Omega CL23A, CL24, CL26, CL25, CL27 User Manual

http://www.omega.com
e-mail: info@omega.com
®
User’s Guide
CL23A, 24, 25, 26, 27
Digital Calibrator/Thermometer
Servicing North America:
ISO 9001 Certified Stamford, CT 06907-0047
Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com
Canada: 976 Bergar
Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada@omega.com
For immediate technical or application assistance:
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
SM
Customer Service: 1-800-622-2378 / 1-800-622-BEST
SM
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
SM
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Mexico and Latin America:
Tel: (95) 800-TC-OMEGA
SM
FAX: (95) 203-359-7807
En Espan˜ ol: (203) 359-1660 ext: 2203 e-mail: espanol@omega.com
Servicing Europe:
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: (31) 20 6418405 FAX: (31) 20 6434643 Toll Free in Benelux: 06 0993344 e-mail: nl@omega.com
Czech Republic: Ostravska 767, 733 01 Karvina
Tel: 42 (69) 6311899 FAX: 42 (69) 6311114 e-mail: czech@omega.com
France: 9, rue Denis Papin, 78190 Trappes
Tel: (33) 130-621-400 FAX: (33) 130-699-120 Toll Free in France: 0800-4-06342 e-mail: france@omega.com
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: 49 (07056) 3017 FAX: 49 (07056) 8540 Toll Free in Germany: 0130 11 21 66 e-mail: germany@omega.com
United Kingdom: 25 Swannington Road, P.O. Box 7, Omega Drive,
ISO 9002 Certified Broughton Astley, Leicestershire, Irlam, Manchester,
LE9 6TU, England M44 5EX, England Tel: 44 (1455) 285520 Tel: 44 (161) 777-6611 FAX: 44 (1455) 283912 FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
e-mail: uk@omega.com
omega.com
TM
OMEGA
®
OMEGAnetSMOn-Line Service Internet e-mail
http://www.omega.com info@omega.com
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
TABLE OF CONTENTS
General Information......................................................................... 2
Common Specifications................................................................... 2
Unique Specifications ...................................................................... 5
Features........................................................................................... 6
Manual Addenda ............................................................................. 6
Unpacking Your Unit........................................................................ 7
Safety Information ........................................................................... 7
Safety Precautions .......................................................................... 7
Battery Installation/Replacement..................................................... 9
Memory Backup.............................................................................. 10
Operation with Rechargeable Battery............................................. 10
Operating Instructions..................................................................... 11
11. Diagnostics and Error Messages...................................... 11
12. Initial Turn On ................................................................... 11
3. Meter Mode Operation (Thermocouple) ........................... 13
14. Meter Mode Operation (RTD, Thermistor, Ohms)............ 14
15. Data Logging & Recall...................................................... 15
16. Calibrator Mode Operation (Thermocouples) ................... 17
17. Calibrator Mode Operation (RTD, Thermistor, Ohms) ..... 18
18. Storage and Recall of Calibrator Settings ........................ 20
9. Step Function.................................................................... 22
10. Ramp Function ................................................................. 22
Service Information......................................................................... 23
11. Calibration Procedures ..................................................... 23
12. Calibration Verification...................................................... 32
13. Mechanical Parts Diagram................................................ 38
Warranty ......................................................................................... 45
Limitation of Warranty..................................................................... 45
Statement of Calibration ................................................................. 45
1
GENERAL INFORMATION
This manual provides operating instructions and maintenance information for five temperature instruments. These instruments are high performance cali­brator-thermometers capable of simulating and measuring a wide-variety of sensors. In addition, features such as data storage and ramping further enhance their versatility.
It is recommended that you read this manual thoroughly, especially the sec­tions on safety, prior to operating these instruments.
COMMON SPECIFICATIONS
THERMOMETER INPUTS: THERMOCOUPLE; Miniature TC jack. RTD, OHMS,
THERMISTOR; T-series instrumentation connector.
