Omega Engineering CN3271, CN33271 User Manual

Page 1
CN3271-add2.pmd 19/08/2004, 16.361
Page 2
Service North AmericaService North America
Service North AmericaService North America
Service North America
USA:USA:
USA:USA:
USA: ISO 9001 Certified One Omega Drive, Box 4047 Stamford CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada:Canada:
Canada:Canada:
Canada: 976 Bergar Laval (Quebec) H7L 5A1, Canada Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.com
OMEGAnetOMEGAnet
OMEGAnetOMEGAnet
OMEGAnet®
On-Line ServiceOn-Line Service
On-Line ServiceOn-Line Service
On-Line Service
www.omega.comwww.omega.com
www.omega.comwww.omega.com
www.omega.com
For immediate technical or For immediate technical or
For immediate technical or For immediate technical or
For immediate technical or
application assistance:application assistance:
application assistance:application assistance:
application assistance:
Internet e-mailInternet e-mail
Internet e-mailInternet e-mail
Internet e-mail
info@omega.cominfo@omega.com
info@omega.cominfo@omega.com
info@omega.com
USA and Canada:USA and Canada:
USA and Canada:USA and Canada:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
®
Customer Service: 1-800-622-2378 / 1-800-622-BEST
®
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
®
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico:Mexico:
Mexico:Mexico:
Mexico: En Español: (001) 203-359-7803 FAX: (001) 203-359-7807 e-mail: espanol@omega.com info@omega.com.mx
Service Europe:Service Europe:
Service Europe:Service Europe:
Service Europe:
Benelux:Benelux:
Benelux:Benelux:
Benelux: Postbus 8034, 1180 LA Amstelveen The Netherlands Tel: +31(0) 20 3472121 Fax: +31(0) 20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@omegaeng.nl
Czech Republic:Czech Republic:
Czech Republic:Czech Republic:
Czech Republic: Frystatska 184, 733 01 Karviná, Czech Republic Tel:+420(0)59 6311899 Fax:+420(0)59 6311114 Toll Free: 0800-1-66342 e-mail: info@omegashop.cz
France:France:
France:France:
France: 11, rue Jacques Cartier, 78280 Guyancourt, France Tel: +33 (0)1 61 37 29 00 Fax: +33 (0)1 30 57 54 27 Toll Free in France: 0800 466 342 e-mail: sales@omega.fr
Germany/Austria:Germany/Austria:
Germany/Austria:Germany/Austria:
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Tel: +49 (0)7056 9398-0 Fax: +49 (0)7056 9398-29 Toll Free in Germany: 0800 6397678 e-mail: info@omega.de
United Kingdom:United Kingdom:
United Kingdom:United Kingdom:
United Kingdom: ISO 9002 Certified One Omega Drive River Bend Technology Centre Northbank, Irlam, Manchester M44 5BD, United Kingdom Tel: +44 (0)161 777-6611 Fax: +44 (0)161 777-6622 Toll Free in UK: 0800-488-488 e-mail: info@omega.co.uk
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: WARNING:
WARNING: WARNING:
WARNING: These products are not designed for use in, and should not be used for, human applications.
CN3271-add2.pmd 19/08/2004, 16.362
Page 3
CONTENTS
MOUNTING REQUIREMENTS ............................ 1
DIMENSIONS AND REAR TERMINALS ................ 2
PANEL CUT OUT ........................................... 3
WIRING GUIDELINES ...................................... 3
PRELIMINARY HARDWARE SETTINGS ............... 8
CONFIGURATION PROCEDURE ........................ 9
OPERATING MODE ...................................... 15
Normal display mode .............................. 15
Indicators ............................................. 16
Key functions in normal display mode ........ 16
Operative parameter display mode ............ 16
Key functions in
operative parameter display mode ............ 17
Operating parameters ............................ 17
Limiter function ..................................... 18
Alarm functions ..................................... 19
Serial link ............................................ 19
ERROR MESSAGES ..................................... 20
GENERAL SPECIFICATIONS .......................... 21
MAINTENANCE ............................................ 24
DEFAULT PARAMETERS .............................. A.1
APPENDIX B .............................................. B.1
APPROVALS This instrument is U.L. and c.U.L. approved as controller.
LHL-COPERTINA.pmd 19/04/2004, 15.252
Page 4
1
MOUNTING REQUIREMENTSMOUNTING REQUIREMENTS
MOUNTING REQUIREMENTSMOUNTING REQUIREMENTS
MOUNTING REQUIREMENTS Select a mounting location with the following
characteristics:
1) Minimal vibration.
2) An ambient temperature range between 0 and
50°C (32 and 122 °F).
3) Easy access to the rear of the instrument.
4) No corrosive gases (sulfuric gas, ammonia,
etc.).
5) No water or other fluid (i.e. condensation).
6) Relative humidity of 20% to 80% non
condensing.
The instrument can be mounted on a panel up to 15 mm (0.591 in) thick with a square cutout of 45 x 45 mm (1.772 x 1.772 in). For outline refer to Dimensions and Panel Cutout.
Panel surface texture must be better than 6.3 µm.
The instrument is shipped with a rubber panel gasket (50 to 60 Sh). To insure the IP65 and NEMA 4 protection, insert the panel gasket between the instrument and the panel as shown below.
Install the instrument as follows:
1) Insert the instrument in the gasket.
2) Insert the instrument in the panel cutout.
3) Pushing the instrument against the panel, insert the mounting bracket.
4) Torque the mounting bracket screws between
0.3 and 0.4 Nm (2.66 and 3.54 lbf-in).
5) To insure NEMA 4X/IP65 protection, make sure the instrument does not move within the cutout .
Panel
bracket
Screw
Gasket
Fig. 1
LHL-1-NEUTRO.pmd 19/04/2004, 15.061
Page 5
2
DIMENSIONS AND REAR TERMINAL BLOCKSDIMENSIONS AND REAR TERMINAL BLOCKS
DIMENSIONS AND REAR TERMINAL BLOCKSDIMENSIONS AND REAR TERMINAL BLOCKS
DIMENSIONS AND REAR TERMINAL BLOCKS
With RS-485
Without RS-485
Fig.2
Chromalox
LHL-1-NEUTRO.pmd 19/04/2004, 15.062
Page 6
3
PANEL CUTOUTPANEL CUTOUT
PANEL CUTOUTPANEL CUTOUT
PANEL CUTOUT
WIRING GUIDELINESWIRING GUIDELINES
WIRING GUIDELINESWIRING GUIDELINES
WIRING GUIDELINES A) Measuring InputsA) Measuring Inputs
A) Measuring InputsA) Measuring Inputs
A) Measuring Inputs NOTE:
Any external components (like Zener di­odes, etc.) connected between sensor and input terminals may cause errors in measurement due to excessive and/or not balanced line resistance or possible leakage currents.
TC InputTC Input
TC InputTC Input
TC Input
Fig. 4 THERMOCOUPLE INPUT WIRING NOTE:
1) Do not run input wires with power cables.
2) For TC wiring use proper compensating cable,
preferably shielded (see Appendix B).
3) Shielded cable should be grounded at one end
only.
Fig.3
10
9
+
_
Shield
10
9
+
_
Shield
LHL-1-NEUTRO.pmd 19/04/2004, 15.063
Page 7
4
8
RTD
10
9 8
RTD
10
9
RTD INPUTRTD INPUT
RTD INPUTRTD INPUT
RTD INPUT
Fig. 5 RTD INPUT WIRING NOTENOTE
NOTENOTE
NOTE:
1) Don’t run input wires together with power cables.
2) Pay attention to the line resistance; a high line resistance may cause measurement errors.
3) When shielded cable is used, it should be grounded at one side only to avoid ground loop currents.
4) The resistance of the 3 wires must be the same.
LINEAR INPUTLINEAR INPUT
LINEAR INPUTLINEAR INPUT
LINEAR INPUT
Fig. 6 mA, mV AND V INPUTS WIRING NOTENOTE
NOTENOTE
NOTE:
1) Don’t run input wires together with power cables.
2) Pay attention to the line resistance; a high line resistance may cause measurement errors.
3) When shielded cable is used, it should be grounded at one side only to avoid ground loop currents.
4) The input impedance is equal to: Less than 5 for 20 mAdc input Greater than 1 M for 60 mVdc input Greater than 400 K for 5 Vdc and 10 Vdc input
Shield
_
+
mA,
mV
or
V
9
+
_
G
mA mV
or
V
10
9
10
LHL-1-NEUTRO.pmd 19/04/2004, 15.074
Page 8
5
14
15
Logic input
B) Logic Input (for models with RS-485 only)B) Logic Input (for models with RS-485 only)
B) Logic Input (for models with RS-485 only)B) Logic Input (for models with RS-485 only)
B) Logic Input (for models with RS-485 only) This input is used for remote acknowledgement
(reset).
