OMEGA BB704 Manual

Page 1
User’s Guide
Shop online at
omega.com
e-mail: info@omega.com
For latest product manuals:
www.omegamanual.info
BB704
Blackbody Calibrator
Page 2
omega.com info@omega.com
Servicing North America:
U.S.A. Omega Engineering, Inc. Headquarters: Toll-Free: 1-800-826-6342 (USA & Canada only)
Customer Service: 1-800-622-2378 (USA & Canada only) Engineering Service: 1-800-872-9436 (USA & Canada only) Tel: (203) 359-1660 Fax: (203) 359-7700 e-mail: info@omega.com
For Other Locations Visit omega.com/worldwide
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
Page 3
Table of Contents
Section ...................................................................................... Page
Table of
BB704 Blackbody Calibrator
Contents
Section 1 Introduction ................................................................................ 1-1
1.1 Precautions ............................................................................. 1-1
1.2 Safety Warnings and IEC Symbols ..................................... 1-1
1.3 General Description .............................................................. 1-2
Section 2 Installation .................................................................................. 2-1
2.1 Unpacking and Inspection ................................................... 2-1
2.2 Mounting ................................................................................ 2-1
2.3 Ambient Temperature .......................................................... 2-1
2.4 Power Connection ................................................................. 2-1
Section 3 Operation ..................................................................................... 3-1
3.1 Front Panel Controls and Indicators .................................. 3-1
3.2 Back Panel Connections ....................................................... 3-2
3.3 Changing the Temperature Setpoint .................................. 3-3
3.4 Changing the Controller Parameter Settings .................... 3-3
3.5 Heat-Up/Cool-Down Cycle Times .................................... 3-5
3.6 Thermal Reset Switch ........................................................... 3-5
Section 4 Serial Communications (RS-232) ............................................ 4-1
Section 5 Maintenance ................................................................................ 5-1
5.1 Calibration ............................................................................. 5-1
5.2 Cleaning ................................................................................. 5-1
5.2.1 Main Body ..................................................................... 5-1
5.2.2 Target Plate ................................................................... 5-1
5.2.3 Fan .................................................................................. 5-1
5.3 Fuse Replacement ................................................................. 5-1
Section 6 Reference Probe ......................................................................... 6-1
Section 7 Specifications .............................................................................. 7-1
Section 8 Troubleshooting Guide ............................................................ 8-1
Section 9 Glossary of Terms Used in This Manual .............................. 9-1
Section 10 The OMEGA Family of Blackbody Calibrators ................ 10-1
i
Page 4
Table of
Figures
BB704 Blackbody Calibrator
Table of Figures
Figure ........................................................................................ Page
1. IEC Symbols ............................................................................................. 1-2
2. Front Panel ................................................................................................ 3-1
3. Back Panel ................................................................................................. 3-2
4. Menu Hierarchy Showing Factory Default Settings ........................... 3-4
5. Programming Procedure ........................................................................ 3-4
6. Heat-Up/Cool-Down Cycle Time Table .............................................. 3-5
7. Connecting the BB704 to a Computer’s Serial Port ............................ 4-1
8. Internal Reference Probe Connections .................................................. 6-1
ii
Page 5
Section 1 - Introduction
Your BB704 Series Blackbody Calibration Source has been designed for ease of use and reliability whenever you have the need to test or calibrate non-contact infrared temperature instruments. It is important that you read this manual completely and follow all safety precautions before operating this instrument.
1.1 Precautions
• Follow all safety precautions and operating instructions outlined in this manual.
• Never leave your calibrator unattended when in use.
• Keep out of reach of children.
• Never touch the target plate when hot.
• Never place any object within 3 inches of the target plate when hot.
• Do not operate in flammable or explosive environments.
• Never operate with a power cord other than the one provided with your unit.
Introduction
1
• Remove and/or disconnect main power cord before attempting any maintenance or fuse replacement.
• Do not connect and/or operate this unit to a non-grounded, non-polarized outlet or power source.
• Do not connect the serial port or reference probe port to equipment with exposed, hazardous, live voltages.
• Protect from moisture and rain.
• The operator of this instrument is advised that if the equipment is used in a manner not specified in this manual, the protection provided by the equipment may be impaired.
