Omega DP41-S, DP41-S-230, DP41-S-GN, DP41-S-230-GN, DP41-S-DC-GN User Manual

...
Page 1
www.omega.com
e-mail: info@omega.com
For latest product manuals
www.omegamanual.info
®
®
User’s Guide
DP41-S
High Performance
Strain Gage Indicator
Page 2
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These pr
oducts are not designed for use in, and should not be used for, patient connected applications.
This device is marked with the international hazard symbol. It is important to read the Setup Guide before installing or commissioning this device as it contains important information relating to safety and EMC.
United Kingdom: One Omega Drive
ISO 9001 Certified River Bend Technology Centre
Northbank, Irlam Manchester M44 5BD United Kingdom TEL: +44 161 777 6611 FAX: +44 161 777 6622 Toll Free in England: 0800 488 488
e-mail: sales@omega.co.uk
France:
TEL: +33 1 61 37 29 00 FAX: +33 1 30 57 54 27 Toll Free in France: 0800 466 342
e-mail: sales@omega.fr
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford CT 06907-0047
TEL: (203) 359-1660 FAX: (203) 359-7700
e-mail: info@omega.com
Canada:
976 Bergar Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886
e-mail: info@omega.ca
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378 / 1-800-622-BEST
®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
®
Mexico and TEL: (001) 203-359-7803 FAX: (001) 203-359-7807 Latin America:
e-mail: espanol@omega.com
Servicing Europe:
Benelux: TEL: +31 20 3472121 FAX: +31 20 6434643
Toll Free in Benelux: 0800 0993344
e-mail: sales@omegaeng.nl
Czech Republic:
Frystatska 184, 733 01 Karviná TEL: +420 59 6311899 FAX: +420 59 6311114
e-mail: info@omegashop.cz
OMEGAnet® On-Line Service
www.omega.com
Internet e-mail
info@omega.com
®
®
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
TEL: +49 7056 9398-0 FAX: +49 7056 9398-29 Toll Free in Germany: 0800 639 7678
e-mail: info@omega.de
Page 3
SECTION 1 INTRODUCTION
1.1 Description.............................................................................1-1
1.2 Features.................................................................................1-1
1.3 Available Models....................................................................1-2
SECTION 2 UNPACKING .........................................................................2-1
SECTION 3 SAFETY CONSIDERATIONS
3.1 Safety Considerations............................................................3-1
SECTION 4 PARTS OF THE METER
4.1 Front of the Meter ..................................................................4-1
4.2 Rear of the Meter...................................................................4-3
SECTION 5 SETUP
5.1 Conditions Requiring Disassembly........................................5-1
5.2 Assembly/Disassembly Opening ...........................................5-1
5.2.1 Safety Precaution/Product ID Label ......................................5-4
5.2.2 Main Board Power Jumpers ..................................................5-5
5.2.3 Printed Circuit Board(s) Installation .......................................5-6
5.2.4 How to Access Jumpers ........................................................5-7
5.2.5 Panel-Mount Assembly ........................................................5-11
SECTION 6 JUMPER POSITIONS
6.1 Introduction............................................................................6-1
6.2 S1 Jumper Positions for Readrate and
Unipolar or Bipolar Input(s)....................................................6-1
6.3 S2 Jumper Positions for Input Range....................................6-1
6.4 Jumper setting(s) for Sensor Excitation.................................6-3
6.4.1 Jumper setting(s) for Sensor Excitation ac-powered unit ......6-3
6.4.2 Jumper setting(s) for Sensor Excitation dc-powered unit ......6-5
SECTION 7 SIGNAL AND POWER INPUT CONNECTIONS
7.1 Introduction............................................................................7-1
7.2 Signal Input Connections.......................................................7-1
7.3 Connecting Main Power.........................................................7-3
SECTION 8 METHODS FOR SCALING THE METER TO DISPLAY IN
ENGINEERING UNITS
8.1 Introduction............................................................................8-1
8.2 Strain Gage (Bridge) Input, programming the meter for a
desired Display ......................................................................8-1
8.3 Voltage Input (Non-Ratiometric), programming the meter
for a desired Display...........................................................8-15
8.4 Current Input, programming the meter for a desired Display ....8-28
i
Table of Contents
Page 4
8.5 Potentiometer Input, programming the meter
for a desired Display............................................................8-40
SECTION 9 METER FUNCTION MENUS
9.1 Individual Lockout Information...............................................9-1
9.2 Meter Function Menus...........................................................9-7
9.2.1 Input.......................................................................................9-7
9.2.2 RDG.CNF (Reading Configuration) .......................................9-7
9.2.3 RDG SC (Reading Scale) and RDG OF (Reading Offset).....9-9
9.2.4 IN CNF (Input Configuration) ...............................................9-11
9.2.5 IN.SC.OF (Input Scale and Offset) ......................................9-12
9.2.6 DEC PT (Decimal Point)......................................................9-14
9.2.7 CNT BY (Count By) .............................................................9-15
9.2.8 FIL.CNF (Filter Configuration) .............................................9-16
9.2.9 FIL TI (Filter Time Constant)................................................9-17
9.2.10 SP CNF (Setpoints 1 & 2 Configuration) .............................9-18
9.2.11 AL CNF (Alarm Configuration).............................................9-19
9.2.12 AL FNC (Alarm Function) ....................................................9-21
9.2.13 AL RDG (Alarm Readings) ..................................................9-22
9.2.14 SP DB (Setpoint Deadband)................................................9-23
9.2.15 AL DB (Alarm Deadband)....................................................9-23
9.2.16 OUT.CNF (Output Configuration).........................................9-24
9.2.17 OT.SC.OF (Output Scale and Offset) ..................................9-25
9.2.18 BAUD (Baud Rate) ..............................................................9-26
9.2.19 SERCNF (Serial Communication Configuration).................9-27
9.2.20 ADDRESS (Multipoint Communications Device Address)...9-28
9.2.21 DAT FT (Data Format) .........................................................9-28
9.2.22 BUS FT (Bus Format)..........................................................9-30
9.2.23 SERCNT (Serial Count).......................................................9-31
9.2.24 Analog Output Calibration Numbers....................................9-32
SECTION 10 SETPOINTS/ALARMS
10.1 Features overview ...............................................................10-1
10.2 Unlocking the Features........................................................10-3
10.3 Selecting “SP CNF” Setpoint Configuration Features..........10-3
10.4 Deviation Function for Alarms..............................................10-4
10.5 Selecting “AL CNF” Alarm Configuration Features ..............10-5
10.6 Selecting “AL FNC” Alarm Configuration Features ..............10-7
10.7 “AL RDG”: Alarm Readings-Select Delay in Alarm Action ...10-8
10.8 “SP DB”: Select “SP 1”and “SP 2” Deadband (Hysteresis) .10-8
10.9 “AL DB”: Select Alarm 1 (“SP 3”)
and Alarm 2 (“SP 4”) Deadband ( Hysteresis) .....................10-8
10.10 “OUT.CNF”: Control Flashing of the Display........................10-9
10.11 Entering Setpoint Levels (in Run Mode)..............................10-9
Table of Contents
ii
Page 5
SECTION 11 PEAK AND VALLEY READINGS .......................................11-1
SECTION 12 ANALOG OUTPUT OPTION
12.1 Features Overview...............................................................12-1
12.2 Unlocking.............................................................................12-2
12.3 “OUT.CNF”: Configuring the Output.....................................12-2
12.4 “OT.SC.OF”: Setting Output Scale and Offset .....................12-2
12.5 Board Installation; Entering the trim Data............................12-3
12.6 Filter Configuration “FIL.CNF” Value to be transmitted .......12-5
12.7 Wiring/Connections..............................................................12-5
SECTION 13 BCD OPTION
13.1 Features Overview...............................................................13-1
13.2 BCD Card Jumper Table......................................................13-3
13.3 Interconnect Board ..............................................................13-3
13.4 50-Line Cable Compatibility.................................................13-4
13.5 Selecting the source of BCD Data: “OUT.CNF” ...................13-4
13.6 Hold Control.........................................................................13-4
13.7 Data Ready Timing Pulses ..................................................13-4
13.8 Bringing out the BCD Overflow line .....................................13-4
13.9 3 Digit at a time Multiplex.....................................................13-4
13.10 6 Digit at a time Card Address .............................................13-5
13.11 Select Data Polarity: Jumper S8..........................................13-6
13.12 Decimal Point Address Code ...............................................13-6
13.13 Applying Non-Isolated/Isolated Power.................................13-6
13.14 Driving a Printer...................................................................13-6
SECTION 14 RELAY OPTIONS
14.1 Features Overview...............................................................14-1
14.2 Wiring/Connections..............................................................14-3
SECTION 15 RS-232 OR RS-485 OPTION BOARD
15.1 Features Overview...............................................................15-1
15.2 Front-Panel Pushbutton Configuration ................................15-3
SECTION 16 EXTERNAL CONTROL LINES
16.1 TARE (PIN 1) .......................................................................16-1
16.2 PEAK (PIN 2).......................................................................16-1
16.3 VALLEY (PIN 3) ...................................................................16-1
16.4 SWLIN2 (PIN 4)...................................................................16-1
16.5 PEAK & VALLEY or EXTERNAL RESET (PIN 5) ................16-1
16.6 PUSH TO CAL (PIN 6).........................................................16-2
16.7 Digital Return (PIN 7)...........................................................16-2
16.8 +5 V (PIN 8).........................................................................16-2
16.9 Display Hold (PIN 9) ............................................................16-2
16.10 LOCKOUT EEPROM (and ‘MENU’ BUTTON) (PIN 10) ......16-2
iii
Table of Contents
Page 6
16.11 Print Command and/or Reset of Alarms (PIN 11)................16-2
16.12 Nonstandard RX (PIN 12) and Nonstandard TX (PIN 13) ...16-2
16.13 PUSH TO CAL (PIN 14).......................................................16-3
16.14 +V EXT (PIN 15)..................................................................16-3
16.15 SP1 (PIN 16)........................................................................16-3
16.16 SP2 (PIN 17)........................................................................16-3
16.17 AL1 (PIN 18)........................................................................16-3
16.18 AL2 (PIN 19)........................................................................16-3
16.19 RTN EXT (PIN 20) ...............................................................16-3
SECTION 17 DISPLAY MESSAGES AND TROUBLESHOOTING GUIDE
17.1 Error Mode Message ...........................................................17-1
17.1.1 Flashing “999999” (Numerical Overflow).............................17-1
17.1.2 Flashing “ERR O1” (Offset Overflow) ..................................17-1
17.1.3 Flashing “ERR O2” (Setpoint Overflow)...............................17-1
17.1.4 “NOSTOR” & “STORED
(Programming Entries In EEPROM)....................................17-1
17.1.5 Flashing “+OVLD” (Positive Input Overload) .......................17-1
17.1.6 Flashing “+OPEN” (Open Sensor Indication) ......................17-1
17.1.7 Flashing “-OPEN” (Open Sensor Indication) .......................17-2
17.1.8 Flashing “I OVSC” (Input Overscale)...................................17-2
17.1.9 Flashing “R OVSC” (Reading Overscale)............................17-2
17.1.10 Flashing “CB OVF” (Count By Overflow).............................17-2
17.1.11 Flashing “UOM.OVF” (Unit Of Measure Overflow) ..............17-2
17.2 Troubleshooting Guide ........................................................17-2
SECTION 18 SPECIFICATIONS
18.1 Current Input........................................................................18-1
18.2 Voltage Input........................................................................18-1
18.3 Potentiometer Input .............................................................18-1
18.4 General................................................................................18-1
SECTION 19 FACTORY PRESET VALUES .............................................19-1
SECTION 20 RECORD YOUR SETUP VALUES......................................20-1
Table of Contents
iv
Page 7
Figure 4-1 Front Detail .........................................................................4-1
Figure 4-2A AC Power - Connector Label for Rear Connectors ............4-3
Figure 4-2B DC Power - Connector Label for Rear Connectors ............4-4
Figure 4-3 Rear View with the optional 4-relay output board
and a serial communications board installed. ....................4-5
Figure 4-4 Rear View with the optional BCD output board
and a serial communications output board installed. .........4-5
Figure 5-1 Meter Exploded View..........................................................5-2
Figure 5-2 Board Assembly Removing/Installing Detail.......................5-3
Figure 5-3 Transformer Jumpers .........................................................5-5
Figure 5-4 Optional Printed Circuit Board Locations ...........................5-6
Figure 5-5 Signal Input Board ..............................................................5-7
Figure 5-6 ac-Powered Main Board .....................................................5-8
Figure 5-7 dc-Powered Main Board .....................................................5-8
Figure 5-8 Relay Option Board ............................................................5-9
Figure 5-9 4-Relay Option Board .........................................................5-9
Figure 5-10 Analog Output Option Board ............................................5-10
Figure 5-11 RS-232 Option Board .......................................................5-10
Figure 5-12 RS-485 Option Board .......................................................5-10
Figure 5-13 BCD Option Board ............................................................5-11
Figure 5-14 Panel Mounting Assembly ................................................5-14
Figure 6-1 S1 and S2 Jumper Locations on Signal Input Board..........6-1
Figure 6-2 ac Main Board Jumper Positions S3 and S4 ......................6-4
Figure 6-3 dc Main Board Jumper Positions S3 and S4 ......................6-5
Figure 7-1 Current Input Without Sensor Excitation ............................7-1
Figure 7-2 Current Input With Sensor Excitation .................................7-1
Figure 7-3 Voltage Input Without Sensor Excitation ............................7-2
Figure 7-4 3-Wire Voltage Input With Sensor Excitation......................7-2
Figure 7-5 4-Wire Voltage Input With Sensor Excitation......................7-3
Figure 7-6 Bridge Input with External Sensor Excitation .....................7-3
Figure 7-7 Potentiometer Connections with Internal Power Supply
and Ratio Measurement. ....................................................7-3
Figure 7-8 Potentiometer Connections With External Power Supply
and Ratio Measurement (Remove jumper S2-T) ...............7-3
v
Figures
Page 8
Figure 7-9 AC Connector Wiring at P1 ................................................7-4
Figure 7-10 DC Connector Wiring at P1 ................................................7-4
Figure 8-1 S3 and S4 Main Board Jumper Positions...........................8-2
Figure 8-2 S1 and S2 Signal Input Jumper Positions ..........................8-3
Figure 8-3 Internally-Powered Bridge Input .........................................8-4
Figure 8-4 S3 and S4 Main Board Jumper Positions.........................8-16
Figure 8-5 S1 and S2 Signal Input Jumper Positions ........................8-16
Figure 8-6 Sensor Input Connections ................................................8-17
Figure 8-7 S3 and S4 Main Board Jumper Positions.........................8-28
Figure 8-8 S1 and S2 Signal Input Jumper Positions ........................8-29
Figure 8-9 Sensor Input Connections ................................................8-29
Figure 8-10 S3 and S4 Main Board Jumper Positions.........................8-41
Figure 8-11 S1 and S2 Signal Input Jumper Positions ........................8-42
Figure 8-12 Sensor Input Connections ................................................8-42
Figure 10-1 Setpoints 1 & 2 Action ......................................................10-2
Figure 10-2 Setpoints 3 & 4 Action ......................................................10-2
Figure 10-3 Process Deviation ............................................................10-4
Figure 10-4 High Deviation for both Active Above and Active Below...10-4
Figure 10-5 Low Deviation for both Active Above and Active Below....10-5
Figure 10-6 Band Deviation for both Active Above and Active Below..10-5
Figure 10-7 AL CNF Hysteresis ...........................................................10-6
Figure 12-1 Analog Option Board and Connection Diagram at P5. .....12-3
Figure 12-2 Isolated Analog Output Board Wiring Connections. .........12-5
Figure 13-1 BCD 40-Pin Cable Connector (P8)...................................13-1
Figure 13-2 BCD Option Board............................................................13-2
Figure 13-3 Address Programming Chart for 4 -line Address..............13-5
Figure 14-1 Dual Relay ........................................................................14-1
Figure 14-2 4 Relay Board Jumpers and Plugs ...................................14-2
Figure 14-3 Dual Relay Output Board Wiring Connections .................14-3
Figure 14-4 4 Relay Output Board Wiring Connections.......................14-3
Figure 15-1 RS-232/RS-485 Option Board and Pin Designations .......15-1
Figure 15-2a Older RS-232 Option Board and Pin Designations ..........15-1
Figures
vi
Page 9
Figure 15-2b Older RS-485 Option Board and Pin Designations ..........15-1
Figure 16-1 Connector Label Detail .....................................................16-1
Figure 16-2 Connection of External Power for Setpoint Transistors....16-3
Figure 18-1 Meter Housing and Panel Cutout. ....................................18-4
Table 4-1 Rear Connector Descriptions .............................................4-5
Table 8-1 Meter’s Natural Gain Chart.................................................8-5
Table 8-2 Reading Scale (RDG SC) vs. Decimal Point Shift............8-13
Table 8-3 Reading Scale (RDG SC) vs. Decimal Point Shift............8-26
Table 8-4 Reading Scale (RDG SC) vs. Decimal Point Shift............8-39
Table 8-5 Meter’s Natural Gain Chart...............................................8-43
Table 8-6 Reading Scale (RDG SC) vs. Decimal Point Shift............8-52
Table 14-1 Dual Relay Board Jumpers ..............................................14-1
Table 14-2 4 Relay Board Jumpers....................................................14-2
Table 14-3 Pin assignments for the P6, P7 and P18 plugs................14-3
vii
Figures
Page 10
Information that is especially important to note is identified by these labels:
NOTE
WARNING
CAUTION
IMPORTANT
NOTE: provides you with information that is important to successfully setup
and use the Programmable Digital Meter.
