One Omega Drive, River Bend Technology CentreNorthbank, Irlam, ManchesterM44TEL: Toll Free in United Kingdom: 0800-488-488e-mail:
8034,118O
(0)20
+31
sales@omegaeng.nl
(0)59
+420
(0)16137
+33
(0)7056 9398-O
+49
info@omega.de
5BD United Kingdom
(0)161777 6611
+44
sales@omega.co.uk
06907-0047
5A1, Canada
H7L
/
l-800-TC-OMEGA”
/
LA Amstelveen, The Netherlands
3472121
Karvina,
6311899
Cartier,
78280 Guyancourt, France
29 00
Czech Republic
Internet e-mail
info@omega.com
FAX: (203) 359-7700
FAX: (514) 856-6886
l-SOO-622-BEST@
/
l-800-USA-WHEN@
e-mail: espanol@omega.com
info@omega.com.mx
(0)20
+31
FAX:
FAX: e-mail:
FAX:
FAX:
FAX:
(0)59
+420
info@omegashop.cz
(0)l 30 57 54 27
+33
(0)7056
+49
(0)161777 6622
+44
6434643
6311114
9398-29
It is the policy of OMEGA to comply with all worldwide safety and apply. OMEGA is constantly pursuing certification of its products to the European New ApproachDirectives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. acceptsno liability for any errors it contains, and reserves the right to alter specifications without notice.WARNING: These products are not designed for use in, and should not be used for, human applications.
EMC/EMI
regulations that
M3594-060%CDE681Ser
-ies.qxd
7/17/03
2:14
PM Page i
m
CDE68 1 Series Conductivity/Resistivity Sensors
.-
HELPFUL
IDENTIFIERS
In addition to information on installation and operation, this instruction manualmay contain WARNINGS pertaining to user safety, CAUTIONS regardingpossible sensor malfunction, and NOTES on important, useful operatingguidelines.
A
warning looks like this. Its purpose is to warn
this sensor of the potential for personal injury.
A
caution looks like this. Its purpose is to alert the user of this
sensor to possible
A
note looks like this. Its
sensor to important operating information.
sensor malfunction or damage.
purpdse
is to alert the user of this
the user of
i
M3594-0603-CDEhElSeries.qxd
Table of Contents
Part One
Part Two
2:14
7/17/03
-
Introduction
Section 1 General Information
Section 2 Specifications
-
Installation
SectionSection 2 Mounting
Section 3 Sensor/Interconnect Cable Details
PM Page ii
1.1 Description:
1.2 Operating Precautions
2.1 Insertion Mounting
2.2Immersion
3.1
3.2Interconnect
3.3
Benefits of Enhanced Performance DesignCompatible Meters
Benefits of Enhanced Performance DesignCDE681-series
tolerances using high quality, rugged materials for demanding ultrapure waterand pure water applications. Each sensor is:
??
Individually tested to determine its absolute cell constant (shown on its labelas K = each sensor’s OMEGA-certified instrument configuration or calibration, ensures the highest possiblemeasurement accuracy
??
Built with a Pt 1000 RTD temperature element located at its tip to provideexceptionally fast response to changes in temperature with high measuringaccuracy
Compatible MetersThese sensors are for use with the CDCN684, CDCN685, CDCN686 Analyzers
and CDTX680 Series transmitters only.Sensor CharacteristicsBasic Cell Constant: Installation Style: For sensors with a 0.05 cell constant, use K-inch or %-inch male
NPT compression fittings made of Kynar (PVDF) or 316 stainless steel. Forsensors with any other cell constant, use a %-inch male NPT compression fittingmade of Kynar or 316 stainless steel. In all cases, the fitting enables the sensor tobe insertion mounted, up to 4 inches/ 102 mm deep, into a pipe tee or vessel.Reversing the fitting enables the sensor to be fastened onto the end of a pipe forimmersion mounting.
Termination Style: An integral 6 m (20 ft.) long cable
compression fitting style sensors are manufactured to exacting
),
and its temperature element value (to the nearest 1.0 ohm). Entering
“K”
value and temperature
0.2”C).
(*
0.05,0.5,1.0,5.0,
or 10.
