Omega DP63000A-I User Manual

LP0694A
User’s Guide
DDPP6633000000AA-II
Digital PPanel MMeter
Shop online at
omega.com
e-mail: info@omega.com
For latest product manuals:
omegamanual.info
OMEGAnet®Online Service
omega.com
Internet e-mail
info@omega.com
Servicing North America:
U.S.A.: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford, CT 06907-0047
TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1, Canada TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca
For immediate technical or application assistance:
U.S.A. and Canada: Sales Service: 1-800-826-6342/1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378/1-800-622-BEST
®
Engineering Service: 1-800-872-9436/1-800-USA-WHEN
®
Mexico: En Español: (001) 203-359-7803 e-mail: espanol@omega.com
FAX: (001) 203-359-7807 info@omega.com.mx
Servicing Europe:
Czech Republic: Frystatska 184, 733 01 Karviná, Czech Republic
TEL: +420 (0)59 6311899 FAX: +420 (0)59 6311114 Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29 Toll Free in Germany: 0800 639 7678 e-mail: info@omega.de
United Kingdom: One Omega Drive, River Bend Technology Centre
ISO 9002 Certified Northbank, Irlam, Manchester
M44 5BD United Kingdom TEL: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622 Toll Free in United Kingdom: 0800-488-488 e-mail: sales@omega.co.uk
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING : These products are not designed for use in, and should not be used for, human applications.
3
z FOUR SELECTABLE D.C. RANGES
200 µA, 2 mA, 20 mA, 200 mA
z MINIMUM AND MAXIMUM DISPLAY CAPTURE
z LCD, REFLECTIVE OR RED/GREEN LED BACKLIGHTING
z 0.48" (12.2 mm) HIGH DIGITS
z OPTIONAL SETPOINT OUTPUT MODULES
z OPTIONAL SERIAL COMMUNICATIONS MODULES (RS232 or RS485)
z OPERATES FROM 9 TO 28 VDC POWER SOURCE
z FRONT PANEL OR SOFTWARE PROGRAMMABLE
z DISPLAY COLOR CHANGE CAPABILITY AT SETPOINT OUTPUT
z NEMA 4X/IP65 SEALED FRONT BEZEL
GENERAL DESCRIPTION
The DP63000 provides the user the ultimate in flexibility, from its complete user programming to the optional setpoint control and communication capability. This unit accepts a DC Current input signal and provides a display in the desired unit of measure. The meter also features minimum and maximum display capture, display offset, units indicator, and programmable user input. The display can be toggled either manually or automatically between the selected displays.
The DP63000 display has 0.48" (12.2 mm) high digits. The LCD is available in two versions, reflective and red/green backlight. The backlight version is user selectable for the desired color and also has variable display intensity.
The capability of the DP63000 can be easily expanded with the addition of option modules. Setpoint capability is field installable with the addition of the setpoint output modules. Serial communications capability for RS232 or RS485 is added with a serial option module.
The DP63000 can be powered from an optional Power Supply (Model Number DP6-MLPS1), that attaches directly to the back of a DP63000. The DP6-MLPS1 is powered from 85 to 250 VAC and provides up to 400 mA to drive the unit and sensors.
CURRENT
This unit is the DC Current meter. It features 4 current input ranges, that are selected by the user via a programming jumper and software input range selection. The ranges consist of following: 200 µA, 2 mA, 20 mA, or 200 mA. Users should select the appropriate current range that covers their maximum signal input.
SAFETY SUMMARY
All safety related regulations, local codes and instructions that appear in this literature or on equipment must be observed to ensure personal safety and to prevent damage to either the instrument or equipment connected to it. If equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Do not use this meter to directly command motors, valves, or other actuators not equipped with safeguards. To do so can be potentially harmful to persons or equipment in the event of a fault to the meter.
CAUTION: Risk of Danger.
Read complete instructions prior to
installationand operation of the unit.
CAUTION: Risk of electric shock.
DIMENSIONS In inches (mm)
Note: Recommended minimum clearance (behind the panel) for mounting clip installation is 2.15" (54.6) H x 3.00" (76.2) W.
C
US LISTEDUS
LISTED
U
L
R
IND. CONT. EQ.
51EB
1. DISPLAY: 5 digit LCD 0.48" (12.2 mm) high digits
DP63000A-I: Reflective LCD with full viewing angle DP63000B-I: Transmissive LCD with selectable red or green LED backlight,
viewing angle optimized. Display color change capability with output state when using an output module.
2. POWER: Input voltage range is +9 to +28 VDC with short circuit and input polarity protection. Must use a DP6-MLPS1 or a Class 2 or SELV rated power supply.