CAUTION or WARNING: tells you about the risk of electric shock.
CAUTION, WARNING or IMPORTANT: tells you of circumstances or
practices that can effect the meter's functionality and must refer to accompanying documents.
Notes, Warnings, and Cautions
viii
Note
Page 11
1.1 DESCRIPTION
This Strain Gage meter is part of a complete line of process indicators/ controllers, offering exceptional performance.
The strain gage meter is front panel programmable to accept unipolar and bipolar DC voltage inputs and 0-20 or 4-20 mAdc inputs. The meter will accept inputs from most of the process sensors in use today such as transmitters, pressure transducers, and potentiometers.
Configuring the meter is accomplished through the 5 front panel buttons. If the RS-232 or RS-485 communication option is installed, the user may remotely set the display parameters.
Options for the meter include analog and BCD outputs, relay outputs, and RS-232 or RS-485 communications.
1.2 FEATURES
The following is a list of features of the meter.
• 6-digit LED display in red or green
• 0.005% accuracy of reading
• 12 DC input ranges: 0-100 mV, 0-1 V, 0-5 V, 1-5 V, 0-10 V, 0-100 V, ±50 mV, ±500 mV, ±5 V, ±50 V, 0-20 mA, or 4-20 mA
• 1.5 to 11 or 24 V dc sensor excitation
• Peak & Valley detection and memory
• Tare
• Up to 13 readings per second
• 4 isolated open collector outputs (standard) and optional isolated relay or isolated BCD and isolated analog outputs
• Optional isolated RS-232 or RS-485 communications
• NEMA 4 Front Panel/IP65
• Non-volatile memory without battery back-up
• 115 Vac, 50/60 Hz, 230 Vac, 50/60 Hz power supply
• Optional 10-32 Vdc power supply
1-1
Introduction
1
Page 12
1.3 AVAILABLE MODELS
The following models and options are available. Optional boards are either installed at the time of purchase, or available as separate items and installed by the user after purchase.
MAIN ASSEMBLIES
MODEL NUMBER DESCRIPTION
DP41-S Red LEDs, 115 V ac, 50/60Hz DP41-S-230 Red LEDs, 230 V ac, 50/60Hz DP41-S-GN Green LED’s, 115 V ac, 50/60 Hz DP41-S-230-GN Green LED’s, 230 V ac, 50/60Hz DP41-S-DC Red LEDs, 10-32 V dc DP41-S-DC-GN Green LED’s, 10-32 V dc
NOTE: The following options are available installed at the time of purchase or as separate items installed by the user after purchase:
Analog Output Board, BCD Output Board, Relay Output Board, RS-232 Communications Board, and RS-485 Communications Board.
CONTROL/BCD OUTPUT OPTIONS
MODEL NUMBER DESCRIPTION
- Standard four open-collector outputs are standard DP40-B Isolated BCD Output Board DP40-R Dual 7A Form-C Relays DP40-R4 Dual 7A & Dual 1A Form-C Relays
NOTE: Choose only one Control/BCD output option per meter. A 40-pin mating connector is included with the BCD option.
ANALOG OUTPUT
MODEL NUMBER DESCRIPTION
- None DP40-A Isolated configurable analog (4-20 mA, 0-1, 0-5, 1-5,
0-10 V dc, 0-20 mA) output
1
Introduction
1-2
Note
Note
Page 13
SERIAL COMMUNICATIONS OPTION
MODEL NUMBER DESCRIPTION
DP40-S2 * Isolated RS-232 Communications DP40-S4 ** Isolated RS-485 Communications
NOTES: Choose only one option per meter. Both computer communications comes with 6 ft. communications cable with phone plug termination.
Free configuration software is available from www.omega.com or on the CD-ROM enclosed with your shipment.
* Recommend purchase of DP40-9SC2 or DP40-25SC2
(see OPTIONS below)
** Recommend purchase of DP40-9SC4 or DP40-25SC4
(see OPTIONS below)
OPTIONS
MODEL NUMBER DESCRIPTION
DP40-9SC2 9-pin Serial Connector for RS-232 DP40-9SC4 9-pin Serial Connector for RS-485 DP40-25SC2 25-pin Serial Connector for RS-232 DP40-25SC4 25-pin Serial Connector for RS-485 SPC4 1/8 DIN NEMA 4 heavy duty cover with thumb screws SPC18 1/8 DIN NEMA 4 Splash Proof lens cover with spring clip
1-3
Introduction
1
Note
Page 14
Unpack all items and make sure that every item on the packing list is present. The items you should receive are listed below. If something is missing, use the phone number for the Customer Service Department nearest you.
Also, inspect the shipping container and enclosed equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
NOTE: The shipping agent will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save all packing material and containers in the event that reshipment is required.
When you ordered your process meter, you will receive the following items in the shipping box:
QTY
DESCRIPTION ILLUSTRATION
1 Basic Meter in a Mounting
Sleeve with Gasket
1 Front-Panel Button Cover
AC Power Connector (orange - P1)
2 Input Connectors
(gray - P3 and P9)
2
Unpacking
2-1
1 2 3
P1
N
L
1 2 3
Note
Page 15
1 Rear Protective
Cover with Screw
1 20-Socket Ribbon
Connector (P2 Connector)
2 Panel-Mounting
Gaskets (1 Spare)
1 Strain Gage Owner’s Guide
1 Quick Start Manual
Other items may also be in the box depending on the options ordered. Refer to specific options described previously.
2-2
Unpacking
2
Page 16
3.1 SAFETY CONSIDERATIONS
This device is marked with the international caution symbol. It is important to read this manual before installing or commissioning this device as it contains important information relating to Safety and EMC
(Electromagnetic Compatibility).
Unpacking & Inspection
Unpack the instrument and inspect for obvious shipping damage. Do not attempt to operate the unit if damage is found.
This instrument is a panel mount device protected in accordance with Class I of EN 61010 (115/230 AC power connections), Class III for the DC power option (10-32Vdc). Installation of this instrument should be done by Qualified personnel. In order to ensure safe operation, the following instructions should be followed.
This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947–1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the mains supply cord.
Furthermore, to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment, an overcurrent protection device shall be installed.
The Protective Conductor must be connected for safety reasons. Check that the power cable has the proper Earth wire, and it is properly connected. It is not safe to operate this unit without the Protective Conductor Terminal connected. Conductor Terminal connected.
Do not exceed voltage rating on the label located on the top of the instrument housing.
Always disconnect power before changing signal and power connections.
Do not use this instrument on a work bench without its case for safety reasons.
Do not operate this instrument in flammable or explosive atmospheres.
Do not expose this instrument to rain or moisture.
Unit mounting should allow for adequate ventilation to ensure instrument does not exceed operating temperature rating.
Use electrical wires with adequate size to handle mechanical strain and power requirements. Install without exposing bare wire outside the connector to minimize electrical shock hazards.
EMC Considerations
Whenever EMC is an issue, always use shielded cables.
Never run signal and power wires in the same conduit.
Use signal wire connections with twisted-pair cables.
Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.
Failure to follow all instructions and warnings may result in injury!
Note
3
Safety Considerations
3-1
Note
Page 17
4.1 FRONT OF THE METER
The following is a brief description of each part of the front of the meter.
Figure 4-1. Front Detail
ITEM DESCRIPTION
1 –.8.8.8.8.8. or 8.8.8.8.8.8.
6-digit, 14 segment, alphanumeric 0.54” high LED display with programmable decimal point.
2 SETPOINT LED
These LEDs, labeled 1 through 4, display the status of setpoints 1, 2, 3 (Alarm 1), and 4 (Alarm 2).
3 SETPTS BUTTON
This button functions only in the run mode. When the Setpoint/Alarm features are unlocked, pressing this button sequentially recalls the previous setpoint settings to the display. After the ‘MIN’ and ‘MAX’ buttons are used to alter those values as desired, pressing the ‘SETPTS’ button, again, stores these new values.
Unless the ‘SETPTS’ button is pressed, each of the four setpoint values is displayed for approximately 10 seconds after the last press of the ‘SETPTS’ button. Holding the ‘SETPTS’ button depressed stalls this automatic sequence, retaining the most recent setpoint number on the display.
SETPTS
MAX
MIN
MENU RESET
1234
2 2 2
3 4
5
6
7
2
1
4-1
Parts of the Meter
4
Page 18
ITEM DESCRIPTION
4 /MAX BUTTON
During run mode, pressing this button displays the “HI RDG” (peak reading) value that has occurred up to the moment the
‘MAX’ button is pressed. This peak reading flashes, to
distinguish its value from the current readings. Since this is a dynamic peak reading, the value will change if the value increases while reviewing it.
To return to display of the current readings without resetting the peak-value memory to zero (0), press the ‘MAX’ button once again.
To reset the peak-value memory (start a new peak determining period), press the ‘RESET’ button once.
During the configuration mode, the ‘MAX’ button is used to change the numerical value of the flashing digit displayed. For submenu items, such as “L1C.1=0”, pressing the ‘MAX’ button toggles the choice from “0” to “1”.
The meter allows rapid changes of a displayed numerical value by making “0” the first value to occur when the ‘MAX’ button is pressed. After that, the numbers increase to “9” and then roll over to “0” again. A negative (“–”) symbol may be displayed in the most significant digit (i.e. the digit at the far left such as shown here “–.8.8.8.8.8.”)
In the SETPT (SETPOINT) mode (SP1, SP2, etc), pressing the ‘MAX’ button causes the flashing digit to increment by 1 from 0 to 9.
5
/MIN BUTTON
During run mode, pressing the ‘MIN’ button recalls the “LO RDG” (valley reading) measured since the last press of the ‘RESET’ button. This lowest value flashes, to distinguish it from the current process display. Since this is a dynamic valley reading, the value will change if the value decreases while reviewing it.
4
Parts of the Meter
4-1
Page 19
ITEM DESCRIPTION
To return to the actual process display, without resetting the low­value memory, press the ‘MIN’ button once again.
To reset the memory for current-value readings (start a new observation period), press the ‘RESET’ button once and the meter will return to the run mode.
In the configuration mode, once in a submenu (like input type) the ‘MIN’ button allows you scroll through the available choices such as, 0-20 mA or 4-20 mA, etc.
In the SETPT mode, the ‘MIN’ button advances the flashing digit to the right.
6 MENU BUTTON
In the run mode, this button terminates the measurement process and allows you to enter the configuration mode, advancing through the configuration menus.
In the configuration mode, this button will store changes in the non-volatile memory at the same time advancing the display to the next menu item.
7 RESET BUTTON
In the run mode, pressing the ‘RESET’ button once erases the memories of peak (“HI RDG”), valley (“LO RDG”), and ALARM latches. The display then returns to the run mode.
WARNING: Pressing the ‘RESET’ button two (2) times will result in a hard reset of the meter. This will clear the Peak & Valley, Alarm latches and meter reading and immediately begin a new measurement.
In the configuration mode, pressing the ‘RESET’ button once, displays the previous selection. For example, if you were in “IN CNF” then pressed the ‘RESET’ button once, the display will then show “RD.SC.OF”. Press the ‘RESET’ button two times to return to the run mode.
4-2
Parts of the Meter
4
Page 20
4.2 REAR OF THE METER
The following is a brief description of each part of the rear of the meter. The label on the top of the mounting sleeve (not the case) identifies the location of the connectors found at the rear of the meter. Figure 4-2A & figure 4-2B shows this label.
