Omega DP25-CRMS, DP25-VRMS User guide

Page 1
DP25-VRMS
DP25-CRMS
True RMS Meter
omega.com
For latest product manuals:
omegamanual.info
User ’s Guide
MADE IN
USA
Page 2
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGAis constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient-connected applications.
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as the guide contains important information relating to safety and EMC.
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as the guide contains important information relating to safety and EMC.
!
®
®
OMEGAnet® On-Line Service
www.omega.com
Internet e-mail
Servicing North America:
USA: One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford CT 06907-0047
 TEL: (203) 359-1660 FAX: (203) 359-7700  e-mail: [email protected]
Canada: 976 Bergar  Laval (Quebec) H7L 5A1  TEL: (514) 856-6928 FAX: (514) 856-6886  e-mail: [email protected]
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
 Customer Service: 1-800-622-2378 / 1-800-622-BEST  Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
Mexico and TEL: (001)800-TC-OMEGA® FAX: (001) 203-359-7807 Latin American: En Espa
ñol: (001) 203-359-7803
 e-mail: [email protected]
®
®
®
Servicing Europe:
Czech Republic: Frystatska 184, 733 01 Karviná
 TEL: +420 59 6311899 FAX: +420 59 6311114  e-mail: [email protected]
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany  TEL: +49 7056 9398-0 FAX: +49 7056 9398-29  Toll Free in Germany: 0800 639 7678  e-mail: [email protected]
United Kingdom: One Omega Drive
ISO 9002 Certified River Bend Technology Centre
 Northbank, Irlam Manchester M44 5BD United Kingdom  TEL: +44 161 777 6611 FAX: +44 161 777 6622  Toll Free in England: 0800 488 488  e-mail: [email protected]
Page 3
i
C
ONTENTS
Section 1: Setup 1
1.1 Before You Begin 2
1.2 Features and Options 3
1.3 Safety Considerations 4
1.4 Getting Started 5
Section 2: Operation 13
2.1 Introduction 14
2.2 SW1 DIP Switch Positions for Input Ranges 15
2.3 Front Panel Display 16
2.4 Meter Buttons 17
2.5 Flow Charts 18 Setpoint Adjustment (Run Mode) 18 Menu Configuration (Configuration Mode) 19
2.6 Menu Configurations 21 Input Menu 22 Reading Menu 24 Setpoint Menu 29 Output Menu 32 Communication Menu 34
Section 3: True RMS Measurement 47
3.1 Introduction 48
3.2 True RMS versus Rectified-average Readings 49
3.3 Voltage-Divider Action of Leads and Meter Input Impedance 50
3.4 Input Current Effects 51
3.5 Overvoltage Heating 51
3.6 Guarding Against AC Pickup 51
3.7 Common-Mode Voltage Effects 52
Section 4: Defaults and Specifications 53
4.1 Back of the Meter 54
4.2 Main Board 56
4.3 W1–W3 Solder Jumpers 57
4.4 Signal Board 58
4.5 Meter Housing 60
4.6 Factory Default Values 61
4.7 Meter Specifications 62
4.8 Glossary 69
Section 5: CE Approvals Information 71
Page 4
L
IST OFFIGURES
Figure 1.1 Panel Cut-out Dimensions 5 Figure 1.2 Exploded View of the Meter 6 Figure 1.3 P6 Volt Meter Connection 7 Figure 1.4 P6 Current Meter Connection 8 Figure 1.4.1 P6 Current Meter Connection with Optional Current Transformer 8 Figure 1.5 P2 & P3 Dual Relay Output Connection 9 Figure 1.6 P4 Analog Output Connection 9 Figure 1.7 P10 & J5 RS232 Output Connections 10 Figure 1.8 J5 Half-duplex RS485 Output Connections 10 Figure 1.9 SW1 Input Range Selection 11 Figure 1.10 P1 AC Power Connection 12 Figure 2.1 Front Panel Display for Volt Meter 16 Figure 2.2 Front Panel Display for Current Meter 16 Figure 2.3 Flow Chart for Setpoint Configuration (Run Mode) 18 Figure 2.4 Flow Chart for Configuration Mode, Part 1 19 Figure 2.5 Flow Chart for Configuration Mode, Part 2 20 Figure 3.1 True RMS versus Rectified-average Readings 49 Figure 3.2 Approximate Input Resistance 50 Figure 4.1 Volt Meter Connections 54 Figure 4.2 Volt Meter Label 54 Figure 4.3 Current Meter Connections 55 Figure 4.4 Current Meter Label 55 Figure 4.5 Jumper Positions on the Main Board 56 Figure 4.6 W1–W3 Jumper Locations 57 Figure 4.7 Signal Board 58 Figure 4.8 Exploded View of Case and Board Assembly 59 Figure 4.9 Meter Housing and Panel Cut-out Dimensions 60
ii
Page 5
iii
L
IST OFTABLES
Table 1.1 RS232 Meter Connection to the Computer 10 Table 1.2 P1 AC Power Wire Color & Terminal Connections 12 Table 2.1 DIP Switch Positions for Voltage Meter 15 Table 2.2 DIP Switch Positions for Current Meter 15 Table 2.3 Display Indicators 16 Table 2.4 Button Functions 17 Table 2.5 Menu Descriptions 21 Table 3.1 Meter Net Loading Impedance 50 Table 4.1 Jumper Functions 56 Table 4.2 W1-W3 Jumper wiring 57 Table 4.3 Menu Item Defaults 61 Table 4.4 Analog Output Scale and Offset Defaults 61 Table 4.5 Signal Input Specifications 62 Table 4.6 Display Specifications 62 Table 4.7 Span Tempco for Voltmeter 63 Table 4.8 Span Tempco for Current Meter 63 Table 4.9 Analog to Digital Specifications 63 Table 4.10 Isolated Analog Output Specifications 64 Table 4.11 Communications Information 64 Table 4.12 Input Power Specifications 64 Table 4.13 Operating Environment 65 Table 4.14 Mechanical Specifications 65 Table 4.15 Alarm Outputs 65 Table 4.16 Voltage Model Accuracy (DC coupled inputs) 66 Table 4.17 Voltage Model Accuracy (AC coupled inputs) 66 Table 4.18 Current Model Accuracy 67 Table 4.19 Decimal Point & Unit of Measure Defaults 67 Table 4.20 Insulation 68
Page 6
P
REFACE
Preface to the Operator's Manual
The latest additions to the digital panel meter series are the VRMS and CRMS True RMS Volt / Current meters. Both the VRMS (voltage meter) and the CRMS (current meter) may be ordered with the isolated analog output, and/or dual relay option. These options are not field installable and must be ordered at the time of purchase. The VRMS measures voltages from 100mV to 750V (see CE Approvals section for full details) in five ranges. All ranges are pre-calibrated and selected via the rear dip switch. The CRMS can measure currents from 1mA to 5amp. The current meter has an internal built in shunt for the 5amp range. The 5amp and the 1amp range have dedicated inputs. The 1mA through 100mA ranges may be selected via the rear dip switch. Again, the meter is shipped pre-calibrated for all the ranges. Users need only connect the meter to the signal source and select the proper range from the menu. The meter will then start measuring and displaying the voltage or current present at its inputs.
The RMS meters are ideal for monitoring complex voltage or current waveforms generated by SCR banks to control the motor speed/torque. If the dual relay option is used, the meters can be programmed to provide an alarm when the motor is overloaded or when the load is disengaged from the motor.
Other typical applications would be to monitor the line voltage or current and use the isolated analog output to transmit the information to a chart recorder. The meters can also communicate with PLC's and computers via an optional isolated serial RS232 or RS485 board. This option may be added at any time as it is field installable.
Range switching for both models is done by selecting the proper opened/closed position for the SW1 DIP switches (see section 2.2) accessible from the back of the meter. Once the SW1 switches have been set to the desired voltage or current range, the same range must also be selected from the Range (RNGE) submenu of the Input (INPT) main menu (see the menu section 2 configuration flowchart ).
You must decide whether the signal is to be AC or DC coupled. AC coupling is the preferred method when the input signal is ridding on some undesirable DC voltage. To AC couple the voltage
signal, connect the input signal to the P6 terminal labeled AC (see section 1.4.2), and change the Coupling (CPLG) submenu of the Input (INPT) main menu to AC. To remove the DC component of the current
signal, SW1-5 DIP switch must be opened (see section 2.2). If the measurement is being done directly on the local power where the frequency may be 50Hz or 60Hz, set the Frequency (FREQ) submenu of the Input (INPT) main menu to the local power frequency for increased accuracy. If the frequency of the input signal is not known, select ALL from the Frequency (FREQ) submenu. The second part of this manual, Operation, explains menu configuration in detail.
iv
Page 7
v
N
OTES
When rescaling the meter either with or without known loads, if a high degree of accuracy is required, do not apply 0 volts or use 0 as a reference point (see section 2.6.2). For any rescaling, the minimum signal should always be 3 to 5% of the full scale of the signal range. The same is true for scaling the analog output option. Do not use a value of 0 on the display if a high degree of accuracy is required, but instead use a value slightly higher than 0 counts on the display.
NOTES, WARNINGS and CAUTIONS
Information that is especially important to note is identified by these labels:
• NOTE
• WARNING
• CAUTION
• IMPORTANT
NOTE: provides you with information that is important to successfully setup and use the Programmable Digital Meter.
CAUTION: tells you about the risk of electric shock.
CAUTION, RISK OF DANGER: tells you of circumstances or practices that can
effect the meter's functionality and documentation needs to be consulted in all cases, in order to find out the nature of potential hazards and actions.
Note
☞
Note
☞
Page 8
C
ONTENTS
vi
Page 9
S
ETUP
Section 1: Setup
Before You Begin
Features and Options
Safety Considerations
Getting Started
1
1
Page 10
1
S
ETUP
1.1 Before You Begin
If there are any questions about the shipment, use the phone number for the Customer Service Department nearest you.
Remove the packing slip and verify that you have received everything listed, including:
• True RMS Meter
• Operator's Manual
• Quick Start Pamphlet
• Connectors
!
Inspect the container and equipment for signs of damage as soon as you receive the shipment. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent. The carrier will not honor damage claims unless all shipping material is saved for inspection. After examining and removing the contents, save the packing material and carton in the event reshipment is necessary.
In addition to the meter and connectors, you will need the following items to setup your unit:
• 110V three-prong power cord
• 1/8" flat head screwdriver
1.1.3 What You Will Need
1.1.2 Inspecting Your Shipment
1.1.1 Customer Service
2
Page 11
3
S
ETUP
1.2 Features and Options
Your meter has several basic features, some of which are listed below. For more detailed information on specifications and factory defaults, see Section 4: Specifications.
• 4 Digit, 14 segment LED display
• NEMA 4 / Type 4 Front Bezel
• Peak and Valley detection
• Nonvolatile memory — no battery backup necessary
• 115Vac or 230Vac 50/60 Hz power source
There are two models available — a current meter and a volt meter. Both of these models are available from the factory in the following configurations:
• Basic Unit
• Basic Unit with Dual Relay Output
• Basic Unit with Isolated Analog Output
• Basic Unit with Dual Relay and Isolated Analog Output
The relay and analog outputs are not field installable options. They must be requested at time of purchase. Menus will only be available for those options that are supported by that meter.
The models can accommodate a communication board, which can be installed at any time. There are two different communication boards available:
• Isolated RS232 Output
• Isolated RS485 Output
If your meter has a communication board installed, then you may control all functions of the meter through a serial communications connection with your computer. The book, Serial Communications, that comes with the communication board itself, has detailed information on setup and configuration procedures. Note that this manual, the Operator's Manual, explains only the more basic options available to users whose meters are equipped with the communication board.
Note
☞
1.2.3 Field Installable Options
Note
☞
1.2.2 Factory Configurations
1.2.1 Basic Features
1
Page 12
1
S
ETUP
1.3 Safety Considerations