3. INPUT RANGES: Jumper Selectable D.C. Currents: 200 µA, 2 mA, 20 mA, or 200 mA
4. SIGNAL INPUTS:
5. OVERRANGE RATINGS, PROTECTION & INDICATION: 9 to 28 VDC power circuit is not isolated from the signal circuit.
Input Overrange Indication: “OLOL”. Input Underrange Indication: “ULUL”. Display Overrange/Underrange Indication: “
.....”/“-.....
6. DISPLAY RESPONSE TIME: 500 msec min.
7. NORMAL MODE REJECTION: 60 dB 50/60 Hz
8. USER INPUT (USR): Programmable input. Connect terminal to common (USR COMM) to activate function. Internal 10Kpull-up resistor to +9 to 28 VDC. Threshold Levels: V
IL
= 1.0 V max; VIH= 2.4 V min; V
MAX
= 28 VDC
Response Time: 5 msec typ.; 50 msec debounce (activation and release)
9. MEMORY: Nonvolatile E2PROM memory retains all programming parameters and max/min values when power is removed.
10. ENVIRONMENTAL CONDITIONS:
Operating Temperature Range for DP63000A-I: -35 to 75°C Operating Temperature Range for DP63000B-I depends on display color
and intensity level as per below:
Storage Temperature: -35 to 85°C Operating and Storage Humidity: 0 to 85% max. relative humidity (non-
condensing)
Vibration According to IEC 68-2-6: 5 to 500 Hz, in X, Y, Z direction for 1.5
hours, 5 g’s.
Shock According to IEC 68-2-27: Operational 30 g, 11 msec in 3 directions Altitude: Up to 2000 meters
11. CONNECTIONS: Wire clamping screw terminals Wire Strip Length: 0.3" (7.5 mm) Wire Gage: 30-14 AWG copper wire Torque: 5 inch-lbs (0.565 N-m) max.
12. CONSTRUCTION: This unit is rated for NEMA 4X/IP65 requirements for
indoor use. Installation Category I, Pollution Degree 2. High impact plastic case with clear viewing window. Panel gasket and mounting clip included.
13. CERTIFICATIONS AND COMPLIANCES:
SAFETY
UL Recognized Component, File #E313607, UL61010A-1, CSA 22.2 No. 61010-1
Recognized to U.S. and Canadian requirements under the Component Recognition Program of Underwriters Laboratories, Inc.
UL Listed, File # E313547, UL508, CSA C22.2 No. 14-M95
LISTED by Und. Lab. Inc. to U.S. and Canadian safety standards IEC 61010-1, EN 61010-1: Safety requirements for electrical equipment
for measurement, control, and laboratory use, Part 1. IP65 Enclosure rating (Face only), IEC 529 Type 4X Indoor Enclosure rating (Face only), UL50
ELECTROMAGNETIC COMPATIBILITY
Emissions and Immunity to EN 61326: Electrical Equipment for Measurement,
Control and Laboratory use.
Notes:
1. Criterion A: Normal operation within specified limits.
14. WEIGHT: 3.2 oz (100 g)
4
GENERAL METER SPECIFICATIONS
MODEL
NO.
DISPLAY COLOR
INPUT CURRENT
@ 9 VDC WITH
DP6-RLY0
DP63000A-I --- 10 mA 40 mA DP63000B-I Red (max intensity) 85 mA 115 mA DP63000B-I Green (max intensity) 95 mA 125 mA
TEMPERATURE
1 & 2 -35 to 75°C
3 -35 to 70°C 4 -35 to 60°C 5 -35 to 50°C
1 & 2 -35 to 75°C
3 -35 to 65°C
5
4
-35 to 35°C
-35 to 50°C
INTENSITY LEVEL
Class AEN 55011Emissions
Emissions:
30 A/m
3 V/rms Criterion A
Criterion A
EN 61000-4-8
EN 61000-4-6RF conducted interference
1 kV L-L,
Criterion AEN 61000-4-5Surge
1 kV signal
2 kV power
Criterion AEN 61000-4-4Fast transients (burst)
2 kV L&N-E power
10 V/m
Criterion AEN 61000-4-3Electromagnetic RF fields
8 kV air discharge
4 kV contact discharge
Criterion AEN 61000-4-2Electrostatic discharge
Immunity to Industrial Locations:
Power frequency magnetic fields
Green Display
Red Display
INPUT CURRENT
@ 9 VDC WITHOUT
DP6-RLY0
INPUT
RANGE
ACCURACY @23 °C, less than 85% RH
INPUT
IMPEDANCE
MAX
INPUT
SIGNAL
RESOLUTION
TEMP.