There are no user serviceable parts inside your unit. Attempting to repair or service your unit may void your warranty.
1.2 Safety Warnings and IEC Symbols
This device is marked with international safety and hazard symbols in ac­cordance with IEC 1010. It is important to read and follow all precautions and instructions in this manual before operating or commissioning this device as it contains important information relating to safety and EMC. Failure to follow all safety precautions may result in injury and/or damage to your calibrator. Use of this device in a manner not specified by the manufacturer may impair protection provided within the unit.
NOTE:
1-1
Page 6
1
Introduction
IEC symbols Description
Caution, risk of electric shock
Caution, refer to accompanying docu­ments
Caution, hot surface
1.3 General Description
The Model BB704 is a portable, rugged, bench-top, hot/cold blackbody calibra­tion source with a built-in precision PID digital controller. The calibrator is used to test and calibrate infrared pyrometers. The large 4 inch diameter target plate has an emissivity of 0.95 and can be set to any temperature between 100 to 398°C (212 to 750°F).
230 VAC~ @ 50Hz (European Models) 115 VAC~ @ 60Hz (Domestic Models)
Figure 1. IEC Symbols
1-2
Page 7
Section 2 - Installation
2.1 Unpacking and Inspection
Remove the packing list and verify that you have received all your equipment. If you have any questions about the shipment, please call our Customer Service Department at: 1-800-622-2378 or 203-359-1660.
We can also be reached on the Internet at:
omega.com e-mail: info@omega.com
When you receive the shipment, inspect the container and equipment for any signs of damage. Note any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
The carrier will not honor any damage claims unless all shipping material is saved for inspection. After examining and removing contents, save packing material and carton in the event reshipment is necessary.
NOTE:
Installation
2
The following items are supplied in the box:
• BB704 Blackbody Calibration Source
• User’s Manual
• Calibration Certificate
• Power Cord
• Reference Port Cord Plug Connector
2.2 Mounting
Mount the unit on a bench, table-top, or shelf in a horizontal position and operate at least 10 inches from any air obstructions to the fan, front panel, rear panel, bottom, and top of the unit, in an ambient environment between the specified 0 to 50°C (32 to 122°F).
2.3 Ambient Temperature
The target plate of the BB704 can achieve any temperature within the specified temperature range when being operated in normal ambient temperature environments. The maximum specified target plate temperature of 398°C (750°F) can be achieved over the entire specified ambient temperature range.
2.4 Power Connection
Standard (115 VAC~, 50/60 Hz models)
The BB704 comes with a standard North American 3-prong AC power cord. Do not use any other power cord other than the one provided. This cord provides the proper grounding and has been safety tested by the proper safety agencies.
2-1
Page 8
2
Installation
International (230 VAC~, 50/60 Hz models)
On 230 VAC~, 50/60Hz models, a European style power cord with the proper color code and approvals is provided with stripped wire ends for connection to the proper connector used in your country or local area. This connector is not provided.
CAUTION:
• Line voltage variations are not to exceed ±10% of the rated input voltage.
• Electrical connections and wiring should be performed only by suitably trained personnel.
• To shut down the unit in an emergency, the power cord can be disconnected from the AC
outlet or rear of the unit.
2-2
Page 9
Section 3 - Operation
SETPOINT TEMPERATURE
3.1 Front Panel Controls and Indicators
PROCESS TEMPERATURE TARGET PLATE
750.0 PV
750.0 SV
Temperatur
e
U
A
C
C
A
100C/212F
Operation
TM
BLACK POINT Blackbody Calibrator
R
U
F
S
A
C
T
E
E
G
R
A
T
:
N
O
I
T
U
T
I
O
N
:
T
A
R
G
M
A
Y
A
M
E
E
T
C
S
A
U
F
R
BB704
Y
B
E
H
O
T
T
O
H
E
B
399C/750F
3
ON/STANDBY POWER SWITCH
PA RAMETER/
ACCESS KEY
MODE KEY TEMPERATURE RANGE
LOWER SETPOINT KEY
REFERENCE BAR
RAISE SETPOINT KEY
Figure 2. Front Panel
Process Temperature:
This field displays the current temperature of the target plate.