Safety note:
- Do not run logic input wiring with AC power cables.
- Use an external dry contact capable of switching 0.5 mA, 5 Vdc.
- The instrument needs 100 ms to recognize a contact status variation.
- The logic inputs are
NOT NOT
NOT NOT
NOT isolated from the
measuring input.
Fig.7
6
7
C
NO
OUT 2 Class 1 (Alarm)
1
2
3
NO - OUT 1
C - OUT 1
NC - OUT 1
OUT 1
Class 1
C.1) Relay OutputsC.1) Relay Outputs
C.1) Relay OutputsC.1) Relay Outputs
C.1) Relay Outputs
Fig. 8 The OUT 1 contact rating is 3A/250V AC on
resistive load. The OUT 2 contact rating is 2A/250V AC on resistive load. The number of operations is 1 x 105 at specified rating. NOTESNOTES
NOTESNOTES
NOTES 1) To a void electric shock, connect power
line at the end of the wiring procedure.
2) For power connections use No 16 AWG or larger wires rated for at last 75 °C.
3) Use cupper conductors only.
4) Don’t run input wires with power cables.
All relay contacts are protected by varistor against inductive load with inductive component up to 0.5 A.
The following recommendations avoid serious problems which may occur, when relay outputs are used with inductive loads.
LHL-1-NEUTRO.pmd 19/04/2004, 15.075
Page 9
6
LOAD
(mA)
<40 mA <150 mA <0.5 A
C
(µF)
0.047
0.1
0.33
R
() 100
22 47
P.
(W) 1/2
2 2
OPERATING
VOLTAGE
260 V AC 260 V AC 260 V AC
LOAD
R
C
POWER
LINE
C.2) Inductive LoadsC.2) Inductive Loads
C.2) Inductive LoadsC.2) Inductive Loads
C.2) Inductive Loads High voltage transients may occur switching inductive loads. Through the internal contacts these transients may introduce disturbances which can affect the performance of the instrument. For all the outputs, the internal protection (varistor) assures a correct protection up to 0.5 A of inductive component.
The same problem may occurs when a switch is used in series with the internal contacts as shown in Fig. 9.
Fig. 9 EXTERNAL SWITCH IN SERIES WITH THE INTERNAL CONTACT
In this case it is recommended to install an additional RC network across the external contact as show in Fig. 9 The value of capacitor (C) and resistor (R) are shown in the following table.
The cable involved in relay output wiring must be as far away as possible from input or communica­tion cables.
D) Serial InterfaceD) Serial Interface
D) Serial InterfaceD) Serial Interface
D) Serial Interface For units built with optional RS-485 communication
interface. RS-485 interface allows to connect up to 30 devices with one remote master unit.
Fig. 10 - RS-485 WIRING The cable length must not exceed 1.5 km at 9600 BAUD. NOTESNOTES
NOTESNOTES
NOTES:
1) This RS 485 serial interface is insulated.
2) The following report describes the signal sense of the voltage appearing across the interconnection cable as defined by EIA for RS-485. a) The ” A ” terminal of the generator shall be
negative with respect to the ” B ” terminal for a binary 1 (MARK or OFF) state.
b) The ” A ” terminal of the generator shall be
positive with respect to the ” B ” terminal for a binary 0 (SPACE or ON).
12
13
COMMON
11
B'/B
B/B'
A/A'
A'/A
M
A S T E R
I N S T R U M E N T
LHL-1-NEUTRO.pmd 19/04/2004, 15.076
Page 10
7
I N S T R U M E N T
12
13
COMMON
11
M A S T E R
12
13
11
I N S T R U M E N T
B'/B
B/B'
A/A'
A'/A
E) Power Line and groundingE) Power Line and grounding
E) Power Line and groundingE) Power Line and grounding
E) Power Line and grounding
Fig.12 NOTES:
1) Before connecting the power line, check that the voltage is correct (see Model Number).
2) For supply connections use 16 AWG or larger wires rated for at least 75 °C.
3) Use copper conductors only.
4) Do not run input wires with power cables.
5) Polarity does not matter for 24 Vdc wiring.
6) The power supply input is
NOT NOT
NOT NOT
NOT fuse protected.
Please provide it externally.
Power supply Type Current Voltage
24 V AC/DC T 500 mA 250 V
100/240 V AC T 125 mA 250 V When fuse is damaged, it is advisable to verify the power supply circuit, so that it is necessary to send back the instrument to your supplier.
7) Safety requirements for permanently connected equipment:
- Include a switch or circuit-breaker in the
installation.
- Place the switch in close proximity to the
equipment and within easy reach of the operator.
- Mark the switch as the disconnecting
device for the equipment.
NOTE:
A single switch or circuit-breaker can
drive more than one instrument.
8) When the NEUTRAL line is present, connect it to terminal 4.
9) To avoid shock and possible instrument damage, connect power last.
4
5
POWER SUPPLY
100 to 240 Vac
24 Vac/Vdc
N, L2
R (S,T), L1
N, L2
R (S,T), L1
Fig.11
LHL-1-NEUTRO.pmd 19/04/2004, 15.077
Page 11
8
PRELIMINARY HARDWAREPRELIMINARY HARDWARE
PRELIMINARY HARDWAREPRELIMINARY HARDWARE
PRELIMINARY HARDWARE SETTINGSSETTINGS
SETTINGSSETTINGS
SETTINGS
1) Remove the instrument from its case.
2) Set J106 according to the desired input type as shown in the following figure.
INPUT J106 TYPE 1-2 3-4 5-6 7-8 TC-RTD close open open open 60 mV close open open open 5 V open close open open 10 V open open close open 20 mA open open open close
High - Low Limiter with RS-485High - Low Limiter with RS-485
High - Low Limiter with RS-485High - Low Limiter with RS-485
High - Low Limiter with RS-485
Fig.13.A
7531
8642
J106
V101
High - Low Limiter without RS-485High - Low Limiter without RS-485
High - Low Limiter without RS-485High - Low Limiter without RS-485
High - Low Limiter without RS-485
7531
8642
J106
V101
Fig.13.B
LHL-1-NEUTRO.pmd 19/04/2004, 15.078
Page 12
9
CONFIGURATION PROCEDURECONFIGURATION PROCEDURE
CONFIGURATION PROCEDURECONFIGURATION PROCEDURE
CONFIGURATION PROCEDURE
CONFIGURATION KEY FUNCTIONSCONFIGURATION KEY FUNCTIONS
CONFIGURATION KEY FUNCTIONSCONFIGURATION KEY FUNCTIONS
CONFIGURATION KEY FUNCTIONS RESET In Configuration Mode, it is used only to scroll
back parameters without to memorize a new parameter value. Used in Configuration Mode to decrease the parameter value. Used in Configuration Mode to increase the parameter value.
FUNC Used to memorize the new parameter
value and go to the next parameter. + Loads the default parameters. + FUNC or + FUNC
Increases/decreases values at a higher rate
when modifying parameters. + RESET or + RESET
Jumps to the Maximum or Minimum param-
eter value when modifying parameters.
CONFIGURATION PROCEDURECONFIGURATION PROCEDURE
CONFIGURATION PROCEDURECONFIGURATION PROCEDURE
CONFIGURATION PROCEDURE
1) Remove the instrument from its case.
2) Open switch V101 (See illustrations under “Preliminary Hardware Settings.”)
3) Re-insert the instrument in its case.
4) Switch on power to the instrument. The upper display will show COnF.
NOTE NOTE
NOTE NOTE
NOTE : If "CAL" indication is displayed, press immediately the V pushbutton and return to the configuration procedure.
5) Press the “
” key and the lower display will
show the firmware version.
Press the "FUNC" key to start the configuration procedure with the first parameter (L1).
The following is a complete list of parameters. The lower display will show the parameter code (L1 to d1) and the upper display will show the selection code or numerical value. No timeout is applied in the configuration mode.
L1L1
L1L1
L1
==
==
=
Serial Interface ProtocolSerial Interface Protocol
Serial Interface ProtocolSerial Interface Protocol
Serial Interface Protocol (Skipped if option is not available.) OFF = No serial interface nbUS = Modbus jbUS = Jbus
L2L2
L2L2
L2
==
==
=
Serial Link Device AddressSerial Link Device Address
Serial Link Device AddressSerial Link Device Address
Serial Link Device Address (Skipped if option is not available or L1 = OFF) From 1 to 255 NOTE:
EIA standard allows no more than 31 device
connected by one RS-485.