CALIBRATOR OUTPUTS: THERMOCOUPLE; Miniature TC jack. RTD, OHMS, THERMISTOR; T-series instrumentation connector.
THERMOCOUPLE TYPES: K, J, T, E, N, B, R, S, G, C, D. RTD TYPES: 1100platinum (= 0.00385)
1000platinum (= 0.00385)
THERMISTOR TYPE: 2252(YSI Series 400).
RANGES: RTK; –200°C to 1372°C, –328°F to 2502°F
RTJ; –210°C to 760°C, –346°F to 1400°F RTT; –200°C to 400°C, –328°F to 752°F RTE; –230°C to 1000°C, –382°F to 1832°F RTN; –200°C to 1300°C, –328°F to 2372°F RTB; 500°C to 1820°C, 932°F to 3308°F RTR; 0°C to 1768°C, 32°F to 3214°F RTS; 0°C to 1768°C, 32°F to 3214°F RTG; 300°C to 2316°C, 572°F to 4201°F RTC; 0°C to 2316°C, 32°F to 4201°F RTD; 0°C to 2316°C, 32°F to 4201°F
RTD; –200°C to 850°C, –328°F to 1562°F
THERMISTOR; –40°C to 150°C, –40°F to 302°F
OHMS; 0-999.99
OHMS; 0-9,999.9 OHMS; 0-99,999
ACCURACY (18°C to 28°C ambient, 2 years)*:
K, J, T, E, N; ± 0.5°F (rdg ≥ –50°F)
K, J, T, E, N; ±1.0°F (rdg < –50°F)
B, R, S, G, C, D; ±1.7°F
2
3
RTD; ±0.2°F (rdg – 50°F)
RTD; ±0.5°F (rdg < – 50°F) RTD; ±0.2°F ±0.04% rdg (meter mode)
THERMISTOR; ±0.5F
OHMS; ±0.02% range
* Exclusive of sensor errors, and lead resistance induced errors.
RESOLUTION: TEMPERATURE; 0.1°/1° F/C
RESOLUTION: OHMS; 0.01/0.1/1
REPEATABILITY (1 week at constant ambient temperature):
K, J, T, E, N; ±0.2°F typ.
B, R, S, G, C, D; ±1.0°F typ.
RTD, THERMISTOR; ±0.1°F typ.
OHMS; ±0.01% rng typ.
TEMPERATURE COEFFICIENT:
From 18°C to 28°C, included in accuracy specifications. From 0°C to 18°C, and 28°C to 50°C:
K, J, T, E, N; ±(0.002% rdg + 0.03°C)/°C B, R, S, G, C, D; ±(0.002% rdg + 0.1°C)/°C
RTD, THERMISTOR, OHMS; ±(0.002% rdg + 0.004% rng)/°C
METER-MODE INPUT-CURRENT (THERMOCOUPLES): 50nA typ.
METER-MODE READING-RATE: 1 reading/second typ.
METER-MODE EXCITATION-CURRENTS:
100RTD; <2mA
1000RTD; <200µA
999.99; <2mA
9999.9; <200µA 99999; <50µA
THERMISTOR; T 113°F <50µA
T> 113°F, <2mA
CALIBRATOR MODE OUTPUT LOAD (THERMOCOUPLES): 500min.
CALIBRATOR-MODE EXCITATION-CURRENTS:
100RTD; 0.5 – 1mA (0.1 – 2mA typ.)
1000RTD; 50 – 100µA (10-200µA typ.)
999.99; Same as 100RTD
9999.9; Same as 1000RTD 99999; 10 – 20µA (1 – 20µA typ.)
THERMISTOR; T ≤113°F, 10 – 20µA
T> 113°F, 0.5 – 1mA
MAXIMUM COMMON MODE VOLTAGE: 42V peak to earth.