“T”
factor during
1.2
Operating Procedures
Always consider the temperature/pressure ratings of the mounting hardwareused to install the sensor. The sensor and hardware combination is an integratedsystem. The hardware material usually limits the system’s temperature/pressurerating. Refer to Section 2 for complete specifications.
for extended sensor body style used with ball valveassembly), PTFE Teflon insulator, and treated VitonO-ring seals
.
When used with Kynar (PVDF) compression fitting:
150°C at 1.7 bar (302°F at 25 psi)
When used with 316 stainless steel compression fitting:
150°C at 13.7 bar (302°F at 200 psi)
.
When used with Kynar (PVDF) compression fitting:
10.3 bar at 36°C (150 psi at 97°F)
When used with 316 stainless steel compression fitting:
13.7 bar at 150°C (200 psi at 302°F)
.
O-3 m (O-10 ft.) per second (fully immersed)
Pt
1000 RTD
.
.6
wire cable (4 conductors and two isolated shieldIntegral
Model No.
Cell
Constant
CDE681 -A-KCDE681 -A-SCDE681 -B-K
CDE681 -B-SCDE681 -C-KCDE681 -C-SCDE681 -D-K5.0
CDE681 -D-S
CDE681 -E-K10.0CDE681-E-S
1
I
1
(
0.05
0.5
1.0
Compression
Material
Kynar (PVDF)
316SS
Kynar
316SS
Kynar
316SS
Kynar
316SS
Kynar
316SS
-
Fitting
Thread
%
NPT
%
NPT
%
NPT
%
NPT
%
NPT
2
M3594m0603mCDE681Series.qxd
PART
TWO
SECTION 1
SECTION 2
2:14
7/17/03
PM Page 3
m
CDE681 Part 2
INSTALLATION
-LOCATION REQUIREMENTS
-
Locate the sensor as close as possible to the measuring instrument. Do notexceed a distance of 91 m (300 feet) between the sensor and instrument.
Mounting
-
lnstallatio
-
.f
CDE681-series
The
compression fitting style sensor may be insertion mountedinto a pipe tee or vessel fitting. By reversing the compression fitting, the sensorcan be fastened to the end of an appropriate length pipe for immersionmounting. The “longer version” compression style sensor is intended to bemounted into a ball valve assembly, enabling sensor removal without stoppingthe process flow. Figure 2-l shows the sensor’s general dimensions.
l/2’
DIA. FOR
00.5 CELL CONSTANT3/4’
DIA. FOR ALL
OTHER CELL CONSTANTS
2.1
l/2’
NPT THREADS
Insertion Mounting
1
3/4’
OR
Figure
\
2-l.
3,-V COMPRESSION
I,?
OR
General Dimensions and Cable Wire Details
To ensure optimum measurement performance, follow these guidelines wheninsertion mounting the sensor:
??
Install the sensor into the pipe run so that the process flows directly into the
end of the sensor (see Figure 2-2).
??
Preferably, mount the sensor in a vertical position to eliminate the possibilityof trapped air bubbles from contacting its electrodes which can causemeasurement error. This also prevents loose pipe line sediment fromaccumulating and obstructing the sensor electrodes.
FIrrING
CLEAR W/BLACK (SENSORBLACK (SENSE)RED (DRIVE)
(TEMP+)
WHITE BLUE (TEMP-)CLEAR (INNER SHIELD)
SHIELD)
3
M3594-0603-CDE681Series.qxd
-
7/17/03
CDE681 Part 2
2:14
PM Page 4
-
Installation
1. Install a pipe tee of appropriate size
m
(‘x
to 2 inch) and material into the process
pipe. If necessary, screw a respectively-sized reducer into the pipe tee.
avoid
leaks.
Recommendation:
sealant with Teflon. (Exception: For higher temperaturesolutions, sealing with Teflon tape may not be adequate.)
Use
Teflon tape or pipe
SENSOR CABLE
FT)
6.1 M (20
CLEARANCE
FOR SENSOR
FLOW
Figure
2-2.
Insertion Mounting Details
SENSOR
3/4
l/2
ORCOMPRESSIONFIrrING
3/4” NPT
i/2
OR
TREADED TEE
DIMENSIONS mm
(in)
u
2. Remove the compression fitting from the sensor and screw it into the pipe tee(or reducer, if used).
3. Electrically connect sensor to analyzer. Refer to analyzer instruction manual
for details.