L3L3
L3L3
L3
==
==
=
Baud Rate for Serial LinkBaud Rate for Serial Link
Baud Rate for Serial LinkBaud Rate for Serial Link
Baud Rate for Serial Link (Skipped if option is not available or L1 = OFF) Set value from 600 to 19200 baud. (19200 baud is shown on display as 1920)
L4L4
L4L4
L4
==
==
=
Byte Format for Serial LinkByte Format for Serial Link
Byte Format for Serial LinkByte Format for Serial Link
Byte Format for Serial Link (Skipped if option is not available or L1 = OFF) 8E = 8 bits + even parity 8O = 8 bits + odd parity 8 = 8 bits without parity
LHL-1-NEUTRO.pmd 19/04/2004, 15.079
Page 13
10
r1r1
r1r1
r1
=Input Type and Range Value=Input Type and Range Value
=Input Type and Range Value=Input Type and Range Value
=Input Type and Range Value 0 = TC J From -100 to 1000 °C 1 = TC K From -100 to 1370 °C 2 = TC T From -200 to 400 °C 3 = TC E From -100 to 800 °C 4 = TC N From -100 to 1400 °C 5 = TC S From -50 to 1760 °C 6 = TC R From -50 to 1760 °C 7 = TC B From 0 to 1820 °C 8 = TC L From -100 to 900 °C 9 = TC U From -200 to 600 °C 10 = TC G From 0 to 2300 °C 11 = TC D From 0 to 2300 °C 12 = TC C From 0 to 2300 °C 13 = TC Pla t . I I From -100 to 1400 °C 14= RTD Pt 100 From -200 to 850 °C 15= Linear From 0 to 60 mV 16= Linear From 12 to 60 mV 17= Linear From 0 to 20 mA 18= Linear From 4 to 20 mA 19= Linear From 0 to 5 V 20= Linear From 1 to 5 V 21= Linear From 0 to 10 V 22= Linear From 2 to 10 V 23 = TC J From -150 to 1830 °F 24 = TC K From -150 to 2500 °F 25 = TC T From -330 to 750 °F 26 = TC E From -150 to 1470 °F 27 = TC N From -150 to 2550 °F 28 = TC S From -60 to 3200 °F 29 = TC R From -60 to 3200 °F 30 = TC B From 32 to 3300 °F 31 = TC L From -150 to 1650 °F 32 = TC U From -330 to 1110 °F 33 = TC G From 0 to 4170 °F 34 = TC D From 0 to 4170 °F 35 = TC C From 0 to 4170 °F 36 = TC Pla t . I I From -150 to 2550 °F 37= RTD Pt100 From -330 to 1560 °F
r2 = Decimal Point Positionr2 = Decimal Point Position
r2 = Decimal Point Positionr2 = Decimal Point Position
r2 = Decimal Point Position (Available only for linear range r1 = 15 to 22)
_ _ _ _ ._ _ _ _ .
_ _ _ _ ._ _ _ _ .
_ _ _ _ . = No decimal _ _ _ . __ _ _ . _
_ _ _ . __ _ _ . _
_ _ _ . _ = One decimal figure _ _ . _ __ _ . _ _
_ _ . _ __ _ . _ _
_ _ . _ _ = Two decimal figures _ . _ _ __ . _ _ _
_ . _ _ __ . _ _ _
_ . _ _ _ = Three decimal figures
r3 = Initial value of the readout scaler3 = Initial value of the readout scale
r3 = Initial value of the readout scaler3 = Initial value of the readout scale
r3 = Initial value of the readout scale (Available only for linear range r1 = 15 to 22) Range: From -1999 to 9999
r4 = final value of the readout scaler4 = final value of the readout scale
r4 = final value of the readout scaler4 = final value of the readout scale
r4 = final value of the readout scale (Available only for linear range r1 = 15 to 22) Range: From -1999 to 9999
r5 = Offset Adjustmentr5 = Offset Adjustment
r5 = Offset Adjustmentr5 = Offset Adjustment
r5 = Offset Adjustment Range: From -500 to 500 Offset value algebraically added to the measured value.
r6 =r6 =
r6 =r6 =
r6 =
Time constant of the filter applied toTime constant of the filter applied to
Time constant of the filter applied toTime constant of the filter applied to
Time constant of the filter applied to the displayed valuethe displayed value
the displayed valuethe displayed value
the displayed value Range: From 0 (filter OFF) to 8 seconds. (First order filter with selected time constant.)
r7 = Alarm action on input fault.r7 = Alarm action on input fault.
r7 = Alarm action on input fault.r7 = Alarm action on input fault.
r7 = Alarm action on input fault. When the instrument detects an input failure condition, the alarm will operate as in presence of: uP = as in presence of the full scale value. doun = as in presence of the initial scale value.
C1 = Type of limit actionC1 = Type of limit action
C1 = Type of limit actionC1 = Type of limit action
C1 = Type of limit action
LHL-1-NEUTRO.pmd 19/04/2004, 15.0710
Page 14
11
Hi. = High limit (for heating process) Lo. = Low limit (for cooling process) Hi.Lo = High and low limit (for special process)
C2 = Rearming ModeC2 = Rearming Mode
C2 = Rearming ModeC2 = Rearming Mode
C2 = Rearming Mode O = Acknowledgements rearm (reset) the limiter (and
restart the process) only if the condition which generated the shutdown status no longer exists (points A and C of the Example 1). It do
notnot
notnot
not generate any effect if the condition which generated the shutdown status still exists (point B of the Example 1).
I = Acknowledgements enable the automatic
rearmament (reset) of the limiter if the condition which generated the shutdown status still exists (point B of the Example 2). (The instrument rearms (reset) automatically when the condition which generated the shutdown status no longer exists).
Notes about limiter functionNotes about limiter function
Notes about limiter functionNotes about limiter function
Notes about limiter function The relay of the output 1 operates in fail-safe mode (relay de-energized during shutdown condition) and latching mode.
The OUT 1 turns OFF when:
- C1 = Hi and the measured value is greater than limiter threshold [“Su” parameter (see Operative parameters”)] or
- C1 = LO and the measured value is less than limiter threshold [“Su” parameter (see Operative parameters”)] or
- C1 = HiLO and the measured value is greater than “Su” parameter (see Operative parameters”) or less than “S1” parameter (see Operative parameters”).
The Out 1 remains OFF until the condition which
generated the shutdown, no longer exists and the acknowledge action has been performed.
The upper display flashes during a shutdown and returns to a steady display when the shutdown condition no longer exists. When C2 = 0 and OUT 1 is OFF, the RESET LED is ON. When C2 = 1 one of the following condition may occur:
- if no acknowledgement has been made, OUT 1 is OFF and the RESET LED is flashing;
- if the acknowledgement has been made but the condition which generated the shutdown status still exists, OUT 1 is OFF and the RESET LED is steady ON.
The shutdown condition can be stored in permanent memory (see C4). Acknowledgment can be performed by pressing the RESET key, by momentarily closing the external dry contact or by a command from the serial link.
The length of the shutdown condition and max/min measured values are stored in memory and available for viewing until the next shutdown condition occurs. These informations are lost at power down. During a shutdown condition the max/min measured values are continuously updated and can be monitored. Example 1 - C1 = Hi and C2 = O
LHL-1-NEUTRO.pmd 19/04/2004, 15.0711
Page 15
12
C4C4
C4C4
C4
==
==
=
Shutdown memoryShutdown memory
Shutdown memoryShutdown memory
Shutdown memory
0 = The shutdown condition will be saved (at
next power up it will be reactivated)
1 = The shutdown condition will be lost in
case of power down
C5 =C5 =
C5 =C5 =
C5 =
Time Constant of the Filter appliedTime Constant of the Filter applied
Time Constant of the Filter appliedTime Constant of the Filter applied
Time Constant of the Filter applied to the Measured Value for Limitto the Measured Value for Limit
to the Measured Value for Limitto the Measured Value for Limit
to the Measured Value for Limit Action.Action.
Action.Action.
Action. Range: From 0 (filter OFF) to 8 seconds Note: First order filter with selected time
Constant.
P1P1
P1P1
P1
= Alarm Function= Alarm Function
= Alarm Function= Alarm Function
= Alarm Function (Skipped when the option is not available) nonE = Not provided AL.P = Process alarm AL.b = Band alarm AL.d = Deviation alarm When C1 = Hi.Lo, “AL.b” and “AL.d” are not available.
P2P2
P2P2
P2
==
==
=
Alarm configurationAlarm configuration
Alarm configurationAlarm configuration
Alarm configuration (Skipped if option is not available or P1 = none) H.A. = High alarm with automatic reset L.A. = Low alarm with automatic reset H.A.Ac = High alarm with automatic reset and
"Silence" function. L.A.Ac = Low alarm with automatic reset and
"Silence" function. H.L. = High alarm with manual reset L.L. = Low alarm with manual reset
HS
(Limiter Hysteresis)
ON
Relay OUT 1
OFF
ON OFF
“Reset” LED
Flash
Steady
Upper display
FLASH
FLASH
Su (Limiter threshold)
HS
(Limiter Hysteresis)
ON
Relay OUT 1
OFF ON
OFF
“Reset” LED
Flash
Steady
Upper display
FLASH
FLASH
FLASH
FLASH
Su (Limiter threshold)
A, B, C = Acknowledgment actions. NOTE: Acknowledgment B has no effect.