Setpoint Temperature:
This field displays the desired target plate temperature. Once the target plate reaches this desired temperature, both displays will read the same value.
Target Plate:
The 4" target plate is a near ideal blackbody source. The emissivity of the plate is 0.95. When calibrating an IR pyrometer, hold the pyrometer perpendicular to the target plate for optimal performance. The proper distance between the IR pyrometer and the target plate depends on the field of view of the pyrometer. If the pyrometer is too far away, it will scan unwanted surfaces outside of the perimeter of the target plate. Holding the pyrometer too close could introduce undesirable heat to the IR detector of the pyrometer.
CAUTION:
• The BB704’s target plate can be set to very high temperatures. Exercise extreme cau­tion when operating the unit. Keep hands and fingers away from the target plate area. Keep flammable products such as paper, plastics, and clothing far from the BB704.
• The BB704 is a Class II instrument. It is intended to be operated in a laboratory environment only. Never operate the unit outdoors or around children.
• Nothing should come in contact with the target plate, even when the unit is off.
• Never unplug the unit while it is on or during “cool down.”
3-1
Page 10
3
Operation
NOTE:
PID Control: Proportional, integral, derivative control (PID) is a temperature control algorithm used in high-end temperature controllers. The controller causes the process to attain the desired temperature by turn­ing the process on or off. The process may be a heater or refrigerator. As the process temperature approaches the setpoint temperature, the hot or cold process will be pulsed to reduce the correc­tive measures and minimize overshooting. The controller provides a visual representation of the process status through LED indicators. An indicator may light continuously, blink, or shut off entirely to indicate that the process is on, being pulsed, or off, respectively.
Parameter/Access Key:
Press to scroll through menu parameters.
Raise Key:
Press to increase the selected parameter or scroll upward in the list of possible settings.
Lower Key:
Press to decrease the selected parameter or scroll downward in the list of pos­sible settings.
Mode Key:
Press to save settings and exit a menu level.
3.2 Back Panel Connections
RTD REFERENCE PROBE CONNECTION
3-2
FAN
RS-232 COMMUNICATIONS PORT
Figure 3. Back Panel
FUSE
AC MAINS POWER INPUT AND FUSE COMPARTMENT
Page 11
AC Power Mains Input and Fuse Compartment:
The customer connects the power cord to the AC Power Input. As a safety pre­caution, the power cord cannot be connected if the fuse compartment is open. Refer to Section 5.3 for information on fuse replacement.
Reference Probe Connection:
The reference probe enables the user to monitor the target plate temperature with an external instrument. The wires are connected to a Class A, 0.00385 RTD Sensor.
Refer to Section 6 for pinout details.
RS-232 Communications Port:
The female DB-9 port allows the customer to make a 3-wire RS-232 interface with the BB704. A detailed description of this port is in Section 4.
3.3 Changing the Temperature Setpoint
The layout of the front panel is shown in Figure 2. The BB704 incorporates a PID digital setpoint controller. The upper display indicates the blackbody target plate temperature known as (PV) Process Variable, while the lower display indicates the programmed setpoint known as (SV) Setpoint Variable. Changes to the setpoint, units of measure and communication settings are made via the
and keys. Holding a key in continuously will cause the setpoint tempera­ture to advance more quickly to a desired value. Three scanning speeds are provided: slow, medium, and fast. The minimum and maximum setpoints are locked.
Operation
3
3.4 Changing the Controller Parameter Settings
The BB704 operates at optimum performance when left with its factory parameter settings. The only internal parameters that the operator should need to change are the engineering units (°F or °C) or the serial communications parameters. See next page for menu hierarchy with factory default settings (Fig. 4) and programming procedure (Fig. 5).