This device is marked with the international caution symbol. It is important to read this manual before installing or commissioning this device as it contains important information relating to Safety and EMC (Electromagnetic Compatibility).
Unpacking & Inspection
Unpack the instrument and inspect for obvious shipping damage. Do not attempt to operate the unit if damage is found.
This instrument is a panel mount device protected in accordance with EN 61010­1:2001, electrical safety requirements for electrical equipment for measurement, control and laboratory. Installation of this instrument should be done by qualified personnel. In order to ensure safe operation, the following instructions should be followed.
This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947–1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the mains supply cord.
Furthermore, to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment, an overcurrent protection device shall be installed.
• The Protective Conductor must be connected for safety reasons. Check that the power cable has the proper Earth wire, and it is properly connected. It is not safe to operate this unit without the Protective Conductor Terminal connected.
• Do not exceed voltage rating on the label located on the top of the instrument housing.
• Always disconnect power before changing signal and power connections.
• Do not use this instrument on a work bench without its case for safety reasons.
• Do not operate this instrument in flammable or explosive atmospheres.
• Do not expose this instrument to rain or moisture.
• Unit mounting should allow for adequate ventilation to ensure instrument does not exceed operating temperature rating.
• Use electrical wires with adequate size to handle mechanical strain and power requirements. Install without exposing bare wire outside the connector to minimize electrical shock hazards.
EMC Considerations
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use signal wire connections with twisted-pair cables.
• Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.
4
Note
☞
Note
☞
Page 13
5
S
ETUP
1.4 Getting Started
Follow these steps to start using your meter right away:
• Mount the meter (section 1.4.1)
• Connect the signal source (section 1.4.2)
• Connect outputs (section 1.4.3)
• Set DIP switches for your input level (section 1.4.4)
• Connect the ~AC power cord to the meter (section 1.4.5)
• Turn on the meter (section 1.4.6)
1. Remove the rear cover at the back of the unit.
2. Cut a hole in your panel, as shown. (see Figure 1.1)
Figure 1.1 — Panel Cut-out Dimensions
1.4.1 Mount the Meter
Overview
1
PANEL THICKNESS
6,4 (.25) MAX 0,8 (.03) MIN
1,5
R
(.06)
4 PLCS
92,00 + 0,81/–0,00
(3.622 + .032/–.000)
NOTE: Dimensions in Millimeters (Inches)
45,00 + 0,61/-0,00
(1.772 + .024/–.000)
Page 14
1
S
ETUP
3. Insert the meter into the hole. Be sure the front bezel is flush to your panel. (see Figure 1.2)
Figure 1.2 — Exploded View of the Meter
6
P
A
N
E
L
C
U
T
-O
U
T
CONNECTOR
LABEL
REAR COVER
(REMOVED)
"N
E
O
M
R
B
G
IN
T
N
U
T
E
K
C
A
E
L
Y
T
" S
W
PRODUCT
LABEL
FRONT BEZEL
GASKET
CASE
B
"O
L
O
M
C
A
R
G
IN
T
N
U
E
K
S
C
P
2
T
E
L
Y
T
" S
R
E
D
Page 15
7
S
ETUP
1
1.4.2 Connect Signal Source
!
WARNING: Do not connect ac power to your meter until you have completed all input and output connections. Failure to do so may result in injury!
!
IMPORTANT: This device must only be installed electrically by specially trained electrician with corresponding qualifications. The mains power input to the unit as well as the AC input signal to be measured must agree with the wiring instruction.
1. Locate the P6 connector. (see Figures 1.3 and 1.4)
2. Make sure the signal source is off.
3. Insert the correct wire in each terminal and tighten the lockdown screw. Tug gently on the wires to verify that the connections are secure.
Measurement Category II. Rated Maximum Working Voltage: 300 Vac/dc. No CE compliance from 300V - 750V. See CE Approvals section for more details.
Figure 1.3 — P6 Volt Meter Connection
Page 16
1
S
ETUP
Measurement Category II. Rated Maximum Working Current: 5A.
Figure 1.4 — P6 Current Meter Connection
Measurement Category II. Rated Maximum Working Current: 5A.
Figure 1.4.1 - P6 Current Meter Connection with Optional Current Transformer
8
Page 17
9
S
ETUP
Your meter may have one or more factory installed outputs. The Dual Relay Output Connection and the Analog Output Connection are shown below. (see Figures 1.5 and 1.6)
Figure 1.5 — P2 & P3 Dual Relay Output Connection
Figure 1.6 — P4 Analog Output Connection
1.4.3 Connect Outputs
1
Page 18
1
S
ETUP
Your meter may also have one of two field installable communication outputs — either a RS232 or RS485 Output Connection. (see Figures
1.7 and 1.8)
Figure 1.7 — P10 & J5 RS232 Output Connections
Table 1.1 — RS232 Meter Connection to the Computer
Figure 1.8 — J5 Half-duplex RS485 Output Connections
Meter (DCE) Computer (DTE)
Pin Signal / Function RJ12 D9 D25
RTS, meter from computer 2 7 4
TX, meter = RX, computer 3 2 3
RX, meter = TX, computer 4 3 2
GND 5 5 7
NC (not connected) 1,6 (all others)
10
Page 19
11
S
ETUP
Locate the SW1 DIP switch on the back of the meter. (see Figure 1.9) To select an input range, SW1-1 through SW1-9 must be in the correct position.
Please refer to the tables on page 16 for the switch positions that match your desired input range. Note that you must also configure the Input menu to reflect your desired input range.
Figure 1.9 — SW1 Input Range Selection
1. Remove the panel at the back of the meter.
2. Locate the P1 connector. (see Figure 1.10)
!
WARNING: Do not connect AC power to your meter until all input and output connections have been properly established. Failure to do so may result in injury.
3. Insert the correct wire in each terminal and tighten the lockdown screw. See Table 1.2, for wire color definitions. Tug gently on the wires to verify that the connections are secure.
1.4.5 Connect the AC Power Cord to the Meter
1.4.4 Set DIP Switches
1
Note
☞
Page 20
1
S
ETUP
Figure 1.10 — P1 AC Power Connection
Table 1.2 — P1 AC Power Wire Color & Terminal Connections
1. Plug the unit into a properly grounded 115V power supply. The unit will initialize, scrolling one of the following messages on the front panel:
TRUE…RMS…VOLT…VER.X TRUE…RMS…CURRENT…VER.X
X = the version number
2. The present value should then display. If a value does not display, follow these steps:
• Unplug the unit
• Verify the P1 power connections
• Check your power source
• Plug the unit in again
1.4.6 Turn On the Meter
CONNECTION EUROPE USA
AC-Line Brown Black AC-Neutral Blue White AC-Earth Ground Green/Yellow Green
12
Page 21
13
O
PERATION
Section 2: Operation
Introduction
SW1 DIP Switch Positions
Front Panel Display
Meter Buttons
Flow Charts
Menu Configurations
2
Page 22
2
O
PERATION
2.1 Introduction
The meter has two different modes of operation. The first, Run Mode, is used to display values for setpoint 1 and setpoint 2 and display or clear peak and valley values. The other mode, Configuration Mode, is used to navigate through the menus options and configure the meter.
This section of the manual, Operation, will explain both the Run Mode and the Configuration Mode and is divided into the following sections:
• SW1 DIP Switch Positions (Section 2.2)
• Front Panel Display (Section 2.3)
• Meter Buttons (Section 2.4)
• Menu Flow Charts (Section 2.5)
• Menu Configurations (Section 2.6)
There are two meter models available — a current meter and a volt meter. Both of these models are available from the factory in the following configurations:
• Basic Unit
• Basic Unit with Dual Relay Output
• Basic Unit with Isolated Analog Output
• Basic Unit with Dual Relay and Analog Output
The relay and analog outputs are not field installable options. They must be requested at time of purchase. Menus will only be available for those options that are supported by that meter.
Note
☞
Overview
14
Page 23
15
O
PERATION
2.2 SW1 DIP Switch Positions for Input Ranges
Before you change the input range of the meter, set the SW1 DIP switches to the correct positions. These switches are located at the back of the meter. Refer to the following tables for the volt meter and current meter settings.
* This range is fully calibrated only when ordered as an FS option. Otherwise only a coarse calibration ± 2% applies.
Table 2.1 — DIP Switch Positions for Voltage Meter
Table 2.2 — DIP Switch Positions for Current Meter
!
WARNING:The 5 Amp and 1 Amp ranges have dedicated inputs. You must always choose the input connector that corresponds to your input range. Failure to do so may
result in injury!
2.2.2 Adjusting SW1 DIP Switches for Current Meter
2.2.1 Adjusting SW1 DIP Switches for Volt Meter
2
Range
750V
100V
10V
1V
0.1V*
SW1-1 to
SW1-5
Open
Open
Open
Open
Open
SW1-6
Open
Closed
Closed
Closed
Open
SW1-7
Closed
Closed
Open
Open
Open
SW1-8
Open
Open
Closed
Open
Open
SW1-9
Closed
Open
Open
Open
Open
SW1-7 to
Open
Open
Open
Open
Open
Closed
SW1-4
Closed
Open
Open
Open
Open Closed
~ AC coupled
Open
Open
Open
Open
Closed
Closed
Closed
Closed
Closed
== DC coupled
SW1-1
Range SW1-5 SW1-6
5A Open
1A Closed
100mA Closed
10mA Closed
1mA
SW1-2
Open
Open
Closed
Open
Open
SW1-3
Open
Open
Open
Closed
Open
SW1-9
Open
Open
Open
Open
Open
Page 24
2
O
PERATION
2.3 Front Panel Display
Figure 2.1 — Front Panel Display for Volt Meter
Figure 2.2 — Front Panel Display for Current Meter
Table 2.3 — Display Indicators
These indicators light when appropriate:
1 Setpoint 1 indicator 2 Setpoint 2 indicator mA, A Milliamp, Amp for Current Meter mV, V Millivolt, Volt for Voltmeter OVRG Overflow indicator
2.3.3 Indicators
2.3.2 Current Meter Display
2.3.1 Volt Meter Display
16
V
Page 25
17
O
PERATION
2.4 Meter Buttons
Table 2.4 — Button Functions
BUTTON DESCRIPTION
SETPTS In Run Mode press this button to display the values for setpoint 1 and
setpoint 2. Press this button repeatedly to recall the previous setpoint settings. After using the /MAX and /MIN buttons to alter these settings, press this button again to store the new values. (Unless you press this button within 20 seconds to store your input, the meter will scroll to setpoint 2 and retain the last value stored.)
In Configuration Mode press this button at the level 1 menus only followed by MENU to enter into the level 2 menus.
/MAX In Run Mode press this button to recall the peak reading since the last
time the reset button was pressed. Press this button again to return to the current readings without resetting the peak reading.
In Configuration Mode press this button to access Level 4 menu items, then press the /MIN button to scroll through Level 4 menu items.
/MIN In Run Mode press this button to recall the valley reading since the last
time the reset button was pressed. Press this button again to return to the current readings without resetting the valley reading.
In Configuration Mode press this button to access Level 3 menu items. When configuring setpoint values, press this button to scroll to the next digit.
MENU In Run Mode press this button to end the current measuring process and
enter configuration mode.
In Configuration Mode pressing this button will store your selections and then advance you to the next menu item.
RESET In Run Mode press MENU button, followed by RESET button to reset
the latched alarms. (If your meter has this option).
In Configuration Mode press this button to go back to the previous menu item. Press this button twice to exit the configuration
2
Page 26
2
O
PERATION
18
2.5 Flow Charts
The following flow charts list all of the menus, submenus, and their respective ranges. The first flow chart demonstrates how to adjust setpoints while in Run Mode. The second and third charts contains all of the menus available while in Configuration Mode. (The third chart is only for those meters that have a communication board installed.)
Figure 2.3 — Flow Chart for Setpoint Configuration (Run Mode)