4. Calibrate analyzer using procedure in analyzer instruction manual.
5. After calibration, mount the sensor into the tee:A. Place
I
’
comtxession nut first.comoression nut and ferrule onto the sensor.
I
4
4
Correctly orient the ferrule
2-3 to
get
an effective seal.
ontoXe
sensor as shown in Figure
M3594m0603mCDE681Series.qxd
-
7/17/03
2:14
PM Page 5
m
lnsta~~ation
-
CDE681 Part 2
OPTIONAL;tYoN
TWO-PIECE FERRULE(ORIENT IN THIS DIRECTION)
;pC$;;ESSION
REDUCER(AS
REQUIRED)
,f
PROLESS
FLOW
Figure
Insert the sensor into the compression fitting. Then adjust the insertion
B.
2-3.
Compression Fitting Parts Arrangement For Insertion Mounting
oosition the sensor’s “cross-flow” holes at the center of the tee,de&h to
direitly in the process
and
Rotate the sensor to align one
flow path.
or
its
“cross-flow” holes with
the exiting process flow path as shown in Figure
After tightening the compression nut, the ferrule will bepermanently crimped. Therefore, make sure that the sensor is
inserted to the proper depth before tightening the compressionnut.
C. With the
sensor properly positioned, tighten the compression nut onto thecompression fitting to crimp the ferrule. Use one wrench to hold the fittingand another to turn the nut 1 to 1% turns. This should provide an effectiveprocess seal. Also, the crimped ferrule becomes a convenient referenceindicator for insertion depth when re-inserting the sensor after cleaning.
This completes the insertion mounting
2-3.
I_
M3594-0603-CDE681Series.qxd
CDE681 Part 2
2:14
7/17/03
-
Installation
PM Page 6
m
Installation Tip!
Re-using Compression Fitting, Nut, and Crimped Ferrule:If the sensor is re-installed, the compression fitting and nut can be re-used. The
crimped ferrule, however, may need to be cut away from the sensor to remove it,making it unusable. If you can remove the crimped ferrule without destroying it,and if it can still provide an effective process seal, you can probably re-use it atleast the number of times listed below:
Recommended Crimped Ferrule Re-use
Ferrule Material
316 stainless steel
Kynai
(PVDF)
Number of Re-uses
1 after initial use3 after initial use
f
2.2
Immersion Mounting
1. Reverse the compression fitting assembly on the sensor so that the longerthreaded section faces towards the sensor cable end (see Figure 2-4).
2. Position the compression fitting assembly at the far cable end of the sensor
body and tighten the compression nut onto the compression fitting. Use one
wrench to hold the fitting and another to turn the nut 1 to 1% turns to crimpthe ferrule. This provides a process seal.
1
INCH PIPE
!
+-+
I
I
I
88.9 TO 139.7(3.50 TO 5.50)
SENSOR
EXTENSION
L---
-:
_.
3/4 OR (CUSTOMER PROVIDED)
1
INCH NPT COUPLING
3/4 OR (CUSTOMER PROVIDED)
/
I
DIMENSIONS mm (I”.)
1
3,4 INCH
l/2
INCH X OR NPT REDUCER
l/2
INCH X
COMPRESSION FITTING
SENSOR
1
3/4 INCH X
INCH
3/4 INCH
.:/
Figure 2-4. Immersion Mounting Details
M3594-0603-CDE681Series.qxd
2:14
l/17/03
3.
Route the sensor cable through a pipe of an appropriate material and length.
Screw the compression fitting onto the end of the pipe (or coupling, if used).
PM Page 7
m
Recommendation:
(Exception: For higher temperature
Teflon tape may not be adequate.)
4.
Fasten a Unilet junction box onto the other end of the pipe.
5. Run interconnect cable from the analyzer into the Unilet junction box. Connectthe sensor and interconnect cable wires, by matching colors, to the terminals inthe junction box. Fasten the cover onto the junction box.
6. Electrically connect the sensor interconnect cable wires to the analyzer. Referto the analyzer instruction manual for details.
7. Calibrate the analyzer using the procedure in the analyzer instruction manual.
8. After calibration, mount the sensor into the process.
This completes the immersion mounting.
Use Teflon tape or
ipe sealant
P
so utions, sealing with
with
Teflon.