3-3
Page 12
3
Operation
Menu 00 Menu 01 Menu 02 Menu 03 Menu 04 Menu 05
Key Lock SETPOINT Ac.Cd = 02 Ac.Cd = 03 Ac.Cd = 04 Ac.Cd = 05
Ac.Cd
Gn.o1 64 ALr1 775 id.no 01 SnSr P
Gr.o2 ALr2 775 BAUd 12.o.7 Sn.00
rAtE 29 Cy.t1 05 CAL.L dEC.P
rSEt 146 Cy.t2 CAL.H FILt
H.Hys SP.tt OFF OUt.1 Ht.P
HyS.1 L.SP. L 212 OUt.2 Alr
C.HyS L.SCL CoL.t nor
HyS.2 U.SP.L 750 A1.HL HI
C.SPr H.SCL A1.Pd Pr
SPr.2 A1.OP LAt
dPnG nl A2.HL Hl
A2.Pd Pr
A2.OP LAt
Unit F
Figure 4. Menu Hierarchy Showing Factory Default Settings
NOTE:
Only the boldface parameters are active for the default mode of operation.
Changing the Controller’s Parameter Settings:
1. Press the
key to enter the programming mode. The lower display will
alternately display the menu level and “Ac.Cd.”
2. Use the
3. Once you have chosen the desired menu use the
and keys to change to the desired menu level.
key to scroll through the parameters. To change the setting of a given parameter, use the and keys.
4. To save settings press the
key. The controller now exits the programming
menu and returns to the normal operating mode.
5. To change settings on other menu levels, you must re-enter the programming menu (from step #1).
Putting the Controller In or Out of Standby Mode:
1. To enter the Standby mode hold the
key for 4 seconds until the window
flashes “StbY.”
2. To exit the Standby mode hold the key for 4 seconds until the window flashes “tUne.” Then press the key for 4 seconds again until “tUne” stops flashing.
3-4
Figure 5. Programming Procedure
Page 13
In the event that the unit has been operated outside of operating range, the controller alarm will disable the heater and the target plate temperature will not rise. Let the unit cool down completely. Turn the power off and then back on. If the problem is not corrected, contact the factory.
3.5 Heat-Up/Cool-Down Cycle Times
Approximate cycle times for heat up and cool down are given in the table above. To find a given transition time from an initial temperature to a second target temperature, follow this procedure: Look for an initial temperature in the left column. Next, look for the target temperature along the top row. The intersec­tion of the row and column provides the approximate transition time.
NOTE:
Operation
3
121°C 260°C 398°C
FROM:
121°C 15 MIN. 25 MIN. 250°F
260°C 25 MIN. 18 MIN. 500°F
398°C 32 MIN. 10 MIN. 750°F
250°F 500°F 750°F
3.6 Thermal Reset Switch
In the event that the target plate temperature and/or ambient temperature exceeds the acceptable limit, the thermal reset switch will pop open. The con­troller will appear to be demanding a temperature increase but the target plate temperature will not rise. Let the unit cool down completely. Press the switch located on the right side of the unit near the handle hinge until a click is heard. If this has no effect, refer to the troubleshooting section.
If the reset switch keeps resetting, call Omega’s Customer Service Department at (203) 359-1660.
TO:
FROM:
Figure 6. Heat-Up/Cool-Down Cycle Time Table
TO:
3-5
Page 14
3
Operation
NOTES:
3-6
Page 15
Serial Communications (RS-232)
Section 4 - Serial Communications (RS-232)
This method allows bi-directional data transfer via a three-conductor cable con­sisting of signal ground, receive input, and transmit output. It is recommended
that less than 50 feet of shielded cable be used between the computer and this instrument. This will assure performance of the BB704 to EN61326 standards, under the EMC. Note that multiple instruments cannot be tied to the same port
in this configuration. The RS-232 port is optically isolated to eliminate ground loop problems.
Below is a pinout diagram for the serial port of the BB704 as well as the pinout for a 9-pin PC serial port. Use a straight DB9 (female) to DB9 (male) connector cable to connect your computer to the BB704. The cable should be attached only when the computer and BB704 are off.
Only parameters in the parameter list should be modified or queried via the serial port. Other parameters should be viewed or queried from the controller directly. It is highly recommended that baud rate for the controller be modified on the controller directly. Note that both the BB704 and the computer must be communicating with the same serial communications parameters to establish a working communication link.
4
The serial communications feature can be tested using a terminal emulation package. Note that this controller does not time-out waiting for the next char­acter to be transmitted. Be sure not to use the XON/XOFF or hardware hand­shaking. Lastly, it should be noted that following a complete transmission to the BB704, a response is sent back. If the message was valid, the changed or queried parameter is echoed back (following the same format). If the message was not according to acceptable format or was attempting to force a parameter out of range, an “ERROR” message is echoed.