4
KEYPAD: 12 momentary switches with tactile feedback select;
• On/Off
• Change/Enter
• Clear/–0
• Sensor Select/1
• °F/°C/2
• Resolution/3
• Operate/4
• Store/5
• Recall/6
• Calibrator/Meter/Up Ramp/7
• Calibrator Step/Down Ramp/8
• Calibrator Ramp/Meter Hold/9
DISPLAY: 5 digit LCD, 0.4" height, and decimal point. Annunciators;
• Loaded Memory Locations (10 for data storage, 4 for ramp parameters)
• Low Battery (BAT)
• Number Change Mode (NUM)
• Calibrator Mode (CALIB)
• Meter Mode (METER)
• Operate Mode (OPR)
• Store Mode (STO)
• Recall Mode (RCL)
• Meter Hold (HOLD)
• Scale (°F/°C)
• Sensor Type
POWER OFF CONFIGURATION RETENTION: Instrument retains last selected;
• Sensor Type
• °F/°C Scale
• 0.1°/1° Resolution
• Calibrator/Meter Mode
• Memory Contents (data storage and ramp parameters).
DIAGNOSTICS: Display codes indicate following conditions;
• 'BAT': Low Battery
• 'OPEN': Open Input Thermocouple, RTD or Thermistor.
• 'E-1' (momentary): Invalid Keypad Entry
• 'E-2' (momentary): Ramp Function Not Installed
• 'HI': Temperature Reading or Keypad Entry Exceeds Rating
• 'LO': Temperature Reading or Keypad Entry Too Low For Accuracy
• LCD Test: During power-up, all segments and annunciators enabled momentarily
ENVIRONMENTAL LIMITS FOR OPERATING: 0°C to 50°C, less than 90% relative humidity (R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35°C to 50°C.
ENVIRONMENTAL LIMITS FOR STORAGE: –35°C to 60°C, less than 95% relative humidity (R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35°C to 60°C.
POWER: 9 volt alkaline battery (NEDA 1604A).
5
BATTERY LIFE, CONTINUOUS: 10 hrs. typical, alkaline; 3 hrs typical, Ni-Cd (rechargeable).
BATTERY INDICATOR: Display indicates BAT when less than 10% of life remains. DIMENSIONS, WEIGHT: 7.0" × 2.9 × 1.1". Net weight 12 oz.
UNIQUE SPECIFICATIONS
NOTE: The specifications below are specific to the model(s) as identified by model
number. The difference specifications are in addition to the OVERALL SERIES SPEC­IFICATIONS found on pages 2-5.
MODEL CL23A/CL23EC:
SENSOR SELECTION: K, J, T/K, J, T, L Thermocouples ACCURACY: (18°C to 28°C ambient, 2 years): ±0.5°F (rdg –50°F).
±0.04% rdg (rdg > 1250°F) ±1.0°F (rdg < –50°F)
TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications. From 0°C to 18°C, and 28°C to 50°C; less than ±(0.006% rdg +0.03°C)/°C.
RAMP FUNCTION: Not Included.
BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd
(rechargeable).
MODEL CL24:
SENSOR SELECTION: K, J, T, E Thermocouples ACCURACY: (18°C to 28°C ambient, 2 years): ±0.5°F (rdg –50°F).
±0.04% rdg (rdg > 1250°F) ±1.0°F (rdg < –50°F)
TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications. From 0°C to 18°C, and 28°C to 50°C; less than ±(0.006% rdg +0.03°C)/°C.
RAMP FUNCTION: Not Included.
BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd
(rechargeable).
MODEL CL25/CL25EC:
SENSOR SELECTION: K, J, T, E, N, B, R, S, G, C, D/K, J, T, E, N, B, R, S, L, C, D Thermocouples
RAMP FUNCTION: Included.
BATTERY LIFE, CONTINUOUS: 16 hrs typical, alkaline; 4 hrs typical, Ni-Cd
(rechargeable)
MODEL CL26:
SENSOR SELECTION: K, J, T, E Thermocouples
100Platinum RTD 2252Thermistor 1,000/100,000Resistance
ACCURACY: (18°C to 28°C ambient, 2 years): ±0.5 °F (rdg ≥ –50°F).
