Omega DPG1100AD User Manual

Digital Pressure Gauges
DPG1100AD Series, NEMA 4X, Low Voltage AC/DC Powered DPG1100ADBL Series, NEMA 4X, Low Voltage AC/DC Powered with Backlit Display
INSTRUCTION
SHEET
M3763/1101
Description
Installation Precautions
Tighten/remove with wrench on hex fitting only. Do not attempt to rotate gauge by turning housing. Use fittings appropriate for the pressure range of the gauge. Do not apply vacuum to gauges not designed for vacuum operation. NEVER insert objects into the gauge por t or blow out with compressed air. Permanent damage not covered by warranty will result to the sensor.
The DPG1100AD and DPG1100ADBL can be powered by any AC source of 8 to 24 VAC 50/60 Hz, or any DC source of 9 to 32 VDC. The gauges can be used with inexpensive unregulated low voltage AC or DC power supplies. The type and magnitude of the supply voltage have negligible effects on the gauge calibration as long as it is within the voltage ranges stated above. No polarity needs to be observed when connecting a DC supply.
After the gauge is installed, route the wires safely and connect the AC adapter’s wires to the gauge cable. Lastly, plug the AC adapter into a 115 VAC outlet.
The only important consideration is to ensure that the gauge supply voltage does not fall below 8 VAC RMS if AC power is used, or 9 VDC if DC power is used. Operation with less than these values may cause erratic or erroneous readings.
If your application requires operation of multiple gauges from the same power supply, consult the factory for wiring recommendations.
Operation
When a supply voltage is applied, the gauge will be ready to use. If the gauge display is off, press the center button to turn the gauge on. If the gauge is in the power­on state and the power is disconnected, the gauge will turn on when power is reapplied. The gauge can be left on continuously or turned off when not in use.
DPG1100ADBL model display backlighting will be on whenever gauge is on.
Calibration
Remove the calibration potentiometer covers on the front of the unit to access the zero and span controls.
GAUGE reference units may be re-zeroed without affecting the span calibration. The gauge port must be open to the ambient with no pressure or vacuum applied. Adjust the Zero control until the gauge reads zero with the minus (–) sign occasionally flashing.
Span calibration should only be attempted if the user has access to a pressure reference of known accuracy. The accuracy of the calibration equipment should be at least four times the gauge accuracy. Zero calibration must be done before span calibration. Record readings at three to five points over the range of gauge and adjust span control to minimize error and meet specifications.
ABSOLUTE reference gauges require vacuum generation and atmospheric pressure measurement equipment for accurate calibration and thus are more difficult to calibrate in the field.
Gauges may be returned to Omega Engineering for recalibration.N.I.S.T.traceability is available.
3 ft long, 2-conductor, 22 AWG shielded cable
Optional Power Supply Kit DPG1000-PS
The kit includes a UL and CSA listed 115 VAC (50/60 Hz) wall-mount power supply with U.S.-style 2-prong plug. Output is 12 VDC at 200 mA.The two-conductor wire is approximately 6 feet long with stripped wire ends. Use a pair of pliers to connect gauge and power supply wires using the included moisture-resistant connector.
2.0"
1
/4" NPT
3.0"
3.5"
0.75"
PSIG
Span
Zero
OMEGA Engineering Inc. / Made in USA
0.75"
Turn at hex fitting only!
SPECIFICATIONS
Ranges & Resolution
–30.0 inHg vac to 15.0 psig –30.0 inHg vac to 100.0 psig
30.0 inHg vacuum, ±15.00 psig
3.00, 5.00, 15.00, 30.0, 100.0, 199.9 psig 300, 500, 1000 psig Absolute reference:15.00, 100.0 psia
Optional Units
Most engineering units such as kPa, atm, bar, mbar, inHg, mmHg, inH2O, ftH2O, torr, kg/cm2, cmH2O, oz/in
2
Display (type, size, update rate)
31/2 digit LCD, 1/2" digit height 3 readings per second nominal display update ADBL: Red LED display backlighting when gauge is on
Accuracy (linearity, hysteresis, repeatability)
±0.25% of full scale or better, ±1 least significant digit
Temperature Stability (relative to 77°F or 25°C)
±1% FS for offset & span, 32 to 158°F (0 to 70°C) typical ±2% FS for offset & span, 32 to 158°F (0 to 70°C) typical for 3 and 5 psi ranges
Controls & Location
Front button turns gauge on or off Front-accessible calibration potentiometers Non-interactive zero & span, ±10% range
Power
AD: Approximately 5 mA ADBL: Approximately 80 mA Operates on 8 to 24 VAC 50/60 Hz or 9 to 32 VDC Optional wall mount power supply is available to operate on 115VAC
Weight (approximate)
Gauge: 9 ounces, shipping weight:1 pound
Housing
NEMA 4X UV stabilized polycarbonate/ABS case, light gray color Clear polycarbonate window to protect display Gasketed rear cover, six captive stainless steel screws
Pressure/Vacuum Connection and Material
1
/
4" NPT male, all wetted parts are 316 SS
Overpressure & Burst
5000 psig for metric units using 3000 psig sensor 7500 psig for metric units using 5000 psig sensor All others 2x rated pressure minimum Burst: 4x rated pressure minimum or 10,000 psi, whichever is less
Storage temperature: –40 to 203°F (–40 to 95°C) Operating temperature: –4 to 185°F (–20 to 85°C) Compensated temperature: 32 to 158°F (0 to 70°C)
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc.accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING:These products are not designed for use in, and should not be used for, patient connected applications.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC.warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase.OMEGA’s Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Depar tment will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge.OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion;or current, heat, moisture or vibration;improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products.However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.LIMITATION OF LIABILITY:The remedies of purchaser set forth herein are exclusive and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based.In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used:(1) as a “Basic Component”under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from a ny liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit. FOR W
ARRANTY RETURNS, please have the following
information available BEFORE contacting OMEGA:
1. P.O. number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges.Have the following information available BEFORE contacting OMEGA:
1. P.O.number to cover the COST of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems relative to
the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2001 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any
electronic medium or machine-readable form, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford CT 06907-0047
TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada@omega.com
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378 / 1-800-622-BEST
®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
®
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico: Tel: (001) 800-826-6342 FAX: (001) 203-359-7807
En Espan˜ol: (001) 203-359-7803 e-mail: espanol@omega.com
info@omega.com.mx
OMEGAnet®On-Line Service Internet e-mail
www.omega.com info@omega.com
Servicing Europe:
Benelux: Postbus 8034, 1180 LAAmstelveen, The Netherlands
Tel: +31 (0)20 6418405 FAX: +31 (0)20 6434643 Toll Free in Benelux: 06 0993344 e-mail: nl@omega.com
Czech Republic: Rudé armády 1868,73301 Karviná
TEL: +420 (0)69 6311899 FAX: +420 (0)69 6311114 Toll Free in Czech Rep.: 0800-1-66342 e-mail: czech@omega.com
France: 9, rue Denis Papin, 78190 Trappes
TEL: +33 (0)130 621 400 FAX: +33 (0)130 699 120 Toll Free in France: 0800-4-06342 e-mail: france@omega.com
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 (0)7056 3017 FAX: +49 (0)7056 8540 Toll Free in Germany: 0800 TC OMEGA
SM
e-mail: germany@omega.com
United Kingdom: P.O. Box 7, Omega Drive,
ISO 9002 Certified Irlam, Manchester,
M44 5EX, United Kingdom TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in England: 0800 488 488 e-mail: sales@omega.co.uk
Loading...