Omega CA450 User Manual

User’s Manual
Model CA450
Process Multimeter
Store this manual in a safe place for future reference.
1st Edition: Aug. 2012
M-5187
“HART” is a registered trademark of the HART Communication Foundation.
■ About This Manual
• Every effort has been made to ensure accuracy in the preparation of this manual. However, should any errors or omissions come to the attention of the user, please contact
Omega Engineering.
• The contents of this manual are subject to change without prior notice because of improvement in performance or function.
• All rights reserved. No part of this manual may be reproduced in any from without permission of an author.
Thank you for purchasing our
This user’s manual describes the specications and handling precaution for this Process
Multimeter.
Before using this Process Multimeter, thoroughly read this manual to get a clear understanding
proper use.
on
Always observe the following instructions. Failure to do so may impair the protection provided by the instrument and probes, and may result in electrical shock or other dangers that may lead to serious injury or the loss of life. Omega Engineering is in no way liable for any damage resulting from the user’s mishandling of the product.
Revision Information
1
st Edition: Aug. 2012
CA450 Process Multimeter
.
1

Regarding Safe Use of This Product

WARNING
CAUTION
Note
Tip
For safe use of this product, the following safety symbols are used on the product and manual:
This indicates that the operator must refer to an explanation in the instruction manual in order to avoid the risk of serious injury or the loss of life.
This indicates that the operator must refer to an explanation in the instruction manual in order to avoid the risk of injury or damage to the product.
This indicates information that is essential for handling the instrument or should be noted in order to familiarize yourself with the instrument’s operating procedures and/or functions.
This indicates supplemental information for explanations.
Danger! Handle with Care
This symbol indicates that the operator must refer to an explanation in the instruction manual in order to avoid risk of injury or death of personnel or damage to the instrument.
This symbol indicates double insulation or reinforced insulation.
This symbol indicates DC voltage/current.
This symbol indicates AC voltage/current.
This symbol indicates AC and DC.
This symbol indicates a fuse.
This symbol indicates a battery.
This symbol indicates ground (earth).
2
WARNING
■ Always observe the following instructions. Failure to do so may result in electrical shock or other dangers that may lead to serious injury or the loss of life.
Test Leads and Lead cables
Use the products provided by Omega Engineering for this device.
• Do not use degraded or damaged test leads or lead cables.
Check the continuity of the test leads and lead cables.
• When you attach or remove the test leads or lead cables or remove the case (for example to change the batteries), be sure to remove the lead cables from the circuit under measurement.
When you remove the case (for example to change the batteries), be sure to remove the test leads and lead cables from the instrument.
• There are caps at the ends of the test leads. To ensure safety (safety standard EN61010-031), be sure to put the caps on the leads when you use them.
Casing
not use the instrument if there is any damage to the casing or when the casing is
• Do removed.
Fuses
Use fuses of the specied rating when the fuse is replaced.
Operating Environment
Do not operate the instrument in an atmosphere where any ammable or explosive gas is
present.
void using the instrument if it has been exposed to rain or moisture or if your hands are
• A wet.
Disassembly
No person, the specied engineer, is authorized to disassemble this instrument.
3
Concernant l’usage en toute sécurité de ce produit
Avertissement
Attention
Remarque
Conseil
Symboles utilisés sur les appareils et dans le manuel d’instruction:
Indique un danger. Attire l’attention sur une utilisation quipourrait engendrer des accidents susceptibles de provoquer des blessures qui peuvent éventuellement s’avérer mortelles.
Indique un danger. Attire l’attention sur une utilisation qui pourrait engendrer une blessure personnelle et/ou être préjudiciable au produit.
Indique les informations essentielles à la manipulation de l’instrument ou qui doivent être prises en compte an de vous familiariser avec les procédures d’utilisation et/ou l es fonctions de l’instrument.
Ce qui suit indique des informations supplémentaires pour les explications.
Danger ! Manipuler avec soin.
Ce symbole indique que l’opérateur doit se reporter à une explicationdonnée par le
manuel d’instruction, an d’éviter tout accident susceptible de provoquer des blessures
au personnel qui peuvent éventuellement s’avérer mortelles, ou de protéger l’appareil.
Ce symbole indique une double isolation ou une isolation renforcée.
Ce symbole indique une tension/intensité C.C.
Ce symbole indique une tension/intensité C.A.
Ce symbole indique le C.A. et le C.C.
Ce symbole indique un fusible.
Ce symbole indique une batterie.
Ce symbole indique la masse (terre).
4
Avertissement
■ Les précautions suivantes doivent être prises. Dans le cas contraire,des accidents susceptibles de provoquer des blessures qui peuvent éventuellement s’avérer mortelles résultant de dangers tels que des chocs électriques, ou un préjudice au produit, risquent de survenir.
Fils de test et câbles de dérivation
Utilisez les produits fournis par Omega Engineering pour ce dispositif.
• N'utilisez pas de ls de test ou de câbles de dérivation dégradés ou endommagés.
Vériez la continuité des ls de test et des câbles de dérivation.
Lorsque vous xez ou retirez les ls de test ou les câbles de dérivation, ou que vous
enlevez le boîtier (par exemple pour changer les batteries), veillez à retirer les ls de test
et les câbles de dérivation du circuit en cours de mesure.
Lorsque vous enlevez le boîtier (par exemple pour changer les batteries), veillez à retirer
les ls de test et les câbles de dérivation de l'instrument.
Un chapeau est fourni sur le bout d’un l de test.
Utilisez un l de test avec le chapeau placé dessus pour une bonne sécurité (normes de
sécurité: EN 61010-031).
Boîtier
• N’utilisez est enlevé.
Fusibles
Utilisez les fusibles avec l’évaluation spéciée quand le fusible est remplacé.
Environnement d’opération
N’utilisez pas l’instrument là où un gaz ou de la vapeur (atmosphère) inammable ou
explosive quelconque est présente.
• Évitez sont humides.
Démontage
Aucune personne, excepté un ingénieur specialisé, n’est autorisée à démonter cet instrument.
pas l’instrument s’il y a un dommage quelconque au boîtier ou quand le boîtier
d’utiliser l’instrument s’il a été exposé à la pluie ou à l’humidité ou si vos mains
5

Contents

1. Overview .............................................................................................................8
2. Operating Environment .......................................................................................9
3.
Specications ....................................................................................................11
3.1
General Specications ........................................................................... 11
3.2 Accuracy ................................................................................................ 13
4. Operation .......................................................................................................... 16
4.1
Precautions before Measurement ........................................................... 16
4.2 Components ........................................................................................... 17
.3 Measuring Instructions ...........................................................................21
4
4.3.1 AC Voltage Measurement ( , ) ...................................... 22
4.3.2 DC Voltage Measurement ( V, mV) .............................. 22
4.3.3 Measuring Using Sensors (SENSOR) .....................................22
4.3.4 Resistance Measurement (Ω) ................................................... 23
4.3.5 Continuity Check ( ) ............................................................ 24
4.3.6 Diode Test (
4.3.7 DC Current Measurement (
4.3.8 Loop Power Measurement ....................................................... 27
.3.9 Frequency Measurement (Hz) .................................................. 28
4
4.3.10 Auto Hold Function .................................................................29
4.3.11 Peak Hold Function .................................................................. 29
4.3.12 Relative Value Display (REL Δ/%) .......................................... 31
4.3.13 MIN/MAX/AVG Function ....................................................... 32
4.4 Output Instructions ................................................................................. 33
4.4.1 Constant Current Output (SOURCE mode) ............................. 34
4.4.2 Constant Current Output (SIMULATE mode) ......................... 36
4.4.3 Current Sweep Output .............................................................. 37
4.4.4 Current Span Setting ................................................................ 39
4.5 Auto Power-Off (Sleep) Function .......................................................... 39
) ........................................................................ 24
) .............................................. 26
6
4.6 Setup Function ....................................................................................... 40
4
.6.1 Turning the Buzzer On and Off ...............................................40
4.6.2 SENSOR Mode Input, Display, and Unit Settings .................. 41
4.6.3 Current Span Setting for the 100 mA
Range for DCmA Measurement ...............................................42
4.6.4 Setting the Slow Step Time ...................................................... 43
4.6.5 Returning the Settings to Their Factory Default Values .......... 43
4.7 Additional Power-on Functions .............................................................44
4.7.1 Making All the LCD Elements Appear .................................... 44
5. User Calibration Function ................................................................................. 45
5
.1 Calibrating Measurements ..................................................................... 45
5.1.1 Peak Hold Calibration .............................................................. 47
5.1.2 Loop Power Calibration ........................................................... 47
5.2 Calibrating Output ................................................................................. 48
5.3 What to Check after Calibration ............................................................ 50
6. Communication
6.1 Cable Connection and Interface Specications ..................................... 51
Communication Commands ................................................................... 52
6.2
6.3 Detailed Command Descriptions ...........................................................54
6.4 List of Errors ..........................................................................................72
6.5 Communication Command Validity Table ............................................. 73
6.6 Status Byte Format ................................................................................. 75
7. Replacing the Batteries and Fuses ....................................................................76
7.
1 Replacing the Batteries .......................................................................... 76
7.2 Replacing the Fuses ............................................................................... 77
8. Calibration and After-Sales Service .................................................................. 78
9.
Disposing This Product ..................................................................................... 78
9.1 Disposing This Product .............................................................................. 78
9.2
How to Replace and Dispose the Batteries ................................................ 79
10. Troubleshooting ................................................................................................ 80
Appendix
1 Fahrenheit ........................................................................................... 81
................................................................................................. 51
7

1. Overview

• Display
Main display: 5-digit LCD, Subdisplay: 5-digit LCD (Measured values other than current values are displayed with 4 digits.)
• Various Functions
Measurement Functions
voltage, AC voltage, DC current (24 V loop power supply possible), resistance,
DC frequency, continuity check, and diode test
Output Functions
current (constant output, current sweep output, and SIMULATE (SINK) function)
DC
Additional Functions
Data hold (D•H); auto hold (A•H); peak hold (P•H); auto range (Auto); range hold
(Range
Hold); maximum, minimum, and average value recording and measurement; zero
adjustment (Ω); relative measured value display (RELΔ, REL%); and backlight
• Communication (Optional PC communication set required)
Measurement data can be transferred to a PC through the use of a USB communication
and a USB cable.
adapter
• Safety
Conforms to IEC 61010-1, Safety requirements for electrical equipment for measurement, control, and laboratory use. “Safety requirements for electrical equipment for measurement, control, and laboratory use” This instrument uses a current-input terminal shutter for preventing wrong input. This instrument uses high-performance UL-standard fuses.
8