press
SETPTS
SP1
-199to9999
press /MAX until the desired value is displayed, then press /MIN to advance to the next digit
SP2
-199to9999
RUN
2.5.1 Setpoints and Menu Configuration Flow Charts
Page 27
19
O
PERATION
2
Figure 2.4 — Flow Chart for Configuration Mode, Part 1
AC for volt meter
DC
IN 1 RD 1 IN 2 RD 2
0.1V 1V 10V 100V 750V
or 1mA or 10mA or 0.1A or 1A or 5A
or
or
or
INPT
F.FFF
FF.FF
FFF.F
FFFF.
64
16
0 4
8
RDNG
mV
VA
mA
* TO SWITCH BETWEEN FIELDS PRESS ITEMS MARKED WITH “
†” OR APPEARING IN PARENTHESIS ARE DEFAULT VALUES
60Hz
50Hz
ALL
-- --
-1999 - 9999 (9999)
-1999 - 9999 (0)
-1999 - 9999 (0)
-1999 - 9999 (9999)
/MIN
/MIN
/MIN
/MIN
/MIN
/MIN
/MIN
MENU LEVEL 1 MENU LEVEL 2 MENU LEVEL 3 MENU LEVEL 4
PRESS PRESS SETPTS ONCE
PRESS
PRESS
/MIN
/MAX
MENU
THEN PRESS MENU
FOR EACH LEVEL 1 ITEM
†
†
†
††
†
for current meter
†
for volt meter
†
for current meter
†
†
RD.S.O.
RNGE
DEC.P
CPLG
FLTR
UNIT
FREQ
ABOV
BELO
NORM
LTCH
LTCH
NORM
BELO
ABOV
SP.CF
0 - 9999 (3 count)
0 - 9999 (3 count)
PROG
DFLT
CURR
VOLT
OUT
OUT2
RD 2
OUT1
RD 1 -1999 - 9999
0 to 10.00V or 0 to 20.00mA
-1999 - 9999 0 to 10.00V or 0 to 20.00mA
*
*
*
*
*
*
†
†
†
†
† †
SPT.1
SPT.2
S1.DB S2.DB
OT.CF
OT.S.O.
Page 28
2
O
PERATION
20
Figure 2.5 — Flow Chart for Configuration Mode, Part 2
300 600 1200 2400 4800 9600
†
19.2K NO
ODD
†
EVEN
NEW.P
†
M.BUS
CMD
†
CONT
PRTY
DATA STOP
PROT
MODE
C.SUM
YES
NO
†
ECHO
LF
SPACE † CR
SEPR
BAUD
STND
485
232C
†
NO YES
†
YES
NO
†
NO YES
†
YES
NO
†
PEAK
NO † YES
STAT RDNG
TIME
VALY
YES
NO
†
DATE
NO † YES
YES
NO
†
NO † YES
YES
NO
ID
UNIT
0 - 199 (-01)
0 - 9999 SEC. (1)
COMM
1BIT † 2BIT
7BIT † 8BIT
COMP
BUS.F
DAT.F
TX.TI
ADDR
MENU LEVEL 1 MENU LEVEL 2 MENU LEVEL 3 MENU LEVEL 4
PRESS PRESS SETPTS ONCE
PRESS
PRESS
/MIN
/MAX
MENU
THEN PRESS MENU
FOR EACH LEVEL 1 ITEM
ITEMS MARKED WITH “
†” OR APPEARING IN PARENTHESIS ARE DEFAULT VALUES
Page 29
21
O
PERATION
2.6 Menu Configurations
Both meters have at least two main menus — Input and Reading. If your meter includes the Dual Relay capability, then it will have a Setpoint Configuration menu. If your meter includes the Analog Output capability, then it will have an Output menu. If your meter has the optional communications board installed, then it will have a Communication Configuration menu. Each of these main menus has submenus.
Menu Name Description
INPT From the Input menu and its submenus you may
choose the range, frequency, and input coupling.
RDNG From the Reading menu and its submenus you may
select decimal point position, filter time constant, display the unit of measure, and adjust the reading scale and offset.
SP.CF From the Setpoint Configuration menu and its
submenus you may specify the active above/below or normal/latch for the alarms and their corresponding dead band.
OUT From the Output menu and its submenus you may
choose analog out for current mode or voltage mode and you may adjust the output levels appropriate to your application.
COMM From the Communication Configuration menu and its
submenus you may set the communications parameters for digital output.
Table 2.5 — Menu Descriptions
The following pages will show, step by step, how to configure all of the menus for your particular application. You may wish to refer to the flow charts on the previous two pages while configuring the menus.
Overview
2
Page 30
2
O
PERATION
2.6.1.0 INPUT MENU
From this menu you may choose the electrical range, frequency, and input coupling. Follow these steps:
1. Press MENU until the meter displays:
INPT
2. Press SETPTS, followed by MENU until the meter displays:
RNGE
2.6.1.1 Input Menu: Electrical Range
Before selecting a range other than the default range of 750V (5A), refer to the Tables in Section 2.2. to set the SW1 DIP switches. To select an input range, follow these steps.
3. Press /MIN. The meter flashes the input range. Press /MAX to scroll through these choices until you find the range that suits your application: Note that there are default decimal point positions and units of measure associated with each selected range. If you change the input range, those two items will be updated automatically according to Section 4.7.15, in Table 4.19. Note that
at any time you can change the decimal point position and or the unit of measure to any position or value you want.
volt meter current meter
0-1V 1 mA
1V 10mA
10V 0.1 A 100V 1 A 750V 5 A
4. Press MENU to select the input range shown. After storing this setting, the meter displays:
FREQ
INPT
RNGE
INPT
22
Page 31
23
O
PERATION
2.6.1.2 Input Menu: Electrical Frequency
To select an electrical frequency, follow these steps. The factory default for electrical frequency is 60Hz if the meter is powered from a 100V power source and 50Hz if the meter is powered from a 220V power source.
5. Press /MIN. The meter flashes the current frequency selection. Press /MAX to scroll through these choices:
60HZ 110V Power 50HZ 220V Power ALL select if the signal frequency is not the
same as the power frequency
6. Press MENU to select the frequency shown. After storing this setting, the meter displays:
CPLG
2.6.1.3 Input Menu: Coupling
To select AC or DC Coupling, follow these steps. The factory default is AC. Note that volt meters have two different inputs for AC and DC coupling. Current meters have only DC coupling. To AC couple the current meter, SW1-5 DIP should be open. (See Table 2.2)
7. Press /MIN. The meter flashes the existing coupling selection. Press /MAX to toggle between these choices:
AC measures only AC component of the input
signal
DC measures AC and DC component of the
input signal.
8. Press MENU to select the coupling shown. After storing this setting, the meter displays:
RDNG
INPT
CPLG
INPT
FREQ
2
Page 32
2
O
PERATION
2.6.2.0 Reading Menu
From this menu you may choose the decimal point position, filter time constant, display unit of measure, and adjust reading scale and offset.
1. Press SETPTS, followed by MENU until the meter displays:
DEC.P
2.6.2.1 Reading Menu:
Decimal Point Position
To select the decimal point location, follow these steps. FFF.F is the factory default.
2. Press /MIN. The meter flashes the current decimal point position. Press /MAX to scroll through these choices:
F.FFF FF.FF FFF.F FFFF.
3. Press MENU to select the decimal point location shown. After storing this setting, the meter displays:
FLTR
2.6.2.2 Reading Menu:
Filter Value
To select a filter value, follow these steps. The factory default is 64.
4. Press /MIN. The meter flashes the filter value. Press /MAX to scroll through these choices:
0 4 8 16 64
5. Press MENU to select the filter value location shown. After storing this setting, the meter displays:
UNIT
RDNG
FLTR
RDNG
DEC.P
RDNG
24
Page 33
25
O
PERATION
2.6.2.3 Reading Menu:
Unit Value
To select a unit value, follow these steps. The factory default is V for volt meters and A for current meters.
6. Press /MIN. The meter flashes the unit value. Press /MAX to scroll through these choices:
volt meter current meter
——
(blank)
VA
mVmA
7. Press MENU to select the unit value location shown. After storing this setting, the meter displays:
RD.S.O.
2.6.2.4 Reading Menu:
Reading Scale and Offset
This step is generally not required since the meters are precalibrated for all ranges. Rescale only if the desired display is different than the actual voltage or current being measured. Typically, this step would apply only to the 5A or 100mV range.
To scale the meter to show readings in engineering units follow these steps. There are two different methods. One method is on-line scaling. The other method is manual scaling. On-line scaling will be described first.
ON-LINE SCALING — This is the most common method since the meter will display actual input from which you may determine the value to be displayed.
1. Set the SW1 DIP Switch Positions if necessary.
Before you begin this procedure, you must set the SW1 DIP switches to the correct positions. Failure to do so may result in an incorrect reading. The SW1 DIP switches are located at the
back of the meter. For voltage and current input SW1 switch positions, refer to the appropriate Table in Section 2.2. For best resolution, find the maximum signal that will be applied to the meter.
Note
☞
RDNG
UNIT
2
RDNG
RD.S.O.
on-line
scaling
Page 34
2
O
PERATION
26
2. Set decimal point position if necessary. The decimal point if configurable under the first submenu of the Reading menu.
3. Press the MENU button, if necessary, until the meter shows:
RD.S.O
4. Press the /MIN button. The meter shows:
IN 1
5. Press the /MIN button to show the existing value at Input 1, with the fourth digit flashing.
6. Apply the low value at the input and press the /MIN button again to show the actual value at the input without a flashing digit. Wait until the value has settled.
Do not apply zero.
7. Press the MENU button. The meter shows:
RD 1
8. Press the /MIN button to show the existing value of Display Reading for Input 1, with the fourth digit flashing.
9. Press the /MIN and /MAX buttons to enter the desired Display Reading for Input 1.
10. Press the MENU button. The meter shows:
IN 2
11. Press the /MIN button to show the existing value at Input 2, with the fourth digit flashing.
12. Apply the second value at the input and press the
/MIN button again to show the actual value at the
input without a flashing digit. Wait until the value has
settled.
13. Press the MENU button. The meter shows:
RD 2
14. Press the /MIN button to show the existing value of Display Reading for Input 2, with the fourth digit flashing.
15. Press the
/MIN and /MAX buttons to enter the
desired Display Reading for Input 2
RD 2
IN 1 IN 2
Read 1
Read 2
Input 1 Input 2
Display Reading versus Input Signal
Display
Reading
RD 1
Input
Signal
When rescaling the meter either with or without known loads, if a high degree of accuracy is required, do not apply 0 volts or use 0 as a reference point.
For any rescaling, the minimum signal should always be 3 to 5% of the full scale of the signal range. The same is true for scaling the analog output option. Do not use a value of 0 on the display if a high degree of accuracy is required, but instead use a value slightly higher than 0 counts on the display.
Note
☞
Page 35
27
O
PERATION
16. Press MENU to store the values. After storing these settings, the meter will display one of four items.
If your meter is equipped with a Dual Relay capabilities, then your meter will now display:
SP.CF GO TO 2.6.3.0 SETPOINT MENU
If your meter does not support Dual Relay capabilities, but has Analog Output capabilities, then you are may skip the next section concerning Dual Relay Output and proceed to the section concerning the Analog Output Menu. Your meter will now display:
OUT SKIP NEXT MENU ITEM AND GO TO
2.6.4.0 OUTPUT MENU
If your meter does not support Dual Relay capabilities or Analog Output capabilities, but has the communications board option installed, then you are may skip the next two sections concerning Dual Relay Output and Analog Output. Proceed to the section concerning the Communications Menu. Your meter will now display:
COMM SKIP NEXT TWO MENU ITEMS AND GO
TO 2.6.5.0 COMMUNICATIONS MENU
If your meter is not equipped with Dual Relay or Analog Output
capabilities and does not have a communications board option installed, then you are finished configuring your meter. It will now display:
RUN if no values were changed RST if values were changed
METER CONFIGURATION IS COMPLETE
2
Page 36
2
O
PERATION
MANUAL SCALING — This is used when you want to enter the values for Input 1 (IN1) and Input 2 (IN2) as well as the Display Readings for Input 1 (RD1) and Input 2 (RD2) manually.
1 Press the /MIN button. The meter shows:
IN 1
2. Press the /MIN button to show the existing value at Input 1, with the fourth digit flashing.
3. Press the /MAX button.
4. Press the /MIN and /MAX buttons to enter the desired value for Input 1.
5. Press the MENU button. The meter shows:
RD 1