Figure 4-2A. AC Power - Connector Label for Rear Connectors
P2 / CABLE CONNECTOR
TARE(T) 1
VALLEY(V) 3
DIG RTN 7
PEAK/VALLEY OR EXT RESET 5
DISPLAY HOLD 9
RESET ALARMS AND/OR 9
PRINT COMMAND 11
NONSTANDARD TX 13
+V EXT 15
SP2 17
AL2 19
2 PEAK(P) 4 SWLIN2 6 PUSH TO CAL 8 +5V
LOCKOUT EEPROM & MENU PUSH BUTTON
12 NONSTANDARD RX 14 PUSH TO CAL 16 SP1 18 AL1 20 RTN EXT
P4 / RS232
6 N/C 5 COMM RTN 4 RX 3 TX 2 RTS 1 N/C
P4 / RS485
6 N/C 5 B,-RX 4 A, +RX 3 COMM RTN 2 B, -TX/RX 1 A, +TX/RX
P5 / ANALOG OUT
1 RETURN 2 4-20MA 3 0-10V
P6 / RELAY
P7 / RELAY
P7
P1
P2
P3
P4
P5
P6
P9
1 2 3
1 2 3
NO1
CM1
NC1
NO2
CM2
NC2
1 2 3 4 5 6
* P18 / 4RELAY
NO3
CM3
NC3
NO4
CM4
NC4
+E +S
- S
- E +R
- R
CONNECT TO LOW VOLTAGE LIMITED ENERGY CIRCUITRY ONLY.
LINE L
NUETRAL N
GND
P1 / AC
POWER
P8 / BCD (see manual for pinout) P9&P3 / SIGNAL IN
1
1
1
1
1
1
1
1
P8 or P16
19
* NOT AVAILBALE
WITH=DC POWER
10
11
4
Parts of the Meter
4-3
Page 21
4-4
Parts of the Meter
4
Figure 4-2B. DC Power - Connector Label for Rear Connectors
P2 / CABLE CONNECTOR
TARE(T) 1
VALLEY(V) 3
DIG RTN 7
PEAK/VALLEY OR EXT RESET 5
DISPLAY HOLD 9
RESET ALARMS AND/OR 9
PRINT COMMAND 11
NONSTANDARD TX 13
+V EXT 15
SP2 17
AL2 19
2 PEAK(P) 4 SWLIN2 6 PUSH TO CAL 8 +5V
LOCKOUT EEPROM & MENU PUSH BUTTON
12 NONSTANDARD RX 14 PUSH TO CAL 16 SP1 18 AL1 20 RTN EXT
P4 / RS232
6 N/C 5 COMM RTN 4 RX 3 TX 2 RTS 1 N/C
P4 / RS485
6 N/C 5 B,-RX 4 A, +RX 3 COMM RTN 2 B, -TX/RX 1 A, +TX/RX
P5 / ANALOG OUT
1 RETURN 2 4-20MA 3 0-10V
P6 / RELAY
P7 / RELAY
P7
P1
P2
P3
P4
P5
P6
P9
1 2 3
1 2 3
NO1
CM1
NC1
NO2
CM2
NC2
1 2 3 4 5 6
* P18 / 4RELAY
NO3
CM3
NC3
NO4
CM4
NC4
+E +S
- S
- E +R
- R
CONNECT TO LOW VOLTAGE LIMITED ENERGY CIRCUITRY ONLY.
N/C
+
P1 / DC
POWER
P8 / BCD (see manual for pinout) P9&P3 / SIGNAL IN
1
1
1
1
1
1
1
1
P8 or P16
19
* NOT AVAILBALE
WITH=DC POWER
10
11
Page 22
Figure 4-3 shows the rear of the meter with the optional 4-relay output board and a serial communications board installed.
Figure 4-3. Rear View
Figure 4-4 shows the rear of the meter with the optional BCD output board and a serial communications output board installed.
Figure 4-4. Rear View
CONNECTOR # DESCRIPTION
P1 AC Power Connector P2 External I/O Connector P3 Input Connector, –E, +R, –R J4 Optional RS-232 or RS-485 Phone Jack Connector P5 Optional Analog Out Connector P6 Optional Form-C Relay #1 Connector P7 Optional Form-C Relay #2 Connector P8 Optional BCD Connector P9 Input Connector, +E, +S, –S P18 Optional Form-C Relay #3 and #4 Connector
TABLE 4-1 Rear Connector Descriptions
1 2 3
P1
N
1
20
P2
1 2 3
P5
1 2 3
P3
P8
U1
L1
U20
L20
1 2 3
P9
J4
L
1 2 3
P1
N
L
P2
P6
P7
J4
20
1
1
2
3
1
2
3
1 2 3
P3
1 2 3
P9
P18
1
2
3
4
5
6
4
Parts of the Meter
4-5
Page 23
5.1 CONDITIONS REQUIRING DISASSEMBLY
You may need to remove the sleeve or open the meter for several reasons:
1. To inspect the rating label on the case (not the same label as on the sleeve) (Section 5.2.1).
2. To check or change the 115 V ac or 230 V ac or main board jumpers (Sections 5.2.2 and 5.2.4).
3. To install optional output board(s) (Section 5.2.3).
4. To mount the meter in a panel (Section 5.2.5).
5.2 ASSEMBLY/DISASSEMBLY OPENING
OPENING THE METER
Your meter is fully assembled, but not wired. See Section 7 for wiring connection for power and sensor inputs. In most cases, if you have ordered optional boards with the meter, these boards will already be installed.
You will need to remove only the rear cover to complete wiring, but you will have to open the meter to do one or more of the following:
WARNING!: You must disconnect and turn-off the power and connector signals before accessing the unit and installing optional boards. Failure to do so may result in injury!
a. Check or reconfigure the Transformer Jumpers on the Main Board so
that they correspond to your line voltage (W1 and W2 for 115 V ac, or W3 for 230 V ac). See Section 5.2.2.
b. Install optional boards. See Section 5.2.3.
c. Access jumpers on the Main and optional boards. See Section 5.2.4.
5-1
Setup
5
Page 24
Figure 5-1. Meter Exploded View
Using Figures 5-1 as a guide, follow these simple instructions to open the meter:
IMPORTANT: Turn-off the power and input signals from the unit before proceeding. Failure to do so may result in injury!
1. Remove the cover mounting screw that secures the rear protective cover to the meter, and remove the Rear Protective Cover.
If you are simply wiring the meter–but not checking jumpers or installing or removing boards–this is as far as the meter needs to be disassembled. Go to Section 5.2.1.
2. Remove all wiring connectors from the rear of the meter.
3. Remove the two thumbnuts that secure the case to the sleeve.
4. Remove the sleeve completely by sliding it back from the front bezel.
REAR PROTECTIVE COVER
COVER MOUNTING SCREW
SLEEVE
AC POWER BOARD
MAIN BOARD ASSY
CASE
BEZEL
GASKET
THUMBNUTS
5
Setup
5-2
Page 25
5. Using Figure 5-2 as a guide, bend the side-panel detents on the case outward and pull the board assembly out of the case by the mounting screw stem.
NOTE: From this point forward, these 5 steps are referred to as “Reveal the main board”.
Figure 5-2. Board Assembly Removing/Installing Detail
The meter is now disassembled to the point where you can check and configure jumpers and install boards.
REINSERTING THE MAIN BOARD ASSEMBLY INTO THE CASE
Reinsert the Main Board into the case once jumpers and optional boards have been configured and installed.
Spread the side-panel detents of the case, and carefully slide the main board all the way in.
MAIN BOARD ASSY
CASE
BEND DETENTS OUTWARD TO INSTALL MAIN BOARD
SIGNAL INPUT JUMPER LABEL
PRODUCT IDENTIFICATION LABEL
MOUNTING SCREW STEM
TO INSTALL
TO REMOVE
5-3
Setup
5
Note
Page 26
5.2.1 Safety Precaution/Product ID Label
To look at the Rating/Product ID label on the case, you must follow the first step as described in Section 5-2. Refer to Figure 5-2 for the location of the Product Identification label.
The meter is protected in accordance with Class I of EN61010. Refer to Safety Considerations page.
WARNING: If your meter is to be wired to sensors to control inputs that could be hazardous potentials, these potentials will be carried to the 20-pin output connector (P2) at the rear. They will also be present on the meter’s circuit boards. Follow all instructions carefully BEFORE connecting the meter to any source of power.
DO NOT contact any exposed metal parts, install optional board(s), change jumpers, or in any way disassemble or assemble the meter while it is connected to AC voltage.
Note the following information and guidelines for safe operation of your meter:
Power Voltage
Your power source voltage rating should agree with the voltage under which the meter has been configured to operate. The first thing you should do is verify this.
The meter’s operating voltage is shown in the VOLTS: entry of the Product Identification and Serial Number Label. It is located on the case, as shown in Figure 5-2, and is clearly visible on the meter packing box.
5
Setup
5-4
Page 27
5.2.2 Main Board Power Jumpers
To check voltage jumpers or to change from 115 V ac to 230 V ac:
CAUTION: The meter has no power-on switch; it will be in operation as soon as you apply power. To change the factory preset jumpers, disconnect the power from the unit. Failure to do so may result in injury! The jumpers must be changed by specifically trained personnel.
1. “Reveal the Main Board” (refer to Section 5.2, Disassembly).
2. Locate the main board assembly and position it in front of you the same way as shown in Figure 5-3.
3. On the main board, locate the transformer jumpers W1, W2, and W3 near the transformer T1. If your power requirement is 115 V ac, jumpers W1 and W2 should be installed. (DO NOT INSTALL W3)
If your power requirement is 230 V ac, jumper W3 should be installed. (DO NOT INSTALL W1 OR W2)
Figure 5-3. Transformer Jumpers
W2
W3
W1
AC POWER BOARD
MAIN BOARD
T1
REAR OF METER
5-5
Setup
5
Page 28
5
Setup
5-6
5.2.3 Printed Circuit Board(s) Installation
To install optional printed circuit board(s):
1. “Reveal the Main Board” (refer to Section 5.2, Disassembly).
2. Using Figure 5-4 as a reference, insert option board(s) into the corresponding slot(s) on the main board. Each circuit board is keyed to fit in it’s own position.
Figure 5-4. Optional Printed Circuit Board Locations
BCD BOARD OR 4 RELAY BOARD
INTERCONNECT BOARD (PART OF BCD ASSY)
REAR PROTECTIVE COVER
COVER MOUNTING SCREW
DUAL RELAY BOARD
RETAINER (ALWAYS USED EXCEPT FOR BCD OR 4 RELAY OPTION)
THUMBNUTS
RS-232/RS-485 BOARD
ISOLATED ANALOG
OUTPUT BOARD
SLEEVE
GASKET
MAIN BOARD ASSY
BEND DETENTS OUTWARD TO INSTALL MAIN BOARD
CASE
J20
P20
P14
P12
P11
J12 J11
J10
J14
P10
AC POWER BOARD
(4 RELAY BOARD ONLY)
P10
Page 29
5-7
Setup
5
5.2.4 How to Access Jumpers
To gain access to jumper S1 and S2 used to configure input type remove the mounting sleeve. The jumpers may be accessed through the slot in the case.
To gain access to jumpers on the main board for power, excitation and lockout selection:
1. “Reveal the main board” (refer to Section 5.2, Disassembly).
NOTE: To access the S1 and S2 jumpers on the Signal Input Board, you only need to remove the mounting sleeve.
2. To re-assemble the meter, follow the steps in reverse order.
Figures 5-5 through 5-13 show the layout of the seven (7) printed circuit boards with respective jumper blocks, where applicable, used in the meter. Figures 5-7 through 5-13 show the optional boards.
Figure 5-5. Signal Input Board
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
TOP VIEWS
S2
J3
J9
S1
R135
Note
Page 30
5
Setup
5-8
Figure 5-6. ac-Powered Main Board
Figure 5-7. dc-Powered Main Board
DISPLAY
BD
A
C
D
S3
SIGNAL INPUT BD
MAIN BD
J9
J2
J1
J3
AC POWER BD
W1
W3
W2
S4
T1
B
C
A
B
R39
SEE
DETAIL A
PINS FOR RS 485
ALL PINS FOR RS232A
DETAIL A
PINS FOR RELAY
PINS FOR
ANALOG OUT
S
3
S
3
A B C D
J12
J11
S4
R34
S5
S4
S5
J10
1A
J2
A
1A
1
11
1
10
A
AA
B
S
1
Page 31
5-9
Setup
5
Figure 5-8. Relay Option Board
Figure 5-9. 4-Relay Option Board
NOTE: Both the Dual Relay Output Board and the 4 Relay Output Board can be installed at J10. However, only one option board can be installed at a time.
4 RELAY BOARD
PIN 1 OF CABLE
PIN 1
MAIN BOARD
J10
AB
C
D
E
S1
P6
P7
P10
PIN 1
Note
Page 32
5
Setup
5-10
Figure 5-10. Analog Output Option Board
Figure 5-11. RS-232 Option Board
Figure 5-12. RS-485 Option Board
S1
A
B
S3
A
S2
A
J4
PIN 1
P11
1
6
S4
A
PIN 10
ABCDE
S
1
S1
A
B
S
3
P
1
1
S3
1
1
1
1
A
J
4
1
6
S
2
S2
A
S
4
S4
A
The RS-232 option board has been updated. The figure shown is the latest version.
Some older versions of the RS-232 cards are not compatible with dc-powered
meters.
Check the J11 connector on the main board to ensure it has a 12 position connector.
The RS-485 option board has been updated. The figure shown is an older version, the newer version is the same figure as the RS-232 option card shown above.
S1
A
1
0
J5
1
1
1A
S1
P12
The Analog option board has been updated. The figure shown is the latest version.
Some older versions of the analog output cards are not compatible with dc-powered
meters.
Check the J12 connector on the main
board to ensure it has a 10
position connector.
S1-A does not need to be installed.
Page 33
5-11
Setup
5
Figure 5-13. BCD Option Board
5.2.5 PANEL-MOUNT ASSEMBLY
The meter can be mounted in a panel so that the front of the meter is flush with the vertical panel surface. Panel mounting can be seen as simply “sandwiching” the panel between the inner case and the outer sleeve in the last phases of assembly. Figure 5-14 shows the panel cutout dimensions, and the dimensions for the panel thickness. It requires that the following already be done:
1. Your line voltage rating has been checked against the meter rating on the Product ID and serial number label on the meter case. See Section
5.2.1.
2. You have configured all jumpers–those on the main board as well as those on any optional boards. See Section 5.2.2 for main board jumper configuration and the appropriate sections for optional board jumper configuration.