2. Operating Environment

WARNING
Internal Wiring
Outlet
Entrance Cable
Fixed Equipment, etc.
Distribution Board
CAT. II
Equipment
CAT.I
CAT.IV
CAT. III
■ Measurement Category of CA450 and Lead cables
There are restrictions on the maximum voltage level at which the CA450 can be used.
These restrictions are based on the measurement categories specied by the safety
standards.
AC/DC 1000 V for CAT.III, AC/DC 600 V for CAT.IV
The terminals for current
all under category I.
f
Do not use the above terminals and lead cables for measuring points that fall under
measurement categories II, III, or IV.
■ Category of Test leads
With Caps: 1000V 10A CAT.III/600V 10A CAT.IV Without Caps: 1000V 10A CAT.II/600V 10A CAT.II
m
easurement and output and the accessory
lead cables
Measurement
Category
I CAT.I
II CAT.II
III CAT.III
IV CAT.IV
Description Remarks
measurements performed on circuits not
For directly connected to MAINS
F
or measurements performed on circuits
directly connected to the low voltage installation For
measurements performed in the building
installation.
or measurements performed at the source of
F the low-voltage installation.
9
Circuits not connected to a mains power source
Appliances, portable equipment, etc.
Distribution
board, circuit
breaker, etc. Overhead wire, cable
systems, etc.
Note
Radiation immunity affects electrical performance (accuracy) under the conditions in EN61326-1 and EN61326-2-2.
The use of this instrument is limited to domestic, commercial, and light industry applications.
The
instrument may malfunction if it is exposed to strong electromagnetic
interference from nearby devices.
10
3. Specications
3.1 General Specications
Measurement functions: DC voltage, AC voltage, DC current, resistance, frequency, continuity
check, diode test
Additional functions: Data hold (D•H); auto hold (A•H); peak hold (P•H); auto range (Auto);
range hold (Range Hold); maximum, minimum, and average value
recording and measurement; zero adjustment (Ω); relative measured value display (RELΔ, REL%); 24 V loop power supply; internal
resistor on/off for HART communication
Output functions: 20 mA DC current for current output SOURCE and current output
SIMULATE(SINK)
Additional functions: Current span switching and current sweep output
Operation methods: Measurement: ΔΣ modulation
Output: Multiplicative DA Display: 5-digit LCD (7 segment) Numeric display
Measurement Output DC current: 33000 DC current: 25000 Frequency: 19999 Other: 6600
Subdisplay Displays supplemental information for various functions
Polarity indicator Automatic display. Only the minus sign “–” appears.
Over range indicator “OL” Low-battery indicator
Measurement cycle: 2.5 to 5 times a second (however, frequency measurement
Operating temperature and humidity: –20°C to 55°C (80% RH or less) with no condensation Within the range of 40°C to 55°C, the humidity must be
Storage temperature and humidity: –40°C to 70°C (70% RH or less) with no condensation Temperature coefcient (typ.): In the ranges of –20°C to 18°C and 28°C to 55°C, add the accuracy of 23°C ± 5°C × 0.1/°C. Power supply: Four AA-size alkaline batteries (1.5 V LR6)
Battery life: When using alkaline batteries DC voltage measurement: Approx. 140 hours DC current output (SIMULATE): Approx. 140 hours
DC current output (SOURCE) 12 mA (500 Ω load):
appears when the battery voltage is below the
operating voltage.
takes place once a second)
70% RH or less.
Approx. 10 hours
11
Insulation resistance: 100 MΩ or greater at 1000 VDC Withstand voltage: 6.88 kVAC for ve seconds (between the input terminals and the case) External dimensions: Approx. 90 (W) × 192 (H) × 49 (D) mm
Weight: Approx. 600 g (including the batteries) Compliant standards:
Safety standards: EN61010-1 and EN61010-031 Measurement Categories: 1000 V CATIII, 600 V CATIV For current measurement and output: 48 V max, 100 mA max CAT I Lead cables : 70 VDC, 100 mA CAT I
Pollution degree 2, indoor use
Vibration: Sweep vibration frequencies 10 Hz to 55 Hz to 10 Hz Amplitude 0.15 mm (peak value) Duration 30 minutes
Shock: 1 m drop test as dened by the safety standards
Altitude: 2000 m or less
EMC standards: EN61326-1 Class B, EN61326-2-2 EN55011 Class B Group 1
Inuence of radiated immunity: In RF electromagnetic elds of 3 V/m EN61326-1 AC voltage measurement, 600 mV range: 1.5% of range DC voltage measurement, 600 mV range: 1% of range DC current measurement, all ranges: 1.5% of range DC current output: 1.5% of range
EN61326-2-2 AC voltage measurement (6 V range or higher):
W
ithin 5 times the accuracy
DC voltage measurement (6 V range or higher):
Within 5 times the accuracy
Standard accessories: AA-size alkaline batteries 4 Test leads , Lead cables 1 set Fuses (inside the CA450) 440 mA/1000 V 2 User’s manual 1 Blank cover 1
12

3.2 Accuracy

Standard test conditions
(1) Ambient temperature: 23°C ± 5°C
(2) Relative humidity: 45% to 75% RH (no condensation)
(3) Orientation: Horizontal or standing
(4) External magnetic elds: Terrestrial magnetic eld only
* Each of the response times indicates the time it takes for the accuracy to fall within the
specied range.
Accuracy: ± (% of reading + digits)
DC Voltage Measurement V, mV
Range Resolution Accuracy Input Resistance
mV 0.1 mV 0.09% + 2 10 MΩ or more
600
6 V 0.001 V
V 0.01 V
60
V 0.1 V
1000 V 1 V 0.1% + 1
NMRR: 60 dB or more, 50/60 Hz ± 0.1%
CMRR: 120 dB or more, 50/60 Hz (Rs = 1 kΩ)
Response
AC Voltage Measurement V, mV
AC coupling, rms value detection: sine wave
For a range of 5 to 100%, the accuracy for the 1000 V range is 200 V to 1000 V CMRR: 60 dB or more, DC to 60 Hz (Rs = 1 kΩ) For nonsinusoidal waveforms whose crest factor is less than 3, add ±(2% of reading + 2% of range)
t For the 1000 V range, the peak voltage is 1500 V or less Response time: Within 2 seconds
time: Within 1 second
Range Resolution
600 mV 0.1 mV
6 V 0.001 V Approx. 11 MΩ, <50 pF
60
V 0.01 V V 0.1 V
1000 V 1 V
o the accuracy.
0.09% + 1
Accuracy
5
0/60 Hz
0.5% + 5 1% + 5
40 Hz to
500 Hz
Approx. 11 MΩ
Approx. 10 MΩ600
500 Hz to
1 kHz
1.5% + 5
Impedance
10 MΩ or more, <200 pF
Approx. 10 MΩ, <50 pF600
Maximum Input
Input
Voltage
1
000 VDC
1000 Vrms AC
Maximum
Voltage
000 VDC
1
1000 V rms AC
Input
13
DC Current Measurement
Range Resolution Accuracy Voltage Drop 30 mA 0.001 mA 0.05%+2 < 0.3 V
1
00 mA*1 0.01 mA 0.05% + 2 < 0.8 V
*1 Only the 30 mA range can be used during LOOP POWER output.
Response
Resistance Measurement Ω
time: Within 1 second
Range Resolution Accuracy
Maximum Measuring
Current
Open-Loop
Voltage
600 Ω 0.1 Ω 0.2%+2 <1.2 mA <3.5 V
6 kΩ 0.001 kΩ
60 kΩ 0.01 kΩ <13µA
0.2%+1*1
600 kΩ 0.1 kΩ <1.3µA
6 MΩ 0.001 MΩ 0.35%+3
60 MΩ 0.01 MΩ 1%+2 *2
*
1 The accuracy after ZERO CAL
<110µA
<1.3 V
<130 nA
*2 For 40 MΩ to 60 MΩ, the accuracy is 2% + 2. Response time: Within 2 seconds for 600 Ω to 600 kΩ,
within 10 seconds for 6 MΩ to 60 MΩ
Continuity Check
Range Resolution Operating Range
The
buzzer sounds at
600 Ω 0.1 Ω
resistances lower than
Measuring
Current
<1.2 mA < 3.5 V 1000 Vrms
Open-Loop
Voltage
50 ± 30 Ω.
Input
Protective
Voltage
1
000 Vrms
Input Protective
Voltage
Diode Test
Range Resolution Accuracy
2
V 0.001 V 1% + 2
Measuring
Current (Vf = 0.6 V) Approx.
0.5
mA
14
Open-Loop
Voltage
<3.5 V 1
Input Protective
Voltage
000 Vrms
Frequency Measurement in Hz AC Coupling
Range Resolution Accuracy Input Voltage Range
10.00 Hz to 199.99 Hz 0.01 Hz
90.0 Hz to 1999.9 Hz 0.1 Hz
0.005% + 1
3 to 600 Vrms
0.
0.900 Hz to 19.999 kHz 0.001 kHz 0.4 to 600 Vrms
Peak Hold (P•H)
Measurement Function Accuracy Minimum Detection Width
DCV ±100 digits >6
DC Output
ms
Range Resolution Accuracy Load Condition
SOURCE
0 to 20 mA
Compliance voltage 28 V
20 mA 0.001 mA 0.05% of range
SIMULATE (SINK) 0 to 20 mA
External power supply 15 to 48 V overrange up to 25 mA
< 10 mH
24 V Loop Power Supply (LOOP POWER)
Range Load Condition
24 V 24 VDC (typ.), load current 20 mA
15

4. Operation

CAUTION

4.1 Precautions before Measurement

■ Checking the Contents of the Package
After receiving the product and opening the package, check the items described below. If the wrong items have been delivered, if items are missing, or if there is a problem with the appearance of the items, contact your nearest Omega Engineering.
■ Operation and Storage Precautions
• When inserting the batteries into the instrument, refer to section 7.1, “Replacing the Batteries.”
• There is a blank cover on the upper part of the back casing.
Only remove it when connecting the USB adapter of the communication package (sold
separately).
Do not use the instrument near devices that produce high levels of noise or in areas subject to sudden changes in temperature. Doing so may result in unstable readings and errors.
Cleaning
Do not wipe the instrument using benzine, paint thinner, or any other solvent (chemical).
Doing so may lead to discoloration and other problems. Use a dry cloth to clean the instrument.
Storage
Do not leave the instrument exposed to direct sunlight or in a hot and humid location,
such as the inside of a vehicle, for a signicant length of time.
If you do not intend to use the instrument for a signicant period of time, remove its
batteries.
16

4.2 Components

Display (LCD)
SHIFT key MIN/MAX key RANGE key HOLD key
LIGHT key SELECT key
RELΔ/% key
Hz key
Function switch
Input and output terminal area
Red
Caps of Test leads
Black
With Caps: 1000V 10A CAT. /600V 10A CAT. Without Caps: 1000V 10A CAT. /600V 10A CAT.
70 VDC100 mA CAT.
• Panel description • Test leads
• Lead cables
17
1) Function switch Use this switch to turn the power on and off and to select the measurement mode.
Function Mode Function Mode
OFF T
V / Hz
V DC voltage measurement
mV / mV
SENSOR
Ω /
If test leads or lead cables are connected to the current measurement and output terminals, you cannot turn the function switch to the OFF position.
2)
SELECT key
By pressing this key, you can select different measurement modes when the function switch is at the positions listed below. You can switch out of the selected mode by pressing the key again.
mV/ mV position: AC voltage mV measurement
(AC SENSOR measurement in SENSOR mode)
Ω position: Diode test
LOOP POWER position: The SELECT key turns the internal series resistor for HART
position: Operation mode ( / and / ) switching
urns the CA450 off. Continuity check AC voltage measurement and frequency measurement
DC/AC
voltage measurement in mV and various SENSOR modes Resistance measurement Diode test LOOP POWER
communication on and off.
OUT
PUT
DC current measurement
DC current output Constant current output
DC current output Current sweep output
DC current measurement Loop power supply
3) RANGE key Use this key to select the measurement range. Fixed range: “Range Hold” appears on the display. The range increases each time you press RANGE. AUTO range: “AUTO” appears on the display. To select AUT
4) HOLD key Use this key to hold the measured values.
Press HOLD to switch between data hold, auto hold, peak hold, and no hold (release).
(Hold modes that are not supported for the selected measurement function are skipped.)
O range, hold down RANGE for 1 second or longer.
18
Data hold: The currently displayed measured values are held. “D•H” appears on
the display.
Auto hold: The measured values are automatically held according to the
operation of the test leads. “A•H”
Peak hold: The peak values are held. “P•H” appears on the display.
5) LIGHT key
Use this key to turn the backlight on and off. Press the key once to turn it on. Press it again
to turn it off. Use this key to control the sleep function. During measurement or output, Hold down LIGHT for 2 seconds or longer to put the CA450 in sleep mode. (While the CA450 is in sleep mode, hold down the LIGHT key for 2 seconds or longer to clear sleep mode.) (You cannot clear sleep mode immediately after the CA450 is put in sleep mode. Wait at least 2 seconds after the CA450 is put in sleep mode before you clear it.)
6) RELΔ/% key
Use this key to display the measured values as relative values (you can display the differences between values using numbers or percentages).
1: Difference display (RELΔ) “Δ” appears on the display. 2: Percentage difference display (REL%) “Δ” and “%” appear on the display.
Use this key during the zero adjustment of the resistance measurement function. With the test leads shorted, press the REL key to execute zero adjustment.
7) MIN/MAX key Use this key to display the minimum (MIN), maximum (MAX), and average (AVG) values within the measurement period. When you press the key, recording starts, and “MIN,” “MAX,” and “AVG” appear on the display. (The AUTO OFF function is disabled.) When peak hold is enabled, press the REL
key to reset the peak value.
8) Hz key When the function switch is turned to ~V, use this key to switch between AC voltage measurement and frequency measurement.
Use this key in calibration mode to conrm the calibration value.
9) SHIFT key If you press SHIFT while a measurement mode is selected, “Shift” appears on the display.
In this state, you can congure various settings by pressing the various keys.
SHIFT+ LIGHT key SETUP mode
SELECT key Switches from mV mode to SENSOR mode
While an output mode is selected, press SHIFT to switch between SOURCE and SIMULATE.
appears on the display.
19
Display (LCD) Description
Symbol or Unit Description
Appears during DC measurement or output Appears during AC measurement
Appears when the polarity is negative (minus) Appears during diode test mode
Appears during continuity check mode Appears when relative values are being displayed
Indicates that the measurement range is set to xed
Indicates that the measurement range is set to auto Indicates the present range Appears when data hold is enabled Appears when auto hold is enabled Appears when peak hold is enabled Appears when the MIN, MAX, and AVG values are being measured Appears when the MAX value is being displayed Appears when the MIN value is being displayed Appears when the AVG value is being displayed Appears when user calibration is being performed Appears when the auto power-off function is enabled Appears during rms value detection Appears when SHIFT has been pressed Shows the main value Shows the input or output value A converted value is shown in SENSOR mode
Indicates the main unit
Indicates the subunits Shows the MIN, MAX, and AVG recording time Shows the input voltage unit in SENSOR mode Shows the subvalue Shows the MIN, MAX, and AVG recording time Shows the reference value during relative calculation
20
Appears when the range has been exceeded
WARNING
Appears when the battery voltage is low Appears during current output
Appears during fast linear output
Appears during slow step output Appears during fast step output
Shows the current sweep output or step output status Appears during slow linear output
Appears in loop power supply mode Appears when the internal resistor for HART communication is on
Appears in SPAN check mode
Appears when the SIMULATE operation is being performed Appears in SENSOR mode