Example 2 - C1 = Hi and C2 = 1
A, B = Acknowledgment actions.
C3C3
C3C3
C3
==
==
=
Rearm at Power-upRearm at Power-up
Rearm at Power-upRearm at Power-up
Rearm at Power-up Auto = Automatic rearm nAn = Manual rearm
LHL-1-NEUTRO.pmd 19/04/2004, 15.0712
Page 16
13
NOTE:
1) For band alarm, H.A./H.A.Ac/H.L. signifies
outside band alarm, while L.A./ L.A.Ac/L.L. signifies inside band alarm.
2) The "Silence" function allows the manual reset
of the alarm even if the alarm condition is still in progress.
Example Example
Example Example
Example for P2 = H.A.
* Alarm Status: Relay energized (P3 = dir)
Relay de-energized (P3 =rEV)
Example Example
Example Example
Example for P2 = H.A.A.c
* Alarm Status: Relay energized (P3 = dir)
Relay de-energized (P3 =rEV)
Example Example
Example Example
Example for P2 = H.L.
* Alarm Status: Relay energized (P3 = dir)
Relay de-energized (P3 =rEV)
P3 = Alarm ActionP3 = Alarm Action
P3 = Alarm ActionP3 = Alarm Action
P3 = Alarm Action (Skipped if option not is available or P1 = none) dir = Direct action
(Relay energized in alarm condition)
rEV = Reverse action
(Relay energized in non-alarm condition)
(Alarm Hysteresis)
Relay
Non alarm status
ON
OFF
FLASH
FLASH
(Alarm threshold)
Manual reset
Alarm status*
ALM LED
Manual reset
(Alarm Hysteresis)
Alarm status*
Relay
Non alarm status
ON
OFF
ALM LED
FLASH
FLASH
(Alarm threshold)
Manual reset
(Alarm Hysteresis)
Alarm status*
Relay
Non alarm status
ON
OFF
FLASH
FLASH
(Alarm threshold)
Manual reset
ALM LED
LHL-1-NEUTRO.pmd 19/04/2004, 15.0713
Page 17
14
P4P4
P4P4
P4
==
==
=
Alarm Standby (mask) FunctionAlarm Standby (mask) Function
Alarm Standby (mask) FunctionAlarm Standby (mask) Function
Alarm Standby (mask) Function (Skipped if option is not available or P1= none) OFF = Standby function disabled On = Standby function enabled
If the alarm is programmed as band or deviation, this function masks the alarm condition at start up and after a “Su” (limit threshold) changement until the process variable reaches the alarm threshold, plus or minus hysteresis. This standby function masks a Process Alarm condition at start up until the process variable reaches the alarm threshold plus or minus hysteresis.
PF =PF =
PF =PF =
PF =
Time Constant of the FilterTime Constant of the Filter
Time Constant of the FilterTime Constant of the Filter
Time Constant of the Filter
applied to the Measured Value forapplied to the Measured Value for
applied to the Measured Value forapplied to the Measured Value for
applied to the Measured Value for Alarm ActionAlarm Action
Alarm ActionAlarm Action
Alarm Action (Skipped if option is not available or P1 = none) Range: From 0 (filter OFF) to 8 seconds (First order filter with selected time constant.)
n 1 = Safety Lockn 1 = Safety Lock
n 1 = Safety Lockn 1 = Safety Lock
n 1 = Safety Lock 0 = UNLOCKed. The device is always
UNLOCKed and all parameters can be modified.
l = LOCKed. The device is always LOCKed
and no parameters can be modified
From 2 to 9999 = This number is a password, to be
used in run time (see “nn”), to LOCK/ UNLOCK the device.
t1 = Timeout Selectiont1 = Timeout Selection
t1 = Timeout Selectiont1 = Timeout Selection
t1 = Timeout Selection tn10 = 10 second timeout tn30 = 30 second timeout
d1 = Digital Input (contact closure)d1 = Digital Input (contact closure)
d1 = Digital Input (contact closure)d1 = Digital Input (contact closure)
d1 = Digital Input (contact closure) (This is a read only parameter) Enb = Digital input enabled dlS = Digital input disabled (The digital input is used as a remote Acknowledgment .)
The configuration procedure is now completed. The display will show "COnF".
LHL-1-NEUTRO.pmd 19/04/2004, 15.0714
Page 18
15
OPERATING MODEOPERATING MODE
OPERATING MODEOPERATING MODE
OPERATING MODE
1) Remove the instrument from its case.
2) Set switch V101 (see fig. 13) to the closed
position.
3) Re-insert the instrument in its case.
4) Switch on the instrument. Normal Display ModeNormal Display Mode
Normal Display ModeNormal Display Mode
Normal Display Mode On powerup the device starts in the "Normal Display Mode."
By pressing the or key, it is possible to change the displayed information; therefore, one of the following display modes can be selected:
1) The upper display shows the measured value
while the lower display shows the "Pu" (Process variable). If this display was active at power down, it will be active at powerup.
2) The upper display shows the limiter threshold
while the lower display shows "Su." If this display was active at power down, it will be active at powerup.
3) The upper display shows the second limiter
threshold while the lower display shows "S1." This information is available only if C1 = Hi.Lo. If this display was active at power down, it will be active at powerup.
4) The upper display shows the total time (hh.mm)
of the last shutdown condition while the lower displays shows “t.” If no shutdown condition was detected, the upper display will show “- - -
-”. The information is lost at power down and at powerup the device will display the process variable.
5) The upper display shows the maximum
measured value detected during the last shutdown condition while the lower display
shows "Ph.". If no shutdown condition was detected, the upper display will show “- - - -”. This information is not available if C1 = HI.Lo. The information is lost at power down and at powerup the device will display the process variable.
NOTE: When the shutdown condition was generated by an input fault condition, the upper display will indicate “m.Err”
6) The upper display shows the minimum mesured value detected during the last shutdown condition while the lower display shows "PL.” If no shutdown condition was detected, the upper display will show “- - - -”. This information is not available if C1 = Hi. The information is lost at power down and at powerup the device will display the process variable.
NOTE: When the shutdown condition was generated by an input fault condition, the upper display will indicate “m.Err”
If, at power off, the device was in shutdown condition and shutdown memoy function is selected (C4 = 0), and/or it was programmed for manual reset at startup (C3 = 1), then at the next power up the lower display will be flashing.
LHL-1-NEUTRO.pmd 19/04/2004, 15.0715
Page 19
16
IndicatorsIndicators
IndicatorsIndicators
Indicators “RESET“ =Indicates control output 1 status as
follows: a) When C2 parameter has been
configured equal to 0, LED ON when Output 1 is OFF LED OFF when Output 1 is ON
b) When C2 parameter has been
configured equal to 1, LED flashes when Output 1 is OFF LED ON when Output 1 is OFF and acknowledged LED OFF when Output is ON
“ALM” = Indicates alarm status as follows:
- Flashes when alarm is ON
- ON when alarm has been resetted but the alarm condition is still present.
- OFF when alarm is OFF
“REM” = Indica t es t h e r e mo t e s t a tu s of t h e
instrument.
- Flashes when instrument is in remote mode.
- OFF when instrument is in local mode.
Key Functions in Normal Display ModeKey Functions in Normal Display Mode
Key Functions in Normal Display ModeKey Functions in Normal Display Mode
Key Functions in Normal Display Mode “FUNC” = By pressing it, the display changes from
“Normal Display Mode” to “Operative Parameter Display Mode.”
= Pressing it for more than ten seconds
initiates the Lamp Test. During the Lamp Test the device function normally while all display segments and LED's are lit with a 50% duty cycle. No timeout is applied to a lamp test. Press the "FUNC" key again to end the Lamp Test.
" " or “ " = By pressing these keys it is possible to
change the displayed information. See “Normal Display Mode” on previous page.
“RESET” = Press and hold for 1 second to rearm
(reset) the limiter.
+ FUNC or + FUNC
Increases/decreases values at a higher rate when modifying parameters.
+ RESET or + RESET
Jumps to the Maximum or Minimum parameter value when modifying parameters.
Operative Parameter Display ModeOperative Parameter Display Mode
Operative Parameter Display ModeOperative Parameter Display Mode
Operative Parameter Display Mode The "FUNC" key initiates the Operative Parameter Display Mode when pressed for less than 10 seconds in the "Normal Display Mode."
The lower display shows the parameter code while the upper display shows the parameter value or status. The value of the selected parameter can be modified with the and keys.
Press the "FUNC" key again to store the new value and advance to the next parameter.
If no keys are pressed within the timeout period (see t1), the instrument will automatically return to the "Normal Display Mode" in the previous display and any modification of the last displayed parameter will be lost.
