It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is
constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the
CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any
errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
TABLE OF CONTENTS
General Information ......................................................................... 2
Common Specifications ................................................................... 2
Limitation of Warranty..................................................................... 45
Statement of Calibration................................................................. 45
1
GENERAL INFORMATION
This manual provides operating instructions and maintenance information for
five temperature instruments. These instruments are high performance calibrator-thermometers capable of simulating and measuring a wide-variety of
sensors. In addition, features such as data storage and ramping further
enhance their versatility.
It is recommended that you read this manual thoroughly, especially the sections on safety, prior to operating these instruments.
THERMOCOUPLE TYPES: K, J, T, E, N, B, R, S, G, C, D.
RTD TYPES: 1100Ω platinum (∝ = 0.00385)
THERMISTOR TYPE: 2252Ω (YSI Series 400).
RANGES:RTK; –200°C to 1372°C, –328°F to 2502°F
THERMISTOR; –40°C to 150°C, –40°F to 302°F
ACCURACY (18°C to 28°C ambient, 2 years)*:
1000Ω platinum (∝ = 0.00385)
RTJ; –210°C to 760°C, –346°F to 1400°F
RTT; –200°C to 400°C, –328°F to 752°F
RTE; –230°C to 1000°C, –382°F to 1832°F
RTN; –200°C to 1300°C, –328°F to 2372°F
RTB; 500°C to 1820°C, 932°F to 3308°F
RTR; 0°C to 1768°C, 32°F to 3214°F
RTS; 0°C to 1768°C, 32°F to 3214°F
RTG; 300°C to 2316°C, 572°F to 4201°F
RTC; 0°C to 2316°C, 32°F to 4201°F
RTD; 0°C to 2316°C, 32°F to 4201°F
RTD; –200°C to 850°C, –328°F to 1562°F
OHMS; 0-999.99Ω
OHMS; 0-9,999.9Ω
OHMS; 0-99,999Ω
K, J, T, E, N; ± 0.5°F (rdg ≥ –50°F)
K, J, T, E, N; ±1.0°F (rdg < –50°F)
B, R, S, G, C, D; ±1.7°F
2
RTD; ±0.2°F (rdg ≥ – 50°F)
RTD; ±0.5°F (rdg < – 50°F)
THERMISTOR; ±0.5F
* Exclusive of sensor errors, and lead resistance induced errors.
RESOLUTION: TEMPERATURE; 0.1°/1° F/C
RESOLUTION: OHMS; 0.01Ω/0.1Ω/1Ω
REPEATABILITY (1 week at constant ambient temperature):
RTD, THERMISTOR; ±0.1°F typ.
TEMPERATURE COEFFICIENT:
From 18°C to 28°C, included in accuracy specifications.
From 0°C to 18°C, and 28°C to 50°C:
SENSOR SELECTION: K, J, T, E, N, B, R, S, G, C, D/K, J, T, E, N, B, R, S, L, C,
RAMP FUNCTION: Included.
D Thermocouples
100Ω/1000Ω Platinum RTD
1,000Ω/10,000Ω Resistance
FEATURES
• Combination calibrator, thermometer and datalogger.
• Function selection and numeric data entered via 12 keyswitch color coded
keypad.
• Keyswitches have tactile feedback.
• 5 digit LCD includes annunciators for operating modes.
• Up to 11 thermocouple types, 2 RTD types, and thermistor.
• 1° or 0.1° resolution over temperature range of each sensor.
• °F and °C scales.
• Reading hold mode.
• 10 memory locations for saving meter readings and/or calibrator outputs.
• Step mode for quick sequential recall of calibrator outputs from memory.
• Manual/automatic ramping for testing controller setpoints.
• Conforms to ITS-90 thermocouple tables.
• Dust proof, splash proof, drop proof construction.
• Built-in tilt stand/hanger for bench use or hands free field measurements.
• User friendly programming.
• Retains data and programming, even when turned off.
• Performs diagnostic tests and indicates fault conditions.
• Low battery and open sensor indications.
MANUAL ADDENDA
Improvements or changes to this manual will be explained on an addendum
included with the instrument. All change information should be incorporated
immediately into the appropriate places in the manual.