Figure 7. Connecting the BB704 to a Computer’s Serial Port
BB704
BLACKBODY
CALIBRATOR
COMPUTER
SERIAL
PORT
(RS-232)
4-1
Page 16
4
Serial Communication (RS-232)
Parameter List (only relevant parameters shown):
PAR#: Parameter: Range/Units:
00 Process Temp. Input determined
01 Setpoint Input determined
19 Baud Selection bAUd
Baud Selections:
Code: Baud: Parity: Data Bits: Stop Bits:
3.o.7 300 odd 7 2
6.o.7 600 odd 7 2
12.o.7 1200 odd 7 2
(factory default settings)
24.o.7 1200 odd 7 2
3.n.8 300 no 8 1
6.n.8 600 no 8 1
12.n.8 1200 no 8 1
24.n.8 2400 no 8 1
General Message Format:
#[controller id][command][parameter number]<new value><units>[CR/LF]
Definitions:
# – This character initiates an “escape sequence” that the controller will recog­nize.
[controller id] – Up to 2 numeric characters, 00 to 99 (factory default = 01)
[command] – 1 character, upper case or lower case
R – To read a parameter from the controller
M – To temporarily modify a controller parameter (lost upon shutdown)
E – To modify a controller parameter in non-volatile memory (saved even
after shutdown)
[parameter #] – Up to 2 numeric characters, 00 to 99
<new value> – This control word is used only when entering or modifying a
parameter. Up to 6 characters may be entered. The first character can be a space, a “+”, or a “-”. The next 4 characters are for enter­ing the new value parameter value. Be sure to use the exact same field format as is currently being used (i.e., if the XXX.X format is used to express temperature, be sure to enter a new value that conforms to the same format).
4-2
<units> – This optional control word is used to specify units, F for °F, C for °C.
[CR/LF] – Every transmission must be terminated with a carriage return [CR] character. The line feed [LF] character is optional.
Page 17
Section 5 - Maintenance
5.1 Calibration
This unit has been fine-tuned at the factory and calibrated to give optimum performance at its full temperature range. It is recommended that the unit be returned annually for re-calibration.
5.2 Cleaning
Remove all electrical connections and power before attempting any maintenance or cleaning.
5.2.1 Main Body
Only a damp, soft rag with a mild cleaning solution should be used when clean­ing the main body of this unit.
CAUTION:
Maintenance
5
5.2.2 Target Plate
Do not attempt to clean the target plate. The target plate has a special coating ap­plied and cleaning may change the emissivity and performance of your unit.
5.2.3 Fan
The fan filter should be cleaned at least monthly by washing the filter with warm water and then blowing dry with air. It can be removed by firmly pulling the black plastic frame outward. The internal protective grill that is seated against the fan can be cleaned with a soft bristle brush.
5.3 Fuse Replacement
Disconnect all power from source before attempting fuse replacement.
For continued protection against the risk of fire, replace with only the same size, type, and fuse rat­ing indicated here and on the rear panel of your unit.
WARNING:
CAUTION:
For model BB704 use: 1 ea. 250 VAC~, T5A (Time-Lag, 5 Amp)
UL/CSA APPROVED (0.25” dia. x 1.25” long).
For model BB704-230VAC use: 2 ea. 250 VAC~, F3.5A (Fast-Acting, 3.15 Amp)
VDE APPROVED (5 mm dia. x 20 mm long).
5-1
Page 18
6
Reference Probe
Section 6 - Reference Probe
An RTD probe has been embedded into the target plate heater assembly to be used as a reference to check the calibrator’s calibration and accuracy. A connec­tor has been provided with your calibrator for connection to this reference probe. Listed below are a few calibration points and the resistance readings you can expect from the reference probe.