±0.04% rdg (rdg > 1250°F) ±1.0°F (rdg < –50°F) OHMS; ±0.04% range
TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications. From 0°C to 18°C, and 28°C to 50°C.
K, J, T, E; ±(0.006% rdg + 0.03°C)/°C RTD, THERMISTOR, OHMS; ±(0.006% rdg +0.012% rng)/°C
RAMP FUNCTION: Not Included.
MODEL CL27/CL27EC:
SENSOR SELECTION: K, J, T, E, N, B, R, S, G, C, D/K, J, T, E, N, B, R, S, L, C,
D Thermocouples
100/1000Platinum RTD 1,000/10,000Resistance
RAMP FUNCTION: Included.
FEATURES
• Combination calibrator, thermometer and datalogger.
• Function selection and numeric data entered via 12 keyswitch color coded keypad.
• Keyswitches have tactile feedback.
• 5 digit LCD includes annunciators for operating modes.
• Up to 11 thermocouple types, 2 RTD types, and thermistor.
• 1° or 0.1° resolution over temperature range of each sensor.
• °F and °C scales.
• Reading hold mode.
• 10 memory locations for saving meter readings and/or calibrator outputs.
• Step mode for quick sequential recall of calibrator outputs from memory.
• Manual/automatic ramping for testing controller setpoints.
• Conforms to ITS-90 thermocouple tables.
• Dust proof, splash proof, drop proof construction.
• Built-in tilt stand/hanger for bench use or hands free field measurements.
• User friendly programming.
• Retains data and programming, even when turned off.
• Performs diagnostic tests and indicates fault conditions.
• Low battery and open sensor indications.
MANUAL ADDENDA
Improvements or changes to this manual will be explained on an addendum included with the instrument. All change information should be incorporated immediately into the appropriate places in the manual.
6
UNPACKING YOUR UNIT
Remove the Packing List and verify that you have received all equipment. If you have any questions about the shipment, please call the OMEGA Customer Service Department at 1-800-622-2378 or (203) 359-1660.
When you receive the shipment, inspect the container and equipment for any signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
NOTE: The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event reshipment is necessary.
Make sure that you have the following items in the packing box.
• 1 CL20 Series Calibrator and Holster.
• 1 Operator’s Manual
• 1 NIST Certificate
• 1 Accessories Kit
SAFETY INFORMATION
SAFETY SYMBOLS AND TERMS
The symbol on the instrument denotes that the user should refer to the
operating instructions.
The WARNING used in this manual explains dangers that could result in
personal injury or death.
The CAUTION used in this manual explains hazards that could damage
the instrument.
SAFETY PRECAUTIONS
WARNING
These instruments are intended for use by qualified person­nel trained in the safe operation of electronic testing equip­ment. Read the instruction manual thoroughly before using, to become familiar with the instrument's operations and capabilities.
!
7
WARNING
Do not touch a temperature probe sheath when measuring excessively high or low temperatures, or toxic substances.
WARNING
Do not attempt to measure temperatures beyond the range of the probe being used. Probe damage or personal injury could result from exceeding a probe's maximum temperature rating.
WARNING
The American National Safety Institute (ANSI) states that a shock hazard exists when probes or sensors are exposed to voltage levels greater than 42VDC or 42V peak AC. Do not use this instrument where voltages at the measurement sur­face exceed these levels.
WARNING
Do not substitute a metal part for the nylon screw in the rear case. Doing so will degrade electrical isolation of the case.
WARNING
The battery is accessible through a cover on the back of the instrument. To avoid electrical shock hazard, disconnect all temperature probes and sensors and turn the unit off before removing the cover.
WARNING
Never use this instrument or any probe or sensor inside a microwave oven.
CAUTION
Avoid making sharp bends in probe or sensor lead wires. Bending lead wires at a sharp angle can damage the wire causing probe failure.
8
CAUTION
Keep inputs electrically isolated from outputs. Do not con­nect an external voltage source between the two connectors, and avoid any external electrical pathways between input and output (T2 & T1).