4.3 Measuring Instructions

To avoid damaging the instrument and other equipment:
• Before you start measuring, make sure that the position of the function switch and position of the input terminals for connecting the test leads or lead cables are appropriate for the desired mode of measurement.
that they are appropriate for the measurement category.)
(Check
• Before you turn the function switch, remove the test leads or lead cables circuit under measurement.
• Before you use this instrument or use its indications as a reference for further procedures, make sure that the instrument functions properly when used with a known power supply.
from
the
21
4.3.1 AC Voltage Measurement (
Black Red
Test leads
Black Red
Test leads
Note
, )
1) Turn the function switch to the “
2) If you turned the function switch to “mV,” press SELECT. (“~” is displayed.)
3) Insert the test leads into the input terminals.
4) Connect the test leads to the circuit under measurement, and then read the value when it stabilizes.
4.3.2 DC Voltage Measurement ( V, mV)
1) Turn the function switch to the “
2) Insert the test leads into the input terminals.
3) Connect the test leads to the circuit under measurement, and then read the value when it stabilizes.
If the
mV
to a circuit, the instrument will still display a value, but measurement will not be adversely affected.
4.3.3 Measuring Using Sensors (SENSOR)
You can use this function to measure the output of a current sensor or other type of sensor that converts its measured values into voltages, to convert the voltages into the original measured values, and to display the converted values.
Before you perform measurement, you need to congure unit, conversion, and other
See “SENSOR Mode Input, Display, and Unit Settings” in section 4.6, “Setup
settings. Function.”
1) Turn the function switch to the “mV” position.
2) Press SHIFT+SELECT to switch to DC SENSOR mode.
To switch to AC SENSOR mode, press SELECT again. The input voltage appears in the subdisplay. In the main display, the value converted
using the settings congured in “SENSOR Mode Input, Display, and Unit Settings” in
section 4.6 appears.
or
mV
range setting is selected and the test leads are not connected
V” or “mV” position.
V” or “mV” position.
22
DC SENSOR range AC SENSOR range
SENSOR
3) Connect the sensor that you will use to the input terminals.
CAUTION
Black Red
Test leads
4) Read the value when it stabilizes.
To return to normal mV measurement, press SHIFT+SELECT again.
Example of measurement when connecting a current clamp sensor with an input-to-output ratio of 10 mV/A
Use the setup function (see section 4.6) to congure the following settings.
Instrument input voltage: 10.0 (mV) Converted value, decimal place: 01.00 Unit: A
If you connect the sensor after conguring the above settings, the current clamp sensor’
s output voltage (in mV) appears in the subdisplay, and the current detected by the current clamp sensor appears in the main display. (You can read the values directly.)
4.3.4 Resistance Measurement (Ω)
To avoid damaging the instrument:
Before measurement, turn off the circuit under measurement so that no voltage is
applied.
Turn the function switch to the “Ω” position.
1)
2) Insert the test leads into the input terminals.
3) Connect the test leads to the circuit under measurement, and then read the value when it stabilizes.
23
Note
Zero adjustment for resistance
CAUTION
Black Red
Test leads
CAUTION
Black Red
Test leads
We recommend that you perform zero adjustment to obtain accurate measurements. After you perform steps 1 and 2 above, short the test leads, and press the REL key.
(When zero adjustment is performed, the instrument will show a reading of 0.0 Ω.)
The zero adjustment value is used until the power is turned off.
4.3.5 Continuity Check (
To avoid damaging the instrument:
Before measurement, turn off the circuit under measurement so that no voltage is applied.
) Turn the function switch to the “
1
2) Insert the test leads into the input terminals.
3) Connect the test leads to the circuit under measurement whose continuity you want to check.
If the circuit is continuous (less than approximately 50Ω),
the buzzer sounds.
4.3.6 Diode Test ( )
To avoid damaging the instrument:
Before measurement, turn off the circuit under measurement so that no voltage is applied.
Turn the function switch to the “Ω” position.
1)
Press SELECT to switch to diode measurement.
(
is displayed.)
2) Insert the test leads into the input terminals.
3) Connect the test leads to the diode, and then read the value when it stabilizes.
)
” position.
24
Forward-Bias Diode Test
Black Red
Figure 1 Forward-bias test
Test leads
Red Black
Figure 2 Reverse-bias test
Test leads
the black test lead to the cathode, and connect the red test lead to the anode.
Connect For a silicon diode, the reading should be approximately 0.5 V. For an LED, the reading should be approximately 1.5 V to 2.0 V.
Reverse-Bias Diode Test
the black test lead to the anode, and connect the red test lead to the cathode.
Connect Normally, “OL” appears. If a voltage value appears, the diode is defective.
25
4.3.7 DC Current Measurement (
WARNING
30 mA range
Main display
Subdisplay (percentage of span)
When the range is 4 to 20 mA
100 mA range
Main display
Subdisplay (percentage of span)
When the range is 10 to 50 mA
BlackRed
Test leads
To avoid damaging the instrument and other equipment:
• Before
you start measuring, make sure that the position of the function switch and
position of the input terminals for connecting the test leads are appropriate for the desired mode of measurement.
The maximum input current for mA measurement is 110 mA.
)
1) Turn the function switch to the “
2) Connect the black test lead to the COM input terminal, and connect the red test lead to the input terminal.
3) Connect the test leads to the circuit under measurement, and then read the value when it stabilizes.
Press the RANGE key to select a range.
Subdisplay: Displaying and Conguring the Span
In addition to the DC current value in the main display, you can also display the span value as a percentage in the subdisplay.
Subdisplay for the 30 mA Range
Input Current Value
(Main display)
-33.000 mA mA -25.0% 0.0%
The section 4.4.4, “Current Span Setting.”
0.000 .000 mA 0.0% 20.0%
4
mA 100.0% 100.0%
20.000
0.000 mA 162.5% 200.0%
3
.000 mA 181.3% 165.0%
33
displayed value varies depending on the setting (4 to 20 mA or 0 to 20 mA) made in
” position.
Subdisplay for the 30 mA Range
4 to 20 mA 0 to 20 mA
-231.3% -165.0%
26
Subdisplay for the 100 mA Range
Transmitter (signal conditioner)
Distributor
LOOP POWER
Two-wire system transmitter
4 to 20 mA
Red Black
Input Current Value
(Main display)
-110.00 mA -110.0% -300.0% -220.0%
0.00
mA 0.0% -25.0% 0.0%
1
0.00 mA 10.0% 0.0% 20.0%
5
0.00 mA 50.0% 100.0% 100.0%
1
00.00 mA 100.0% 225.0% 200.0%
1
10.00 mA 110.0% 250.0% 220.0%
The
displayed value varies depending on the setting (0-100mA, 10-50mA, or 0-50mA) made in
section 4.6, “Setup Function.”

4.3.8 Loop Power Measurement

You can use this function to measure the current that ows when a constant voltage of 24 VDC
applied.
is 24 V loop measurement can be used for transmitter loop testing. (You can connect the CA450 to the transmitter instead of connecting a transmitter or signal conditioner.)
0 to 100 mA 10 to 50 mA 0 to 50 mA
Subdisplay for the 100 mA Range
27
1) Turn the function switch to the “
Tip
Black Red
Test leads
Tip
(“LOOP POWER” appears on the display.) A constant voltage of 24 VDC is output between the “SOURCE(+)” and “SOURCE(-)”
measurement terminals.
2) Connect the red test lead to “SOURCE(+)” and the black test lead to “SOURCE(-).”
3) Connect the test leads to the circuit under measurement, and then read the values when they stabilize.
Just as in mA measurement, the main display shows the measured value, and the
subdisplay shows the value of the measured value divided by the current span times 100
(%). (See chapter 4.3.7.)
4)
Press SELECT to turn the internal 250 Ω series resistor for HART communication on.
(“250ΩHART” appears on the display.) Press SELECT again to turn the internal resistor
off.
The loop power is supplied at the typical value of 24 VDC. Depending on different
conditions, such as the loop current value or whether there is a internal series resistor, the voltage across the terminals may be greater than 24 V.
4.3.9 Frequency Measurement (Hz)
1) Turn the function switch to the “~V “ position.
2) Press the Hz key to select frequency measurement.
(The measured frequency value and unit appear in the
main display, and the measured voltage appears in the subdisplay.)
3
) Insert the test leads into the input terminals.
4) Connect the test leads to the circuit under measurement, and then read the values when they stabilize.
LOOP POWER” position.
Note that range hold and selection operations performed using the Range key only
apply to the subdisplay (measured voltages).
28

4.3.10 Auto Hold Function

Note
The CA450 can automatically detect measured values and hold them according to the operation of the test leads.
1) During normal measurement, press the HOLD key twice to select “A•H.” (“A•H”
appears on the display.)
2) Connect the test leads to the circuit under measurement.
3) When the measured value stabilizes, the buzzer makes a beeping sound.
4) Remove the test leads from the circuit under measurement.
5) The measured value display is held down.
While “A•H” is displayed, you can repeat steps 2 through 4 as often as you want.
• You cannot use this function during DCmV or ACmV measurement (this includes SENSOR mode), frequency measurement, or during a continuity check.
• You cannot use this function at the same time as the relative value display function or when the MIN, MAX, and AVG values are being recorded.
• Unstable input signals (measured values) cannot be held.