Target Plate Temperature Reference Probe Resistance (Ohms)
100°C (212°F) 138.51
149°C (300°F) 156.96
204°C (400°F) 177.33
260°C (500°F) 197.71
316°C (600°F) 217.73
371°C (700°F) 237.05
399°C (750°F) 246.71
TARGET PLATE ASSEMBLY
BLACK WIRE
RED WIRE
BLACK WIRE
RTD REFERENCE PROBE
Figure 8. Internal Reference Probe Connections
1
2
3
REFERENCE PROBE CONNECTOR
6-1
Page 19
Section 7 - Specifications
Target Plate Temperature Range: 100 to 398°C (212 to 750°F)
Overall Accuracy: ±0.8°C (±1.4°F)
Display Accuracy: ±0.5°C (±1.0°F)
Stability: ±0.1°C or less (±0.2°F or less)
Ambient Environmental Conditions
Temperature: 0 to 50°C (32 to 122°F ) Humidity: 0 to 90% RH, Non-condensing
Target Plate Size: 101.6 mm (4")
Cavity Emissivity: 0.95*
Internal Control Sensor: RTD, Class A, Alpha = 0.00385
Reference Sensor: RTD, Class A, Alpha = 0.00385 Accuracy: ±1.8°F (±1.0°C)
Warm-up Time: See Table, Page 3-5
Cool-down Time: See Table, Page 3-5
Specifications
7
Power BB704: 115 VAC~, 50/60 Hz, 425W BB704-230 VAC: 230 VAC~, 50/60 Hz, 425W
Dimensions: 190 mm H x 410 mm W x 264 mm L (7.5 H x 16.1 W x 10.4" L)
Weight: 7.2kg (16 lbs)
Installation Category II
* Reference to 8-14 microns wavelength bandwidth.
7-1
Page 20
8
Troubleshooting Guide
Section 8 - Troubleshooting Guide
Problem Solution
1. Unit will not turn on. a. Check all electrical connections.
b. Check rear panel fuses.
c. If previous steps don’t solve the problem,
2. Unit turns on but the target a. Check that you have entered a setpoint
plate will not get hot. above the ambient temperature.
b. Verify that the controller is set to its factory default settings.
c. Unit may have been operated outside of
contact our customer service department.
operating range. The controller alarm may have disabled the heater. Let the unit cool down completely. Turn the power off and back on. If the problem is not corrected, contact the factory.
d. Unit may have been operated outside
of operating range. Safety cutoff switch may have disabled heater. Let the unit cool down completely. Press the reset switch located on the left side of the unit, behind the carry handle, until a click is heard. If the problem is not cor­rected, contact the factory.
e. If previous steps don’t solve the problem,
contact our customer service department.
3. Controller display shows “Error” a. Unit requires service; contact our
and the target plate will not get customer service department. hot or cold.
4. Target plate temperature will not a. Verify that the controller is set to its
stabilize to within ±0.5°F of the factory default settings. setpoint temperature.
5. Controller has no output. a. Controller may be in STANDBY mode.
Pressing buttons has no effect. Hold down the TUNE blinks on the display. until “TUNE" stops blinking.
b. If previous steps don’t solve the problem,
contact our customer service department.
key continuously
8-1
6. Controller has no output. a. Controller may be in LOCK mode.
Controller does not respond to Hold down the any key presses. until “Ac.Cd” is displayed. Change to a
menu level other than level 1 and exit the menu selection mode by pressing the
key once.
key continuously
Page 21
Glossary of Terms Used in This Manual
Section 9 - Glossary of Terms Used in This Manual
Blackbody
A theoretical object that radiates the maximum amount of energy at a given tem­perature, and absorbs all the energy incident upon it.
Calibration
The process of adjusting an instrument or compiling a deviation chart so that its reading can be correlated to the actual value being measured.
Emissivity
The ratio of energy emitted by a surface to the energy emitted by a blackbody at the same temperature.
IEC
9
International Electrotechnical Commission.
Infrared (IR)
A range of the electromagnetic spectrum extending beyond red visible light from 760 nanometers to 1000 microns.
PID
Proportional, Integral, Derivative. A three-mode control action where the con­troller has time proportioning, integral (auto reset), and derivative rate action.
RTD
Resistance temperature detector.