All parameters (except a ) can be modified only when the device is UNLOCKed.
The LOCK/UNLOCK status can be selected in configuration using “n1” parameter or during the operating mode with the “nn” parameter (pass­word).
6
LHL-1-NEUTRO.pmd 19/04/2004, 15.0716
Page 20
17
nn
Software KeySoftware Key
Software KeySoftware Key
Software Key (Skipped if n1 = 0 or 1) ON = the device is LOCKED. OFF = the device is UNLOCKED. When it is desired to switch from LOCK to UNLOCK condition, set a value equal to “n1” parameter. When it is desired to switch from UNLOCK to LOCK condition, set a value different from “n1” parameter.
Su
Limiter ThresholdLimiter Threshold
Limiter ThresholdLimiter Threshold
Limiter Threshold Range: Span limits (From "S1" to full scale value when C1 = Hi.Lo)
S1
Second Limiter ThresholdSecond Limiter Threshold
Second Limiter ThresholdSecond Limiter Threshold
Second Limiter Threshold (Available when C1 = Hi.Lo) Range: From initial scale value to "Su"
HS
Limiter HysteresisLimiter Hysteresis
Limiter HysteresisLimiter Hysteresis
Limiter Hysteresis Range: From 0.1% to 10.0% of the input span or 1 LSD
AL
Alarm Threshold (optional)Alarm Threshold (optional)
Alarm Threshold (optional)Alarm Threshold (optional)
Alarm Threshold (optional) (Available only if the option is fitted and P1= AL.P, AL.b or AL.d.) Ranges: span limits for process alarm (P1 = AL.P) from 0 to 500 for band alarm (P1 = AL.b) from -500 to 500 for deviation alarm (P1 = AL.d)
HA
Alarm Hysteresis (optional)Alarm Hysteresis (optional)
Alarm Hysteresis (optional)Alarm Hysteresis (optional)
Alarm Hysteresis (optional) (Available only if the option is fitted and P1 = AL.P, AL.b or AL.d) Range: From 0.1% to 10.0% of the input span or 1 LSD.
To switch from LOCKED to UNLOCKED, assign to the “nn” parameter a value equal to the “n1” parameter setting. To switch from UNLOCKED to LOCKED, assign to the “nn” parameter any number other than the n1 parameter setting. When the device is in remote mode (the serial link controls the device) no parameters can be modified.
Key Functions inKey Functions in
Key Functions inKey Functions in
Key Functions in Operative Parameter Display ModeOperative Parameter Display Mode
Operative Parameter Display ModeOperative Parameter Display Mode
Operative Parameter Display Mode
FUNC = Pressing the “FUNC” key, the
instrument stores the new setting (if changed) and goes to the next parameter.
or = Changes the setting of the
selected parameter.
RESET = Press and hold for more than 1
second for limiter rearmament.
OPERATING PARAMETERSOPERATING PARAMETERS
OPERATING PARAMETERSOPERATING PARAMETERS
OPERATING PARAMETERS Some of the following parameters may not appear, depending on the configuration.
Lower Description Display
a
Manual reset of the alarm.Manual reset of the alarm.
Manual reset of the alarm.Manual reset of the alarm.
Manual reset of the alarm. (Available only if P1 = AL.p, AL.b or AL.d) ON = Starts the manual reset of the alarm OFF = Do not start the alarm reset. Select ON and press the FUNC key in order to reset the alarm. After a manual reset of the alarm the instrument returns in Normal Display Mode.
6
LHL-1-NEUTRO.pmd 19/04/2004, 15.0717
Page 21
18
Limiter functionLimiter function
Limiter functionLimiter function
Limiter function The relay of the output 1 operates in fail-safe mode (relay de-energized during shutdown condition) and latching mode.
The OUT 1 turns OFF when:
- The instrument is configured as a high limiter (C1 = Hi) and the measured value is greater than limiter threshold [“Su” parameter (see Operative parameters”)] or
- The instrument is configured as a low limiter (C1 = LO) and the measured value is less than limiter threshold [“Su” parameter (see Operative parameters”)] or
- The instrument is configured as a high/low limiter (C1 = HiLO) and the measured value is greater than “Su” parameter (see Operative parameters”) or less than “S1” parameter (see Operative parameters”).
The Out 1 remains OFF until the condition which generated the shutdown, no longer exists and the acknowledge action has been performed.
The upper display flashes during a shutdown and returns to a steady display when the shutdown condition no longer exists.
When the OUT 1 is OFF the RESET LED is ON [if the selected rearming mode is equal to 0 (C2 = 0)] or flashes [if the selected rearming mode is equal to 1 (C2 = 1)]. When the selected rearming mode is equal to 1 (C2 = 1) the RESET LED is steady ON when OUT 1 is OFF and acknowledged. The shutdown condition can be stored in permanent memory (see C4). Acknowledgment can be performed by pressing the RESET key, by momentarily closing the external dry contact or by a command from the serial link.
The length of the shutdown condition and max/min measured values are stored in memory and available for viewing (see “Normal Display Mode”) until the next shutdown condition occurs. These informations are lost at power down.
During a shutdown condition the max/min measured values are continuously updated and can be monitored.
Example 1 - C1 = Hi and C2 = O
Example 2 - C1 = Hi and C2 = 1
LHL-1-NEUTRO.pmd 19/04/2004, 15.0718
Page 22
19
Alarm functionsAlarm functions
Alarm functionsAlarm functions
Alarm functions (Skipped if option is not available or P1 = none) The alarm can be programmed as:
- process alarm
- band alarm
- deviation alarm.
Band and deviation alarms are referred to the limiter threshold and are possible only if an high limiter or a low limiter function has been selected. For all the alarm types, it is possible to select automatic or manual reset or the “Silence” function. The "Silence" function is a typical function of the alarm annunciators (see ISA “Alarm annunciator operational sequence”) and it is usually applied to audible alarm indications (horn). This function allows the manual reset of the alarm even if the alarm condition is still in progress.
It is also possible to assign to the alarm a stand by (mask) function. If the alarm is programmed as band or deviation alarm, this function masks the alarm condition after a safety threshold change or at the instrument start-up until process variable reaches the alarm threshold plus or minus hysteresis. If the alarm is programmed as a process alarm, this function masks the alarm condition at instrument start-up until process variable reaches the alarm threshold plus or minus hysteresis. Graphic example of the alarm behaviour are shown at pages 12 and 13.
Serial LInk (optional)Serial LInk (optional)
Serial LInk (optional)Serial LInk (optional)
Serial LInk (optional) The device can be connected to a host computer via serial link.
The host can put the device in LOCAL (parameters are controlled via keyboard) or in REMOTE (functions and parameters are controlled via serial link). REMOTE is shown by the decimal point to the left of "REM" which is on the right side of the numerical display.
Via serial link it is possible to read and/or to modify all the operative and configuration parameters. The following conditions must apply to implement this function:
1) Configure parameters L1 through L4 with the front keyboard.
2) The device must be in the Operating mode.
For other details require ENG 816-E document.
LHL-1-NEUTRO.pmd 19/04/2004, 15.0719
Page 23
20
ERROR MESSAGESERROR MESSAGES
ERROR MESSAGESERROR MESSAGES
ERROR MESSAGES Overrange, Underrange andOverrange, Underrange and
Overrange, Underrange andOverrange, Underrange and
Overrange, Underrange and Sensor Break IndicationsSensor Break Indications
Sensor Break IndicationsSensor Break Indications
Sensor Break Indications This device detects input fault conditions. (OVERRANGE, UNDERRANGE OR SENSOR BREAK). When the process variable exceeds the span limits an OVERRANGE condition will appear as:
An UNDERRANGE condition will appear as:
A sensor break is signalled as "OPEn”. On the mA/ V input, a sensor break can be detected only when the range selected has a zero elevation (4/ 20 mA, 12/60 mV, 1/5 V or 2/10 V.)
On the RTD input "shrt" is signalled when input resistance is less than 15 (short circuit sensor detection).
This device detects reference junction errors or errors on the internal autozero measurement. When a fault is detected the output goes OFF and the alarm assumes an upscale/downscale reading in accordance with r7.
Error MessagesError Messages
Error MessagesError Messages
Error Messages On power up, the instrument performs a self­diagnostic test. When an error is detected, the lower display shows an "Er" indication while the upper display shows the code of the detected error.
Error ListError List
Error ListError List
Error List 100 Error in EEPROM writing 150 Short circuit on CPU's outputs 200 Error on "protect register" in EEPROM XXX Configuration parameter error. 301 Error on calibration of selected input. 307 rj input calibration error. 400 Error on operative parameters. 500 Error on autozero measurement. 502 Error on reference junction measurement. 510 Error during calibration procedure.