3. You have installed all optional boards and inserted the main board assembly back into the case. See Section 5.2.3.
4. You have wired P1, the AC power connector, and P2 the input output control connector; connectors are not installed in the meter, but are ready to be installed. See Section 7.
5. You have wired all connectors for optional boards; connectors are not connected to the meter, but are ready to be installed.
If all of these steps are done, use Figure 5-14 as a guide:
CAUTION: Connectors with the wiring connections will be installed after mounting the unit.
S1
S2
S4
S5
A A
A B
A
C
E
G
S3
A B
S6
B
A
B
A
S7
S8
A
B
J20
PIN 1
L1
U1
J8
Page 34
5
Setup
5-12
Figure 5-14. Panel Mounting Assembly
6. Punch or cut a hole in the panel using the panel cutout dimensions in
Figure 5-14. Remove burrs and paint the panel as required.
7. Insert the panel-mount gasket around the rear of the case and slide it forward to the bezel (if it’s not already in place).
8. Working from the front of the panel, insert the case assembly, rear end first, all the way into the panel cutout such that the gasket firmly backs the panel surface.
9. Working from the rear of the panel, slide the sleeve forward over the case and up to the panel surface.
The panel should now be sandwiched between the bezel-backed gasket in front and the sleeve in back.
10. Replace the thumbnuts that secure the sleeve tabs to the case.
REAR PROTECTIVE COVER
COVER MOUNTING SCREW
THUMBNUTS
SLEEVE
GASKET
CASE
BEZEL
PANEL
METER
0.25 [6.4] MAX
0.03 [0.8] MIN
PANEL THICKNESS
3.622 +0.032/-.000 [92.00 +0.81/-0.00]
1.772 +0.024/-.000 [45.00 +0.61/-0.00]
0.06
[1.5]
4PLCS
R
Page 35
WARNING: Do not “turn-on” the ac power and input signals until all connections are connected to the meter.
11. Set P1, the AC power connector, aside and connect or reconnect all other connectors to the back of the meter using Figures 4-3 and 4-4 in Section 4.2 as guides.
Connect P1 last.
NOTE: The P1 connector is “keyed”; it is shaped in such a way that it fits only the J1 male pins.
12. Replace the rear protective cover and secure it with the cover mounting screw.
Your meter is now ready for operation and you can turn-on the power.
The meter display should light, and pass through “RESET 2” to run or display mode. If the meter flashes an overscale or overload message, press the ‘MENU’ button to advance to the configuration mode. Do not be concerned about overloads (the +S input can stand 120 V continuously and current inputs can handle ten times rated current).
5-13
Setup
5
Note
Page 36
6
Jumper Positions
6-1
6.1 INTRODUCTION
This section is for the configuration and setup of your jumper positions for readrate, unipolar or bipolar signal input, sensor input signal jumpers, sensor excitation jumpers, pushbutton lockouts and lockout of lockout configuration menus.
6.2 S1 JUMPER POSITIONS FOR READRATE AND UNIPOLAR OR BIPOLAR INPUT(S)
The typical readrate for your meter is 3/per second. This requires that no jumper has been installed in the S1A position and Input Configuration (“IN CNF”) bit “INP.2” has been set to equal “0”. Your meter is capable of a fast readrate of 13/per second. This requires that you install a jumper in the S1A position and the Input Configuration (“IN CNF”) bit “INP.2” has been set to equal “1”. Refer to Figure 6-1 for the location of the S1 jumpers.
The typical setting for your meter is unipolar. For unipolar input, no jumper is installed in the S1B position and Input Configuration (“IN CNF”) bit “INP.3” must be set to equal “0”. For bipolar inputs, install a jumper in S1B and set Input Configuration (“IN CNF”) bit “INP.3” to equal “1”.
6.3 S2 JUMPER POSITIONS FOR INPUT RANGES
The following are the input signal jumper positions required to be installed in the “S2” position on your meter for the CURRENT or VOLTAGE input ranges you require. These jumper positions include those that are required for sensor excitation. Jumpers S2-N & S2-T are for either 1.5 to 11 Vdc or 24 Vdc sensor excitation. To select desired excitation see Section 6.4. Refer to Figure 6-1 for the location of the S2 jumpers.
Figure 6-1. S1 and S2 Jumper Locations on Signal Input Board
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
TOP VIEWS
S2
J3
J9
S1
Page 37
6-2
Jumper Positions
6
BRIDGE - UNIPOLAR
BRIDGE - BIPOLAR
Voltage (Non-Ratiometric) - Unipolar/Bipolar similiar to Bridge inputs above except substitute jumper M for L.
CURRENT
POTENTIOMETER
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
Jumpers for 0 to 100 mV range: (meter supplied excitation)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
Jumpers for 0 to 1 V range: (meter supplied excitation)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
Jumpers for 0 to 10 V range: (meter supplied excitation)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
Jumpers for 0 to 100 V range: (meter supplied excitation)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
Jumpers for -50 to +50 mV range: (meter supplied excitation)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
Jumpers for -500 to +500 mV range: (meter supplied excitation)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
Jumpers for -5 to +5 V range: (meter supplied excitation)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
Jumpers for -50 to +50 V range: (meter supplied excitation)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
Jumpers for 0-20 mA or 4-20 mA: (meter supplied excitation)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
Jumpers for 0 to 10 V range: (using 10 Vdc drive)
NOTE: Remove S2T jumper
when you use external sensor excitation.
Page 38
6
Jumper Positions
6-3
6.4 JUMPER SETTING(S) FOR SENSOR EXCITATION
1). Maximum excitation for “Ratiometric” measurement is 10Vdc.
2). Remove S2T jumper when you use external sensor excitation.
6.4.1 JUMPER SETTING(S) FOR SENSOR EXCITATION- ac Powered Unit
Your ac-powered meter is capable of supplying either 1.5 to 11Vdc or 24 Vdc sensor excitation. (refer to Figure 6-2.).
For 1.5 to 11Vdc excitation, install S4A and C jumpers, then adjust the potentiometer (R39) for proper voltage.
For 24Vdc excitation, remove S4A and C jumpers and install S4B.
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
24 Vdc meter excitation (S2N, S2T, & S4B)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
10 Vdc meter excitation (S2N, S2T, & S4A, S4C)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
1.25 Vdc meter excitation (S2Q)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
160 µA meter excitation (S2P)
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
S1
1.6 mA meter excitation (S2P, S2V)
Note
Page 39
6-4
Jumper Positions
6
Figure 6-2. ac-Powered Main Board Jumper Positions S3 and S4
NOTE: S3B should NOT be installed. This jumper is only used when recalibrating the meter (e.g. an annual, careful performance by the calibration lab). When this jumper is installed, calibration coefficients can be changed via digital communications.
Note
DISPLAY
BD
A
C
D
S3
SIGNAL INPUT BD
MAIN BD
J9
J2
J1
J3
AC POWER BD
W1
W3
W2
S4
T1
B
C
A
B
R39
S4A shown in storage position
S4 On ac-powered units, main board
S4B Installed For 24 Vdc excitation.
(S4A located in storage position).
S4A ,S4C Installed For 1.5 to 10 Vdc excitation.
S3 On ac-powered units, main board
S3A Installed Unlocks “MENU” button for programming. S3B Omit See note below. S3C Installed Unlocks lockout menu (L1 through L4). S3D Installed Unlocks Front pushbuttons.
Page 40
6.4.2 JUMPER SETTING(S) FOR SENSOR EXCITATION - dc Powered Unit
Your dc-powered meter is capable of supplying either 1.25 to 12Vdc or 24 Vdc sensor excitation. (refer to Figure 6-3).
For 1.25 to 12Vdc excitation, install S1A and S4A jumpers, then adjust the potentiometer (R34) for proper voltage.
For 24Vdc excitation, remove S4A jumpers and install S1B.
Figure 6-3. dc-Powered Main Board Jumper Positions S3 and S4
6
Jumper Positions
SEE
DETAIL A
PINS FOR RELAY
PINS FOR
ANALOG OUT
S3
S3
A
B
C
D
J12
J1
1
S
4
R
3
4
S
5
S
4
S
5
J1
0
1A
J2
A
1
10
A
AA
B
S1
6-5
S1 On dc-powered unit, side power board
S1A Installed For 1.25 to 12 Vdc excitation at 120 mA
.
S1B Installed For 24 Vdc excitation at 35 mA.
S3, S4, S5 On dc-powered unit, main board
S3A Installed To store data and setup parameters in nonvolatile memory. S3B Omit See note in previous Section 6.4.1. S3C Installed Unlocks lockout menu (L1 through L4). S3D Installed Unlocks Front pushbuttons.
S4A Installed Along with the S1 jumper to program the excitation
output. Adjust excitation with R34 surface mount pot from 1.25 to 12 volts, with an output current up to 120mA.
S4A Removed For 24 Vdc excitation.
(S4A located in storage position).
S5A Installed To enable the RESET front panel pushbutton. S5A Removed To secure against unauthorized meter reset.
Page 41
7-1
Signal and Power Input Connections
7
7.1 INTRODUCTION
The following describes how to connect your sensors to your meter with and without sensor excitation and how to connect the AC power to your meter. Prior to wiring the sensor to the meter, check with a multimeter that a proper excitation exists.
WARNING: Do not connect ac power to your meter until you have completed all input and output connections. Failure to do so may result in injury! This device must only be installed electrically by specially trained electrician with corresponding qualifications.
7.2 SIGNAL INPUT CONNECTIONS
The following figures (7-1 through 7-5) show the connections for voltage, current and potentiometer inputs:
Figure 7-1. Current Input Without Sensor Excitation
Figure 7-2. Current Input With Sensor Excitation
NC
+
S
S
P9
1
2
3
1
2
3
P3
NC
+R
(+)
(-)
(4-20mA)
+
E
E
JUMPER USER PROVIDED
METER
-R
NC
+
S
S
P9
1
2
3
1
2
3
P3
NC
NC
NC
+ -
(4-20mA)
(+)
(
-)
METER
Page 42
7
Signal and Power Input Connections
7-2
Figure 7-3. Voltage Input Without Sensor Excitation
Figure 7-4. 3-Wire Voltage Input With Sensor Excitation
Figure 7-5 4-Wire Voltage/Bridge Input With Sensor Excitation
+
E
+
S
S
E
P9
P3
1
2
3
1
2
3
GND
VOLTAGE
+EXCITATION
+OUTPUT
-EXCITATION
-OUTPUT
METER
NC
+
E
+
S
S
E
P9
P3
1
2
3
1
2
3
GND
JUMPER USER PROVIDED
VOLTAGE
+EXCITATION
+OUTPUT
COMMON
METER
NC
METER
NC
+
S
S
P9
P3
1
2
3
1
2
3
VOLTAGE
+OUTPUT
-OUTPUT
NC
NC
NC
Page 43
7-3
Signal and Power Input Connections
7
Figure 7-6 Bridge Input with External Sensor Excitation
7.3 CONNECTING MAIN POWER
Wire your power (from a wall socket or other source) to P1, the orange, 3-socket connector that plugs into the 3 pins on the left side as you view the meter from the rear. The orange (power) connector must be wired according to the following table (also refer to Figure 7-8):
USA EUROPE PIN # ON WIRING WIRING ORANGE CODE CODE CONNECTION CONNECTOR
Black Brown ~ AC Line (L) 1 White Blue ~ AC Neutral (N) 2 Green Green/Yellow ~ AC Protective
Earth Ground 3
NC
+
S
S
E
+
R
P9
P3
1
2
3
1
2
3
GND
JUMPER USER PROVIDED
METER
+E
+
S
S
E
NC
P9
P3
1
2
3
1
2
3
NC
-ex
+ex
+R
+
E
+
S
S
-E
P9
P3
1
2
3
1
2
3
GND
JUMPER USER PROVIDED
METER
Figure 7.7 Potentiometer Connections with Internal Power Supply & Ratio Measurement.
Figure 7.8
Potentiometer Connections with External Power Supply & Ratio Measurement (Remove jumper S2-T)
Page 44
7
Signal and Power Input Connections
7-4
Figure 7-9. AC Connector Wiring at P1
Connect your AC meter power as described above.
CAUTION: As mentioned in Section 5.2.2, the meter has no power ON/OFF switch. The meter will be ON when power is applied.
WARNING: Do not connect ac power to your meter until you have completed all input and output connections. Failure to do so may result in injury! This device must only be installed electrically by specially trained electrician with corresponding qualifications. The main power input to the unit as well as the AC input signal to be measured must agree with the wiring instruction. The meter is factory set to the power specified by the customer at the time of ordering. The voltage is printed on the Product ID Label.
Figure 7-10. DC Connector Wiring at P1
You are now ready to proceed with scaling your meter to display in engineering units as described in Section 8.
1 2 3
P1
N
~AC LINE ~AC NEUTRAL
PROTECTIVE EARTH GND
L
L
FUSE
N
SWITCH
GREEN WIRE
EARTH
LINE
NEUTRAL
Check for proper Earth grounding in the power distribution system (single phase).
1 2 3
P1
NC +DC -DC
NO CONNECTION +DC SHIELD/-DC RETURN
DC POWER
Page 45
8. Methods for Scaling the Meter to Display in Engineering Units
8.1 INTRODUCTION
This section is for scaling the meter to read in desired engineering units. This meter is capable of producing any engineering units up to 6 digits regardless of decimal point. Even though this is a strain gauge input meter it is capable of working with high voltage, current and potentiometric inputs. The difference between strain gauge input and a voltage input is ratiometric versus non-ratiometric inputs. A strain gauge based sensor is ratiometric which means its output is directly proportional to the supply voltage powering it. Most millivolt sensors are ratiometric (strain gauge input)
The decimal point is completely programmable.
Performing the following steps will ensure that the meter is prop­erly scaled.
1. Make sure that all jumpers are in their correct location(s). (Refer to Section 6)
2. Provide AC power to the instrument (Refer to Section 7.3)
3. Check with a multimeter that the proper sensor excitation exists.
4. Wire the sensor to the meter (Refer to Section 7.2)
5. Program the unit for the desired display with strain gauge inputs (Bridge inputs, refer to Section 8.2), Voltage inputs (Refer to Section 8.3), Current inputs (Refer to section 8.4), or Potenti­ometer inputs (Refer to Section 8.5).
8.2 STRAIN GAUGE (BRIDGE) INPUT, PROGRAMMING THE METER FOR A DESIRED DISPLAY
An easy way to program your unit is to follow the example and step by step procedure below and just modify it for your exact require­ments.