COEFFICIENT
200 µA
0.1% of span
1.111 K
15 mA 10 nA 70 ppm / °C
2 mA 0.1% of span 50 mA .1 µA 70 ppm / °C
20 mA 0.1% of span
11
150 mA 1 µA 70 ppm / °C
200 mA 0.1% of span
1
500 mA 10 µA 70 ppm / °C
111
PART NUMBER INFORMATION
DP6-SNK0Dual Sinking Open Collector Output card
DP6-SOFT
DP6-485-CABLERS485 Programming Cable (DB9-RJ11)
DP63000B-I
PC Configuration Software for Windows 98, ME, 2000, XP
*
DC Current Meter with backlight display
DP6-232-CABLE
DP6-RLY0
DP6-MLPS1Micro-Line Power Supply, 85 to 250 VAC
Single Relay Output Card
RS232 Programming Cable (DB9-RJ11)
DP6-COM2RS232 Serial Communications Card
DP6-COM1RS485 Serial Communications Card
DP63000A-IDC Current Meter with reflective display
PART NUMBERDESCRIPTION
5
ADDING OPTION CARDS
The DP63000 meters can be fitted with optional output cards and/or serial communications cards. The details for the plug-in cards can be reviewed in the specification section below. The plug-in cards, that are sold separately, can be installed initially or at a later date.
WARNING: Disconnect all power to the unit before
installing Plug-in card.
Note: Measurement errors may occur if signal input common is shared with
another circuit common (ie, serial common, Dual Sinking Output option card,
or Power Supply common) on multiple units.
SINGLE RELAY CARD
Type: Single FORM-C relay
Isolation To Sensor & User Input Commons: 1400 Vrms for 1 min.
Working Voltage: 150 Vrms
Contact Rating: 1 amp @ 30 VDC resistive; 0.3 amp @ 125 VAC resistive
Life Expectancy: 100,000 minimum operations
Response Time:
Turn On Time: 4 msec max. Turn Off Time: 4 msec max.
DUAL SINKING OUTPUT CARD
Type: Non-isolated switched DC, N Channel open drain MOSFET Current Rating: 100 mA max. V
DS ON
: 0.7 V @ 100 mA
V
DS MAX
: 30 VDC
Offstate Leakage Current: 0.5 mA max.
RS485 SERIAL COMMUNICATIONS CARD
Type: RS485 multi-point balanced interface (non-isolated) Baud Rate: 300 to 38.4k Data Format: 7/8 bits; odd, even, or no parity Bus Address: 0 to 99; max 32 meters per line Transmit Delay: Selectable (refer to DP6-COM bulletin)
RS232 SERIAL COMMUNICATIONS CARD
Type: RS232 half duplex (non-isolated) Baud Rate: 300 to 38.4k Data Format: 7/8 bits; odd, even, or no parity
OPTIONAL PLUG-IN CARDS
1.0 I
NSTALLING THE M
ETER
INSTALLATION
The meter meets NEMA 4X/IP65 requirements when properly installed. The unit is intended to be mounted into an enclosed panel. Prepare the panel cutout to the dimensions shown. Remove the panel latch from the unit. Slide the panel gasket over the rear of the unit to the back of the bezel. The unit should be installed fully assembled. Insert the unit into the panel cutout.
While holding the unit in place, push the panel latch over the rear of the unit so that the tabs of the panel latch engage in the slots on the case. The panel latch should be engaged in the farthest forward slot possible. To achieve a proper seal, tighten the latch screws evenly until the unit is snug in the panel (Torque to approx. 28 to 36 in-oz [0.202 to 0.26 N-m]). Do not over-tighten the screws.
INSTALLATION ENVIRONMENT
The unit should be installed in a location that does not exceed the operating temperature and provides good air circulation. Placing the unit near devices that generate excessive heat should be avoided.
The bezel should only be cleaned with a soft cloth and neutral soap product. Do NOT use solvents. Continuous exposure to direct sunlight may accelerate the aging process of the bezel.
Do not use tools of any kind (screwdrivers, pens, pencils, etc.) to operate the keypad of the unit.
2.0 SETTING THE JUMPERS
INPUT RANGE JUMPER
This jumper is used to select the proper input range. The input range selected in programming must match the jumper setting. Select a range that is high enough to accommodate the maximum signal input to avoid overloads. To access the jumper, remove the rear cover of the meter.
Warning: Exposed line voltage exists on the circuit boards. Remove
all power to the meter and load circuits before accessing inside of the meter.
REMOVING THE REAR COVER
To remove the rear cover, locate the cover locking tab below the 2nd and 3rd input terminals. To release the tab, insert a small, flat blade screwdriver between the tab and the plastic wall below the terminals. Inserting the screwdriver will provide enough pressure to release the tab locks. To replace the cover, align the cover with the input terminals and press down until the cover snaps into place.
Loading...