Dealing with Error MessagesDealing with Error Messages
Dealing with Error MessagesDealing with Error Messages
Dealing with Error Messages
1) When a configuration parameter error is detected, repeat the configuration procedure of that specific parameter.
2) If an error 400 is detected, press and hold the
key and press the key and load the default parameters; then repeat the control parameter setup.
3) For all other errors, contact your Service Representative.
LHL-1-NEUTRO.pmd 19/04/2004, 15.0720
Page 24
21
GENERAL SPECIFICATIONSGENERAL SPECIFICATIONS
GENERAL SPECIFICATIONSGENERAL SPECIFICATIONS
GENERAL SPECIFICATIONS CaseCase
CaseCase
Case: Polycarbonate grey case Self extinguishing degreeSelf extinguishing degree
Self extinguishing degreeSelf extinguishing degree
Self extinguishing degree: V-0 according to UL94. Front protectionFront protection
Front protectionFront protection
Front protection - designed and tested for IP 65 (*) and NEMA 4X (*) for indoor locations (when panel gasket is installed). (*) Test were performed in accordance with IEC 529, CEI 70-1 and NEMA 250-1991 STD. InstallationInstallation
InstallationInstallation
Installation: panel mounting. Rear terminal boardRear terminal board
Rear terminal boardRear terminal board
Rear terminal board: 15 screw terminals (screw M3, for cables from φ 0.25 to φ 2.5 mm
2
or from AWG 22 to AWG 14 ), connection diagram and safety rear cover. DimensionsDimensions
DimensionsDimensions
Dimensions: 48 x 48 mm (according to DIN
43700); depth
- 122 mm for models with RS-485.
- 105 mm for models without RS-485 WeightWeight
WeightWeight
Weight: 250 g. max. (8.75 oz.). Power supply Power supply
Power supply Power supply
Power supply : (switching mode) from 100 to 240 V AC. 50/60 Hz (+10 % to -15 % of the nominal value) or 24 V DC/AC (
+10 % of the nominal value).
Power consumptionPower consumption
Power consumptionPower consumption
Power consumption: 8 VA. Insulation resistanceInsulation resistance
Insulation resistanceInsulation resistance
Insulation resistance: > 100 M according to IEC 1010-1. Isolation voltageIsolation voltage
Isolation voltageIsolation voltage
Isolation voltage: 1500 V r.m.s. according to IEC 1010-1. Common mode rejection ratioCommon mode rejection ratio
Common mode rejection ratioCommon mode rejection ratio
Common mode rejection ratio: 120 dB @ 50/60 Hz. Normal mode rejection ratioNormal mode rejection ratio
Normal mode rejection ratioNormal mode rejection ratio
Normal mode rejection ratio: 60 dB @ 50/60 Hz. Electromagnetic compatibility and safetyElectromagnetic compatibility and safety
Electromagnetic compatibility and safetyElectromagnetic compatibility and safety
Electromagnetic compatibility and safety requirementsrequirements
requirementsrequirements
requirements: This instrument is marked CE. Therefore, it is conforming to council directives 89/336/EEC (reference harmonized standard EN 50081-2 and EN 50082-2) and to council directives 73/23/EEC and 93/68/EEC (reference harmonized standard EN 61010-1). Installation categoryInstallation category
Installation categoryInstallation category
Installation category: II
D/A conversionD/A conversion
D/A conversionD/A conversion
D/A conversion: dual slope integration. Sampling timeSampling time
Sampling timeSampling time
Sampling time :
- for linear inputs = 250 ms.
- for TC or RTD inputs = 500 ms. Display updating timeDisplay updating time
Display updating timeDisplay updating time
Display updating time: 500 ms. ResolutionResolution
ResolutionResolution
Resolution: 30000 counts. Temperature Drift Temperature Drift
Temperature Drift Temperature Drift
Temperature Drift (CJ excluded)
- Less than 200 ppm/°C of full span for mV and TC ranges 0, 1, 3, 4, 8, 13, 23, 24, 26, 27, 31, 36 (CJ excluded).
- Less than 300 ppm/°C of full span for mA, V and TC ranges 10, 11, 12, 33, 34, 35 (CJ excluded)
- Less than 400 ppm/°C of full span for RTD and TC range 9, 32 (CJ excluded).
- Less than 500 ppm/°C of full span for TC ranges 2, 5, 6, 25, 28, 29 (CJ excluded).
- Less than 600 ppm/°C of full span for TC ranges 7, 30. NOTE: Precision and drift guaranteed (for T>300°C/570°F).
AccuracyAccuracy
AccuracyAccuracy
Accuracy:
+ 0.2% f.s.v. @ 25 °C (77 °F) and nominal power supply voltage. Operative temperatureOperative temperature
Operative temperatureOperative temperature
Operative temperature: from 0 to +50 °C (32 to 122 °F). Storage temperatureStorage temperature
Storage temperatureStorage temperature
Storage temperature: from -20 to +70 °C (-4 to 158 °F). HumidityHumidity
HumidityHumidity
Humidity: from 20% to 85 % RH not condensing.
LHL-1-NEUTRO.pmd 19/04/2004, 15.0721
Page 25
22
INPUTSINPUTS
INPUTSINPUTS
INPUTS A) A)
A) A)
A)
THERMOCOUPLETHERMOCOUPLE
THERMOCOUPLETHERMOCOUPLE
THERMOCOUPLE
Type :Type :
Type :Type :
Type : J, K, T, E, N, S, R, B, L, U, G(W), D(W3), C(W5), Platinel II, °C/°F selectable. External resistanceExternal resistance
External resistanceExternal resistance
External resistance: 100 max, maximum error 0,1% of span. Burn outBurn out
Burn outBurn out
Burn out: It is shown as an overrange condition (standard). It is possible to obtain an underrange indication by cut and short. Cold junctionCold junction
Cold junctionCold junction
Cold junction: automatic compensation from 0 to 50 °C. Cold junction accuracy Cold junction accuracy
Cold junction accuracy Cold junction accuracy
Cold junction accuracy : 0.1 °C/°C Input impedanceInput impedance
Input impedanceInput impedance
Input impedance: > 1 M CalibrationCalibration
CalibrationCalibration
Calibration : according to IEC 584-1 and DIN 43710 - 1977. STANDARD RANGES TABLE
(*) P. equal to Platinel II
B) RTDB) RTD
B) RTDB) RTD
B) RTD (
RR
RR
Resistance
TT
TT
Temperature
DD
DD
Detector)
InputInput
InputInput
Input: for RTD Pt 100 , 3 wire connection. Input circuitInput circuit
Input circuitInput circuit
Input circuit: current injection. °C/°F selection°C/°F selection
°C/°F selection°C/°F selection
°C/°F selection: via front pushbuttons or serial link. Line resistanceLine resistance
Line resistanceLine resistance
Line resistance: automatic compensation up to 20 /wire with no measurable error. CalibrationCalibration
CalibrationCalibration
Calibration: according to DIN 43760 Burn out Burn out
Burn out Burn out
Burn out : The instrument detect the open condition of one or more wires. It is able to detect also the short circuit of the sensor.
STANDARD RANGES TABLE
C) LINEAR INPUTSC) LINEAR INPUTS
C) LINEAR INPUTSC) LINEAR INPUTS
C) LINEAR INPUTS Read-outRead-out
Read-outRead-out
Read-out: keyboard programmable between
-1999 and +9999. Decimal pointDecimal point
Decimal pointDecimal point
Decimal point: programmable in any position Burn outBurn out
Burn outBurn out
Burn out: the instrument shows the burn out condition as an underrange condition for 4-20 mA, 1-5 V and 2-10 V input types. It shows the burn out condition as an underrange or an overrange condition (selectable by soldering jumper) for 0-60 mV and 12-60 mV input types. No indication are available for 0-20 mA, 0-5 V and 0-10 V input types.