9-1
Page 22
10
The OMEGA Family of Blackbody Calibrators
The OMEGA Family of Blackbody Calibrators
Listed below is a selection guide to OMEGA’s current line of blackbody calibration sourc­es including the one you have selected. This family of rugged, portable, and accurate calibrators covers a wide range of temperatures, target plate sizes, and features, making them perfect for infrared pyrometer field service testing and laboratory calibrations.
BB701 Hot/Cold Blackbody Calibration Source Calibration Range: -18 to 149°C (0 to 300°F) Accuracy: ±0.8°C (±1.4 °F) Emissivity: 0.95 Ambient Temp.: 4 to 43°C (40 to 110°F) Cavity Size: 63.5 mm (2.5 in.) Power: 115/230V, 50/60 Hz, 175W
BB702 Blackbody Calibration Source Calibration Range: 32 to 215°C (amb. 90 to 420°F) Accuracy: ± 0.5°C (±0.9°F), ±0.25% rdg Emissivity: 0.95 Ambient Temp.: 5 to 45°C (41 to 113°F) Cavity Size: 63.5 mm (2.5 in.) Power: 115/230V, 50/60 Hz, 175W
BB703 Mini Blackbody Calibration Source Calibration Range: 32 to 400°C (90 to 752°F) Accuracy: ±1.4°C (±2.5°F) Emissivity: 0.95 Ambient Temp.: 0 to 40°C (32 to 104°F) Cavity Size: 28.6 mm (1.125 in.) Power: 115/230V, 50/60 Hz, 175W
BB704 4” Target Plate Blackbody Calibration Source Calibration Range: 100 to 398°C (212 to 750°F) Accuracy: ±0.8°C (±1.4°F) Emissivity: 0.95 Ambient Temp.: 0 to 50°C (32 to 122°F) Cavity Size: 101.6 mm (4 in.) Power: 115/230V, 50/60 Hz, 425W
BB705 Laboratory Grade Blackbody Calibration Source Calibration Range: 100 to 1046°C (212 to 1915°F) Accuracy: ±1.0°C (±1.8°F), ±0.25% rdg Emissivity: 0.99 Ambient Temp.: 0 to 35°C (32 to 95°F) Cavity Size: 44 mm (1.75 in.) Power: 115/230V, 50/60 Hz, 1100W
BB-4A High Temperature Blackbody Calibration Source Calibration Range: 100 to 982°C (212 to 1800°F) Accuracy: ±1.0°C (±1.8°F), ±0.25% rdg Emissivity: 0.99 Ambient Temp.: 0 to 50°C (32 to 122°F) Cavity Size: 22.2 mm (0.88 in.) Power: 115/230V, 50/60 Hz, 400W
For a complete, updated specification sheet and price on any of the calibrators listed here, visit our web site at omega.com. Please call our sales or customer service department for information and pricing on any new models available.
9-2
Page 23
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 25 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal two (2) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.
OMEGA is a trademark of OMEGA ENGINEERING, INC. © Copyright 2017 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.
FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.
Page 24
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
TEMPERATURE
MU
Thermocouple, RTD & Thermistor Probes, Connectors,
Panels & Assemblies
MU
Wire: Thermocouple, RTD & Thermistor
MU
Calibrators & Ice Point References
MU
Recorders, Controllers & Process Monitors
MU
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
MU
Transducers & Strain Gages
MU
Load Cells & Pressure Gages
MU
Displacement Transducers
MU
Instrumentation & Accessories
FLOW/LEVEL
MU
Rotameters, Gas Mass Flowmeters & Flow Computers
MU
Air Velocity Indicators
MU
Turbine/Paddlewheel Systems
MU
Totalizers & Batch Controllers
pH/CONDUCTIVITY
MU
pH Electrodes, Testers & Accessories
MU
Benchtop/Laboratory Meters
MU
Controllers, Calibrators, Simulators & Pumps
MU
Industrial pH & Conductivity Equipment
DATA ACQUISITION
MU
Communications-Based Acquisition Systems
MU
Data Logging Systems
MU
Wireless Sensors, Transmitters, & Receivers
MU
Signal Conditioners
MU
Data Acquisition Software
HEATERS
MU
Heating Cable
MU
Cartridge & Strip Heaters
MU
Immersion & Band Heaters
MU
Flexible Heaters
MU
Laboratory Heaters