Failure to do so will disturb meter readings and calibrator output levels. Damage to the instrument is also possible.
Where the meter and calibrator connectors are simultane­ously wired into a system lacking the above isolation, it is necessary to use an "ungrounded" (i.e. electrically isolated) probe at the thermometer input.
BATTERY INSTALLATION/REPLACEMENT
A 9V alkaline battery is supplied with the instrument but not installed. Read the following installation instructions before attempting to install or remove the battery. Replacement batteries should also be alkaline types.
WARNING
Turn the unit off and disconnect any input/output connec­tions before replacing the battery. Put the cover back into place on the battery compartment before resuming use of the instrument.
1. Remove the cover from the battery compartment by sliding it off in the direction of the arrow located on the battery cover.
9
Figure 1. Battery Installation
2. Remove the old battery by lifting the battery extractor loop.
3. Place the new battery in the battery compartment. Be sure to observe proper polarity.
4. Re-install the battery cover before resuming use of the instrument.
NOTES:
• Less than 10% of battery life remains when the BAT annuncia­tor turns on.
• When the battery is excessively discharged, the display is blanked.
• If the instrument is going to be stored for a long period of time or in a high temperature environment, remove the battery to prevent leakage damage.
After a new battery is installed, allow approximately 5 sec-
onds for display turn-on the first time the unit is turned on. During this period, microcomputer initialization is per­formed.
MEMORY BACKUP
During battery replacement, the contents of user-programmed memory (data, operating modes, etc.) can be saved. Prior to removing the old battery, turn off the instrument, and connect an HH22-AC battery charger. Then exchange batteries, and disconnect the battery charger. Do not leave the bat­tery charger connected to instruments with non-rechargeable batteries.
OPERATION WITH RECHARGEABLE
BATTERY
HH22-AC provides a 9-volt Ni-Cd battery and recharger suitable for use with
the unit. This battery provides 3-4 hours of continuous operation. This dura­tion can be extended to several days by operating simultaneously from both the battery and recharger.
Turn off the instrument to recharge the battery. Allow 7 hours to fully recharge the battery. Then unplug the recharger to avoid overcharging.
10
OPERATING INSTRUCTIONS
1. DIAGNOSTICS AND ERROR MESSAGES
Unit self-diagnostics provide fault condition readouts which are described below:
PROBLEM: LIKELY CAUSE:
Blank display, unit does not (1) Improper battery installation. power-up. (1) Check battery polarity.
(2) Dead battery.
BAT Annunciator. (1) Low battery voltage, install a new
battery. If problem persists, con­sult factory.
Display reads OPEN. (1) No thermocouple, RTD or thermis-
tor is plugged into the meter input.
Display reads E1 momentarily. (1) This indicates that an invalid key-
pad entry has been made. Review keystroke sequence, or consult manual for input instructions.
Display reads E2 momentarily. (1) Attempt made to activate non-
installed ramp function. NOTE: RAMP Function enabled on Models CL25 and CL27 only.
Display reads HI. (1) Meter-mode input-temperature
exceeds rating of selected sensor.
(2) Calibrator-mode keypad-entry
exceeds rating of sensor type.
Display reads LO. (1) Meter-mode input-temperature too
low for accurate measurement.
(2) Calibrator-mode keypad-entry too
low for accurate simulation.
2. INITIAL TURN ON
Turn on the instrument by depressing the ON-OFF keyswitch once (Figure
2). At power-up, the unit first performs a display test. All segments and annunciators of the liquid crystal (LCD) are momentarily turned on for visual confirmation by the user (Figure 3).
NOTE: In case of missing or poor-contrast segments, contact
factory.
11
If the unit is turning on for the first time after a new battery is installed, it will automatically default to the METER mode (ie. temperature measurement operation). It also defaults to K-type thermocouples, and °F readings with
0.1° resolution. The corresponding LCD annunciators are enabled:
12
Figure 3. Display Test
Figure 2. Keypad
Loading...