6. Press the /MIN button to show the existing value of Display Reading for Input 1, with the fourth digit flashing.
7. Press the /MIN and /MAX buttons to enter the desired Display Reading for Input 1.
8. Press the MENU button. The meter shows:
IN 2
9. Press the /MIN button to show the existing value at Input 2, with the fourth digit flashing.
10. Press the /MAX button.
11. Press the /MIN and /MAX buttons to enter the desired value for Input 2.
12. Press the MENU button. The meter shows:
RD 2
13. Press the /MIN button to show the existing value of Display Reading for Input 2, with the fourth digit flashing.
14. Press the /MIN and /MAX buttons to enter the desired Display Reading for Input 2.
15. Press MENU to store the values. After storing these settings, the meter will display one of four items.
If your meter is equipped with a Dual Relay capabilities, then your meter will now display:
SP.CF go to 2.6.3.0 SETPOINT menu
RDNG
RD.S.O.
manual
scaling
28
Page 37
29
O
PERATION
If your meter does not support Dual Relay capabilities, but has Analog Output capabilities, then you are may skip the next section concerning Dual Relay Output and proceed to the section concerning the Analog Output Menu. Your meter will now display:
OUT skip next menu item and go to
2.6.4.0 Output menu
If your meter does not support Dual Relay capabilities or Analog Output capabilities, but has the communications board option installed, then you are may skip the next two sections concerning Dual Relay Output and Analog Output. Proceed to the section concerning the Communications Menu. Your meter will now display:
COMM GO TO 2.6.5.0
COMMUNICATIONS MENU
If your meter is not equipped with Dual Relay or Analog Output capabilities and does not have a communications board option installed, then you are finished configuring your meter. It will now display:
RUN if no values were changed RST if values were changed
meter configuration is complete
GAIN MULTIPLIER TABLE
CURRENT MODEL VOLTAGE MODEL
RANGE MULTIPLIER TOTAL COUNTS RANGE MULTIPLIER TOTAL COUNTS
5 A 1,000 cts/A 10,000 or 9999 750 V 10 cts/V 10,000 or 9999
(10 A f.s.) (1,000 V f.s.)
1 A 10,000 cts/A 100 V 100 cts/V
100 mA 100 cts/mA 10 V 1,000 cts/V
10 mA 1,000 cts/mA 1 V 10,000 cts/V
1 mA 10,000 cts/mA 0.1 V 100 cts/mV
2.6.3.0 Setpoint Menu:
This menu is only available for meters equipped with the Dual Relay Output option. From this menu you may set the active band,
alarm zones, and deadband values for both setpoints.
1. Press SETPTS, followed by MENU until the meter displays:
SPT.1
SP.CF
2
Page 38
2
O
PERATION
2.6.3.1 Setpoint Menu:
Setpoint 1 — Band & Alarm
To set the active band and alarm zone for setpoint 1, follow these steps. The ABOV active band position sets the band above the setpoint and BELO sets it below the setpoint. The NORM alarm position indicates the alarm is unlatched. The LTCH alarm position indicates the alarm has to be reset if it is energized.
2. Press
/MIN. The meter flashes the last saved active band
position. Press
/MAX to toggle between these choices:
ABOV BELO
3. Press /MIN. The meter flashes the last saved alarm position. Press
/MAX to toggle between these choices:
NORM LTCH
4. Press MENU to save the active band and alarm position. After storing these settings, the meter displays:
SPT.2
2.6.3.2 Setpoint Menu:
Setpoint 2 — Band & Alarm
To set the active band and alarm zone for setpoint 2, follow these steps.
5. Press /MIN. The meter flashes the last saved active band position. Press
/MAX to toggle between these choices:
ABOV BELO
6. Press /MIN. The meter flashes the last saved alarm position. Press
/MAX to toggle between these choices:
NORM LTCH
7. Press MENU to save the active band and alarm positions for setpoint 2. After storing these settings, the meter displays:
S1.DB
SP.CF
SPT.2
SP.CF
SPT.1
30
Page 39
31
O
PERATION
2.6.3.3 Setpoint Menu:
Setpoint 1 — Deadband
To set the deadband value for setpoint 1, follow these steps:
8. Press /MIN. The meter shows the last saved deadband value, with the left-most digit flashing. Press /MIN and /MAX to program a deadband value.
9. Press MENU to save the deadband value for setpoint 1. After storing this setting, the meter displays:
S2.DB
2.6.3.4 Setpoint Menu:
Setpoint 2 — Deadband
To set the deadband value for setpoint 2, follow these steps:
10. Press /MIN. The meter shows the last saved deadband value, with the left-most digit flashing. Press /MIN and /MAX to program a deadband value.
11. Press MENU to save the deadband value for setpoint 2. After storing this settings, the meter will display one of three items.
If your meter is equipped with the Analog Output option, then your meter will now display:
out go to next menu item
2.6.4.0 Output Menu
If your meter is not equipped with the Analog Output option, but your meter is equipped with a communications board, then your meter will now display:
COMM SKIP THE NEXT MENU ITEM
GO TO 2.6.5.0 COMMUNICATIONS MENU
if your meter is not equipped with the Analog Output option, and
your meter is not equipped with a communications board, then you are finished configuring your meter. It will now display:
RUN if no values were changed RST if values were changed
METER CONFIGURATION IS COMPLETE
SP.CF
S2.DB
SP.CF
S1.DB
2
Page 40
2
O
PERATION
2.6.4.0 OUTPUT MENU
This menu is only available for meters equipped with the Analog Output option. From this menu you may select current or voltage analog output, select regular or programmable analog output. Note that regular output
uses the default ranges for Output Scale and Offset. Programmable output allows you to set up different ranges. If you select programmable output, the OT.S.O. submenu displays, allowing you to enter values manually.
1. Press SETPTS, followed by MENU until the meter displays:
OT.CF
2.6.4.1 Output Menu: Analog Power Output Options
To set the power output option to either current or voltage, follow these steps.
2. Press /MIN. The meter shows the last saved power output option. Press /MAX to toggle between these choices:
VOLT CURR
3. Once the correct power output option is shown, press /MIN. The meter shows the last saved programming option. Press /MAX to toggle between these choices:
DFLT PROG
4. Press MENU to save the Analog Power Output Options. After storing these settings, the meter will display one of three items.
If you selected the programmable option in step 3, your meter will display:
O.T.S.O GO TO 2.6.5.0
COMMUNICATIONS MENU
If you did not select the programmable option in the previous step, and your meter is equipped with a communications board, then you may skip the next section concerning Output Scale and Offset and proceed to the Communications section
2.6.5.0. Your meter will now display:
COMM GO TO 2.6.5.0
COMMUNICATIONS MENU
If you did not select the programmable option in the previous step, and your meter is not equipped with a communications board, then you are finished configuring your meter. It will now display:
RST METER CONFIGURATION IS COMPLETE
OUT
OT.CF
OUT
32
Page 41
33
O
PERATION
2.6.4.2 Output Menu: Reading Scale & Offset
Follow these steps to program the Output Reading Scale and Offset.
1. Press
/MIN. The meter shows:
RD 1 (THIS IS YOUR FIRST POINT OF DISPLAY
READING
.)
2. Press
/MIN again to show the last Read 1 value, with the
left-most digit flashing. Press
/MIN and /MAX to enter
the desired value.
3. Press /MIN. The meter shows:
OUT1 (THIS STARTING ANALOG SIGNAL
CORRESPONDS TO YOUR
READ 1 DISPLAY.)
4. Press
/MIN again to show the last Out 1 value, with the
left-most digit flashing. Press
/MIN and /MAX to enter
the desired value.
5. Press /MIN. The meter shows:
RD 2 (THIS IS YOUR SECOND POINT OF DISPLAY
READING
.)
6. Press
/MIN again to show the last Read 2 value, with the
left-most digit flashing. Press
/MIN and /MAX to enter
the desired value.
7. Press /MIN. The meter shows:
OUT2 (THIS ANALOG SIGNAL CORRESPONDS TO
YOUR
READ 2 DISPLAY.)
8. Press
/MIN again to show the last Out 2 value, with the
left-most digit flashing. Press
/MIN and /MAX to enter
the desired value.
OUT
OT.S.O.
2
20mA
4mA
RD 1 RD 2
Read 1 Read 2
Analog Output versus
Display Reading
Display
Reading
Analog Output
Page 42
2
O
PERATION
9. Press MENU to save the Reading Scale and Offset Values. After storing these values, the meter will display one of two items.
If your meter is not equipped with a communications board, then your meter will now display:
rst METER CONFIGURATION IS COMPLETE
If your meter is equipped with a communications board, then your meter will now display:
COMM CONTINUE
2.6.5.0 Communications Menu
The Communications menu is only available if your meter has a communication board installed. There two types of communication items: 1) Items you can access through the front panel or through serial communications; and 2) Items you can access only through serial communications. With the communication board installed, all menu items configurable through the front panel push buttons are configurable through serial communication as well.
From this menu you may configure the communications protocol, bus format, data format, address, and transmission time interval.
To configure the communications protocol, follow these steps.
1. Press MENU until the meter displays:
COMM
2. Press SETPTS, followed by MENU until the meter displays:
COM.P
COMM
34
Page 43
35
O
PERATION
2.6.5.1 Comm Menu: Computer Interface
3. The first submenu item controls the baud rate. Press /MIN. The meter will display:
BAUD
Press /MAX to scroll through these choices:
300 600 1200 2400 4800 9600 19.2K
Once you have selected the appropriate baud rate, you may choose to store the value and move on to BUS.F, or continue setting the other three submenus for the COM.P menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store the baud rate.
4. To adjust parity, press /MIN again. The meter will display:
PRTY
Press /MAX to scroll through these choices:
NO ODD EVEN
Once you have selected the appropriate parity setting, you may store the value and move on to BUS.F, or continue setting the other two submenus for the COM.P menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store the parity setting.
5. To adjust data bits, press /MIN again. Meter will display:
DATA
Press /MAX to scroll through these choices:
7.BIT 8.BIT
When you 8 bit is selected, you must choose no parity. Once you have selected one or the other, you may store the value and move on to BUS.F, or set the last submenu for the COM.P menu item. Unless you have chosen 7 data bits and no parity, the number of stop bits should always be 1. The meter will set the appropriate number of stop bits.
COMM
COM.P
DATA
COMM
COM.P
PRTY
COMM
COM.P
BAUD
2
Page 44
2
O
PERATION
36
6. Press MENU to store the data bits setting. Press /MIN again. The meter will display:
STOP
Press /MAX The meter will automatically select the appropriate number of stop bits based on the parity and data settings you have chosen. You can not manually change this item.
Once you have selected one or the other, store the current settings and move on to BUS.F by pressing MENU.
2.6.5.2 Comm Menu: Bus Settings
To configure the bus format, follow these steps.
7. The first submenu item, checksum, is an error detection scheme that you choose to activate or disable. Press
/MIN. The meter
displays:
C.SUM
Press /MAX to toggle between these two choices:
YES NO
Once you have activated or disabled error detection, you may choose to store the setting and move on to DAT.F, or continue setting the other six submenus for the BUS.F menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store your checksum selection.
8. The second submenu item, linefeed, determines whether the meter will print on every line or on every other line. Choosing yes will cause data to be printed on every other line. Press
/MIN. The
meter displays:
LF
Press /MAX to toggle between these two choices:
YES NO
Once you have activated or disabled echo, you may choose to store the current settings and move on to DAT.F, or continue setting the other four submenus for the BUS.F menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store your selections.
