration mode before different engineering units can be selected.
The default engineering units are mathematically converted to the newly selected engineering unit. When the gauge is powered up, the originally configured range is displayed and then the conversion with the selected engineering unit is displayed.
To change engineering units remove the rear cover to gain access to the two internal buttons located near the lower right and left corners of the circuit board.
With the gauge powered up, press and hold the UP button. Release the button when the engineering units begin to flash.
Use the UP and DOWN but­tons to scroll through the list of engineering units available for the pressure range of the sensor.
When the desired units are displayed, press and release the front button to save the selection and return to normal operation.
Note: The gauge will automatically revert to normal operation if no buttons are operated for approximately 15 seconds.
Replace the rear cover taking care not to pinch the power wires between the cover and the case.
Basic Configuration—continued
Auto Shutoff Time Selection
Auto shutoff time selection is done via internal buttons to help prevent accidental or unauthorized changes. The selected shut off time is stored in non-volatile memory and will be retained even with the battery off or batteries removed.
Remove the rear cover to gain access to the two internal buttons located near the lower right and left corners of the circuit board.
With the gauge powered up, press and hold the DOWN button. Release the button when the auto shutoff time is displayed.
The lower display will indicate AST M if the time displayed is in minutes, and AST H if it in hours.
An auto shutoff time of 0 signifies that the auto shutoff feature is disabled and the front button turns the gauge on and off.
Use the UP and DOWN buttons to select 0, 1, 2, 5, 10, 15, 20 or 30 minutes, or 1, 2, 4, or 8 hours.
When the desired time is displayed, press and release the front but­ton to save the selection and return to normal operation.
Note: The gauge will automatically revert to normal operation if no buttons are operated for approximately 15 seconds.
Replace the rear cover taking care not to pinch the power wires between the cover and the case.
Advanced Configuration
Advanced configuration requires a pass code for access and allows more features to be configured.
Remove the rear cover to gain access to the buttons located near the lower right and left corners of the circuit board.
With the gauge off, press and hold the UP button. Then press the front button. Release all buttons when the display indicates CFG and the program version then the full-scale range is indicated and the display is tested.
The display then indicates _ _ _ _ with the first underscore blink­ing, and with CFGPC (configuration pass code) on the character segments.
Note: The gauge will automatically revert to normal operation if no buttons are operated for approximately 15 seconds. To cancel and return to normal operation, press and release the front button without entering any pass code characters.
Pass Code Entry
The factory default is 3510, but this may be changed by the user under the Pass Code Configuration section.
1. Use the UP or DOWN buttons to set the left-most digit to 3.
2. Press and release the front button to move to the next position. The 3 will remain, and the second position will be blinking.
3. Use the UP or DOWN buttons to select 5.
4. Press and release the front button to index to the next position. 35 will remain, and the third position will be blinking.
5. Use the UP or DOWN buttons to select 1.
6. Press and release the front button to index to the next position. 351 will remain, and the fourth position will be blinking.
7. Use the UP or DOWN buttons to select 0.
8. Press and release the front button to proceed with configuration procedures.
If an incorrect pass code is entered, the gauge will return to the start of the pass code entry sequence.
Factory/User Configuration
The upper display section will be blank, and the lower section will display either USER_ or FCTRY.
If USER_ is selected, the user configuration can be modified as described in the following steps.
To select USER_, press and release the DOWN button. The lower display will indicate USER_. Press and release the front button to continue. If FCTRY is selected, the existing user configuration will be replaced
by the original factory configuration. To select FCTRY, press and release the UP button. The lower display will indicate FCTRY. Press and release the front button to restore the factory configura-
tion and restart the gauge.
Max/Min Configuration
Use the UP and DOWN buttons to select from the following: MX/MN Both highest and lowest values will be captured MX/--- Only highest value will be captured
--/MN Only lowest value will be captured
--/-- Capture feature is disabled
Press and release the front button to move to the next parameter.
Loading...
+ 2 hidden pages