:I
7
I_
M3594m0603mCDE681Series.qxd
CDL681
Part 2
SECTION 3
2:14
7/17/03
-
Installation
SENSOR/INTERCONNECT CABLE DETAILS
-
PM Page 8
m
3.1
3.2
Sensor Cable Details
The sensor’s integral cable is a 6-wire crosslinked polyethylene-jacketed cablewith 4 conductors and two isolated shield wires. Refer to Figure 2-l for thefunction and color of each wire in the sensor’s integral cable.
Interconnect Cable Details
The OMEGA interconnect cable
CDE3600-CAB
is provided with unfinishedends since it must often be shortened during installation. The cable is verysimilar to the sensor’s integral cable except that it has two additional conductors(green and yellow) which are not required. When stripping the interconnectcable during termination, purposely cut off these green and yellow wires fromeach end of the stripped-back cable. This ensures the same wire color codingused by the sensor’s integral cable.
only its
OMEGA strongly recommendsinterconnect cable. If a different cable is used, it must haveequivalent construction: four conductors, and two separateisolated shieldsthe overall cable. These specific cable characteristics protectthe measurement signal from electromagnetic interference.Using a cable with different construction may interfere withthe measurement system’s ability to properly measure.
-
one
shielding the signal, and one shielding
ising
CDE3600-CAB
i
To correctly terminate the ends of the interconnect cable, refer to Figure 2-6 andfollow this procedure:
Carefully
strip back 2% inches of the outer cable jacket, the outer shield foil,and the cellophane binder. This exposes the sensor shield wire, the innershield wire, and the three foil-wrapped wire pairs.
Cut off the exposed
Peel back and cut off the exposed inner shield foil from the red/black andblue/white wire pairs.
Carefully strip back an additional
shield foil.
_
Be careful not to damage thecellophane binder.
2%
inches of only the yellow and green wire pair.
!4
inch of the outer cable jacket and outer
exposedsection of the
M3594m0603mCDE681Series.qxd
2:14
7/17/03
5.
Carefully position a
PM Page 9
m
CDE68 1 Part 2
2%
inch long piece of shrink tubing or tape on the baresensor shield wire %-inch from the end as shown in Figure 2-6 to insulate anddistinguish it from the inner shield wire. Doing this exposes X-inch of bareshield wire beyond the tubing or tape for connection purposes.
6. Carefully position a l-inch long piece of shrink tubing or tape on the cable asshown in Figure 2-6 to secure all wires.
”
-
Installation
Do not fold back the cellophane
binder exposed in step 4.
7. Using an ohmmeter or test light, verify that the sensor shield wire youinsulated is not shorted to the bare inner shield wire. If the wires are shorted,cut the cable to get a new unfinished end and start over at step 1.
8. Strip
%
inch of insulation from the ends of the red, black, white, and bluewires. Tin these leads, the insulated sensor shield wire, and the bare innershield wire with solder.
9. Connect the interconnect cable to the analvzer in the same wav as the sensor
_I
,
cable, by matching colors as indicated.
Typ
_
0.25
L6.41
SHIELD
SENSOR RED (DRIVE)
BLACK
(
SENSE
)
WHITE (TEMP
StuEINNER SHIELD
+)
-)
(TEMP
OUTER CABLE JACKET
OUTER
3.3
Connecting Interconnect Cable
To Analyzer:
interconnect cable wires to appropriate SENSOR terminals in the same way as
the sensor wires would be directly connected.
BARE INNER SHIELD WIRE
-
FOlL
SHlELO
CELLOPHANE BINDER
INNER
SHIELO
FOIL
ALL
[mm]
OlMS
INCH
Figure 2-5.Interconnect Cable Termination Details
Refer to the instrument instruction manual and connect the
9
I_
M3594&0603-CDE681Series.qxd
CDE681 Part 3
PART THREESECTION 1
2:14
7/17/03
-
SERVICE AND MAINTENANCE
-
RECOMMENDED CLEANING PROCEDURE
Keep the sensor reasonably clean to maintain measurement accuracy. The timebetween cleanings (days, weeks, etc.) is affected by the characteristics of theprocess solution and can only be determined by operating experience.
1.
Remove most contaminate buildup by carefully wiping the inner electroderod, and the concentric outer electrode tube (inner and outer surfaces) with asoft clean cloth. Then rinse the sensor with clean, warm water.