COMM
BUS.F
LF
COMM
BUS.F
C.SUM
COMM
COM.P
STOP
Page 45
37
O
PERATION
9. The third submenu item, echo, determines whether the meter will echo the command parameter along with the response. Choosing yes will activate this feature. Press /MIN. The meter displays:
ECHO
Press /MAX to toggle between the two choices:
YES NO
Once you have activated or disabled echo, you may choose to store the current settings and move on to DAT.F, or continue setting the other four submenus for the BUS.F menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store your selections.
10. The fourth submenu item, cabling standard, allows you to select the type of communications standard to be used, either RS232C or RS485. Press /MIN. The meter displays:
STND
Press /MAX to toggle between the two choices:
232C 485
Once you have selected a cabling standard, you may choose to store the current settings and move on to DAT.F, or continue setting the other three submenus for the BUS.F menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store your selections.
11. The fifth submenu item, mode, determines whether the meter will operate in ommand mode (CMD) or continuous data mode (CONT). Note that while in command mode, the meter will only respond to valid commands that are transmitted to it. While in continuous mode, the meter will transmit data for a period of time specified in the last submenu, transmission interval time (TX.TI), of this main menu item (COMM).
There is no continuous mode for the RS485 standard. This prevents collisions on the bus line if there is more than one meter installed on the bus line. If you want to use the long capability of the RS485 board and you only have one meter on the bus, then set the meter for the RS232 standard and choose continuous mode. Press
/MIN. The meter displays:
MODE
COMM
BUS.F
MODE
COMM
BUS.F
STND
COMM
BUS.F
ECHO
2
continued on next page
Page 46
2
O
PERATION
Press /MAX to toggle between the two choices:
CMD CONT
Once you have selected a mode, you may choose to store the current settings and move on to DAT.F, or continue setting the other two submenus for the BUS.F menu item. If you do not want to change the last two submenus under this menu, then press MENU to store your selections.
12. The sixth submenu item, separator, determines what character will follow each piece of data when the meter is in continuous mode. Choosing CR will cause a carriage return to be inserted after each piece of data. Choosing SPCE will cause a space to be inserted after each piece of data. See Section 2.6.5.3 for examples of transmitted data. Press /MIN. The meter displays:
SEPR
Press /MAX to toggle between these two choices:
CR SPCE
Once you have chosen a separator, you may choose to store the current settings and move on to DAT.F, or set the last submenu for the BUS.F menu item. If you do not want to change the last submenu, then press the MENU button to store your selections.
13. The last submenu item under the BUS.F menu, protocol, allows you to select the type of communications protocol to be used, either Default Protocol or ModBus Protocol to be developed in the future. Press /MIN. The meter displays:
PROT
Press /MAX to toggle between the two choices:
NEW.P M.BUS
Once you have selected a protocol, store the current settings and move on to DAT.F by pressing MENU.
COMM
BUS.F
PROT
COMM
BUS.F
SEPR
38
Page 47
39
O
PERATION
2.6.5.3 Comm Menu: Data Format
The third submenu item under the COMM main menu is data format. This item controls which data is to be printed and how when using RS232 continuous mode or when using external print in command mode. (See the explanation of the BUS.F submenus, above, for details.) All items of this submenu have two possible settings — YES to enable or transmit or NO to disable or not transmit. Follow these steps to configure the data format.
14. Activating the first submenu item, status, enables the transmission of alarms and peak/valley status characters. Press /MIN. The meter displays:
STAT (refer to the example at end of Section
2.6.5.3 for detailed information)
Press /MAX to toggle between the two choices:
YES NO
Once you have selected yes or no, you may choose to store the current setting and move on to ADDR, or continue setting the other seven submenus for the DAT.F menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store your selection.
15. Activating the second submenu item, reading, enables the transmission of the reading value. Press /MIN. The meter displays:
RDNG
Press /MAX to toggle between the two choices:
YES NO
Once you have selected yes or no, you may choose to store the current settings and move on to ADDR, or continue setting the other six submenus for the DAT.F menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store your selections.
COMM
DAT.F
RDNG
COMM
DAT.F
STAT
2
Page 48
2
O
PERATION
16. Activating the third submenu item, peak, enables the transmission of the peak value. Press /MIN. The meter displays:
PEAK
Press /MAX to toggle between the two choices:
YES NO
Once you have selected yes or no, you may choose to store the current settings and move on to ADDR, or continue setting the other five submenus for the DAT.F menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store your selections.
17. Activating the fourth submenu item, valley, enables the transmission of the valley value. Press /MIN. The meter displays:
VALY
Press /MAX to toggle between the two choices:
YES NO
Once you have selected yes or no, you may choose to store the current settings and move on to ADDR, or continue setting the other four submenus for the DAT.F menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store your selections.
18. Activating the fifth submenu item, time, enables the transmission of the time of day. Press /MIN. The meter displays:
TIME
Press /MAX to toggle between the two choices:
YES NO
Once you have selected yes or no, you may choose to store the current settings and move on to ADDR, or continue setting the other three submenus for the DAT.F menu item. If you do not want to change any of the other submenus under this menu, then press MENU to store your selections.
COMM
DAT.F
TIME
COMM
DAT.F
VALY
COMM
DAT.F
PEAK
40
Page 49
41
O
PERATION
19. Activating the sixth submenu item, date, enables the transmission of the date. Press /MIN. The meter displays:
DATE
Press /MAX to toggle between the two choices:
YES NO
Once you have selected yes or no, you may choose to store the current settings and move on to ADDR, or continue setting the last two submenus for the DAT.F menu item. If you do not want to change the last two submenus under this menu, then press MENU to store your selections.
Date and Time Formats
Transmitting both the date and time provides full logging capability. The date will be transmitted in one of the following formats:
• month/day/year
• day/month/year
Refer to the book Serial Communication for further instructions.
20. Activating the seventh submenu item, unit, enables the transmission of the unit of measurement. Press /MIN. The meter displays:
UNIT
Press /MAX to toggle between the two choices:
YES NO
Once you have selected yes or no, you may choose to store the current settings and move on to ADDR, or set the last submenu for the DAT.F menu item. If you do not want to change the last submenu under this menu, then press MENU to store your selections.
COMM
DAT.F
UNIT
COMM
DAT.F
DATE
2
Page 50
2
O
PERATION
42
2
O
PERATION
21. Activating the last submenu item, ID, enables the transmission of a character representing which value is being sent. Press /MIN. The meter displays:
ID
Press /MAX to toggle between the two choices:
YES NO
Examples of Transmitted Data
A. Note that if activated, the following IDs will be sent:
R: Reading values P: Peak values V: Valley values
B. Suppose you have the following data:
Reading Value: 125.4V Peak Value: 130.8V Valley Value: 121.2V Time: 9:32:15 AM Date: 12/20/94
C. Now suppose that you have selected continuous mode
(MODE), RS232 cabling (STND), no error detection (C.SUM), and no linefeed (LF) in the bus format (BUS.F) submenus. Suppose you are including the reading, peak, valley, time, and date values.
D. if space is the selected separator (SEPR), and
ID (ID) is not enabled, and echo (ECHO) is not enabled, and unit of measure (UNIT) is enabled,
then meter will transmit:
09:32:15 12/20/94 125.4 130.8 121.2 V <CR>
E. if carriage return (CR) is enabled then meter will transmit:
09:32:15 12/20/94
125.4
130.8
121.2 V
COMM
DAT.F
ID
Page 51
43
O
PERATION
F. if ID (ID) and carriage return (CR) are enabled
then meter will transmit:
09:32:15 12/3/94 R: 125.4 P: 130.8 V: 121.2 V
Once you have selected YES or NO, store the current settings and move on to ADDR by pressing MENU.
2.6.5.4 Comm Menu: Multipoint Address
The fourth submenu item under the COMM main menu is the multipoint address. This item has no submenus. Follow these steps to configure the multipoint address.
22. When choosing RS485 (485) in the Standard (STND) submenu of the Bus Format (BUS.F) menu item, an address will be included in the transmitted data. Any command transmitted by DTE should include the address to ensure a proper response from the addressed meter.
Press /MIN and the meter shows a value from 001 to 199 with the third digit flashing. You may choose any address from 001 to 199 for the meter. Use /MAX to change the value of the flashing digit and /MIN to change which digit flashes.
The 00 address is reserved by the host for broadcasting. Any command from the DTE with an address of 00 will be performed without a response by all the meters on that bus.
Once you have selected an address, store the address and move on to TX.TI by pressing MENU to store the address.
COMM
ADDR
2
Page 52
2
O
PERATION
2.6.5.6 Comm Menu: Transmission Time Interval
The last submenu item under the COMM main menu is the transmission time interval. This item has no submenus. Follow these steps to configure the transmission time interval.
23. When choosing continuous mode (CONT) in the mode (MODE) submenu of the Bus Format (BUS.F) menu item, and RS232 (232C) in the Cabling Standard (STND) submenu of the Bus Format (BUS.F) menu item, you may specify the period of transmission in seconds.
Press /MIN and the meter will show a value from 0000 to 5999 with the fourth digit flashing. You may choose any number from 0000 to 5999 for a transmission time. Use
/MAX to change the
value of the flashing digit and /MIN to change which digit flashes.
Setting the time to 0000 (zero) seconds will cause data to be transmitted after each reading.
Once you have selected a transmission time, store all values by pressing MENU. After storing your settings, the meter displays RST indicating that the meter is initializing itself with the new values. It will now enter run mode.
2.6.5.7 Comm Menu: Items not displayed by the meter
Communications Items Not Displayed by the Meter. For complete information on communication items not displayed by the meter, refer to Serial Communications, the book that came with your communication board. These items are not accessible through the meter display, but are accessible through serial communications:
1. Recognition character: Any command issued by the computer should be proceeded by the recognition character otherwise the meter ignores the command and chooses any character in the ASCII table except “^, A, E”.
2. Time and date: These should be set or initialized through serial communication.
3. External Print: The P10 connector at the rear of the meter is for sending print commands from the meter. Whenever Pin 1 to Pin 2 of P10 are connected, a string of data is transmitted according to how the DAT.F submenu of the communication menu is programmed.
COMM
TX.TI
44
Page 53
45
O
PERATION
2.6.5.8 Accessing the Communication Board From Your Computer