8
Scaling to Display Engineering Units
EXAMPLE: An end user wants to scale the meter for a 200 PSI, 0-100 mV voltage output, pressure transducer. You want the meter to display in hundreds of a PSI from 0 to 200.0 PSI. The pressure sensor requires 10 Vdc excitation power.
8-1
Page 46
The two most common methods of scaling, Method I scaling without simulating a known load, and Method II scaling by simu­lating known loads will be demonstrated. Sometimes end users simulate known voltages instead of known physical standards (pressures); this is covered in Method II. A third method is for a unique case where you are scaling with a known load and you want the display to have 3 or more decimal places. Method III is a little complicated and should only be used, if you are looking for really accurate results and you desire a display with 3 or more places to the right of the decimal point.
A. Set the correct jumpers on the meter. (Refer to Section 6).
Install jumpers S3A & S3C and S4A & S4C on the main board. Installing jumpers S4A and S4C will provide a factory set 10 Vdc. Adjusting R38 will vary the excitation. Use a multimeter while adjusting R38.
8
Scaling to
Display
Engineering
Units
S4
DISPLAY
BD
A
BC
CBA
S3
R38
T1
W2
W3
W1
AC POWER BD
(FACTORY CONFIGURED)
SIGNAL INPUT BD
MAIN BD
J10
J2
J1
J9
J3
Figure 8-1. S3 and S4 Main Board Jumper Positions
8-2
Page 47
Install jumpers on S1 and S2 as shown in Figure 8-2.
M
B
E
A
B
S1
A
G
D
F
C
K
H
L
J
N
R
P
Q
T
S
V
U
S2
TOP VIEWS
J9
J3
8
Scaling to Display Engineering Units
Figure 8-2. S1 and S2 Signal Input Jumper Positions
B. Wire the transducer as shown in Figure 8-3 (Refer to Section 7
for additional information).
Always check the meter’s excitation prior to wiring your sensor to the meter.
8-3
Page 48
+
+
SIG
EXC
SIG
+
E
+
S
S
1
P9
2
3
METER
8
Scaling to
Display
Engineering
Units
EXC
NC
NC
E
1
P3
2
3
Figure 8-3. Internally-Powered Bridge Input
C. The unit comes with all lockout bytes set in the unlocked
position(“0”). Refer to Section 9.1 for information on lockout bytes.
If the unit has been programmed previously make sure that at least the following lockout bytes are unlocked. If they are Not unlocked you will not be able to scale the meter.
MENU SUBMENU
“L1CNF” L1C.7=0, L1C.8=0 “L2CNF” All sub menus must be unlocked “0”
Scroll to the menu by pressing the ‘MENU’ button. Enter a submenu by pressing the ‘MIN’ button. select “0” or “1” by pressing the ‘MAX’ button. Always press the ‘MENU’ button when leaving a submenu to store the values you entered.
METHOD I: SCALING WITHOUT SIMULATING A KNOWN LOAD.
NOTE: All of the following steps are accomplished by using the front panel buttons.
1. Scale the instrument by entering only 3 parameters.
The 3 parameters are “RDG SC” (Reading Scale), “RDG OF” (Reading Offset), and “DEC PT” (Decimal Point). Calculate “RDG SC” using the following formula:
8-4
Page 49
RDG SC= display span/((sensor’s mV/Voutput) (natural gain)) RDG SC= 200/(10 x 10000) = 0.002 Where: display span= 200 – 0 = 200 Sensor’s output = 100 mV/10V EXC =10 mV/V Meter’s natural gain= 10000 (Obtained from Table 8-1)
TABLE 8-1
METER’S NATURAL GAIN CHART
MAX SENSOR JUMPER RANGE METER’S OUTPUT (mV/V) @ 10 V EXC NATURAL GAIN
10 mV/V, ±5 mV/V 100 mV, ±50 mV 10000 100 mV/V, ±50 mV/V 1 V, ±500 mV 1000 1 V/V, ±0.5 V/V 10 V, ±5 V 100000 10 V/V, ±5 V/V 100 V, ±50 V 10000
Pressing the ‘MENU’ button will take you through all the different menus listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more than one item to select use the ‘MIN’ button to scroll to the next item of choice and again use the ‘MAX’ button to make your selection.
After making all your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED” on the display. If the item selected in submenu 1 has a submenu 2 associated with it, press the ‘MIN’ button to go to submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button. Pressing the ‘RESET’ button one time brings you back one ‘MENU’ item, pressing it a second time returns you to the run mode.
8
Scaling to Display Engineering Units
WARNING! this also results in a hard reset of peak and valley and your setpoints.
Proceed through the flow chart below entering the values listed in submenu 1 and 2 columns.
8-5
Page 50
8
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input.
RDG.CNF RDG.1=0 Scaling y=mx+b
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
peak reset
RDG SC 0.00200 See calculations above
Scaling to
Display
Engineering
Units
RDG OF 0000000
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S1A jumper omitted)
INP.3=0 Unipolar input
(S1B jumper omitted) INP.4=0 Std, for BRIDGE inputs INP.5= Disregard not used INP.6=0 Disables IN.SC.OF INP.7=1 Ratiometric input
IN.SC.OF Skip past
DEC PT FFFF.FF Sets decimal point
CNT BY Press ‘RESET’ two times
to return to the run mode
8-6
Page 51
2. Now you are in the run mode and the transducer is connected to the meter. Simulate no load on the transducer (just leave the pressure port open to atmosphere) and note the display reading. Assume the display shows “1.25”. To make the display indicate zeros you must go to the menu “RDG OF” and enter the value “–001.25”. After entering this value press the ‘MENU’ button to store it, and then press the ‘RESET’ button two times to return to the run mode. The scaling is now complete.
METHOD II: SCALING BY SIMULATING A KNOWN LOAD.
1. Prior to simulating a known load you must set up the unit as shown below.
The ‘MENU’ button will take you through all the different menus listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more than one item to select use the ‘MIN’ button to scroll to the next item of choice and again press the ‘MAX’ button to make your selection. After making all of your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED” on the display. If the item selected in submenu 1 has a submenu 2 associated with it, press the ‘MIN’ button to enter the submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button.
8
Scaling to Display Engineering Units
Pressing the ‘RESET’ button one time brings you back one menu item, pressing it two times returns you to the run mode.
WARNING! this also results in a hard reset of your peak and valley and setpoints.
Proceed through the flow chart below entering the values listed in submenu 1 and 2 columns.
8-7
Page 52
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input
RDG.CNF RDG.1=0 2-point scaling IN.SC.OF
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
peak reset
RDG SC 1.00000 Preset for IN.SC.OF
8
Scaling to
Display
Engineering
Units
RDG OF 0000000 Preset for IN.SC.OF
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S2A jumper omitted)
INP.3=0 Unipolar input
(S2B jumper omitted) INP.4=0 Std, for current inputs INP.5= Disregard not used INP.6=0 Disables IN.SC.OF INP.7=1 Ratiometric input
IN.SC.OF Skip past
DEC PT FFFFFF No decimal point for
calibration run
(Desired decimal point
will be entered later)
CNT BY Press ‘RESET’ two times to
get return to the run mode
8-8
Page 53
2. Now you are in the run mode and the transducer is connected to the meter. Simulate two loads on the transducer and record the meter display as “INPUT1” and “INPUT2” and record what each of these readings (“READ 1” and “READ 2”) should display on a piece of scrap paper or in the space provided below. These numbers will be the INPUT values entered in the “IN.SC.OF” scaling. Normally the two points that are calibrated are the no load condition and 3/4 full scale. However, you may use any two measurements.
“INPUT1” “INPUT2”
“READ 1” “READ 2”
Assume that you simulate no load and the display indicates “13”, and when you simulate 160 PSI the display indicates “80567”.
The following data will now be entered into the 2-point scaling “IN.SC.OF” method.
“INPUT1”: 13 “READ 1”: 000000
“INPUT2”: 80567 “READ 2”: 000160
Proceed back into the set up mode and follow the flowchart below entering the values in the submenus shown.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past INPUT Skip past RDG.CNF Skip past RDG SC Skip past (Leave as 1.0000) RDG OF Skip past (Leave as 000000) IN CNF INP.6=1 Activates IN.SC.OF
(2-point scaling)
8
Scaling to Display Engineering Units
continued next page
8-9
Page 54
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
IN.SC.OF INPUT 1 000013 This is where you
READ 1 000000 enter the data INPUT2 080567 obtained during the READ 2 000160 calibration run
(see above).
DEC PT FFFF.FF Decimal point selection
CNT BY Press the ‘RESET’ button
two times to return to the run mode
The unit is now completely scaled. If during verifying the calibra­tion, the display is not correct, you can use the offset function (“RDG OF” if all values read high or low) or perform this calibra­tion again.
8
Scaling to
Display
Engineering
Units
METHOD III: SCALING BY SIMULATING A KNOWN LOAD WITH THREE OR MORE
DECIMAL POINT POSITIONS.
EXAMPLE: An end user has a 5 pound load cell with a 2 mV/V output. The end user wants to display in ten thousandths of a pound (desired display 0.0000 to 5.0000). Proceed to Section
8.2 performing A, B and C sections.
1. Prior to simulating a known load you must set up the unit as shown below.
Pressing the ‘MENU’ button will step you through all the different menu’s listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more than one item to select, press the ‘MIN’ button to scroll to the next item of choice and again press the ‘MAX’ button to make your selection.
After making all your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED”, on the display.
If the item selected in submenu 1 has a submenu 2 associated with it, you can press the ‘MIN’ button to enter the submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button.
8-10
Page 55
Pressing the ‘RESET’ button one time brings you back one menu item, pressing the ‘RESET’ button two times returns you to the run mode.
WARNING! this also results in a hard reset of your peak and valley and setpoints.
Proceed through the flow chart below entering the values listed in submenu 1 and 2 columns.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input.
RDG.CNF RDG.1=0 2-point scaling IN.SC.OF
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
only peak reset
8
Scaling to Display Engineering Units
RDG SC 1.00000 Preset for IN.SC.OF
RDG OF 0000000 Preset for IN.SC.OF
continued next page
8-11
Page 56
8
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S2A jumper omitted)
INP.3=0 Unipolar input
(S2B jumper omitted) INP.4=0 Std, for current inputs INP.5= Disregard not used INP.6=0 Disables IN.SC.OF INP.7=1 Ratiometric input
IN.SC.OF Skip past
DEC PT FFFFF.F By shifting the decimal
point, it makes the meter
10 times more sensitive
(desired decimal point
will be entered later)
Scaling to
Display
Engineering
Units
CNT BY Press the ‘RESET’ button
two times to return to
the run mode
2. Now you are in the run mode and the load cell is connected to the meter. Simulate two loads on the transducer and record the meters display as “INPUT1” and “INPUT2” and record what each of these readings (“READ 1” and “READ 2”) should display in the spaces provided below. These numbers will be the INPUT values entered in the “IN.SC.OF” scaling. Normally the two points that are calibrated are the no load condition and 3/4 full scale. However, you may use any two measurements.
“INPUT1” “INPUT2”
“READ 1” “READ 2”
Assume that you simulate no load the display indicates “–55.6”, and when you simulate 4.0000 pounds the display indicates “16136.6”.
8-12
Page 57
The following data will now be entered into the 2-point scaling “IN.SC.OF” method.
“INPUT1”: –55.6 “READ 1”: 00000.0 “INPUT2”: 16136.6 “READ 2”: 04000.0
Ignore the decimal point position for the read values. Just note how many zero’s (or places to the right of the decimal) you need.
In this example there were 4 decimal places needed and there are 4 zeros after the “4” in the READ 2 value (ignore the decimal point).
You must alter the “RDG SC” factor as follows. Note that the decimal point above in the “READ 1” and “READ 2” values is not in the correct position. You need to move it 3 places to the left to make it correct.
By looking at Table 8-2 below, you can see that a value of 0.00100 must be entered for a new “RDG SC” value.
TABLE 8-2
READING SCALE (RDG SC) vs DECIMAL POINT SHIFT
NUMBER OF SPACES RDG SC DECIMAL POINT MOVES VALUE TO THE LEFT
0 1.00000 1 0.10000 2 0.01000 3 0.00100 4 0.00010
Proceed back into the setup mode and enter the new “RDG SC” value, the two point calibration data and move the decimal point.
You must proceed as follows.
8
Scaling to Display Engineering Units
8-13
Page 58
8
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past INPUT Skip past RDG.CNF Skip past RDG SC 0.00100 Enter new scale
value from above.
RDG OF Skip past
(Leave as 000000)
IN CNF INP.6=1 Activates “IN.SC.OF”
(2-point scaling)
IN.SC.OF INPUT 1 –0055.6 This is where you
READ 1 00000.0 enter the data INPUT2 16136.6 obtained during the READ 2 04000.0 calibration run
(see above).
Scaling to
Display
Engineering
Units
DEC PT FF.FFFF Correct decimal
point selection
CNT BY Press the ‘RESET’
button two times to return to the run mode
The unit is now completely scaled. If during verifying the calibration the display is not correct you can use the offset function (“RDG OF” if all values read high or low) or perform this calibra­tion again.
8-14
Page 59
8.3 VOLTAGE INPUT (NON RATIOMETRIC), PROGRAMMING THE METER FOR A DESIRED DISPLAY
An easy way to program your unit is to follow the example and step by step procedure below and just modify it for your exact require­ments.
EXAMPLE: An end user wants to scale the meter for a 4000 PSI, 1-5 Vdc voltage output, pressure transducer. You want the meter to display in tenths of a PSI from 0 to 4000.0 PSI. The pressure sensor requires 12-30 Vdc excitation power.
The two most common methods of scaling, Method I scaling without simulating a known load, and Method II scaling by simu­lating known loads will be demonstrated. Sometimes end users simulate known voltages instead of known physical standards (pressures); this is covered in Method II. A third method is for a unique case where you are scaling with a known load and you want the display to have 3 or more decimal places. Method III is a little complicated and should only be used, if you are looking for really accurate results and you desire a display with 3 or more places to the right of the decimal point.
A. Set the correct jumpers on the meter. (Refer to Section 6)
8
Scaling to Display Engineering Units
Install jumpers S3A, S3C and S4B on the main board.
8-15
STRAIN
Page 60
DISPLAY
BD
S3
R38
CBA
(FACTORY CONFIGURED)
SIGNAL INPUT BD
MAIN BD
J9
8
Scaling to
Display
Engineering
Units
S4
T1
J10
A
B
BC
W2
W3
AC POWER BD
W1
J1
Figure 8-4 S3 Main Board Jumper Positions
Install jumpers on S1 and S2 as shown in Figure 8-2.