Prepare a mild soap solution. Use warm water and dishwashing detergent,
2.
Borax hand soap, or a similar soap.
3.
Soak the sensor for 2 to 3 minutes in the soap solution.
4.
Use a small bristle brush, cotton swab (Q-tip), or pipe cleaner to scrub theentire measuring end of the sensor, thoroughly cleaning the electrode surfaces.If detergent solution cleaning cannot remove surface deposits, use muriaticacid (or another dilute acid) to dissolve the deposits. The acid should be asdilute as possible, but yet strong enough to clean. Experience will helpdetermine which acid to use and how dilute it can be. Some stubborn coatingsmay require a different cleaning agent. For assistance in these difficult cases,contact the OMEGA Service Department (see page 12).
Before cleaning with acid, determine if any hazardous reaction products couldform. (Example: A sensor used in a cyanide bath should not be put directlyinto a strong acid for cleaning because poisonous cyanide gas could beproduced.) Acids are hazardous. Wear appropriate eye protection and clothingin accordance with Material Safety Data Sheet recommendations.
Soak the sensor in dilute acid for no morewith clean, warm water and then place the sensor back into the mild soapsolution for 2 to 3 minutes to neutralize any remaining acid.
5.
Rinse the sensor in clean, warm water.
6.
Calibrate the analyzer using the procedure in the analyzer instruction manual.If calibration cannot be attained, check the sensor using the procedure in PartThree, Section
PM Page10
-
Service and Maintenance
2.1.
m
than 5 minutes. Rinse
I
the sensor
10
I_
M3S94m0603-CDE681Series.qxd
7/17/03
2:14
PM Page11
m
2.1
Checking Sensor Operation
Use the troubleshooting section in the analyzer instruction manual to determinewhether the sensor or analyzer is in-operative. If you suspect the sensor, check itusing this procedure:
1. Disconnect the sensor from the analyzer (or junction box, if using interconnectcable).
2. Clean the sensor using the procedure in Part Three, Section 1.
3. Using an ohmmeter, check all of the measurement point resistance readingsshown in Table A below.
Be sure that the ohmmeter is set to its highest range for allinfinite (open circuit) resistance readings shown in Table A.
Measurement PointsCorrect Resistance Readings
Between blue and white wires
Between red wire and sensor bodyLess than
Between black
wire and inner electrode
1089-l 106 ohms at
Less
than 5 ohms
5 ohms
23-27°C
3
Between black and red wires
Between black and white wires
Between red and white wires
Between red and inner shield wires
Between black and inner
Between white and inner shield wires
Between
4.
If you cannot get the required readings for one or more of the resistance
checks in step 3, the sensor is probably inoperative. Refer to OMEGA’swarranty/replacement plan for sensor replacement details. If all resistance
checks are okay, the sensor may still be inoperative. In this case, moreextensive troubleshooting is required. Please consult the OMEGA CustomerService Department for details (see page 12).
outer and inner shield wiresInfinite (open circuit)
shield
wires
Infinite (open circuit)
Infinite (open circuit)
Infinite (open circuit)
Infinite (open circuit)
Infinite (open circuit)
Infinite (open circuit)
11
I_
M3594-0603-CDE681Series.qxd
CDE681 Part 3
2:14
7/17/03
-
Service and Maintenance
PM Page 12
m
2.2
Customer Assistance
If you need assistance in troubleshooting or repair service, please contact yourlocal OMEGA representative, or the OMEGA Customer Service Department at:
l-800-633-2378 or l-203-359-1660.
We can also be reached on the Internet at
email: info at omega.com
All sensors returned for repair or replacement must be freight prepaid and
include the following information:
1. A clearly written description of the malfunction.
2. Name of person to contact and the phone number where they can be reached.
3. Proper return address for shipping sensor(s) back. Include preferred shippingmethod (UPS, Federal Express, etc.) if applicable.