4.3.11 Peak Hold Function

The CA450 always detects and updates the instantaneous peak value during DC voltage measurement (DCV). You can determine the peak value of a waveform.
1) Set the function switch to
2) Connect the test leads to the circuit under measurement.
3) Press the HOLD key three times to select “P•H.” (“P•H” appears on the display, and the voltage range is xed.)
) The peak value display is held.
4
5) You can reset the held peak value by pressing the MIN/MAX key. You can hold a new peak value.
V.
29
Note
• Even if the polarity of the input signal (DCV) is negative (minus), as long as the peak (change) direction is positive, the peak value can be measured.
• While the peak value is being measured, you can display relative values based on a measured reference value.
1) Press the REL Δ /% key while the peak is held. (“Δ” is displayed.)
Main display value: The current peak value – the peak value when the
REL key was pressed
2) Press the REL Δ /% key again. (The unit changes to ‘%’.)
Main display value: (The current peak value – the peak value when the
REL key was pressed) /the peak value when the REL key was pressed × 100 (%)
To no longer show relative values, press REL Δ /% again.
3)
The “Δ” and “%” display disappear, and the CA450 returns to the normal peak
hold mode.
• You can reset the peak value by pressing MIN/MAX.
You can hold a new peak value.
30
4.3.12 Relative Value Display (REL Δ/%)
The CA450 can calculate and display relative values that represent the difference or percentage from a reference value.
(In this display, the range is xed.)
Difference Display (REL Δ)
The
CA450 shows the value of the measured value minus the reference value.
1) Measure
2) Press REL Δ/% to conrm the reference value. (“Δ” is displayed, and the range is held.)
3) Measure (input) the measured values. The displayed values vary depending on the
Percentage Difference Display (REL%)
The CA450 computes the value of (Measured value – reference value)/reference value and
displays
1) Measure (input) the reference value.
2) Press REL Δ/% to conrm the reference value. (“Δ” is displayed, and the range is held.)
3) Press REL Δ /% again. (“%” appears.)
4) Measure (input) the measured values. The displayed values vary depending on the
(input) the reference value.
measurement function.
• For DCmA measurement (including loop power measurement):
Main display value: Measured value – reference value
Subdisplay value: (Measured value – reference value)/current span
setting × 100 (%)
• For measurement types other than DCmA: Main display value: Measured value – reference value Subdisplay value: Reference value
the result as a percentage.
measurement function.
• For DCmA measurement (including loop power measurement): Main display value: (Measured value – reference value)/current span
setting × 100 (%)
Subdisplay value: Reference value
• For measurement types other than DCmA: Main display value: (Measured value – reference value)/reference
value × 100 (%)
Subdisplay value: Reference value
31

4.3.13 MIN/MAX/AVG Function

Note
You can use this function to display the minimum (MIN), maximum (MAX), and average (AVG) values within the current measurement period.
(In this display, the range is xed.)
The
averaged value that is displayed is the total of the recorded data divided by the number of recorded values. When you press the MIN/MAX key, recording starts, and “MIN,” “MAX,” and “AVG” appear on the display. (The AUTO OFF function is disabled.)
Recording Time
The
timer starts, and the time from the start of recording is displayed as well as the MIN/ MAX record update time. The recorded time display shows times in the range of 0 seconds to 99 minutes, 59 seconds in units of seconds. It shows times greater than 100 minutes in units of minutes. To stop recording, press the HOLD key. (“D•H” appears on the display.)
Checking the Recording Time
ou can check the recorded data by pressing the MIN/MAX key to switch between the
Y display of the current maximum (MAX), minimum (MIN), and average (AVG) values. To restart recording, press the HOLD key again. To disable the MIN/MAX/AVG function, hold down MIN/MAX for 1 second or longer. (“MAX,” “MIN,” and “AVG” disappear.)
• While recording is stopped, removing the test leads does not affect the recorded data.
• When an out-of-range input value is recorded, the minimum or maximum value is displayed as “OL,” and the average value data becomes inaccurate.
• When you are measuring a widely uctuating signal, set the range to contain the
maximum and minimum values so that these values are not displayed as “OL.”
32

4.4 Output Instructions

WARNING
CAUTION
The CA450 can output DC current. There are two output modes. SOURCE mode: Current is supplied from the CA450. SIMULATE mode (SINK): The CA450 sinks current from an external voltage
source.
There are two conguration modes. Constant current output: The specied current is output continuously.
Current sweep output: The output current is increased and decreased
within the specied span.
Do not apply a voltage of 48 V or greater to the output terminals. Doing so may lead to electric shock.
Also, keep the voltage between the circuit and the ground below 48 V. Always use the attached lead cables and lead cables.
that they are appropriate for the measurement category.)
(Check
In modes other than SIMULATE mode, do not apply voltages to the output terminals. Mistakenly applying voltage to the output terminals could damage the internal circuitry.
If the batteries.
mark appears while you are using the current output function, replace the
33

4.4.1 Constant Current Output (SOURCE mode)

BlackRed
Lead cables
Tip
1) Turn the function switch to the “
” position.
After you turn the function switch, “OUTPUT” appears on the screen, and the output is
set to 0% of the current span setting.
(If you turn the function switch from the “
” (current sweep) position to the “
position, the output value setting and SOURCE/SIMULATE setting are retained.)
2) If “SIMULATE” is displayed, press the SHIFT key. (“SIMULATE” disappears, the CA450 switches to SOURCE mode, and current output
starts.)
3)
Connect the lead cables to the positive (+) and negative (–) SOURCE terminals.
4) Use the output value setting keys to set the output value.
Output Value Setting Key
*
%STEP
Output Setting (Increase and decrease)
▲ 1-step (25%) increase ▼ 1-step (25%) decrease
COARSE 0.
FINE 0.00
0.
0.00
1 mA increase 1 mA decrease
1 mA increase 1 mA decrease
* For details about using the %STEP key, see “Output
Values for Each Step” and “Span Check Mode.”
5) Connect the lead cables to the circuit under test.
• Because the CA450 outputs current in SOURCE mode, it consumes more current than in other modes. You can keep the current consumption down in SIMULATE mode.
Hold down the %STEP
, COARSE, or FINE key to increase or decrease
each digit consecutively.
Output Values for Each Step
You can press %STEP▲ or %STEP▼ to increase or decrease the output by one step (25%).
ou can check the step value (as a percentage) in the subdisplay. (The output current
Y appears in the main display.)
34
Step
When the range is 4 to 20 mA
Tip
(Subdisplay)
-25% 0.000 0% 4
25% 8.000 50% 1
75% 1 100% 20.000 125% 2
131.2% 2
For the method for setting the current span, see section 4.4.4, “Current Span Setting.”
Span Check Mode
ou can switch the output current between 20 mA and 4 mA (or 20 mA and 0 mA) using
Y
just the %STEP keys.
When you hold down %STEP▲ or %STEP▼ for 1 second or longer, “SPAN” appears, and
the CA450 enters into span check mode.
In this mode, the output switches to 100% when you press %STEP▲ and to 0% when you press %STEP▼.
Setting Key Output Setting
20.000 mA (100%)
%STEP
The COARSE (0.1 mA increase and decrease) and FINE (0.001 mA increase and decrease) keys can also be used.
To exit span check mode, hold down %STEP▲ or %STEP▼ for 1 second or longer again. (“SPAN” disappears.)
4 to 20 mA 0 to 20 mA
.000 mA 0.000 mA
2.000 mA 10.000 mA
6.000 mA 15.000 mA
5.000 mA
4
to 20 mA
4.000 mA (0%)
Output Value
mA
mA 5.000 mA
mA 20.000 mA
5.000 mA
to 20 mA
0
0.000 mA (0%)
• If the set current is 0.1 mA or more and the terminals are open, “-----” appears in the main display, and “----” appears in the subdisplay.
35

4.4.2 Constant Current Output (SIMULATE mode)

CAUTION
Transmitter (signal conditioner)
Distributor
SIMULATE
(SINK)
Two-wire system transmitter
4 to 20 mA
Black
Red
When the SIMULATE (SINK) function is enabled, the CA450 can sink the specied amount of current from an external voltage source (such as a distributor) through the SIMULATE(+)
terminal,
and you can use the CA450 to simulate a two-wire system transmitter in a loop test.
(You can connect the CA450 instead of a transmitter to test a transmitter or signal conditioner.)
• When sinking 20 mA from an external power source, keep the voltage within the range of 15 to 48 V.
Set the polarity of the applied voltage as shown in the gure below, and be careful not
to apply voltage in the opposite direction.
1) Turn the function switch to the “ ” position.
After you turn the function switch, “OUTPUT” appears on the screen, and the output is
set to 0% of the current span setting.
(If you turn the function switch from the “
2) Press the SHIFT key. “SIMULATE” appears. (Current sinking starts.)
output value setting and SOURCE/SIMULATE setting are retained.)
” position to the “ ” position, the
36
3) Connect the lead cables to the positive (+) and negative (–) SIMULATE terminals.
Tip
Tip
(Connect the positive SIMULATE terminal to the positive terminal of the external DCV
power source. Make sure that the power source voltage is within 15 to 48 V.)
4) Just as in SOURCE mode, use the output setting keys to set the sink current value.
(To enter span check mode, hold down %STEP▲ or %STEP▼ for one second or
longer.)
5
) Connect the lead cables to the circuit under test.
Press the SHIFT key again. “SIMULATE” disappears, and the CA450 returns to constant
current output (SOURCE) mode.
• If the set current is 0.1 mA or more and the terminals are open, “-----” appears in the main display, and “----” appears in the subdisplay.

4.4.3 Current Sweep Output

You can make the output current (in SOURCE or SIMULATE mode) automatically uctuate between 0% and 100% of the current span setting.
1) Turn the function switch to the “
After you turn the function switch, “OUTPUT” appears on the screen, and the output
starts rising from 0% of the current span setting.
(If you turn the function switch from the “
output value setting and SOURCE/SIMULATE setting are retained.)
2) Press the SHIFT key to select the current output mode (SOURCE or SIMULATE).
3) Press the SELECT key to select an output mode.
(When you change the output mode, the current starts rising from the value currently
being output according to the new mode.)
4) After you connect the lead cables to the appropriate terminals for the output mode (SOURCE or SIMULATE), connect them to the circuit under test.
• If you want to stop a current sweep that is being output, turn the function switch to “ being output when the switch is turned.)
.” (The constant current output value is set to the value of the current
” position.
” position to the “ ” position, the
37
Operation Mode Operation
Output current
Span
(%)
100
0
Time (s)
4015
SlowFast
Output current
Span
(%)
100
0
Slow: 15 to 60 s* Fast: 5 s
Time (s)
Tip
Slow
Fast
Slow
Fast
Linear Mode Current Output
Step Mode Current Output
0% to 100% to 0% in 40 seconds
0% to 100% to 0% in 15 seconds
0% to 100% to 0% in 25% steps
step is 15 to 60 seconds long.*
Each
0% to 100% to 0% in 25% steps
step is 5 seconds long.
Each
* In slow step mode, you can select 15, 30, 45, or 60 seconds. See section 4.6.4, “Setting the Slow Step Time.”
38

4.4.4 Current Span Setting

You can select one of the following two current spans. The following two percentage displays vary depending on the selected range: the output current value and the measured current value for mA measurement mode in the 30 mA range.
Current Span Setting
to 20 mA 4.000 mA 20.000 mA Factory default
4 0 to 20 mA 0.000 mA 20.000 mA
You can use the following method to set (change) the current span.
1) Turn the function switch to the “OFF” position, or put the CA450 into sleep mode.
2) While holding down the RANGE key, turn the function switch to any position. If or longer.
3) After all the LCD elements appear and then disappear, release the key.
4) After setting the current span, turn the function switch to the "OFF" position, or put the CA450 into sleep mode. If the current span was set to 4 to 20 mA, this operation changes it to 0 to 20 mA. If the current span was set to 0 to 20 mA, this operation sets it to 4 to 20 mA. The setting you make remains valid until you change it.
Output Current and Measured Current (30 mA range) Notes
When the display is 0% When the display is 100%
the CA450 is in sleep mode, hold down the RANGE and LIGHT keys for 2 seconds

4.5 Auto Power-Off (Sleep) Function

When the Auto Power-Off Function Is Enabled
“AUTO OFF” appears.
• The CA450 has an auto power-off function that automatically switches itself to sleep mode 20 minutes after the last key operation.
The buzzer makes a beeping sound approximately 30 seconds before the CA450 enters
sleep mode. (This period is referred to as the warning period.)
• If you press a key or switch during the warning period, the time before auto power-off is extended. You can clear sleep mode by turning the function switch to the OFF position, waiting 1
• second or longer, and then turning the switch to any position other than OFF. (You can also hold down the LIGHT key for 2 seconds or longer to clear sleep mode.
When the Auto Power-Off Function Is Disabled
1
) Turn the function switch to the “OFF” position, or put the CA450 into sleep mode.
2) While holding down HOLD, turn the function switch to any position. the CA450 is in sleep mode, hold down the HOLD and LIGHT keys for 2 seconds or
If longer. The auto power-off function is disabled, and “AUTO OFF” disappears.
39
Restoring the Auto Power-Off Function
Note
the function switch is set to OFF or when the CA450 enters sleep mode, the auto
When power-off function is restored. (When the power is turned on or when sleep mode is cleared, “AUTO OFF” appears.)