To access any menu item via the communication board, you must establish serial communication between the meter and your computer. To do this, plug in either a RS232 or RS485 board into the meter and connect the board to the communication port on your computer. If you have a RS232 connection, then either serial port 1 or 2 will work. If you have a RS485 connection, then either you have to install a RS485 card inside your computer, or use a stand alone box which converts RS232 to RS485. Call our Sales Department for a high quality converter box for this purpose.
After configuring the communication menus (as described in the previous section) and connecting the cables, either, 1) use a communication software package (terminal emulator) like PROCOM and set the same parameters on your computer as your meter; or, 2) write your own software that satisfies your application needs. The following is an example of a QuickBasic program for communicating with the meter.
2
Page 54
2
O
PERATION
2.6.5.9 QuickBasic Program for Communicating with the Meter
Items in the brackets are comments only.
1000 CLS [Clear screen command] 1010 COMM$ = "COM1:" [Label for communications port
COM1] 1020 BAUD$ = "9600" [Label for choice of 9600 baud] 1030 PARITY$ = "O" [Label for odd parity] 1040 STOPBT$ = "1" [Label for one stop bit] 1050 CLOSE#1 [Get ready to reopen number 1] 1060 OPEN COMM$ + BAUD$ + "," + PARITY$ + ",7," +
STOPBT$ + "," + "CS,DS,RS" FOR RANDOM AS#1
[Opens COM1] 1070 INPUT "ENTER A COMMAND>",CMD$ [Prompts and
Labels Input] 1080 IF CMD$ <>"^AE" THEN CMD$ = "*" + CMD$ [Add
recognition character unless command is that
for setup data] 1090 PRINT CMD$ [Screen echoes command] 1100 N = 0 [Initialize watchdog] 1120 SOUND 32767,27:SOUND 32767,1:
SOUND 32767,1 [Delay 0.5 seconds] 1130 N = N + 1 [Increment watchdog] 1140 IF N <> 20 THEN 1180 [Test for timeout] 1150 INPUT "NO RESPONSE YET: TYPE 'C' TO CONTINUE,
'Q' TO QUIT",B$ [10 second timeout] 1160 IF B$ = "Q" THEN 1230 [Exit if desired] 1170 N = 0 [Restart watchdog] 1180 IF LOC(1) = 0 THEN 1120 [Stall until buffer
has contents] 1190 A$ = INPUT$(LOC(1),#1) [Read and label contents
of COM1] 1200 PRINT A$ [Put data onto screen] 1210 INPUT "TYPE 'M' FOR MORE COMMANDS OR 'Q' TO
QUIT", B$ 1220 IF B$ = "M" THEN 1070 1230 END
46
Page 55
47
A
BOUTTRUE
RMS M
EASUREMENT
Section 3: About True RMS
Measurement
Introduction
True RMS versus
Rectified-average Readings
Voltage-Divider Action of Leads
and Meter Input Impedance
Input Current Effects
Overvoltage Heating
Guarding Against AC Pickup
Common-Mode Voltage Effects
3
Page 56
3
A
BOUTTRUE
RMS M
EASUREMENT
3.1 Introduction
When voltage is applied across a resistor, the power supplied is:
Similarly, if a resistor is fed a current, the power supplied is:
For a voltage or current that has AC components, the true average power supplied is according to the formulas above, integrated and averaged over a time interval considerably longer than the period of the lowest AC component. This power, then, is proportional to the Mean Square of the voltage or current waveform.
This true power is always equal to, or greater than, that power computed from the average of the rectified voltage or current. Any AC voltage or current components when squared do not average to zero but add to the DC component squared. Correspondingly, the most useful value for a voltage or current is its Root Mean Square. RMS is the “heating” capability of the waveform.
P=
V
2
—
R
P=
I
2
—
R
48
Page 57
49
A
BOUTTRUE
RMS M
EASUREMENT
3.2 True RMS versus Rectified-average Readings
Rectify-and-average meters are less complex to build than true RMS meters. A meter with a waveform of only a single frequency (pure sine wave without DC) can be calibrated to read out the equivalent RMS value. A meter with waveforms of more than one frequency component, however, make significant errors and a true RMS meter is required for accuracy.
The figure below illustrates some examples of true RMS versus rectified-average readings for sample waveforms.
Figure 3.1 — True RMS versus Rectified-average Readings
Our meter has a complete waveform input, AC plus DC, for total RMS measurements, and another input, AC-Coupled, for measurements if the DC component is to be ignored, as when measuring a DC power supply ripple.
+V
0
-V
+V
0
-V
+V
0
-V
3
Waveform Pattern Average
1 Volt Peak Crest Factor RMS
1.414 0.707 Calibrated
1.732 0.580 -3.9%
Responding Error
for 0 error
1.000 1.000 -48% for C.F. = 4
Page 58
3
A
BOUTTRUE
RMS M
EASUREMENT
3.3 Voltage-Divider Action of Leads and Meter Input Impedance
The meter’s voltage input impedance is nearly one meg ohm in parallel with 10pF of capacitance, drawing negligible current from most voltage sources. Significant impedance from high-frequency sources should be taken into account when the greatest possible accuracy is needed. Also, consider the capacity from lead to lead and leakage resistance, if significant. Table 3.1 below gives the net loading impedance of the meter at different frequencies.
Table 3.1 — Meter Net Loading Impedance
Figure 3.2 — Approximate Input Resistance
INPUT FREQUENCY INPUT IMPEDANCE
100 Hz 1 MΩ
1 kHz 940 kΩ
10 kHz 614 kΩ
50
Source Resistance
Meter
Page 59
51
A
BOUTTRUE
RMS M
EASUREMENT
3.4 Input Current Effects
The meter’s input amplifier has a DC bias current less than 350pA for an ambient temperature less than 30°C (86°F). This current doubles about every 8°C above this point. This small current can produce small DC offset voltages in the impedance of the source. These offsets are readily measured by the meter when the AC stimulus is removed or a dummy source impedance is connected across the input terminals. The resulting offsets can be added or subtracted from the DC source value to increase the data accuracy.
Any AC pickup voltages measured in the above condition must not be added or subtracted linearly from the AC source value. They must be RMS-combined.
3.5 Overvoltage Heating
If you apply more than the rated voltage on any range it will increase the heating in the meter’s signal conditioner. The effects of momentary overvoltage will not add more than 0.02%. This overvoltage error will dissipate in a few minutes after removal of the overvoltage.
3.6 Guarding Against AC Pickup
The meter has a high input impedance to accurately measure sources which have high source impedance of up to 100kΩ. This means that the meter can respond to extraneous sources of AC voltage or current if they have significant capacitance or mutual inductance to the input leads.
For example, the space surrounding florescent lighting usually has a strong alternating electrostatic field (volts per meter). This field will push andor pull charge in any nearby exposed conductor. Within its vicinity, use shielded cable for input leads to drain away any capacitive current to ground. (Tie the shield to the input RTN terminal.) This is particularly important when measuring small signals at line frequency. Be especially careful to shunt away any pickup currents.
3
Note
☞
Page 60
3
A
BOUTTRUE
RMS M
EASUREMENT
If the signal leads must pass by equipment with large AC currents, use twisted-pair cable to minimize induced currents. You may also use magnetic shielding materials to reduce these effects, but the best solution is to reroute the signal wires away from such currents.
Ground-loop currents are another possible source of error. If the AC signal source has resistive or capacitive coupling to an AC power source, and that source is not supplying the meter, current can flow in the loop formed by the source, meter and power circuits, causing a reading error.
Whenever possible, drive the meter and the signal source from the same power outlet. When different outlets must be used, both should be from the same AC line-power phase. A convenient way to check for outlet voltage differences is with a hand-held multimeter.
3.7 Common-Mode Voltage Effects
Although DC-isolated, the meter’s RTN terminal has approximately 100pF of capacity to power-supply ground. If the signal source return has an AC voltage with respect to that ground, some current will flow in that lead, and possibly degrade the data. In this situation, reverse the connections of the source to the INPUT and RTN leads, then observe any differences in meter output. These differences in meter output may show which polarity offers superior performance.
52
Page 61
53
D
EFAULTS ANDSPECIFICATIONS
4
Section 4: Defaults and
Specifications
Back of the Meter
Main Board
W1–W3 Solder Jumpers
Signal Board
Meter Housing
Factory Default Values
Meter Specifications
Glossary
Page 62
4
D
EFAULTS ANDSPECIFICATIONS
4.1 Back of the Meter
Figure 4.1 — Volt Meter Connections
Figure 4.2 – Volt Meter Label
54
Page 63
55
D
EFAULTS ANDSPECIFICATIONS
Figure 4.3 — Current Meter Connections
Figure 4.4 – Current Meter Label
4
Page 64
4
D
EFAULTS ANDSPECIFICATIONS
4.2 Main Board
Figure 4.5 — Jumper Positions on the Main Board
Table 4.1 — Jumper Functions
Jumper Function
J7 Used for the optional serial communications card. The default is not
installed.
S1 Install where load impedance of analog output voltage is <1KΩ.
S2A Install to prevent access to menu items. If S2A is installed you may change
setpoints only and view peak and valley. The default is not installed.
S2B Install to enable access to all front panel push-buttons. Remove to disable
access to all front-panel push-buttons. The default is installed.
S2C Install S2C for calibration and make sure that S2A is removed if you want to
calibrate the meter. The default is not installed.
56
Page 65
57
D
EFAULTS ANDSPECIFICATIONS
4.3 W1–W3 Solder Jumpers
Figure 4.6 — W1–W3 Jumper Locations
Figure 4.6 shows the location of solder jumpers W1 through W3.
CAUTION: The meter has no power-on switch, so it will be in operation as soon as you apply power.
IMPORTANT: These changes must be performed by a qualified technician. If you want to change the Factory preset jumpers, do the following steps; otherwise go to Section 4.5.
Disconnect the power from the unit before proceeding.
1. Remove the main board from the case. (See Figure 4.8)
2. Locate the solder jumpers W1, W2, and W3 (located near the edge of the main board alongside the transformer).
3. If your power requirement is 115 V ac, wire jumpers W1 and W3 should be
wired, but jumper W2 should not.
If your power requirement is 230 V ac, wire jumper W2 should be wired, but jumpers W1 and W3 should be open. (See Table 4.2)
Table 4.2 W1-W3 Jumper wiring
AC VOLTAGE INSTALL REMOVE
115VAC W1, W3 W2 230VAC W2 W1, W3
4
W2W1 W3
T1
MAIN BOARD
DISPLAY BOARD
Page 66
4
D
EFAULTS ANDSPECIFICATIONS
58
4.4 Signal Board
The signal board fits on top of the main board and must be properly secured to the main board for the meter to function. See Figure 4.8.
For faster response time, remove the S1 jumper. The default is jumper installed.
Figure 4.7 — Signal Board
Note
5
☞
4
3
2
1
Page 67
59
D
EFAULTS ANDSPECIFICATIONS
4
Figure 4.8 — Exploded View of Case and Board Assembly
Page 68
4
D
EFAULTS ANDSPECIFICATIONS
4.5 Meter Housing
Figure 4.9 — Meter Housing and Panel Cut-out Dimensions
60
20,3 (.80)
151,4 (5.96)
96,0 (3.78)48,0 (1.89)
FRONT BEZEL
RETAINER
CASE
REAR COVER
SIDE VIEW
PANEL THICKNESS
6,4 (.25) MAX 0,8 (.03) MIN
1,5
R
(.06)
4 PLCS
92,00 + 0,81/–0,00
(3.622 + .032/–.000)
NOTE: Dimensions in Millimeters (Inches)
45,00 + 0,61/-0,00
(1.772 + .024/–.000)
TOP VIEW
Page 69
61
D
EFAULTS ANDSPECIFICATIONS
4.6 Factory Default Values
4.6.1 Menu Item Defaults
Menu Item Description Volt Current
RNGE Input Type 750 V** 5A
FREQ Frequency 60 HZ
*
60 HZ
*
50 HZ
†
50 HZ
†
CPLG Coupling Type ~AC ==DC
DEC.P Decimal Point FFF.F F.FFF
FLTR Filter Value 64 64
UNIT Volt / Amp V A
Table 4.3 — Menu Item Defaults
4.6.2 Analog Output Scale and Offset Defaults
Menu Item Description Volt Current
RD 1 Reading 1 000.0 0.000
OUT1 Output 1 4.00 (mA) 04.00 (mA)
RD 2 Reading 2 750.0 5.000
OUT2 Output 2 20.00 (mA) 20.00 (mA)
Table 4.4 — Analog Output Scale and Offset Defaults