6
UNPACKING YOUR UNIT
Remove the Packing List and verify that you have received all equipment. If
you have any questions about the shipment, please call the OMEGA
Customer Service Department at 1-800-622-2378 or (203) 359-1660.
When you receive the shipment, inspect the container and equipment for any
signs of damage. Note any evidence of rough handling in transit. Immediately
report any damage to the shipping agent.
NOTE: The carrier will not honor any claims unless all shipping
material is saved for their examination. After examining and
removing contents, save packing material and carton in the event
reshipment is necessary.
Make sure that you have the following items in the packing box.
• 1 CL20 Series Calibrator and Holster.
• 1 Operator’s Manual
• 1 NIST Certificate
• 1 Accessories Kit
SAFETY INFORMATION
SAFETY SYMBOLS AND TERMS
The symbol on the instrument denotes that the user should refer to the
▲▲
!
operating instructions.
The WARNING used in this manual explains dangers that could result in
personal injury or death.
The CAUTION used in this manual explains hazards that could damage
the instrument.
SAFETY PRECAUTIONS
WARNING
These instruments are intended for use by qualified personnel trained in the safe operation of electronic testing equipment. Read the instruction manual thoroughly before using,
to become familiar with the instrument's operations and
capabilities.
7
WARNING
Do not touch a temperature probe sheath when measuring
excessively high or low temperatures, or toxic substances.
WARNING
Do not attempt to measure temperatures beyond the range of
the probe being used. Probe damage or personal injury
could result from exceeding a probe's maximum temperature
rating.
WARNING
The American National Safety Institute (ANSI) states that a
shock hazard exists when probes or sensors are exposed to
voltage levels greater than 42VDC or 42V peak AC. Do not
use this instrument where voltages at the measurement surface exceed these levels.
WARNING
Do not substitute a metal part for the nylon screw in the rear
case. Doing so will degrade electrical isolation of the case.
WARNING
The battery is accessible through a cover on the back of the
instrument. To avoid electrical shock hazard, disconnect all
temperature probes and sensors and turn the unit off before
removing the cover.
WARNING
Never use this instrument or any probe or sensor inside a
microwave oven.
CAUTION
Avoid making sharp bends in probe or sensor lead wires.
Bending lead wires at a sharp angle can damage the wire
causing probe failure.
8
CAUTION
Keep inputs electrically isolated from outputs. Do not connect an external voltage source between the two connectors,
and avoid any external electrical pathways between input
and output (T2 & T1).
Failure to do so will disturb meter readings and calibrator
output levels. Damage to the instrument is also possible.
Where the meter and calibrator connectors are simultaneously wired into a system lacking the above isolation, it is
necessary to use an "ungrounded" (i.e. electrically isolated)
probe at the thermometer input.
BATTERY INSTALLATION/REPLACEMENT
A 9V alkaline battery is supplied with the instrument but not installed. Read
the following installation instructions before attempting to install or remove
the battery. Replacement batteries should also be alkaline types.
WARNING
Turn the unit off and disconnect any input/output connections before replacing the battery. Put the cover back into
place on the battery compartment before resuming use of
the instrument.
1. Remove the cover from the battery compartment by sliding it off in the
direction of the arrow located on the battery cover.
Figure 1. Battery Installation
9
2. Remove the old battery by lifting the battery extractor loop.
3. Place the new battery in the battery compartment. Be sure to observe
proper polarity.
4. Re-install the battery cover before resuming use of the instrument.
NOTES:
• Less than 10% of battery life remains when the BAT annunciator turns on.
• When the battery is excessively discharged, the display is
blanked.
• If the instrument is going to be stored for a long period of time
or in a high temperature environment, remove the battery to
prevent leakage damage.
• After a new battery is installed, allow approximately 5 sec-
onds for display turn-on the first time the unit is turned on.
During this period, microcomputer initialization is performed.
MEMORY BACKUP
During battery replacement, the contents of user-programmed memory
(data, operating modes, etc.) can be saved. Prior to removing the old battery,
turn off the instrument, and connect an HH22-AC battery charger. Then
exchange batteries, and disconnect the battery charger. Do not leave the battery charger connected to instruments with non-rechargeable batteries.