4.6 Setup Function

You can use the setup function to:
1 Turn the buzzer on and off.
2 Congure SENSOR mode input, display, and unit settings.
3 Set the current range of the 100 mA range for DCmA measurement. 4 Set the slow step time. 5 Return the settings to their factory default values.
1) Press the SHIFT key. “Shift” appears on the display.
2) While “Shift” is displayed, press the LIGHT key to switch to setup mode. After “SEtUP”
appears, the screen for turning the buzzer on and off appears.
3) The setting to congure changes each time you press the LIGHT or SHIFT key.
4) To change a setting, use the ▲ (RANGE) and ▼ (REL) keys.
5) To save (conrm) a setting, press the HOLD key. (“SEt” appears, the setting is saved,
and the setting item is redisplayed.)
6) To return to normal measurement from setup mode, hold down the LIGHT key for 1
second or longer.
• To cancel a setting, press the LIGHT key for 1 second or longer without pressing the HOLD key, or use the function switch to turn the power off.
• You cannot enter setup mode from one of the current output modes (constant current output or current sweep output mode).

4.6.1 Turning the Buzzer On and Off

You can turn the buzzer (beeping) sound on and off. Even if you turn the buzzer off, it will sound when it is performing important functions such as those listed below.
• Continuity check buzzer
• Excess input alarm buzzer
• Auto power-off warning buzzer
40
1) Press the LIGHT or SHIFT key to display “bEEP.”
HOLD
:ON :OFF
Main display 1
Subdisplay 1Subdisplay 4
Main unit
Main decimal place
Main display 4
(Turn the buzzer on and off in the subdisplay. The initial setting is on.)
2) Use the ▲ (RANGE) and ▼ (REL) keys to select ON or OFF.
3) Press the HOLD key to save the setting.
(After “SEt” appears, “bEEP” will appear again.)
4) To return to normal measurement from setup mode, hold down the LIGHT key for 1 second or longer.

4.6.2 SENSOR Mode Input, Display, and Unit Settings

For SENSOR mode (mV position on the function switch), you can set the input voltage and the corresponding main display and unit for that input voltage.
1) Press the LIGHT or SHIFT key to display the screen shown below. The parameter that
can be changed blinks.
Keep pressing the LIGHT key to switch between the following parameters. Subdisplay 4, subdisplay 3, subdisplay 2, subdisplay 1, main display 4, main display 3,
(You can press the SHIFT key to switch parameters in the opposite direction.)
2) Use the ▲ (RANGE) and ▼ (REL) keys to set the various values, the decimal place,
3) Press the HOLD key to save the settings. (After “SEt” appears, the setting screen will
4) To return to normal measurement from setup mode, hold down the LIGHT key for 1
main display 2, main display 1, main decimal place, and main unit
and the unit.
appear again.)
second or longer.
41
Settings
0
1 A
10 mV
66 A
660 mV
Display
Input
HOLD
Subdisplay value: 000.0 to 999.9 Main display value: 0000 to 9999 Main display decimal place: XXXX, X.XXX, XX.XX, or XXX.X
Main display unit: A, mA, μA, °C, MΩ, kΩ, Ω, kHz, Hz, μF, nF, %, lx, no unit, V, or mV
Example For a current clamp sensor with an input-to-output ratio of 10 mV/A, set the subdisplay to 10.0 mV, and set the main display (value, decimal place, and unit) to 01.00 A.
With these settings, a maximum input of 660.0 mV will be displayed as
66.00 A.

4.6.3 Current Span Setting for the 100 mA Range for DCmA Measurement

When you select the 100 mA range in DCmA measurement mode (
), you can change the
current span (0 to 100 mA, 10 to 50 mA, or 0 to 50 mA).
Input Current Value
(Main display)
0 to 100 mA 10 to 50 mA 0 to 50 mA
Subdisplay for the 100 mA Range
-110.00 mA -110.0% -300.0% -220.0%
0.00 mA 0.0% -25.0% 0.0%
10.00 mA 10.0% 0.0% 20.0%
50.00 mA 50.0% 100.0% 100.0%
100.00 mA 100.0% 225.0% 200.0%
110.00 mA 110.0% 250.0% 220.0%
1) Press the LIGHT or SHIFT key to display “SPAn.”
(The selected current span appears in the subdisplay. The default span is 0 to 100 mA.)
2) Use ▲ (RANGE) and ▼ (REL) to select 0 to 100 mA, 10 to 50 mA, or 0 to 50 mA.
3) Press the HOLD key to save the setting. (After “SEt” appears, “SPAn” will appear
again.)
4) To return to normal measurement from setup mode, hold down the LIGHT key for 1 second or longer.
42

4.6.4 Setting the Slow Step Time

HOLD
HOLD
If you select slow step mode for current sweep output, you can select the hold time (in seconds) of each step.
1) Press the LIGHT or SHIFT key to display “t.Int.”
2) Use ▲ (RANGE) and ▼ (REL) to select 15s, 30s, 45s, or 60s.
3) Press the HOLD key to save the setting. (After “SEt” appears, “t.lnt” will appear again.)
4) To return to normal measurement from setup mode, hold down the LIGHT key for 1 second or longer.

4.6.5 Returning the Settings to Their Factory Default Values

You can reset all the items that you have changed (set) in setup mode to their factory default values.
Press the LIGHT or SHIFT key to display the screen for initialization, which shows
1)
“dEF.”
2) The settings return to their default values when you press the HOLD key. (After “donE” appears, “dEF” will appear again.)
3) To return to normal measurement from setup mode, hold down the LIGHT key for 1 second or longer.
Initial Settings (Default Values) for the Setting Items
Setting Item Initial Setting Buzzer sound ON SENSOR
mode Subdisplay 1
0.0 Main display 01.00 Unit A
current span of the 100 mA range for
The
0 to 100 mA DCmA measurement Slow step time setting 15 s
43

4.7 Additional Power-on Functions

Hold one of the keys below, and turn the function switch from OFF to any position to
congure various functions. (If the CA450 is in sleep mode, hold down one of the keys below and the LIGHT key for 2 seconds or longer to congure various functions.)
Key Congured Function Details
RANGE Current
SELECT All
HOLD The
HOLD + RELΔ/% The SELECT + RANGE Calibration

4.7.1 Making All the LCD Elements Appear

If you hold down the SELECT key and turn the function switch from OFF to any position, all the LCD segments and marks are displayed. The LCD segments and marks are only displayed while the SELECT key is held. (If the CA450 is in sleep mode, hold down the SELECT and LIGHT keys for 2 seconds or longer to turn the entire LCD on.)
output and measurement (30 mA range) span
(0 to 20 mA, 4 to 20 mA)
the LCD elements appear (only while the SELECT
key is pressed).
auto power-off function is disabled. Section 4.5 calibration data is returned to its factory default.
function Chapter 5
Section
4.4.4 Section
4.7.1
44

5. User Calibration Function

CAUTION
Subdisplay Calibration point
Main display Measured value
Calibration mode
To make sure that the CA450 is highly accurate, we recommend that you calibrate it regularly. You can use the user calibration function to perform calibration.
To avoid electrical shock:
• Make sure that specialized technicians calibrate the instrument using the proper equipment.
To connect the CA450 to a signal generator (reference device), use the test leads and
• lead cables that come with the reference device.
• Be sure to read the reference device’s instruction manual before you calibrate the CA450.
• When you switch measurement modes during calibration, be sure to remove the test
leads and lead cables rst, and then change the mode and terminal locations.
Calibration
Reference device: Use a reference device that fully satises the accuracy of the CA450.
DC voltage and current generator or digital multimeter (DMM)
Environment: Temperature: 23 ± 5°C, Humidity: 45 to 75% RH (no condensation)
Before you perform calibration, leave the CA450 in the environment

5.1 Calibrating Measurements

Conditions
described above for 30 minutes or more.
Perform calibration according to table 1.
1) While holding down the SELECT and RANGE keys, turn the function switch from the OFF position to the ACV position. The CA450 will start in “calibration mode.” (The “CAL” segment will be visible.) If the CA450 is in sleep mode, hold down the SELECT, RANGE, and LIGHT keys for 2 seconds or longer to start the CA450 in calibration mode.
2) Use the test leads to connect the CA450 to the reference device.
3) From the reference device, apply the values in Table 1.
The calibration point appears in the subdisplay.
45
4) Press the Hz key to conrm the value.
Note
Press the calibration conrmation key after the reference device’s value stabilizes.
5) To calibrate other ranges, press the RANGE key to switch the range, check the function switch and the input terminals, refer to table 1, and perform steps 3 and 4 again.
To perform peak hold and loop power calibration, see section 5.1.1, “Peak Hold
Calibration,” and section 5.1.2, “LOOP POWER calibration.”
6)
To end calibration, turn the function switch off.
You cannot calibrate the continuity check. Even in calibration mode, normal measurement is performed. (Keys other than RANGE and SELECT cannot be used.)
Table 1 Calibration Table (Measurements)
Function Range Calibration Point (Input value)
AC voltage measurement
DC voltage measurement
DC voltage measurement
6.000 V 6.000 V 50.00/60.00 Hz
60.00 V 60.00 V 50.00/60.00 Hz
600.0 V 600.0 V 50.00/60.00 Hz 1000 V 1000 V 50.00/60.00 Hz*1
6.000 V 6.000 V
60.00 V 60.00 V
600.0 V 600.0 V 1000 V 1000 V 600 mV 600.0 mV
AC voltage measurement
Diode test
Resistance measurement
DC
voltage measurement
*1 If the power supply frequency is 50 Hz, calibrate using 60.00 Hz. If the power supply
frequency
is 60 Hz, calibrate using 50.00 Hz.
600 mV 600.0 mV
2 V 2.000 V
600.0 Ω 600.0 Ω
6.000 kΩ 6.000 kΩ
60.00 kΩ 60.00 kΩ
600.0 kΩ 600.0 kΩ
6.000 MΩ 6.000 MΩ
60.00 MΩ 40.00 MΩ
30.000 mADC 30.000 mA
100.00 mADC 100.00 mA
46