M
B
A
B
S2
S1
E
A
G
D
F
C
K
H
L
J
TOP VIEWS
R
P
Q
N
T
S
J2
J3
V
U
J9
J3
Figure 8-5 S1 and S2 Signal Input Jumper Positions
B. Wire the transducer as shown in Figure 8-6 (Refer to Section 7
for additional information).
Always check the meter’s excitation prior to wiring your sensor to the meter.
8-16
Page 61
+
+
NC
GND
E
S
S
E
1
P9
2
3
DP41
METER
1
P3
2
3
+EXCITATION
+OUTPUT
VOLTAGE
COMMON
JUMPER USER PROVIDED
Figure 8-6 Sensor Input Connections
C. The unit comes with all lockout bytes set in the unlocked
position(“0”). Refer to Section 9.1 for information on lockout bytes.
If the unit has been programmed previously make sure that at least the following lockout bytes are unlocked. If they are Not unlocked you will not be able to scale the meter.
MENU SUBMENU
“L1CNF” L1C.7=0, L1C.8=0 “L2CNF” All sub menus must be unlocked “0”
Scroll to the menu by pressing the ‘MENU’ button. Enter a submenu by pressing the ‘MIN’ button. select “0” or “1” by pressing the ‘MAX’ button. Always press the ‘MENU’ button when leaving a submenu to store the values you entered.
METHOD I: SCALING WITHOUT SIMULATING A KNOWN LOAD.
NOTE: All of the following steps are accomplished by using the front panel buttons.
1. Scale the instrument by entering only 3 parameters.
8
Scaling to Display Engineering Units
The 3 parameters are “RDG SC” (Reading Scale), “RDG OF” (Reading Offset), and “DEC PT” (Decimal Point). Calculate “RDG SC” using the following formula:
8-17
STRAIN
Page 62
8
RDG SC= display span/((sensor’s output span) (natural gain)) RDG SC= 4000/(4 x 10000) = 0.100 Where: display span= 4000 – 0 = 4000 Sensor’s output span= 5 Vdc – 1 Vdc = 4 Vdc Meter’s natural gain= 10000 (Obtained from Table 8-1)
VOLTAGE SENSOR’S METER’S RANGE OUTPUT UNITS NATURAL GAIN
100 mV, ±50 mV Millivolts 1000 1 V, ±500 mV Millivolts 100 10 V, ±5 V Volts 10000 100 V, ±50 V Volts 1000
Pressing the ‘MENU’ button will take you through all the different menus listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more than one item to select you use the ‘MIN’ button to scroll to the next item of choice and again use the “‘MAX’” button to make your selection.
Scaling to
Display
Engineering
Units
After making all your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED” on the display. If the item selected in submenu 1 has a submenu 2 associated with it, press the ‘MIN’ button to go to submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button. Pressing the “‘RESET’” button one time brings you back one ‘MENU’ item, pressing it a second time returns you to the run mode.
WARNING! this also results in a hard reset of peak and valley and your setpoints.
Proceed through the flow chart below entering the values listed in submenu 1 and 2 columns.
8-18
Page 63
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input.
RDG.CNF RDG.1=0 Scaling y=mx+b
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
peak reset
RDG SC 0.10000 See calculations above
RDG OF 0000000
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S1A jumper omitted)
INP.3=0 Unipolar input
(S1B jumper omitted) INP.4=0 Std, for voltage inputs INP.5= Disregard not used INP.6=0 Disables IN.SC.OF. INP.7=0 Non ratiometric input
8
Scaling to Display Engineering Units
IN.SC.OF Skip past
DEC PT FFFFF.F Sets decimal point
CNT BY Press ‘RESET’ two times
to return to the run mode
8-19
Page 64
8
Scaling to
Display
Engineering
Units
2. Now you are in the run mode and the transducer is connected to
the meter. Simulate no load on the transducer (just leave the pressure port open to atmosphere) and note the display reading. Assume the display shows “996.5”. To make the display indicate zeros you must go to the menu “RDG OF” and enter the value “–996.5”. After entering this value press the ‘MENU’ button to store it, and then press the ‘RESET’ button two times to return to the run mode. The scaling is now complete.
METHOD II: SCALING BY SIMULATING A KNOWN LOAD.
1. Prior to simulating a known load you must set up the unit as
shown below.
The ‘MENU’ button will take you through all the different menus listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more than one item to select you use the ‘MIN’ button to scroll to the next item of choice and again press the ‘MAX’ button to make your selection. After making all of your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED” on the display. If the item selected in submenu 1 has a submenu 2 associated with it, press the ‘MIN’ button to enter the submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button.
Pressing the ‘RESET’ button one time brings you back one menu item, pressing it two times returns you to the run mode
WARNING! this also results in a hard reset of your peak and valley and setpoints.
Proceed through the flow chart below entering the values listed in submenu 1 and 2 columns.
8-20
Page 65
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input
RDG.CNF RDG.1=0 2-point scaling IN.SC.OF
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
peak reset
RDG SC 1.00000 Preset for IN.SC.OF
RDG OF 0000000 Preset for IN.SC.OF
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S2A jumper omitted) INP.3=0 Unipolar input
(S2B jumper omitted) INP.4=0 Std, for current inputs INP.5= Disregard not used INP.6=0 Disables IN.SC.OF INP.7=0 Non ratiometric input
IN.SC.OF Skip past
DEC PT FFFFFF No decimal point for
calibration run
(Desired decimal point
will be entered later)
CNT BY Press ‘RESET’ two times to
return to the run mode
8
Scaling to Display Engineering Units
8-21
Page 66
2. Now you are in the run mode and the transducer is connected to
the meter. Simulate two loads on the transducer and record the meter display as “INPUT1” and “INPUT2” and record what each of these reading (“READ 1” and “READ 2”) should display on a piece of scrap paper or in the space provided below. These numbers will be the INPUT values entered in the “IN.SC.OF” scaling. Normally the two points that are calibrated are the no load condition and 3/4 full scale. However, you may use any two measurements.
“INPUT1” “INPUT2”
“READ 1” “READ 2”
Assume that you simulate no load and the display indicates “9997”, and when you simulate 3000 PSI the display indicates “40998”.
The following data will now be entered into the 2-point scaling “IN.SC.OF” method.
8
Scaling to
Display
Engineering
Units
“INPUT1”: 9997 “READ 1”: 000000
“INPUT2”: 40998 “READ 2”: 003000
Proceed back into the set up mode and follow the flowchart below entering the values in the submenus shown.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past INPUT Skip past RDG.CNF Skip past RDG SC Skip past (Leave as 1.0000) RDG OF Skip past (Leave as 000000)
IN CNF INP.6=1 Activates IN.SC.OF
(2-point scaling)
continued next page
8-22
Page 67
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
IN.SC.OF INPUT 1 009997 This is where you
READ 1 000000 enter the data INPUT2 40998 obtained during the READ 2 003000 calibration run
(see above).
DEC PT FFFFF.F Decimal point selection
CNT BY Press the ‘RESET’ button
two times to return to the run mode
The unit is now completely scaled. If during verifying the calibra­tion, the display is not correct, you can use the offset function (“RDG OF” if all values read high or low) or perform this calibra­tion again.
METHOD III: SCALING BY SIMULATING A KNOWN LOAD WITH THREE OR MORE
DECIMAL POINT POSITIONS.
EXAMPLE: An end user has a 3 inch of water pressure transducer with a 1-5 Vdc output. The end user wants to display in ten thousandths of an inch of water (desired display 0.0000 to 3.0000). Proceed to Section 8.2 performing A,B and C sections.
1. Prior to simulating a known load you must set up the unit as shown below.
Pressing the ‘MENU’ button will step you through all the different menu’s listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more than one item to select, press the ‘MIN’ button to scroll to the next item of choice and again press the ‘MAX’ button to make your selection.
After making all your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED”, on the display.
8
Scaling to Display Engineering Units
If the item selected in submenu 1 has a submenu 2 associated with it, you can press ‘MIN’ to enter the submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button.
8-23
Page 68
8
Scaling to
Display
Engineering
Units
Pressing the ‘RESET’ button one time brings you back one menu item, pressing the ‘RESET’ button two times returns you to the run mode.
WARNING! this also results in a hard reset of your peak and valley and setpoints.
Proceed through the flow chart below entering the values listed in submenu 1 and 2 columns.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input.
RDG.CNF RDG.1=0 2-point scaling IN.SC.OF
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
only peak reset
RDG SC 1.00000 Preset for IN.SC.OF
RDG OF 0000000 Preset for IN.SC.OF
continued next page
8-24
Page 69
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S2A jumper omitted) INP.3=0 Unipolar input
(S2B jumper omitted) INP.4=0 Std, for current inputs INP.5= Disregard not used INP.6=0 Disables IN.SC.OF INP.7=0 Non ratiometric input
IN.SC.OF Skip past
DEC PT FFFFF.F By shifting the decimal
point, it makes the meter
10 times more sensitive
(desired decimal point
will be entered later)
8
CNT BY Press the ‘RESET’ button
two times to return to
the run mode
2. Now you are in the run mode and the transducer is connected to the meter. Simulate two loads on the transducer nd record the meters display as “INPUT1” and “INPUT2” and record what each of these reading (“READ 1” and “READ 2”) should display on a piece of scrap paper or in the spaces provided below. These numbers will be the INPUT values entered in the “IN.SC.OF” scaling. Normally the two points that are calibrated are the no load condition and 3/4 full scale. However, you may use any two measurements.
“INPUT1” “INPUT2”
“READ 1” “READ 2”
Assume that you simulate no load and the display indicates “9949.5”, and when you simulate 2.0000 inches of water and the display indicates “36614.2”.
Scaling to Display Engineering Units
8-25
Page 70
The following data will now be entered into the two point scaling “IN.SC.OF” method.
“INPUT1”: 9949.5 “READ 1”: 00000.0 “INPUT2” 36614.2 “READ 2” 02000.0
Ignore the decimal point position for the read values. Just note how many zero’s (or places to the right of the decimal) you need.
In this example there were 4 decimal places needed and there are 4 zeros after the “2” in the READ 2 value (ignore the decimal point).
You must alter the “RDG SC” factor as follows. Note that the decimal point above in the “READ 1” and “READ 2” values is not in the correct position. You need to move it 3 places to the left to make it correct.
By looking at Table 8-3 below, you can see that a value of 0.00100 must be entered for a new “RDG SC” value.
8
Scaling to
Display
Engineering
Units
TABLE 8-3
READING SCALE (RDG SC) vs DECIMAL POINT SHIFT
NUMBER OF SPACES DECIMAL POINT MOVES RDG SC TO THE LEFT VALUE
0 1.00000 1 0.10000 2 0.01000 3 0.00100 4 0.00010
Go back into the setup mode and enter the new “RDG SC” value, the two point calibration data and move the decimal point.
You must proceed as follows.
8-26
Page 71
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT Skip past
RDG.CNF Skip past
RDG SC 0.00100 Enter new scale
value from above.
RDG OF Skip past
(Leave as 000000)
IN CNF INP.6=1 Activates “IN.SC.OF”
(2-point scaling)
8
IN.SC.OF INPUT 1 09949.5 This is where you
READ 1 00000.0 enter the data INPUT2 336614.2 obtained during the READ 2 02000.0 calibration run
(see above).
DEC PT FF.FFFF Correct decimal
point selection
CNT BY Press the ‘RESET’
button two times to return to the run mode
The unit is now completely scaled. If during verifying the calibration the display is not correct you can use the offset function (“RDG OF” if all values read high or low) or perform this calibra­tion again.
Scaling to Display Engineering Units
8-27
Page 72
8
Scaling to
Display
Engineering
Units
8.4 CURRENT INPUT, PROGRAMMING THE METER FOR A DESIRED DISPLAY
An easy way to program your unit is to follow the example and step by step procedure below and just modify it for your exact requirements.
EXAMPLE: An end user wants to scale the meter for a 5000 PSI, 4-20 mA current output, pressure transmitter. He wants the meter to display in tenths of a PSI from 0 to 5000.0 PSI. The pressure sensor requires 12-30 Vdc excitation power.
The two most common methods of scaling, Method I scaling without simulating a known load, and Method II scaling by simulating known loads will be demonstrated. Sometimes end users simulate known currents instead of known physical standards (pressures), which is covered in Method II. A third method, Method III, is for a unique case where you are scaling with a known load and you want the display to have 3 or more decimal places. Method III is a little complicated and should only be used, if you are looking for really accurate results and you desire a display with 3 or more places to the right of the decimal point.
A. Set the correct jumpers on the meter. (Refer to Section 6 for
additional information.) Install jumpers S3A & S3C and S4B on the main board.
S3
CBA
(FACTORY CONFIGURED)
S4
SIGNAL INPUT BD
DISPLAY
BD
R38
T1
A
B
BC
W2
W3
W1
AC POWER BD
Figure 8-7. S3 and S4 Main Board Jumper Positions
8-28
J10
J1
MAIN BD
J9
J3
J2
Page 73
Install jumpers on S1 and S2 as shown in Figure 8-8.
M
B
E
A
B
S1
A
G
D
F
C
K
H
L
J
N
R
P
Q
T
S
U
TOP VIEWS
Figure 8-8. S1 and S2 Signal Input Jumper Positions
V
S2
J9
J3
8
Scaling to Display Engineering Units
B. Wire the transducer as shown in Figure 8-9. (Refer to Section 7
for additional information)
Always check the meter’s excitation prior to wiring your sensor to the meter.
(4-20mA)
(+)
(
-)
JUMPER USER PROVIDED
NC
+
+
NC
GND
E
S
S
1
P9
2
3
METER
E
1
P3
2
3
Figure 8-9. Sensor Input Connection
8-29
Page 74
8
C. The unit comes with all lockout bytes set in the unlocked
position(“0”). Refer to Section 9.1 for information on lockout bytes.
If the unit has been program previously please make sure that at least the following lockout codes are unlocked. If they are not unlocked you will not be able to scale the meter.
MENU SUB MENU
“L1CNF” L1C.7=0, L1C.8=0 “L2CNF” All sub menus must be unlocked “0”
Scroll to the desired menu by pressing the ‘MENU’ button. You enter a submenu by pressing the ‘MIN’ button. You select “0” or “1” by pressing the ‘MAX’ button. Always press the ‘MENU’ button when leaving a submenu to store the values or parameters you entered.
METHOD I: SCALING WITHOUT SIMULATING A KNOWN LOAD.
Scaling to
Display
Engineering
Units
All of the following steps are accomplished by using the front panel buttons.