* — if powered by 115V, 60 Hz power source
** -- for CE compliance, the max range is 300V, see CE Approvals Section for full details.
†
— if powered by 230V, 50 Hz power source
4
Page 70
4
D
EFAULTS ANDSPECIFICATIONS
4.7 Meter Specifications
4.7.1 Signal Input
Input Ranges: Voltage Model: 100mV, 1V, 10V, 100V, 750V*
Current Model: 1mA, 10mA, 100mA, 1A, 5A
Input Impedance for Volt Meter:
For all voltage ranges 1.0MegΩ ± 1%
Input Impedance for Current Meter:
For 5 AMP 0.05Ω ± 5% For 1 AMP 0.15Ω ± 2% For 100 mA 1.15Ω ± 1% For 10 mA 11.1Ω ± 1% For 1 mA 111.1Ω ± 1%
Table 4.5 — Signal Input Specifications
4.7.2 Display
4-digit Red or Green LED 14-segment
13.8mm (0.54”) Symbols: 8.8.8.8.
Table 4.6 — Display Specifications
62
Page 71
63
D
EFAULTS ANDSPECIFICATIONS
4.7.3 Tempco for Volt Meter
100 mV Range Typical 0.015% / °C
1,10,100, 750V Range Typical 0.01% / °C
Table 4.7 — Span Tempco for Voltmeter
4.7.4 Tempco for Current Meter
1, 10, 100 mA Range Typical 0.015% / °C 400 PPM/°C
1 AMP Typical 0.02% / °C 600 PPM/°C
5 AMP Typical 0.025% / °C 800 PPM/°C
Table 4.8 — Span Tempco for Current Meter
4.7.5 Analog to Digital
Technique Dual Slope
Internal Resolution 15 bits
Read Rate 2.5 Readings per second
Crest Factor 4
Warm Up Time to Rated Accuracy 30 minutes
Table 4.9 — Analog to Digital Specifications
4
Page 72
4
D
EFAULTS ANDSPECIFICATIONS
4.7.6 Isolated Analog Output (If Applicable)
Signal Type Current or Voltage
Signal Level Current: 10 V max compliance at 20 mA output
Voltage: 20 mA max for 0–10 V output
Function Assigned to a display range
Linearity 0.2%
Step Response Time 2–3 seconds to 99% of the final value
Table 4.10 — Isolated Analog Output Specifications
4.7.7 Communications Information (If Applicable)
RS-232: RJ12 6-wire connection; complete program setup and message display
capability; programmable to transmit current display, alarm status, and peak/valley status.
Baud Rates: 300, 600, 1200, 2400, 4800, 9600, 19200
RS-485: RJ12 6-wire connection; addressable from “000” to “199”
Baud Rates: 300, 1200, 2400, 4800, 9600, 19200
Table 4.11 — Communications Information
4.7.8 Input Power Information
ac Voltage 115/230 Vrms 10%
Frequency 50/60 Hz
Power 7 watts max.
Table 4.12 — Input Power Specifications
64
Page 73
65
D
EFAULTS ANDSPECIFICATIONS
4.7.9 Operating Environment
Operating Temperature: °0 to 60°C (32 °F to 140 °F)
(115/230 Vrms 10%)
°0 to 40°C (32 °F to 104 °F) (115/230 Vrms 15%) When Relay Output Option is used.
Storage Temperature: -40°C to 85°C
(-40°F to 185°F)
Relative Humidity: 90% at 40°C
(non-condensing)
Table 4.13 — Operating Environment
4.7.10 Mechanical
Panel Cutout: 1/8 DIN 3.62" x 1.78"
(45 x 92mm)
Weight: 1.27 lb (575 g)
Case Material Polycarbonate, 94 V-O
UL rated
Table 4.14 — Mechanical Specifications
4.7.11 Alarm Outputs (If Applicable)
2 Form C on/off relays
Latched and unlatched status configurable by software
Maximum current: 5 AMPS
Maximum Voltage: 250 Vac or 30 Vdc
Table 4.15 — Alarm Outputs
4
Page 74
4
D
EFAULTS ANDSPECIFICATIONS
4.7.12 Voltage Model Accuracy ( ≠ DC coupled inputs)
Frequency Accuracy
At 50, 60 Hz 0.1%R ±10 counts
At ≠DC 0.1%R ±15 counts
from 50Hz To 1 KHz 0.1%R ±20 counts
from 40Hz To 3 KHz 0.1%R ±30 counts
from 25Hz To 5 KHz (except 100mV) 0.1%R ±40 counts
from 25Hz To 4 KHz (100mV only) 0.1%R ±50 counts
The specifications are valid from 5% to 100% of each range.
Table 4.16 — Voltage Model Accuracy (≠DC coupled inputs)
4.7.13 Voltage Model ( ~AC coupled inputs)
Frequency Accuracy
at 50, 60 Hz 0.1%R ±10 counts
from 50Hz to 1 KHz 0.1%R ±30 counts
from 40Hz to 3 KHz 0.1%R ±40 counts
from 25Hz to 5 KHz (except 100mV) 0.1%R ±50 counts
from 25Hz to 4 KHz (100mV only) 0.1%R ±60 counts
The specifications are valid from 5% to 100% of each range.
Table 4.17 — Voltage Model Accuracy (~AC coupled inputs)
66
Page 75
67
D
EFAULTS ANDSPECIFICATIONS
4
4.7.14 Current Model ( ~AC or ≠DC coupled inputs)
Frequency Accuracy
at 50, 60 Hz 0.1%R ±10 counts
at ≠DC 0.1%R ±15 counts
from 50Hz to 1 KHz 0.1%R ±20 counts
from 40Hz to 3 KHz 0.1%R ±30 counts
from 25Hz to 5 KHz (except 5 AMP) 0.1%R ±40 counts
from 25Hz to 4 KHz (5 AMP only) 0.1%R ±50 counts
The specifications are valid from 5% to 100% of each range.
Table 4.18 — Current Model Accuracy
4.7.15 Decimal Point & Unit of Measure Defaults
Range Decimal Point Unit of Measure
Volt Meter: 100mV 99.99 mV
1V 999.9 mV
10V 9.999 V
100V 99.99 V
750V* 750.0 V * see CE Approvals section for full details.
Range Decimal Point Unit of Measure
Current Meter: 1mA 999.9 no unit
10mA 9.999 mA
100mA 99.99 mA
1A 999.9 mA
5A 5.000 A
Note that at any time you can change the decimal point position and or the unit of measure to any position or value you want.
Table 4.19 — Decimal Point & Unit of Measure Defaults
Page 76
4
D
EFAULTS ANDSPECIFICATIONS
4.7.16 Insulation
Power to Input/Output 2500 Vrms per 1 min. test
Power to Relay Outputs
2500 Vrms per 1 min. test
Relays to Inputs/Outputs 2500 Vrms per 1 min. test
RS-232/RS-485 to Inputs/Outputs
500 Vrms (720 Vdc) per 1 min. test
Analog Output to Inputs
500 Vrms (720 Vdc) per 1 min. test
Table 4.20 — Insulation Table
68
Page 77
69
D
EFAULTS ANDSPECIFICATIONS
4.8 Glossary
The substanitive portion of this glossary lists and defines all of the display messages and error display messages that may appear on your meter. Please keep this information with the Operator's Manual at all times.
Message Description
232C RS232C Standard
485 RS485 Standard
ABOV Setpoint Active Above
ADDR RS485 Device Address
ALL All Input Signal Frequencies
other than 50Hz & 60Hz
BAUD Baud Rate
BELO Setpoint Active Below
BUS.F Communication Bus Format
C.SUM Checksum
CMD Transmission on Command
COM.P Communication Parameters
COMM Communication Top Menu
CONT Continuous Transmission
CPLG Input Coupling
CR Carriage Return
CURR Analog Output Current Mode
DAT.F Communication Data Format
DATA Data Bits
Message Description
DATE Date
DEC.P Decimal Point Position
DFLT Analog Output Default Mode
ECHO Echo Response
FLTR Filter Value
FREQ Input Signal Frequency
ID Measurement Value ID
INPT Input Top Menu
LF Linefeed
LTCH Setpoint Latch
M.BUS ModBus Protocol
MODE Mode of Communication
NEW.P Protocol
NORM Setpoint Nonlatch
OT.CF Analog Output Configuration
OT.S.O Analog Output 2 Point
Calibration
OUT Output Configuration Top
Menu
PEAK Peak Value
4
Page 78
4
D
EFAULTS ANDSPECIFICATIONS
Please note that all of the following error messages flash while displayed.
Error Message Description
+OL Input Signal Overload
-OL Input Signal Underload
9999 Reading is Overflowed
ER1 2-Coordinate Programming
Error
Error Message Description
RD.OF The reading output is larger
than 11000 due to scaling.
(value) If the displayed value is
flashing, then the reading is overrange (between 10000 to 11000)
Message Description
PK.RS Reset Peak Value
PROG Analog Output
Programming Mode
PROT Communication Protocol
PRTY Parity Bit
RD.S.O Reading Scale and Offset (2
point Calibration)
RDNG Reading Configuration Top
Menu
RNGE Input Signal Range
S1.DB Setpoint 1 Deadband
S2.DB Setpoint 2 Deadband
SEPR Data Separator
SP.CF Setpoint Configuration Top
Menu
SPCE Space
Message Description
SPT.1 Setpoint 1 Deadband and
Alarm
SPT.2 Setpoint 2 Deadband and
Alarm
STAT Status Character
STND Communication Standard
STOP Stop Bits
STRD Stored
TIME Time Value
TX.TI Time Between Transmission
UNIT Unit of Measurement
VALY Valley Value
VL.RS Reset Valley Value
VOLT Analog Output Volt Mode
70
Page 79
71
CE A
PPROVALSINFORMATION
5
This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory. Current Ranges: 1mA, 10mA, 100mA, 1A, and 5A.
Voltage Ranges: 0.1V, 1V, 10V, 100V, and 750V*. *For the 750V range, see the following page of Special Safety Ratings.
Double Insulation Pollution Degree 2 Dielectric withstand Test per 1 min (50/60 Hz)
• Power to Input/Output: 2500Vrms
• Power to Relay Output: 2500Vrms
• Relay Output to Inputs/Outputs: 2500Vrms
• Isolated RS232/485 to Inputs: 500Vrms (720Vdc)
• Analog Output to Inputs: 500Vrms (720Vdc)
Measurement Category II
Category II are measurements performed on circuits directly connected to the Mains Supply (power). Maximum Line-to-Neutral working voltage is 100 Vac/dc.
This unit should not be used in Measurement Categories III and IV, but Catagory I (measurements performed on circuits not directly connected to the Mains Supply) is allowed.
For current and voltage connections, the Installation Requirements of this manual must be followed.
Transients Overvoltage Surge (1.2 / 50uS pulse)
• Input Power: 2500V
• Output Relays: 2500V
• Input/Output Signals: 500V
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement, control and laboratory.
• EMC Emissions Class B of EN61326
• EMC Immunity** Table 1 of EN61326
Note: **I/O signal and control lines require shielded cables and these cables must be located on conductive cable trays or in conduits. Furthermore, the length of these cables should not exceed 30 meters
Refer to the EMC and Safety installation considerations (Guidelines) of this manual for additional information.
Page 80
5
CE A
PPROVALSINFORMATION
72
This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory. Voltage Range of 750V The 750V is de-rated to 300V max. range to meet Double Insulation requirements for CE. No Analog or Serial Communication should be used when operating this de-rated range of 300V.
Input Range 300V max. 50/60Hz (on 750V range) Double Insulation Pollution Degree 2 Dielectric withstand Test per 1 min (50/60 Hz)
• Power to Input: 2500Vrms
• Power to Relay Output: 2500Vrms
• Relay Output to Inputs: 2500Vrms
• Isolated RS232/485 to Inputs: not used
• Analog Output to Inputs: not used
Measurement Category II
Category II are measurements performed on circuits directly connected to the Mains Supply (power). Maximum Line-to-Neutral working voltage is 300 Vac/dc.
This unit should not
be used in Measurement Categories III and IV.
For current and voltage connections, the Installation Requirements of this manual must be followed.
The CE compliance described above are null when Analog or Serial options are used or the Input Voltage is beyond the de-rated 300V max scale range.
This unit will work 300 - 750 V but the isolation is only Basic, provided that the relays are not used to activate DC loads requiring Double Insulation to any hazardous voltages (Mains).
Transients Overvoltage Surge (1.2 / 50uS pulse)
• Input Power: 2500V
• Output Relays: 2500V
• Input Signals: 2500V
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement, control and laboratory.
• EMC Emissions Class B of EN61326
• EMC Immunity** Table 1 of EN61326
Note: **I/O signal and control lines require shielded cables and these cables must be located on conductive cable trays or in conduits. Furthermore, the length of these cables should not exceed 30 meters
Refer to the EMC and Safety installation considerations (Guidelines) of this manual for additional information.
Page 81
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of one (1) year from the date of purchase. In addition to OMEGA’s standard warranty period, OMEGA Engineering will extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is returned to OMEGA.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST
of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
© Copyright 2006 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
TRADEMARK NOTICE:
®
,
omega.com
®
and are trademarks of OMEGA Engineering, Inc.
PATENT NOTICE:
This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577; 6,243,021 / CANADA 2052599; 2052600 / ITALY 1249456; 1250938 / FRANCE BREVET No. 91 12756 / SPAIN 2039150; 2048066 / UK PATENT No. GB2 249 837; GB2 248 954 / GERMANY DE 41 34398 C2.
The “Meter Case Bezel Design” is a trademark of
NEWPORT Electronics, Inc., registered in the U.S.
USED UNDER LICENSE. Other US and International Patents pending or
applied for.
®
Page 82
M2295/1106 11969ML-98 Rev D
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop on line at omega.com
TEMPERATURE

Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies

Wire: Thermocouple, RTD & Thermistor

Calibrators & Ice Point References

Recorders, Controllers & Process Monitors

Infrared Pyrometers
PRESSURE, STRAIN AND FORCE

Transducers & Strain Gauges

Load Cells & Pressure Gauges

Displacement Transducers

Instrumentation & Accessories
FLOW/LEVEL

Rotameters, Gas Mass Flowmeters & Flow Computers

Air Velocity Indicators

Turbine/Paddlewheel Systems

Totalizers & Batch Controllers
pH/CONDUCTIVITY

pH Electrodes, Testers & Accessories

Benchtop/Laboratory Meters

Controllers, Calibrators, Simulators & Pumps

Industrial pH & Conductivity Equipment
DATA ACQUISITION

Data Acquisition & Engineering Software

Communications-Based Acquisition Systems

Plug-in Cards for Apple, IBM & Compatibles

Datalogging Systems

Recorders, Printers & Plotters
HEATERS

Heating Cable

Cartridge & Strip Heaters

Immersion & Band Heaters

Flexible Heaters

Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL

Metering & Control Instrumentation

Refractometers

Pumps & Tubing

Air, Soil & Water Monitors

Industrial Water & Wastewater Treatment

pH, Conductivity & Dissolved Oxygen Instruments
Loading...