5.1.1 Peak Hold Calibration

1) In calibration mode (when the “CAL” segment is displayed), turn the function switch to
V position.
the
2) Press the HOLD key to enable peak hold mode.
3) Use the test leads to connect the CA450 to the reference device.
4) From the reference device, apply the value under ZERO in Table 2.
The calibration point appears in the subdisplay. You can clear the peak value by pressing the MIN/MAX key.
5) Press the Hz key to conrm the value.
6) From the reference device, apply the value under Full Scale in Table 2.
The calibration point appears in the subdisplay. You can clear the peak value by pressing the MIN/MAX key.
7) Press the Hz key to conrm the value.
8) To end calibration, turn the function switch off.
Table 2 Calibration Table (Peak Hold)
Function Range
DCV peak
hold
5.1.2

Loop Power Calibration

1) In calibration mode (when the “CAL” segment is displayed), turn the function switch to
the LOOP POWER position.
2)
Connect the CA450 to the reference device as shown in the diagram.
3) Set the reference device (the sink) to -30.000 mA.
The calibration point appears in the subdisplay.
4) Adjust the reference device, which was set to -30.000 mA in the previous step, so that the DMM display value (measured value) is 3.0000 V.
5) Press the Hz key to conrm the value.
6) To end calibration, turn the function switch off.
6.000 V 0 V 6.000 V
Calibration Point (Input value)
ZERO +Full Scale
Condition
Table 3 Calibration Table (Loop Power)
Function Range Calibration Point DMM Display (Measured)
Value
LOOP POWER 3
0.000 mADC 30.000 mA 3.0000 V
47
Connection Diagram (Loop power)
LOOP POWER
LOOP POWER
Black
Red
Digital multimeter
(DMM)
H (V)
L
DC voltage and current generator (Sink function)
Load
510 Ω/1 W
100 Ω
3.0000 V
H
L
-30.000 mA
Subdisplay Calibration point
Main display Measured value
Calibration mode
Subdisplay Fourth decimal place
Main display Up to the third decimal place
Fourth decimal placeSecond decimal place

5.2 Calibrating Output

Perform calibration according to Table 4.
1) While holding down the SELECT and RANGE keys, turn the function switch from the OFF position to the
2) Connect the CA450 to the reference device as shown in the diagram.
3) Start +5% calibration:
In the main display and the subdisplay, 0.0000 mA appears. (The display in the
subdisplay shows the value up to the fourth decimal place.)
position. (The “CAL” segment will be visible.)
48
4) Use %STEP▲, %STEP▼, COARSE▲, and COARSE▼ to adjust the value that
Note
appears in the main display and the subdisplay so that it is the same as the value read by the multimeter connected to the CA450.
You can use %STEP▲ and %STEP▼ to adjust the second decimal place (values above
the second decimal place are also affected), and you can use COARSE▲ and COARSE
▼ to adjust the fourth decimal place (the third decimal place is also affected).
5)
After you have nished adjusting the value, press the Hz key to conrm it.
6) Start full scale calibration:
In the main display and the subdisplay, 20.0000 mA appears. (The display in the
subdisplay shows the value up to the fourth decimal place.)
7) Use %STEP▲, %STEP▼, COARSE▲, and COARSE▼ to adjust the value that
appears in the main display and the subdisplay so that it is the same as the value read by the multimeter connected to the CA450.
You can use %STEP▲ and %STEP▼ to adjust the second decimal place, and you can
use COARSE▲ and COARSE▼ to adjust the fourth decimal place.
8) After you have nished adjusting the value, press the Hz key to conrm it.
The calibrated range value is computed and saved in the internal memory. (The previous calibration data is overwritten.)
9) To stop calibration, turn the function switch off.
10) Perform steps 2 through 7 for the SOURCE and SIMULATE output modes.
You cannot calibrate current sweep output. Even in calibration mode, normal output is performed.
(Keys other than SHIFT and SELECT cannot be used.)
Table 4 Calibration Table (Output)
Function Output Direction
Constant current output
SOURCE 1 SIMULATE 1
* Adjust at 5% of the range.
Calibration Point
*
±5%
±Full Scale
Condition
.0000 mA 20.0000 mA — .0000 mA 20.0000 mA 28 V external
voltage
49
Connection Diagram (SOURCE, SIMULATE)
SIMULATE
(SINK)
Black
Red
Digital multimeter
(DMM)
SOURCE
Red Black
I (A)
L
Digital multimeter
(DMM)
I (A)
L
H
L
DC voltage and current generator

5.3 What to Check after Calibration

After you have nished calibrating the CA450, make sure (through inspection) that the
calibration has been performed correctly and that the calibration values have been saved to the memory.
Inspection Method
After calibration has nished, turn off the CA450.
turn on the CA450, and check the calibration in the normal output and measurement
Then, modes (not calibration mode). Use the reference device that you used for calibration to check the calibration. (Refer to the calibration points.)
50

6. Communication

Note
You can congure the CA450 and check its settings, measured values, and other information from a PC.
Through the use of a communication adapter, the CA450 can communicate with a PC
through USB.
6.1 Cable Connection and Interface Specications
Cable Connection
Remove the blank cover on the back of the CA450, and connect a USB adapter.
Conguring the RS232C Interface
Baudrate: 9600 bps
Parity: None
bits: 2 bits
Stop Data length: 8 bits Flow control: None Terminator: CrLf only
51

6.2 Communication Commands

Item
Data
acquisition Measurement
settings
Output
settings
Command
OD
MF MR
HD
SM DI
IO
MH RP MI MC
CC
RG
SF SR SD
AS
UP DW SP
UQ
DQ
RA SS
Outputs measured values
Queries the measurement function Sets or queries the measurement range Sets or queries the measured-value-display
hold function Sets or queries SENSOR mode Switches or queries diode test mode Sets or queries whether the internal resistor for
HART communication is on or off Sets or queries frequency measurement
Sets or queries the RELΔ/% display Sets or queries the MIN/MAX/AVG function Switches or queries the MIN/MAX/AVG function
Sets or queries whether the measurement
mode is DCmV or ACmV (DC sensor or AC
sensor in SENSOR mode) Sets or queries the auto range and range hold
settings Queries the output function Sets or queries the output range (span) Sets or queries the output value Sets or queries whether the current output
function is set to SOURCE or SIMULATE mode Increases the m
th
Decreases the m Sets or queries SPAN check mode Increases the output value by one step or sets
the output value to 100%
Decreases the output value by one step or sets
the output value to 0%
Sets or queries the current sweep output Sets or queries the slow step time
Description
digit of the output value by 1
th
digit of the output value by 1
Normal
Î 
Î 
Î
Î  Î  Î  Î
Î
Î    Î
Î
Î  Î  Î
Î
Î
Cal.
52
Item
Calibration
Other
commands
Command
SY
CL CP
CMF
CS
CSF
CR CD CW OE OS
ESC C /
RC
ESC S
H
IM
BL
HC BZ
MP
SI
ST
SL
SU
Description
Sets or queries whether the mode is normal
mode or calibration mode Queries the calibration item Sets the calibration point Queries the measurement mode being
calibrated Confirms the measurement calibration value Queries the output mode being calibrated Sets the output calibration value Confirms the output calibration value Saves the calibration value Outputs error data Outputs setup data Initializes setup data
Outputs the status byte Sets or queries whether a header is attached to
the output data (OD) Sets or queries whether the bits in the status
byte are detected or masked Sets or queries whether the backlight is on or
off Resets the peak hold value Sets or queries whether the buzzer is on or off Sets or queries the span setting for the 100 mA
range for DCmA measurement Sets or queries the input voltage value for
SENSOR mode Sets or queries the converted value for
SENSOR mode Sets or queries the decimal place for SENSOR
mode Sets or queries the unit for SENSOR mode
Normal
Î  Î
Î
Î  Î  Î  Î  Î     Î
Î
Î
Î  Î
Î
Î
Î
Î
Î
Cal.
53