4. A purchase order if sensor(s) is out of warranty to cover costs of repair.
inadequate
for repair costs. It is recommended to use the originalOMEGA shipping carton or an equivalent. Also, OMEGA
will not accept sensors returned for repair or replacement
unless they are thoroughly cleaned and all process chemicalsare removed.
packaging,
the customer assumes
www.omega.com
responsibility
li
12
V
I_
M3594-0603-CDE681Series.qxd
-
OMEGA ENGINEERING, INC. warrants this unit to be free of defects
period of 13 grace period to the normal ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at nocharge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser,
including but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operatingconditions outside of OMEGA’s control. Components which wear are not warranted, including but not
limited to contact points, fuses, and
OMEGA is pleased to offer suggestions on the use of its various products. However,OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for anydamages that result from the use of its products in accordance with information provided byOMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will beas specified and free of defects.
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS ORAND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability ofOMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of thecomponent upon which liability is based. In no event shall OMEGA be liable forconsequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used:
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
f
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility
as set forth
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
,_;
I$+$$>
:.:z..:,
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN
(AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
(f,
cov5
2~14
one
PM Page
WARRANTY/DISCLAIMER
(1)
year
triacs.
product warranty
REQUESTS/INQUIRIES
~~~~:~~~~~~~~~~~~~~
7/17/03
%,
Fq,F..
.%
~~~~~~~~~~~~~~~:
in
,$
,
>y*
.........
\
months
our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify
from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month
RETURN
~r~~~~~~~~~~~~~:~~
in materials and workmanship for a
to cover handling and shipping time. This
OMEGA MAKES NO OTHER WARRANTIES OR
IMPUED, EXCEPT THAT OF
(I)
-‘a*‘,,~
..
&ww+”
~~~~~~~~~~~~~~~~~~~~~~~
TITLE,
as a “Basic
’
q,+b
%
d
d
*%
,
*<:
%
i
f
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR
WARRANTY
following information available BEFOREcontacting OMEGA:
1.
Purchase Order number under which the
was PURCHASED,
2.
Model and serial number of the product underwarranty, and
3.
Repair instructions and/or specific problemsrelative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affordsour customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.0
Copyright 2003 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without theprior written consent of OMEGA ENGINEERING, INC.
RETURNS, please have the
FOR for current repair charges. Have the followinginformation available BEFORE contacting OMEGA:
product
NON-WARRANTY
Purchase Order number to cover the COST
of the repair,Model and serial number of the product, andRepair instructions and/or specific problems
relative to the product.
REPAIRS, consult OMEGA
M3594-0603-CDE681Series.qxd
-
2:14
l/17/03
Where Do
PM Page
I
cov2
Find Everything
h
I
Need for
Process Measurement and Control?
OMEGA...Of Course!
Shop online at
www,omega.com
TEMPERATURE
&
B
Thermocouple, RTD
m
Wire: Thermocouple, RTD
&
@
Calibrators
@
Recorders, Controllers
G?
Infrared Pyrometers
Ice Point References
Thermistor Probes, Connectors, Panels
&
Thermistor
&
Process Monitors
PRESSURE, STRAIN AND FORCE
&
B’
Transducers
[a
Load Cells
@
Displacement Transducers
0’
Instrumentation
Strain Gages
&
Pressure Gages
&
Accessories
FLOW/LEVEL
&
@
Rotameters, Gas Mass Flowmeters
m
Air Velocity Indicators
Turbine/Paddlewheel
II?t’Ji?
Totalizers
&
Batch Controllers
Systems
Flow Computers
pH/CONDUCTIVITY
&
pH
Electrodes, Testers
B
G?
BenchtoplLaboratory
lZi’
Controllers, Calibrators, Simulators
&
Conductivity Equipment
pH
w
Industrial
Accessories
Meters
&
Pumps
DATA ACQUISITION
&
B
Data Acquisition
B
Communications-Based Acquisition Systems
Gl’
Plug-in Cards for Apple, IBM
@
Datalogging Systems
@
Recorders, Printers
Engineering Software
&
Compatibles
&
Plotters
&
Assemblies
HEATERS0’
Heating Cable
B
Cartridge
B
Immersion
0’
Flexible Heaters
m
Laboratory Heaters
ENVIRONMENTAL
MONITORING AND CONTROL
m
Metering
m
Refractometers
m
Pumps
m
Air, Soil
m
Industrial Water
pH,
w
&
Conductivity
&
Strip Heaters
&
Band Heaters
&
Control Instrumentation
Tubing
&
Water Monitors
&
Wastewater Treatment
&
Dissolved Oxygen Instruments
M3594/0603
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.