1. Scale the instrument by entering only 3 parameters. The 3 parameters are “RDG SC” (Reading Scale), “RDG OF” (Reading Offset), and “DEC PT” (Decimal Point).
Calculate your “RDG SC” (Reading Scale) using the following formula:
RDG SC= display span/ ((sensors span normally 16 mA) (5000))
RDG SC= 5000/(16 x 5000) = 0.06250 Where: display span= 5000 – 0 = 5000 Sensor’s output span= 20 mA – 4 mA = 16 mA
Pressing the ‘MENU’ button will take you through all the different menus listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more than one item to select use the ‘MIN’ button to scroll to the next item of choice and again use the ‘MAX’ button to make your selection.
8-30
Page 75
After making all your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED” on the display.
If the item selected in submenu 1 has a submenu 2 associated with it, press the ‘MIN’ button to go to submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button.
Pressing the ‘RESET’ button one time brings you back one ‘MENU’ item, pressing the ‘RESET’ button two times returns you to the run mode (this also results in a hard reset of peak and valley and your setpoints).
Proceed through the flow chart below entering the values listed in the Sub Menu 1 and 2 columns.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input.
8
Scaling to Display Engineering Units
RDG.CNF RDG.1=0 Scaling y=mx+b
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
only peak reset
RDG SC 0.06250 See calculations above
RDG OF 0000000
continued next page
8-31
Page 76
8
Scaling to
Display
Engineering
Units
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S1A jumper omitted)
INP.3=0 Unipolar input
(S1B jumper omitted) INP.4=0 Std, for current inputs INP.5= Not used INP.6=0 Disables IN.SC.OF INP.7=0 Non ratiometric input
IN.SC.OF Skip past
DEC PT FFFFF.F Sets decimal point
CNT BY Press ‘RESET’ two times to
return to the Run mode
2. Now you are in the run mode and the transducer is connected to the meter. Simulate no load on the transducer (just leave the pressure port open to atmosphere) and note the display reading. Assume the display shows “1013.3”. To make the display indi­cate zeros you must go to the menu “RDG OF” and enter the value “–1013.3”. After entering this value press the ‘MENU’ button to store it, and then press the ‘RESET’ button two times to return to the run mode. The scaling is now complete.
METHOD II: SCALING BY SIMULATING A KNOWN LOAD.
1. Prior to simulating a known load you must set up the unit as shown below.
The ‘MENU’ button will take you through all the different menu’s listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more than one item to select, use the ‘MIN’ button to scroll to the next item of choice and again press the ‘MAX’ button to make your selection.
After making all your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED”, on the display. If the item selected in submenu 1 has a
8-32
Page 77
submenu 2 associated with it, press the ‘MIN’ button to enter the submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button.
Pressing the ‘RESET’ button one time brings you back one ‘MENU’ item, press the ‘RESET’ button two times to return to the run mode.
WARNING! this also results in a hard reset of your peak and valley and setpoints.
Proceed through the following flow chart entering the values listed in submenu 1 and 2 columns.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
8
INPUT BRIDGE Meter’s input
RDG.CNF RDG.1=0 2-point scaling IN.SC.OF
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
only peak reset
RDG SC 1.00000 Preset for IN.SC.OF
RDG OF 0000000 Preset for IN.SC.OF
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S2A jumper omitted)
continued next page
Scaling to Display Engineering Units
8-33
Page 78
8
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
INP.3=0 Unipolar input
(S2B jumper omitted INP.4=0 Std, for current inputs INP.5= Disregard not used INP.6=0 Disables IN.SC.OF INP.7=0 Non ratiometric input
IN.SC.OF Skip past
DEC PT FFFFFF No decimal point for
calibration run
(Desired decimal point
will be entered later)
CNT BY Press the ‘RESET’ button
two times to return to
the run mode
Scaling to
Display
Engineering
Units
2. Now you are in the run mode and the transducer is connected to the meter. Simulate two loads on the transducer and record the meter’s display as “INPUT1” and “INPUT2” and record what each of these readings should display (“READ 1” and “READ 2”) on a piece of scrap paper or in the spaces provided below. These numbers will be the INPUT values entered in the “IN.SC.OF” scaling. Normally the two points that are calibrated are the no load condition and 3/4 full scale. However, you may use any two measurements.
“INPUT1” “INPUT2”
“READ 1” “READ 2”
Assume that you simulate no load and the display indicates “19590”, and when you simulate 4000 PSI the display indicates “80150”.
The following data will now be entered into the 2-point scaling “IN.SC.OF” method.
“INPUT1”: 019590 “READ 1”: 0000
“INPUT2”: 080150 “READ 2”: 4000
8-34
Page 79
Go back into the set up mode and follow the flow chart below entering the values in the submenus shown.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past INPUT Skip past RDG.CNF Skip past RDG SC Skip past
(Leave as 1.00000)
RDG OF Skip past
(Leave as 000000)
IN CNF INP.6=1 Activates IN.SC.OF
(2-point scaling)
IN.SC.OF INPUT 1 019590 This is where you
READ 1 000000 enter the data INPUT2 080150 obtained during the READ 2 004000 calibration run (see above).
DEC PT FFFFF.F Decimal point selection
CNT BY Press the ‘RESET’ button
two times to return to the run mode
The unit is now completely scaled. If during verifying the calibration, the display is not correct, you can use the offset function (“RDG OF” if all values read high or low) or perform this calibration again.
8
Scaling to Display Engineering Units
8-35
Page 80
8
METHOD III: SCALING BY SIMULATING A KNOWN LOAD WITH THREE OR
MORE DECIMAL POINT POSITIONS.
An end user has a 3 inch of water pressure transducer with a 4-20 mA output. The end user wants to display in ten thousandths of an inch of water (desired display 0.0000 to 3.0000). Proceed to Section 8.3 performing A, B and C sections.
1. Prior to simulating a known load you must set up the unit as shown below:
Pressing the ‘MENU’ button will step you through all the different menus listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more than one item to select press the ‘MIN’ button to scroll to the next item of choice and again press the ‘MAX’ button to make your selection.
After making all your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED”, on the display.
Scaling to
Display
Engineering
Units
If the item selected in submenu 1 has a submenu 2 associated with it, press the ‘MIN’ button to enter the submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button.
Pressing the ‘RESET’ button one time brings you back one menu, pressing it a second time returns you to the run mode.
WARNING! this also results in a hard reset of your peak and valley and setpoints.
Proceed through the flow chart below entering the values listed in submenu 1 and 2 columns.
8-36
Page 81
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input
RDG.CNF RDG.1=0 2-point scaling IN.SC.OF
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
only peak reset
RDG SC 1.00000 Enter for IN.SC.OF scaling
RDG OF 0000000 Enter for IN.SC.OF scaling IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S2A jumper omitted)
INP.3=0 Unipolar input
(S2B jumper omitted) INP.4=0 Std, for current inputs INP.5= Disregard not used INP.6=0 Disables IN.SC.OF INP.7=0 Non ratiometric input
IN.SC.OF Skip past
DEC PT FFFFF.F By shifting the decimal
point, it makes the meter
10 times more sensitive
(desired decimal point will
be entered later)
CNT BY Press the ‘RESET’ button
two times to return to
the run mode
8
Scaling to Display Engineering Units
8-37
Page 82
2. Now you are in the run mode and the transducer is connected to the meter. Simulate two loads on the transducer and record the meter display as “INPUT1” and “INPUT2” and record what each of these readings should display (“READ 1” and “READ 2”) on a piece of scrap paper or in the spaces provided below, these numbers will be the INPUT values entered in the “IN.SC.OF” scaling.
Normally the two points that are calibrated are the no load condition and 3/4 full scale. However, you may use any two measurements.
“INPUT1” “INPUT2”
“READ 1” “READ 2”
Assume that you simulate no load and the display indicates “20054.3”, and when you simulate 2.0000 inches of water, the display indicates “66675.6” .
8
Scaling to
Display
Engineering
Units
The following data will now be entered into the 2-point scaling “IN.SC.OF” method.
“INPUT1”: 20054.3 “READ 1”: 00000.0 “INPUT2”: 66675.6 “READ 2”: 02000.0
Ignore the decimal point position for the READ values. Just note how many zero’s (or places to the right of the decimal) you need.
In this example there were 4 decimal places needed and there are 4 zeros after the “2” (in the “READ 2” value).
You must alter the “RDG SC” factor as follows. Note that the decimal point above in the “READ 1” and “READ 2” values is not in the correct position. You need to move it 3 places to the left to make it correct.
By looking at Table 8-4, you can see that a value of 0.00100 must be entered for a new “RDG SC” value.
8-38
Page 83
TABLE 8-4
READING SCALE (RDG SC) vs DECIMAL POINT SHIFT
NUMBER OF SPACES DECIMAL POINT MOVES RDG SC TO THE LEFT VALUE
0 1.00000 1 0.10000 2 0.01000 3 0.00100 4 0.00010
Go back into the setup mode and enter the new “RDG SC” value, the 2-point calibration data and move the decimal point.
You must proceed as follows:
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past INPUT Skip past RDG.CNF Skip past
RDG SC 0.00100 Enter new scale value
from above.
RDG OF Skip past (Leave as 000000)
IN CNF INP.6=1 Activates “IN.SC.OF”
(2-point scaling)
IN.SC.OF INPUT 1 20054.3 This is where you
READ 1 00000.0 enter the data INPUT2 66675.6 obtained during the READ 2 02000.0 calibration
run (see above)
8
Scaling to Display Engineering Units
DEC PT FF.FFFF Correct decimal point
CNT BY Press the ‘RESET’ button
selection
two times to return to the run mode
8-39
Page 84
8
Scaling to
Display
Engineering
Units
The unit is now completely scaled. If during verifying the calibration the display is not correct you can use the offset function (“RDG OF” if all values read high or low) or perform this calibra­tion again.
8.5 POTENTIOMETER, PROGRAMMING THE METER FOR A DESIRED DISPLAY
An easy way to program your unit is to follow the example and step by step procedure below and just modify it for your exact requirements.
EXAMPLE: An end user wants to scale the meter for a 200 PSI, 0-500 ohms output, pressure transducer. You want the meter to display in hundredths of a PSI from 0 to 200.00 PSI. The pressure sensor requires 10 Vdc excitation power.
The two most common methods of scaling, Method I scaling without simulating a known load, and Method II scaling by simu­lating known loads will be demonstrated. Sometimes end users simulate known voltages instead of known physical standards (pressures); this is covered in Method II. A third method is for a unique case where you are scaling with a known load and you want the display to have 3 or more decimal places. Method III is a little complicated and should only be used, if you are looking for really accurate results and you desire a display with 3 or more places to the right of the decimal point.
8-40
Page 85
A. Set the correct jumpers on the meter as shown in Figure 8-10. (Refer to Section 6 for additional information)
Configure jumpers on S1 and S2 as shown in Section 6.
Install jumpers S3A & S3C and S4A & S4C on the main board. S4A & S4C will provide a factory set 10 Vdc. Adjusting R38 will vary excitation. Use a multimeter when adjusting R38.
S4
DISPLAY
BD
A
BC
CBA
S3
R38
T1
W2
W3
W1
AC POWER BD
(FACTORY CONFIGURED)
SIGNAL INPUT BD
MAIN BD
J10
J2
J1
J9
J3
8
Scaling to Display Engineering Units
Figure 8-10. S3 and S4 Main Board Jumper Positions
Install jumpers on S1 and S2 as shown in Figure 8-11.
8-41
Page 86
M
E
A
B
S2
S1
B
A
G
D
F
C
K
H
L
J
N
R
P
S
Q
V
T
U
TOP VIEWS
J9
J3
8
Scaling to
Display
Engineering
Units
Figure 8-11. S1 and S2 Signal Input Jumper Positions
B. Wire the transducer as shown in Figure 8-12. (Refer to Section 7 for additional information).
Always check the meter’s excitation prior to wiring your sensor to the meter.
+
JUMPER USER PROVIDED
+
NC
NC
GND
E
S
S
E
1
P9
2
3
METER
1
P3
2
3
Figure 8-12. Sensor Input Connections
8-42
Page 87
C. The unit comes with all lockout bytes set in the unlocked
position (“0”). Refer to Section 9.1 for information on lockout bytes.
If the unit has been programmed previously make sure that at least the following lockout bytes are unlocked. If they are Not unlocked you will not be able to scale the meter.
MENU SUBMENU
“L1CNF” L1C.7=0, L1C.8=0 “L2CNF” All sub menus must be unlocked “0”
Scroll to the menu by pressing the ‘MENU’ button. Enter a submenu by pressing the ‘MIN’ button. Select “0” or “1” by pressing the ‘MAX’ button. Always press the ‘MENU’ button when leaving a submenu to store the values you entered.
METHOD I: SCALING WITHOUT SIMULATING A KNOWN LOAD.
All of the following steps are accomplished by using the front panel buttons.
1. Scale the instrument by entering only 3 parameters. The 3 parameters are “RDG SC” (Reading Scale), “RDG OF” (Reading Offset), and “DEC PT” (Decimal Point).
Calculate “RDG SC” using the following formula:
RDG SC= display span/((sensor’s output span) (natural gain))
RDG SC= 200/(10 x 10000) = 0.002 Where: display span= 200 – 0 = 200 Sensor’s output span= 10 volts Meter’s natural gain= 10000 (from Table 8-5)
TABLE 8-5
METER’S NATURAL GAIN CHART
VOLTAGE SENSOR’S METER’S RANGE OUTPUT UNITS NATURAL GAIN
8
Scaling to Display Engineering Units
100 mV, ±50 mV Millivolts 1000 1 V, ±500 mV Millivolts 100 10 V, ±5 V Volts 10000 100 V, ±50 V Volts 1000
8-43
Page 88
Pressing the ‘MENU’ button will take you through all the different menu’s listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more than one item to select use the ‘MIN’ button to scroll to the next item of choice and again use the ‘MAX’ button to make your selection.
After making all of your choices, in submenu 1, enter these selec­tions by pressing the ‘MENU’ button which will momentarily flash “STORED” on the display.
If the item selected in submenu 1 has a submenu 2 associated with it, press the ‘MIN’ button to enter the submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice with the ‘MENU’ button. Pressing the ‘RESET’ button one time brings you back one menu item, pressing the ‘RESET’ button two times returns you to the run mode.
8
Scaling to
Display
Engineering
Units
WARNING! this also results in a hard reset of peak and valley and your setpoints.
Proceed through the flow chart below entering the values listed in submenu 1 and 2 columns.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input.