6.3 Detailed Command Descriptions

Conguration and Control
Command: The format for sending the command.
Answer: The format of replies to commands, such as conguration and control
commands, that are not queries for information. (Some commands do not receive any reply.) When an error occurs, the same error message that appears on the LCD (“ERRm,” where m is the error number) is returned.
Queries
Command: Return: The format of replies to commands that are queries for information.
Mode Conditions
Normal: Cal.: The command can be used during user calibration (see chapter 5).
The other conditions for each command are listed in section 6.5, “Communication Command Validity Table.”
AS
The format for sending the command.
The command can be used during normal measurement and output.
Sets or queries whether the current output function is set to
SOURCE or SIMULATE mode
Command = ASm<CRLF> -> Answer = ASm<CRLF>
Command = AS?<CRLF> -> Return = ASm<CRLF>
Parameters
m =
0: SOURCE (default)
1: SIMULATE (SINK)
Normal Cal.
Î
BL
This command returns ERR13 when the CA450 is in constant
current output mode or current sweep output mode.
Sets or queries whether the backlight is on or off Normal Cal.
Command = BLm<CRLF> -> Answer = BLm<CRLF>
Command = BL?<CRLF> -> Return = BLm<CRLF>
Parameters
m =
0: Off (default)
1: On
54
Î
BZ
Sets or queries whether the buzzer is on or off Normal Cal.
Command = BZm<CRLF> -> Answer = BZm<CRLF>
Command = BZ?<CRLF> -> Return = BZm<CRLF>
Parameters
m =
0: OFF
1: On (default)
Î
CC
CD
CL
Sets or queries whether the measurement mode is DCmV or
ACmV
Command = CCm<CRLF> -> Answer = CCm<CRLF>
Command = CC?<CRLF> -> Return = CCm<CRLF>
Parameters
0: DCmV measurement (default)
m =
1: ACmV measurement
In SENSOR mode
m = 0: DC sensor measurement (default)
1: AC sensor measurement
If the measurement mode is neither DCmV nor ACmV, ERR13
is returned.
Confirms the output calibration value Normal Cal.
Command = CD<CRLF> -> Answer = CD<CRLF>
In output calibration mode, the current output settings are
used as the output calibration values for the currently selected
function, range, and calibration point (+FS, +5%, -FS, or -5%).
If the CA450 is in a measurement mode, ERR13 is returned.
Queries the calibration item Normal Cal.
Command = CL?<CRLF> -> Return = CLm<CRLF>
Normal Cal.
Î
Î
Parameters
3: Output calibration
m =
4: Measurement calibration
55
CMF
Queries the measurement mode being calibrated Normal Cal.
Command = CMF?<CRLF> -> Return = CMFm<CRLF>
Parameters
m =
0: DCV
2: Ω
5: Hz
7: ACV
8: DCmV
9: ACmV
10: Diode test
11: Continuity check
12: DCmA
13: LOOP POWER
If the CA450 is in constant current output, current sweep
output, or continuity check function, ERR13 is returned.
Î
CP
Sets the calibration point Normal Cal.
Command = CPm<CRLF> -> Answer = CPm<CRLF>
Parameters
0: +FS calibration
m =
1: Peak hold zero calibration or +5% output
calibration
2: –FS calibration
3: –5% calibration
–FS calibration and –5% calibration are used to calibrate the
SIMULATE output mode.
If –FS calibration or –5% calibration is specified in
measurement mode, ERR13 is returned.
56
Î
CR
Sets the output calibration value Normal Cal.
Command = CRm<CRLF> -> Answer = CRm<CRLF>
Command = CR?<CRLF> -> Return = CRm<CRLF>
Parameters
m =
18.0000 to 22.0000 for ±FS calibration
m = 0.9600 to 1.0400 for ±5% calibration
Î
CS
CSF
CW
Confirms the measurement calibration value Normal Cal.
Command = CS<CRLF> -> Answer = CS<CRLF>
In measurement calibration mode, the current measured values
are used as the measurement calibration values for the currently
selected function, range, and calibration point (+FS or +0).
If the CA450 is in an output mode, ERR13 is returned.
Queries the output mode being calibrated Normal Cal.
Command = CSF?<CRLF> -> Return = CSFm<CRLF>
Parameters
14: Constant current output
m =
15: Current sweep output
If the CA450 is in a measurement mode, ERR13 is returned.
Saves the calibration value Normal Cal.
Command = CW<CRLF> -> Answer = CW,OK<CRLF>
(normal completion)
Saves the data after a mode or range is calibrated. If you turn
off the CA450 without executing this command, the calibrations
made immediately before are discarded.
Î
Î
Î
57
DI
Switches or queries diode test mode Normal Cal.
Command = DIm<CRLF> -> Answer = DIm<CRLF>
Command = DI?<CRLF> -> Return = DIm<CRLF>
Parameters
m =
0: Resistance measurement (default)
1: Diode test
If the mode is neither Ω nor diode test, ERR13 is returned.
DQ
DW
ESC C/RC
Decreases the output value by one step or sets the output
value to 0%
Command = DQ<CRLF> -> Answer = DQ,OK<CRLF>
Decreases the output value by one step (25%).
Sets the output value to 0% of the span in span check mode.
If the CA450 is in any mode other than constant current output
mode, ERR13 is returned.
th
Decreases the m
digit of the output value by 1 Normal Cal.
Command = DWm<CRLF> -> Answer = DW,OK<CRLF>
Parameters
m = 1 to 5 (1 is the least significant digit while 5 is the most
significant digit)
If the CA450 is in any function other than constant current
output function, ERR13 is returned.
If the CA450 is in span check mode, ERR13 is returned.
Initializes setup data
"ESC" = ASCII 0x1B
(
)
Command = ESC C<CRLF>
or
Command = RC<CRLF>
Normal Cal.
Î
Î
Normal Cal.
Î
However, when this command is used to initialize the settings:
The output current span setting (0 to 20 mA or 4 to 20 mA) is
not initialized.
58
ESC S
Outputs the status byte
"ESC" = ASCII 0x1B
(
Command = ESC S<CRLF> -> Answer = m<CRLF>
Outputs the status byte (as a decimal number).
See section 6.6, “Status Byte Format.”
)
Normal Cal.
H
HC
Sets or queries whether a header is attached to the output data
(OD)
Command = Hm<CRLF> -> Answer = Hm<CRLF>
Command = H?<CRLF> -> Return = Hm<CRLF>
Parameters
m =
0: Î header (default)
1: Header
The changed setting is valid until the power is turned off.
Resets the peak hold value Normal Cal.
Command = HC<CRLF> -> Answer = HC,OK<CRLF>
When peak hold is not enabled, ERR13 is returned.
Normal Cal.
Î
59
HD
* During calibration, only the HD3 command, which switches the mode so that
Sets or queries the measured-value-display hold function Normal Cal.
Command = HDm<CRLF> -> Answer = HDm<CRLF>
Command = HD?<CRLF> -> Return = HDm<CRLF>
Parameters
m =
0: Display update enabled (default)
1: Data hold
2: Auto hold
3: Peak hold
• If the CA450 is in an output mode, ERR13 is returned.
• During MIN/MAX measurement, if HD1 is sent, recording and
the updating of the display of the elapsed time are stopped. If
HD0 is sent, recording and the updating of the display of the
elapsed time are resumed.
• If HD1 is sent while peak hold is enabled, ERR13 is returned.
• To clear peak hold, use the HC command.
peak hold calibration can be performed, is valid.
*
60
IM
Sets or queries whether the bits in the status byte are detected
or masked
Command = IMm<CRLF> -> Answer = IMm<CRLF>
Command = IM?<CRLF> -> Return = IMm<CRLF>
Sets whether each of the bits in the status byte is detected or
masked. (Querying is performed using the ESC S command.
See section 6.6.)
If you specify IM0, all the bits are masked. If you specify IM63,
all the data bits reflect the current status.
Parameter
m = 0 to 63
1:
bit0 (measurement completion) is detected
2: bit1 (output adjustment completion) is detected
4: bit2 (syntax error) is detected
8: bit3 (overrange) is detected
16: bit4 (24 V loop output error) is detected
32: bit5 (output error) is detected
(Bits 6 and 7 of the status byte are fixed bits.)
The default is m = 63 (no masking)
Normal Cal.
IO
Sets or queries whether the internal resistor for HART
communication is on or off
Command = IOm<CRLF> -> Answer = IOm<CRLF>
Command = IO?<CRLF> -> Return = IOm<CRLF>
Parameters
0: 250 Ω resistor off (default)
m =
1: 250 Ω resistor on
If the CA450 is in any mode other than loop power
measurement mode, ERR13 is returned.
61
Normal Cal.
Î
MC
Switches or queries the MIN/MAX/AVG function Normal Cal.
Command = MCm<CRLF> -> Answer = MCm<CRLF>
Command = MC?<CRLF> -> Return = MCm<CRLF>
Parameters
m =
0: Measured value display (default)
1: MAX value display
2: MIN value display
3: AVG value display
If MIN/MAX/AVG recording is disabled, ERR13 is returned.
Î
MF
MH
Queries the measurement function Normal Cal.
Command = MF?<CRLF> -> Return = MFm<CRLF>
Parameters
0: DCV
m =
2: Ω
5: Hz
7: ACV
8: DCmV
9: ACmV
10: Diode test
11: Continuity check
12: DCmA
13: LOOP POWER
Switches or queries frequency measurement Normal Cal.
Command = MHm<CRLF> -> Answer = MHm<CRLF>
Command = MH?<CRLF> -> Return = MHm<CRLF>
Parameters
m = 0: ACV measurement (default)
1: Frequency measurement
If the CA450 is in any mode other than ACV mode, ERR13 is
returned.
Î
Î
62
MI
Sets or queries the MIN/MAX/AVG function Normal Cal.
Command = MIm<CRLF> -> Answer = MIm<CRLF>
Command = MI?<CRLF> -> Return = MIm<CRLF>
Parameters
m =
0: MIN/MAX/AVG recording is disabled (default)
1: MIN/MAX/AVG recording starts
If the CA450 is in an output mode, ERR13 is returned.
Î
MP
Sets or queries the span setting for the 100 mA range for
DCmA measurement
Command = MPm<CRLF> -> Answer = MPm<CRLF>
Command = MP?<CRLF> -> Return = MPm<CRLF>
Parameters
0: 0 to 100 mA (default)
m =
1: 10 to 50 mA
2: 0 to 50 mA
If the CA450 is in an output mode, ERR13 is returned.
The changed setting is valid until it is changed again.
Normal Cal.
Î
63
MR
Sets or queries the measurement range Normal Cal.
Command = MRm<CRLF> -> Answer = MRm<CRLF>
Command = MR?<CRLF> -> Return = MRm<CRLF>
Parameters
[DCV]
m = 1: 6 V
3: 60 V
4: 600 V
5: 1000 V
[DCA] m = 0: 30 mA
1: 100 mA
[ACV] m = 0: 6 V
1: 60 V
2: 600 V
3: 1000 V
[Ω] m = 0: 600 Ω
1: 6 kΩ
2: 60 kΩ
3: 600 kΩ
4: 6 MΩ
5: 60 MΩ
[DCmV] m = 0: 600 mV
[ACmV] m = 0: 600 mV
[Diode test] m = 0: 2 V
[Continuity check] m = 0: 600 Ω
[LOOP POWER] m = 0: 30 mA
64
OD
Outputs measured values Normal Cal.
Command = OD<CRLF> -> Return = abbcsddddddeee<CRLF>
Parameters
Header (4 bytes)*
a = V: voltage, A: current, O: resistance, T: temperature,
F: frequency, S: SENSOR mode
bb = DC: direct current, AC: alternating current,
R3: resistance measurement
c = N: normal, O: overrange, E: no data, B: burnout
* The header will not appear if you specify 0 (no header) in
the H command.
Data Area (10 bytes)
s (sign) = _: plus or –: minus (_ represents a space)
dddddd = measured value (6 digits)
eee = E+0, E+3, E–3, E+6, E–6
(For overrange, no data, and burnout, ddddddeee =
_99999.E+6)
OE
Outputs error data Normal Cal.
Command = OE<CRLF> -> Return = ERRm<CRLF>
Outputs the most recent output error.
After a value is returned, the saved error number is overwritten
with ‘ERR00<CRLF>’.
Even if there is no error, ‘ERR00<CRLF>’ is returned.
Parameters
m = Error code
See section 6.4, “List of Errors.”
65
OS
Outputs setup data Normal Cal.
Command = OS<CRLF> -> Return = Measure a<CRLF>
Function b<CRLF>
Range c<CRLF>
Source d<CRLF>
Function e<CRLF>
Range f<CRLF>
Data g<CRLF>
24V Output h<CRLF>
Light i<CRLF>
Charge j<CRLF>
Parameters
a (measurement) = ON or OFF
b (measurement function) =
c (measurement range) = 1000V, 600V, 60V, 6V for
600 mV for DCmV
600 mV for ACmV
60 MOHM, 6MOHM, 600kOHM,
60kOHM, 6kOHM, 600OHM for OHM
2 V for diode test
600OHM for continuity check
100 mA or 30 mA for DCmA
30 mA for loop power
d (output) = ON or OFF
e (output function) = DCmA
f (output range) = 20 mA for DCmA
g (output value)
h (output for 24 V loop measurement) = ON or OFF
i (backlight) = ON or OFF
j (charging) = OFF (always OFF)
DCV, OHM, FREQ, ACV, DCmV,
ACmV,
DIODE, CONTINUITY,
DCmA, LOOP POWER
DCV/FREQ
Î
66
RA
Sets or queries the current sweep output Normal Cal.
Command = RAm<CRLF> -> Answer = RAm<CRLF>
Command = RA?<CRLF> -> Return = RAm<CRLF>
Parameters
m =
0: Slow Linear (default)
1: Fast Linear
2: Slow Step
3: Fast Step
When the CA450 is in a measurement mode or constant
current output mode, ERR13 is returned.
RG
RP
Sets or queries the auto range and range hold settings Normal Cal.
Command = RGm<CRLF> -> Answer = RGm<CRLF>
Command = RG?<CRLF> -> Return = RGm<CRLF>
Parameters
0: Auto range (default)
m =
1: Range hold
When the CA450 is in an output mode or DCmV, ACmV, diode
test, or continuity check mode, ERR13 is returned.
Sets or queries the RELΔ/% display Normal Cal.
Command = RPm<CRLF> -> Answer = RPm<CRLF>
Command = RP?<CRLF> -> Return = RPm<CRLF>
Parameters
m = 0: Normal measurement (default)
1: RELΔ display
2: REL% display
If the CA450 is in an output mode, ERR13 is returned.
Î
Î
67
SD
Sets or queries the output value. Normal Cal.
Command = SDm<CRLF> -> Answer = SDm<CRLF>
Command = SD?<CRLF> -> Return = SDm<CRLF>
Parameters
m (output value) = 0.000 to 25.000
If the CA450 is in a measurement mode, ERR13 is returned.
Î
SF
SI
SL
Queries the output function Normal Cal.
Command = SF?<CRLF> -> Return = SFm<CRLF>
Parameters
m =
14: Constant current output
15: Current sweep output
If the CA450 is in a measurement mode, ERR13 is returned.
Sets or queries the input voltage value for SENSOR mode Normal Cal.
Command = SIm<CRLF> -> Answer = SIm<CRLF>
Command = SI?<CRLF> -> Return = SIm<CRLF>
Parameters
m (input voltage value) = 0000 to 9999 (4 digits)
If the CA450 is in an output mode, ERR13 is returned.
Because the decimal place of the input voltage value is fixed,
only the number is entered.
Sets or queries the decimal place for SENSOR mode Normal Cal.
Command = SLm<CRLF> -> Answer = SLm<CRLF>
Command = SL?<CRLF> -> Return = SLm<CRLF>
Parameters
m (decimal place) = 0: X.XXX, 1: XX.XX, 2: XXX.X, 3: XXXX
Î
Î
Î
If the CA450 is in an output mode, ERR13 is returned.
68
SM
Sets or queries SENSOR mode Normal Cal.
Command = SMm<CRLF> -> Answer = SMm<CRLF>
Command = SM?<CRLF> -> Return = SMm<CRLF>
Parameters
m =
0: SENSOR mode is disabled (default)
1: SENSOR mode is enabled
If the CA450 is in a mode other than DCmV or DCmA mode,
ERR13 is returned.
Î
SP
SR
Sets or queries SPAN check mode Normal Cal.
Command = SPm<CRLF> -> Answer = SPm<CRLF>
Command = SP?<CRLF> -> Return = SPm<CRLF>
Parameters
0: SPAN check mode is disabled (default)
m =
1: SPAN check mode is enabled
When you switch to span check mode, the previously
output values are retained. Change the output values using
commands such as the UQ and DQ commands.
If the CA450 is in any mode other than constant current output
mode, ERR13 is returned.
Sets or queries the output range (span) Normal Cal.
Command = SRm<CRLF> -> Answer = SRm<CRLF>
Command = SR?<CRLF> -> Return = SRm<CRLF>
Parameters
m = 0: 4 to 20 mA span (factory default setting)
1: 0 to 20 mA span
The changed setting is valid until it is changed again.
Î
69
SS
Sets or queries the slow step time Normal Cal.
Command = SSm<CRLF> -> Answer = SSm<CRLF>
Command = SS?<CRLF> -> Return = SSm<CRLF>
Parameters
m = 0: 15 seconds (default)
1: 30 seconds
2: 45 seconds
3: 60 seconds
If you select slow step mode for current sweep output, this
command sets or queries the hold time of each step.
If the CA450 is in an output mode, ERR13 is returned.
Î
ST
SU
Sets or queries the converted value for SENSOR mode Normal Cal.
Command = STm<CRLF> -> Answer = STm<CRLF>
Command = ST?<CRLF> -> Return = STm<CRLF>
Parameters
m (converted value) = 0000 to 9999 (4 digits)
If the CA450 is in an output mode, ERR13 is returned.
Sets or queries the unit for SENSOR mode Normal Cal.
Command = SUm<CRLF> -> Answer = SUm<CRLF>
Command = SU?<CRLF> -> Return = SUm<CRLF>
Parameters
m (unit) = 0: A, 1: mA, 2: uA, 3: °C, 5: MΩ, 6: kΩ,
7: Ω, 8: kHz, 9: Hz, 10: uF, 11: nF, 12: %
13: lx, 14: none, 15: V, 16: mV
If the CA450 is in an output mode, ERR13 is returned.
70
Î
Î
SY
Sets or queries whether the mode is normal mode or calibration
mode
Command = SYm<CRLF> -> Answer = SYm<CRLF>
Command = SY?<CRLF> -> Return = SYm<CRLF>
Parameters
m =
0: Normal mode (default)
1: Calibration mode
Normal Cal.
UP
UQ
th
Increases the m
digit of the output value by 1 Normal Cal.
Command = UPm<CRLF> -> Answer = UP,OK<CRLF>
Parameters
m = 1 to 5 (1 is the least significant digit while 5 is the most
significant digit)
If the CA450 is in any mode other than constant current output
mode, ERR13 is returned.
If the CA450 is in span check mode, ERR13 is returned.
Increases the output value by one step or sets the output value
to 100%
Command = UQ<CRLF> -> Answer = UQ,OK<CRLF>
Increases the output value by one step (25%).
Sets the output value to 100% of the span in span check mode.
If the CA450 is in any mode other than constant current output
mode, ERR13 is returned.
Î
Normal Cal.
Î
71