OPERATION WITH RECHARGEABLE
BATTERY
HH22-AC provides a 9-volt Ni-Cd battery and recharger suitable for use with
the unit. This battery provides 3-4 hours of continuous operation. This duration can be extended to several days by operating simultaneously from both
the battery and recharger.
Turn off the instrument to recharge the battery. Allow 7 hours to fully
recharge the battery. Then unplug the recharger to avoid overcharging.
10
OPERATING INSTRUCTIONS
1. DIAGNOSTICS AND ERROR MESSAGES
Unit self-diagnostics provide fault condition readouts which are described
below:
PROBLEM:LIKELY CAUSE:
Blank display, unit does not(1) Improper battery installation.
power-up.(1) Check battery polarity.
BAT Annunciator.(1) Low battery voltage, install a new
Display reads OPEN.(1) No thermocouple, RTD or thermis-
Display reads E1 momentarily.(1) This indicates that an invalid key-
Display reads E2 momentarily.(1) Attempt made to activate non-
Display reads HI.(1) Meter-mode input-temperature
Display reads LO.(1) Meter-mode input-temperature too
(2) Dead battery.
battery. If problem persists, consult factory.
tor is plugged into the meter input.
pad entry has been made. Review
keystroke sequence, or consult
manual for input instructions.
installed ramp function. NOTE:
RAMP Function enabled on
Models CL25 and CL27 only.
exceeds rating of selected sensor.
(2) Calibrator-mode keypad-entry
exceeds rating of sensor type.
low for accurate measurement.
(2) Calibrator-mode keypad-entry too
low for accurate simulation.
2. INITIAL TURN ON
Turn on the instrument by depressing the ON-OFF keyswitch once (Figure
2). At power-up, the unit first performs a display test. All segments and
annunciators of the liquid crystal (LCD) are momentarily turned on for visual
confirmation by the user (Figure 3).
NOTE: In case of missing or poor-contrast segments, contact
factory.
11
Figure 2. Keypad
Figure 3. Display Test
If the unit is turning on for the first time after a new battery is installed, it will
automatically default to the METER mode (ie. temperature measurement
operation). It also defaults to K-type thermocouples, and °F readings with
0.1° resolution. The corresponding LCD annunciators are enabled:
12
METER: (unit functioning as a thermometer)
K:(K-type thermocouple)
°F:(Fahrenheit readings)
OPR:(Instrument is operational)
If no input thermocouple is connected to the unit at this time, the numeric
segments will indicate "OPEN".
Plugging a thermocouple into the meter jack (connector T2, on the top right
side) will give actual thermocouple temperature readings.
NOTE: To obtain full accuracy, allow 1-2 minutes after connecting a thermocouple plug, for thermal setting.
If the instrument is to be used as a thermometer or ohmmeter, refer to the
Meter Mode Operation section for detailed instructions. For use as a temperature-calibrator or ohms simulator, refer to Calibrator Mode Operation.
3. METER MODE OPERATION (Thermocouples)
Set-Up:
1. Turn on instrument by depressing the ON/OFF key.
2. Check that the BAT annunciator turns off following the power-up LCD test.
If it does not turn off, less than 10% of battery remains. Refer to Battery
Installation/Replacement instructions.
3. If the meter is already in the METER mode, the METER annunciator will
be on. Otherwise the calibrator mode is enabled and the CALIB annunciator is on. Depress the METER key to return to meter mode.
4. Connect an input thermocouple (TC) to connector T2. Select the appropriate TC Type with the select-key. A display annunciator will indicate the
selected TC-type.
NOTE: Space and legibility constraints on the display preclude
the use of dedicated annunciators for the less common tungstenrhenium thermocouple types (G, C & D). Instead, these types are
annunciated momentarily on the alpha-numeric display when first
selected. Both T and R annunciators remain on to indicate a
Tungsten-Rhenium selection. To recall the actual G, C, D selection, turn the unit off, then on again for another momentary readout.
5. Select temperature scale with °F/°C key. The corresponding annunciator
is enabled.
13
Loading...
+ 33 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.