T/CT/C
T/CT/C
T/C
typetype
typetype
type
RangesRanges
RangesRanges
Ranges
0 1 2 3 4 5 6 7 8
9 10 11 12 13
23 24 25 26 27 28 29 30 31 32 33 34 35 36
JJ
JJ
J KK
KK
K
TT
TT
T EE
EE
E NN
NN
N
SS
SS
S RR
RR
R
BB
BB
B
LL
LL
L UU
UU
U
G(W)G(W)
G(W)G(W)
G(W) D(W3)D(W3)
D(W3)D(W3)
D(W3) C(W5)C(W5)
C(W5)C(W5)
C(W5)
P.(*)P.(*)
P.(*)P.(*)
P.(*)
-150 / 1830 °F
-150 / 2500 °F
-330 / 750 °F
-150 / 1470 °F
-150 / 2550 °F
-60 / 3200 °F
-60 / 3200 °F 32 / 3300 °F
-150 / 1650 °F
-330 / 1110 °F 0 / 4170 °F 0 / 4170 °F 0 / 4170 °F
-150 / 2550 °F
-100 / 1000 °C
-100 / 1370 °C
-200 / 400 °C
-100 / 800 °C
-100 / 1400 °C
-50 / 1760 °C
-50 / 1760 °C 0 / 1820 °C
-100 / 900 °C
-200 / 600 °C 0 / 2300 °C 0 / 2300 °C 0 / 2300 °C
-100 / 1400 °C
InputInput
InputInput
Input
typetype
typetype
type
RangesRanges
RangesRanges
Ranges
- 200 / 850 °C
- 330 / 1560 °F
14 37
RTD Pt 100 RTD Pt 100
RTD Pt 100 RTD Pt 100
RTD Pt 100
ΩΩ
ΩΩ
DIN 43760DIN 43760
DIN 43760DIN 43760
DIN 43760
LHL-1-NEUTRO.pmd 19/04/2004, 15.0722
Page 26
23
STANDARD RANGES TABLE
D) LOGIC INPUTSD) LOGIC INPUTS
D) LOGIC INPUTSD) LOGIC INPUTS
D) LOGIC INPUTS (for models with RS-485 only) (for models with RS-485 only)
(for models with RS-485 only) (for models with RS-485 only)
(for models with RS-485 only) This instrument is provided of 1 logic input used for remote acknowledgement. NOTESNOTES
NOTESNOTES
NOTES
1) Use an external dry contact capable of switching 0.5 mA, 5 V DC.
2) The instrument needs 100 ms to recognize a contact status variation.
3) The logic inputs are
NOT NOT
NOT NOT
NOT isolated by the
measuring input.
OUTPUTSOUTPUTS
OUTPUTSOUTPUTS
OUTPUTS Output updating time Output updating time
Output updating time Output updating time
Output updating time :
- 250 ms when a linear input is selected
- 500 ms when a TC or RTD input is selected. OUTPUT 1OUTPUT 1
OUTPUT 1OUTPUT 1
OUTPUT 1 Type: relayType: relay
Type: relayType: relay
Type: relay SPDT contact . Contact ratedContact rated
Contact ratedContact rated
Contact rated: 3 A at 250 V AC on resistive load. FunctionFunction
FunctionFunction
Function: Safety limiter output. ActionAction
ActionAction
Action: reverse (fail-safe). OUTPUT 2OUTPUT 2
OUTPUT 2OUTPUT 2
OUTPUT 2 Type: relayType: relay
Type: relayType: relay
Type: relay SPST contact . Contact ratedContact rated
Contact ratedContact rated
Contact rated: 2 A at 250 V AC on resistive load. FunctionFunction
FunctionFunction
Function: Alarm output ActionAction
ActionAction
Action: direct/reverse programmable by front
keyboard.
ALARMALARM
ALARMALARM
ALARM ActionAction
ActionAction
Action: Direct or reverse acting. Alarm functionsAlarm functions
Alarm functionsAlarm functions
Alarm functions: configurable as process alarm, band alarm or deviation alarm. Alarm resetAlarm reset
Alarm resetAlarm reset
Alarm reset: automatic reset, manual reset or "Silence" function is programmable. Stand by (mask) alarmStand by (mask) alarm
Stand by (mask) alarmStand by (mask) alarm
Stand by (mask) alarm: the alarm can be configured with or without stand by (mask) function.
Process alarmProcess alarm
Process alarmProcess alarm
Process alarm: Operative mode Operative mode
Operative mode Operative mode
Operative mode : High or low programmable. Threshold Threshold
Threshold Threshold
Threshold : programmable in engineering unit within the readout span. HysteresisHysteresis
HysteresisHysteresis
Hysteresis: programmable from 0.1 % to 10.0 % of the readout span.
Input typeInput type
Input typeInput type
Input type
15 16 17 18 19 20 21 22
0 - 60 mV 0 - 60 mV
0 - 60 mV 0 - 60 mV
0 - 60 mV
12 - 60 mV12 - 60 mV
12 - 60 mV12 - 60 mV
12 - 60 mV
0 - 20 mA 0 - 20 mA
0 - 20 mA 0 - 20 mA
0 - 20 mA 4 - 20 mA 4 - 20 mA
4 - 20 mA 4 - 20 mA
4 - 20 mA
0 - 5 V0 - 5 V
0 - 5 V0 - 5 V
0 - 5 V 1 - 5 V1 - 5 V
1 - 5 V1 - 5 V
1 - 5 V 0 - 10 V0 - 10 V
0 - 10 V0 - 10 V
0 - 10 V 2 - 10 V2 - 10 V
2 - 10 V2 - 10 V
2 - 10 V
impedanceimpedance
impedanceimpedance
impedance
> 1 M
< 5
> 400 k
> 400 k
AccuracyAccuracy
AccuracyAccuracy
Accuracy
0.2 % + 1 digit @ 25°C
(77 °F)
LHL-1-NEUTRO.pmd 19/04/2004, 15.0723
Page 27
24
Band alarmBand alarm
Band alarmBand alarm
Band alarm Operative modeOperative mode
Operative modeOperative mode
Operative mode: Inside or outside band programmable. ThresholdThreshold
ThresholdThreshold
Threshold: programmable from 0 to 500 units. Hysteresis Hysteresis
Hysteresis Hysteresis
Hysteresis : programmable from 0.1 % to 10.0 % of the readout span.
Deviation alarmDeviation alarm
Deviation alarmDeviation alarm
Deviation alarm Operative mode Operative mode
Operative mode Operative mode
Operative mode : High or low programmable. Threshold Threshold
Threshold Threshold
Threshold : programmable from - 500 to +500 units. Hysteresis Hysteresis
Hysteresis Hysteresis
Hysteresis : programmable from 0.1 % to 10.0 % of the readout span.
SERIAL COMMUNICATION INTERFACESERIAL COMMUNICATION INTERFACE
SERIAL COMMUNICATION INTERFACESERIAL COMMUNICATION INTERFACE
SERIAL COMMUNICATION INTERFACE (OPTION)(OPTION)
(OPTION)(OPTION)
(OPTION) TypeType
TypeType
Type: insulated RS-485. Protocol typeProtocol type
Protocol typeProtocol type
Protocol type: MODBUS or JBUS. Baud rateBaud rate
Baud rateBaud rate
Baud rate: programmable from 600 to 19200 BAUD. Byte formatByte format
Byte formatByte format
Byte format: 8 bit. ParityParity
ParityParity
Parity: even, odd or none programmable. Stop bitStop bit
Stop bitStop bit
Stop bit: one. AddressAddress
AddressAddress
Address: from 1 to 255. Output voltage levelsOutput voltage levels
Output voltage levelsOutput voltage levels
Output voltage levels: according to EIA standard.
MAINTENANCEMAINTENANCE
MAINTENANCEMAINTENANCE
MAINTENANCE
1) REMOVE POWER FROM THE POWER SUPPLY TERMINALS AND FROM RELAY OUTPUT TERMINALS
2) Remove the instrument from case.
3) Using a vacuum cleaner or a compressed air jet (max. 3 kg/cm
2
) remove all deposit of dust and dirt which may be present on the louvers and on the internal circuits trying to be careful for not damage the electronic components.
4) To clean external plastic or rubber parts use only a cloth moistened with:
- Ethyl Alcohol (pure or denatured) [C
2H5
OH] or
- Isopropil Alcohol (pure or denatured) [(CH
3)2
CHOH] or
- Water (H
2
O)
5) Verify that there are no loose terminals.
6) Before re-inserting the instrument in its case, be sure that it is perfectly dry.
7) re-insert the instrument and turn it ON.