6.4 List of Errors

When ERR13 appears
in the main display
Number Description ERR00 No errors ERR11 An undened command has been received. ERR12 The parameter specication for the command is incorrect. ERR13 A command has been received that cannot be executed because of the
status of the CA450. ERR16 An error was detected during calibration. ERR20 Error in the power supply for LOOP POWER measurement ERR23 The output current or voltage has exceeded its limit. ERR60 The saved setting data in the internal ash memory is not correct. ERR61 The saved measurement calibration value data in the internal ash
memory is not correct. ERR62 The saved output calibration value data in the internal ash memory is
not correct.
If error number 11, 12, or 13 occurs during communication, the error number appears
in the LCD main display.
72

6.5 Communication Command Validity Table

Measurement Output
Command
AS BL BZ CC CD CL CMF CP CS CSF CR CW DQ DI DW ESC C/RS ESC S H HC HD IM IO MC MF MH MI MP MR
Normal
RELΔ/% MIN/MAX
Î Î Î Î Î Î   Î Î Î  Î Î Î Î  Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î  Î Î Î Î Î Î Î Î Î Î Î Î Î Î  Î Î Î Î      Î Î Î Î Î Î Î  Î Î Î Î Î    Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î  Î Î Î Î Î Î  Î Î Î Î Î Î Î  Î Î Î Î Î Î Î  Î Î Î Î Î Î  Î Î Î Î Î Î Î
Cal. SETUP
Constant
Current Output Normal Cal. Normal Cal.
Current Sweep
Output
73
Measurement Output
Command
OD OE OS RA RG RP SD SF SI SL SM SP SR SS ST SU SY UP UQ
Normal
There are communication commands that may not be executable depending on the
selected mode and the status of the CA450.
RELΔ/%
MIN/MAX
Î Î Î Î    Î Î Î Î Î Î Î Î Î Î   Î Î Î Î Î Î Î Î  Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î  Î Î Î Î Î Î  Î Î Î Î Î Î  Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î  Î   Î Î Î Î Î Î  Î Î Î Î Î Î  Î Î Î Î Î Î  Î Î Î  Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î
Cal. SETUP
Constant
Current Output Normal Cal. Normal Cal.
Current Sweep
Output
74

6.6 Status Byte Format

Status byte format (see the explanation of the ESC S command)
bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0
LOOP
0
(xed)1(xed)
bit7: Fixed at 0 bit6: Fixed at 1 bit5: Becomes 1 if an error occurs during output The data is held until the status byte is read.
bit4: Becomes 1 if an error occurs in the power supply for LOOP POWER measurement
The data is held until the status byte is read. bit3: Becomes 1 if a measured value is outside of its limits (overrange) Becomes 1 if the CA450 switches to a higher range when auto range is enabled The data is held until the status byte is read. bit2: Becomes 1 when a forbidden operation or command is processed, when the CA450
cannot parse a command, or when a parameter is outside of its range. The data is held until the status byte is read. bit1: Becomes 1 when output is on after the output value has changed and the output has
stabilized. The data is held until the status byte is read.
bit0: Becomes 1 when measured data is conrmed during measurement.
The data is held until the status byte is read.
Output
error
POWER
supply
error
Overrange
Syntax
error
Output
adjustment
completion
Measurement
end
75

7. Replacing the Batteries and Fuses

WARNING
CAUTION
Battery slots
Screw position
Back of the case With the case open
Blank cover (USB adapter attachment position)

7.1 Replacing the Batteries

When the battery voltage falls below the operating voltage, appears on the display. Follow the procedure below to replace the batteries.
• To avoid electrical shock, be sure to disconnect the CA450 from the circuit under measurement.
(Remove the test leads and lead cables as well.)
• Turn the CA450 off.
• Do not measure while the case is open. Doing so may lead to electric shock.
• Do not mix different types of batteries or mix old batteries with new batteries.
• Insert the batteries with their polarities in the directions indicated by the holder.
Replacement Procedure
1
) Remove the fastening screw on the back of the CA450’s case.
2) Open the case.
3) Replace each of the four batteries with a new battery.
4) Return the case to its original position, and tighten the fastening screw.
76

7.2 Replacing the Fuses

WARNING
F1: (440 mA/1000 V)
F2: (440 mA/1000 V)
Screw position
Back of the case With the case open
During current range measurement, the fuses may melt (blow out) because of excessive current. If this happens, replace the fuses.
• To avoid electrical shock, be sure to disconnect the CA450 from the circuit under measurement.
(Remove the test leads and lead cables as well.)
• Turn the CA450 off.
• Do not measure while the case is open. Doing so may lead to electric shock.
• To avoid damaging the CA450, be sure to use the specied fuses.
Fuse specications (rating and maker):
(440 mA/1000 V, Littlefuse, FLU .440)
Replacement Procedure
F1 and F2
1) Remove the fastening screw on the back of the CA450’s case.
2) Open the case.
3) Remove the fuses from their holders.
4) Insert the new fuses into the holders. (Check the ratings of the fuses.)
5) Return the case to its original position, and tighten the fastening screw.
77

8. Calibration and After-Sales Service

For a calibration or after-purchase servicing, contact Omega Engineering.
Calibration
T
o use the CA450 properly, we recommend that you calibrate it regularly. (Refer to the information about the user calibration function.) Recommended calibration period: 1 year

9. Disposing This Product

9.1 Disposing This Product

Waste Electrical and Electronic Equipment (WEEE), DIRECTIVE 2002/96/EC
(This directive is only valid in the EU.)
This product complies with the WEEE Directive (2002/96/EC) marking requirement. The following marking indicates that you must not discard this electrical/electronic product
domestic household waste.
in
Product Category
W
ith reference to the equipment types in the WEEE directive Annex 1,
this product is classied as a “Monitoring and Control Instrumentation” product.
disposing products in the EU, contact Omega Engineering.
When Do not dispose in domestic household waste.
78

9.2 How to Replace and Dispose the Batteries

New EU Battery Directive, DIRECTIVE 2006/66/EC
(This directive is only valid in the EU.)
This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This directive is valid only in the EU. Batteries are included in this product. When you remove batteries from this product and dispose them, discard them in accordance Take a right action on waste batteries, because the collection system in the EU on waste batteries are regulated.
Battery type: Alkaline dry cell
Notice:
This is marking means they shall be sorted out and collected as ordained in ANNEX
How to remove batteries safely:
For
with domestic law concerning disposal.
II in DIRECTIVE 2006/66/EC.
further details, see “7.1 Replacing the Batteries”.
79

10. Troubleshooting

What to Check When There Is a Problem
If the CA450 does not work properly even after you check the items below or you have a problem that is not listed below, contact Omega Engineering.
Symptom Items to Check
Even when the power is on,
nothing appears on the LCD. Current output is not working. • Have the fuses melted (blow out)?
The output or measured values
are strange.
Control through the USB interface
is not working.
“Err60” appears when the power
is turned on.
“Err61” or “Err62” appears when
the power is turned on.
• Are the batteries inserted properly?
• Have the batteries worn out?
SOURCE mode:
When the current is output, is the load
resistance too high?
(Open output terminals or leads)
SIMULATE mode:
• Is the input voltage too low?
(Check the external DCV power supply.)
Is it possible that the values are being affected by
noise?
• Are the communication settings correct?
• The settings that were specified with the setup
function are incorrect.
Refer to section 4.6, “Setup Function,” to return
the CA450 to its factory default settings.
• The calibration data is incorrect.
Refer to section 4.7, “Additional Power-on
Functions,” to return the calibration data to its
factory defaults.
If after you perform the above operations, errors
still appear every time you turn on the power,
repairs are necessary.
80

Appendix 1 Fahrenheit

The default temperature read-out of the Process Multimeter is in Celsius (°C).
To change it to Fahrenheit (°F), it is necessary to proceed as follows:
Changing the temperature unit setting to Fahrenheit
SEE ALSO
For more information on operation, see this manual.
4.3.3 Measuring Using Sensors (SENSOR)
4.6.2 SENSOR Mode Input, Display, and Unit Settings
Displaying “ ºC ” only is congured at factory before shipment.
Carry
out the following setting procedure to display “ ºF ”. While pressing the SELECT, RANGE and HOLD keys simultaneously, turn the function switch to any position.
In the unit area of the main display in SENSOR mode of the setup
function, °F will appear after °C. You will be able to select °F.
6.3 Detailed Command Descriptions
“4: °F” will be added as another m (unit) parameter value for the
command (which sets or queries the unit for SENSOR mode).
SU
81
Loading...