RDG.CNF RDG.1=0 Scaling y=mx+b
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
only peak reset
continued next page
8-44
Page 89
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
RDG SC 0.00200 See calculations above
RDG OF 0000000
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S1A jumper omitted)
INP.3=0 Unipolar input
(S1B jumper omitted) INP.4=0 Std, for voltage inputs INP.5= Disregard not used INP.6=0 Disables IN.SC.OF INP.7=1 Ratiometric input
IN.SC.OF Skip past
DEC PT FFFF.FF Sets decimal point
CNT BY Press the ‘RESET’ button
two times to return to
the run mode
2. Now you are in the run mode and the transducer is connected to the meter. Simulate no load on the transducer (just leave the pressure port open to atmosphere) and note the display reading. Assume the display shows “1.25”. To make the display indicate zeros you must go to the menu “RDG OF” and enter the value “–001.25”. After entering this value press the ‘MENU’ button to store it, and then press the ‘RESET’ button two times to return to the run mode. The scaling is now complete.
METHOD II: SCALING BY SIMULATING A KNOWN LOAD.
1. Prior to simulating a known load you must set up the unit as
shown below.
8
Scaling to Display Engineering Units
The ‘MENU’ button will take you through all the different menu’s listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more then one item to select use the ‘MIN’ button to scroll to the next item of choice and
8-45
Page 90
again press the ‘MAX’ button to make your selection. After making all of your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED”, on the display. If the item selected in submenu 1 has a submenu 2 associated with it, press the ‘MIN’ button to enter the submenu 2.
Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button.
Pressing the ‘RESET’ button one time brings you back one menu item, pressing the ‘RESET’ button two times returns you to the run mode
WARNING! this also results in a hard reset of your peak and valley and setpoints.
Proceed through the flow chart below entering the values listed in submenu 1 and 2 columns.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
8
Scaling to
Display
Engineering
Units
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input
RDG.CNF RDG.1=0 2-point scaling IN.SC.OF
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed. RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset
vs only peak reset
RDG SC 1.00000 Preset for IN.SC.OF
RDG OF 0000000 Preset for IN.SC.OF
continued next page
8-46
Page 91
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S2A jumper omitted)
INP.3=0 Unipolar input
(S2B jumper omitted) INP.4=0 Std INP.5= Disregard not used INP.6=0 Disables IN.SC.OF INP.7=1 Ratiometric input
IN.SC.OF Skip past
DEC PT FFFFFF No decimal point for
calibration run
(Desired decimal point will
be entered later)
CNT BY Press the ‘RESET’ button
two times to return to
the run mode
8
2. Now you are in the run mode and the transducer is connected to the meter. Simulate two loads on the transducer and record the meter’s display as “INPUT1” and “INPUT2” and record what each of these readings (“READ 1” and “READ 2”) should display on a piece of scrap paper or in the spaces provided below. These numbers will be the INPUT values entered in the “IN.SC.OF” scaling. Normally the two points that are calibrated are the no load condition and 3/4 full scale. However, you may use any two measurements.
“INPUT1” “INPUT2”
“READ 1” “READ 2”
Assume that you simulate no load and the display indicates “13”, and when you simulate 160 PSI the display indicates “80567”.
The following data will now be entered into the two point scaling “IN.SC.OF” method.
Scaling to Display Engineering Units
“INPUT1”: 000013 “READ 1”: 000000
“INPUT2”: 080567 “READ 2”: 000160
8-47
Page 92
8
Scaling to
Display
Engineering
Units
Proceed back into the set up mode and follow the flow chart below entering the values in the submenus shown.
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past INPUT Skip past RDG.CNF Skip past RDG SC Skip past
(Leave as 1.0000)
RDG OF Skip past
(Leave as 000000)
IN CNF INP.6=1 Activates IN.SC.OF
(2-point scaling)
IN.SC.OF INPUT 1 000013 This is where you
READ 1 000000 enter the data INPUT2 080567 obtained during the READ 2 000160 calibration
run (see above)
DEC PT FFFFF.F Decimal point
selection
CNT BY Press the ‘RESET’ button
two times to return to the run mode
The unit is now completely scaled. If during verifying the calibration the display is not correct you can use the offset function (“RDG OF” if all values read high or low) or perform this calibra­tion again.
8-48
Page 93
METHOD III: SCALING BY SIMULATING A KNOWN LOAD WITH THREE OR MORE
DECIMAL POINT POSITIONS.
EXAMPLE: An end user has a displacement sensor with 5 inches of travel and a 500 ohm output. You want to display in one ten thousandths of an inch (desired display 0.0000 to 5.0000). Proceed to Section 8.4 performing A, B and C sections.
1. Prior to simulating a known load you must set up the unit as
shown below.
Pressing the ‘MENU’ button will step you through all the different menus listed below. When you wish to enter a menu, to the submenu 1 stage, press the ‘MIN’ button. To make a selection, in submenu 1, press the ‘MAX’ button. If the submenu has more then one item to select press the ‘MIN’ button to scroll to the next item of choice and again press the ‘MAX’ button to make your selection.
After making all your choices, in submenu 1, enter these selections by pressing the ‘MENU’ button which will momentarily flash “STORED”, on the display.
If the item selected in submenu 1 has a submenu 2 associated with it, press the ‘MIN’ button to enter the submenu 2. Select your choice by pressing the ‘MAX’ button and enter your choice by pressing the ‘MENU’ button.
Pressing the ‘RESET’ button one time brings you back one menu item, pressing the ‘RESET’ button two times returns you to the run mode.
WARNING! this also results in a hard reset of your peak and valley and setpoints.
Proceed through the flow chart below entering the values listed in submenu 1 and 2 columns.
8
Scaling to Display Engineering Units
8-49
Page 94
MENU SUBMENU 1 SUBMENU 2 DESCRIPTION
L1CNF Skip past L2CNF Skip past L3CNF Skip past L4CNF Skip past
INPUT BRIDGE Meter’s input
RDG.CNF RDG.1=0 2-point scaling IN.SC.OF
RDG.2=0 Active decimal point RDG.3=0 Normal display brightness RDG.4=1 Leading zeros suppressed RDG.5= Disregard not used RDG.6= Disregard not used RDG.7=0 External hard reset vs
only peak reset
RDG SC 1.00000 Preset for IN.SC.OF
8
Scaling to
Display
Engineering
Units
RDG OF 0000000 Preset for IN.SC.OF
IN CNF INP.1=0 60 Hz ac power
INP.2=0 Slow reading
(S2A jumper omitted)
INP.3=0 Unipolar input
(S2B jumper omitted) INP.4=0 Std, for current inputs INP.5= Disregard not used INP.6=0 Disables IN.SC.OF INP.7=1 Ratiometric input
IN.SC.OF Skip past
DEC PT FFFFF.F By shifting the decimal
point, it makes the meter
10 times more sensitive
(desired decimal point
will be entered later)
CNT BY Press the ‘RESET’ two
times to return to the
run mode
8-50
Page 95
2. Now you are in the run mode and the transducer is connected to the meter. Simulate two loads on the transducer and record the meter displays as “INPUT1” and “INPUT2” and record what each of these readings (“READ 1” and “READ 2”) should display on a piece of scrap paper or in the spaces provided below. These numbers will be the INPUT values entered in the “IN.SC.OF” scaling. Normally the two points that are calibrated are the no load condition and 3/4 full scale. However, you may use any two measurements.
“INPUT1” “INPUT2”
“READ 1” “READ 2”
Assume that you simulate no displacement and the display indi­cates “-55.6”, and when you simulate 4.0000 inches the display indicates “16136.6”.
The following data will now be entered into the two point scaling “IN.SC.OF” method.
8
“INPUT1”: -0055.6 “READ 1”: 00000.0
“INPUT2”: 16136.6 “READ 2” 04000.0
Ignore the decimal point position for the READ values. Just note how many zero’s (or places to the right of the decimal) you need.
In this example there were 4 decimal places needed and there are 4 zeros after the “4” in the “READ 2” value (ignore the decimal point).
You must alter the “RDG SC” factor as follows. Note that the decimal point above in the “READ 1” and “READ 2” values is not in the correct position. You need to move it 3 places to the left to make it correct.
By looking at Table 8-6, you can see that a value of 0.00100 must be entered for a new “RDG SC” value.
Scaling to Display Engineering Units
8-51
Page 96
Page 97
9-1
Meter Function Menus
9
EXPLANATION OF LOCKOUT CONFIGURATIONS AND METER FUNCTION MENUS
HOW TO USE THE TABLES IN SECTION 9
MIN/MAX/MENU These are the buttons on the meter you are to
press to access the parameters given in the same column.
MAIN MENU/ These are headings for the table columns. SUBMENU:
DISPLAYED These are parameters seen on the display INFORMATION: after pressing either ‘MIN’, ‘MAX’, or ‘MENU’
button(s).
NOTE: If you press the ‘RESET’ button two times while the meter is in the run mode, all Setpoints, Alarms, Peak & Valley will be reset and the meter will begin new measurements.
If you press the ‘RESET’ button one time while in the configuration mode, you will move one MAIN MENU backwards and any selection will not be saved. If you press the ‘RESET’ button two times while in the configuration mode, you will reset the meter and only those menu items saved by pressing the ‘MENU’ button will be saved.
9.1 Individual Lockout Information
To restrict access to different parameters of the program in the meter, you may want to lockout parts of the meter. When you lock out a parameter, it will no longer appear when you scroll through the menu. To lock out specific parameters of the meter (setpoint, scaling), refer to the following tables.
Once set (to unlock useful features for a given application and to lock out any features), these four “L1C”, “L2C”, “L3C”, and “L4C” can be rapidly skipped over by pressing the ‘MENU’ button four times.
Note
Page 98
9
Meter Function Menus
9-2
MENU MIN/MAX* BUTTON BUTTON MAIN MENU SUB MENU CONDITION
“L1 CNF”: LOCKOUT CONFIGURATION #1
“L1C.1=0” Setpoint 1 change unlocked.
“L1C.1=1” Setpoint 1 change locked out.
“L1C.2=0” Setpoint 2 change unlocked.
“L1C.2=1” Setpoint 2 change locked out.
“L1C.3=0” Setpoint 3 (Alarm 1) change unlocked.
“L1C.3=1” Setpoint 3 (Alarm 1) change locked out.
“L1C.4=0
Setpoint #4 (Alarm 2) change unlocked.
“L1C.4=1” Setpoint #4 (Alarm 2) change locked out.
“L1C.5=0” Valley-value (LO RDG) display is permitted.
“L1C.5=1” Displays microprocessor revision.
“L1C.6=0” Peak-value (HI RDG) display is permitted.
“L1C.6=1” Peak-value (HI RDG) display is not
permitted.
“L1C.7=0” INPUT CLASS (BRIDGE) is selected.
“L1C.7=1” INPUT CLASS is locked out.
“L1C.8=0” Not used in “BRIDGE” mode,
must be set to “0”.
“L1C.8=1” Not used in “BRIDGE” mode.
*The ‘MIN’ button allows you to sequence through L1C.1, L1C.2, L1C.3,
L1C.4, L1C.5, L1C.6, L1C.7 and L1C.8.
The ‘MAX’ button allows you to select the “0” or “1” state for each “L1C” condition.
The ‘MENU’ button stores the selected values for all “L1C” condition(s) changed and advances the meter to “L2 CNF”. Do not press the ‘MENU’ button after each change within the submenu or the meter will advance to the next menu item.
Every underlined “0” or “1” state is the factory preset value.
Page 99
9-3
Meter Function Menus
9
MENU MIN/MAX* BUTTON BUTTON MAIN MENU SUB MENU CONDITION
“L2 CNF”: LOCKOUT CONFIGURATION #2
“L2C.1=0” RDG.CNF (scale/offset method and display
features) may be chosen.
“L2C.1=1” RDG.CNF (scale/offset method and display
features) is locked out.
“L2C.2=0” Either RDG SC (computed input-to-display
scale factor) or RD.SC.OF (two data points, which determine the reading scale/offset) may be entered.
“L2C.2=1” Either RDG SC (computed input-to-display
scale factor) or RD.SC.OF (two data points, which determine the reading scale/offset) is locked out.
“L2C.3=0” RDG OF (offset computed in display digits)
may be entered.
“L2C.3=1” RDG OF (offset computed in display digits)
is locked out.
“L2C.4=0
INP.CNF (meter rates, front-end features,
prelinearizing scale/offset) may be chosen.
“L2C.4=1” INP.CNF (meter rates, front-end features,
prelinearizing scale/offset) may not be locked out.
“L2C.5=0” IN.SC.OF (two data points for additional
scale/offset) may be entered.
“L2C.5=1” IN.SC.OF (two data points for additional
scale/offset) may not be entered.
“L2C.6=0” DEC PT (decimal-point location) may be
chosen
“L2C.6=1” DEC PT (decimal-point location) may not
be chosen.
continued next page
Page 100
9
Meter Function Menus
9-4
MENU MIN/MAX* BUTTON BUTTON MAIN MENU SUB MENU CONDITION
“L2 CNF”
“L2C.7=0” CNT BY (round off of display) can be
specified.
“L2C.7=1” CNT BY (round off of display) cannot be
specified.
“L2C.8=0” FIL CNF (adaptive/fixed filtering and for
which output(s)) can be chosen.
“L2C.8=1” FIL CNF (adaptive/fixed filtering and for
which output(s)) cannot be chosen.
* The ‘MIN’ button allows you to sequence through L2C.1, L2C.2, L2C.3,
L2C.4, L2C.5, L2C.6, L2C.7, and L2C.8.
The ‘MAX’ button allows you to select the “0” or “1” state for each “L2C” condition.
The ‘MENU’ button stores the selected values for all “L2C” condition(s) changed and advances the meter to “L3 CNF”. Do not press the ‘MENU’ button after each change within the submenu or the meter will advance to the next menu item.
Every underlined “0” or “1” state is the factory preset value.
MENU MIN/MAX* BUTTON BUTTON MAIN MENU SUB MENU CONDITION
“L3 CNF”: LOCKOUT CONFIGURATION#3
“L3C.1=0
FIL TI (# of samples in average) can be
chosen.
“L3C.1=1” FIL TI (# of samples in average) cannot be
locked out.
“L3C.2=0” SP CNF (mode of action of setpoints 1 & 2
LEDs, transistors and relays) can be selected.
“L3C.2=1” SP CNF (mode of action of setpoints 1 & 2
LEDs, transistors and relays) cannot be locked out.
“L3C.3=0
AL CNF (mode of action of Setpoints 3 & 4,
often used as alarms) can be locked out.
Loading...