LHL-1-NEUTRO.pmd 19/04/2004, 15.0724
Page 28
Appendix A.1
APPENDIX AAPPENDIX A
APPENDIX AAPPENDIX A
APPENDIX A DEFAULT PARAMETERSDEFAULT PARAMETERS
DEFAULT PARAMETERSDEFAULT PARAMETERS
DEFAULT PARAMETERS
DEFAULT PARAMETERSDEFAULT PARAMETERS
DEFAULT PARAMETERSDEFAULT PARAMETERS
DEFAULT PARAMETERS Loading Default Operating ParametersLoading Default Operating Parameters
Loading Default Operating ParametersLoading Default Operating Parameters
Loading Default Operating Parameters The control parameters can be loaded with predetermined default values. These are the settings loaded into the instrument prior to shipment from the factory. To load the default values proceed as follows:
a ) Press and hold the key and press the
key; the displays will show:
b) Press either the or key; the display will
show:
c) Press the "FUNC" key; the display will show:
This indicates that the loading procedure has been initiated. After about 3 seconds the loading
procedure is complete and the instrument reverts to the “Normal Display Mode.” The following is a list of the default operating parameters loaded during the procedure:
Default Operating Parameters ListDefault Operating Parameters List
Default Operating Parameters ListDefault Operating Parameters List
Default Operating Parameters List Parameter Default Value Alarm Acknowledge OFF Software Key Unlock Setpoint Threshold Low range value (if low limit)
High range value (if high or
high/low limit) Setpoint1 Threshold Low range value Setpoint Threshold Hysteresis 0.1% Alarm Threshold Low range (if process alarm)
100 (if deviation or band
alarm) Alarm Hysteresis 0.1%
LHL-A-NEUTRO.pmd 19/04/2004, 15.281
Page 29
Appendix A.2
Loading Default Configuration ParametersLoading Default Configuration Parameters
Loading Default Configuration ParametersLoading Default Configuration Parameters
Loading Default Configuration Parameters The configuration parameters can be loaded with predetermined default values. These are the settings loaded into the instrument prior to shipment from the factory. To load the default values proceed as follows:
a ) Internal switch V101 must be open. b) The upper display will show:
c) Press the key; the lower display will show
the firmware version.
d) Still holding the key, press the key; the
display will show:
e ) Press the key to select Table 1 (Euro-
pean) or Table 2 (American) default parameters; the display will show:
f) Press the FUNC key; the display will show:
This indicates that the loading procedure has been initiated. After about 3 seconds the procedure is complete and the instrument reverts to the “COnF” display. The following is a list of the default configuration parameters loaded during the procedure:
LHL-A-NEUTRO.pmd 19/04/2004, 15.282
Page 30
Appendix A.3
PARA.
L1 L2 L3 L4 r1
r2 r3 r4 r5 r6 r7 c1 c2 c3 c4 c5 P1 P2 P3 P4 PF n1 t1
Table 1 European nbUS 1 19200 8E Type J (-100 to 1000 °C)
----.
-100 1000 0 1 second uP Hi 1 Auto 0 1 second nonE H.A. rEV OFF 1 second 0 10 seconds
Table 2 American nbUS 1 19200 8E Type J (-150 to 1830 °F)
----.
-150 1830 0 1 second uP Hi 0 Auto 0 1 second nonE H.A.Ac rEV OFF 1 second 0 30 seconds
LHL-A-NEUTRO.pmd 19/04/2004, 15.283
Page 31
APPENDIX BAPPENDIX B
APPENDIX BAPPENDIX B
APPENDIX B THERMOCOUPLE COMPENSATING CABLE COLOR CODES.THERMOCOUPLE COMPENSATING CABLE COLOR CODES.
THERMOCOUPLE COMPENSATING CABLE COLOR CODES.THERMOCOUPLE COMPENSATING CABLE COLOR CODES.
THERMOCOUPLE COMPENSATING CABLE COLOR CODES.
ThermocoupleThermocouple
ThermocoupleThermocouple
Thermocouple
MaterialMaterial
MaterialMaterial
Material
TT
TT
T Copper Constantan
J/L J/L
J/L J/L
J/L Iron Constantan
KK
KK
K Nickel Chromium
Nickel Aluminum
RR
RR
R Platinum/Platinum
13% Rhodium
SS
SS
S Platinum/Platinum
10% Rhodium
EE
EE
E Chromel
Constantan
BB
BB
B Platinum 30% Rh
Platinum 6% Rh
NN
NN
N Nicrosil / Nisil
BritishBritish
BritishBritish
British
BS 1843BS 1843
BS 1843BS 1843
BS 1843
+ White
- Blue Blue
+ Yellow
- Blue Black
+ Brown
- Blue Red
+ White
- Blue Green
+ White
- Blue Green
+ Brown
- Blue Brown – –
AmericanAmerican
AmericanAmerican
American
ANSI MC 96.1ANSI MC 96.1
ANSI MC 96.1ANSI MC 96.1
ANSI MC 96.1
+ Blue
- Red Blue
+ White
- Red Black
+ Yellow
- Red Yellow
+ Black
- Red Green
+ Black
- Red Green
+ Violet
- Red Violet
+ Grey
- Red Grey –
GermanGerman
GermanGerman
German
DIN 43710DIN 43710
DIN 43710DIN 43710
DIN 43710
+ Red
- Brown Brown
+ Red
- Blue Blue
+ Red
- Green Green
+ Red
- White White
+ Red
- White White – –
– – – –
FrenchFrench
FrenchFrench
French
NFE 18-001NFE 18-001
NFE 18-001NFE 18-001
NFE 18-001 + Yellow
- Blue Blue
+ Yellow
- Black Black
+ Yellow
- Purple Yellow
+ White
- Green Green
+ White
- Green Green – –
– – – –
LHL-B-NEUTRO.pmd 19/04/2004, 15.301
Page 32
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR
WARRANTY WARRANTY
WARRANTY WARRANTY
WARRANTY RETURNS, please have the following
information available BEFORE contacting OMEGA:
1.Purchase Order number under which the product was PURCHASED,
2.Model and serial number of the product under warranty, and
3.Repair instructions and/or specific problems relative to the product.
FOR
NON-WARRANTY NON-WARRANTY
NON-WARRANTY NON-WARRANTY
NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3.Repair instructions and/or specific problems relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2003 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
WARRANTY/DISCLAIMERWARRANTY/DISCLAIMER
WARRANTY/DISCLAIMERWARRANTY/DISCLAIMER
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period
of 37 months from date of purchase. The OMEGA® Warranty adds an additional one (1) month grace period to the normal three (3) years product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; or misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for anyOMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for anyOMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with informationomissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information
omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with informationomissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information
omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified andprovided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and
provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified andprovided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and
provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS ORfree of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR
free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS ORfree of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR
free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY ANDIMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND
IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY ANDIMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND
IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forthFITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth
FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forthFITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth
FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty,herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty,
herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty,herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty,
herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon whichnegligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which
negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon whichnegligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which
negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
CN3271-add2.pmd 19/08/2004, 16.363
Page 33
Where Do I Find Everything I Need for Process MeasurementWhere Do I Find Everything I Need for Process Measurement
Where Do I Find Everything I Need for Process MeasurementWhere Do I Find Everything I Need for Process Measurement
Where Do I Find Everything I Need for Process Measurement
and Control? OMEGA…Of Course!and Control? OMEGA…Of Course!
and Control? OMEGA…Of Course!and Control? OMEGA…Of Course!
and Control? OMEGA…Of Course!
Shop online at www.omega.comShop online at www.omega.com
Shop online at www.omega.comShop online at www.omega.com
Shop online at www.omega.com
TEMPERATURETEMPERATURE
TEMPERATURETEMPERATURE
TEMPERATURE
Thermocouple, RTD & Thermistor Probes, Connectors,
Panels & Assemblies
Wire: Thermocouple, RTD & ThermistorCalibrators & Ice Point ReferencesRecorders, Controllers & Process MonitorsInfrared Pyrometers
PRESSURE, STRAIN AND FORCEPRESSURE, STRAIN AND FORCE
PRESSURE, STRAIN AND FORCEPRESSURE, STRAIN AND FORCE
PRESSURE, STRAIN AND FORCE
Transducers & Strain GagesLoad Cells & Pressure GagesDisplacement TransducersInstrumentation & Accessories
FLOW/LEVELFLOW/LEVEL
FLOW/LEVELFLOW/LEVEL
FLOW/LEVEL
Rotameters, Gas Mass Flowmeters & Flow ComputersAir Velocity IndicatorsTurbine/Paddlewheel SystemsTotalizers & Batch Controllers
pH/CONDUCTIVITYpH/CONDUCTIVITY
pH/CONDUCTIVITYpH/CONDUCTIVITY
pH/CONDUCTIVITY
pH Electrodes, Testers & AccessoriesBenchtop/Laboratory MetersControllers, Calibrators, Simulators & PumpsIndustrial pH & Conductivity Equipment
DATA ACQUISITIONDATA ACQUISITION
DATA ACQUISITIONDATA ACQUISITION
DATA ACQUISITION
Data Acquisition & Engineering SoftwareCommunications-Based Acquisition SystemsPlug-in Cards for Apple, IBM & CompatiblesDatalogging SystemsRecorders, Printers & Plotters
HEATERSHEATERS
HEATERSHEATERS
HEATERS
Heating CableCartridge & Strip HeatersImmersion & Band HeatersFlexible HeatersLaboratory Heaters
ENVIRONMENTAL MONITORINGENVIRONMENTAL MONITORING
ENVIRONMENTAL MONITORINGENVIRONMENTAL MONITORING
ENVIRONMENTAL MONITORING AND CONTROLAND CONTROL
AND CONTROLAND CONTROL
AND CONTROL
Metering & Control InstrumentationRefractometersPumps & TubingAir, Soil & Water MonitorsIndustrial Water & Wastewater TreatmentpH, Conductivity & Dissolved Oxygen Instruments
M4084
CN3271-add2.pmd 19/08/2004, 16.364
Loading...