• Do not exceed voltage rating on the label located on the top of the instrument
housing.
• Always disconnect power before changing signal and power connections.
• Do not use this instrument on a work bench without its case for safety
reasons.
• Do not operate this instrument in flammable or explosive atmospheres.
• Do not expose this instrument to rain or moisture.
EMC:
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use signal wire connections with twisted-pair cables.
• Install Ferrite Bead(s) on signal wire close to the instrument if EMC problems
persist.
Mount the Unit
1. Cut a panel opening using the dimensions shown to the right.
2. Position the unit in the opening, making
sure the front bezel is flush with the
panel.
3. Install retaining clips on both sides of
the meter and tighten against the panel.
Wiring
Warning: Do not connect AC power to your device until you
have completed all input and output connections. This device
must only be installed by a specially trained electrician with
corresponding qualifications. Failure to follow all instructions
and warnings may result in injury!
1. Remove the panel at the back of the unit.
2. Locate the TB1 connector.
3. Insert the correct wire in each terminal as shown in the following figure
and tighten the lockdown screws.
4. Tug gently on the wires to verify the connections.
External Fuse Required:
Time-delay, UL 248-14 listed Time-lag, IEC 127-3 recognized
175 mA (115 Vac line) 125 mA (115 Vac line)
80 mA (230 Vac line) 63 mA (230 Vac line)
AC Powered Unit Connections
START HERE
Wiring (continued)
When using DC power, do not use internal excitation or Isolated
Analog Output for high color brightness. For low or medium
brightness, internal excitation is limited to 24 V @ 25 mA; 5 V, 10
V, 12 V @ 35 mA.
In order to maintain the same degree of protection as the AC units,
always use a Safety Agency Approval DC source with the same
Overvoltage Category and Pollution Degree.
DC Powered Unit Connections
Connect the Sensor
1. Locate the TB2 connector on the rear of the unit.
2. Attach the sensor wires and tighten the lockdown screws.
The diagram below shows the wiring for bridge sensors with internal
excitation.
Refer to the Operator’s Manual for setup requirements for other
sensor types.
Bridge Sensor with Internal Excitation
3. Tug gently on the wires to verify the connections.
4. Replace the panel at the back of the unit.
Using the Configuration Menu
To configure the meter, you use the buttons on the front panel.
To: Take This Action:
Display the Press the MENU button. The first function
Configuration Menu on the menu, INPT, displays.
Select a submenu 1. Press MENU until the function you
function want is shown.
2. Press /TARE.
The information you can change flashes.
Select a value 1. Press /NT/GRS to display the option
for that submenu you want.
function 2. Press MENU to store it.
STRD quickly flashes, indicating that
the selection has been stored in memory.
Then the next menu function displays.
Go back to previous Press RESET once.
menu function
Exit the Press RESET twice. The unit displays
Configuration RST as it reinitializes. When a numeric
Menu value displays, the unit is in run mode.
(Optionally, you can press MENU to
move through all the menu functions
until the unit reinitializes.)
2 3 4
To Set the Decimal Point (continued)
3.
Press /NT/GRS to move the decimal point to
the desired location. The factory settings is
FFFF. The other choices are F.FFF, FF.FF,
and FFF.F.
4. Press MENU to select the decimal point position shown. The
unit displays:
To Scale the Meter
You can scale the meter in one of two ways:
1. With a known load — This method uses input (load) information
sent from another device such as a scale or a simulator for
voltage or current.
2. Without a known load — This involves calculating the load
based on transducer specifications and manually entering it to
the meter.
For both methods, you must first identify the minimum input load
IN!1 and the corresponding display reading you want RD!1. Then
you identify the maximum input load IN!2 and its corresponding
display reading RD!2 .
The decimal point is for display purposes only — you set it
where you want it to display for your application.
When entering IN!1 and IN!2 values, ignore any decimal
point on the display. However, you must enter RD!1 and
RD!2 values with the decimal point in the desired position.
Scaling With Known Loads
To identify the minimum known load (IN!1 and RD!1
):
1. If it's not already shown, press MENU until the unit displays:
2. Apply the minimum known load (0%).
3. Press /TARE. The unit displays:
4. Press /TARE again. The unit displays the last setting for
IN!1
.
5. Press /TARE again. The unit displays the actual reading
being received from the sending device.
MENU SUBMENU /TARE DESCRIPTION
INPT 100M, ±50M, !1 0 V , !±5 V , 0-20*Input
DEC.P FFFF*, F.FFF, FF.FF, FFF.F Decimal Point
RD.S.O IN!1, RD!1, IN!2 , RD!2 Scale and Offset
RD.CF R.1=T*, R.1=N
Reading Configuration
R.2=0, R.2=1, R.2=2 , R.2=3, R.2=4
*
R.3=F*, R.3=U
R.4=P, R.4=G
COLR GRN, RED, AMBR Display Color
S1.CF S.1=A*, S.1=B
Setpoint 1 Configuration
S.2=U*, S.2=L
S.3=N*, S.3=G
shown if R.4=G
S2.CF S.1=A*, S.1=B
Setpoint 2 Configuration
S.2=U*, S.2=L
S.3=N*, S.3=G
shown if R.4=G
S1.DB 0003
*
Setpoint 1, Deadband
S2.DB 0003
*
Setpoint 2, Deadband
OT.CF O.1=E*, 0.1=D Analog Output
O.2=C*, 0.2=V
Configuration
O.3=A*, 0.3=P
OT.S.O RD!1, OUT1, RD!2, OUT2
Output Scale & Offset
LK.CF RS=E*, RS=D
Lockout Configuration
SP=E*, SP=D
L3=0*, L3=1
BRIT M.BrT, .BrT, H.BrT
Display Brightness
* Factory Default Settings
To Set the Input Type
1. Press MENU until the unit displays:
2. Press /TARE. The unit displays:
3. For this application you want 100M. If 100M is not displayed,
press /NT/GRS until it appears. Other choices are 50M, 10V,
5V and 0-20.
Refer to the Operator's Manual for more information on
changing ranges.
4. Press MENU to select the sensor shown. The meter displays the
next menu item. If you changed input type, the meter displays:
To Set the Decimal Point
1. If it's not already shown, press MENU until the
unit displays:
2. Press /TARE. The unit displays:
Using the Configuration Menu (continued)